K Number
K192308

Validate with FDA (Live)

Date Cleared
2020-12-21

(483 days)

Product Code
Regulation Number
890.3800
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The device is intended for medical purposes to provide mobility to persons restricted to a sitting position.

Device Description

The Wu's Electrical Scooter, WT-M4JKx, is an indoor-use electrical scooter that is battery operated. It has a base with 4 wheels, a seat and 2 armrests. The movement of the electrical scooter is controlled by the rider who uses speed control lever to control the direction and speed of an electrical scooter. The motor power is 470W. The device uses a PG S-Drive 70A electronic controller. The device is provided with an off-board battery charger (Input: AC100-240V, Output: DC 24V, 5 Amp). The maximum weight capacity of WT-M4JKx is 400 lbs (182 kg), and its maximum forward speed is 5.9 mph (9.5 km/h).

AI/ML Overview

The provided text describes the regulatory clearance for a medical device called "Wu's Electrical Scooter, WT-M4JKx". The document focuses on demonstrating substantial equivalence to a predicate device (Wu's Electrical Scooter, WT-M4Jr, K032489) through compliance with recognized standards and comparison of specifications.

Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are implicitly defined by the safety and performance standards the device was tested against (primarily RESNA WC-1 and ISO 7176 series), and the comparison to the predicate device's performance. The reported device performance is the outcome of these tests.

Acceptance Criterion (Standard / Feature)Reported Device Performance (Subject Device: WT-M4JKx)
Safety Standards
RESNA WC-1:2009 Wheelchairs - Volume 1 (General)Compliant
RESNA WC-2:2009 Wheelchairs - Volume 2 (Electrical Systems)Compliant
RESNA WC-2:2009 Section 21 (EMC)Compliant
ISO 7176-16:2012 (Resistance to ignition of postural support)Compliant
ISO 7176-25: 2013 (Batteries and Chargers)Compliant
ISO 7176-15:1996 (Dimensions and Mass)Compliant
Biocompatibility
Cytotoxicity (various parts)Passed
Sensitization (various parts)Passed
Skin Irritation tests (various parts)Passed
Delayed-type hypersensitivity (Epicutan test) (various parts)Passed
Performance (RESNA WC-1 Section 1:2009 - Static Stability)
Static stability downhill (Max.)26°
Static stability downhill (Min.)16°
Static stability uphill (Max.)16°
Static stability uphill (Min.)15°
Static stability sideways (Max.)9.5°
Performance (RESNA WC-1 Section 2:2009 - Dynamic Stability)
Dynamic stability downhill
Performance (RESNA WC-1 Section 3:2009 - Braking)
Minimum braking distance from max speed1.4m
Performance (RESNA WC-1 Section 4:2009 - Cruising Range)
Cruising range29km
Performance (RESNA WC-1 Section 5:2009 - Dimensions & Turn)
Overall length1210mm
Overall width580mm
Total mass85.5kg
Minimum turning radius1400mm
Minimum turn-around width1600mm
Performance (RESNA WC-1 Section 6:2009 - Speed)
Maximum speed forward9.5km (5.9 mph)
Performance (RESNA WC-1 Section 7:2009 - Seat Dimensions)
Seat plane angle (Max.)
Effective seat depth430mm
Effective seat width460mm
Seat surface height at front edge (Min.)570mm
Backrest angle10°
Backrest height360mm
Performance (RESNA WC-1 Section 10:2009 - Obstacle Climbing)
Obstacle climbing (Max.)50mm
Obstacle climbing (Min.)15mm
Performance (ISO 7176-15:1996 - Footrest & Armrest)
Footrest to seat distance (Max.)475mm
Footrest to seat distance (Min.)425mm
Leg to seat surface angle90°
Armrest to seat distance210mm
Front location of armrest structure320mm
Other Specifications
Maximum weight capacity400 lbs (182 kg)
Curb climbing ability1.96"/50mm
Operating environmentsIndoor use

2. Sample Size Used for the Test Set and Data Provenance

The document does not specify a "test set" in the context of a typical clinical study with human subjects. Instead, the device's performance is demonstrated through bench testing against recognized industry standards (RESNA, ISO) and biocompatibility testing of its components.

  • Sample Size: The sample size for these engineering and materials tests is not explicitly stated, but it is typically one or a small number of physical units of the device and its components, sufficient to perform the required measurements and evaluations per the standards.
  • Data Provenance: The tests were conducted according to international and national standards. The country of origin of the data is not directly mentioned beyond the manufacturer being in Taiwan. The nature of these tests is prospective in the sense that the new device was built and then rigorously tested to meet pre-defined engineering and safety benchmarks. There is no indication of retrospective data.

3. Number of Experts Used to Establish Ground Truth and Their Qualifications

This information is typically relevant for studies involving subjective human assessment (e.g., image interpretation). For this device, the "ground truth" is established by the objective measurements and adherence to the specified engineering and safety standards.

