K Number
K192216

Validate with FDA (Live)

Date Cleared
2020-01-02

(140 days)

Product Code
Regulation Number
890.3860
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Invacare® AVIVA FX Power Wheelchair is indicated to provide mobility and positioning to persons limited to a sitting position.

Device Description

The subject device is a front wheel drive version of the existing previously cleared Invacare TDX SP2 (Center Wheel Drive) Power Wheelchair (K170507) with LiNX Electronics and Ultra Low Maxx Seating System. The subject version of the Invacare TDX SP2 Power Wheelchair consists of the following changes: Alternative Front Wheel Drive Base, Seating Brackets to connect the Ultra Low Maxx Seating System to the base, Wiring Harness to connect the LiNX controller, the AVIVA FX base and the Ultra Low Maxx Seating System. The Invacare® AVIVA FX Power Wheelchair is a 24V DC battery-powered, motor-driven wheelchair, utilizing the predicate device LiNX® Control System and Ultra Low Maxx Seating System. The subject device consists of a rigid or "non-folding" type power wheelchair base with front wheel drive with 2 casters in the rear and two anti-tippers in the front. It is powered by two 12-volt DC batteries and two 4-pole single stage drive motors.

AI/ML Overview

This document describes the Invacare® AVIVA FX Power Wheelchair (Model: IFX-20MP), a front-wheel-drive version of the previously cleared Invacare TDX SP2 Power Wheelchair, with LiNX Electronics and Ultra Low Maxx Seating System.

Here's an analysis of the acceptance criteria and the study proving the device meets them:

1. Table of Acceptance Criteria and Reported Device Performance:

The document primarily relies on demonstrating substantial equivalence to a predicate device (Invacare TDX SP2 Power Wheelchair, K170507) and adherence to recognized international and national standards for safety and performance. The acceptance criteria are implicitly derived from these standards and the characteristics of the predicate device.

Acceptance Criteria CategorySpecific Criteria (from standards/predicate)Reported Device Performance (as demonstrated by testing)
Indications for UseProvide mobility and positioning to persons limited to a sitting position (same as predicate).Met (subject device shares the same Indications for Use as the predicate).
Material CompositionSame as predicate device.Met (subject device utilizes the same material composition).
Technological CharacteristicsSimilar to predicate device, with differences not raising new questions of safety/effectiveness.Met (differences, primarily the front-wheel-drive base, were evaluated and found not to raise new safety/effectiveness concerns).
Static StabilityAdherence to ISO 7176-1:2014.Demonstrated compliance through non-clinical testing.
Dynamic StabilityAdherence to ISO 7176-2:2017.Demonstrated compliance through non-clinical testing.
Effectiveness of BrakesAdherence to ISO 7176-3:2012.Demonstrated compliance through non-clinical testing.
Energy Consumption (Theoretical Distance Range)Adherence to ISO 7176-4:2008.Demonstrated compliance through non-clinical testing.
Dimensions, Mass, Maneuvering SpaceAdherence to ISO 7176-5:2008 / -7:1998.Demonstrated compliance through non-clinical testing.
Maximum Speed, Acceleration, DecelerationAdherence to ISO 7176-6:2018.Demonstrated compliance through non-clinical testing (Maximum Speed 6.25mph, predicate 5mph/5.8mph).
Static, Impact, Fatigue StrengthsAdherence to ISO 7176-8:2014.Demonstrated compliance through non-clinical testing.
Climatic TestsAdherence to ISO 7176-9:2009.Demonstrated compliance through non-clinical testing.
Obstacle Climbing AbilityAdherence to ISO 7176-10:2008.Demonstrated compliance through non-clinical testing.
Power and Control SystemsAdherence to ISO 7176-14:2008.Demonstrated compliance through non-clinical testing.
Information Disclosure, Documentation, LabelingAdherence to ISO 7176-15:1996.Demonstrated compliance through non-clinical testing.
Resistance to Ignition of Postural Support DevicesAdherence to ISO 7176-16:2012.Demonstrated compliance through non-clinical testing.
Wheeled Mobility Devices for Use as Seats in Motor VehiclesAdherence to ISO 7176-19:2008.Demonstrated compliance through non-clinical testing.
Electromagnetic Compatibility (EMC)Adherence to ISO 7176-21:2008.Demonstrated compliance through non-clinical testing.
Batteries and ChargersAdherence to ISO 7176-25:2013.Demonstrated compliance through non-clinical testing.
Software Life CycleAdherence to IEC 62304:2006.Software Verification Testing performed.
FlammabilityAdherence to CAL117:2013, Section 1 and EN 1021-2:2014.Demonstrated compliance through non-clinical testing.
Wireless CoexistenceAdherence to ANSI C63.27.Demonstrated compliance through non-clinical testing.
BiocompatibilityAdherence to AAMI/ANSI/ISO 10993-5:2009 (Cytotoxicity) and AAMI/ANSI/ISO 10993-10:2010 (Irritation).Demonstrated compliance through non-clinical testing.

