K Number
K192342

Validate with FDA (Live)

Date Cleared
2020-05-21

(267 days)

Product Code
Regulation Number
876.4300
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Electrosurgical Generator is intended to deliver High Frequency (HF) electrical current for the cutting and/or coagulation of tissue.

Bipolar Polypectomy Snare Combination has been designed to be used with endoscope and electrosurgical unit for the removal of sessile polyps, pedunculated polyps, tissue and foreign bodies from within the GI tract using high-frequency current.

Single Use Electrosurgical Knife Combination has been designed to be used with endoscopes and electrosurgical units to cut tissue within the digestive tract using high-frequency current.

Device Description

The Electrosurgical System comprises:

  • ◆ Electrosurgical Generator;
  • ◆ Bipolar polypectomy snare combination;
  • Single use electrosurgical knife combination.

The Electrosurgical Generator is designed to deliver bipolar high-frequency current for the purpose of cutting and coagulating tissue. It consists of High frequency generator, footswitch, bipolar connecting cable and power cord. The Electrosurgical Generator output is actuated via a two-pedal Footswitch. One pedal activated the bipolar high-frequency current output for cut; the other pedal activates the bipolar high-frequency current output for coagulation. The Electrosurgical Generator incorporates proprietary software developed by AGS for generating and controlling the two energies delivered. The Bipolar connecting cable connects electrosurgical instruments to the Electrosurgical Generator, it transfer electrical energy - without changing the energy - from devices to instruments. The Electrosurgical Generator, Footswitch, Bipolar connecting cable and Power cord are non-sterile and reusable.

Bipolar polypectomy snare combination is an accessory to Electrosurgical Generator. It is a sterile device consists of Polypectomy snare (Bipolar) and Distal attachment. It is for endoscopic use. It is used for removal of polyps within the digestive tract using high-frequency current. EO sterilization and use for single use only.

Single use electrosurgical knife combination is an accessory to Electrosurgical Generator. It is a sterile device consists of Single use electrosurgical knife and Distal attachment. It is for endoscopic use. It is used for cutting tissues within the digestive tract using high-frequency current. EO sterilization and use for single use only.

AI/ML Overview

The provided text is a 510(k) summary for an Electrosurgical System, comprising an Electrosurgical Generator, a Bipolar polypectomy snare combination, and a Single use electrosurgical knife combination. However, the document does not describe a study involving acceptance criteria and associated device performance for an AI/ML-based medical device.

Instead, it focuses on demonstrating substantial equivalence to predicate electrosurgical devices through:

  • Comparison of Technology Characteristics: This section details differences in output parameters for the electrosurgical generator (e.g., frequency, voltage, power) and differences in the principles of operation, energy use, and structure for the accessories (bipolar vs. monopolar, materials).
  • Performance Testing: This section lists the types of performance testing conducted to address the differences between the proposed device and predicate devices. These tests are primarily focused on electrical safety, thermal effects on tissue, biocompatibility, and bench testing.

Crucially, the document explicitly states "No Clinical test is included in this submission" (Page 12, Section 5.9). This strongly indicates that the device's clearance was based on demonstrating equivalence through non-clinical performance data and comparison to existing predicate devices, rather than a clinical study evaluating its acceptance criteria in human subjects, especially not a study involving human readers or AI assistance.

Therefore, I cannot extract the requested information (acceptance criteria table, sample size for test set with provenance, number/qualifications of experts, adjudication method, MRMC study details, standalone performance, ground truth type for test set, training set details) because the document does not describe an AI/ML device or a clinical study that would have generated such data in the context of human-in-the-loop performance or diagnostic accuracy.

