K Number
K192516

Validate with FDA (Live)

Date Cleared
2019-12-10

(88 days)

Product Code
Regulation Number
878.4810
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Diode Laser 808nm is intended for hair removal, permanent hair reduction on all skin type I-VI), including tanned skin.

Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.

Device Description

The Diode Laser 808nm is intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin.

Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.

The proposed device has six main function modules as following:

Water cooling system:

Through refrigeration compressor to achieve heat and cold exchange of air, the internal water tank by the water and water pump for cooling water cycle, with cooling water to cool the parts of the tools, the system to achieve this function is called water cooling system.

Power system:

Through the main circuit control system, according to the control requirements to provide the size of laser energy, the system to achieve this function is called the power system.

User Interface system:

Through the display screen and touch screen, 1. Provide users with relevant information about display equipment; 2. Users operate or set relevant parameters or functions; 3. Display fault information. The system that realizes this function is called human-machine switching system.

Handpiece system:

According to the choice of different treatment methods to choose the appropriate treatment tools, the system to achieve this function is called the handpiece system.

Main circuit power supply system:

The system that realizes the power supply of equipment and the control switch or stops the power supply is called the main circuit power supply system.

Main circuit control system:

It is used to control the parts of the equipment to work in a reasonable sequence and provide fault or warning sound. The system that realizes this function is called the main circuit control system.

AI/ML Overview

The provided document is a 510(k) Premarket Notification from the U.S. FDA for a Diode Laser 808nm, Model: SL-HR10. This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device rather than proving clinical effectiveness through detailed acceptance criteria and human studies for AI/software devices.

Therefore, the document does not contain the information requested in your prompt regarding acceptance criteria and studies for an AI/software device. Specifically, it does not include:

  • A table of acceptance criteria and reported device performance for an AI/software system.
  • Sample sizes for test sets, data provenance, or details on ground truth establishment for AI/software.
  • Information on experts used to establish ground truth or adjudication methods for AI/software.
  • Details about Multi-Reader Multi-Case (MRMC) comparative effectiveness studies or standalone algorithm performance.
  • Sample sizes or ground truth establishment methods for training sets of an AI/software.

Instead, the document details:

  • Device Name: Diode Laser 808nm, Model: SL-HR10
  • Intended Use: Hair removal, permanent hair reduction on all skin types (I-VI), including tanned skin. Permanent hair reduction is defined as the long-term, stable reduction in hair regrowth measured at 6, 9, and 12 months post-treatment.
  • Basis for SE: Non-clinical tests verifying compliance with relevant medical device safety and performance standards (e.g., AAMI/ANSI/ES 60601-1, IEC 60601-2-22, IEC 60825-1, IEC 60601-1-2, ISO 10993-5, ISO 10993-10).
  • Comparison to Predicates: The device is compared to two predicate devices (K180353 and K181019) on product code, regulation number, intended use, configuration, principle of operation, laser type, classification, wavelength, spot size, fluence, frequency, pulse duration, power supply, dimension, weight, patient contact materials, and biocompatibility.
  • No Clinical Study: The document explicitly states: "No clinical study is included in this submission."

In summary, this FDA submission is for a laser device, and its approval is based on demonstrating substantial equivalence to existing devices through non-clinical testing against established safety and performance standards, not through clinical trials or AI/software specific performance metrics.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 10, 2019

Beijing Superlaser Technology Co., Ltd. % Ray Wang General Manager Beijing Believe-Med Technology Service Co., Ltd. Rm. 912, Building #15, XiYueHui, No.5, YiHe North Rd., Fangshan District Beijing, 102401 China

Re: K192516

Trade/Device Name: Diode Laser 808nm, Model: SL-HR10 Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: September 10, 2019 Received: September 13, 2019

Dear Ray Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of

{1}------------------------------------------------

Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Purva Pandya Acting Assistant Director DHT4A: Division of General Surgery Devices OHT5: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

510(k) Number (if known)

K192516

Device Name Diode Laser 808nm, Model: SL-HR10

Indications for Use (Describe)

The Diode Laser 808nm is intended for hair removal, permanent hair reduction on all skin type I-VI), including tanned skin.

Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.

Type of Use (Select one or both, as applicable)
❌ Prescription Use (Part 21 CFR 801 Subpart D) ❍ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services

Food and Drug Administration

Office of Chief Information Officer

Paperwork Reduction Act (PRA) Staff

PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92.

The assigned 510(k) Number:

    1. Date of Preparation
      09/10/2019
  • Sponsor 2.

Beijing Superlaser Technology Co., Ltd.

No.2 Zhongfu Street, Economic and Technological Industrial Zone, Xihongmen Town, Daxing District, Beijing, 100076, China

Contact Person: Shi Shuang Position: Registration Specialist Tel: 86-10-81284899 to 806 Fax:86-10-81284899 Email: 672257488@qq.com

3. Submission Correspondent

Ray Wang General Manager

Beijing Believe-Med Technology Service Co., Ltd. Rm.912, Building #15, XiYueHui, No.5, YiHe North Rd., FangShan District, BeiJing, China 102401

Tel: +86-18910677558 Fax: +86-10-56335780 Email: ray.wang@believe-med.com

{4}------------------------------------------------

4. Identification of Proposed Device

Trade Name: Diode Laser 808nm Common Name: Powered Laser Surgical Instrument Model(s): SL-HR10

Regulatory Information: Classification Name: Powered Laser Surgical Instrument Classification: II; Product Code: GEX; Regulation Number: 21 CFR 878.4810; Review Panel: General & Plastic Surgery;

Indication For Use:

The Diode Laser 808nm is intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin.

Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.

5. Device Description

The Diode Laser 808nm is intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin.

Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.

The proposed device has six main function modules as following:

Water cooling system:

Through refrigeration compressor to achieve heat and cold exchange of air, the internal water tank by the water and water pump for cooling water cycle, with cooling water to cool the parts of the tools, the system to achieve this function is called water cooling system.

Power system:

Through the main circuit control system, according to the control requirements to provide the size of laser energy, the system to achieve this function is called the power system.

User Interface system:

Through the display screen and touch screen, 1. Provide users with relevant information about display equipment; 2. Users operate or set relevant parameters or functions; 3. Display fault information. The system that realizes this function is called human-machine switching system.

Handpiece system:

According to the choice of different treatment methods to choose the appropriate treatment tools, the system to achieve this function is called the handpiece system.

{5}------------------------------------------------

Main circuit power supply system:

The system that realizes the power supply of equipment and the control switch or stops the power supply is called the main circuit power supply system.

Main circuit control system:

It is used to control the parts of the equipment to work in a reasonable sequence and provide fault or warning sound. The system that realizes this function is called the main circuit control system.

{6}------------------------------------------------

6. Identification of Predicate Device

510(k) Number: K181019 Product Name: Diode laser System Manufacturer: Guangzhou Huafei Tongda Technology Co., Ltd.

510(k) Number: K180353 Product Name: Diode laser hair removal device Manufacturer: Zhengzhou PZ Laser Slim Technology Co., Ltd

7. Non-Clinical Test Conclusion

Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:

  • AAMI/ANSI/ES 60601-1:2012 Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance;

  • IEC 60601-2-22 Edition 3.1 2012-10, Medical Electrical Equipment - Part 2-22: Particular Requirements For Basic Safety And Essential Performance Of Surgical, Cosmetic, Therapeutic And Diagnostic Laser Equipment.

  • IEC 60825-1:2014, Safety of laser products - Part 1: Equipment classification and requirements.

  • IEC 60601-1-2:2014 , Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances -Requirements and tests

  • ISO 10993-5 Third Edition 2009-06-01, Biological Evaluation Of Medical Devices - Part 5: Tests For In Vitro Cytotoxicity. (Biocompatibility)

  • ISO 10993-10 Third Edition 2010-08-01, Biological Evaluation Of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization. (Biocompatibility)

    1. Clinical Test Conclusion

No clinical study is included in this submission.