  • Number of Experts: Not applicable in the sense of establishing a subjective "ground truth." The experts involved would be the test engineers and laboratory personnel from the accredited testing facilities that performed the RESNA and ISO standard tests.
  • Qualifications of Experts: Not specified, but implied to be qualified and accredited by virtue of conducting tests according to international standards.

4. Adjudication Method for the Test Set

Not applicable. Adjudication methods (like 2+1, 3+1) are for resolving discrepancies in human expert opinions, which is not relevant for objective engineering and biocompatibility testing. The "adjudication" here is the pass/fail result against the criteria set by the standards.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No. An MRMC comparative effectiveness study is designed to evaluate differences in diagnostic accuracy or effectiveness among human readers, often with and without AI assistance, using a set of cases. This type of study is not relevant for a motorized scooter.

  • Effect Size: Not applicable.

6. Standalone Performance (Algorithm Only without Human-in-the-Loop Performance)

Yes, in the sense that the device itself (the "algorithm only") was tested against established performance and safety standards, independent of direct interaction with a human in a clinical efficacy setting. The performance metrics listed above (e.g., dynamic stability, braking distance, cruising range, maximum speed) are intrinsic to the device's design and function and were measured directly from the physical product. There is no "human-in-the-loop performance" analysis detailed for this device in the context of its regulatory clearance.

7. Type of Ground Truth Used

The ground truth used for this device is based on:

  • Established engineering and safety standards: RESNA WC-1, RESNA WC-2, ISO 7176 series. The performance of the device is measured against the requirements and test methods defined by these standards.
  • Objective material properties and biological response: For biocompatibility, the ground truth is established by the results of cytotoxicity, sensitization, skin irritation, and delayed-type hypersensitivity tests performed according to ISO 10993 standards.
  • Comparative equivalence to a predicate device: The regulatory submission establishes that the subject device's performance is "as safe, as effective, and performs as well as" a legally marketed predicate device (K032489 Wu's Tech WT-M4Jr). This comparison itself forms a type of ground truth for regulatory purposes.

8. Sample Size for the Training Set

Not applicable. "Training set" refers to data used to train machine learning models. This device is a physical medical device (electrical scooter), not an AI/ML-driven software product.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no training set for this type of device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 21, 2020

Wu's Tech Co., Ltd. % Jen Ke-Min Chinese-European Industrial Research Society No. 58, Fu-Chiun St Hsin-Chu City, Hsinchu 30067 Taiwan

Re: K192308

Trade/Device Name: Wu's Electrical Scooter Regulation Number: 21 CFR 890.3800 Regulation Name: Motorized Three-Wheeled Vehicle Regulatory Class: Class II Product Code: INI Dated: September 29, 2020 Received: October 5, 2020

Dear Jen Ke-Min:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance)and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Heather Dean, Ph.D. Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K192308

Device Name

Wu's Electrical Scooter. WT-M4JKx

Indications for Use (Describe)

The device is intended for medical purposes to provide mobility to persons restricted to a sitting position.

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows a logo for "WU'S Tech". The logo is inside of a green rounded rectangle. The text "WU'S" is in red, and the word "Tech" is in black.

五氏科技股份有限公司
WU'S TECH CO., LTD

5382191 lenhone/886-3-5382105 .225 Yuan Peir St. Hsinchu, Taiwan R -MAIL:wustech@wustech.com.tw

510(k) Summary of Safety and Effectiveness

(per 21 CFR 807.92)

Submitter's Name:Wu's Tech Co., Ltd.
No.225, Yuan Peir St., Hsinchu, Taiwan ROC 30093
Date summary prepared:December 11, 2020
Proprietary Name:Wu's Electrical Scooter, WT-M4JKx
Common or Usual Name:Electrical Scooter
Classification NameMotorized 3-Wheeled Vehicle
Class II, 21 CFR 890.3800
Product Code:INI
Contact Person:Dr. Jen, Ke-Min
Email:ceirs.jen@msa.hinet.net
TEL: +886-3-5208829, FAX: +886-3-5209783
Predicate DeviceWu's Tech Co., Ltd.
Wu's Electrical Scooter, WT-M4Jr
K032489

Indications for Use: ●

The device is intended for medical purposes to provide mobility to persons restricted to a sitting position.

● Device Description:

The Wu's Electrical Scooter, WT-M4JKx, is an indoor-use electrical scooter that is battery operated. It has a base with 4 wheels, a seat and 2 armrests. The movement of the electrical scooter is controlled by the rider who uses speed control lever to control the direction and speed of an electrical scooter. The motor power is 470W. The device uses a PG S-Drive 70A electronic controller. The device is provided with an off-board battery charger (Input: AC100-240V, Output: DC 24V, 5 Amp). The maximum weight capacity of WT-M4JKx is 400 lbs (182 kg), and its maximum forward speed is 5.9 mph (9.5 km/h).

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows a logo with the text "WiU's Tech" in red letters. The letters are stylized and have a rounded appearance. The logo is enclosed in a green rounded rectangle. The word "Tech" is smaller and located below the "WiU's" part of the logo.

