K Number
K192539

Validate with FDA (Live)

Device Name
Deka Luxea
Date Cleared
2019-12-16

(91 days)

Product Code
Regulation Number
878.4810
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

VIVID LASER handpiece: Indicated for the treatment of benign vascular lesions, benign pigmented lessions, hair removal and permanent harr reduction. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime. SPARKS LASER handpiece : Indicated for removal of unwanted hair, for stable long term or permanent hair reduction (Skin Types Fitzpatrick I-VI), photocoagulation and hemostasis of benign pigmented and vascular lesions, such as but not limited to warts, teleangiectasia, leg veins and spider veins, treatment of benign pigmented lesions. PRISMA LASER handpiece: Indicated for removal of dark tattoos and treatment of benign pigmented lesions. RUBER, LILAC Pulsed Light handpieces: Permanent hair reduction, photocoagulation of benign vascular lesions, photothermolysis of blood vessels, treatment of benign pigmented lesions. Different wavelength ranges of Pulsed Light attachment are indicated for the various treatments and skin types. as indicated in the following table: 500-1200nm: Benign Vascular lesions Skin Types I, II; Blood vessels Skin Types I, II. 520-1200nm: Benign Vascular lesions SkinType III; Blood Type III; Benign Pigmented lesions Skin Types I, II. 550-1200nm: Hair reduction Skin Types I, II; Benign Pigmented lesions Skin Type Ill. 600-1200nm: Hair reduction Skin Type Ill. 650-1200nm: Hair reduction Skin Type IV; Benign Pigmented lesions Skin Type IV. LAZUR Pulsed Light handpiece: The handpiece (with and without contact-cooling) is indicated for: -The treatment of moderate inflammatory acne vulgaris. -The treatment of benign pigmented epidermal lesions including dyschromia, hyperpigmentation, melasma, and ephelides (freckles). -The treatment of benign cutaneous lesions including warts, scars and striae. -The treatment of benign cutaneous vascular lesions including port wine stains, hemangiomas, facial, truncal and leg telangiectasias, rosacea, erythema of rosacea, angiomas and spider angiomas, poikiloderma of Civatte, leg veins and venous malformations. - Use on all skin types (Fitzpatrick I-VI). VIRIDIS Pulsed Light handpiece: Photocoagulation of benign dermatological vascular lesions, photothermolysis of blood vessels (treatment of facial andleg veins), and the treatment of benign pigmented lesions.

Device Description

The DEKA LUXEA is a versatile medical platform that can be equipped with various types of laser and pulsed light handpieces. Thanks to its universal connector it is possible to quickly insert and deactivate the various handpieces to cover a wide spectrum of application. The available handpieces are - VIVID: 808±10 nm Laser - SPARKS: 1064nm Laser - PRISMA: 1064nm O-Switched laser - VIRIDIS: Pulsed Light handpiece 15mmx13mm with integrated 500-1200 nm filter - LILAC: Pulsed Light handpiece 48mmx13mm with interchangeable filters - RUBER: Pulsed Light handpiece 48mmx17mm with integrated 550mm filter - LAZUR: Pulsed Light handpiece 48mmx13mm with integrated 420-950 nm filter Emission activation is either by footswitch or fingerswitch. Electrical specifications are 115-240V~, 50-60Hz, 2300VA max.

AI/ML Overview

The provided text is a 510(k) Pre-Market Notification from the FDA regarding the DEKA LUXEA device. This document focuses on demonstrating substantial equivalence to previously cleared predicate devices, rather than presenting a study to prove the device meets specific performance acceptance criteria for a new clinical claim.

Therefore, the document does not contain the information requested in your prompt regarding:

  1. A table of acceptance criteria and reported device performance: This document explicitly states "Clinical Performance Data: None." The performance metrics provided (fluence, spot size, pulse duration, repetition rate) are technical specifications for comparison to predicate devices, not performance against clinical acceptance criteria.
  2. Sample size used for the test set and the data provenance: No clinical test set data is presented.
  3. Number of experts used to establish the ground truth... and their qualifications: No ground truth establishment is described as there's no clinical data.
  4. Adjudication method for the test set: Not applicable.
  5. If a multi reader multi case (MRMC) comparative effectiveness study was done: Not applicable.
  6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
  7. The type of ground truth used: Not applicable.
  8. The sample size for the training set: Not applicable.
  9. How the ground truth for the training set was established: Not applicable.

Explanation of the Document's Purpose:

This 510(k) submission primarily relies on the "substantial equivalence" pathway for regulatory clearance. This means the manufacturer is asserting that their new device is as safe and effective as a legally marketed device (predicate device) that was on the market before May 28, 1976, or has been reclassified.

