(97 days)
SICAT Implant is a software application of imaging information of the oral-maxillofacial region. The imaging data originates from medical scanners such as CT or CBCT scanners. SICAT Implant is intended for use as planning software to aid qualified dental professionals in the placement of dental implants and the planning of surgical treatments. The dental professionals' planning data can be exported from SICAT Implanning data includes in particular the positions and types of implants and drill sleeves to be used in the surgical procedures. This data may be used as input to design and manufacture surgical guides for dental implants.
SICAT Implant V2.0 is a pure software device. SICAT Implant V2.0 is a software application for the visualization of imaging information of the oral-maxillofacial region. The imaging data originates from medical scanners such as CT or CBCT scanners. SICAT Implant V2.0 is intended for use as planning software to aid qualified dental professionals in the placement of dental implants and the planning of surgical treatments. SICAT Implant V2.0 allows to name, position, move, rotate, resize and visualize dental implants and other planning objects (i.e. nerves) within the visualized 3D volume. Thus, dental professionals like implantologists are enabled to precisely plan the positions, orientations, types and sizes of implants to be placed in the patient's mandible/maxilla together with the related surgical procedures. The dental professionals' planning data can be exported from SICAT Implant. This planning data includes in particular the positions, orientations and types of implants and drill sleeves to be used in the surgical procedures. This data may be used as input to design and manufacture surgical quides for dental implants.
The provided text, a 510(k) summary for SICAT Implant V2.0, primarily focuses on demonstrating substantial equivalence to a predicate device (SICAT Implant V1.2) rather than presenting a detailed performance study with explicit acceptance criteria. Therefore, several requested pieces of information are not available in this document.
However, based on the information provided, we can deduce some aspects of the device's validation and the general nature of its acceptance criteria.
-
Table of Acceptance Criteria and Reported Device Performance:
The document does not provide a quantitative table of acceptance criteria and reported device performance in the format typically seen for novel AI algorithms demonstrating clinical effectiveness. Instead, it relies on demonstrating consistency with the predicate device and success in general software verification and validation activities.
Inferred Acceptance Criteria (Qualitative) and Reported Performance:
| Feature/Metric | Acceptance Criteria (Implied) | Reported Device Performance |
|---|---|---|
| Accuracy of Measurements (Length) | 100 µm | The device (SICAT Implant V2.0) performs at 100 µm accuracy for length measurement, consistent with the predicate device. |
| Accuracy of Measurements (Angular) | 1 degree | The device (SICAT Implant V2.0) performs at 1 degree accuracy for angular measurement, consistent with the predicate device. |
| Functional Equivalence | The device should perform core functions (visualization, planning, measurement, implant manipulation, nerve visualization, data import/export) consistently with or improve upon the predicate device (SICAT Implant V1.2). | The document details that SICAT Implant V2.0 has "complete reengineering including significant software re-write," "added planning functionality," and "updated user interface" while maintaining the same intended use. The extensive "Device Comparison Table" indicates that all functionalities of the predicate are present in V2.0, with additions like abutment and sleeve visualization/manipulation. It states "functions work as designed" and "performance requirements and specifications have been met." |
| Safety and Effectiveness | Modifications compared to the predicate device should not adversely affect the safety and effectiveness of the device. The device should comply with relevant medical device safety, quality, and software standards (e.g., ISO 14971, IEC 62304, IEC 62366, 21 CFR 820). | "The modifications to the predicate device SICAT Implant V1.2 do not affect safety and effectiveness of the proposed device SICAT Implant V2.0." "Safety and effectiveness of the product has been demonstrated in the context of its intended use." Compliance with ISO 14971, IEC 62304, IEC 62366, and 21 CFR 820 is explicitly stated. |
| Usability | The device should be usable for its intended purpose by qualified dental professionals. | "Usability Tests" were performed, indicating that usability was assessed as part of verification and validation. |
-
Sample size used for the test set and the data provenance:
The document does not specify a "test set" in the context of a clinical performance study with a particular sample size of patient data. The testing described is primarily in the realm of software verification and validation (unit tests, integration tests, system verification and validation tests, usability tests). Data provenance (country of origin, retrospective/prospective) related to patient data is not mentioned because this is a software update (V2.0 of SICAT Implant) and its substantial equivalence is being demonstrated against a previous version (V1.2).
