K Number
K192264

Validate with FDA (Live)

Manufacturer
Date Cleared
2019-11-08

(79 days)

Product Code
Regulation Number
890.5850
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

These Electrical Muscle Stimulation units are indicated for the improvement of muscle tone and firmness, for strengthening muscles in arms, abdomen, thighs and buttocks areas. Not intended for use in any therapy or for the treatment of any medical conditions or diseases.

Device Description

EMS, Electrical Muscle Stimulator, which improves, tones, firms & strengthens muscle and relaxes stiff muscle through the skin. It is recognized as a clinically proven, effective, non-medication method of training muscle from certain causes. It manages muscle strengthen, toning and firming. It is also free from side effects when used properly, and can also be used as a simple means of self-training.

HIVOX Spopad EMS SP series, SP-911 and SP-921 are the proposed devices for the 510(k) submission. These Electrical Muscle Stimulation units are indicated for the improvement of muscle tone and firmness, for strengthening muscles in arms, abdomen, thighs, and buttocks areas. Not intended for use in therapy or for the treatment of any medical conditions or diseases.

SP-911 and SP-921 are 1-channel battery-operated-user-friendly muscle stimulation system specifically designed to exercise the muscles. Each device comprises namely an electronic stimulator module which generates the required stimulation signals. SP-911 comprises 2 electrodes, which connects the signals from the stimulator to the skin. SP-921 comprises 4 electrodes, which connect the signals from the stimulator to the skin. Power is supplied from one battery, CR2032, located in a compartment protected by a removable battery cover. The user cannot access the wiring or connectors.

AI/ML Overview

The provided text is a 510(k) Summary for the HIVOX Spopad EMS SP-911 and SP-921. It details the safety and performance testing conducted to demonstrate substantial equivalence to a predicate device, rather than proving the device meets specific acceptance criteria through a clinical study for a new, complex algorithm or AI.

The document explicitly states: "No clinical test data was used to support the decision of substantial equivalence." (Section 5.9). This means there was no clinical study conducted where the device's performance was measured against established acceptance criteria using human data, adjudicated ground truth, or expert readers.

Instead, the submission relies on non-clinical testing to demonstrate performance and safety, specifically for a "Powered Muscle Stimulator." The acceptance criteria referenced are for electrical and physical performance characteristics of the device itself, rather than diagnostic accuracy or clinical effectiveness in the context of AI.

Therefore, many of the requested details regarding sample size, data provenance, expert qualifications, ground truth, MRMC studies, or standalone algorithm performance are not applicable to this type of device clearance and the information provided.

I can, however, extract the acceptance criteria for the non-clinical tests that were performed, as implied by the statement "All the test result demonstrate HIVOX Spopad EMS SP-911 and SP-921 meet the requirement of its pre-defined acceptance criteria and intended use". The document also implicitly defines acceptance criteria by noting when differences do not adversely impact safety and effectiveness, and by comparing the subject device's performance against the predicate and relevant standards.

Here's a breakdown of the relevant information provided:

1. A table of acceptance criteria and the reported device performance:

The document doesn't present a clean "acceptance criteria" table with "reported performance" directly. Instead, it compares the subject device's performance characteristics to those of its predicate device, and implicitly (though not explicitly numerically) against relevant international standards. The determination of "Identical" implies meeting the same performance characteristics as the cleared predicate, which serves as the de facto acceptance benchmark for substantial equivalence in many parameters. For parameters that are "Different but does not adversely impact safety and effectiveness," the standard or a justified value serves as the acceptance criteria.