To establish substantial equivalence, manufacturers typically compare their new device to predicate devices in terms of:

  • Indications for Use: The DEKA LUXEA handpieces have indications for use that are very similar to their respective predicate devices.
  • Technological Characteristics: The document provides detailed comparisons of technical specifications like laser wavelength, fluence, spot sizes, pulse duration, and pulse repetition rate for each handpiece against its chosen predicate. The core argument is that these characteristics are either identical or sufficiently similar that the device is as safe and effective as the predicate.
  • Non-Clinical Performance Data: The document explicitly mentions conformance with safety and electromagnetic compatibility (EMC) standards (IEC 60601-1, IEC 60601-1-2, IEC 60601-2-22, IEC 60825-1, IEC 60601-2-57). This demonstrates electrical safety and performance within established engineering norms, not clinical performance for specific outcomes.
  • Clinical Performance Data: The document explicitly states "None." This is common for 510(k) submissions where substantial equivalence can be established through non-clinical means and comparison of technological characteristics.

In summary, the provided FDA document is a regulatory submission demonstrating substantial equivalence, not a clinical study proving device performance against specific acceptance criteria for a new clinical claim.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue. Underneath the FDA logo is the word "ADMINISTRATION".

El..EN. Electronic Engineering SPA Paolo Peruzzi Regulatory Affairs Manager via Baldanzese 17 Calenzano, 50041 It

December 16, 2019

Re: K192539

Trade/Device Name: Deka Luxea Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument for use in General and Plastic Surgery and in Dermatology Regulatory Class: Class II Product Code: GEX, ONF Dated: September 12, 2019 Received: September 16, 2019

Dear Paolo Peruzzi:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Neil R.P. Ogden, MS Assistant Director THT4A4 DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K192539

Device Name DEKA LUXEA

Indications for Use (Describe)

VIVID LASER handpiece:

Indicated for the treatment of benign vascular lesions, benign pigmented lessions, hair removal and permanent harr reduction.

Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime

SPARKS LASER handpiece :

Indicated for removal of unwanted hair, for stable long term or permanent hair reduction (Skin Types Fitzpatrick I-VI), photocoagulation and hemostasis of benign pigmented and vascular lesions, such as but not limited to warts, teleangiectasia, leg veins and spider veins, treatment of benign pigmented lesions.

PRISMA LASER handpiece:

Indicated for removal of dark tattoos and treatment of benign pigmented lesions.

RUBER, LILAC Pulsed Light handpieces:

Permanent hair reduction, photocoagulation of benign vascular lesions, photothermolysis of blood vessels, treatment of benign pigmented lesions.

Different wavelength ranges of Pulsed Light attachment are indicated for the various treatments and skin types. as indicated in the following table:

Pulsed lightWavelength rangeHair reductionBenignVascular lesionsBlood vesselsBenignPigmented lesions
500-1200nm-Skin Types I, IISkin Types I, II-
520-1200nm-SkinType IIISkin Type IIISkin Types I, II
550-1200nmSkin Types I, II--Skin Type Ill
600-1200nmSkin Type Ill---
650-1200nmSkin Type IV--Skin Type IV

LAZUR Pulsed Light handpiece:

The handpiece (with and without contact-cooling) is indicated for:

-The treatment of moderate inflammatory acne vulgaris.

-The treatment of benign pigmented epidermal lesions including dyschromia, hyperpigmentation, melasma, and ephelides (freckles).

-The treatment of benign cutaneous lesions including warts, scars and striae.

-The treatment of benign cutaneous vascular lesions including port wine stains, hemangiomas, facial, truncal and leg telangiectasias, rosacea, erythema of rosacea,

angiomas and spider angiomas, poikiloderma of Civatte, leg veins and venous malformations.

  • Use on all skin types (Fitzpatrick I-VI).

{3}------------------------------------------------

VIRIDIS Pulsed Light handpiece:

Photocoagulation of benign dermatological vascular lesions, photothermolysis of blood vessels (treatment of facial andleg veins), and the treatment of benign pigmented lesions.

Type or Use (Select one or both, as applicable)
---------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.qov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(K) Summary DEKA LUXEA

Submitter:

El.En. S.p.A. Via Baldanzese, 17 50041 Calenzano (FI), Italy

Contact:

Paolo Peruzzi Regulatory Affairs Manager & Official Correspondent Phone: +39.055.8826807 E-mail: p.peruzzi@elen.it

Date Summary Prepared:

December 11, 2019

Device Trade Name:

DEKA LUXEA

Common Name:

Medical Laser and pulsed Light platform

Classification Name:

Laser Surgical Instrument for use in General and Plastic Surgery and in Dermatology

Product Code :

GEX, ONF

Classification Number:

21 CFR 878.4810

Equivalent Devices:

{5}------------------------------------------------

Device Description:

The DEKA LUXEA is a versatile medical platform that can be equipped with various types of laser and pulsed light handpieces.

