K Number
K192152

Validate with FDA (Live)

Date Cleared
2019-12-13

(126 days)

Product Code
Regulation Number
892.1550
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-86/DC-86S/DC-89/DC-TV/DC-TQ Diagnostic Ultrasound System is applicable for adults, pregnant woments and neonates. It is intended for use in fetal, abdominal. intra-operative (abdominal, thoracic, and vascular), pediatric, small organ(breast, thyroid, testes), neonatal and adult cephalic, trans-rectal, trans-vaginal, musculo-skeletal(conventional, cardiac(adult, pediatric), trans-soph. (Cardiac), peripheral vessel, and urology exams.

Device Description

DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85/DC-86/DC-86S/DC-89/DC-TV/ DC-TQ is a software controlled, ultrasonic diagnostic system. Its function is to acquire and display ultrasound data in B-Mode, M-Mode, PW-Mode, CCW-Mode, Color-Mode , Power/Dirpower Mode, THI, Smart3D, 4D, iScape, TDI, Color M, Biopsy Guidance, Strsin Elastography, Contrast imaging (Contrast agent for Liver and LVO), STE, STQ, Ultrasound Fusion Imaging, GYN/Pelvic and the combined mode (i.e. B/M-Mode, B/PW-mode, B/PW/Color). This system is a Track 3 device that employs an array of probes that include Linear array, Convex array, Phased array, Volume and Pencil probe.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study information for the DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85/DC-86/DC-86S/DC-89/DC-TV/DC-TQ Diagnostic Ultrasound System, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance:

Acceptance Criteria CategorySpecific Criteria/StandardReported Device Performance
BiocompatibilityISO 10993-1Evaluated and conducted in accordance with ISO 10993-1.
Electrical SafetyIEC 60601-1Complies with IEC 60601-1.
Electromagnetic Compatibility (EMC)IEC 60601-1-2Complies with IEC 60601-1-2.
Ultrasound Specific SafetyIEC 60601-2-37Complies with IEC 60601-2-37.
Acoustic Power OutputFDA Guidance for Industry and FDA Staff "Marketing Clearance of Diagnostic Ultrasound Systems and Transducers"Below the limits recommended by FDA guidance.
Software General SafetyIEC 62304 and FDA Guidance "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices"Software verification and validation testing were conducted and documentation was provided.
UsabilityIEC 62366 and IEC 60601-1-6Designed in compliance with these standards.
Risk ManagementISO 14971Application of risk management as per ISO 14971.
Quality Systems21 CFR 820, ISO 9001, ISO 13485Conforms with these quality systems.

2. Sample Size Used for the Test Set and Data Provenance:

The provided document does not explicitly state specific sample sizes for test sets for the performance data. The sections on performance data refer to compliance with standards and guidelines rather than specific study parameters involving patient data.

Regarding data provenance, the document mentions non-clinical tests (biocompatibility, electrical safety, EMC, acoustic power, software verification/validation) and compliance with international standards. There is no mention of patient data (retrospective or prospective) or countries of origin for test data in the provided text.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications:

This information is not provided in the document. The non-clinical tests described involve compliance with established standards, which don't typically involve human expert ground truth for comparative performance in the same way clinical studies do.

4. Adjudication Method for the Test Set:

This information is not provided in the document. The tests performed are primarily technical and regulatory compliance checks.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and the effect size of human readers improving with AI vs. without AI assistance:

This information is not provided in the document. The document explicitly states: "Not applicable. The subject of this submission... does not require clinical studies to support substantial equivalence." An MRMC study would be a type of clinical study.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

The document describes the device as a "software controlled, ultrasonic diagnostic system" with various "Main Added Features" (e.g., Ultrasound Fusion Imaging, Smart Pelvic Floor, STE/STQ, Strain Elastography, Smart Face, Smart Planes FH, Glazing Flow, Fusion RESP). While these features are algorithm-based, the document does not explicitly detail standalone performance studies for these specific algorithms, nor does it refer to them as "AI" in the context of standalone performance evaluation. The evaluation focuses on overall system compliance with safety and effectiveness standards, and the capabilities of these features as part of the system.

7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.):

For the non-clinical tests detailed, the "ground truth" is implied by the international and regulatory standards themselves (e.g., IEC, ISO, FDA guidance limits). For instance, acoustic power output is compared against specified limits, and electrical safety against defined safety requirements. There is no mention of clinical ground truth (like pathology or outcomes data) because clinical studies were not conducted.

8. The Sample Size for the Training Set:

This information is not provided in the document. The document describes specific software features but does not delve into the development or training of any potential machine learning components within these features.

9. How the Ground Truth for the Training Set Was Established:

This information is not provided in the document, as details about training sets and their ground truth establishment are absent.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue box, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font below that.

December 13, 2019

Shenzhen Mindray Bio-Medical Electronics Co., Ltd. % Zhang Wei Engineer of Technical Regulation Keji 12th Road South, Hi-tech Industrial Park Shenzhen, Guangdong 518057 CHINA

Re: K192152

Trade/Device Name: DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85/DC-86/DC-86S/DC-89/ DC-TV/DC-TQ Diagnostic Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: Class II Product Code: IYN, IYO, ITX and LLZ Dated: October 30, 2019 Received: November 21, 2019

Dear Zhang Wei:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K192152

Device Name

DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-86/DC-865/DC-89/DC-TV/DC-TQ Diagnostic Ultrasound System

Indications for Use (Describe)

The DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-86/DC-86S/DC-89/DC-TV/DC-TQ Diagnostic Ultrasound System is applicable for adults, pregnant woments and neonates. It is intended for use in fetal, abdominal. intra-operative (abdominal, thoracic, and vascular), pediatric, small organ(breast, thyroid, testes), neonatal and adult cephalic, trans-rectal, trans-vaginal, musculo-skeletal(conventional, cardiac(adult, pediatric), trans-soph. (Cardiac), peripheral vessel, and urology exams.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and comblete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

System:DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85/DC-86/DC-86S/DC-89/DC-TV/DC-TQ Diagnostic UltrasoundSystem
Transducer:
Intended Use:Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation
General (Track 1 Only)Specific (Track 1 & 3)BMPWDCWDColor DopplerAmplitude DopplerCombinedOther (specify)
OphthalmicOphthalmic
Fetal Imaging& OtherFetalPPPPPPNote 1, 2,3, 4,6,7,13
AbdominalPPPPPPPNote 1, 2,3,4,6,7,10,11,12,13,14
Intra-operative (Specify*)PPPPPPNote1,2,4
Intra-operative (Neuro)
Laparoscopic
PediatricPPPPPPNote 1, 2,3, 4,6,7,13
Small Organ (Specify**)PPPPPPNote 1,2,4,7,8,11,12,13
Neonatal CephalicPPPPPPNote 1,2, 4,6,7,13
Adult CephalicPPPPPPNote 1, 2,4, 6,7,13
Trans-rectalPPPPPPNote 1, 2,3, 4,6,7,8
Trans-vaginalPPPPPPNote 1, 2,3, 4,6,7,8,14
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)PPPPPPNote 1, 2,4,6,7,8,11,12,13
Musculo-skeletal(Superficial)PPPPPNote 1,2,4,7,8,11,12,13
Intravascular
CardiacCardiac AdultPPPPPPNote 1, 2,4,5,6,7,9,13
Cardiac PediatricPPPPPPNote 1, 2,4,5,6,7,13
Intravascular (Cardiac)Trans-esoph. (Cardiac)PPPPPPNote 1, 5,6
Intra-cardiac
PeripheralvesselPeripheral vesselPPPNPPPNote 1, 2, 4,6,7,13
Other (Specify***)PPPPPPNote 1, 2, 4,6,7
N=new indication; P=previously cleared by FDA; E=added under Appendix E
Additional comments: Combined modes--B+M、PW+B、Color + B、Power + B、PW +Color+B、Power + PW +B.
*Intraoperative includes abdominal, thoracic, and vascular
**Small organ-breast, thyroid, testes.
***Other use includes Urology.
Note 1: Tissue Harmonic Imaging. The feature does not use contrast agents.
Note 2: Smart3D
Note 3:4D(Real-time 3D)
Note 4: iScape
Note 5: TDI
Note 6: Color M
Note 7: Biopsy Guidance
Note 8: Strain Elastography
Note 9: Contrast imaging(contrast agent for LVO)
Note 10: Contrast imaging (Contrast agent for Liver)
Note 11: STE
Note 12: STQ
Note 13: Ultrasound Fusion Imaging
Note 14: GYN/Pelvic
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
System:DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85/DC-86/DC-86S/DC-89/DC-TV/DC-TQ Diagnostic UltrasoundSystem
Transducer:C5-1E
Intended Use:Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation
General (Track 1 Only)Specific (Track 1 & 3)BMPWDCWDColorDopplerAmplitude DopplerCombined(specify)Other (specify)
OphthalmicOphthalmic
Fetal Imaging& OtherFetalPPPPPPNote 1, 2, 4,6,7,13
AbdominalPPPPPPNote 1, 2,4,6,7,10,11,12,13,14
Intra-operative (Specify*)
Intra-operative (Neuro)
Laparoscopic
PediatricPPPPPPNote 1, 2, 4,6,7,13
Small Organ (Specify**)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)PPPPPPNote 1, 2, 4,6,7,13
Musculo-skeletal(Superficial)
Intravascular
CardiacCardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
PeripheralvesselPeripheral vesselPPPPPPNote 1, 2, 4,6,7,13
Other (Specify***)
N=new indication; P=previously cleared by FDA(K173471); E=added under Appendix E
Additional comments: Combined modes--B+M\PW+B\Color + B\Power + B\PW +Color+B\Power + PW +B.
*Intraoperative includes abdominal, thoracic, and vascular
**Small organ-breast, thyroid, testes.
***Other use includes Urology.
Note 1: Tissue Harmonic Imaging. The feature does not use contrast agents.
Note 2: Smart3D
Note 3:4D(Real-time 3D)
Note 4: iScape
Note 5: TDI
Note 6: Color M
Note 7: Biopsy Guidance
Note 8: Strain Elastography
Note 9: Contrast imaging(contrast agent for LVO)
Note 10: Contrast imaging (Contrast agent for Liver)
Note 11: STE
Note 12: STQ
Note 13: Ultrasound Fusion Imaging
Note 14: GYN/Pelvic
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE NEEDED)
System:DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85/DC-86/DC-86S/DC-89/DC-TV/DC-TQ Diagnostic Ultrasound System
System
Transducer:L14-5WE
Intended Use:Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical Application
General (Track 1 Only)Specific (Track 1 & 3)BMPWDCWDColorDopplerAmplitudeDopplerCombined(specify)Other (specify)
OphthalmicOphthalmic
Fetal Imaging& OtherFetal
AbdominalPPPPPPNote 1,2, 4,7,13
Intra-operative (Specify*)
Intra-operative (Neuro)
Laparoscopic
PediatricPPPPPPNote 1,2, 4,7,13
Small Organ (Specify**)PPPPPPNote 1,2,4,7,8,11,12,13
Neonatal CephalicPPPPPPNote 1,2, 4,7,13
Adult Cephalic
Trans-rectalTrans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)PPPPPPNote 1,2,4,7,11,12,13
Musculo-skeletal(Superficial)PPPPPPNote 1,2,4,7,11,12,13
Intravascular
CardiacCardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
PeripheralvesselPeripheral vesselPPPPPPNote 1,2, 4,7,13
Other (Specify***)
N=new indication; P=previously cleared by FDA(K173471); E=added under Appendix E
Additional comments: Combined modes--B+M, PW+B, Color + B, Power + B, PW+Color+B, Power + PW +B.
*Intraoperative includes abdominal, thoracic, and vascular
**Small organ-breast, thyroid, testes.
***Other use includes Urology.
Note 1: Tissue Harmonic Imaging. The feature does not use contrast agents.
Note 2: Smart3D
Note 3:4D(Real-time 3D)
Note 4: iScape
Note 5: TDI
Note 6: Color M
Note 7: Biopsy Guidance
Note 8: Strain Elastography
Note 9: Contrast imaging(contrast agent for LVO)
Note 10: Contrast imaging (Contrast agent for Liver)
Note 11: STE
Note 12: STQ
Note 13: Ultrasound Fusion Imaging
Note 14: GYN/Pelvic
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE NEEDED)
System:DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85/DC-86/DC-86S/DC-89/DC-TV/DC-TQ Diagnostic Ultrasound System
Transducer:L9-3E
Intended Use:Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation
BMPWDCWDColorDopplerAmplitude DopplerCombined(specify)Other (specify)
General (Track 1 Only)Specific (Track 1 & 3)
OphthalmicOphthalmic
Fetal Imaging& OtherFetal
AbdominalPPPPPPNote 1, 2, 4,7,10
Intra-operative (Specify*)
Intra-operative (Neuro)
Laparoscopic
PediatricPPPPPPNote 1, 2, 4,7
Small Organ (Specify**)PPPPPPNote 1,2, 4,7,8,11,12
Neonatal CephalicPPPPPPNote 1, 2, 4,7
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)PPPPPPNote 1, 2, 4,7,11,12
Musculo-skeletal(Superficial)PPPPPPNote 1, 2, 4,7,11,12
Intravascular
CardiacCardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
PeripheralvesselPeripheral vesselPPPPPPNote 1, 2, 4,7
Other (Specify***)
N=new indication; P=previously cleared by FDA(K173471); E=added under Appendix E
Additional comments: Combined modes--B+M、PW+B、Color + B、Power + B、PW+Color+B、Power + PW +B
*Intraoperative includes abdominal, thoracic, and vascular
**Small organ-breast, thyroid, testes.
***Other use includes Urology.
Note 1: Tissue Harmonic Imaging. The feature does not use contrast agents.
Note 2: Smart3D
Note 3:4D(Real-time 3D)
Note 4: iScape
Note 5: TDI
Note 6: Color M
Note 7: Biopsy Guidance
Note 8: Strain Elastography
Note 9: Contrast imaging(contrast agent for LVO)
Note 10: Contrast imaging (Contrast agent for Liver)
Note 11: STE
Note 12: STQ
Note 13: Ultrasound Fusion Imaging
Note 14: GYN/Pelvic
System:DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85/DC-86/DC-86S/DC-89/DC-TV/DC-TQ Diagnostic UltrasoundSystem
Transducer:V11-3HE
Intended Use:Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation
General (Track 1 Only)Specific (Track 1 & 3)BMPWDCWDColor DopplerAmplitude DopplerCombined (specify)Other (specify)
OphthalmicOphthalmic
Fetal Imaging & OtherFetalPPPPPPNote 1, 2, 4,6,7
Abdominal
Intra-operative (Specify*)
Intra-operative (Neuro)
Laparoscopic
Pediatric
Small Organ (Specify**)
Neonatal Cephalic
Adult Cephalic
Trans-rectalPPPPPPNote 1, 2, 4,6,7,8
Trans-vaginalPPPPPPNote 1, 2, 4,6,7,8,14
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
CardiacCardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
PeripheralvesselPeripheral vessel
Other (Specify***)PPPPPPNote 1, 2, 4,6,7
N=new indication; P=previously cleared by FDA(K173471); E=added under Appendix E
Additional comments: Combined modes--B+M\、PW+B\、Color + B\、Power + B\、PW+Color+B\、Power + PW +B.
*Intraoperative includes abdominal, thoracic, and vascular
**Small organ-breast, thyroid, testes.
***Other use includes Urology.
Note 1: Tissue Harmonic Imaging. The feature does not use contrast agents.