{7}------------------------------------------------

9. Substantially Equivalent (SE) Comparison

ItemProposed DevicePredicate DeviceK180353Predicate DeviceK181019Remark
Product CodeGEXGEXGEXSAME
Regulation Number21 CFR 878.481021 CFR 878.481021 CFR 878.4810SAME
Intended UseThe Diode Laser 808nm isintended for hair removal,permanent hair reduction onall skin types (Fitzpatrickskin type I-VI), includingtanned skin.Permanent hair reduction isdefined as the long-term,stable reduction in thenumber of hairs regrowingwhen measured at 6, 9, and12 months after thecompletion of a treatmentregime.The Diode laser hair removaldevice is intended for hairremoval, permanent hairreduction on all skin types(Fitzpatrick skin type I-VI),including tanned skin.Permanent hair reduction isdefined as the long-term, stablereduction in the number ofhairs regrowing when measuredat 6, 9, and 12 months after thecompletion of a treatmentregime.The Diode Laser System isintended for hair removal,permanent hair reductionon all skin types(Fitzpatrick skin typeI-VI), including tannedskin.Permanent hair reduction isdefined as the long-term,stable reduction in thenumber of hairs regrowingwhen measured at 6, 9, and12 months after thecompletion of a treatmentregime.SAME
ConfigurationMain UnitMain UnitMain UnitSAME
HandpieceHandpieceHandpieceSAME
Foot ControlFoot ControlFoot ControlSAME
Principle ofOperationDiode LaserDiode LaserDiode LaserSAME

Table 7-1 General Comparison

{8}------------------------------------------------

ItemProposed DevicePredicate DeviceK180353Predicate DeviceK181019Remark
Laser TypeDiode LaserDiode LaserDiode LaserSAME
LaserClassificationClass IVClass IVClass IVSAME
Laser Wavelength808 nm808 nm808 nmSAME
Spot Size$1.2 cm^2$$1.44 cm^2$$1.2 cm^2$SAME
Fluence$1-70J/ cm^2$$1-100J/ cm^2$$5-40J/ cm^2$Discussion
Frequency1-20 Hz1-20 Hz1-5 HzSAME
Pulse Duration5~400ms10~400ms30-200msSAME
Power Supply110V 60 Hz or 230V50HzAC 110V/60Hz100-240V 50/60HzSAME
Dimension(LWH)598mm x 440mm x1093mm560mm x 380mm x 180mm450mm x 550mm x380mmSIMILAR
Weight60Kg60Kg50 KgSAME

Table 7-2 Performance Comparison

Discussion

The proposed device is different in fluence from the predicate device, the difference is very slight, and only in the range, the max. fluence of proposed device is between the two predicate devices, which means the proposed device could accomplishment the same indication for use with two predicate device. And the proposed device has passed the IEC60601-1 test, IEC60601-1-2 test, IEC60601-2-22 test , IEC60825-1 test and performance test(Energy density and Spot Size), the safety and performance of the product can be ensured.

So the proposed device is determined to be substantially equivalency with predicate device.

{9}------------------------------------------------

ItemProposed DevicePredicate DevicePredicate DeviceRemark
K180353K181019
Patient Contact Materials and Biocompatibility
Patient ContactMaterialsSapphire in handpieceSapphire in handpieceSapphire in handpieceSAME
CytotoxicityNo CytotoxicityNo CytotoxicityNo CytotoxicitySAME
SensitizationNo evidence of sensitizationNo evidence of sensitizationNo evidence of sensitizationSAME
IrritationNo evidence of irritationNo evidence of irritationNo evidence of irritationSAME
EMC, Electrical and Laser Safety
ElectricalSafetyComply with IEC 60601-1, IEC60601-2-22Comply with IEC 60601-1, IEC 60601-2-22Comply with IEC 60601-1, IEC 60601-2-22SAME
EMCComply with IEC 60601-1-2Comply with IEC 60601-1-2Comply with IEC 60601-1-2SAME
Laser SafetyComply with IEC 60601-2-22, IEC60825-1Comply with IEC 60601-2-22, IEC 60825-1Comply with IEC 60601-2-22, IEC 60825-1SAME

Table 11-3 Safety Comparison

10. Substantially Equivalent (SE) Conclusion

Based on the comparison and analysis above, the proposed device is determined to be Substantially Equivalent (SE) to the predicate device.

§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.

(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.