K Number
K192498

Validate with FDA (Live)

Date Cleared
2020-04-28

(230 days)

Product Code
Regulation Number
876.1500
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

This instrument has been designed to be used with an Olympus endoscope to deliver injectable materials into the urinary bladder wall during the transurethral endoscopic procedures.

Device Description

This instrument has been designed to be used with an Olympus endoscope to deliver injectable materials into the urinary bladder wall during the transurethral endoscopic procedures. The subject device is sterilized and packaged in a sterilization package.

AI/ML Overview

The provided document is a 510(k) premarket notification for a medical device called the "Single Use Injector NM-221C-0427." This is a device used in conjunction with an endoscope to deliver injectable materials into the urinary bladder wall.

Based on the content, specifically Section 7 "PERFORMANCE DATA," it is clear that this submission focuses on bench testing, sterilization/shelf-life testing, and biocompatibility testing to demonstrate substantial equivalence to a predicate device. There is no mention of a study involving AI, human readers, or image analysis.

Therefore, many of the requested details about acceptance criteria and studies related to AI performance, such as MRMC studies, expert ground truth establishment, training sets, and data provenance, are not applicable to this specific device submission as described in the provided text.

However, I can extract the acceptance criteria and performance data that were provided for this traditional medical device.

1. Table of Acceptance Criteria and Reported Device Performance

For this device, the "acceptance criteria" are implied by the successful completion of the specified performance tests. The document states that the tests were conducted "to ensure that the subject device performs as intended and meet design specifications."

Acceptance Criteria CategorySpecific Tests ConductedStated Performance (Implicit Acceptance)
Sterilization/Shelf-lifeSterilization in accordance with FDA GuidanceConducted in accordance with FDA Guidance. Accelerated aging test done (ASTM F1980-16). Real-time aging (3 years) ongoing.
BiocompatibilityCytotoxicity, ISO Intracutaneous, ISO Guinea Pig Maximization Sensitization, ISO Acute Systemic Toxicity, USP Rabbit Pyrogen StudyConducted in accordance with FDA Guidance (ISO 10993-1). Device considered externally communicating, limited exposure (up to 24 hrs).
Performance (Bench)Endoscope compatibilityConducted to ensure device performs as intended and meets design specifications.
Needle slidabilityConducted to ensure device performs as intended and meets design specifications.
Puncture performanceConducted to ensure device performs as intended and meets design specifications.
Liquid leakageConducted to ensure device performs as intended and meets design specifications.
DurabilityConducted to ensure device performs as intended and meets design specifications.
Flow rate (BS EN 1618: 1997)Conducted.
Corrosion Testing (ISO 9626: 2016)Conducted.
Risk AnalysisConducted in accordance with ISO 14971:2007Design verification tests and their acceptance criteria identified and performed as a result.

2. Sample size used for the test set and the data provenance

  • Sample Size: Not explicitly stated for each test, but standard engineering practices for medical devices (e.g., ISO, ASTM standards) imply sufficient samples were tested to demonstrate performance and meet statistical confidence. This is not a data-driven AI model, so "test set" in the context of images or AI performance isn't applicable.
  • Data Provenance: Not applicable in the context of image data. The testing was conducted by Olympus Medical Systems Corp. and Aomori Olympus Co., Ltd. (Japan) to support their 510(k) submission to the FDA (USA).

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • Not applicable. This device is a physical instrument, not an AI or image analysis system. Ground truth is established through adherence to engineering specifications and performance standards (e.g., "Needle slidability" is assessed against a defined mechanical threshold, not expert consensus on an image).

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Not applicable. See point 3.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No. The document explicitly states: "No clinical study was performed to demonstrate substantial equivalence." and "No animal study was performed to demonstrate substantial equivalence." This is not an AI-assisted diagnostic device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Not applicable. This is a physical medical device, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

  • For physical performance characteristics (e.g., flow rate, durability, puncture performance), engineering specifications and validated testing methods serve as the "ground truth." For biocompatibility and sterilization, the "ground truth" is adherence to established international standards (e.g., ISO, ASTM, USP) and FDA guidance.

8. The sample size for the training set

  • Not applicable. No training set was used as this is not an AI/machine learning device.