Note 2: Smart3D
Note 3:4D(Real-time 3D)
Note 4: iScape
Note 5: TDI
Note 6: Color M
Note 7: Biopsy Guidance
Note 8: Strain Elastography
Note 9: Contrast imaging(contrast agent for LVO)
Note 10: Contrast imaging (Contrast agent for Liver)
Note 11: STE
Note 12: STQ
Note 13: Ultrasound Fusion Imaging
Note 14: GYN/Pelvic
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE NEEDED)
System:DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85/DC-86/DC-86S/DC-89/DC-TV/DC-TQ Diagnostic Ultrasound System
Transducer:D8-2E
Intended Use:Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation
General (Track Specific (Track 1 & 3)BMPWDCWDColorDopplerAmplitude DopplerCombined(specify)Other (specify)
1 Only)
OphthalmicOphthalmic
Fetal Imaging& OtherFetalPPPPPPPNote 1, 3, 4, 6,7
AbdominalPPPPPPPNote 1, 3, 4, 6,7,14
Intra-operative (Specify*)
Intra-operative (Neuro)
Laparoscopic
Pediatric
Small Organ (Specify**)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
CardiacCardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
PeripheralvesselPeripheral vessel
Other (Specify***)
N=new indication; P=previously cleared by FDA(K173471); E=added under Appendix E
Additional comments: Combined modes--B+M、PW+B、Color + B、Power + B、PW +Color+B、Power + PW +B.
*Intraoperative includes abdominal, thoracic, and vascular
**Small organ-breast, thyroid, testes.
***Other use includes Urology.
Note 1: Tissue Harmonic Imaging. The feature does not use contrast agents.
Note 2: Smart3D
Note 3:4D(Real-time 3D)
Note 4: iScape
Note 5: TDI
Note 6: Color M
Note 7: Biopsy Guidance
Note 8: Strain Elastography
Note 9: Contrast imaging(contrast agent for LVO)
Note 10: Contrast imaging (Contrast agent for Liver)
Note 11: STE
Note 12: STQ
Note 13: Ultrasound Fusion Imaging
Note 14: GYN/Pelvic
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE NEEDED)
System:DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85/DC-86/DC-86S/DC-89/DC-TV/DC-TQ Diagnostic Ultrasound System
Transducer:D7-2E
Intended Use:Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation
General (Track Specific (Track 1 & 3) 1 Only)BMPWDCWDColor DopplerAmplitude DopplerCombined (specify)Other (specify)
OphthalmicOphthalmic
Fetal Imaging & OtherFetalPPPPPPNote 1, 3, 4, 6
AbdominalPPPPPPNote 1, 3, 4, 6,14
Intra-operative (Specify*)
Intra-operative (Neuro)
Laparoscopic
Pediatric
Small Organ (Specify**)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
CardiacCardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Peripheral vesselPeripheral vessel
Other (Specify***)
N=new indication; P=previously cleared by FDA(K173471); E=added under Appendix E
Additional comments: Combined modes--B+M、PW+B、Color + B、Power + B、PW +Color+B、Power + PW +B
*Intraoperative includes abdominal, thoracic, and vascular
**Small organ-breast, thyroid, testes.
***Other use includes Urology.
Note 1: Tissue Harmonic Imaging. The feature does not use contrast agents.
Note 2: Smart3D
Note 3:4D(Real-time 3D)
Note 4: iScape
Note 5: TDI
Note 6: Color M
Note 7: Biopsy Guidance
Note 8: Strain Elastography
Note 9: Contrast imaging(contrast agent for LVO)
Note 10: Contrast imaging (Contrast agent for Liver)
Note 11: STE
Note 12: STQ
Note 13: Ultrasound Fusion Imaging
Note 14: GYN/Pelvic
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE NEEDED)
System:DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85/DC-86/DC-86S/DC-89/DC-TV/DC-TQ Diagnostic UltrasoundSystem
Transducer:C7-3E
Intended Use:Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation
General (Track1 Only)Specific (Track 1 & 3)BMPWDCWDColorDopplerAmplitude DopplerCombined(specify)Other (specify)
OphthalmicOphthalmic
Fetal Imaging& OtherFetalPPPPPPNote 1, 2, 4.6.7
AbdominalPPPPPPNote 1, 2, 4,6,7,10,14
Intra-operative (Specify*)
Intra-operative (Neuro)
Laparoscopic
PediatricPPPPPPNote 1, 2, 4,6,7
Small Organ (Specify**)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
CardiacCardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
PeripheralvesselPeripheral vesselPPPPPPNote 1, 2, 4,6,7
Other (Specify***)
N=new indication; P=previously cleared by FDA(K173471); E=added under Appendix E
Additional comments: Combined modes--B+M\、PW+B\、Color + B\、Power + B\、PW +Color+B\、Power + PW +B.
*Intraoperative includes abdominal, thoracic, and vascular
**Small organ-breast, thyroid, testes.
***Other use includes Urology.
Note 1: Tissue Harmonic Imaging. The feature does not use contrast agents.
Note 2: Smart3D
Note 3:4D(Real-time 3D)
Note 4: iScape
Note 5: TDI
Note 6: Color M
Note 7: Biopsy Guidance
Note 8: Strain Elastography
Note 9: Contrast imaging(contrast agent for LVO)
Note 10: Contrast imaging (Contrast agent for Liver)
Note 11: STE
Note 12: STQ
Note 13: Ultrasound Fusion Imaging
Note 14: GYN/Pelvic
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE NEEDED)
System:DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85/DC-86/DC-86S/DC-89/DC-TV/DC-TQ Diagnostic UltrasoundSystem
Transducer:L12-3E
Intended Use:Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation
General (Track Specific (Track 1 & 3)BMPWDCWDColorDopplerAmplitude DopplerCombined(specify)Other (specify)
1 Only)e Doppler
(specify)
OphthalmicOphthalmic
Fetal Imaging& OtherFetal
AbdominalPPPPPPNote 1, 2, 4,7
Intra-operative (Specify*)
Intra-operative (Neuro)
Laparoscopic
PediatricPPPPPPNote 1, 2, 4,7
Small Organ (Specify**)PPPPPPNote 1,2, 4,7,8,11,12
Neonatal CephalicPPPPPPNote 1, 2, 4,7
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)PPPPPPNote 1, 2, 4,7,11,12
Musculo-skeletal(Superficial)PPPPPPNote 1, 2, 4,7,11,12
Intravascular
CardiacCardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
PeripheralvesselPeripheral vesselPPPPPPNote 1, 2, 4,7
Other (Specify***)
N=new indication; P=previously cleared by FDA(K173471); E=added under Appendix E
Additional comments: Combined modes--B+M、PW+B、Color + B、Power + B、PW +Color+B、Power + PW +B.
*Intraoperative includes abdominal, thoracic, and vascular
**Small organ-breast, thyroid, testes.
***Other use includes Urology.
Note 1: Tissue Harmonic Imaging. The feature does not use contrast agents.
Note 2: Smart3D
Note 3:4D(Real-time 3D)
Note 4: iScape
Note 5: TDI
Note 6: Color M
Note 7: Biopsy Guidance
Note 8: Strain Elastography
Note 9: Contrast imaging(contrast agent for LVO)
Note 10: Contrast imaging (Contrast agent for Liver)
Note 11: STE
Note 12: STQ
Note 13: Ultrasound Fusion Imaging
System:Note 14: GYN/PelvicDC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85/DC-86/DC-86S/DC-89/DC-TV/DC-TQ Diagnostic UltrasoundSystem
Transducer:DE11-3E
Intended Use:Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation
General (Track1 Only)Specific (Track 1 & 3)BMPWDCWDColorDopplerAmplitude DopplerCombined(specify)Other (specify)
OphthalmicOphthalmic
Fetal Imaging& OtherFetalPPPPPPNote 1, 3,4,6,7
Abdominal
Intra-operative (Specify*)
Intra-operative (Neuro)
Laparoscopic
Pediatric
Small Organ (Specify**)
Neonatal Cephalic
Adult Cephalic
Trans-rectalPPPPPPNote 1, 3,4,6,7,8
Trans-vaginalPPPPPPNote 1, 3,4,6,7,8,14
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
CardiacCardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
PeripheralvesselPeripheral vesselOther (Specify***)
N=new indication; P=previously cleared by FDA(K173471); E=added under Appendix E
Additional comments: Combined modes--B+M、PW+B、Color + B、Power + B、PW+Color+B、Power + PW +B.
*Intraoperative includes abdominal, thoracic, and vascular
**Small organ-breast, thyroid, testes.
***Other use includes Urology.
Note 1: Tissue Harmonic Imaging. The feature does not use contrast agents.
Note 2: Smart3D
Note 3:4D(Real-time 3D)
Note 4: iScape
Note 5: TDI
Note 6: Color M
Note 7: Biopsy Guidance
Note 8: Strain Elastography
Note 9: Contrast imaging(contrast agent for LVO)
Note 10: Contrast imaging (Contrast agent for Liver)
Note 11: STE
Note 12: STQ
Note 13: Ultrasound Fusion Imaging
Note 14: GYN/Pelvic
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE NEEDED)Concurrence of CDRH, Office of Device Evaluation(ODE)