統一編號/70711910 新竹市元培街225號 FAX /886-3-5382191 Telephone/886-3-5382105 No.225 Yuan Peir St. Hsinchu, Taiwan R.O.C. E-MAIL:wustech@wustech.com.tw

The following surfaces are recommended NOT to operate on:

  • � Sand surface
  • Wet or icy surface
  • � Road maintenance hole metal cover
  • � Avoid going up multiple steps.
  • � Avoid using escalators. Use the elevator.
  • � Too steep incline over 6 degrees.
  • Ground clearance to battery: 1.77"/45 mm
  • Curb climbing ability: 1.96"/50 mm
  • Compliant Testing Standards

Safety standards:

  • RESNA WC-1:2009 Wheelchairs - Volume 1 Requirements and test methods for wheelchairs (including scooter)

  • RESNA WC-2:2009 Wheelchairs - Volume 2 Additional requirements for wheelchairs (including scooter) with electrical systems

EMC standard:

  • RESNA WC-2:2009 Section 21: Requirements and test methods for electromagnetic compatibility of electrically powered wheelchairs and motorized scooters

Bench Testing:

  • ISO 7176-16:2012 - Wheelchairs - Part 16: Resistance to ignition of postural support devices.

  • く ISO 7176-25: 2013 - Wheelchairs - Part 25: Batteries and Chargers for Powered Wheelchairs
  • イ RESNA WC-1 Section 1:2009
  • く RESNA WC-1 Section 2:2009
  • く RESNA WC-1 Section 3:2009
  • イ RESNA WC-1 Section 4:2009
  • く RESNA WC-1 Section 5:2009
  • く RESNA WC-1 Section 6:2009
  • く RESNA WC-1 Section 7:2009
  • イ RESNA WC-1 Section 10:2009
  • イ RESNA WC-1 Section 11:2009
  • イ RESNA WC-1 Section 13:2009
  • イ ISO 7176-15:1996
  • イ ISO 7176-14:2008

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image is a logo for "WiLi'S Tech". The logo is in red and green, with the text "WiLi'S" in large, bold letters. The word "Tech" is in smaller letters below "WiLi'S". The logo has a rounded rectangular border around it. The logo appears to be for a technology company.

伍氏科技股份有限公司
WU'S TECH CO., LTD

統一編號/70711910 新竹市元培街225號 新刊サポリンセーム(3)-----------------------------------------------------------------------------------------------------------------------------------------------------------------No.225 Yuan Peir St. Hsinchu,Taiwan R.O.C. E-MAIL:wustech@wustech.com.tw

Biocompatibility evaluation of patient-contacting parts

Patient-contacting PartsMaterial's name & Color additiveDirect or Indirect contact / Contact classification & contact durationTests conductedRational for the tests conducted
1. Headrest Seatback Seat leather Safety beltPVC Vinyl & Color additive: Diphenylmeth-ane 4Direct-contact / Surface contacting less than 24 hours durationCytotoxicitySensitizationSkin Irritation testsISO 10993-1:2009
2. ArmrestPU Foam & Color additive: Diphenylmeth-ane 4Direct-contact / Surface contacting less than 24 hours durationCytotoxicitySensitizationSkin Irritation testsISO 10993-1:2009
3. Horn buttonDow Corning RBB-6671-70 + RBB-6671-70+ XE20-523-5U + Colourant Techi Black 1801 & Color additive: Solvent YellowDirect-contact / Surface contacting less than 24 hours durationCytotoxicity + skin Irritation & delayed-type hypersensitivity (Epicutan test) TestsISO 10993-5:2009ISO 10993-10:2013
4. Speed dialMakrolon 6557 (PC material ) & Color additive: Carbon blackDirect-contact / Surface contacting less than 24 hours durationCytotoxicity testISO 10993-5:2009
5. REM060/1 10 bottom caseMakroblend EL700 (PC+PET materials) & Color additive: Carbon blackDirect-contact / Surface contacting less than 24 hours durationCytotoxicity + Skin irritation & delayed-type hypersensitivity (Epicutan test) TestsISO 10993-5:2009ISO 10993-10:2013
6. REM050/1 00 (Full joystick control module)Makroblend EL 700 + Makrolon 6557 & Color addtitive: Carbon blackDirect contact / Surface contacting less than 24 hours durationSkin irritation and delayed-type hypersensitivity Tests (Epicutan test)ISO 10993-10:2013
7. FootrestPolylac ABS & Color additive: Carbon blackIndirect-contact / Surface contacting less than 24 hours durationMSDS*Polylac ABS material is a common and mature material for wheelchair's part, and was adopted by many wheelchair manufacturers in the current market.*There are no customer claims with adverse conditions for more than thousands of units sold in the market according to our sale experience.
8. Delta handle barPVC Vinyl & Color additive: Polyol, PigmentDirect-contact / Surface contacting less than 24 hours durationCytotoxicitySkin irritationSensitizationISO 10993-5:2009ISO 10993-10:2010
9. Speedleverε-Caprolactam resinDirect-contact /Surfacecontacting lessthan 24 hoursdurationCytotoxicitySkin irritationSensitizationISO 10993-5:2009ISO 10993-10:2010
10. LightswitchPolyamide -6 &Color additive:Solvent 37Direct-contact /Surfacecontacting lessthan 24 hoursdurationMSDS*Low frequency ofhuman contact and nocustomer claims withadverse conditionsaccording to previous saleexperience and ourexpertise
11.IndicatorswitchPhenolic & Color additive: Carbon blackDirect-contact /Surfacecontacting lessthan 24 hoursdurationMSDS*Low frequency ofhuman contact and nocustomer claims withadverse conditionsaccording to previous saleexperience and ourexpertise
12.Emergency switchPhenolic & Color additive: Solvent red 135Direct-contact /Surfacecontacting lessthan 24 hoursdurationMSDS*Low frequency ofhuman contact and nocustomer claims withadverse conditionsaccording to previous saleexperience and ourexpertise
13. Deltahandlebar knobPolylac & Color additive: Carbon blackDirect-contact /Surfacecontacting lessthan 24 hoursdurationMSDS*Low frequency ofhuman contact and nocustomer claims withadverse conditionsaccording to previous saleexperience and ourexpertise
14. SeatremoveleverPolyvinyl Chloride & Color additive:Polyol ,PigmentDirect-contact /Surfacecontacting lessthan 24 hoursdurationMSDS*Low frequency ofhuman contact and nocustomer claims withadverse conditionsaccording to previous saleexperience and ourexpertise
15. Seatslide leverε-CaprolactamresinDirect-contact /Surfacecontacting lessthan 24 hoursdurationMSDS*Low frequency ofhuman contact and nocustomer claims withadverse conditionsaccording to previous saleexperience and ourexpertise
16.Freewheel leverPolyvinyl chloride & Color additive:Solvent YellowDirect-contact /Surfacecontacting lessthan 24 hoursdurationMSDS*Low frequency ofhuman contact and nocustomer claims withadverse conditionsaccording to previous saleexperience and ourexpertise
17.Chargingsocketcoverε-CaprolactamresinDirect-contact /Surfacecontacting lessthan 24 hoursdurationMSDS*Low frequency ofhuman contact and nocustomer claims withadverse conditionsaccording to previous saleexperience and ourexpertise

