K Number
K192464

Validate with FDA (Live)

Date Cleared
2020-08-25

(351 days)

Product Code
Regulation Number
880.5570
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Injection Pen Needle is intended for use with pen injector devices for the subcutaneous injection of drugs.

Device Description

The Injection Pen Needle is designed for use with pen injectors for subcutaneous injection of a desired dose of drugs approved for delivery using a pen needle. It consists of a seal, needle hub, needle tube, needle sheath and primary container. The Injection Pen Needle is offered in the following sizes: Model(s): 0.30x6mm/30G, 0.30x8mm/30G, 0.25x6mm/31G, 0.25x8mm/31G, 0.23x4mm/32G. It is a single-use disposable device that is provided sterile and has a shelf life of 3 years. The needle hub is connects to the pen injector and the needle tube punctures the sealed medicine container. The medicine solution is pushed out through the needle tube by the injection mechanism of the pen injector. The needle tube is a stainless steel needle tube with lubricant coating on the surface and a beveled tip. The seal (medical glue-coated dialysis paper) and the primary container (medical polypropylene) constitute the sterile barrier system of the product.

AI/ML Overview

This document describes the non-clinical testing performed to demonstrate that the Injection Pen Needle (K192464) is substantially equivalent to the predicate device (Droplet Pen Needles, K171982).

Here's the breakdown of the acceptance criteria and study information provided:

1. Acceptance Criteria and Reported Device Performance

Test CategoryAcceptance Criteria (Standards Met)Reported Device Performance
Performance- ISO 7864:2016 Sterile hypodermic needles for single use Requirements and test methods.Met all acceptance criteria for all gauges and lengths of needles, supporting a 3-year shelf life.
- ISO 9626:2016 Stainless steel needle tubing for the manufacture of medical devices - Requirements and test methods.
- ISO 11608-2:2012 Needle-based injection systems for medical use Requirements and test methods - Part 2: Needles.
Compatibility- Demonstrated with FDA cleared/approved injectors via ISO 11608-2:2012 clauses 4.4 and 4.9: PEN INJECTOR - HumaPen Luxura (K142518); PEN INJECTOR - NovoPen Echo (K182387); PEN INJECTOR - NOVOLOG MIX 70/30 (BLA 021172); PEN INJECTOR - LANTUS SOLOSTAR (BLA 021081).Demonstrated compatibility.
Biocompatibility- Evaluated per FDA Guidance: "Use of International Standard ISO 10993-1. Biological Evaluation of Medical Devices Part 1: Evaluation and Testing within a risk management process, June 2016" for limited contact (≤ 24 hrs) external communicating, circulating blood contacting device.Met all acceptance criteria for endpoints.
- Endpoints evaluated: Cytotoxicity (ISO 10993-5:2009), Sensitization (ISO 10993-10:2010), Irritation (ISO 10993-10:2010), Acute Systemic Toxicity (ISO 10993-11:2017), Hemocompatibility (ISO 10993-4:2017, ASTM F756), Pyrogenicity (USP41 <85> Bacterial Endotoxins Test).
Sterility & Shelf Life- Sterilized by Ethylene Oxide to achieve a SAL of 10-6.Met all acceptance criteria for sterility and ensured sterility maintained during a 3-year shelf life.
- Validated and tested in accordance with: ISO 11135:2014 Sterilization of health-care products Ethylene oxide Requirements for the development, validation and routine control of a sterilization process for medical devices; ISO 10993-7:2008 Biological evaluation of medical devices - Part 7: Ethylene oxide sterilization residuals; ASTM F1980-16 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices.

2. Sample size used for the test set and the data provenance

The document does not specify sample sizes for the conducted tests (performance, biocompatibility, and sterility). It also does not mention the data provenance (e.g., country of origin, retrospective or prospective) for these tests, as they are likely laboratory-based non-clinical tests rather than clinical studies with human data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This question is not applicable. The listed tests are non-clinical (laboratory/bench testing) of a medical device (Injection Pen Needle). They do not involve human experts establishing ground truth in the way a diagnostic study would. The acceptance criteria are based on recognized international and national standards and FDA guidance documents.

4. Adjudication method for the test set

This question is not applicable to non-clinical laboratory testing.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This question is not applicable, as the device is an Injection Pen Needle, not an AI-powered diagnostic tool requiring human reader studies.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This question is not applicable, as the device is an Injection Pen Needle, not an algorithm or AI system.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

For the non-clinical tests, the "ground truth" is defined by the acceptance criteria specified in the referenced international and national standards (e.g., ISO 7864, ISO 9626, ISO 11608-2, ISO 10993 series, ASTM F756, USP41 <85>, ISO 11135, ASTM F1980). The device's performance is measured against these established quantitative or qualitative requirements.