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):
This information is not provided. Since the device is a planning software, and not an AI diagnostic algorithm that produces a "ground truth" for disease detection, the concept of establishing ground truth by expert consensus would not apply in the same way. The software's "accuracy" is defined by its ability to correctly measure and manipulate 3D data as per its specifications (e.g., bone density, nerve canals, implant positions). The ground truth for measurement accuracy would typically be established through phantom or calibrated physical models.
-
Adjudication method (e.g. 2+1, 3+1, none) for the test set:
Not applicable in the context of this 510(k) submission, as it focuses on software verification and equivalence rather than a reader study adjudicating findings.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
No MRMC study is mentioned. This device is described as "planning software to aid qualified dental professionals in the placement of dental implants and the planning of surgical treatments." It's not presented as a device with AI assistance that's being compared to unassisted human readers for diagnostic improvement.
-
If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
The document states "SICAT Implant V2.0 is a pure software device." It's a planning tool for dental professionals, not an AI algorithm that makes a standalone decision or diagnosis. Its performance is always in the context of aiding a human professional. The stated measurement accuracies (100 µm, 1 degree) could be considered "standalone" technical performance metrics for the software's internal calculations.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc):
For the core functionalities like measurement accuracy, the ground truth would typically be established through validation against known physical phantoms or highly precise imaging standards, as indicated by the stated accuracies of 100 µm and 1 degree. For general software function, the "ground truth" is that the software performs according to its functional specifications, as verified by various software tests.
-
The sample size for the training set:
This information is not applicable and not provided. The document describes a traditional software development process ("complete reengineering including significant software re-write") rather than a machine learning model that requires a training set.
-
How the ground truth for the training set was established:
Not applicable, as no machine learning training set is described.
{0}------------------------------------------------
Image /page/0/Picture/10 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). The logo features a stylized caduceus symbol on the left, followed by a blue square. To the right of the blue square, the text "U.S. Food & Drug ADMINISTRATION" is displayed in a clear, sans-serif font. The text is arranged in three lines, with "U.S." and "Drug" on the top line, "Food" on the second line, and "ADMINISTRATION" on the third line.
December 4, 2019
SICAT GmbH & Co. KG % Manfred Breuer Head of Quality Management and Regulatory Affairs Brunnenallee 6 53177 Bonn. NRW GERMANY
Re: K192348
Trade/Device Name: SICAT Implant V2.0 Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: Class II Product Code: LLZ Dated: October 2, 2019 Received: October 7, 2019
Dear Manfred Breuer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
510(k) Number (if known) K192348
Device Name SICAT Implant V2.0
Indications for Use (Describe)
SICAT Implant is a software application of imaging information of the oral-maxillofacial region. The imaging data originates from medical scanners such as CT or CBCT scanners. SICAT Implant is intended for use as planning software to aid qualified dental professionals in the placement of dental implants and the planning of surgical treatments. The dental professionals' planning data can be exported from SICAT Implanning data includes in particular the positions and types of implants and drill sleeves to be used in the surgical procedures. This data may be used as input to design and manufacture surgical guides for dental implants.
Type of Use (Select one or both, as applicable)
× Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
{3}------------------------------------------------
510(k) Summary for Implant V2.0
Content and format as required by section 21 CFR 807.92
( http://www.fda.gov/ MedicalDevices/ DeviceRegulationandGuidance/ HowtoMarket YourDevice/PremarketSubmissions/PremarketNotification510k/ucm142651.htm)
1. SUBMITTER/510(K) HOLDER
SICAT GmbH & Co. KG Brunnenallee 6 53177 Bonn Germany
| EstablishmentRegistration Number: | 3006098230 |
|---|---|
| Telephone: | +49 (228) 854697-82 |
| Facsimile: | +49 (228) 854697-99 |
| Primary Contact: | Mr. Dr. Manfred Breue |
| Secondary Contact: | Mr. Markus Gratzfeld |
| Date Prepared: | December 4th, 2019 |
2. DEVICE NAME AND DEVICE CLASSIFICATION
| Proprietary Name: | SICAT Implant V2.0 |
|---|---|
| Common/Usual Name: | Radiological Visualization Software for Diagnosis and DentalImplant Planning |
| Classification Name: | System, Image Processing, Radiological |
| Regulation Description: | Picture archiving and communications system |
| Product Code: | LLZ |
| Classification Panel: | Radiology |
| Regulation Number: | 21 CFR 892.2050 |
| Device Class: | Class II |
3. PREDICATE DEVICE
| Proprietary Name: | SICAT Implant V1.2 |
|---|---|
| Manufacturer: | SICAT GmbH & Co. KG |
| 510(k) number: | K103723 |
4. DEVICE DESCRIPTION
SICAT Implant V2.0 is a pure software device.