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows a logo with the text "WU'S Tech" in red letters. The letters "WU" are connected, and there are two green dots above the "U". The text "Tech" is smaller and located below the "S". The logo is surrounded by a green border with rounded corners.

国限公司 O.. ILITD

新竹市元培街225號

統一編號/70711910
FAX :886-2-522214

新竹市元培街225號 No.225 Yuan Peir St. Hsinchu,Taiwan R.O.C. E-MAIL:wustech@wustech.com.tw

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows a logo with the text "WU'S Tech" in red with a green border. The "WU'S" part of the logo is larger and more prominent than the "Tech" part. The logo has a rounded rectangular shape.

伍氏科技股份有限公司
WU'S TECH CO., LTD

統一編號/70711910 新竹市元培街225號 FAX /886-3-5382191 Telephone/886-3-5382105 No.225 Yuan Peir St. Hsinchu, Taiwan R.O.C. E-MAIL:wustech@wustech.com.tw

Specifications comparison table ●

ModelSpecificationPredicate DeviceSubject Device
ManufacturerWu's Tech Co., Ltd.Same
Proprietary name, modelWu's Electrical Scooter,WT-M4JrWu's Electrical Scooter,WT-M4JKx
510(k) numberK032489TBA
Device classification nameMotorized 3-wheeled vehicleMotorized 3-wheeled vehicle
Classification regulationsClass II, 21 CFR 890.3800Class II, 21 CFR 890.3800
Product codeINIINI
Same or Similarity
Indications for UseThe device is intended formedical purposes to providemobility to persons restricted toa sitting position.The device is intended for medicalpurposes to provide mobility topersons restricted to a sittingposition.
Number of Wheels44
Movement Control MethodSpeed control leverSpeed control lever
Driving SystemDirect drive rear wheel(with differential gear)Direct drive rear wheels(with differential gear)
Brake SystemElectromagnetic brakeElectromagnetic brake
Charger typeOff-board chargerInput: AC 100-240 V,Output: DC 24V, 5AOff-board chargerInput AC: 100-240V,Output DC 24V, 5 Amp
Electronic ControllerPG S-Drive 70APG S-Drive 70A
Operating environmentsIndoor useIndoor use
Differences
Front Wheel Size2.5 - 4" solid tire10.2"x3.3" solid tire
Rear Drive Wheel Size2.5 - 4" solid tire10.2" × 3.3" solid tire
Dimension (LxW x D inch)46 x 22 x 35.4 inches40.6 × 23.8 × 49 inches
(LxW x D mm)1170 × 560 × 900 mm1210 × 580 × 1040 mm
Motor Output350W × 24V × 1 pcs470W x 24V x 1 pcs
Weight with Battery94.7 lbs. /43 kg188.1 lbs. / 85.5 kg
Cruising Range20 miles /32 km18 miles / 29 km
Battery type34 Ah × 12V × 2 pcs36 Ah × 12V × 2 pcs
Slope Grade Ability10
Max Speed Forward4 mph (6.4 km/h)5.9 mph (9.5 km/h)
Max Loading250 lbs. (113.5 kg )400 lbs. (182 kg)
Folding mechanismNo folding functionFoldable backrest and armrests