8. The sample size for the training set

This question is not applicable, as the device is a physical medical device and not an AI model that requires a training set.

9. How the ground truth for the training set was established

This question is not applicable, as the device is a physical medical device and not an AI model.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath. The logo is simple and professional, and it is easily recognizable.

August 25, 2020

Gangan Medical Technology Jiangsu Co., Ltd. % Ray Wang General Manager Beijing Believe-Med Technology Service Co., Ltd. R912, Building #15, XiYueHui, No.5, YiHe North Rd., FangShan District BeiJing. China 102401

Re: K192464

Trade/Device Name: Injection Pen Needle Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: Class II Product Code: FMI Dated: July 21, 2020 Received: July 23, 2020

Dear Ray Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For CAPT Alan Stevens Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

{3}------------------------------------------------

K192464 510(k) Summary

This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of Title 21, CFR Section 807.92.

1. Date of Preparation: August 12, 2020

2. Sponsor Identification/Contact Information:

Submitter Name: Gangan Medical Technology Jiangsu Co., Ltd. Addresss: 1st-4th floor, Building G130, fifth phase of standard workshop, China Medical City, 225300 Taizhou City, China

Contact Person: Wu Shuhui Position: Quality Manager Telephone number: +86-010-56965029 Fax Number:: +86-523-80108778 Email: ShuHui.Wu@ganlee.com

3. Identification of Subject Device

510(k) Number: K192464 Trade Name: Injection Pen Needle Common Name: Hypodermic single lumen needle Classification Name: Needle, hypodermic, single lumen Regulatory Classification: 2 Product Code: FMI Regulation Number: 21 CFR 880.5570 Review Panel: General Hospital;

Identification of Predicate Device(s) 4.

510(k) Number: K171982 Product Name: Droplet Pen Needles Common Name: Hypodermic single lumen needle Classification Name: Needle, hypodermic, single lumen Regulatory Classification: 2 Product Code: FMI Regulation Number: 880.5570 Review Panel: General Hospital

5. Indications for Use Statement:

The Injection Pen Needle is intended for use with pen injector devices for the subcutaneous injection of drugs.

Device Description 6.

The Injection Pen Needle is designed for use with pen injectors for subcutaneous injection of a desired dose

{4}------------------------------------------------

of drugs approved for delivery using a pen needle. It consists of a seal, needle hub, needle tube, needle sheath and primary container. The Injection Pen Needle is offered in the following sizes: Model(s): 0.30x6mm/30G, 0.30x8mm/30G, 0.25x6mm/31G, 0.25x8mm/31G, 0.23x4mm/32G

It is a single-use disposable device that is provided sterile and has a shelf life of 3 years. The needle hub is connects to the pen injector and the needle tube punctures the sealed medicine container. The medicine solution is pushed out through the needle tube by the injection mechanism of the pen injector.

The needle tube is a stainless steel needle tube with lubricant coating on the surface and a beveled tip. The seal (medical glue-coated dialysis paper) and the primary container (medical polypropylene) constitute the sterile barrier system of the product.

7. Comparison of Technology to Predicate Device

A technological comparison table is provided below that compares the subject device and predicate device:

ItemProposed Device(s)Predicate Device(K171982)Remark
Device nameInjection Pen NeedleDroplet Pen NeedleN/A
Classification NameNeedle, hypodermic, single lumenNeedle, hypodermic, single lumenSAME
Product CodeFMIFMISAME
Regulation Number21 CFR 880.557021 CFR 880.5570SAME
Intended UseThe Injection Pen Needle is intended for use with a pen injector device for the subcutaneous injection of drugs.The Droplet® Pen Needle is intended for use with a pen injector device for the subcutaneous injection of drugs.SAME
Prescription/OTCPrescription UsePrescription UseSAME
ComponentsPrimary Container, Needle Sheath, Needle Tube and HubPrimary Container, Needle Shield, Needle Tube and HubSAME
Supplied SterileYesYesSAME
Single useYesYesSAME
Principle of operationManualManualSAME
CompatibilityPen Injector DevicePen Injector DeviceSAME
Method of attachment to pen injectorScrew threadsScrew threadsSAME
Length4mm, 6mm, 8mm4mm, 5mm, 6mm, 8mm, 10mm, 12mmSAME
Gage30G, 31G, 32G29G, 31G, 32GDifferent
Sterilization MethodEOGamma irradiationDifferent
SterilitySAL=10-6SAL=10-6SAME
Shelf Life3 years5 yearsDifferent
Unit PackagingPolypropylene container with seal made of medical gradePolypropylene container with seal made of medical gradeSAME