2. Sample size used for the test set and the data provenance:

The document describes non-clinical laboratory testing. For such tests, the "test set" typically refers to the devices or components manufactured and tested according to the specified standards. The document doesn't explicitly state the exact number of units tested for each standard, but it implies that tests were conducted on the subject device (Invacare® AVIVA FX Power Wheelchair).

  • Sample Size: Not explicitly stated as a number of devices, but testing was performed on the "subject Invacare® AVIVA FX Power Wheelchair" (singular/representative unit for engineering testing).
  • Data Provenance: Non-clinical laboratory data generated to specifically evaluate the subject device against recognized standards. It is prospective testing designed to confirm compliance. The country of origin of the data is not specified but would typically be from accredited testing laboratories, likely within the manufacturer's region or designated by them.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

This device is a powered wheelchair and its evaluation primarily relies on engineering and performance standards, not on interpretation of images or clinical diagnoses by human experts to establish "ground truth" in the medical imaging sense. Therefore, the concept of "experts establishing ground truth" as typically seen in AI/diagnostic device submissions using a test set of patient data, is not directly applicable here.

Instead, the "ground truth" for the performance criteria is derived from:

  • The requirements set forth in the international and national standards (e.g., ISO, IEC, CAL, ANSI).
  • The published specifications and validated performance of the predicate device.

The "experts" involved would be:

  • Engineers and quality assurance personnel from Invacare Corporation responsible for design, manufacturing, and internal testing.
  • Personnel at accredited testing laboratories who perform the tests according to the standards and provide certified results. These individuals would have expertise in mechanical engineering, electronics, software validation, biocompatibility, etc., relevant to the specific test being conducted. Their qualifications are inherent in their accreditation to perform such tests.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

Adjudication methods like 2+1 or 3+1 are typically used in clinical studies where human readers interpret data (e.g., medical images) and their interpretations need to be reconciled to establish a consensus "ground truth." This is not relevant for the non-clinical performance and engineering tests described for this powered wheelchair.

For the non-clinical tests, the "adjudication" is essentially:

  • Direct measurement and observation: The device either meets the quantitative specification (e.g., a stability angle, a speed, a dimension) or it does not.
  • Compliance to a specified test protocol: The testing laboratory follows a defined protocol from the standard, and the results are reported as compliant or non-compliant. Deviations and failures would be investigated and resolved by the manufacturer to achieve compliance.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

No. An MRMC comparative effectiveness study is not mentioned and would not be applicable to a powered wheelchair. Such studies are typically conducted for diagnostic devices (e.g., AI in medical imaging) to assess how AI assistance impacts human reader performance. This submission focuses on the mechanical and electrical safety and performance of the wheelchair as a standalone mobility device.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