Feature / Acceptance Criteria (Implied)Subject Device Performance (SP-911 / SP-921)Predicate Device Performance (SP-910 / SP-920)Substantial Equivalence Determination / Justification of Acceptance
General
Intended UseIdentical to PredicateIndicated for the improvement of muscle tone and firmness, for strengthening muscles in arms, abdomen, thighs, and buttocks areas. Not intended for use in any therapy or for the treatment of any medical conditions or diseases.Identical.
Technology/Principle of OperationElectrical Muscle StimulationElectrical Muscle StimulationIdentical.
Compliance with Voluntary StandardsIEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-2-10IEC 60601-1, IEC 60601-1-2, IEC 60601-2-10Identical (SP-911/921 also comply with IEC 60601-1-11, which extends compliance).
Compliance with 21 CFR 898YesYesIdentical.
Housing material and constructionSiliconeSiliconeIdentical.
Electrical / Output Characteristics
Output waveformSymmetrical biphasicSymmetrical biphasicIdentical.
ShapeRectangularRectangularIdentical.
Duration of primary (depolarizing) phase00Identical.
Pulse duration (μs)400400Identical.
Max Output Voltage (V, ±10%) at 500 ΩSP-911: 52 / SP-921: 58.4SP-910: 52 / SP-920: 58.4Identical.
Max Output Voltage (V, ±10%) at 2k ΩSP-911: 102 / SP-921: 106SP-910: 102 / SP-920: 106Identical.
Max Output Voltage (V, ±10%) at 10k ΩSP-911: 150 / SP-921: 154SP-910: 150 / SP-920: 154Identical.
Max Output Current (mA, ±10%) at 500 ΩSP-911: 104 / SP-921: 117SP-910: 104 / SP-920: 117Identical.
Max Output Current (mA, ±10%) at 2k ΩSP-911: 51 / SP-921: 53SP-910: 51 / SP-920: 53Identical.
Max Output Current (mA, ±10%) at 10k ΩSP-911: 15 / SP-921: 15.4SP-910: 15 / SP-920: 15.4Identical.
Frequency (Hz)SP-911: 3/4/5 / SP-921: 2/4/25SP-910: 3/4/5 / SP-920: 2/4/25Identical.
Net charge per pulse at 500 Ω (μC)SP-911: 0.416 / SP-921: 0.468SP-910: 0.416 / SP-920: 0.468Identical.
Maximum charge at 500 Ω (μC)SP-911: 41.6 / SP-921: 46.8SP-910: 41.6 / SP-920: 46.8Identical.
Maximum current density at 500 Ω (mA/cm²)SP-911: 1.672 / SP-921: 1.057SP-910: 1.187 / SP-920: 1.057SP-911: Different (higher) but does not adversely impact safety and effectiveness. Justified against 0.25 W/cm² acceptance criterion for power density. SP-921: Identical.
Maximum average power density at 500 Ω (W/cm²)SP-911: 0.0869 / SP-921: 0.0617SP-910: 0.0617 / SP-920: 0.0617SP-911: Different (higher) but does not adversely impact safety and effectiveness. Still below the acceptance criterion of 0.25 W/cm² from the Class II Special Control Guidance Document. SP-921: Identical.
Safety Features
Shelf-Life testingMet requirementsN/A (implied by predicate clearance)Met pre-defined acceptance criteria.
Performance verificationMet requirementsN/A (implied by predicate clearance)Met pre-defined acceptance criteria.
Usability validationMet requirementsN/A (implied by predicate clearance)Met pre-defined acceptance criteria.
Method of Line Current IsolationBattery supplyBattery supplyIdentical.
Patient Leakage Current Normal Condition (μΑ)2.02.0Identical.
Patient Leakage Current Single Fault Condition (μΑ)2.12.1Identical.
Method of channel isolationSingle channelSingle channelIdentical.
Average DC current through electrodes when device is on but no pulses are being applied (μΑ)00Identical.
Number of output modes11Identical.
Regulated current or regulated voltage?VoltageVoltageIdentical.
Software / Firmware / Microprocessor control?YesYesIdentical.
Automatic overload trip?NoNoIdentical.
Automatic no-load trip?NoNoIdentical.
Automatic shut-off?YesYesIdentical.
User overrides control?YesYesIdentical.
Indicator display – On / Off StatusNoNoIdentical.
Indicator display – Low battery?NoNoIdentical.
Indicator display – Voltage / Current levelNoNoIdentical.
Timer range (minutes)2020Identical.