{4}------------------------------------------------
SICAT Implant V2.0 is a software application for the visualization of imaging information of the oral-maxillofacial region. The imaging data originates from medical scanners such as CT or CBCT scanners.
SICAT Implant V2.0 is intended for use as planning software to aid qualified dental professionals in the placement of dental implants and the planning of surgical treatments.
SICAT Implant V2.0 allows to name, position, move, rotate, resize and visualize dental implants and other planning objects (i.e. nerves) within the visualized 3D volume. Thus, dental professionals like implantologists are enabled to precisely plan the positions, orientations, types and sizes of implants to be placed in the patient's mandible/maxilla together with the related surgical procedures.
The dental professionals' planning data can be exported from SICAT Implant. This planning data includes in particular the positions, orientations and types of implants and drill sleeves to be used in the surgical procedures. This data may be used as input to design and manufacture surgical quides for dental implants.
The modifications to SICAT Implant V1.2 (K103723) include the following:
- Complete reengineering including significant software re-write -
- -Added planning functionality
- -Updated user interface
5. Indications for Use
Indications for use of SICAT Implant V2.0 are:
SICAT Implant is a software application for the visualization of imaging information of the oral-maxillofacial region. The imaging data originates from medical scanners such as CT or CBCT scanners. SICAT Implant is intended for use as planning software to aid qualified dental professionals in the placement of dental implants and the planning of surgical treatments. The dental professionals' planning data can be exported from SICAT Implant. This planning data includes in particular the positions, orientations and types of implants and drill sleeves to be used in the surgical procedures. This data may be used as input to design and manufacture surgical guides for dental implants.
6. Device Comparison Table
The modified device SICAT Implant V2.0 has the same intended use and fundamental scientific technology. A comparison of the proposed device SICAT Implant V2.0 to the currently marketed predicate device SICAT Implant V1.2 including intended use. technological characteristics, design and function is provided in the following table.
| SICAT Implant V2.0 | SICAT Implant V1.2,SICAT(marketed in EU / USAcleared under 510(k)number K103723) | ||
|---|---|---|---|
| Common/Usual Name | |||
| Radiological VisualizationSoftware for Diagnosis andDental Implant Planning | Radiological VisualizationSoftware for Diagnosis andDental Implant Planning | ||
| SICAT Implant V2.0 | SICAT Implant V1.2,SICAT(marketed in EU / USAcleared under 510(k)number K103723) | ||
| Device Description | |||
| General Description | SICAT Implant V2.0 is apure software device.SICAT Implant V2.0 is asoftware application for thevisualization of imaginginformation of the oral-maxillofacial region. Theimaging data originatesfrom medical scanners suchas CT or CBCT scanners.SICAT Implant V2.0 isintended for use asplanning software to aidqualified dentalprofessionals in theplacement of dentalimplants and the planningof surgical treatments.SICAT Implant V2.0 allowsto name, position, move,rotate, resize and visualizedental implants and otherplanning objects (i.e.nerves) within thevisualized 3D volume. Thus,dental professionals likeimplantologists are enabledto precisely plan thepositions, orientations,types and sizes of implantsto be placed in the patient'smandible/maxilla togetherwith the related surgicalprocedures.The dental professionals'planning data can beexported from SICATImplant. This planning dataincludes in particular thepositions, orientations andtypes of implants and drillsleeves to be used in thesurgical procedures. Thisdata may be used as inputto design and manufacture | SICAT Implant V1.2 is apure software device.SICAT Implant V1.2 is asoftware application for thevisualization of imaginginformation of the oral-maxillo-facial region. Theimaging data originatesfrom medical scanners suchas CT or DVT scanners.SICAT Implant V1.2 isintended for use asplanning and simulationsoftware to aid qualifieddental professionals in theplacement of dentalimplants and the planningof surgical treatments.SICAT Implant V1.2 allowsto name, position, move,rotate, resize and visualizedental implants and otherplanning objects (i.e. nervecanals) within the visualized3D volume. Thus, dentalprofessionals likeimplantologists are enabledto precisely plan thepositions, orientations,types and sizes of implantsto be placed in the patient'smandible/maxilla togetherwith the related surgicalprocedures.The dental professionals'planning data may beexported from SICATImplant V1.2 and used asinput data for CAD or RapidPrototyping Systems. | |