Table 1 Comparison Table

{5}------------------------------------------------

paperpaper
MaterialsNeedle Tube304 austenitic stainless steelMedical Grade Stainless SteelSimilar
Hub, PrimaryContainer,Needle SheathPolypropylenePolypropyleneSimilar
LubricantMedical grade siliconeMedical grade siliconeSimilar
AdhesiveUV Cured AdhesiveNot AvailableDifferent
PerformanceISO 11608-2, ISO 9626, ISOISO 11608-2, ISO 9626, ISOSAME
Compliance78647864SAME
BiocompatibilityBiocompatibility evaluation perISO 10993-1Biocompatibility evaluation perISO 10993-1SAME

The indications for use and intended use of the subject device is equivalent to that of the predicate device. The differences that exist between the subject device and predicate device are the following:

  • . Gauge size
  • Shelf life
  • . Sterilization method
  • Materials of composition

Based on the aforementioned modifications to the subject device, the subject device does not raise different types of safety and effectiveness questions when compared to the predicate device. Differences between the devices were verified or validated through non-clinical testing.

Non-Clinical Test Summary 8.

Non clinical tests were conducted to verify that the Injection Pen Needle met all design specifications and is Substantially Equivalent (SE) to the predicate device. Testing was conducted in accordance with recognized standards and FDA Guidance.

Performance:

The Injection Pen Needle was tested for device performance with the following recognized standards. Results of the testing demonstrated that the acceptance criteria were met for all gauges and lengths of needles and would support a shelf life of 3 years.

  • ISO 7864:2016 Sterile hypodermic needles for single use Requirements and test methods. .
  • . ISO 9626:2016 Stainless steel needle tubing for the manufacture of medical devices -Requirements and test methods.
  • ISO 11608-2:2012 Needle-based injection systems for medical use Requirements and test methods -● Part 2: Needles.

Compatibly was demonstrated to the following FDA cleared/approved injectors via ISO 11608-2: 2012 clauses 4.4 and 4.9: PEN INJECTOR - HumaPen Luxura (K142518); PEN INJECTOR - NovoPen Echo (K182387); PEN INJECTOR - NOVOLOG MIX 70/30 (BLA 021172); PEN INJECTOR - LANTUS SOLOSTAR (BLA 021081).

Biocompatibility:

{6}------------------------------------------------

The Injection Pen Needle was evaluated for biocompatibility endpoints associated with a limited contact (≤ 24 hrs) external communicating, circulating blood contacting device per FDA Guidance: "Use of International Standard ISO 10993-1. Biological Evaluation of Medical Devices Part 1: Evaluation and Testing within a risk management process, June 2016'' and the list of recognized standards below. The endpoints evaluated include: Cyotoxicity, Sensitization, Irritation, Acute Systemic Toxicity, Hemocompatibility, Pyrogenicity. Results of the testing demonstrated that the acceptance criteria were met.

  • . ISO 10993-5:2009, Biological Evaluation of Medical Device, Part 5-Tests for Vitro cytotoxicity.
  • . ISO 10993-10:2010 Standard, Biological Evaluation of Medical Device, Part 10-Test for irritation and delay-type hypersensitivity.
  • . ISO 10993-11:2017 Biological evaluation of medical devices - Part 11: Tests for systemic toxicity.
  • ISO 10993-4:2017 Biological evaluation of medical devices--Part 4: Selection of tests for interactions . with blood.
  • . ASTM F756, Standard Practice for Assessment of Hemolytic Properties of Materials
  • USP41 < 85> Bacterial Endotoxins Test.

Sterility

The Injection Pen Needle is supplied sterile in sterile packaging for single use and is sterilized by Ethylene Oxide to achieve a SAL of 10-6 . Sterilization was validated and tested to ensure that sterility is maintained during the shelf life of 3 years in accordance with the following recognized standards. Results of the testing demonstrated that the acceptance criteria were met.

  • ISO 11135:2014 Sterilization of health-care products Ethylene oxide Requirements for the . development, validation and routine control of a sterilization process for medical devices
  • ISO 10993-7:2008 Biological evaluation of medical devices - Part 7: Ethylene oxide sterilization residuals
  • . ASTM F1980-16 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices

9. Substantially Equivalent Conclusion

Differences between the technological characteristics of the subject device as compared to the predicate do not raise different questions of safety and effectiveness. The differences between the device are supported by non-clinical testing. The Injection Pen needle is Substantially Equivalent (SE) to the Droplet Pen Needles, cleared under K171982.

§ 880.5570 Hypodermic single lumen needle.

(a)
Identification. A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.(b)
Classification. Class II (performance standards).