9. How the ground truth for the training set was established

  • Not applicable. See point 8.

In summary, the provided document describes the regulatory submission for a physical medical device (a single-use injector) that demonstrates substantial equivalence through bench testing, biocompatibility testing, and sterilization validation, rather than studies focused on AI performance or human reader interaction with an AI system.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

April 28, 2020

Olympus Medical Systems Corp. Sheri Musgnung Manager, Regulatory Affairs Olympus Corporation of the Americas 3500 Corporate Parkway PO Box 610 Center Valley, PA 18034-0610

Re: K192498

Trade/Device Name: Single Use Injector NM-221C-0427 Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: II Product Code: FBK Dated: March 19, 2020 Received: March 20, 2020

Dear Sheri Musgnung:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Martha W. Betz, Ph.D. Acting Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K192498

Device Name Single Use Injector NM-221C-0427

Indications for Use (Describe)

This instrument has been designed to be used with an Olympus endoscope to deliver injectable materials into the urinary bladder wall during the transurethral endoscopic procedures.

Type of Use (Select one or both, as applicable)

Research Use (Part 81 CFR 801.3 Subject to Review)
Genetic Testing (81.420)

Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The letters are all capitalized and evenly spaced. A registered trademark symbol is located to the right of the letter "S".

March 19, 2020

Single Use Injector NM-221C-0427

Revised 510(k) Summary

1 GENERAL INFORMATION

■ 510(k) Submitter:OLYMPUS MEDICAL SYSTEMS CORP.2951 Ishikawa-cho, Hachioji-shi, Tokyo 192-8507,Japan
-----------------------------------------------------------------------------------------------------------------
  • Contact Person: Sheri L. Musgnung Olympus Corporation of the Americas 3500 Corporate Parkway PO Box 610 Center Valley, PA 18034-0610, USA Phone: 484-896-3147 Fax: 484-896-7128 Email: sheri.musgnung@olympus.com
  • Manufacturing site: Aomori Olympus Co., Ltd. ■ 2-248-1 Okkonoki, Kuroishi-shi, Aomori 036-0357, Japan

2 DEVICE IDENTIFICATION

■ Device NameSingle Use Injector NM-221C-0427
■ Common Nameendoscopic injection needle, gastroenterology-urology
■ Regulation Number21 CFR 876.1500
■ Regulation NameEndoscope and accessories
■ Regulatory ClassII
■ Product CodeFBK
■ Classification PanelGastroenterology/Urology

3 PREDICATE DEVICE

■ Predicate device

Device name510(k) Submitter510(k) No.
BONEE NEEDLE FOR BLADDERINJECTIONS NB1070COLOPLAST A/SK090217

{4}------------------------------------------------

4 DEVICE DESCRIPTION

- General Description of the subject device

This instrument has been designed to be used with an Olympus endoscope to deliver injectable materials into the urinary bladder wall during the transurethral endoscopic procedures. The subject device is sterilized and packaged in a sterilization package.

Principle of Operation

When the operation portion of the needle section is pushed into the Sheath section, the needle protrudes from the distal end of the sheath section. Pulling the operation portion of the needle section toward the proximal end with respect to the sheath section pulls the needle into the coil sheath. The subject device is inserted into the endoscope in the retracted state. After the subject device reaches the urinary bladder wall, puncture is performed with the needle protruding from the coil sheath. A commercially available marketed syringe (not included in the subject device) is attached to the proximal end of the needle section to deliver injectable materials to the target site of the urinary bladder wall.

5 INDICATIONS FOR USE

This instrument has been designed to be used with an Olympus endoscope to deliver injectable materials into the urinary bladder wall during the transurethral endoscopic procedures.

6 COMPARISON OF TECHNOLOGY CHARACTERISTICS WITH THE PREDICATE DEVICE

The NM-221C-0427 has the same technological characteristics and design as the predicate device except for the following new features:

  • Needle design
  • Sheath design
  • Working length

Validation from non-clinical testing demonstrated that these technological features do not raise any new issues of safety or effectiveness of the subject device.

{5}------------------------------------------------

K192498

A side by side comparison of the subject device and the predicate device is provided below.