| SICAT Implant V2.0surgical guides for dentalimplants. | SICAT Implant V1.2,SICAT(marketed in EU / USAcleared under 510(k)number K103723) | ||
| Indications for Use/Intended use | SICAT Implant is a softwareapplication for thevisualization of imaginginformation of the oral-maxillofacial region. Theimaging data originatesfrom medical scanners suchas CT or CBCT scanners.SICAT Implant is intendedfor use as planningsoftware to aid qualifieddental professionals in theplacement of dentalimplants and the planningof surgical treatments. Thedental professionals'planning data can beexported from SICATImplant. This planning dataincludes in particular thepositions, orientations andtypes of implants and drillsleeves to be used in thesurgical procedures. Thisdata may be used as inputto design and manufacturesurgical guides for dentalimplants. | SICAT Implant is a softwareapplication for thevisualization of imaginginformation of the oral-maxillofacial region. Theimaging data originatesfrom medical scanners suchas CT or DVT scanners.SICAT Implant is intendedfor use as planning andsimulation software to aidqualified dentalprofessionals in theplacement of dentalimplants and the planningof surgical treatments. Thedental professionals'planning data may beexported from SICATImplant and used as inputdata for CAD or RapidPrototyping Systems. | |
| Contraindications | None | None | |
| Medical data viewing | |||
| Data typesvisualized | 3D volume data, 3D surfacedata (optical impressions,restorations, implantobjects) | 3D volume data, 3D surfacedata (optical impressions,implant objects) | |
| Imaging datavisualization region | Oral-maxillofacial region | Oral-maxillofacial region | |
| Viewing modes | |||
| *Orthogonal slices | Axial, coronal, sagittal | Axial, coronal, sagittal | |
| SICAT Implant V2.0 | SICAT Implant V1.2,SICAT(marketed in EU / USAcleared under 510(k)number K103723) | ||
| *Panoramic view | Panoramic view (ray sum)based on a panoramiccurve.Panoramic curve can bemanually adjusted. | Panoramic view (ray sum)based on a panoramiccurve.Panoramic curve can bemanually adjusted. | |
| *Panoramic slice | Movable small or notmovable maximized"Panoramic Slicing Window"embedded in panoramicview | Movable "Panoramic SlicingWindow" embedded inpanoramic view | |
| *TSA and LSA | Transversal slice (TSA) andlongitudinal slice (LSA),both with respect topanoramic curve panoramicview | Transversal slice (TSA) andlongitudinal slice (LSA),both with respect topanoramic curve | |
| * 3D volumerendering | Yes | Yes | |
| * 3D surfacerendering of volumedata | Yes | Yes | |
| Volume Data Navigation and Manipulation | |||
| *View manipulatingtools | Zoom, Pan, Change oforientation | Zoom, Pan, Change oforientation | |
| *Color manipulatingtools | Brightness, Contrast | Brightness, Contrast | |
| *Scrolling throughslices | Yes | Yes | |
| Volume orientation | Yes | Yes | |
| Measurements | |||
| *Lengthmeasurement | Yes | Yes | |
| *Anglemeasurement | Yes | Yes | |
| *Overall LengthMeasurementAccuracy | 100 μm | 100 μm | |
| *Overall AngularMeasurementAccuracy | 1 degree | 1 degree | |
| SICAT Implant V2.0 | SICAT Implant V1.2, SICAT(marketed in EU / USA cleared under 510(k) number K103723) | ||
| Optical surface data visualization | |||
| *Optical impression(maxilla, mandible,or a part thereof) | Polygonal mesh in 3D view and panorama view, cut through polygonal mesh (contour) in some slice views | Polygonal mesh in 3D view and panorama view, cut through polygonal mesh (contour) in some slice views | |
| Optical impressionidentification | Names of loaded optical impression in object browser. | A list view in a dedicated toolbar contains the names of all loaded optical impressions. | |