The provided text pertains to a traditional electrosurgical device, not an AI/ML-driven solution that would involve complex data analysis or human reader performance studies.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

May 21, 2020

Hangzhou AGS MedTech Co., Ltd. Chunqi Han R&D Director Building 5, Building 6, No.597 Kangxin Road Yuhang District Hangzhou, Zhejiang, 311106 CHINA

Re: K192342

Trade/Device Name: Electrosurgical System (Electrosurgical Generator with Bipolar polypectomy snare combination and Single use electrosurgical knife combination accessories) Regulation Number: 21 CFR 876.4300 Regulation Name: Endoscopic electrosurgical unit and accessories Regulatory Class: II Product Code: KNS Dated: August 23, 2019 Received: August 28, 2019

Dear Chunqi Han:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for

Shani P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K192342

Device Name

Electrosurgical System (Electrosurgical Generator with Bipolar polypectomy snare combination and Single use electrosurgical knife combination accessories)

Indications for Use (Describe)

Electrosurgical Generator is intended to deliver High Frequency (HF) electrical current for the cutting and/or coagulation of tissue.

Bipolar Polypectomy Snare Combination has been designed to be used with endoscope and electrosurgical unit for the removal of sessile polyps, pedunculated polyps, tissue and foreign bodies from within the GI tract using high-frequency current.

Single Use Electrosurgical Knife Combination has been designed to be used with endoscopes and electrosurgical units to cut tissue within the digestive tract using high-frequency current.

Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K192342 Page 1 of 10

Image /page/3/Picture/1 description: The image shows the logo for AGS Medtech. The logo consists of a circular design with three curved shapes in blue and green. To the right of the circular design, the text "AGS MEDTECH" is written in English, followed by the same text in Chinese characters. The logo appears to be for a medical technology company.

Section 5 510(k) Summary Electrosurgical System

510(k) Summary

We submit this 510(k) Summary as per 21 CFR 807.92, it meets the content and format regulatory requirements.

5.1 Submitter

Submitted by/Owner:Hangzhou AGS MedTech Co., Ltd.Building 5, Building 6, NO.597 Kangxin Road YuhangDistrict, Hangzhou, Zhejiang 311106 China
EstablishmentRegistration Number:3010288205
Registration Status:Active
Contact Person:Yanping FuPhone: 0086-15958493282Fax: 0086-0571-87671225Email: fuyp@bioags.com
Date Prepared:July 18, 2019

5.2 Proposed Device

Trade Name:Electrosurgical System (Electrosurgical Generator withBipolar polypectomy snare combination and Single useelectrosurgical knife combination accessories)
Device Name:Electrosurgical System (Electrosurgical Generator withBipolar polypectomy snare combination and Single useelectrosurgical knife combination accessories)
Common Name:Electrosurgical System
Regulation class:Class II
Regulation Number:876.4300
Regulation Description:Endoscopic electrosurgical unit and accessories
Review Panel:Gastroenterology/Urology
Product Code:KNS
Product Code Name:Unit, Electrosurgical, Endoscopic (With Or WithoutAccessories)

5.3 Predicate Device

Trade Name:ERBE ESU Model VIO 300D with Accessories
Device Name:ERBE ESU Model VIO 300D with Accessories
Common Name:Electrosurgical Unit (ESU/Generator) System
510(k) Number:K083452
Regulation class:Class II
Regulation Number:21 CFR 878.4400
Regulation Description:Electrosurgical cutting and coagulation device and accessories.

{4}------------------------------------------------

K192342 Page 2 of 10

Image /page/4/Picture/1 description: The image shows a logo for AGS Medtech. The logo consists of a circular design with three blue curved shapes forming a triangle, with a green triangle in the center. To the right of the circular design, the text "AGS MEDTECH" is written in English, followed by "安杰思医学" in Chinese.

