K Number
K192167

Validate with FDA (Live)

Device Name
CuraRad-ICH
Manufacturer
Date Cleared
2020-04-13

(248 days)

Product Code
Regulation Number
892.2080
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

CuraRad-ICH is a software workflow tool designed to aid in prioritizing the clinical assessment of adult non-contrast head CT cases with features suggestive of acute intracranial hemorrhage. CuraRad-ICH analyzes cases using deep learning algorithms to identify suspected ICH findings. It makes case-level output available to a PACS/workstation for worklist prioritization or triage.

CuraRad-ICH is not intended to direct attention to specific portions of an image or to anomalies other than acute ICH. Its results are not intended to be used on a stand-alone basis for clinical decision-making nor is it intended to rule out hemorrhage or otherwise preclude clinical assessment of CT studies.

Device Description

CuraRad-ICH is software as a medical device (SaMD) that detects intracranial hemorrhage (ICH) condition by analyzing non-contrast CT images. The software needs to be integrated with a third-party worklist application to receive analysis requests and the corresponding DICOM images, and return the ICH findings (whether suspected ICH is found) to the worklist to alert the radiologists.

To help radiologists triage and prioritize reading of images for patients with ICH, CuraRad-ICH uses deep learning methods to automatically detect acute ICH in non-contrast head CT scans. The software analyzes the input image and returns a binary prediction as to whether the exam suggests the presence of acute ICH.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study proving the device meets them, based on the provided text:

1. Table of Acceptance Criteria & Reported Device Performance

MetricAcceptance Criteria (Implied)Reported Device Performance (95% CI)
SensitivityClinically Meaningful90.6% (85.9%-94.2%)
SpecificityClinically Meaningful93.1% (88.3%-96.4%)
Positive Predictive Value (PPV)Reasonable given prevalenceVaries by prevalence: 1% prevalence: 11.8% (7.2%-18.8%) 15% prevalence: 70% (57.4%-80.1%) 54.9% prevalence (actual): 89.1% (84.3%-92.5%)
Negative Predictive Value (NPV)Very HighVaries by prevalence: 1% prevalence: 99.9% (99.8%-100%) 15% prevalence: 98.3% (97.4%-98.8%) 54.9% prevalence (actual): 91.8% (88.6%-94.3%)
System Processing TimeNot explicitly stated but implies quick processingAverage: 43 seconds (39-46) Median: 33 seconds Min: 16 seconds Max: 301 seconds

Note: The document states that the performance "met the pre-specified criteria for study success," implying that the reported values for sensitivity and specificity were the acceptance criteria.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: 388 CT studies (213 positives and 175 negatives)
  • Data Provenance: Retrospective, collected from 296 imaging facilities across 48 states in the US.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

The document states ground truth was "provided by experienced radiologists" but does not specify the number of experts or their specific qualifications (e.g., years of experience).

4. Adjudication Method for the Test Set

The document does not explicitly state the adjudication method used for establishing ground truth for the test set. It only mentions "ground truth provided by experienced radiologists."

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, an MRMC comparative effectiveness study was not reported. The study focused on the standalone performance of the AI algorithm.

6. If a Standalone (i.e., Algorithm Only Without Human-in-the-Loop Performance) Was Done

Yes, a standalone performance study was done. The reported sensitivity, specificity, PPV, and NPV are for the CuraRad-ICH algorithm's performance in identifying ICH.

7. The Type of Ground Truth Used

The ground truth was established by "experienced radiologists," implying an expert consensus or expert-determined ground truth.

8. The Sample Size for the Training Set

The document does not specify the sample size used for the training set. It only states that the deep learning algorithm was "trained on non-contrast head CT scans with ICH ground truth."

9. How the Ground Truth for the Training Set Was Established

The ground truth for the training set was "provided by experienced radiologists." This implies expert labeling or consensus, similar to the test set, though no further details are given.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

April 13, 2020

CuraCloud Corp. % Yarmela Pavlovic Regulatory Counsel Manatt, Phelps & Phillips, LLP One Embarcadero Center 30th Floor SAN FRANCISCO CA 94111

Re: K192167

Trade/Device Name: CuraRad-ICH Regulation Number: 21 CFR 892.2080 Regulation Name: Radiological computer aided triage and notification software Regulatory Class: Class II Product Code: QAS Dated: March 24, 2020 Received: March 24, 2020

Dear Yarmela Pavlovic:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

510(k) Number (if known) K192167

Device Name

CuraRad-ICH

Indications for Use (Describe)

CuraRad-ICH is a software workflow tool designed to aid in prioritizing the clinical assessment of adult non-contrast head CT cases with features suggestive of acute intracranial hemorrhage. CuraRad-ICH analyzes cases using deep learning algorithms to identify suspected ICH findings. It makes case-level output available to a PACS/workstation for worklist prioritization or triage.