2. Sample size used for the test set and the data provenance:

  • Sample Size: Not applicable. The testing was non-clinical (shelf-life, performance verification, usability validation) and involves device samples, not patient data or clinical subjects. The document does not specify the number of devices tested for these non-clinical assessments.
  • Data Provenance: The testing was conducted by HIVOX Biotek Inc., based in New Taipei City, Taiwan, R.O.C. (as per "Submitter Address" in Section 5.3). The data is from the manufacturer's internal testing as part of their 510(k) submission. It is not retrospective or prospective clinical data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Not applicable. As this was a non-clinical submission, there was no "ground truth" derived from expert consensus on medical images or clinical outcomes. The "ground truth" for the device's electrical and physical parameters would be defined by engineering specifications and international standards, measured using appropriate test equipment and procedures.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

  • Not applicable for the reasons stated above.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • Not applicable. This device is a powered muscle stimulator, not an AI-enabled diagnostic or assistive tool. No MRMC study was performed.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

  • Not applicable. This is a hardware device, not a standalone algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

  • For the non-clinical tests, the "ground truth" implicitly refers to engineering specifications, regulatory requirements (e.g., 21 CFR 898), and international standards (e.g., IEC 60601 series). Performance verification would involve measuring parameters against these defined acceptable ranges. Usability validation typically involves testing by representative users against predefined usability criteria.

8. The sample size for the training set:

  • Not applicable. This device does not involve machine learning or AI that requires a "training set."

9. How the ground truth for the training set was established:

  • Not applicable for the reasons stated above.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

November 8, 2019

Hivox Biotek Inc. Marx Lee Regulatory Affairs Representative 5F., No. 123, Xinde Rd. Sanchong Dist. New Taipei City, TW 24158

Re: K192264

Trade/Device Name: HIVOX Spopad EMS SP-911, SP-921 Regulation Number: 21 CFR 890.5850 Regulation Name: Powered Muscle Stimulator Regulatory Class: Class II Product Code: NGX Dated: October 7, 2019 Received: October 9, 2019

Dear Marx Lee:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

for: Vivek Pinto, Ph.D. Acting Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K192264

Device Name HIVOX Spopad EMS SP-911, SP-921

Indications for Use (Describe)

These Electrical Muscle Stimulation units are indicated for the improvement of muscle tone and firmness, for strengthening muscles in arms, abdomen, thighs and buttocks areas. Not intended for use in any therapy or for the treatment of any medical conditions or diseases.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Section 6

510(k) Summary

{4}------------------------------------------------

510(k) SUMMARY

5.1 Type of SubmissionSpecial
5.2 Date of SummaryNovember 8, 2019
5.3 SubmitterHIVOX BIOTEK INC.
Address:5F., No. 123, Xingde Rd., Sanchong Dist., New Taipei City 24158, Taiwan, R.O.C.
Phone:+886-2-8511-2668
Fax:+886-2-8511-2669
Contact:Marx Lee
(Marx.Lee@hivox-biotek.com)

5.4 Identification of the Subject Device

Proprietary/Trade name:HIVOX Spopad EMS
Models:SP-911, SP-921
Classification product code:NGX
Regulation number:890.5850
Regulation description:Power muscle stimulator
Review panel:Physical Medicine
Device class:II

5.5 Identification of the Predicate Device

Proprietary/Trade name:HIVOX Spopad EMS
Models:SP-910, SP-920
510(k) number:K141921
Classification product code:NGX
Regulation number:890.5850
Regulation description:Power muscle stimulator
Review panel:Physical Medicine
Device class:II

{5}------------------------------------------------

5.6 Intended Use / Indication for Use of the Device

These Electrical Muscle Stimulation units are indicated for the improvement of muscle tone and firmness, for strengthening muscles in arms, abdomen, thighs and buttocks areas. Not intended for use in any therapy or for the treatment of any medical conditions or diseases.

5.7 Description of the Device

EMS, Electrical Muscle Stimulator, which improves, tones, firms & strengthens muscle and relaxes stiff muscle through the skin. It is recognized as a clinically proven, effective, non-medication method of training muscle from certain causes. It manages muscle strengthen, toning and firming. It is also free from side effects when used properly, and can also be used as a simple means of self-training.