Thanks to its universal connector it is possible to quickly insert and deactivate the various handpieces to cover a wide spectrum of application.

The available handpieces are

  • VIVID: 808±10 nm Laser
  • SPARKS: 1064nm Laser ●
  • PRISMA: 1064nm O-Switched laser ●
  • VIRIDIS: Pulsed Light handpiece 15mmx13mm with integrated 500-1200 nm filter ●
  • LILAC: Pulsed Light handpiece 48mmx13mm with interchangeable filters
  • RUBER: Pulsed Light handpiece 48mmx17mm with integrated 550mm filter ●
  • LAZUR: Pulsed Light handpiece 48mmx13mm with integrated 420-950 nm filter ●

Emission activation is either by footswitch or fingerswitch. Electrical specifications are 115-240V~, 50-60Hz, 2300VA max.

Indications for Use:

VIVID LASER handpiece:

Indicated for the treatment of benign vascular lesions, benign pigmented lesions, hair removal and permanent hair reduction.

Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.

SPARKS LASER handpiece :

Indicated for removal of unwanted hair, for stable long term or permanent hair reduction (Skin Types Fitzpatrick I-VI), photocoagulation and hemostasis of benign pigmented and vascular lesions, such as but not limited to warts, teleangiectasia, leg veins and spider veins, treatment of benign pigmented lesions.

PRISMA LASER handpiece:

Indicated for removal of dark tattoos and treatment of benign pigmented lesions

RUBER, LILAC Pulsed Light handpieces:

Permanent hair reduction. photocoagulation of benign vascular lesions, photothermolysis of blood vessels, treatment of benign pigmented lesions.

Different wavelength ranges of Pulsed Light attachment are indicated for the various treatments and skin types. as indicated in the following table:

{6}------------------------------------------------

Pulsed lightWavelengthrangeHair reductionBenignVascular lesionsBlood vesselsBenignPigmented lesions
500-1200nm-Skin Types I, IISkin Types I, II-
520-1200nm-Skin Type IIISkin Type IIISkin Types I, II
550-1200nmSkin Types I, II--Skin Type III
600-1200nmSkin Type III---
650-1200nmSkin Type IV--Skin Type IV

LAZUR Pulsed Light handpiece:

The handpiece (with and without contact-cooling) is indicated for:

-The treatment of moderate inflammatory acne vulgaris.

-The treatment of benign pigmented epidermal lesions including dyschromia, hyperpigmentation, melasma, and ephelides (freckles).

-The treatment of benign cutaneous lesions including warts, scars and striae.

-The treatment of benign cutaneous vascular lesions including port wine stains, hemangiomas, facial, truncal and leg telangiectasias, rosacea, erythema of rosacea, angiomas and spider angiomas, poikiloderma of Civatte, leg veins and venous malformations.

  • Use on all skin types (Fitzpatrick I-VI).

VIRIDIS Pulsed Light handpiece:

Photocoagulation of benign dermatological vascular lesions, photothermolysis of blood vessels (treatment of facial and-leg veins), and the treatment of benign pigmented lesions.

{7}------------------------------------------------

Substantial equivalence discussion:

The VIVID Laser Handpiece is substantially equivalent to the Quanta Forte 808nm laser (cleared with K152714):

Device TradeNameDEKA LUXEAVIVID LASER handpiecePredicate DeviceK152714Quanta Forte
Product code andregulationGEX21 CFR 878.4810GEX21 CFR 878.4810
Laser Wavelength808±10 nm808±10 nm
MAX Fluence40 J\cm244 J/cm²
Handpiece SpotSizes10x12 mm12 x 14 mm
Pulse DurationUp to 200msUp to 200ms
Pulse RepetitionRate (Hz)Single shot to 10 HzSingle shot to 10 Hz

The SPARKS Laser Handpiece is substantially equivalent to the DEKA Synchro FT (cleared with K103288):

Device TradeNameDEKA LUXEASPARKS LASER handpiecePredicate DeviceK103288DEKA Synchro FT
Product code andregulationGEX21 CFR 878.4810GEX21 CFR 878.4810
Laser Wavelength1064 nm1064 nm
MAX Fluence80-700 J/cm2 (2.5 mm)30-250 J/cm2 (4 mm)14-170 J/cm2 (6 mm)5-60 J/cm2 (10 mm)70-700J/cm2 (2.5mm)30 243J/cm2 (5mm)25-124J/cm2 (7mm)20 60J/cm2 (10mm)15 36J/cm2 (13mm)10 25J/cm2 (16mm)
Handpiece SpotSizesØ 2.5, 4, 6, 10 mmØ 2.5, 5, 7, 10, 13, 16 mm
Pulse Duration1- 280 ms0.8 – 285 ms
Pulse RepetitionRateSingle shot to 10 Hz0.5 to 10 Hz