CuraRad-ICH is not intended to direct attention to specific portions of an image or to anomalies other than acute ICH. Its results are not intended to be used on a stand-alone basis for clinical decision-making nor is it intended to rule out hemorrhage or otherwise preclude clinical assessment of CT studies.

Type of Use (Select one or both, as applicable)

网 Prescription Use (Part 21 CFR 801 Subpart D)

□ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page

{3}------------------------------------------------

510(k) SUMMARY

CuraCloud's CuraRad-ICH

Submitter

K192167

CuraCloud Corp. 999 Third Ave. Suite 700 Seattle, WA 98104

Phone: (206) 508-1036 Contact Person: Xiaoxiao Liu, Ph.D.

Date Prepared: April 3, 2020

Name of Device: CuraRad-ICH Classification Name: Radiological Computer-Assisted Triage and Notification Software Regulatory Class: Class II Product Code: QAS Predicate Device: MaxQ-Al Ltd. Accipolx (K182177).

Device Description

CuraRad-ICH is software as a medical device (SaMD) that detects intracranial hemorrhage (ICH) condition by analyzing non-contrast CT images. The software needs to be integrated with a third-party worklist application to receive analysis requests and the corresponding DICOM images, and return the ICH findings (whether suspected ICH is found) to the worklist to alert the radiologists.

For ICH patients, immediate emergency diagnosis and treatment is critical for saving their lives and later recovery. Thus, it is very important to triage and identify ICH patients in a speedy manner in order to prioritize their treatment. Computed tomography (CT) is a non-invasive and effective diagnosis imaging approach to detect ICH. Acute Intracranial hemorrhage can be recognized on non-contrast CT scans since blood has higher density (Hounsfield unit, HU) than other brain tissues but lower than that of bones. Radiologists are able to identify ICH and determine the location and severity of any such bleeding from the intensity patterns presented in the images.

To help radiologists triage and prioritize reading of images for patients with ICH, CuraRad-ICH uses deep learning methods to automatically detect acute ICH in non-contrast head CT scans. The software analyzes the input image and returns a binary prediction as to whether the exam suggests the presence of acute ICH.

Intended Use / Indications for Use

CuraRad-ICH is a software workflow tool designed to aid in prioritizing the clinical assessment of adult non-contrast head CT cases with features suggestive of acute intracranial hemorrhage. CuraRad-ICH analyzes cases using deep learning algorithms to

{4}------------------------------------------------

identify suspected ICH findings. It makes case-level output available to a PACS/workstation for worklist prioritization or triage.

CuraRad-ICH is not intended to direct attention to specific portions of an image or to anomalies other than acute ICH. Its results are not intended to be used on a stand-alone basis for clinical decision-making nor is it intended to rule out hemorrhage or otherwise preclude clinical assessment of CT studies.

Summary of Technological Characteristics

The core technology of this software is a deep learning algorithm trained on non-contrast head CT scans with ICH ground truth provided by experienced radiologists. The algorithm utilizes an end-to-end trainable 3D classification framework for automatic ICH detection.

The predicate device and the subject device both use deep learning algorithms to detect a predefined clinical condition of the brain. Both devices analyze head CT scans immediately after the patient is scanned and works in parallel to and in conjunction with the standard of care workflow. Both devices use the results of its respective detection algorithm to automatically triage and enable prioritization of CT scans for radiologist review, thereby creating an opportunity for earlier diagnosis and treatment of ICH. Although the algorithms differ between the two devices, this difference does not raise any new questions of safety or effectiveness as the core question remains the accuracy of the respective algorithms in detecting ICH.

CuraRad-ICHAccipiolx
ManufacturerCuraCloud CorporationMaxQ Al
Intended ClinicalEnd UserRadiologists/Trained CliniciansRadiologists/Trained Clinicians
Clinical ConditionAcute Intracranial hemorrhageAcute Intracranial hemorrhage
Independent ofstandard of careworkflowYes; No cases are removedfrom worklistYes; No cases are removedfrom worklist
Al UsedYesYes
Input ImageModalityNon-contrast Head CTNon-contrast Head CT
Non-DiagnosticPreviewNoNo
OutputSuspected ICH (Yes or No)Suspected ICH (Yes or No)
Results ReceiverPACS/WorkstationPACS/Workstation

In addition, both devices are PACS agnostic and can work with any PACS and RIS system. Both devices also flag the suspected ICH on the DICOM images for the radiologist to review.

{5}------------------------------------------------

Performance Data

CuraCloud conducted a retrospective, blinded, multisite clinical validation study with CuraRad-ICH. The primary endpoint evaluates the performance of the software in identifying ICH findings from non-contrast head CT scans on a validation dataset of 388 CT studies (213 positives and 175 negatives) from 296 imaging facilities across 48 states in the US.

The observed ICH detection sensitivity was 90.6% (95% Cl: 85.9%-94.2%), and specificity was 93.1% (95% Cl: 88.3%-96.4%), demonstrating that CuraRad-ICH yielded clinical meaningful results and met the pre-specified criteria for study success on the clinical validation dataset.