Section 5 510(k) Summary Electrosurgical System

Review Panel:General & Plastic Surgery
Product Code:GEI
Product Code Name:Electrosurgical, Cutting & Coagulation & Accessories
Trade Name:Polypectomy snare
Device Name:Polypectomy snare
Common Name:Polypectomy snare
510(k) Number:K172729
Regulation class:Class II
Regulation Number:21 CFR 876.4300
Regulation Description:Endoscopic electrosurgical unit and accessories
Review Panel:Gastroenterology/Urology
Product Code:FDI
Product Code Name:Snare, Flexible
Trade Name:Single use Electrosurgical Knife Series
Device Name:Single use Electrosurgical Knife Series
Common Name:Electrosurgical Knife
510(k) Number:K092309
Regulation class:Class II
Regulation Number:21 CFR 876.4300
Regulation Description:Endoscopic electrosurgical unit and accessories
Review Panel:Gastroenterology/Urology
Product Code:KNS
Product Code Name:Unit, Electrosurgical, Endoscopic (With Or Without Accessories)
Trade Name:Single use Electrosurgical knife
Device Name:Single use Electrosurgical knife
Common Name:Single use Electrosurgical knife
510(k) Number:K171158
Regulation class:Class II
Regulation Number:21 CFR 876.4300
Regulation Description:Endoscopic electrosurgical unit and accessories
Review Panel:Gastroenterology/Urology
Product Code:KNS
Product Code Name:Unit, Electrosurgical, Endoscopic (With Or Without Accessories)

5.4 Device Description

The Electrosurgical System comprises:

  • ◆ Electrosurgical Generator;
  • ◆ Bipolar polypectomy snare combination;

{5}------------------------------------------------

K192342 Page 3 of 10

Image /page/5/Picture/1 description: The image shows the logo for AGS Medtech. The logo consists of a blue and green circular design on the left, with the text "AGS MEDTECH" and "安杰思医学" on the right. The circular design appears to be a stylized representation of three interconnected shapes, with the green shape in the center.

Section 5 510(k) Summary Electrosurgical System

Single use electrosurgical knife combination. �

The Electrosurgical Generator is designed to deliver bipolar high-frequency current for the purpose of cutting and coagulating tissue. It consists of High frequency generator, footswitch, bipolar connecting cable and power cord. The Electrosurgical Generator output is actuated via a two-pedal Footswitch. One pedal activated the bipolar high-frequency current output for cut; the other pedal activates the bipolar high-frequency current output for coagulation. The Electrosurgical Generator incorporates proprietary software developed by AGS for generating and controlling the two energies delivered. The Bipolar connecting cable connects electrosurgical instruments to the Electrosurgical Generator, it transfer electrical energy - without changing the energy - from devices to instruments. The Electrosurgical Generator, Footswitch, Bipolar connecting cable and Power cord are non-sterile and reusable.

Bipolar polypectomy snare combination is an accessory to Electrosurgical Generator. It is a sterile device consists of Polypectomy snare (Bipolar) and Distal attachment. It is for endoscopic use. It is used for removal of polyps within the digestive tract using high-frequency current. EO sterilization and use for single use only.

Single use electrosurgical knife combination is an accessory to Electrosurgical Generator. It is a sterile device consists of Single use electrosurgical knife and Distal attachment. It is for endoscopic use. It is used for cutting tissues within the digestive tract using high-frequency current. EO sterilization and use for single use only.

5.5 Indication for use statement

Electrosurgical Generator is intended to deliver High Frequency (HF)

electrical current for the cutting and/or coagulation of tissue.

Bipolar Polypectomy Snare Combination has been designed to be used with endoscope and electrosurgical unit for the removal of sessile polyps, pedunculated polyps, tissue and foreign bodies from within the GI tract using high-frequency current.

Single Use Electrosurgical Knife Combination has been designed to be used with endoscopes and electrosurgical units to cut tissue within the digestive tract using high-frequency current.

5.6 Comparison of Technology Characteristics

Our proposed device Electrosurgical System is substantially equivalent to the predicate devices. The differences between the Electrosurgical System and the predicate devices do not raise any questions regarding its safety and effectiveness. The differences are listed in the table below.