The concept of "standalone performance" for an "algorithm only" device is not directly applicable here in the same way it would be for an AI diagnostic algorithm. While the wheelchair does incorporate the LiNX Control System (which has software and control algorithms), its performance is evaluated as an integrated system (device + software) through the various ISO, IEC, CAL, and ANSI standards. The software verification testing (IEC 62304:2006) ensures the software itself functions correctly and safely within the device's overall design. The overall device's performance is tested, essentially, "without human-in-the-loop" in the sense that the tests are designed to objectively measure the device's inherent capabilities against a standard, rather than evaluating a human operator's improved performance with the device.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

The "ground truth" for the evaluation of this powered wheelchair is primarily:

  • Industry and regulatory performance standards: International (ISO, IEC) and national (CAL, ANSI) standards define acceptable performance limits, test methods, and safety requirements.
  • Predicate device characteristics: The specifications and previously cleared performance of the Invacare TDX SP2 Power Wheelchair (K170507) serve as a benchmark for substantial equivalence.
  • Engineering specifications and design verification: Internal technical requirements define the expected operating parameters and safety features.

8. The sample size for the training set:

Not applicable. This submission is for a physical medical device (powered wheelchair), not an AI algorithm that requires a training set of data.

9. How the ground truth for the training set was established:

Not applicable. As there is no "training set," there is no "ground truth for the training set."

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym along with the full name of the agency on the right. The seal features a stylized depiction of a human figure, while the text is in a clear, sans-serif font, with "FDA" in a blue square.

January 2, 2020

Invacare Corporation Elijah Wreh Regulatory Affairs Manager One Invacare Way Elyria, Ohio 44035

Re: K192216

Trade/Device Name: Invacare® Aviva FX Power Wheelchair, Model: IFX-20MP Regulation Number: 21 CFR 890.3860 Regulation Name: Powered Wheelchair Regulatory Class: Class II Product Code: ITI Dated: December 5, 2019 Received: December 6, 2019

Dear Elijah Wreh:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Heather Dean, Ph.D. Acting Assistant Director, Acute Injury Devices DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K192216

Device Name Invacare® AVIVA FX Power Wheelchair Model: IFX 20MP

Indications for Use (Describe)

The Invacare® AVIVA FX Power Wheelchair is indicated to provide mobility and positioning to persons limited to a sitting position.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary – K192216

Submitter Information per 21 CFR 807.92(a)(1)

SPONSOR:Invacare CorporationOne Invacare WayElyria, OH 44035
------------------------------------------------------------------------
PRIMARY CONTACT:Elijah WrehRegulatory Affairs ManagerPhone: (440) 329-6840Email: ewreh@invacare.com
-------------------------------------------------------------------------------------------------------------------
MANUFACTURER:Invacare Corporation1200 Taylor StreetElyria, Ohio 44035
---------------------------------------------------------------------------------
DATE PREPARED per 21 CFR 807.92(a)(1):2 January 2020
--------------------------------------------------------

Device Information per 21 CFR 807.92(a)(2)

NAME OF SUBJECT DEVICE:Invacare® AVIVA FX Power WheelchairModel: IFX 20MP
COMMON/USUAL NAME:Power Wheelchair
CLASSIFICATION NAME:Powered Wheelchair [21 CFR §890.3860]
REGULATORY CLASS:2
PRODUCT CODE:ITI: Wheelchair, Mechanical
PREDICATE DEVICES:Primary: Invacare TDX SP2 Power Wheelchair(K170507)No reference devices were used in this submission.