| * Restoration model | Polygonal mesh in 3D view and panorama view, cut through polygonal mesh (contour) in slice views. | Polygonal mesh in 3D view and panorama view, cut through polygonal mesh (contour) in slice views. | |
| Restoration modelidentification | Names of loaded restoration model in object browser, hierarchically sorted according to the corresponding optical impressions. | A list view in a dedicated toolbar contains the names of all loaded restorations, hierarchically sorted according to the corresponding optical impressions. | |
| *Show/Hide asurface object | YesVisibility applies to all views simultaneously. | YesVisibility applies to all views simultaneously. | |
| *Rendering ofoptical surface data | Yes, 3D | Yes, 3D | |
| Optical surface data manipulation | |||
| *Registration ofsurface data (opticalimpressions) tovolume data | Semi-automatic, using user specified reference points. | Semi-automatic, using user specified reference points. | |
| *Fine tuning theaccuracy of theregistration | Restart of the registration with a different set of reference points. | Restart of the registration with a different set of reference points. | |
| Delete | Yes, each loaded optical surface dataset separately. Restorations are deleted together with their associated optical impression. | Yes, each loaded optical impression separately. Restorations are deleted together with their associated optical impression. | |
| Input/Output (I/O) | |||
| *Volume dataimport | DICOM | DICOM | |
| SICAT Implant V2.0 | SICAT Implant V1.2,SICAT(marketed in EU / USAcleared under 510(k)number K103723) | ||
| *Optical surfacedata / opticalimpression import | Standard STL format andproprietary SSI or SIXDcontainer format. | Standard STL format andproprietary SSI. | |
| Data Import/Export | A "study" consisting ofplanning data, volume dataand optical surface datamay be exported andimported in a SICATproprietary format. | Proprietary: Planning,volume and surface data. | |
| *Data Compression | Lossless ZIP compression | Lossless ZIP compression | |
| * Export of data forrealization ofplanned therapy | For Surgical guide designand production by theSICAT dental laboratory ora third party manufacturer. | For Surgical guide designand production by theSICAT dental laboratory | |
| Export of planningreport | PNG | ||
| Implant visualization | |||
| * Body | Polygonal mesh (conicalfrustum) in 3D view and cutthrough polygonal mesh inplanar slice views. | Polygonal mesh (conicalfrustum) in 3D view and cutthrough polygonal mesh inplanar slice views | |
| * Axis | Line in 3D view andprojected line in planar sliceviews | Line in 3D view andprojected line in planar sliceviews | |
| Identification | Name (tooth number) ofimplants in object browser | Name (tooth number) ofactive implant in toolbar | |
| Dimensions(occlusal diameter,apical diameter,length)/Models | Dimensions/Models ofimplants in object browserand of active implant inproperties area. | Dimensions of (active)implant in toolbar. | |
| Implant database | |||
| Implant database | Realistic geometric data ofmanufacturer implants,abutments and sleeves | Realistic geometric data ofmanufacturer implants | |
| Implant manipulation and planning | |||
| * Plan/insertrealisticmanufacturer andgeneric implants | Via workflow toolbar1-click in all planar sliceviews and in the panoramicview. | 1-click in all planar sliceviews and in the panoramicview. | |
| * Plan/insertrealistic | Via workflow toolbar1-click with a rough implantpre-positioning based on | 1-click with a rough implantpre-positioning at the siteof the restoration | |
| SICAT Implant V2.0 | SICAT Implant V1.2,SICAT(marketed in EU / USAcleared under 510(k)number K103723) | ||
| manufacturer andgeneric implantsbased on arestoration includedin the opticalsurface data | the position and orientationof the restoration. | ||
| Delete implants | All implants or activeimplant via object browser | Active implant via toolbaror context menu | |
| * Move implants | Active implant in all planarslice views and in thepanoramic view | Active implant in all planarslice views | |
| * Rotate implants | Active implant about bothend points in all planar sliceviews | Active implant about bothend points in all planar sliceviews | |
| Change implantdimensions/models | Via object browser andproperties area. | Via toolbar or context menu | |