{6}------------------------------------------------

Image /page/6/Figure/0 description: The image shows a logo for AGS Medtech. The logo consists of a blue circular design with a green triangle in the center. To the right of the logo, the text "AGS MEDTECH" is written in English, followed by the same name written in Chinese characters.

5.6.1 Electrosurgical Generator

DifferenceProposed devicePredicate deviceK083452Performance Testing
Output parametersCUT I:Rated Output frequency: 460 kHz±10%;Max. voltage output: 480Vp;Bipolar CUT:Rated Output frequency: 350kHz ±10%;Max. voltage output: 740Vp;Different rated output frequency, but in the range of200k-500MHz;Electrical safety reports.Thermal Effect on Tissue study report;The difference raises no new questions regardingsafety and efficacy.
CUT II:Working mode: Bipolar;Rated Output frequency: 460 kHz±10%;Rated Power: 120W±20%;Max. voltage output: 650Vp;Crest factor: 1.4~6.3ENDO CUT Q:Working mode: Monopolar;Rated Output frequency: 350kHz ±10%;Rated Power: 400W+0%/-20%;Max. voltage output: 770Vp;Crest factor: 1.4 ( $RL=500Ω$ )Electrical safety reports.Thermal Effect on Tissue study report;Animal study report;The difference raises no new questions regardingsafety and efficacy.
CUT III:Working mode:Bipolar;Rated Output frequency: 460kHz±10%;Rated Power: 60W±20%;Max. voltage output: 560Vp;Crest factor: 1.9~7.6ENDO CUT I:Working mode:Monopolar;Rated Output frequency: 350kHz ±10%;Rated Power: 170W±20%;Max. voltage output: 550Vp;Crest factor: 1.4 ( $RL=500Ω$ )Electrical safety reports.Thermal Effect on Tissue study report;Animal study report;The difference raises no new questions regardingsafety and efficacy.
COAG I:Rated Output frequency: 460 kHz±10%;Rated load: 100Ω;Bipolar Soft COAG:Rated Output frequency: 350kHz ±10%;Rated load: 75Ω;Different rated output frequency, but in the range of200k-500MHz;Electrical safety reports.Thermal Effect on Tissue study report;
DifferenceProposed devicePredicate deviceK083452Performance Testing
The difference raises no new questions regardingsafety and efficacy.
COAG II:Rated Output frequency: 460 kHz±10%;Bipolar Forced COAG:Rated Output frequency: 350kHz±10%;Different rated output frequency, but in the range of200k-500MHz;Electrical safety reports.Thermal Effect on Tissue study report;The difference raises no new questions regardingsafety and efficacy.

{7}------------------------------------------------

Image /page/7/Figure/0 description: The image shows the logo for AGS Medtech. The logo consists of a circular design with three blue curved shapes and a green shape in the center. To the right of the logo is the text "AGS MEDTECH" in English, followed by the same name in Chinese characters. The text is in a simple, sans-serif font.