Subject DevicePredicate Device
ItemSingle Use InjectorNM-221C-0427BONEE NEEDLE FORBLADDER INJECTIONSNBI070
Indications for UseThis instrument has beendesigned to be used with anOlympus endoscope todeliver injectable materialsinto the urinary bladderwall during the transurethralendoscopic procedures.The Bonee Needle forBladder Injections is usedto deliver injectablematerials into the urinarybladder wall during thetransurethral endoscopicprocedures.
Regulation Number21 CFR 876.150021 CFR 876.1500
Regulation NameEndoscope and accessoriesEndoscope and accessories
Regulatory ClassIIII
Product CodeFBKFBK
Classification PanelGastroenterology/UrologyGastroenterology/Urology
Compatible OlympuscystoscopeWorking length: less than380mmChannel inner diameter:2.0mm or moreWorking length: unknownChannel inner diameter: 5French(= 1.7mm) or more
Needle width27G22G
Typical needle length4mm4mm
Maximum insertion portiondiameterφ1.84mmunknown
Working length971 mm700mm
Sheath typeSheath structure: dualstructure (outer: coil sheath,inner: tube sheath)Sheath structure: singlestructure (tube sheath only)
StyletEquipped (Diameter:φ0.3mm, Length: 700mm)Not equipped
Tip shape of needleTip shape: Lancet PointTip shape: Unknown

{6}------------------------------------------------

7 PERFORMANCE DATA

The following performance data were provided in support of the substantial equivalence determination.

1) Sterilization/Shelf-life testing

Sterilization/shelf-life testing for the NM-221C-0427 were conducted in accordance with the FDA's Guidance for Industry and Food and Drug Administration Staff, "Submission and Review of Sterility Information in Premarket Notification (510(k)) Submissions for Devices Labeled as Sterile".

Accelerated aging test for the NM-221C-0427 was conducted in accordance with ASTM F1980-16, the standard guide for accelerated aging of sterile barrier systems for medical devices. The real-time aging test for three-years will be performed to demonstrate longer stability and support the results of the accelerated aging test.

2) Biocompatibility testing

Biocompatibility testing for the NM-221C-0427 were conducted in accordance with the FDA's Guidance for Industry and Food and Drug Administration Staff. Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process". The Single Use Injector NM-221C-0427 is considered an externally communicating medical device in contact with Tissue/bone/dentin. The contact duration is limited exposure (i.e. contact is up to 24 hours). The biocompatibility testing included the following tests:

  • Cytotoxicity Study Using the Colony Assay ।
  • ISO Intracutaneous Study in Rabbits ।
  • ISO Guinea Pig Maximization Sensitization Test
  • ISO Acute Systemic Toxicity Study in Mice
  • USP Rabbit Pyrogen Study, Material Mediated ।

3) Performance testing - Bench

Bench testing for the NM-221C-0427 as listed below was conducted to ensure that the subject device performs as intended and meet design specifications.

  • Endoscope compatibility
  • Needle slidability
  • Puncture performance
  • Liquid leakage
  • Durability
  • Flow rate (BS EN 1618: 1997 Catheters other than intravascular catheters Test methods for common properties)
  • Corrosion Testing (ISO 9626: 2016 Stainless steel needle tubing for the manufacture of medical devices - Requirements and test methods)

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The letters are all capitalized and have a slightly three-dimensional appearance. A registered trademark symbol is located to the right of the letter "S".

K192498 Page 5 of 5

4) Performance testing - Animal

No animal study was performed to demonstrate substantial equivalence.

5) Performance testing - Clinical

No clinical study was performed to demonstrate substantial equivalence.

6) Risk analysis

Risk analysis for the NM-221C-0427 was conducted in accordance with established in-house acceptance criteria based on ISO 14971:2007. The design verification tests and their acceptance criteria were identified and performed as a result of this risk analysis assessment.

8 CONCLUSIONS

Based on the indications for use, technological characteristics, performance testing and technological comparison to the predicate device, the NM-221C-0427 raise no new issue of safety and effectiveness and are substantially equivalent to the predicate device in terms of safety, effectiveness and performance.

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.