{4}------------------------------------------------

{5}------------------------------------------------

{6}------------------------------------------------

{7}------------------------------------------------

{8}------------------------------------------------

{9}------------------------------------------------

{10}------------------------------------------------

{11}------------------------------------------------

{12}------------------------------------------------

{13}------------------------------------------------

System:DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85/DC-86/DC-86S/DC-89/DC-TV/DC-TQ Diagnostic Ultrasound
System
Transducer:SP5-1E
Intended Use:Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation
General (Track1 Only)Specific (Track 1 & 3)BMPWDCWDColorDopplerAmplitudee DopplerCombined(specify)Other (specify)
OphthalmicOphthalmic
Fetal Imaging& OtherFetal
AbdominalPPPPPPPNote 1, 2,4, 6,7,13
Intra-operative (Specify*)
Intra-operative (Neuro)
Laparoscopic
PediatricPPPPPPPNote 1, 2,4, 6,7,13
Small Organ (Specify**)
Neonatal Cephalic
Adult CephalicPPPPPPPNote 1, 2,4, 6,7,13
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
CardiacCardiac AdultPPPPPPPNote 1, 2,4,5,6,7,9,13
Cardiac PediatricPPPPPPPNote 1, 2,4,5,6,7,13
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
PeripheralvesselPeripheral vessel
Other (Specify***)
N=new indication; P=previously cleared by FDA(K173471);E=added under Appendix E
Additional comments: Combined modes--B+M、PW+B、Color + B、Power + B、Power + B、PW+Color+B、Power + PW +B.
*Intraoperative includes abdominal, thoracic, and vascular
**Small organ-breast, thyroid, testes.
Other use includes Urology.
Note 1: Tissue Harmonic Imaging. The feature does not use contrast agents
Note 2: Smart3D
Note 3:4D(Real-time 3D)
Note 4: iScape
Note 5: TDI
Note 6: Color M
Note 7: Biopsy Guidance
Note 8: Strain Elastography
Note 9: Contrast imaging(contrast agent for LVO)
Note 10: Contrast imaging (Contrast agent for Liver)
Note 11: STE
Note 12: STQ
Note 13: Ultrasound Fusion Imaging
Note 14: GYN/Pelvic
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE NEEDED)
Concurrence of CDRH, Office of Device Evaluation(ODE)
System:DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85/DC-86/DC-86S/DC-89/DC-TV/DC-TQ Diagnostic Ultrasound System
Transducer:SC5-1E
Intended Use:Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation
General (Track 1 Only)Specific (Track 1 & 3)BMPWDCWDColorDopplerAmplitude DopplerCombin ed(specify)Other (specify)
OphthalmicOphthalmic
Fetal Imaging & OtherFetalPPPPPPNote 1, 2, 4,6,7,13
AbdominalPPPPPPNote 1, 2,4,6,7,10,11,12,13,14
Intra-operative (Specify*)
Intra-operative (Neuro)
Laparoscopic
PediatricPPPPPPNote 1, 2, 4,6,7,13
Small Organ (Specify**)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)PPPPPPNote 1, 2, 4,6,7,13
Musculo-skeletal(Superficial)
Intravascular
CardiacCardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Peripheral vesselPeripheral vesselPPPPPPNote 1, 2, 4,6,7,13
Other (Specify***)
N=new indication; P=previously cleared by FDA(K173471); E=added under Appendix E
Additional comments: Combined modes--B+M、PW+B、Color + B、Power + B、PW+Color+B、Power + PW +B
*Intraoperative includes abdominal, thoracic, and vascular
**Small organ-breast, thyroid, testes.
***Other use includes Urology.
Note 1: Tissue Harmonic Imaging. The feature does not use contrast agents.
Note 2: Smart3D
Note 3:4D(Real-time 3D)
Note 4: iScape
Note 5: TDI
Note 6: Color M
Note 7: Biopsy Guidance
Note 8: Strain Elastography
Note 9: Contrast imaging(contrast agent for LVO)
Note 10: Contrast imaging (Contrast agent for Liver)
Note 11: STE
Note 12: STQ
Note 13: Ultrasound Fusion Imaging
Note 14: GYN/Pelvic
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE NEEDED)
Concurrence of CDRH, Office of Device Evaluation(ODE)
System:DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85/DC-86/DC-86S/DC-89/DC-TV/DC-TQ Diagnostic UltrasoundSystem
Transducer:D6-2NE
Intended Use:Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation
General (Track1 Only)Specific (Track 1 & 3)BMPWDCWDColorDopplerAmplitude DopplerCombined (specify)Other (specify)
OphthalmicFetal Imaging& OtherOphthalmic
FetalPPPPPPNote1,3, 4,6
AbdominalPPPPPPNote1,3, 4,6,14
Intra-operative (Specify*)
Intra-operative (Neuro)
Laparoscopic
PediatricPPPPPPNote1,3, 4,6
Small Organ (Specify**)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
CardiacCardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
PeripheralvesselPeripheral vessel
Other (Specify***)
N=new indication; P=previously cleared by FDA(K173471); E=added under Appendix E
Additional comments: Combined modes--B+M、PW+B、Color + B、Power + B、PW +Color+B、Power + PW +B.
*Intraoperative includes abdominal, thoracic, and vascular
**Small organ-breast, thyroid, testes.
***Other use includes Urology.
Note 1: Tissue Harmonic Imaging. The feature does not use contrast agents.
Note 2: Smart3D
Note 3:4D(Real-time 3D)
Note 4: iScape
Note 5: TDI
Note 6: Color M
Note 7: Biopsy Guidance
Note 8: Strain Elastography
Note 9: Contrast imaging(contrast agent for LVO)
Note 10: Contrast imaging (Contrast agent for Liver)
Note 11: STE
Note 12: STQ
Note 13: Ultrasound Fusion Imaging
Note 14: GYN/Pelvic
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE NEEDED)
Concurrence of CDRH, Office of Device Evaluation(ODE)
System:DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85/DC-86/DC-86S/DC-89/DC-TV/DC-TQ Diagnostic Ultrasound System
Transducer:C6-2GE
Intended Use:Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation
General (Track Specific (Track 1 & 3) 1 Only)BMPWDCWDColorDopplerAmplitude DopplerCombined(specify)Other (specify)
OphthalmicOphthalmic
Fetal Imaging & OtherFetalPPPPPPNote1,2,4,6,7
AbdominalPPPPPPNote1,2,4,6,7, 10 , 14
Intra-operative (Specify*)
Intra-operative (Neuro)
Laparoscopic
PediatricPPPPPPNote1,2,4,6,7
Small Organ (Specify**)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
CardiacCardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
PeripheralvesselPeripheral vessel
Other (Specify***)
N=new indication; P=previously cleared by FDA(K173471); E=added under Appendix E
Additional comments: Combined modes--B+M、PW+B、Color + B、Power + B、PW+Color+B、Power + PW +B.
*Intraoperative includes abdominal, thoracic, and vascular
**Small organ-breast, thyroid, testes.
***Other use includes Urology.
Note 1: Tissue Harmonic Imaging. The feature does not use contrast agents.
Note 2: Smart3D
Note 3:4D(Real-time 3D)
Note 4: iScape
Note 5: TDI
Note 6: Color M
Note 7: Biopsy Guidance
Note 8: Strain Elastography
Note 9: Contrast imaging(contrast agent for LVO)
Note 10: Contrast imaging (Contrast agent for Liver)
Note 11: STE
Note 12: STQ
Note 13: Ultrasound Fusion Imaging
Note 14: GYN/Pelvic
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE NEEDED)
Concurrence of CDRH, Office of Device Evaluation(ODE)
System:DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85/DC-86/DC-86S/DC-89/DC-TV/DC-TQ Diagnostic UltrasoundSystem
Transducer:C11-3E
Intended Use:Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation
General (Track 1 Only)Specific (Track 1 & 3)BMPWDCWDColorDopplerAmplitudeDopplerCombined(specify)Other (specify)
OphthalmicOphthalmic
Fetal Imaging& OtherFetal
AbdominalPPPPPPNote 1, 2, 4,6,7
Intra-operative (Specify*)
Intra-operative (Neuro)
Laparoscopic
PediatricPPPPPPNote 1, 2, 4,6,7
Small Organ (Specify**)
Neonatal CephalicPPPPPPNote 1, 2, 4,6,7
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
CardiacCardiac AdultPPPPPPNote 1, 2, 4,6,7
Cardiac PediatricPPPPPPNote 1, 2, 4,6,7
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
PeripheralvesselPeripheral vesselPPPPPPNote 1, 2, 4,6,7
Other (Specify***)
N=new indication; P=previously cleared by FDA(K173471); E=added under Appendix E
Additional comments: Combined modes--B+M、PW+B、Color + B、Power + B、PW+Color+B、Power + PW +B.
*Intraoperative includes abdominal, thoracic, and vascular
**Small organ-breast, thyroid, testes.
***Other use includes Urology.
Note 1: Tissue Harmonic Imaging. The feature does not use contrast agents.
Note 2: Smart3D
Note 3:4D(Real-time 3D)
Note 4: iScape
Note 5: TDI
Note 6: Color M
Note 7: Biopsy Guidance
Note 8: Strain Elastography
Note 9: Contrast imaging(contrast agent for LVO)
Note 10: Contrast imaging (Contrast agent for Liver)
Note 11: STE
Note 12: STQ
Note 13: Ultrasound Fusion Imaging
Note 14: GYN/Pelvic
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE NEEDED)
Concurrence of CDRH, Office of Device Evaluation(ODE)
System:DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85/DC-86/DC-86S/DC-89/DC-TV/DC-TQ Diagnostic Ultrasound System
Transducer:L14-6NE
Intended Use:Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation
General (Track 1 Only)Specific (Track 1 & 3)BMPWDCWDColor DopplerAmplitude DopplerCombined (specify)Other (specify)
OphthalmicOphthalmic
Fetal Imaging & OtherFetal
AbdominalPPPPPPNote 1,2, 4,7
Intra-operative (Specify*)
Intra-operative (Neuro)
Laparoscopic
PediatricPPPPPPNote 1,2, 4,7
Small Organ (Specify**)PPPPPPNote 1,2, 4,7,8
Neonatal CephalicPPPPPPNote 1,2, 4,7
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Conventional)PPPPPPNote 1,2, 4,7
Musculo-skeletal (Superficial)PPPPPPNote 1,2, 4,7
Intravascular
CardiacCardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Peripheral vesselPeripheral vesselPPPPPPNote 1,2, 4,7
Other (Specify***)
N=new indication; P=previously cleared by FDA(K173471);E=added under Appendix E
Additional comments: Combined modes--B+M、PW+B、Color + B、Power + B、PW +Color+B、Power + PW +B.
*Intraoperative includes abdominal, thoracic, and vascular
**Small organ-breast, thyroid, testes.
***Other use includes Urology.
Note 1: Tissue Harmonic Imaging. The feature does not use contrast agents.
Note 2: Smart3D
Note 3:4D(Real-time 3D)
Note 4: iScape
Note 5: TDI
Note 6: Color M
Note 7: Biopsy Guidance
Note 8: Strain Elastography
Note 9: Contrast imaging(contrast agent for LVO)
Note 10: Contrast imaging (Contrast agent for Liver)
Note 11: STE
Note 12: STQ
Note 13: Ultrasound Fusion Imaging
Note 14: GYN/Pelvic
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE NEEDED)
System:DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85/DC-86/DC-86S/DC-89/DC-TV/DC-TQ Diagnostic UltrasoundSystem
Transducer:L14-6WE
Intended Use:Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation
General (Track 1 Only)Specific (Track 1 & 3)BMPWDCWDColorDopplerAmplitudee DopplerCombined(specify)Other (specify)
OphthalmicOphthalmic
Fetal Imaging& OtherFetal
AbdominalPPPPPPNote 1,2, 4,7
Intra-operative (Specify*)
Intra-operative (Neuro)
Laparoscopic
PediatricPPPPPPNote 1,2, 4,7
Small Organ (Specify**)PPPPPPNote 1,2, 4,7,8
Neonatal CephalicPPPPPPNote 1,2, 4,7
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)PPPPPPNote 1,2, 4,7
Musculo-skeletal(Superficial)PPPPPPNote 1,2, 4,7
Intravascular
CardiacCardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
PeripheralvesselPeripheral vesselPPPPPPNote 1,2, 4,7
Other (Specify***)
N=new indication; P=previously cleared by FDA(K173471); E=added under Appendix E
Additional comments: Combined modes--B+M、PW+B、Color + B、Power + B、PW+Color+B、Power + PW +B.
*Intraoperative includes abdominal, thoracic, and vascular
**Small organ-breast, thyroid, testes.
***Other use includes Urology.
Note 1: Tissue Harmonic Imaging. The feature does not use contrast agents.
Note 2: Smart3D
Note 3:4D(Real-time 3D)
Note 4: iScape
Note 5: TDI
Note 6: Color M
Note 7: Biopsy Guidance
Note 8: Strain Elastography
Note 9: Contrast imaging(contrast agent for LVO)
Note 10: Contrast imaging (Contrast agent for Liver)
Note 11: STE
Note 12: STQ
Note 13: Ultrasound Fusion Imaging
Note 14: GYN/Pelvic
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE NEEDED)
System:DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-86/DC-86/DC-865/DC-89/DC-TV/DC-TQ Diagnostic Ultrasound
System
Transducer:V11-3E
Intended Use:Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation
General (Track Specific (Track 1 & 3)BMPWDCWD ColorAmplitudCombinOther (specify)
1 Only)Dopplere Dopplered
(specify)
OphthalmicOphthalmic
Fetal ImagingFetalPPPPPPNote 1, 2, 4,6,7
& OtherAbdominal
Intra-operative (Specify*)
Intra-operative (Neuro)
Laparoscopic
Pediatric
Small Organ (Specify**)
Neonatal Cephalic
Adult Cephalic
Trans-rectalPPPPPPNote 1, 2, 4,6,7,8
Trans-vaginalPPPPPPNote 1, 2, 4,6,7,8,14
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal
(Conventional)
Musculo-skeletal
(Superficial)
Intravascular
CardiacCardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
PeripheralPeripheral vessel
vesselOther (Specify***)PPPPPPNote 1, 2, 4,6,7
N=new indication; P=previously cleared by FDA(K173471);E=added under Appendix E
Additional comments: Combined modes-B+M、PW+B、Color + B、PW +Color+B、PW +Color+B、PW+B.
*Intraoperative includes abdominal, thoracic, and vascular
**Small organ-breast, thyroid, testes.
***Other use includes Urology.
Note 1: Tissue Harmonic Imaging. The feature does not use contrast agents
Note 2: Smart3D
Note 3:4D(Real-time 3D)
Note 4: iScape
Note 5: TDI
Note 6: Color M
Note 7: Biopsy Guidance
Note 8: Strain Elastography
Note 9: Contrast imaging(contrast agent for LVO)
Note 10: Contrast imaging (Contrast agent for Liver)
Note 11: STE
Note 12: STQ
Note 13: Ultrasound Fusion Imaging
Note 14: GYN/Pelvic
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE NEEDED)
Concurrence of CDRH, Office of Device Evaluation(ODE)