Specification comparison discussions ●

Both devices have the same indications for use, same technological designs, same number of wheels, same speed control lever, same driving system, and same brake system,

In the below, we discuss the differences between them,

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image is a logo for a company called "Wi's Tech". The logo is in a green rounded rectangle. The text "Wi's" is in red, with two green dots above the "i" and the "s". The word "Tech" is in black and is located below the "s".

  • √ Wheel size: The wheel sizes for the predicate and subject devices are different. The predicate device's tires are 2.5'' x 4" solid tires and the subject device's tires are 10.2"x3.3" solid tire. The wheels of predicate device are smaller than those of the subject device. Larger wheel will give more comfort to the users, and will not raise any safety and effectiveness concerns due to the different wheel size.
  • Dimensions: The dimensions of the two devices are different. The larger depth will > give more support to the legs and will bring more comfort to user's ride, and since the subject device has passed the RESNA WC-1 and RESNA WC-2 standards with respect to safety and performance aspects, and this difference does not raise any new safety and effectiveness concerns for the subject device.
  • Weight of wheelchair: The weight of subject device is 188.1 lbs. (85.5 kg) and it is heavier than the predicate device. This difference will make a lower ratio of the wheelchair weight/motor power output for the subject device. The ratio of the weight/power for the subject device is 188.1 1bs / 470 W = 0.400 lbs/W and the predicate device is 94.7 lbs/ 350 W = 0.270 lbs./W. The subject device will use one ioule energy per second to push 0.400 pounds of wheelchair weight, and the predicate device will use one joule energy per second to push 0.270 pounds of wheelchair weight. That is to say, the subject device consumes more battery's electrical energy than the predicate device. There are no new safety and effectiveness concerns raised for the subject device with respect to the larger weights.

  • ✓ Motor output: The motor output for the subject device is 470W, it is larger than 350W of the predicate device. Since the volume and weight of the subject device is larger than the predicate device, the motor output of the subject device should be designed to be larger than the predicate device. The difference will not raise any new safety and effectiveness concerns for the subject device.
  • V Cruising range: In general, the cruising range depends on the wheelchair weight, battery charging condition, motors output, and cruising surface conditions. The cruising range of the subject device is 18 miles and that of the predicate device is about 20 miles. The difference between subject device and predicate device is not large. As the user knew the cruising range of the subject device is 18 miles from the user manual, they will drive the scooter in consideration of 18 miles range, and avoid driving exceeding this cruising range for each travel. There are no new safety

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image is a logo for a company called "WU'S Tech". The logo is red text with a green border. The text is in a bold, sans-serif font. The word "WU'S" is on the top line, and the word "Tech" is on the bottom line.

新竹市元培街225號 統一編號/70711910 Telephone/886-3-5382105 FAX /886-3-5382191 No.225 Yuan Peir St. Hsinchu, Taiwan R.O.C. E-MAIL:wustech@wustech.com.tw

and effectiveness concerns raised due to this smaller cruising range.

  • ✔ Battery type: The battery capacity of the subject device is 36 Ah and the predicate device is 34 Ah. A larger battery capacity means this battery can drive a larger range or last for a longer time. Also, the batteries of the subject device have passed the testing of ISO 7176-25:2013. These facts demonstrate there are no any new safety and effectiveness concerns raised due to using a larger battery.
  • Slope grade ability: The slope grade ability is 10 degrees for predicate device and 6 degrees for subject device. The relevant statement of 6 degrees for the slope grade ability is placed in the user manual. If the user operates the electrical scooter according to the instructions for use, any safety and effectiveness concerns are not raised due to the smaller slope grade ability.

  • The Max Speed Forward for the subject device is 5.9 mph, which is larger than 4 mph of the predicate device. As we know the maximum safety speed for the motorized scooter is 6 mph, there is no safety and effectiveness concerns for the difference of maximum forward speed.

  • The maximum loading of the subject device is 400 lbs. (182 kg) and it is 250 lbs (113.5 kg) for the predicate device. The maximum loading of the subject device is much larger than the predicate device, which will bring more comfort and safety to the users. There is no any new safety and effectiveness concerns due to the difference of the maximum loading for the subject device.

  • The backrest and armrests of the subject device are foldable and the predicate device has no any folding function. These more functions bring more conveniences of storage or transportation than the predicate device. There are no any safety and effectiveness concerns raised for the subject device due to the differences of folding functions.