{4}------------------------------------------------

Device Description per 21 CFR 808.92(a)(4)

The subject device is a front wheel drive version of the existing previously cleared Invacare TDX SP2 (Center Wheel Drive) Power Wheelchair (K170507) with LiNX Electronics and Ultra Low Maxx Seating System. The subject version of the Invacare TDX SP2 Power Wheelchair consists of the following changes:

  • . Alternative Front Wheel Drive Base
  • Seating Brackets to connect the Ultra Low Maxx Seating System to the base .
  • . Wiring Harness to connect the LiNX controller, the AVIVA FX base and the Ultra Low Maxx Seating System

The Invacare® AVIVA FX Power Wheelchair is a 24V DC battery-powered, motor-driven wheelchair, utilizing the predicate device LiNX® Control System and Ultra Low Maxx Seating System. The subject device consists of a rigid or "non-folding" type power wheelchair base with front wheel drive with 2 casters in the rear and two anti-tippers in the front. It is powered by two 12-volt DC batteries and two 4-pole single stage drive motors.

Each accessory connects to the LiNX Control system either directly by connecting to the LiNX communication bus (direct access) or indirectly by connecting to an Input module (indirect access).

The following components are equipped with wireless technology:

  • Primary Driver Controls ●
  • . Direct Access
  • . Display Modules
  • Indirect Access (via input module) Wireless mouse emulator .

The following components are not equipped with wireless technology:

  • Power Modules .
  • Stability Control
  • Indirect Access (via input module) Compact ●
  • Attendant Driver Controls
  • Actuator Control
  • Alternative Driver Controls (LiNX Electronic options) ●
  • Input Module with Integrated Sip-n-Puff
  • USB Charger Module ●

Intended Use per 21 CFR 807.92(A)(5)

The intended use of the device is to provide mobility and positioning to persons limited to a sitting position.

{5}------------------------------------------------

Indications for Use per FORM FDA 3881

The Invacare® AVIVA FX Power Wheelchair is indicated to provide mobility and positioning to persons limited to a sitting position.

Indications for Use Characteristics Comparison

Both the subject and predicate device share the same Indications for Use and Intended use.

Technological Characteristics Comparison with the predicate device per 21 CFR 807.92(a)(6) The technological characteristics comparison demonstrates that the subject device is substantially equivalent in intended use, design, materials, and operational principles to the previously cleared predicate device.

Basis of Substantial Equivalence per 21 CFR 807.100(b)(2)(ii)(A)

The substantial equivalence of the subject device was determined as per the FDA guidance document, "The 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications [510(k)]" and the technological characteristics which include materials, design, energy source, and other device features, as defined in section 513(i)(1)(B) of the FD&C Act and 21 CFR 807.100(b)(2)(ii)(A).

The subject device is a modification to the previously cleared Invacare TDX SP2 Power Wheelchair (K170507) and contains the same components and features as the predicate device. The modification, use of a front wheel drive base does not raise new questions of safety and effectiveness.

The performance testing, device comparison, and dimensional analysis demonstrate that the subject device components and features are the same or substantially equivalent to the predicate device regarding the following:

  • . Static Stability
  • Dynamic Stability of Electric Wheelchairs .
  • Effectiveness of Brakes .
  • Energy Consumption of Electric Wheelchairs and Scooters for Determination of . Theoretical Distance Range
  • Dimensions Mass and Maneuvering Space .
  • Maximum Speed Acceleration and Deceleration of Electric Wheelchairs ●
  • . Seating and Wheel Dimensions
  • . Methods for Static Impact and Fatigue Strengths
  • . Climatic Tests for Electric Wheelchairs

{6}------------------------------------------------

  • Climbing Ability of Electrically Powered Wheelchairs .
  • . Power and Control Systems for Electrically Powered Wheelchairs and Scooters
  • Information Disclosure Documentation and Labeling ●
  • Wheeled Mobility Devices for Use as Seats in Motor Vehicles ●
  • . Methods for Electromagnetic Compatibility of Electrically Powered Wheelchairs and Scooters and Battery Chargers
  • Batteries and chargers for powered wheelchairs ●
  • Software Life Cycle
  • Flammability Testing
  • Assessment of the Ignitability of Upholstered Furniture Ignition Source Smoldering . Cigarette
  • Wireless Coexistence .