{14}------------------------------------------------

{15}------------------------------------------------

{16}------------------------------------------------

{17}------------------------------------------------

{18}------------------------------------------------

{19}------------------------------------------------

{20}------------------------------------------------

{21}------------------------------------------------

System:DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85/DC-86/DC-86S/DC-89/DC-TV/DC-TQ Diagnostic UltrasoundSystem
Transducer:V11-3BE
Intended Use:Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationGeneral (Track Specific (Track 1 & 3)1 Only)BMPWDCWDColorDopplerAmplitude DopplerCombined(specify)Other (specify)
OphthalmicOphthalmic
Fetal Imaging& OtherFetalPPPPPPNote 1, 2, 4,6,7
Abdominal
Intra-operative (Specify*)
Intra-operative (Neuro)
Laparoscopic
Pediatric
Small Organ (Specify**)
Neonatal Cephalic
Adult Cephalic
Trans-rectalPPPPPPNote 1, 2, 4,6,7,8
Trans-vaginalPPPPPPNote 1, 2, 4,6,7,8,14
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
CardiacCardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
PeripheralvesselPeripheral vessel
Other (Specify***)PPPPPPNote 1, 2, 4,6,7
N=new indication; P=previously cleared by FDA(K173471); E=added under Appendix E
Additional comments: Combined modes--B+M、PW+B、Color + B、Power + B、PW+Color+B、Power + PW +B.
*Intraoperative includes abdominal, thoracic, and vascular
**Small organ-breast, thyroid, testes.
***Other use includes Urology.
Note 1: Tissue Harmonic Imaging. The feature does not use contrast agents.
Note 2: Smart3D
Note 3:4D(Real-time 3D)
Note 4: iScape
Note 5: TDI
Note 6: Color M
Note 7: Biopsy Guidance
Note 8: Strain Elastography
Note 9: Contrast imaging(contrast agent for LVO)
Note 10: Contrast imaging (Contrast agent for Liver)
Note 11: STE
Note 12: STQ
Note 13: Ultrasound Fusion Imaging
Note 14: GYN/Pelvic
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE NEEDED)
Concurrence of CDRH, Office of Device Evaluation(ODE)
System:DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85/DC-86/DC-86S/DC-89/DC-TV/DC-TQ Diagnostic Ultrasound
System
Transducer:DE10-3E
Intended Use:Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation
General (Track 1 Only)Specific (Track 1 & 3)BMPWDCWDColorDopplerAmplitude DopplerCombined (specify)Other (specify)
OphthalmicOphthalmic
Fetal Imaging& OtherFetalPPPPPPNote 1, 3,4,6,7
Abdominal
Intra-operative (Specify*)
Intra-operative (Neuro)
Laparoscopic
Pediatric
Small Organ (Specify**)
Neonatal Cephalic
Adult Cephalic
Trans-rectalPPPPPPNote 1, 3,4,6,7,8
Trans-vaginalPPPPPPNote 1, 3,4,6,7,8,14
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
CardiacCardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
PeripheralvesselPeripheral vessel
Other (Specify***)
N=new indication; P=previously cleared by FDA(K173471); E=added under Appendix E
Additional comments: Combined modes--B+M、PW+B、Color + B、Power + B、PW+Color+B、Power + PW +B.
*Intraoperative includes abdominal, thoracic, and vascular
**Small organ-breast, thyroid, testes.
***Other use includes Urology.
Note 1: Tissue Harmonic Imaging. The feature does not use contrast agents.
Note 2: Smart3D
Note 3:4D(Real-time 3D)
Note 4: iScape
Note 5: TDI
Note 6: Color M
Note 7: Biopsy Guidance
Note 8: Strain Elastography
Note 9: Contrast imaging(contrast agent for LVO)
Note 10: Contrast imaging (Contrast agent for Liver)
Note 11: STE
Note 12: STQ
Note 13: Ultrasound Fusion Imaging
Note 14: GYN/Pelvic
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE NEEDED)
Concurrence of CDRH, Office of Device Evaluation(ODE)
System:DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85/DC-86/DC-86S/DC-89/DC-TV/DC-TQ Diagnostic UltrasoundSystem
Transducer:P4-2NE
Intended Use:Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation
General (Track Specific (Track 1 & 3)1 Only)BMPWDCWDColorDopplerAmplitudee DopplerCombined(specify)Other (specify)
OphthalmicOphthalmic
Fetal Imaging& OtherFetal
AbdominalPPPPPPPNote 1, 2,4,6,7
Intra-operative (Specify*)
Intra-operative (Neuro)
Laparoscopic
PediatricPPPPPPPNote 1, 2,4,6,7
Small Organ (Specify**)
Neonatal Cephalic
Adult CephalicPPPPPPPNote 1, 2,4,6,7
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
CardiacCardiac AdultPPPPPPPNote 1, 2,4,5,6,7
Cardiac PediatricPPPPPPPNote 1, 2,4,5,6,7
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
PeripheralvesselPeripheral vessel
Other (Specify***)
N=new indication; P=previously cleared by FDA(K173471); E=added under Appendix E
Additional comments: Combined modes--B+M, PW+B, Color + B, Power + B, PW+Color+B, Power + PW+B.
*Intraoperative includes abdominal, thoracic, and vascular
**Small organ-breast, thyroid, testes.
***Other use includes Urology.)
Note 1: Tissue Harmonic Imaging. The feature does not use contrast agents.
Note 2: Smart3D
Note 3:4D(Real-time 3D)
Note 4: iScape
Note 5: TDI
Note 6: Color M
Note 7: Biopsy Guidance
Note 8: Strain Elastography
Note 9: Contrast imaging(contrast agent for LVO)
Note 10: Contrast imaging (Contrast agent for Liver)
Note 11: STE
Note 12: STQ
Note 13: Ultrasound Fusion Imaging
Note 14: GYN/Pelvic
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE NEEDED)
Concurrence of CDRH, Office of Device Evaluation(ODE)
System:DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85/DC-86/DC-86S/DC-89/DC-TV/DC-TQ Diagnostic Ultrasound System
Transducer:P7-3E
Intended Use:Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation
General (Track Specific (Track 1 & 3) 1 Only)BMPWDCWDColorDopplerAmplitudeDopplerCombined(specify)Other (specify)
OphthalmicOphthalmic
Fetal Imaging& OtherFetal
AbdominalPPPPPPPNote 1, 2,4,6
Intra-operative (Specify*)
Intra-operative (Neuro)
Laparoscopic
PediatricPPPPPPPNote 1, 2.4.6
Small Organ (Specify**)
Neonatal CephalicPPPPPPPNote 1, 2,4,6
Adult CephalicPPPPPPPNote 1, 2,4,6
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)PPPPPPPNote 1, 2,4,6
Musculo-skeletal(Superficial)
Intravascular
CardiacCardiac AdultPPPPPPPNote 1, 2,4,5,6
Cardiac PediatricPPPPPPPNote 1, 2,4,5,6
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
PeripheralvesselPeripheral vessel
Other (Specify***)
N=new indication; P=previously cleared by FDA(K173471); E=added under Appendix E
Additional comments: Combined modes--B+M、PW+B、Color + B、Power + B、PW+Color+B、Power + PW +B.
*Intraoperative includes abdominal, thoracic, and vascular
**Small organ-breast, thyroid, testes.
***Other use includes Urology.
Note 1: Tissue Harmonic Imaging. The feature does not use contrast agents.
Note 2: Smart3D
Note 3:4D(Real-time 3D)
Note 4: iScape
Note 5: TDI
Note 6: Color M
Note 7: Biopsy Guidance
Note 8: Strain Elastography
Note 9: Contrast imaging(contrast agent for LVO)
Note 10: Contrast imaging (Contrast agent for Liver)
Note 11: STE
Note 12: STQ
Note 13: Ultrasound Fusion Imaging
Note 14: GYN/Pelvic
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE NEEDED)
Concurrence of CDRH, Office of Device Evaluation(ODE)
System:DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85/DC-86/DC-86S/DC-89/DC-TV/DC-TQ Diagnostic Ultrasound System
System
Transducer:P10-4E
Intended Use:Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation
General (Track Specific (Track 1 & 3)BMPWDCWDColorAmplitude DopplerCombined (specify)Other (specify)
1 Only)Dopplere Dopplered
(specify)
OphthalmicOphthalmic
Fetal ImagingFetal
& OtherAbdominalPPPPPPPNote 1, 2,4,6
Intra-operative (Specify*)
Intra-operative (Neuro)
Laparoscopic
PediatricPPPPPPPNote 1, 2,4,6
Small Organ (Specify**)
P
Neonatal CephalicPPPPPPPNote 1, 2,4,6
Adult CephalicPPPPPPPNote 1, 2,4,6
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal
(Conventional)
Musculo-skeletal
(Superficial)
Intravascular
CardiacCardiac AdultPPPPPPPNote 1, 2,4,5,6
Cardiac PediatricPPPPPPPNote 1, 2,4,5,6
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
PeripheralPeripheral vessel
vesselOther (Specify***)
N=new indication; P=previously cleared by FDA(K173471);E=added under Appendix E
Additional comments: Combined modes--B+M、PW+B、Color + B、Power + B、PW+Color+B、Power + PW +B.
*Intraoperative includes abdominal, thoracic, and vascular
**Small organ-breast, thyroid, testes.
***Other use includes Urology.
Note 1: Tissue Harmonic Imaging. The feature does not use contrast agents.
Note 2: Smart3D
Note 3:4D(Real-time 3D)
Note 4: iScape
Note 5: TDI
Note 6: Color M
Note 7: Biopsy Guidance
Note 8: Strain Elastography
Note 9: Contrast imaging(contrast agent for LVO)
Note 10: Contrast imaging (Contrast agent for Liver)
Note 11: STE
Note 12: STQ
Note 13: Ultrasound Fusion Imaging
Note 14: GYN/Pelvic
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE NEEDED)
Concurrence of CDRH, Office of Device Evaluation(ODE)