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows a logo with the text "WU'S Tech" in red letters with green accents. The letters are stylized and have a rounded appearance. The word "Tech" is smaller and located below the "WU'S" part of the logo. The entire logo is enclosed in a rounded green border.

in

統一編號/70711910 新竹市元培街225號 新刊サポリントロッコース-5382105 FAX /886-3-5382191 No.225 Yuan Peir St. Hsinchu,Taiwan R.O.C. E-MAIL:wustech@wustech.com.tw

● RESNA standard tests comparison table

Predicate Device:Subject Device:
Wu's Electrical Scooter,Wu's Electrical Scooter,
SpecificationWT-M4JrWT-M4JKx
K032489K192308
Test StandardResultTest StandardResult
Static stability downhill(Max.)RESNA WC/Vol.1Section 1:1998NARESNA WC/Vol.1Section 1:200926°
Static stability downhill(Min.)RESNA WC/Vol.1Section 1:199812°RESNA WC/Vol.1Section 1:200916°
Static stability uphill(Max.)RESNA WC/Vol.1Section 1:1998NARESNA WC/Vol.1Section 1:200916°
Static stability uphill(Min.)RESNA WC/Vol.1Section 1:199811°RESNA WC/Vol.1Section 1:200915°
Static stability sideways(Max.)RESNA WC/Vol.1Section 1:1998RESNA WC/Vol.1Section 1:20099.5°
Dynamic stabilitydownhillRESNA WC/Vol.1Section 2:199810°RESNA WC/Vol.1Section 2:20096 °
Minimum braking distancefrom max speedRESNA WC/Vol.1Section 3:19981.15mRESNA WC/Vol.1Section 3:20091.4m
Cruising rangeRESNA WC/Vol.1Section 4:199832kmRESNA WC/Vol.1Section 4:200929km
Overall lengthRESNA WC/Vol.1Section 5:19981170mmRESNA WC/Vol.1Section 5:20091210mm
Overall widthRESNA WC/Vol.1Section 5:1998560mmRESNA WC/Vol.1Section 5:2009580mm
Folded lengthRESNA WC/Vol.1Section 5:1998NARESNA WC/Vol.1Section 5:20091210mm
Folded widthRESNA WC/Vol.1Section 5:1998NARESNA WC/Vol.1Section 5:2009580mm
Total massRESNA WC/Vol.1Section 5:199843kgRESNA WC/Vol.1Section 5:200985.5kg
Horizontal location ofaxleRESNA WC/Vol.1Section 5:1998NARESNA WC/Vol.1Section 5:2009750mm
Minimum turning radiusRESNA WC/Vol.1Section 5:19981100mmRESNA WC/Vol.1Section 5:20091400mm
Minimum turn-aroundwidthRESNA WC/Vol.1Section 5:19981200mmRESNA WC/Vol.1Section 5:20091600mm
Maximum speed forwardRESNA WC/Vol.1Section 6:19986.4kmRESNA WC/Vol.1Section 6:20099.5km
Seat plane angle (Max.)RESNA WC/Vol.1Section 7:1998RESNA WC/Vol.1Section 7:2009
Effective seat depthRESNA WC/Vol.1Section 7:1998420mmRESNA WC/Vol.1Section 7:2009430mm
Effective seat widthRESNA WC/Vol.1Section 7:1998430mmRESNA WC/Vol.1Section 7:2009460mm
Seat surface height atfront edge (Max.)RESNA WC/Vol.1Section 7:1998NARESNA WC/Vol.1Section 7:2009620mm
Seat surface height atfront edge (Min.)RESNA WC/Vol.1Section 7:1998500mmRESNA WC/Vol.1Section 7:2009570mm
Backrest angleRESNA WC/Vol.1Section 7:199810°RESNA WC/Vol.1Section 7:200910°
Backrest heightRESNA WC/Vol.1Section 7:1998360mmRESNA WC/Vol.1Section 7:2009360mm
Obstacle climbing (Max.)RESNA WC/Vol.1Section 10:199850mmRESNA WC/Vol.1Section 10:200950mm
Obstacle climbing (Min.)RESNA WC/Vol.1Section 10:199820mmRESNA WC/Vol.1Section 10:200915mm
Footrest to seat distance(Max.)ISO7176-15:1996460mmISO7176-15:1996475mm
Footrest to seat distance(Min.)ISO7176-15:1996410mmISO7176-15:1996425mm
Leg to seat surface angleISO7176-15:199690°ISO7176-15:199690°
Armrest to seat distanceISO7176-15:1996210mmISO7176-15:1996210mm
Front location of armreststructureISO7176-15:1996305mmISO7176-15:1996320mm

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image shows a logo with the text "WiU'S Tech" in a stylized font. The letters are red with a green outline, and there are two green dots above the "i" and "U". The word "Tech" is written in a smaller font below the main text. The entire logo is enclosed in a rounded green rectangle.

五氏科技股份有限公司
WU'S TECH CO., LTD

統一編號/70711910 新竹市元培街225號 Telephone/886-3-5382105 FAX /886-3-5382191 No.225 Yuan Peir St. Hsinchu, Taiwan R.O.C. E-MAIL:wustech@wustech.com.tw

● RESNA standards test comparison discussions

1) RESNA WC-1 Section 1

The standard used for testing the static stability of the subject device is RESNA WC-1 Section 1:2009, but the standard used for the predicate device is RESNA WC-1 Section 1:1998. The subject device used updated standard.