{8}------------------------------------------------

The PRISMA Laser Handpiece is substantially equivalent to the Lumenis M22 (cleared with K142860):

Device TradeNameProposed 510(k) DeviceDEKA LUXEAPRISMA LASER handpiecePredicate DeviceK142860Lumenis M22
Product code andregulationGEX21 CFR 878.4810GEX21 CFR 878.4810
Laser Wavelength1064nm1064nm
MAX Fluence14 J/cm² (2.5x2.5mm)9 J/cm² (3x3mm)14 J/cm²
Handpiece SpotSizes2.5x2.5 mm3x3 mmØ 2-8 mm
Pulse Duration6ns6-8 ns
Pulse RepetitionRate1 to 5 Hz5Hz

The RUBER and LILAC Pulsed light Handpieces are substantially equivalent to the DEKA Synchro Replay FT Pulsed light handpiece (cleared with K150516):

Device TradeNameDEKA LUXEARUBER, LILAC Pulsed LighthandpiecesPredicate DeviceK150516DEKA Synchro Replay
Product code andONFONF
regulation21 CFR 878.481021 CFR 878.4810
Pulsed light500 - 1200nm (LILAC)500 - 1200nm
Emission520 - 1200nm (LILAC)520 - 1200nm
spectrum550 - 1200nm (RUBER, LILAC)550 - 1200nm
600 - 1200nm (LILAC)600 - 1200nm
650 - 1200nm (LILAC)650 - 1200nm
Fluence1 - 25 J/cm² (LILAC)1 - 20 J/cm² (RUBER)3 - 25 J/cm²
Spot Sizes48x13 mm (LILAC)48x17 mm (RUBER)48x13 mm
Pulse Duration3 - 124 ms (LILAC)8 - 50 ms (RUBER)3 – 124 ms
Repetition Rate0.5 Hz max.0.5 Hz max.
Method of SkinCoolingIntegrated - provided via the handpiecelightguideIntegrated - provided via the handpiecelightguide

{9}------------------------------------------------

The LAZUR Pulsed light Handpiece is substantially equivalent to the Alma Laser Harmony XLTm Multi-Application Platform (cleared with K072564):

Device TradenameProposed 510(k) DeviceDEKA LUXEALAZUR Pulsed Light handpiecePredicate DeviceK072564ALMA LASER Harmony XLTm Multi-Application Platform
Product code andregulationONF21 CFR 878.4810ONF21 CFR 878.4810
Pulsed lightEmissionspectrum420-950 nm420 -950 nm
Fluence3 - 20 J/cm²5 - 25 J/cm²
Spot Sizes6.2 cm² (48 mm x 13 mm)6.4 cm²
Pulse Duration30, 40, 50 ms30, 40, 50 ms
Repetition Rate0.5 Hz max0.5 Hz
Method of SkinCoolingOptional, integrated, provided via thehandpiece lightguideOptional

The VIRIDIS Pulsed light Handpiece is substantially equivalent to the Palomar LuxG (cleared with K020941):

Device TradeNameProposed 510(k) DeviceDEKA LUXEAVIRIDIS Pulsed light handpiecePredicate DeviceK020941Palomar LuxG
Product code andregulationGEX21 CFR 878.4810GEX21 CFR 878.4810
Pulsed lightEmission spectrum500–677 & 854 1200nm500–670 & 870–1400nm
Fluenceup to 50 J/cm²up to 50 J/cm²
Spot Sizes15x13 mm15x10 mm
Pulse Duration3–116 ms1–100 ms
Repetition Rate1 Hz max.2 Hz max.
Method of SkinCoolingIntegrated - provided via the handpiecelightguideIntegrated - provided via the handpiecelightguide

{10}------------------------------------------------

The DEKA LUXEA laser and pulsed light handpieces have indications for use as the abovementioned predicate devices, with same principle of operation and essentially the same performances.

Clinical Performance Data:

None.

Non-Clinical Performance Data:

The LUXEA was tested for conformance with the following standards:

IEC 60601-1- Medical electrical equipment - Part 1: General requirements for basic safety and essential performance.

IEC 60601-1-2 Medical electrical equipment – Part 1: General requirements for basic safety and essential performance- Collateral standard: Electromagnetic compatibility - Requirements and tests.

IEC 60601-2-22 Medical electrical equipment - Part 2-22: Particular requirements for basic safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment.

IEC 60825-1- Safety of laser products - Part 1: Equipment classification and requirements.

IEC 60601-2-57 - Medical electrical equipment - Part 2-57: Particular requirements for the basic safety and essential performance of non-laser light source equipment intended for therapeutic, diagnostic, monitoring and cosmetic/aesthetic use

Conclusion:

We can conclude that the DEKA LUXEA laser system is substantially equivalent to the predicate devices .

Additional Information:

None.

§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.

(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.