{22}------------------------------------------------

{23}------------------------------------------------

{24}------------------------------------------------

{25}------------------------------------------------

{26}------------------------------------------------

System:DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85/DC-86/DC-86S/DC-89/DC-TV/DC-TQ Diagnostic Ultrasound System
Transducer:CB10-4E
Intended Use:Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation
General (Track1 Only)Specific (Track 1 & 3)BMPWDCWDColorDopplerAmplitudee DopplerCombined(specify)Other (specify)
OphthalmicOphthalmic
Fetal Imaging& OtherFetal
Abdominal
Intra-operative (Specify*)
Intra-operative (Neuro)
Laparoscopic
Pediatric
Small Organ (Specify**)
Neonatal Cephalic
Adult Cephalic
Trans-rectalPPPPPPNote 1, 2, 4,6,7
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
CardiacCardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
PeripheralvesselPeripheral vessel
Other (Specify***)PPPPPPNote 1, 2, 4,6,7
N=new indication; P=previously cleared by FDA(K173471); E=added under Appendix E
Additional comments: Combined modes--B+M、PW+B、Color + B、Power + B、PW+Color+B、Power + PW +B.
*Intraoperative includes abdominal, thoracic, and vascular
**Small organ-breast, thyroid, testes.
***Other use includes Urology.
Note 1: Tissue Harmonic Imaging. The feature does not use contrast agents.
Note 2: Smart3D
Note 3:4D(Real-time 3D)
Note 4: iScape
Note 5: TDI
Note 6: Color M
Note 7: Biopsy Guidance
Note 8: Strain Elastography
Note 9: Contrast imaging(contrast agent for LVO)
Note 10: Contrast imaging (Contrast agent for Liver)
Note 11: STE
Note 12: STQ
Note 13: Ultrasound Fusion Imaging
Note 14: GYN/Pelvic
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE NEEDED)
System:DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85/DC-86/DC-86S/DC-89/DC-TV/DC-TQ Diagnostic Ultrasound System
Transducer:L16-4HE
Intended Use:Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation
General (Track 1 Only)Specific (Track 1 & 3)BMPWDCWDColor DopplerAmplitude DopplerCombined (specify)Other (specify)
OphthalmicOphthalmic
Fetal Imaging & OtherFetal
AbdominalPPPPPPPNote1,2,4
Intra-operative (Specify*)PPPPPPPNote1,2,4
Intra-operative (Neuro)
Laparoscopic
PediatricPPPPPPPNote1,2,4
Small Organ (Specify**)PPPPPPPNote1,2,4
Neonatal CephalicPPPPPPPNote1,2,4
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Conventional)PPPPPPPNote1,2,4
Musculo-skeletal (Superficial)PPPPPPPNote1,2,4
Intravascular
CardiacCardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Peripheral vesselPeripheral vesselPPPPPPPNote1,2,4
Other (Specify***)
N=new indication; P=previously cleared by FDA(K173471); E=added under Appendix E
Additional comments: Combined modes--B+M、PW+B、Color + B、Power + B、PW+Color+B、Power + PW +B.
*Intraoperative includes abdominal, thoracic, and vascular
**Small organ-breast, thyroid, testes.
***Other use includes Urology.
Note 1: Tissue Harmonic Imaging. The feature does not use contrast agents
Note 2: Smart3D
Note 3:4D(Real-time 3D)
Note 4: iScape
Note 5: TDI
Note 6: Color M
Note 7: Biopsy Guidance
Note 8: Strain Elastography
Note 9: Contrast imaging(contrast agent for LVO)
Note 10: Contrast imaging (Contrast agent for Liver)
Note 11: STE
Note 12: STQ
Note 13: Ultrasound Fusion Imaging
Note 14: GYN/Pelvic
System:DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85/DC-86/DC-86S/DC-89/DC-TV/DC-TQ Diagnostic UltrasoundSystem
Transducer:P7-3TE
Intended Use:Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation
General (Track1 Only)Specific (Track 1 & 3)BMPWDCWDColorDopplerAmplitude DopplerCombined(specify)Other (specify)
Ophthalmic& OtherOphthalmic
Fetal ImagingFetal
Abdominal
Intra-operative (Specify*)
Intra-operative (Neuro)
Laparoscopic
Pediatric
Small Organ (Specify**)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
CardiacCardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)PPPPPPPNote 1, 5,6
Intra-cardiac
PeripheralvesselPeripheral vessel
Other (Specify***)
N=new indication; P=previously cleared by FDA(K173471); E=added under Appendix E
Additional comments: Combined modes--B+M、PW+B、Color + B、Power + B、PW+Color+B、Power + PW +B.
*Intraoperative includes abdominal, thoracic, and vascular
**Small organ-breast, thyroid, testes.
***Other use includes Urology.
Note 1: Tissue Harmonic Imaging. The feature does not use contrast agents.
Note 2: Smart3D
Note 3:4D(Real-time 3D)
Note 4: iScape
Note 5: TDI
Note 6: Color M
Note 7: Biopsy Guidance
Note 8: Strain Elastography
Note 9: Contrast imaging(contrast agent for LVO)
Note 10: Contrast imaging (Contrast agent for Liver)
Note 11: STE
Note 12: STQ
Note 13: Ultrasound Fusion Imaging
Note 14: GYN/Pelvic
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE NEEDED)
System:DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85/DC-86/DC-86S/DC-89/DC-TV/DC-TQ Diagnostic Ultrasound System
Transducer:CW5s
Intended Use:Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation
General (Track 1 Only) Specific (Track 1 & 3)BMPWDCWDColor DopplerAmplitude DopplerCombined (specify)Other (specify)
OphthalmicOphthalmic
Fetal Imaging & OtherFetal
Abdominal
Intra-operative (Specify*)
Intra-operative (Neuro)
Laparoscopic
PediatricP
Small Organ (Specify**)
Neonatal Cephalic
Adult CephalicP
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Conventional)
Musculo-skeletal (Superficial)
Intravascular
CardiacCardiac AdultP
Cardiac PediatricP
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Peripheral vesselPeripheral vesselN
Other (Specify***)
N=new indication; P=previously cleared by FDA(K173471); E=added under Appendix E
Additional comments: Combined modes--B+M、PW+B、Color + B、Power + B、PW +Color+B、Power + PW +B.
*Intraoperative includes abdominal, thoracic, and vascular
**Small organ-breast, thyroid, testes.
***Other use includes Urology.
Note 1: Tissue Harmonic Imaging. The feature does not use contrast agents.
Note 2: Smart3D
Note 3:4D(Real-time 3D)
Note 4: iScape
Note 5: TDI
Note 6: Color M
Note 7: Biopsy Guidance
Note 8: Strain Elastography
Note 9: Contrast imaging(contrast agent for LVO)
Note 10: Contrast imaging (Contrast agent for Liver)
Note 11: STE
Note 12: STQ
Note 13: Ultrasound Fusion Imaging
Note 14: GYN/Pelvic
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE NEEDED)
System:DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85/DC-86/DC-86S/DC-89/DC-TV/DC-TQ Diagnostic Ultrasound System
Transducer:C5-2E
Intended Use:Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation
General (Track 1 Only)Specific (Track 1 & 3)BMPWDCWDColor DopplerAmplitude DopplerCombined (specify)Other (specify)
OphthalmicOphthalmic
Fetal Imaging & OtherFetalNNNNNNNote 1, 2, 4,6,7
AbdominalNNNNNNNote 1, 2, 4,6,7,14
Intra-operative (Specify*)
Intra-operative (Neuro)
Laparoscopic
PediatricNNNNNNNote 1, 2, 4,6,7
Small Organ (Specify**)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Conventional)NNNNNNNote 1, 2, 4,6,7
Musculo-skeletal (Superficial)
Intravascular
CardiacCardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Peripheral vesselPeripheral vesselNNNNNNNote 1, 2, 4,6,7
vesselOther (Specify***)
N=new indication; P=previously cleared by FDA; E=added under Appendix E
Additional comments: Combined modes--B+M, PW+B, Color + B, Power + B, PW +Color+B, Power + PW +B.
*Intraoperative includes abdominal, thoracic, and vascular
**Small organ-breast, thyroid, testes.
***Other use includes Urology.
Note 1: Tissue Harmonic Imaging. The feature does not use contrast agents.
Note 2: Smart3D
Note 3: 4D(Real-time 3D)
Note 4: iScape
Note 5: TDI
Note 6: Color M
Note 7: Biopsy Guidance
Note 8: Strain Elastography
Note 9: Contrast imaging(contrast agent for LVO)
Note 10: Contrast imaging (Contrast agent for Liver)
Note 11: STE
Note 12: STQ
Note 13: Ultrasound Fusion Imaging
Note 14: GYN/Pelvic
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE NEEDED)
Concurrence of CDRH, Office of Device Evaluation(ODE)