HIVOX Spopad EMS SP series, SP-911 and SP-921 are the proposed devices for the 510(k) submission. These Electrical Muscle Stimulation units are indicated for the improvement of muscle tone and firmness, for strengthening muscles in arms, abdomen, thighs, and buttocks areas. Not intended for use in therapy or for the treatment of any medical conditions or diseases.

SP-911 and SP-921 are 1-channel battery-operated-user-friendly muscle stimulation system specifically designed to exercise the muscles. Each device comprises namely an electronic stimulator module which generates the required stimulation signals. SP-911 comprises 2 electrodes, which connects the signals from the stimulator to the skin. SP-921 comprises 4 electrodes, which connect the signals from the stimulator to the skin. Power is supplied from one battery, CR2032, located in a compartment protected by a removable battery cover. The user cannot access the wiring or connectors.

5.8 Non-clinical Testing

A series of safety and performance tests were conducted on the subject device, HIVOX Spopad EMS SP-911 and SP-921.

  • Shelf-Life testing
  • Performance verification
  • Usability validation

All the test result demonstrate HIVOX Spopad EMS SP-911 and SP-921 meet the requirement of its pre-defined acceptance criteria and intended use, and is substantially equivalent to the predicate device.

{6}------------------------------------------------

5.9 Clinical Testing

No clinical test data was used to support the decision of substantial equivalence.

5.10 Comparison of Differences and Substantial Equivalence Determination 5.10.1 Comparison between SP-910 and SP-911

ItemSubject devicePredicate deviceSubstantialequivalencedetermination
Device nameHIVOX SpopadEMSHIVOX SpopadEMSN/A
Model nameSP-911SP-910N/A
510(k) numberK192264K141921N/A
Product codeNGXNGXIdentical
Classification namePowered MuscleStimulatorPowered MuscleStimulatorIdentical
Regulation number21 CFR 890.585021 CFR 890.5850Identical
Indication for useIndicated for theimprovement ofmuscle tone andfirmness, forstrengtheningmuscles in arms,abdomen, thighs,and buttocksareas.Not intended foruse in anytherapy or for thetreatment of anymedicalconditions ordiseases.Indicated for theimprovement ofmuscle tone andfirmness, forstrengtheningmuscles in arms,abdomen, thighs,and buttocksareas.Not intended foruse in anytherapy or for thetreatment of anymedicalconditions ordiseases.Identical
TechnologyElectrical MuscleStimulationElectrical MuscleStimulationIdentical
Size (L x W x H, inch)8.15 × 4.06 ×9.41 × 2.76 ×Different but
0.510.45does notadverselyimpact safetyandeffectiveness ofsubject device
Weight (g)24.235.8Different butdoes notadverselyimpact safetyandeffectiveness ofsubject device
Power source3 V battery × 13 V battery × 1Identical
Method of Line Current IsolationBattery supplyBattery supplyIdentical
Patient Leakage CurrentNormal Condition (μΑ)2.02.0Identical
Patient Leakage CurrentSingle Fault Condition (μΑ)2.12.1Identical
Method of channel isolationSingle channelSingle channelIdentical
Average DC current throughelectrodes when device is on butno pulses are being applied (μΑ)00Identical
Number of output modes11Identical
Regulated current or regulatedvoltage?VoltageVoltageIdentical
Software / Firmware /Microprocessor control?YesYesIdentical
Automatic overload trip?NoNoIdentical
Automatic no-load trip?NoNoIdentical
Automatic shut-off?YesYesIdentical
User overrides control?YesYesIdentical
Indicator display –On / Off StatusNoNoIdentical
Indicator display –NoNoIdentical
Low battery?
Indicator display –Voltage / Current levelNoNoIdentical
Timer range (minutes)2020Identical
Compliance with voluntarystandards?IEC 60601-1IEC 60601-1-2IEC 60601-1-11IEC 60601-2-10IEC 60601-1IEC 60601-1-2IEC 60601-2-10Identical
Compliance with 21 CFR 898?YesYesIdentical
Housing material and constructionSiliconeSiliconeIdentical
Output waveformSymmetricalbiphasicSymmetricalbiphasicIdentical
ShapeRectangularRectangularIdentical
Duration of primary(depolarizing) phase00Identical
Pulse duration (μs)400400Identical
Maximum output voltage(V, ±10%) at 500 Ω5252Identical
Maximum output voltage(V, ±10%) at 2k Ω102102Identical
Maximum output voltage(V, ±10%) at 10k Ω150150Identical
Maximum output current(mA, ±10%) at 500 Ω104104Identical
Maximum output current(mA, ±10%) at 2k Ω5151Identical
Maximum output current(mA, ±10%) at 10k Ω1515Identical
Frequency (Hz)3/4/53/4/5Identical
Net charge per pulse at 500 Ω(μC)0.4160.416Identical
Maximum charge at 500 Ω (μC)41.641.6Identical
Conductive surface area (cm²)62.20(Two electrodes)87.58(Two electrodes)Identical
Maximum current density at 5001.6721.187Different but
Ω (mA/cm²)does notadverselyimpact safetyandeffectiveness ofsubject device
Maximum average power densityat 500 Ω (W/cm²)0.08690.0617Different butdoes notadverselyimpact safetyandeffectiveness ofsubject device
BurstmodeA. Pulse per burstN/AN/AIdentical
B. Burst per secondN/AN/AIdentical
C. Burst duration (sec)N/AN/AIdentical
D. Duty cycleN/AN/AIdentical