| Changeidentification | Tooth number of (active)implant via tooth chart | Tooth number via toothchart | |
| Safety marginvisualization | For all implants in all views | For active implant in allviews | |
| * Nerve | Polygonal mesh (tube) in3D view and panoramicview and cut throughpolygonal mesh in planarslice views | Polygonal mesh (tube) in3D view and panoramicview and cut throughpolygonal mesh in planarslice views | |
| Insert | Manual definition of asequence of spline controlpoints in all planar sliceviews | Manual definition of asequence of spline controlpoints in all planar sliceviews | |
| Delete | Via object browser. | Via toolbar | |
| Insert control point | In all planar slice views. | In all planar slice views | |
| Delete control point | Via properties area. | Via toolbar | |
| Move control point | In all planar slice views. | In all planar slice views | |
| Change diameter | Via properties area. | Via toolbar | |
| Abutment visualization | |||
| Body | Polygonal mesh in 3D viewand cut through polygonalmesh in planar slice views | No | |
| SICAT Implant V2.0 | SICAT Implant V1.2,SICAT(marketed in EU / USAcleared under 510(k)number K103723) | ||
| Identification | Tooth number ofcorresponding implant inobject browser | No | |
| Dimensions/Models | Dimensions/Models ofabutments in objectbrowser and of activeabutment in propertiesarea. | No | |
| Abutment manipulation and planning | |||
| Add compatiblerealisticmanufacturerabutments | Via workflow toolbar forcompatible plannedimplants. | No | |
| Add genericabutments forplanned implants | Via workflow toolbar forplanned implants. | No | |
| Delete abutments | Active abutment via objectbrowser | No | |
| Change abutmentdimensions/models | Via object browser andproperties area. | No | |
| Changeidentification | N/A, Identification viacorresponding implant | No | |
| Sleeve visualization | |||
| Body | Polygonal mesh in 3D viewand cut through polygonalmesh in planar slice views. | No | |
| Identification | Tooth number ofcorresponding implant inobject browser | No | |
| Dimensions/Models | Dimensions/models ofsleeve in object browserand for active sleeve inproperties area. | No | |
| Sleeve manipulation | |||
| Add realisticmanufacturersleeves | Via workflow toolbar orplan area for compatibleplanned implants | No | |
| Delete sleeves | All sleeves via objectbrowser and plan area | No | |
| Change sleevedimensions/models | Via object browser,properties area and planarea. | No | |
| SICAT Implant V2.0 | SICAT Implant V1.2,SICAT(marketed in EU / USAcleared under 510(k)number K103723) | ||
| Change sleeveposition | Via planar slice views,object browser andproperties area | No | |
| Changeidentification | N/A, Identification viacorresponding implant | No | |
| Generation ofplanning report fordocumentation | Yes | Yes | |
| ProgrammingLanguage | C# and C++ | C# and C++ | |
| PC Hardware &Softwarerequirements(minimum) | Operating System:Windows 7 64-bit SP1+"Platform Update"KB2670838, Windows 8.164-bit (Desktop), Windows10 64-bit (Desktop)CPU: Quad Core 2.3 GHzRAM: 8 GBHDD: 20 GB free spaceVideo card: 2 GB RAM(dedicated), must supportDirectX 11 or higher, featdriver WDDM 1.0 or higher,e.g. NVidia GTX 960 (2GBRAM)Screen:Resolution min. 1980x1080(Full HD), Scaling 100 % -125 %,Resolution max.3840x2160, Scaling 100 %- 200 % must pass testwith SMPTE Test patternAdditional requirementswhen running as SIDEXIS 4Module: SIDEXIS 4.3.1 orhigher, Sirona Planning API | Operating System:Windows XP with SP2 and.NET 2.0 Runtime Librariesor Windows VistaCPU: Dual Core 1.6 GHzRAM: 2 GBFree space on HDD: 5 GBVideo card: 128 MBexternal video card, 24-bitcolor, Shader Model 3 (foradvanced 3D rendering)Screen resolution:1280x1024 pixelsCD-Writer, networkinterface card (100 MBit/s,1 GBit/srecommended), keyboard,mouse |
{5}------------------------------------------------
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
{10}------------------------------------------------
{11}------------------------------------------------
{12}------------------------------------------------
- for functions and algorithms used in SICAT Implant V2.0 and SICAT Implant V2.0 having the same principle;
The software algorithms used in SICAT Implant V2.0 are equal or similar to algorithms used in the predicate device SICAT Implant V2.0.