{27}------------------------------------------------

{28}------------------------------------------------

{29}------------------------------------------------

{30}------------------------------------------------

{31}------------------------------------------------

System:DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85/DC-86/DC-86S/DC-89/DC-TV/DC-TQ Diagnostic Ultrasound
System
Transducer:P4-2E
Intended Use:Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation
General (Track Specific (Track 1 & 3)BMPWDCWDColorDopplerAmplitude DopplerCombined(specify)Other (specify)
1 Only)e
OphthalmicOphthalmic
Fetal Imaging& OtherFetal
AbdominalNNNNNNNNote 1,2,4,6,7
Intra-operative (Specify*)
Intra-operative (Neuro)
Laparoscopic
PediatricNNNNNNNNote 1,2,4,6,7
Small Organ (Specify**)
Neonatal CephalicNNNNNNNNote 1,2,4,6,7
Adult CephalicNNNNNNNNote 1,2,4,6,7
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
CardiacCardiac AdultNNNNNNNNote 1,2,4,5,6,7
Cardiac PediatricNNNNNNNNote 1,2,4,5,6,7
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
PeripheralvesselPeripheral vessel
Other (Specify***)
N=new indication; P=previously cleared by FDA;E=added under Appendix E
Additional comments: Combined modes--B+M、PW+B、Color + B、Power + B、PW+Color+B、Power + PW +B.
*Intraoperative includes abdominal, thoracic, and vascular
**Small organ-breast, thyroid, testes.
***Other use includes Urology.
Note 1: Tissue Harmonic Imaging. The feature does not use contrast agents.
Note 2: Smart3D
Note 3:4D(Real-time 3D)
Note 4: iScape
Note 5: TDI
Note 6: Color M
Note 7: Biopsy Guidance
Note 8: Strain Elastography
Note 9: Contrast imaging(contrast agent for LVO)
Note 10: Contrast imaging (Contrast agent for Liver)
Note 11: STE
Note 12: STQ
Note 13: Ultrasound Fusion Imaging
Note 14: GYN/Pelvic
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE NEEDED)
Concurrence of CDRH, Office of Device Evaluation(ODE)
System:DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85/DC-86/DC-86S/DC-89/DC-TV/DC-TQ Diagnostic Ultrasound
System
Transducer:D6-2E
Intended Use:Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation
General (Track 1 Only)Specific (Track 1 & 3)BMPWDCWDColorDopplerAmplitude DopplerCombined(specify)Other (specify)
OphthalmicOphthalmic
Fetal Imaging& OtherFetalNNNNNNNote 1,3,4,6,7
AbdominalNNNNNNNote 1,3,4,6,7,14
Intra-operative (Specify*)
Intra-operative (Neuro)
Laparoscopic
PediatricNNNNNNNote 1,3,4,6,7
Small Organ (Specify**)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
CardiacCardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
PeripheralvesselPeripheral vessel
Other (Specify***)
N=new indication; P=previously cleared by FDA; E=added under Appendix E
Additional comments: Combined modes--B+M、PW+B、Color + B、Power + B、PW+Color+B、Power + PW +B.
*Intraoperative includes abdominal, thoracic, and vascular
**Small organ-breast, thyroid, testes.
***Other use includes Urology.
Note 1: Tissue Harmonic Imaging. The feature does not use contrast agents.
Note 2: Smart3D
Note 3:4D(Real-time 3D)
Note 4: iScape
Note 5: TDI
Note 6: Color M
Note 7: Biopsy Guidance
Note 8: Strain Elastography
Note 9: Contrast imaging(contrast agent for LVO)
Note 10: Contrast imaging (Contrast agent for Liver)
Note 11: STE
Note 12: STQ
Note 13: Ultrasound Fusion Imaging
Note 14: GYN/Pelvic
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE NEEDED)
Concurrence of CDRH, Office of Device Evaluation(ODE)
System:DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85/DC-86/DC-86S/DC-89/DC-TV/DC-TQ Diagnostic UltrasoundSystem
Transducer:SC6-1E
Intended Use:Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation
General (Track 1 Only)Specific (Track 1 & 3)BMPWDCWDColorDopplerAmplitudeDopplerCombined(specify)Other (specify)
OphthalmicOphthalmic
Fetal Imaging& OtherFetalNNNNNNNote 1,2,4,6,7,13
AbdominalNNNNNNNote1,2,4,6,7,10,11,12,13,14
Intra-operative (Specify*)
Intra-operative (Neuro)
Laparoscopic
PediatricNNNNNNNote 1,2,4,6,7,13
Small Organ (Specify**)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)NNNNNNNote 1,2,4,6,7,13
Musculo-skeletal(Superficial)
Intravascular
CardiacCardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
PeripheralvesselPeripheral vesselNNNNNNNote 1,2,4,6,7,13
Other (Specify***)
System:DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85/DC-86/DC-86S/DC-89/DC-TV/DC-TQ Diagnostic Ultrasound
System
Transducer:L20-5E
Intended Use:Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation
General (Track1 Only)Specific (Track 1 & 3)BMPWDCWDColorDopplerAmplitude DopplerCombined(specify)Other (specify)
OphthalmicOphthalmic
Fetal Imaging& OtherFetal
Abdominal
Intra-operative (Specify*)
Intra-operative (Neuro)
Laparoscopic
Pediatric
Small Organ (Specify**)NNNNNNNote 1,2,4
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)NNNNNNNote 1,2,4,8
Musculo-skeletal(Superficial)NNNNNNNote 1,2,4,8
Intravascular
CardiacCardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
PeripheralvesselPeripheral vesselNNNNNNNote 1,2,4
Other (Specify***)
N=new indication; P=previously cleared by FDA; E=added under Appendix E
Additional comments: Combined modes--B+M、PW+B、Color + B、Power + B、PW+Color+B、Power + PW +B.
*Intraoperative includes abdominal, thoracic, and vascular
**Small organ-breast, thyroid, testes.
***Other use includes Urology.
Note 1: Tissue Harmonic Imaging. The feature does not use contrast agents.
Note 2: Smart3D
Note 3:4D(Real-time 3D)
Note 4: iScape
Note 5: TDI
Note 6: Color M
Note 7: Biopsy Guidance
Note 8: Strain Elastography
Note 9: Contrast imaging(contrast agent for LVO)
Note 10: Contrast imaging (Contrast agent for Liver)
Note 11: STE
Note 12: STQ
Note 13: Ultrasound Fusion Imaging
Note 14: GYN/Pelvic
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE NEEDED)
Concurrence of CDRH, Office of Device Evaluation(ODE)
System:DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85/DC-86/DC-86S/DC-89/DC-TV/DC-TQ Diagnostic UltrasoundSystem
Transducer:SD8-1E
Intended Use:Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation
General (Track1 Only)Specific (Track 1 & 3)BMPWDCWDColorDopplerAmplitude DopplerCombined(specify)Other (specify)
OphthalmicOphthalmic
Fetal Imaging& OtherFetalNNNNNNNote 1, 3, 4, 6,7
AbdominalNNNNNNNote 1, 3, 4, 6,7,14
Intra-operative (Specify*)
Intra-operative (Neuro)
Laparoscopic
Pediatric
Small Organ (Specify**)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
CardiacCardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
PeripheralvesselPeripheral vessel
Other (Specify***)
N=new indication; P=previously cleared by FDA; E=added under Appendix E
Additional comments: Combined modes--B+M、PW+B、Color + B、Power + B、PW +Color+B、Power + PW +B.
*Intraoperative includes abdominal, thoracic, and vascular
**Small organ-breast, thyroid, testes.
***Other use includes Urology.
Note 1: Tissue Harmonic Imaging. The feature does not use contrast agents.
Note 2: Smart3D
Note 3:4D(Real-time 3D)
Note 4: iScape
Note 5: TDI
Note 6: Color M
Note 7: Biopsy Guidance
Note 8: Strain Elastography
Note 9: Contrast imaging(contrast agent for LVO)
Note 10: Contrast imaging (Contrast agent for Liver)
Note 11: STE
Note 12: STQ
Note 13: Ultrasound Fusion Imaging
Note 14: GYN/Pelvic
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE NEEDED)
Concurrence of CDRH, Office of Device Evaluation(ODE)
System:DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85/DC-86/DC-86S/DC-89/DC-TV/DC-TQ Diagnostic UltrasoundSystem
Transducer:DE10-3WE
Intended Use:Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation
General (Track Specific (Track 1 & 3)BMPWDCWDColorDopplerAmplitude DopplerCombined(specify)Other (specify)
1 Only)
(specify)
OphthalmicOphthalmic
Fetal Imaging& OtherFetalNNNNNNNote 1, 3, 4, 6,7
Abdominal
Intra-operative (Specify*)
Intra-operative (Neuro)
Laparoscopic
Pediatric
Small Organ (Specify**)
Neonatal Cephalic
Adult Cephalic
Trans-rectalNNNNNNNote 1, 3, 4, 6,7
Trans-vaginalNNNNNNNote 1, 3, 4, 6,7,14
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
CardiacCardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
PeripheralvesselPeripheral vessel
Other (Specify***)
N=new indication; P=previously cleared by FDA; E=added under Appendix E
Additional comments: Combined modes--B+M、PW+B、Color + B、Power + B、PW+Color+B、Power + PW +B.
*Intraoperative includes abdominal, thoracic, and vascular
**Small organ-breast, thyroid, testes.
***Other use includes Urology.
Note 1: Tissue Harmonic Imaging. The feature does not use contrast agents.
Note 2: Smart3D
Note 3:4D(Real-time 3D)
Note 4: iScape
Note 5: TDI
Note 6: Color M
Note 7: Biopsy Guidance
Note 8: Strain Elastography
Note 9: Contrast imaging(contrast agent for LVO)
Note 10: Contrast imaging (Contrast agent for Liver)
Note 11: STE
Note 12: STQ
Note 13: Ultrasound Fusion Imaging
Note 14: GYN/Pelvic
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE NEEDED)
System:DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85/DC-86/DC-86S/DC-89/DC-TV/DC-TQ Diagnostic Ultrasound
System
Transducer:V11-3WE
Intended Use:Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:
Mode of Operation
General (Track 1 Only)Specific (Track 1 & 3)BMPWDCWDColorDopplerAmplitude DopplerCombined(specify)Other (specify)
OphthalmicOphthalmic
Fetal Imaging& OtherFetalNNNNNNNote 1,2,4,6,7
Abdominal
Intra-operative (Specify*)
Intra-operative (Neuro)
Laparoscopic
Pediatric
Small Organ (Specify**)
Neonatal Cephalic
Adult Cephalic
Trans-rectalNNNNNNNote 1,2,4,6,7,8
Trans-vaginalNNNNNNNote 1,2,4,6,7,8
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
CardiacCardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
PeripheralvesselPeripheral vessel
Other (Specify***)NNNNNNNote 1,2,4,6,7
N=new indication; P=previously cleared by FDA; E=added under Appendix E
Additional comments: Combined modes--B+M、PW+B、Color + B、Power + B、PW+Color+B、Power + PW +B.
*Intraoperative includes abdominal, thoracic, and vascular
**Small organ-breast, thyroid, testes.
***Other use includes Urology.
Note 1: Tissue Harmonic Imaging. The feature does not use contrast agents.
Note 2: Smart3D
Note 3:4D(Real-time 3D)
Note 4: iScape
Note 5: TDI
Note 6: Color M
Note 7: Biopsy Guidance
Note 8: Strain Elastography
Note 9: Contrast imaging(contrast agent for LVO)
Note 10: Contrast imaging (Contrast agent for Liver)
Note 11: STE
Note 12: STQ
Note 13: Ultrasound Fusion Imaging
Note 14: GYN/Pelvic
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE NEEDED)
Concurrence of CDRH, Office of Device Evaluation(ODE)
System:DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85/DC-86/DC-86S/DC-89/DC-TV/DC-TQ Diagnostic Ultrasound System
Transducer:SC5-1NE
Intended Use:Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation
General (Track 1 Only)Specific (Track 1 & 3)BMPWDCWDColorDopplerAmplitude DopplerCombined(specify)Other (specify)
OphthalmicOphthalmic
Fetal Imaging& OtherFetalNNNNNNNote 1,2,4,6,7
AbdominalNNNNNNNote 1,2,4,6,7,14
Intra-operative (Specify*)
Intra-operative (Neuro)
Laparoscopic
PediatricNNNNNNNote 1,2,4,6,7
Small Organ (Specify**)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)NNNNNNNote 1,2,4,6,7
Musculo-skeletal(Superficial)
Intravascular
CardiacCardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
PeripheralvesselPeripheral vesselNNNNNNNote 1,2,4,6,7
Other (Specify***)
N=new indication; P=previously cleared by FDA; E=added under Appendix E
Additional comments: Combined modes--B+M、PW+B、Color + B、Power + B、PW +Color+B、Power + PW +B.
*Intraoperative includes abdominal, thoracic, and vascular
**Small organ-breast, thyroid, testes.
***Other use includes Urology.
Note 1: Tissue Harmonic Imaging. The feature does not use contrast agents.
Note 2: Smart3D
Note 3:4D(Real-time 3D)
Note 4: iScape
Note 5: TDI
Note 6: Color M
Note 7: Biopsy Guidance
Note 8: Strain Elastography
Note 9: Contrast imaging(contrast agent for LVO)
Note 10: Contrast imaging (Contrast agent for Liver)
Note 11: STE
Note 12: STQ
Note 13: Ultrasound Fusion Imaging
Note 14: GYN/Pelvic
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE NEEDED)
Concurrence of CDRH, Office of Device Evaluation(ODE)
System:DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85/DC-86/DC-86S/DC-89/DC-TV/DC-TQ Diagnostic UltrasoundSystem
Transducer:6LB7E
Intended Use:Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation
General (Track1 Only)Specific (Track 1 & 3)BMPWDCWDColorDopplerAmplitudeDopplerCombined(specify)Other (specify)
OphthalmicOphthalmic
Fetal Imaging& OtherFetal
Abdominal
Intra-operative (Specify*)
Intra-operative (Neuro)
Laparoscopic
Pediatric
Small Organ (Specify**)
Neonatal Cephalic
Adult Cephalic
Trans-rectalNNNNNNNote 1,2,4,6,7
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
CardiacCardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
PeripheralvesselPeripheral vessel
Other (Specify***)NNNNNNNote 1,2,4,6,7
N=new indication; P=previously cleared by FDA; E=added under Appendix E
Additional comments: Combined modes--B+M、PW+B、Color + B、Power + B、Power +Color+B、Power + PW +B.
*Intraoperative includes abdominal, thoracic, and vascular
**Small organ-breast, thyroid, testes.
***Other use includes Urology.
Note 1: Tissue Harmonic Imaging. The feature does not use contrast agents.
Note 2: Smart3D
Note 3:4D(Real-time 3D)
Note 4: iScape
Note 5: TDI
Note 6: Color M
Note 7: Biopsy Guidance
Note 8: Strain Elastography
Note 9: Contrast imaging(contrast agent for LVO)
Note 10: Contrast imaging (Contrast agent for Liver)
Note 11: STE
Note 12: STQ
Note 13: Ultrasound Fusion Imaging
Note 14: GYN/Pelvic
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE NEEDED)
Concurrence of CDRH, Office of Device Evaluation(ODE)

{32}------------------------------------------------

{33}------------------------------------------------

{34}------------------------------------------------

{35}------------------------------------------------

{36}------------------------------------------------

{37}------------------------------------------------

{38}------------------------------------------------

{39}------------------------------------------------

{40}------------------------------------------------

510(K) SUMMARY

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92(c).