The testing results show that the static stability downhill for the subject device is 26 degrees (Max.) / 16 degrees (Min.), the static stability uphill is 16 degrees (Max.) / 15 degrees (Min.) and the static stability sideways is 9.5 degrees. All of the static testing results for the subject device are larger than the predicate device. The differences of the angles show that the subject device has more static stability in three ways than the predicate device. So there are no any new safety and effectiveness concerns raised for the subject device due to the static stability differences.

{12}------------------------------------------------

Image /page/12/Picture/0 description: The image shows a logo with the text "Wi's Tech". The text is in red and is surrounded by a green border. The word "Tech" is smaller than the rest of the text and is located at the bottom right of the logo. The logo is simple and clean, with a focus on the text.

新竹市元培街225號 統一編號/70711910 FAX /886-3-5382191 Telephone/886-3-5382105 No.225 Yuan Peir St. Hsinchu, Taiwan R.O.C. E-MAIL:wustech@wustech.com.tw

2) RESNA WC-1 Section 2

The standard used for testing the dynamic stability of the subject device is RESNA WC-1 Section 2:2009, but the standard used for the predicate device is RESNA WC-1 Section 2:1998. The subject device used updated standard.

The dynamic stability downhill is 6 degrees for the subject device and is 10 degrees for the predicate device. AS we knew the Americans with Disabilities Act Accessibility (ADAAG) Section 4.7.2 Slope has regulated that the angle of the slopes that may be encountered in public place and accessible environments that comply with the ADAAG will be required not to have slopes greater than 1:12 (4.8 degrees). The risk level of tip over due to the 4-degree less dynamic stability has been mitigated to an acceptable level by the slope limitation requirements of ADAAG and the disclosure of the dynamic stability information in the user manual. There are no safety and effectiveness concerns raised by the 4-degree less dynamic stability difference for the subject device.

RESNA WC-1 Section 3 3)

The standard used for testing minimum braking distance of the subject device is RESNA WC-1 Section 3:2009, but the standard used for the predicate device is RESNA WC-1 Section 3:1998. The subject device used updated standard.

The minimum braking distance from maximum speed is 1.40 m for subject device and is 1.15 m for predicate device. The subject device needs more 0.25 m distance to stop the wheelchair from the maximum speed than the predicate device. As we knew, the max forward speed for the subject device is 5.9 mph, and it is 4 mph for the predicate device. The higher-speed wheelchair normally needs more braking distance to stop the wheelchair. Since the minimum braking distance information was disclosed in the user manual, the risk level of larger braking distance was reduced to an acceptable level. There are no any safety and effectiveness concerns raised for the subject device due to this braking difference of 0.25 meters.

4) RESNA WC-1 Section 4

The standard used for testing the cruising range of the subject device is RESNA WC-1 Section 4:2009, but the standard used for the predicate device is RESNA WC-1 Section 4:1998. The subject device used updated standard.

{13}------------------------------------------------

Image /page/13/Picture/0 description: The image shows a logo with the text "Wi's Tech" in red letters. The letters "Wi's" are larger than the word "Tech", and there are two green dots above the "i" and the "s". The logo is surrounded by a green border.

The cruising range for the subject device is 29 km and it is 32 km for the predicate device. If the user got used to the 29 km cruising range and already knew when the wheelchair needed to be charged, the user would not meet the low-battery situation, which resulted to abnormal driving situation. Thus, based on the above, there are no any new safety and effectiveness concerns raised by this shorter cruising range for the subject device.

ર) RESNA WC-1 Section 5

The standard used for testing the overall dimensions of the subject device is RESNA WC-1 Section 5:2009, but the standard used for the predicate device is RESNA WC-1 Section 5:1998. The subject device used updated standard.

The overall length / width is tested to be 1210 / 580 mm for the subject device and it is 1170 / 560 mm for predicate device. These minor differences mean the subject device is bigger than the predicate device, thus not related to the safety aspects.

The total mass for the subject device is 85.5 kg and it is 43 kg for the predicate device, and the difference comes from bigger sizes of the subject device. The heavier wheelchair will lead to more comfortable feeling and less cruising range. There are no any new safety and effectiveness concerns raised by this difference for the subject device.

The minimum turning radius and minimum turn-around width are 1400 / 1600 mm for the subject device and they are 1100 / 1200 mm for the predicate device. The larger minimum turning radius and minimum turn-around width for the subject device mean that it needs considering more space to make a turn or a turn-around. Since this information was disclosed in the user manual to the users, the risk level of shortage of space was reduced to an acceptable level. There are no any new safety and effectiveness concerns raised by these differences for the subject device.

6) RESNA WC-1 Section 6

The standard used for testing the maximum forward speed of the subject device is RESNA WC-1 Section 6:2009, but the standard used for the predicate device is RESNA WC-1 Section 6:1998. The subject device used updated standard.