The data generated from the subject Invacare® (front wheel drive) AVIVA FX Power

Wheelchair design verification test reports support a finding of substantial equivalence regarding the device comparison, dimensional analysis, device specifications, design characteristics and to provide mobility and positioning to persons limited to a sitting position.

DeviceSubject DeviceInvacare® AVIVA FX Power WheelchairPredicate DevicesInvacare TDX SP2 Power Wheelchair(K170507)
Indications forUseThe Invacare® AVIVA FX PowerWheelchair is indicated to provide mobilityand positioning to persons limited to asitting position.The indication for use of the Invacare®TDX® SP2 Power Wheelchair is to providemobility and positioning to persons limited toa sitting position.
Intended UseTo provide mobility and positioning topersons limited to a sitting positionTo provide mobility and positioning topersons limited to a sitting position
Type of UsePrescription (RX Only)Prescription (RX Only)

Indications for Use Comparison Table

{7}------------------------------------------------

ComponentDescriptionPredicate DeviceInvacare® TDX SP2 Power Wheelchair(K170507)Subject DeviceInvacare® AVIVA FX Power Wheelchair(K192216)
SEATINGUltra LowMaxxPowered PositioningConfigurationsFixed, Tilt/Recline/Elevate, Tilt/Recline, Recline,Elevate, Tilt/Elevate, Tilt Only.Tilt/Recline/Elevate
Seat Widths16" to 22"16" to 22"
Seat Depths15" to 23"15" to 23"
Back Heights18" to 25" (tilt) or 20" to 27" (tilt and recline)18" to 25" (tilt) or 20" to 27" (tilt and recline)
UpholsteryMeshtex, Startex, Spacetex, O-Vinyl, PolyesterMeshtex, Startex, Spacetex, O-Vinyl, Polyester
Elevating SeatRange12"12"
Tilt Range50°50°
Recline Range168°168°
Seat CushionAccessoriesMatrx Libra, Matrx PS, Matrx Vi, Matrx Flo-techLite and ContourMatrx Libra, Matrx PS, Matrx Vi, Matrx Flo-techLite and Contour
Back TypesHigh Back, Matrx PB, Matrx PB Elite, Matrx PBDeep and Matrx PB Elite TRHigh Back, Matrx PB, Matrx PB Elite, Matrx PBDeep and Matrx PB Elite TR
Back and HeadrestAccessoriesMotion Concepts standard, Elan standard, ElanOccipital, Elan 4-point and Motion Concepts OnyxMotion Concepts standard, Elan standard, ElanOccipital, Elan 4-point and Motion Concepts Onyx
Tilt ArmrestsTilt Armrests
Arm TypesDual post adjustable, Ultra Rail mounted flip backcantilever Maxx tilt armDual post adjustable, Ultra Rail mounted flip backcantilever Maxx tilt arm
Recline ArmrestsRecline Armrests
Adjustable, Maxx style cane mounted straight andcurved.Adjustable, Maxx style cane mounted straight andcurved.
ArmpadsModular, Standard, Waterfall, Flat and ErgonomicModular, Standard, Waterfall, Flat and Ergonomic
Leg Rest TypesBasic fixed center mount, Invacare Action fixedswing away receiver, Invacare 70° fixed swingaway, LNX powered center mount, Maxx styleBasic fixed center mount, Invacare Action fixedswing away receiver, Invacare 70° fixed swingaway, LNX powered center mount, Maxx style
ComponentDescriptionPredicate DeviceInvacare® TDX SP2 Power Wheelchair(K170507)Subject DeviceInvacare® AVIVA FX Power Wheelchair(K192216)
powered swing away, Maxx style manual swingaway, Heavy duty 70° swing away.powered swing away, Maxx style manual swingaway, Heavy duty 70° swing away.
Leg RestAccessoriesFlip-up foot platform, Individual foot plates forcenter mount, Foot plate options for elevating andswing-away, Single foot plate options (adjustableand multi-axis adjustable, Heel loops, Calf panelFlip-up foot platform, Individual foot plates forcenter mount, Foot plate options for elevating andswing-away, Single foot plate options (adjustableand multi-axis adjustable, Heel loops, Calf panel
LateralsMatrx standard fixed and offset fixed, Matrx swingaway, Matrx Elite swing away, Matrx Offset Eliteswing away, Matrx lateral trunk support with fixedmounting, Maxx Style swing-awayMatrx standard fixed and offset fixed, Matrx swingaway, Matrx Elite swing away, Matrx Offset Eliteswing away, Matrx lateral trunk support with fixedmounting, Maxx Style swing-away
Hip SupportsLateral, Lift-off removable, Maxx style quickrelease, Swing away removableLateral, Lift-off removable, Maxx style quickrelease, Swing away removable
CONTROL SYSTEMSystem NameLiNX ElectronicLiNX Electronic
LINXCablesVariable cable lengthsA range of standard cable lengths availableVariable cable lengthsA range of standard cable lengths available
System ArchitectureMicroprocessor ControlledMicroprocessor Controlled
Non-ExpandableOptionsYesYes
Expandable OptionsYesYes
Wireless DevicesBluetoothBluetooth
Power Source24V nominal24V nominal
Bus InterfaceCANCAN
POWER BASEBase ConfigurationCentre Wheel DriveFront Wheel Drive
Base Width24" or 25.5"(depending on narrow/wide version)24.3"(single option)
Length(without leg rests)31.5" to 45.3"(depending on seat configuration)42.71"
ComponentDescriptionPredicate DeviceSubject Device
Invacare® TDX SP2 Power Wheelchair(K170507)Invacare® AVIVA FX Power Wheelchair(K192216)
OTHERSPECIFICATIONSGround Clearance> 2.5"> 2.5"
BatteriesGP24 BatteriesGP24 Batteries
Braking SystemElectro-mechanical Friction BrakeElectro-mechanical Friction Brake
Drive Wheel Size14" x 3"14" x 3"
Incline Capability
Maximum Speed5mph, or 5.8mph6.25mph
Motors4-Pole SSD4-Pole SSD
Weight Capacity300lbs.300lbs.
SuspensionEnhanced SureStep® SuspensionFour-bar linkage Independent Suspension System