The assigned 510(k) number is: K192152

1. Submitter:

Shenzhen Mindray Bio-medical Electronics Co., LTD Mindray Building, Keji 12th Road South, Hi-tech Industrial Park, Nanshan, Shenzhen, 518057, P. R. China Tel: +86 755 8188 5655 Fax: +86 755 2658 2680

Contact Person:

Zhang Wei Shenzhen Mindray Bio-medical Electronics Co., LTD Mindray Building, Keji 12th Road South, Hi-tech Industrial Park, Nanshan, Shenzhen, 518057, P. R. China

Date Prepared: July 25, 2019

  1. Device Name: DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85/DC-86/DC-865/ DC-89/DC-TV/DC-TQ Diagnostic Ultrasound System

Classification

Regulatory Class: II Review Category: Tier II 21 CFR 892.1550 Ultrasonic Pulsed Doppler Imaging System (IYN) 21 CFR 892.1560 Ultrasonic Pulsed Echo Imaging System (IYO) 21 CFR 892.1570 Diagnostic Ultrasound Transducer (ITX) 21 CFR 892.2050 Picture Archiving and Communications System(LLZ)

{41}------------------------------------------------

3. Device Description:

DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85/DC-86/DC-86S/DC-89/DC-TV/

DC-TQ is a software controlled, ultrasonic diagnostic system. Its function is to acquire and display ultrasound data in B-Mode, M-Mode, PW-Mode, CCW-Mode, Color-Mode , Power/Dirpower Mode, THI, Smart3D, 4D, iScape, TDI, Color M, Biopsy Guidance, Strsin Elastography, Contrast imaging (Contrast agent for Liver and LVO), STE, STQ, Ultrasound Fusion Imaging, GYN/Pelvic and the combined mode (i.e. B/M-Mode, B/PW-mode, B/PW/Color).

This system is a Track 3 device that employs an array of probes that include Linear array, Convex array, Phased array, Volume and Pencil probe.

4. Intended Use:

The DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85/DC-86/DC-86S/DC-89/DC-TV/ DC-TO Diagnostic Ultrasound System is applicable for adults, pregnant women, pediatric patients and neonates. It is intended for use in fetal, abdominal, intra-operative (abdominal, thoracic, and vascular), pediatric, small organ(breast, thyroid, testes), neonatal and adult cephalic, trans-rectal, trans-vaginal, musculo-skeletal(conventional, superficial), cardiac(adult, pediatric) , trans-esoph. (Cardiac), peripheral vessel, and urology exams.

5. Summary of Modifications

  • . New Added Transducers: C5-2E, P4-2E, D6-2E, SC6-1E, L20-5E, SD8-1E, DE10-3WE, V11-3WE, SC51-NE, 6LB7E;
  • . New Added Needle-Guided Bracket: NGB-009, NGB-015, NGB-020, NGB-027, NGB-035, NGB-039;
  • Main Added Features:
      1. Ultrasound Fusion Imaging to C5-1E, SP5-1E, SC5-1E, L14-5WE, SC6-1E;
      1. Smart Pelvic Floor to C5-1E, V11-3HE, D8-2E, D7-2E, DE11-3E, SC5-1E, D6-2NE, C6-2GE, V11-3E, V11-3BE, DE10-3E, C5-2E, D6-2E, SC6-1E, SD8-1E, DE10-3WE, SC5-1NE, C7-3E;
      1. STE/STQ to L12-3E, L14-5WE, L9-3E, SC6-1E;

{42}------------------------------------------------

    1. Strain Elastography to L20-5E, V11-3WE;
    1. Add the peripheral vessel to the Clinical Application of CW5s ;;
    1. Smart Face;
    1. Smart Planes FH;
    1. Glazing Flow;
    1. Fusion RESP;
    1. The DC-80 series has two had disks (128G SSD+1T HDD);
    1. The battery can provide power supply when scanning;
    1. DC-80 series can support four or five transducer ports configuration;
    1. The signal channel between the probe and the main host is 128.

6. Comparison with Predicate Devices:

DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85/DC-86/DC-86S/DC-89/DC-TV/

DC-TQ Diagnostic Ultrasound System is comparable with and substantially equivalent to these predicate devices:

Predicate DeviceManufacturerModel510(k)ControlNumber
1. Primary predicate deviceMindrayDC-70K181637
2. Reference deviceGEVoluson E8K181985
3. Reference deviceMindrayResona 7K171233
4. Reference deviceMindrayDC-90K190206
5. Reference deviceMindrayDC-80K173471
6. Reference deviceMindrayM9K171034

DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85/DC-86/DC-86S/DC-89/DC-TV/

DC-TQ has the same technological characteristics, is comparable in key safety and effectiveness features, and has the same intended uses and basic operating modes as the predicate devices. All systems transmit ultrasonic energy into patients and perform

{43}------------------------------------------------

post processing of received echoes to generate onscreen display of anatomic structures and fluid flow within the body. All systems allow for specialized measurements of structures and flow, and calculations.

  • Subject device
    DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85/DC-86/DC-86S/DC-86S/DC-89 /DC-TV/DC-TQ has the samilar intended uses and transduer as the predicated device DC-70 (K181637),DC-90 (K190206) , DC-80 (K173471 ) , M9 (K171034)
SubjectDeviceDC-80DC-70(predicatedevice)DC-80(Referencedevice)Resona7(Referencedevice)M9(Referencedevice)
C5-2EC5-2E///
P4-2EP4-2E///
L20-5EL20-5E///
6LB7E6LB7E///
SD8-1ESD8-1E///
SC5-1NEC5-2E///
D6-2E/D6-2NE//
SC6-1E/C5-1E//
DE10-3WE//DE10-3WU/
V11-3WE/V11-3WE//
CW5s///CW2s

DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85/DC-86/DC-86S/DC-89/DC-TV/

DC-TQ has the same function as the predicated device Resona 7 (K171233)

Subject Device DC-80Predicate device Resona7 (171233)
Ultrasound Fusion ImagingUltrasound Fusion Imaging
Smart PelvicSmart Pelvic
STE/STQSTE/STQ
elastographyelastography
Smart FaceSmart Face

{44}------------------------------------------------

Smart Planes FHSmart Planes FH
Fusion RESPFusion RESP

DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85/DC-86/DC-86S/DC-89/DC-TV/

DC-TQ has the samilar function as the Reference device Voluson™ E8 K181985

Subject Device DC-80Predicate device( Voluson TM E8 K181985)
Glazing FlowGlazing Flow

7. PERFORMANCE DATA

  • . Biocompatibility testing
    The biocompatibility evaluation for the DC-80 device was conducted in accordance with International Standard ISO 10993-1 "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA.

  • . Electrical safety and electromagnetic compatibility (EMC) DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85/DC-86/DC-86S/DC-89/ DC-TV/DC-TQ is Electrical safety and electromagnetic compatibility (EMC) Electrical safety and EMC testing were conducted on the DC-80 Diagnostic Ultrasound. The system complies with the IEC 60601-1, and IEC 60601-2-37 standards for safety and the IEC 60601-1-2 standard for EMC.

  • . DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85/DC-86/DC-86S/DC-89/ DC-TV/DC-TQ is designed in compliance with the FDA recognized electrical and physical safety standard, which is the same as the predicated device DC-70 (K181637), DC-90(K190206), DC-80(K173471), M9(K171034)

  • . Acoustic Power Testing

The acoustic power levels of DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85

/DC-86/DC-86S/DC-89/DC-TV/DC-TQ are below the limits of FDA, as recommended by FDA's Guidance for Industry and FDA Staff "Marketing Clearance of Diagnostic Ultrasound Systems and Transducers - Guidance for Industry and Food and Drug Administration Staff".

{45}------------------------------------------------

Software Verification and Validation Testing ●

Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices."

8. Non-clinical Tests:

DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85/DC-86/DC-86S/DC-89/DC-TV/ DC-TQ Diagnostic Ultrasound System has been evaluated for acoustic output, biocompatibility, cleaning and disinfection effectiveness as well as thermal, electrical and mechanical safety, and has been designed to conform with applicable medical safety standards. This device has been tested and evaluated under the following standards:

  • AAMI/ANSI ES60601-1: Medical electrical equipment Part 1: General requirements for basic safety and essential performance
  • . IEC 60601-1-2: Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests (Edition 3)
  • . IEC 60601-2-37: Medical electrical equipment - Part 2-37: Particular requirements for the basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment
  • IEC 62304: Medical device software Software life cycle processes ●
  • IEC:62366:Medical devices - application of usability engineering to medical devices
  • IEC 60601-1-6: medical electrical equipment - part 1-6: general requirements for basic safety and essential performance - collateral standard: usability.
  • ISO14971: Medical devices - Application of risk management to medical devices
  • UD 2: Acoustic Output Measurement Standard for Diagnostic Ultrasound Equipment ● Revision 3
  • ISO 10993-1 Biological evaluation of medical devices -- Part 1: Evaluation and testing within a risk management process

{46}------------------------------------------------

These non-clinical tests relied on in this premarket notification submission can support the determination of substantial equivalence of the subject device.

9. Clinical Studies

Not applicable. The subject of this submission, DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85/DC-86/DC-86S/DC-89/DC-TV/DC-TQ Diagnostic Ultrasound System, does not require clinical studies to support substantial equivalence.

10. Summary

Based on the performance data as documented in the study, the DC-80 Diagnostic Ultrasound system was found to have a safety and effectiveness profile that is similar to the predicate device.

Conclusion:

Intended uses and other key features are consistent with traditional clinical practices, FDA guidelines and established methods of patient examination. The design, development and quality process of the manufacturer confirms with 21 CFR 820, ISO 9001 and ISO 13485 quality systems. The device conforms to applicable medical device safety standards. Therefore, the DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85/DC-86/DC-86S/DC-89/DC-TV/DC-TQ Diagnostic Ultrasound System is substantially equivalent with respect to safety and effectiveness to devices currently cleared for market.

§ 892.1550 Ultrasonic pulsed doppler imaging system.

(a)
Identification. An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.