{7}------------------------------------------------

{8}------------------------------------------------

{9}------------------------------------------------

5.10.2 Comparison between SP-920 and SP-921

ItemSubject devicePredicate deviceSubstantialequivalencedetermination
Device nameHIVOX SpopadEMSHIVOX SpopadEMSN/A
Model nameSP-921SP-920
510(k) numberK192264K141921
Product codeNGXNGXIdentical
Classification namePowered MuscleStimulatorPowered MuscleStimulatorIdentical
Regulation number21 CFR 890.585021 CFR 890.5850Identical
Indication for useIndicated for theimprovement ofmuscle tone andfirmness, forstrengtheningIndicated for theimprovement ofmuscle tone andfirmness, forstrengtheningIdentical
muscles in arms,abdomen, thighs,and buttocksareas.Not intended foruse in anytherapy or for thetreatment of anymedicalconditions ordiseases.muscles in arms,abdomen, thighs,and buttocksareas.Not intended foruse in anytherapy or for thetreatment of anymedicalconditions ordiseases.
TechnologyElectrical MuscleStimulationElectrical MuscleStimulationIdentical
Different butdoes notadverselyimpact safetyandeffectiveness ofsubject device
Size (L × W × H, inch)$7.17 × 6.81 ×0.51$$6.69 × 6.69 ×0.51$
Different butdoes notadverselyimpact safetyandeffectiveness ofsubject device
Weight (g)48.352.6
Power source3 V battery × 13 V battery × 1Identical
Method of Line Current IsolationBattery supplyBattery supplyIdentical
Patient Leakage CurrentNormal Condition (μΑ)2.02.0Identical
Patient Leakage CurrentSingle Fault Condition (μΑ)2.12.1Identical
Method of channel isolationSingle channelSingle channelIdentical
Average DC current through00Identical
electrodes when device is on butno pulses are being applied (μΑ)
Number of output modes11Identical
Regulated current or regulatedvoltage?VoltageVoltageIdentical
Software / Firmware /Microprocessor control?YesYesIdentical
Automatic overload trip?NoNoIdentical
Automatic no-load trip?NoNoIdentical
Automatic shut-off?YesYesIdentical
User overrides control?YesYesIdentical
Indicator display –On / Off StatusNoNoIdentical
Indicator display –Low battery?NoNoIdentical
Indicator display –Voltage / Current levelNoNoIdentical
Timer range (minutes)2020Identical
Compliance with voluntarystandards?IEC 60601-1IEC 60601-1-2IEC 60601-1-11IEC 60601-2-10IEC 60601-1IEC 60601-1-2IEC 60601-2-10Identical
Compliance with 21 CFR 898?YesYesIdentical
Housing material and constructionSiliconeSiliconeIdentical
Output waveformSymmetricalbiphasicSymmetricalbiphasicIdentical
ShapeRectangularRectangularIdentical
Duration of primary(depolarizing) phase00Identical
Pulse duration (μs)400400Identical
Maximum output voltage(V, ±10%) at 500 Ω58.458.4Identical
Maximum output voltage(V, ±10%) at 2k Ω106106Identical
Maximum output voltage154154Identical
(V, ±10%) at 10k Ω
Maximum output current(mA, ±10%) at 500 Ω117117Identical
Maximum output current(mA, ±10%) at 2k Ω5353Identical
Maximum output current(mA, ±10%) at 10k Ω15.415.4Identical
Frequency (Hz)2/4/252/4/25Identical
Net charge per pulse at 500 Ω(μC)0.4680.468Identical
Maximum charge at 500 Ω (μC)46.846.8Identical
Conductive surface area (cm²)110.68(four electrodes)110.68(four electrodes)Identical
Maximum current density at 500Ω (mA/cm²)1.0571.057Identical
Maximum average power densityat 500 Ω (W/cm²)0.06170.0617Identical
BurstmodeE. Pulse per burst2525Identical
F. Burst per second11Identical
G. Burst duration (sec)2020Identical
H. Duty cycle2020Identical