5.6.2 For Bipolar polypectomy snare combination

DifferenceProposed devicePredicate deviceK172729Performance Testing
Principles ofoperationBipolar Polypectomy Snare Combination is an appliedpart of electrosurgical generator, using bipolarhigh-frequency current delivered by theelectrosurgical generator to cut and coagulate tissuewith the electrode.We designed a diffusion electrode attached to the tipof its sheath and an active electrode attached to theelectrode on the end of the sheath, and designed theDistal Attachment to connect with the diffusionelectrode to provide a larger contact area. Thehigh-frequency electricity flows from the electrode tothe Distal Attachment. Distal Attachment provides areturn path for high-frequency current with a lowMonopolar Polypectomy Snare manufacturedis an applied part of electrosurgical generator,using monopolar high-frequency currentdelivered by the electrosurgical generator tocut and coagulate tissue with the electrode.The high-frequency electricity flows from theactive electrode to the neutral electrode placedon patient skin.Electrical safety reports.Thermal effect on tissue study report.The difference raises no newquestions regarding safety andefficacy.
DifferenceProposed devicePredicate deviceK172729Performance Testing
current density and closer to the other electrode, toreduce the area of human tissue wherehigh-frequency currents circulate and prevent theoccurrence of anticipated burns in human tissue. Noneutral electrode is needed.
Energy UseBiopolar High-Frequency CurrentMonopolar Radio Frequency CurrentElectrical safety reports.Thermal effect on tissue study report.The difference raises no newquestions regarding safety andefficacy.
StructureBipolar Polypectomy snare combination consists ofPolypectomy snare (Bipolar) and Distal Attachment.Polypectomy snare (Bipolar) consists of Electrode,External tube, Sheath tube, Base, Connector, Handleand Conductor.Distal Attachment consists of Plastic cap, Connectpiece, Metal tube and Connect tube.Polypectomy snare consists of Electrode,Sheath, Fixed/Rotary part, Sliding handle,Connector plug and handle.Bench testing report.Thermal effect on tissue study report.The difference raises no newquestions regarding safety andefficacy.
Materials orsubstances incontact with thesame human tissueor body fluids1. Polypectomy Snare (Bipolar):Sheath: PTFE;External tube:SUS304;Electrode:SUS304;2.Distal Attachment:Plastic Cap: TPE;Connect piece: SUS304;Sheath: PTFE;Electrode: SUS304;Biocompatibility tests reports.The difference raises no newquestions regarding safety andefficacy.
DifferenceProposed devicePredicate devicePerformance Testing
K172729
Metal tube: SUS303;
Connect tube:SUS304;
5.6.3 For Single use electrosurgical knife combination
DifferenceProposed devicePredicate devicePerformance Testing
K171158;
K092309;
Principles ofAGS's Bipolar Electrosurgical Knife Combination isMonopolar Single Use Electrosurgical KnifeElectrical safety reports.
operationan applied part of electrosurgical generator, usingis an applied part of electrosurgical generator,Thermal effect on tissue study report.
bipolar high-frequency current delivered by theusing monopolar high-frequency currentdifference raises noThenew
electrosurgical generator to cut and cauterize tissuedelivered by thequestionsregarding safetyand
with the electrode.electrosurgical generator to cut and cauterizeefficacy.
The Bipolar Electrosurgical Knife Combination hastissue with the electrode. The high-frequency
a diffusion electrode (named 'External tube')electricity flows from the active electrode to
attached to the tip of its sheath and an activethe neutral electrode placed on patient skin.
electrode attached to the knife on the end of the
sheath. The active electrode is used to perform cut
and cauterize while the external tube provides a
return path for bipolar high-frequency current.
However, the contact area of External tube is so
small, we design the Distal Attachment to provide a
larger contact area. Distal Attachment is installed at
the distal end of the endoscope, Bipolar Single Use
Electrosurgical Knife pass through the channel of
endoscope, connect tube of the Distal Attachment
DifferenceProposed devicePredicate deviceK171158;K092309;Performance Testing
connect with the External tube of the Bipolar SingleUse Electrosurgical Knife. The high-frequencyelectricity flows from the knife to the DistalAttachment. Distal Attachment provides a returnpath for high-frequency current with a low currentdensity and closer to the other electrode, to reducethe area of human tissue where high-frequency
Energy UseBiopolar High-Frequency CurrentMonopolar Radio Frequency CurrentElectric safety reports.Thermal effect on tissue study report.The difference raises no newquestions regarding safety andefficacy.
StructureSingle use electrosurgical knife combination consistsof single use electrosurgical knife and distalattachment. Single use electrosurgical knife consistsof electrode part, sheath part and handle part.Distal Attachment consists of Plastic cap, Connectpiece, Metal tube and Connect tube.Single use electrosurgical knife consists ofelectrode part, sheath part and handle part.Bench test report.Thermal effect on tissue study report.The difference raises no newquestions regarding safety andefficacy.
Maximum insertionportion diameter$\Phi$ 2.4mmKD-655 L: $\Phi$ 2.7mm;KD-620LR: $\Phi$ 2.6mm;Bench test report.The difference raises no newquestions regarding safety andefficacy.
DifferenceProposed devicePredicate devicePerformance Testing
(mm)K171158;K092309;
Working Length5521: 1650mm,1950mm, 2300mm;5522: 1650mm,1950mm;5524: 1650mm,1950mm, 2300mm;KD-612L: Φ2.6mm.KD-655L: 1650mm;KD-620LR: 1650mm;KD-611IL: 1650mm;questions regarding safety and efficacy.
Cutting knife length5521: 1.4mm, 2.1mm5522: 4.0mm5524: 3.0mm, 3.5mmKD-655 L: 2.0mm;KD-620LR: 4.5mm;KD-611L: 4mm;
Tip diameter5524: Ф1.2mm, Ф1.7mmKD-612: Φ2.2mm
Materials orsubstances incontact with thesame human tissueor body fluidsPTFE, SUS304, SUS303No information.Biocompatibility tests reports.The difference raises no newquestions regarding safety andefficacy.