Design and Technological Characteristics Comparison – Finished Device

{8}------------------------------------------------

{9}------------------------------------------------

Design Characteristics Comparison - LiNX® Electronics

DescriptionPredicate DeviceInvacare® TDX SP2 Power Wheelchair(K170507)Subject DeviceInvacare® AVIVA FX Power Wheelchair (K192216)
System NameLiNXLiNX
CablesVariable cable lengthsA range of standard cable lengths availableVariable cable lengthsA range of standard cable lengths available
System ArchitectureMicroprocessor ControlledMicroprocessor Controlled
Non-Expandable OptionsYesYes
Expandable OptionsYesYes
Wireless DevicesBluetoothBluetooth
Power Source24V nominal24V nominal
Bus InterfaceCANCAN

{10}------------------------------------------------

DescriptionPredicate DeviceInvacare® TDX® SP2 Power WheelchairSubject DeviceInvacare® AVIVA FX Power Wheelchair (K192216)
Seat TypesFixed, Tilt/Recline/Elevate, Tilt/Recline, Recline, Elevate,Tilt/Elevate, Tilt Only.Tilt/Recline/Elevate
Seat Widths16" to 24"16" to 24"
Seat Depths16" to 23"16" to 23"
Back Heights18" to 25" (tilt only)or20" to 27" (tilt and recline)18" to 25" (tilt only)or20" to 27" (tilt and recline)
UpholsteryMeshtex, Startex, Spacetex, O-Vinyl, PolyesterMeshtex, Startex, Spacetex, O-Vinyl, Polyester
Elevating Seat Range12"12"
Tilt Range50°50°
Recline Range168°168°