{13}------------------------------------------------
The added abutment planning functionality is for visualization purposes and the dentist determines if the abutment offers a reasonable starting point for possible future prosthetics. Within SICAT Implant, the user is not able to design any abutments or parts thereof. Furthermore, within SICAT Implant the user is not able to design a dental restauration such as a crown. There is no kind of output at all for the design of the abutment or a dental restauration such as a crown.
The modifications to the predicate device SICAT Implant V1.2 do not affect safety and effectiveness of the proposed device SICAT Implant V2.0.
Based on this information, SICAT GmbH & Co. KG believes that SICAT Implant V2.0 described in this submission is substantially equivalent to the predicate device SICAT Implant V1.2.
7. Non-Clinical Performance Testing and Verification and Validation Activities
No performance standards applicable to this device have been adopted under Section 514 of the Act.
Risk management has been ensured via risk analysis in compliance with ISO 14971:2007 to identify and provide mitigation to potential hazards beginning early in the design cycle and continuing throughout the development of the product. SICAT GmbH & Co. KG adheres to recognized and established medical industry standards for development including ISO 13485, IEC 62304 and IEC 62366. The device is designed and manufactured in accordance with Quality Systems Regulations as outlined in 21 CFR 820.
Verification and validation activities have been successfully performed on the software, including assurance that functions work as designed, performance requirements and specifications have been met, and that all hazard mitigations have been fully implemented.
Among others the following verification and validation activities were performed:
- Requirements Reviews
- . Design Validation/Reviews
- Risk Management ●
- Software verification and validation including ●
- O Unit Tests
- Code Reviews O
- Usability Tests O
- Integration Tests O
- System Verification and Validation Tests O
For SICAT Implant V2.0, software verification and validation activities were performed, in accordance with the following quidance and standards:
- Principles of Software Validation; Final Guidance for Industry and FDA Staff, ● January 2002
- ISO 14971:2007-Ed.2 Medical devices Application of risk management to medical ● devices
- AAMI/ANSI/IEC 62304:2006, Medical device software Software life cycle processes ●
- IEC 62304 Edition 1.1 2015-06 Medical device software - Software life cycle processes
{14}------------------------------------------------
- . IEC 62366-1 Edition 1.0 2015-02 Medical Devices - Part 1: Application Of Usability Engineering To Medical Devices [Including CORRIGENDUM 1 (2016)]
The above-mentioned verification and validation activities concluded that SICAT Implant V2.0 passed all verification and validation activities and that safety and effectiveness of the product has been demonstrated in the context of its intended use.
8. Conclusion
Based on the information and supporting documentation provided in the premarket notification, SICAT Implant V2.0 is considered to be comparable in design, material and function to the predicate device SICAT Implant V1.2. It is believed to perform as well as the predicate device for the visualization of imaging information of the oral-maxillofacial region toqether with optical impression data, the diagnostic purpose and the planning functionality to aid qualified dental professionals in the placement of dental implants and the planning of surgical treatments. Any minor differences between SICAT Implant V2.0 and the predicate device do not affect safety and effectiveness.
N/A