{10}------------------------------------------------

{11}------------------------------------------------

{12}------------------------------------------------

The HIVOX Spopad EMS SP-911 and SP-921 submitted in this 510(k) file are substantially equivalent in intended use, design, technology/principles of operation, materials and performance to the cleared HIVOX Spopad EMS SP-910 and SP-920 (K141921) respectively. Differences between the devices cited in the following do not raise any new issue of substantially equivalence.

FeatureSubstantial equivalence justification
LabelingDifferent in product model.However, it does not affect the intended use or alter thefundamental scientific technology of subject device.
AppearanceChanged for the sake of product segmentation.However, it does not affect the intended use or alter thefundamental scientific technology of subject device.
(1) Comparison of SP-910 and SP-911
-----------------------------------------

{13}------------------------------------------------

Conductive surface areaCompared with predicate device (SP-910), the calculatedpower density value of the subject device (SP-911) israised from 0.0617 W/cm² to 0.0869 W/cm² due to thesmaller size of gel pad. However, we believe that theproduct safety will not be affected because the powerdensity is still far below than the acceptance criterion (0.25W/cm²) based on the recommendation of Class II SpecialControl Guidance Document: Powered Muscle Stimulatorfor Rehabilitation issued on April 5, 2010.Certainly, the intended use and the fundamental scientifictechnology of subject device will not be affected andaltered by this difference.
WeightDifference because of different design of productappearance.However, it does not affect the intended use or later thefundamental scientific technology of subject device.

(2) Comparison of SP-920 and SP-921:

FeatureSubstantial equivalence justification
LabelingDifferent in product model.However, it does not affect the intended use or alter thefundamental scientific technology of subject device.
AppearanceChanged for the sake of product segmentation.However, it does not affect the intended use or alter thefundamental scientific technology of subject device.
WeightDifference because of different design of productappearance.However, it does not affect the intended use or alter thefundamental scientific technology of subject device.

5.11 Discussion

The HIVOX Spopad EMS SP-911 and SP-921 have been compared with HIVOX Spopad EMS SP-910 and SP-920 respectively. The subject devices have the same intended use, principle of operation and technological characters as the predicate device. We have completed the design control process and the validation tests to confirm the safety and performance of subject device. Although there are some specifications that are different between the subject devices and

{14}------------------------------------------------

predicate devices, the test results complied with the test requests, and this demonstrates that the differences between these parameters would not impact the safety and effectiveness of the subject device. Therefore, the difference between the subject devices and predicate devices did not raise any problem of substantial equivalence. The subject devices are substantially equivalent to the predicate devices intended use, safety and performance claims.

5.12 Conclusion

After analyzing non-clinical studies and related testing data, it can be concluded that the HIVOX Spopad EMS SP-911 and SP-921 are substantially equivalent to the predicated devices.

§ 890.5850 Powered muscle stimulator.

(a)
Identification. A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.(b)
Classification. Class II (performance standards).