Design Characteristics Comparison – Seating

General Comparison of the Performance Characteristics Associated with the Control System

DescriptionSubject DeviceInvacare® TDX® SP2 Power Wheelchair (K170507)Subject DeviceInvacare® AVIVA FX Power Wheelchair (K192216)
CablesVariable cable lengthsA range of standard cable lengths availableVariable cable lengthsA range of standard cable lengths available
System ArchitectureMicroprocessor ControlledMicroprocessor Controlled
Non-ExpandableOptionsYesYes
Expandable OptionsYesYes
Wireless DevicesBluetoothBluetooth
Power Source24V nominal24V nominal
Bus InterfaceCANCAN
DescriptionSubject DeviceInvacare® TDX® SP2 Power Wheelchair (K170507)Subject DeviceInvacare® AVIVA FX Power Wheelchair (K192216)
REM400REM400
Mounting2 x M5 screws suitable for both tube and plate.2 x M5 screws suitable for both tube and plate.
ConnectionDirectDirect
User DisplayLCD - colour.LCD - colour.
Viewable LCD Size49 x 74mm49 x 74mm
JoystickMagneticMagnetic
Text & GraphicsIcons & translations.Customisable.Icons & translations.Customisable.
Touch InterfaceYes - capacitiveYes - capacitive
On/Off ButtonYesYes
HornYesYes
Mode/FunctionButton or TouchButton or Touch
Programmable Multi-Function Keys2 x Configurable2 x Configurable
Jack Sockets2 x Stereo.2 x Stereo.
Speed SelectionVirtual speed dial operated by touch.Virtual speed dial operated by touch.
Speed IndicationYesYes
Number of Drive Functions3636
Battery GaugeBar on LCD (continuous)Bar on LCD (continuous)
Seating ControlUp to 8 actuatorsUp to 8 actuators
Lighting ControlYesYes
Real Time ClockYesYes
Charger PortXLRXLR
Status IndicatorError codes on display with supporting icons. Additional LEDflash codes for faulty LCD.Yes.Error codes on display with supporting icons. AdditionalLED flash codes for faulty LCD.Yes.
Mouse MoverBuilt-in Bluetooth connection.Built-in Bluetooth connection.
Remote DiagnosticsBuilt-in Bluetooth connection (Yes).Built-in Bluetooth connection (Yes).

{11}------------------------------------------------

Comparison of the Performance Characteristics Associated with the Enhanced Rehabilitation Primary Remotes

{12}------------------------------------------------

Performance Data

Non-Clinical Test per 21 CFR 807.92(b)(1)

International Organization of Standardization (ISO) testing, California Technical (CAL), American National Standards Institute (ANSI) and European (EN) standards testing were performed to demonstrate that the subject Invacare® AVIVA FX Power Wheelchair meet the performance requirements and is substantially equivalent to the predicate device identified throughout this submission and do not raise any new questions of safety and effectiveness.