The Positive Predictive Value (PPV) and negative predictive value (NPV) of CuraRad ICH were also evaluated. Given that ICH prevalence is not well established and may vary from site to site, analysis was conducted using both low (1%) and high (15%) estimated clinical prevalence, as well as the actual prevalence from the retrospective study. At a prevalence rate of 1%, PPV was 11.8% (95% Cl: 7.2%-18.8%) and NPV was 99.9% (95% Cl: 99.8%-100%). At a prevalence rate of 15%, PPV was 70% (95% Cl: 57.4%-80.1%) and NPV was 98.3% (95% Cl: 97.4%-98.8%). At a prevalence rate of 54.9%, PPV was 89.1% (95% Cl: 84.3%-92.5%) and NPV was 91.8% (88.6%-94.3%). The NPV was very high across all estimated prevalence rates and, as expected, PPV varied significantly, but was reasonable across prevalence estimates given the relative expected rarity of the condition.

In accordance with FDA's recommendations, stratified analyses were also performed by slice thickness, the number of detector rows and scanner manufacturers. Examining slice thickness up to and including 4.0 mm versus slice thickness greater than 4.0 mm, ICH detection sensitivity was 92.2% and 89.1% respectively, while specificity was 98.7% and 93% respectively. No significant statistical difference was observed between the performance of the two slice thickness groups. Evaluating performance by detector rows also did not produce any statistically notable differences, with sensitivity ranging from 88.4% to 92.9% and specificity ranging from 89.8% to 97.4%. Additionally, no statistically significant differences were revealed when analyzing performance by imaging equipment manufacturer.

The secondary endpoint evaluates the system processing time of the device based on the same validation dataset. The observed system processing time per study is 43 seconds (95%C1: 39-46) in average, with a median time of 33 seconds and a standard deviation of 32 seconds. The minimum observed system processing time was 16 seconds, and the maximum observed processing time was 301 seconds.

Based on the clinical performance as documented in the pivotal clinical study, CuraRad-ICH has a safety and effectiveness profile that is similar to the predicate device.

{6}------------------------------------------------

Conclusions

CuraRad-ICH is as safe and effective as the MaxQ's Accipiolx. CuraRad-ICH has the same intended uses and indications, technological characteristics, and principles of operation as its predicate device. In addition, the minor technological differences between CuraRad-ICH and its predicate device raise no new issues of safety or effectiveness. Performance data demonstrate that CuraRad-ICH is as safe and effective as the predicate. Thus, CuraRad-ICH is substantially equivalent.

§ 892.2080 Radiological computer aided triage and notification software.

(a)
Identification. Radiological computer aided triage and notification software is an image processing prescription device intended to aid in prioritization and triage of radiological medical images. The device notifies a designated list of clinicians of the availability of time sensitive radiological medical images for review based on computer aided image analysis of those images performed by the device. The device does not mark, highlight, or direct users' attention to a specific location in the original image. The device does not remove cases from a reading queue. The device operates in parallel with the standard of care, which remains the default option for all cases.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Design verification and validation must include:
(i) A detailed description of the notification and triage algorithms and all underlying image analysis algorithms including, but not limited to, a detailed description of the algorithm inputs and outputs, each major component or block, how the algorithm affects or relates to clinical practice or patient care, and any algorithm limitations.
(ii) A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will provide effective triage (
e.g., improved time to review of prioritized images for pre-specified clinicians).(iii) Results from performance testing that demonstrate that the device will provide effective triage. The performance assessment must be based on an appropriate measure to estimate the clinical effectiveness. The test dataset must contain sufficient numbers of cases from important cohorts (
e.g., subsets defined by clinically relevant confounders, effect modifiers, associated diseases, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals for these individual subsets can be characterized with the device for the intended use population and imaging equipment.(iv) Stand-alone performance testing protocols and results of the device.
(v) Appropriate software documentation (
e.g., device hazard analysis; software requirements specification document; software design specification document; traceability analysis; description of verification and validation activities including system level test protocol, pass/fail criteria, and results).(2) Labeling must include the following:
(i) A detailed description of the patient population for which the device is indicated for use;
(ii) A detailed description of the intended user and user training that addresses appropriate use protocols for the device;
(iii) Discussion of warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (
e.g., poor image quality for certain subpopulations), as applicable;(iv) A detailed description of compatible imaging hardware, imaging protocols, and requirements for input images;
(v) Device operating instructions; and
(vi) A detailed summary of the performance testing, including: test methods, dataset characteristics, triage effectiveness (
e.g., improved time to review of prioritized images for pre-specified clinicians), diagnostic accuracy of algorithms informing triage decision, and results with associated statistical uncertainty (e.g., confidence intervals), including a summary of subanalyses on case distributions stratified by relevant confounders, such as lesion and organ characteristics, disease stages, and imaging equipment.