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image contains a logo with the text "AGS MEDTECH" and Chinese characters below it. The logo consists of three blue curved shapes forming a triangle with a green shape in the center. The text "AGS MEDTECH" is in a sans-serif font, and the Chinese characters appear to be a translation of the company name. The logo and text are positioned side by side, with the logo on the left and the text on the right.

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image contains a logo with the text "AGS MEDTECH" and Chinese characters below it. The logo consists of three blue curved shapes forming a triangle with a green shape in the center. The text "AGS MEDTECH" is in a sans-serif font, and the Chinese characters appear to be a translation of the company name. The logo and text are positioned side by side, with the logo on the left and the text on the right.

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image contains a logo with a circular design on the left and text on the right. The circular design consists of three blue curved shapes that form a triangle, with a green shape in the center. To the right of the logo, there is the text "AGS MEDTECH" in English, followed by the same text in Chinese characters. The text is aligned to the right of the logo and appears to be the name of a company or organization.

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image shows the logo for AGS Medtech. The logo consists of a circular design with three blue curved shapes and a green shape in the center. To the right of the logo is the text "AGS MEDTECH" in English, followed by the same name in Chinese characters. The text is in a simple, sans-serif font.

5.7 Applicable Guidance Document

NA

5.8 Performance Data

The Electrosurgical System meets all device standards for electrical safety and EMC (IEC 60601), biocompatibility (ISO 10993) and sterility (ISO 11135). The performance in ex-vivo and in-vivo meets the design specification and shows substantial equivalence to the predicated devices.

5.9 Clinical Test

{12}------------------------------------------------

Image /page/12/Picture/0 description: The image contains a logo for AGS Medtech. The logo consists of a circular design with three curved shapes in blue and green. To the right of the logo, the text "AGS MEDTECH" is written in English, followed by the same name written in Chinese characters. The logo and text are positioned side by side, creating a clear and concise brand representation.

K192342 Page 10 of 10

Section 5 510(k) Summary Electrosurgical System

No Clinical test is included in this submission.

5.10 Conclusion

In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807, Based on the information provided in this premarket notification, Hangzhou AGS Medtech Co., Ltd has demonstrated that proposed device is substantially equivalent to the predicate devices.

§ 876.4300 Endoscopic electrosurgical unit and accessories.

(a)
Identification. An endoscopic electrosurgical unit and accessories is a device used to perform electrosurgical procedures through an endoscope. This generic type of device includes the electrosurgical generator, patient plate, electric biopsy forceps, electrode, flexible snare, electrosurgical alarm system, electrosurgical power supply unit, electrical clamp, self-opening rigid snare, flexible suction coagulator electrode, patient return wristlet, contact jelly, adaptor to the cord for transurethral surgical instruments, the electric cord for transurethral surgical instruments, and the transurethral desiccator.(b)
Classification. Class II (performance standards).