{13}------------------------------------------------

Test StandardTest Description
ISO 7176-1:2014Wheelchairs Part 1: Determination of Static Stability
ISO 7176-2:2017Wheelchairs Part 2: Determination of Dynamic Stability of Electrically Powered Wheelchairs
ISO 7176-3:2012Wheelchairs Part 3: Determination of Effectiveness of Brakes
ISO 7176-4:2008Wheelchairs Part 4: Energy Consumption of Electrical Wheelchairs and Scooters for Determination of Theoretical DistanceRange
ISO 7176-5:2008Wheelchairs Part 5: Determination of Dimensions, Mass and Maneuvering Space
ISO 7176-6:2018Wheelchairs Part 6: Determination of Maximum Speed, Acceleration of Electric Wheelchairs
ISO 7176-7:1998Wheelchairs Part 7: Measurement of Seating and Wheel Dimensions
ISO 7176-8:2014Wheelchairs Part 8: Requirements and Test Methods for Static, Impact and Fatigue Strengths
ISO 7176-9:2009Wheelchairs Part 9: Climatic Tests for Electric Wheelchairs
ISO 7176-10:2008Wheelchairs Part 10: Determination of Obstacle Climbing Ability of Electrically Powered Wheelchairs
ISO 7176-11:2012Wheelchairs Part 11: Test Dummies
ISO 7176-13:1989Wheelchairs Part 13: Determination of Coefficient of Friction of Test Surface
ISO 7176-14:2008Wheelchairs Part 14: Power and Control Systems for Electrically Powered Wheelchairs and Scooters – Requirements andTest Methods
ISO 7176-15:1996Wheelchairs Part 15: Requirements for Information Disclosure, Documentation and Labeling
ISO 7176-16:2012Wheelchairs Part 16: Resistance to Ignition of Postural Support Devices
ISO 7176-19:2008Wheelchairs Part 19: Wheeled Mobility Devices for Use as Seats in Motor Vehicles
ISO 7176-21:2008Wheelchairs Part 21: Requirements and test methods for electromagnetic compatibility of electrically powered wheelchairsand scooters, and battery chargers
ISO 7176-22:2014Wheelchairs Part 22: Set-up Procedures
ISO 7176-25:2013Wheelchairs Part 25: Batteries and chargers for powered wheelchairs

{14}------------------------------------------------

Test StandardTest Description
IEC 62304:2006Medical Device Software – Software Life Cycle
CAL117: 2013, Section 1Requirements, Test Procedure and Apparatus for Testing the Smolder Resistance of Materials Used in Upholstered Furniture
EN 1021-2:2014Furniture Assessment of the Ignitability of Upholstered Furniture: Ignition Source: Match Flame Equivalent
ANSI C63.27Wireless Coexistence

{15}------------------------------------------------

Performance Data Conclusions per 21 CFR 807.92(b)(3)

The subject device utilizes the same intended use, same material composition, and similar technological characteristics as the predicate device. The non-clinical laboratory data support the safety and performance of the subject device and demonstrate that any differences in technological characteristics do not raise any new questions of safety and effectiveness. Therefore, the subject Invacare® AVIVA FX Power Wheelchair is substantially equivalent to the predicate devices identified throughout this submission.

Software Verification Testing

Software Verification Testing was performed to evaluate the functionality of the design, materials, and operational principles of the subject device. Software verification testing was conducted on the subject device as recommended by the FDA's guidance document "FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, May 11, 2005." and IEC 62304:2006, Medical Device Software – Software Life Cycle.

Level of Concern: The Level of Concern for the subject device software is moderate. This determination is based on answering the questions in the FDA Guidance Document "FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, May 11, 2005." All of the questions related to a Major Level of Concern were answered "No." One question in the Moderate Level of Concern was answered "Yes" because "prior to mitigation of hazards, a failure of the Software Device could result in Minor Injury, either to a patient or to a user of the subject device

Biocompatibility Testing

The biocompatibility evaluation for the subject Invacare® AVIVA FX Power Wheelchair were conducted in accordance with the FDA Blue Book Memorandum #G95 - 1 "Use of International Standard ISO – 10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing," May 1, 1995, and International Standard ISO 10993 – 1 "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. The battery of testing included the following tests:

  • AAMI / ANSI / ISO 10993-5:2009. Biological Evaluation of Medical Devices Part 5: . Tests for in vitro Cytotoxicity

{16}------------------------------------------------

  • AAMI / ANSI / ISO 10993-10:2010, Biological Evaluation of Medical Devices Part 10: Tests for skin irritation

Animal Study

Animal testing was not required for this submission.

Clinical Testing

Clinical testing was not required for this submission.

§ 890.3860 Powered wheelchair.

(a)
Identification. A powered wheelchair is a battery-operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.(b)
Classification. Class II (performance standards).