{14}------------------------------------------------

Image /page/14/Picture/0 description: The image shows a logo with the text "WU'S Tech" in a stylized font. The letters "WU'S" are large and red, with green circles above the "U" and "S". The word "Tech" is smaller and located below the "S" in a different font. The entire logo is enclosed in a green rounded rectangle.

氏科技股份有限公司
U'S TECH CO., LTD

統一編號/70711910 新竹市元培街225號 Telephone/886-3-5382105 FAX /886-3-5382191 No.225 Yuan Peir St. Hsinchu, Taiwan R.O.C. E-MAIL:wustech@wustech.com.tw

The forward speed for the subject device is 5.9 mph and the predicate device is 4 The higher forward speed of the subject device requires more braking mph. distance than the predicate device. The maximum forward speed information was disclosed in the user manual to the users, and the risk level of higher speed was reduced to an acceptable level. There are no any new safety and effectiveness concerns raised by the difference of higher speed for the subject device.

7) RESNA WC-1 Section 7

The standard used for testing the angle and the dimensions of the subject device is RESNA WC-1 Section 7:2009, but the standard used for the predicate device is RESNA WC-1 Section 7:1998. The subject device used updated standard.

The seat plane angle is tested to be 5 degrees for the subject device and it is 0 degree for the predicate device. This difference will result in backward inclination of the user or less happening chance of forward tip-over for the subject device and no any new safety and effectiveness concerns were raised due to the seat plane angle difference for the subject device.

The effective seat depth / effective seat width are tested to be 430 / 460 mm for the subject device and 420 / 430 mm for the predicate device. These differences can result in different comfort feeling during driving period, and they are not related to the safety aspects.

The minimum / maximum seat surface heights at front edge is 570 / 620 mm for the subject device and 500 / 500 mm for the predicate device. The seat surface height of the predicate device is fixed. It means the minimum seat height for the subject device is larger than the predicate device by 70 mm. The seat surface height of the subject device can be adjusted by 50 mm according to the comfortable feeling of users, and it is not related to the safety aspects. There are no any new safety and effectiveness concerns raised by the difference of larger seat surface height for the subject device.

8) RESNA WC-1 Section 10

The standard used for testing the climbing obstacle of the subject device is RESNA WC-1 Section 10:2009, but the standard used for the predicate device is RESNA WC-1 Section 10:1998. The subject device used updated standard.

{15}------------------------------------------------

Image /page/15/Picture/0 description: The image shows a logo with the text "WiU'S Tech" in red letters with a green outline. The letters are stylized and have a rounded appearance. The word "Tech" is smaller and located below the "S" in "WiU'S".

科技股份有限公司
TECH CO., LTD
新竹市
Teleph
No.225

統一編號/70711910 新竹市元培街225號 Telephone/886-3-5382105 FAX /886-3-5382191 No.225 Yuan Peir St. Hsinchu, Taiwan R.O.C. E-MAIL:wustech@wustech.com.tw

The maximum / minimum obstacles climbing are 50 / 15 mm for the subject device and 50 / 20 mm for the predicate device. The maximum obstacles climbing are the same and the minimum obstacles are different by 5 mm. The obstacles climbing were disclosed in the user manual and the users will notice the obstacles climbing range, 50 mm to 15 mm, and they are instructed to follow the instructions for use to operate the wheelchair. Thus, there are no any new safety and effectiveness concerns raised due to the minor difference of the climbing obstacle for the subject device.

9) ISO 7176-15

The standard used for testing the relevant distances and angle of the subject device is ISO 7176-15:1996, and the standard used for the predicate device is ISO 7176-15:1996. Both devices used the same standard.

Footrest to seat distances (Max./ Min.) were tested to be 475 /425 mm for subject device and 460 / 410 mm for predicate device. These distances for both devices can be adjusted 50 mm by changing the vertical position of the footrest between the minimum position and maximum position. The users of two devices can adjust the positions of the footrest according to the comfort feeling at various positions. These differences are not related to any safety aspects. There are no any new safety and effectiveness concerns raised due to these differences for the subject device.

Front location of armrest structure is 320 mm for the subject device and is 305 mm for the predicate device. The armrest is intended for use for resting arms, and people's arms have minor differences of lengths, so the front location of armrest structure for the larger overall design of the subject device is 15 mm larger than the predicate device accordingly. It is not related to any safety aspect. There are no any new safety and effectiveness concerns raised by the difference of larger front location of armrest structure for the subject device.

● CONCLUSIONS

The conclusions drawn from the non-clinical tests demonstrate that the subject device is as safe, as effective, and performs as well as the legally marketed predicate device, K032489 Wu's Tech WT-M4Jr. identified in the submission. Thus, the subject device is substantially equivalent to the predicate device.

§ 890.3800 Motorized three-wheeled vehicle.

(a)
Identification. A motorized three-wheeled vehicle is a gasoline-fueled or battery-powered device intended for medical purposes that is used for outside transportation by disabled persons.(b)
Classification. Class II (performance standards).