K Number
K192056

Validate with FDA (Live)

Manufacturer
Date Cleared
2020-03-27

(239 days)

Product Code
Regulation Number
880.6850
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Aesculap JJ Series Container System is a reusable rigid sterilization container system intended to be used to enclose another medical device that is to be sterilized by a healthcare provider. It is intended to allow sterilization of the enclosed device and also maintain sterility of the enclosed device during transport and until used. This container system is compatible for use in steam sterilization according to the configurations listed in the tables below.

The JJ Series Container System container styles: a solid bottom container and enhanced drying system (EDS) container.

The Aesculap JJ Series Container System includes accessories such as baskets, trays, silicone mats, instrument organizational accessories, holders, indicator cards, tamper evident locks, lid covers, faceplate holders, and tags.

The attached tables identifies the load configurations.

Device Description

The Aesculap JJ Series Container System is a reusable rigid sterilization container system intended to be used to enclose another medical device that is to be sterilized by a healthcare provider. It is intended to allow sterilization of the enclosed device and also maintain sterility of the enclosed device until used. This container system is compatible for use in Prevac steam with and without dry time.

The JJ Series Container System consists of two different container styles: a solid bottom container and enhanced drying system (EDS) container.

The containers are made from anodized aluminum and utilize a single-use filter. The JJ Series Container System includes accessories such as baskets, trays, silicone mats, instrument organizational accessories, holders, indicator cards, tamper evident locks, lid covers, faceplate holders, and tags.

AI/ML Overview

The provided text is a 510(k) Summary for the Aesculap JJ Series Container System. This document focuses on demonstrating substantial equivalence to a predicate device based on nonclinical performance data, rather than on a comparative effectiveness study with human readers (MRMC) or a standalone AI algorithm. Therefore, many of the requested categories (2, 3, 4, 5, 6, 7, 8, 9) are not applicable or cannot be extracted from this document as they are specific to studies involving AI or human interpretation of data, which is not the nature of this device (a sterilization container system).

However, I can provide the acceptance criteria and reported device performance from the nonclinical testing, as well as general information about the device and its testing.

Here's the information that can be extracted:

1. Table of Acceptance Criteria and the Reported Device Performance

Performance TestingAcceptance CriteriaReported Device Performance
Sterilization EfficacyA sterility assurance level (SAL) of 10-6 will be achieved post sterilization using the BI overkill method and half cycle validation indicated in section 5.6 of AAMI ST-77. Biological indicators must be negative for growth after incubation period.Pass
Dry TimePer AAMI ST-77 section 4.4.2, (in association with EN868-8 section G.3/G.4 and EN 285 section G20.3), the system shall demonstrate an average pre and post sterilization weight difference of less than 0.2% within five 5 minutes of cycle completion using final validated parameters and be free of visible moisture following a cooling period.Pass
Microbial Aerosol ChallengeThe container load maintains sterility after exposure to a defined amount of aerosol microorganisms per ST-77 sections 5.9.1.1 & 5.9.1.2. No presence of growth after incubation period.Pass
Sterility MaintenanceSterility of container contents is maintained under conditions which simulate hospital sterile package handling and storage conditions and were tested per ISO 11167-1 section 6.4. Test articles stored for their post processing shelf life. Biological indicators must be negative for growth after incubation period.Pass
Material CompatibilityNo degradation or impact to functionality at the completion of multiple sterilization cycles according to section 5.0 of TIR17.Pass
CytotoxicityTesting completed in accordance with ISO 10993-5:2009. Using the ISO Elution Method, the response to the article must not be greater than 2 (mild reactivity).Pass
Cleaning ValidationIn accordance with Annex A and section 7.5 of TIR30, test samples shall show no visible soil after cleaning and shall have protein or hemoglobin levels less than the predetermined amounts.Pass

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

This information is not provided in the document. The document describes the acceptance criteria for various performance tests, but does not specify the sample sizes used for these tests or the data provenance. It states "Testing was conducted in accordance with applicable FDA guidance and standards," which implies a prospective testing approach conducted by the manufacturer to meet regulatory requirements.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This information is not applicable/provided. The device is a sterilization container. The "ground truth" for the performance tests listed (e.g., sterility, dry time, material compatibility) is established through standardized laboratory testing protocols and objective measurements, not by expert consensus or interpretation of medical images.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This information is not applicable/provided. Adjudication methods like 2+1 or 3+1 are typically used in studies where human judgment is involved in establishing ground truth (e.g., reading medical images) and disagreements need to be resolved. This is not relevant for the type of objective performance testing described for a sterilization container.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This information is not applicable. An MRMC study is relevant for evaluating the performance of AI-assisted diagnostic tools with human readers. The Aesculap JJ Series Container System is a physical medical device (sterilization container), not an AI diagnostic tool, so such a study was not conducted.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

This information is not applicable. This question pertains to the performance of an AI algorithm in isolation. The Aesculap JJ Series Container System is a physical product and does not involve an AI algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The "ground truth" for the tests performed on the Aesculap JJ Series Container System is established through objective, standardized laboratory measurements and scientific principles as defined by recognized standards (e.g., AAMI ST-77, EN868-8, EN 285, ISO 11167-1, TIR17, ISO 10993-5:2009, TIR30). For example:

  • Sterilization Efficacy: Achieved Sterility Assurance Level (SAL) of 10-6, confirmed by negative biological indicators.
  • Dry Time: Measured weight difference and absence of visible moisture.
  • Microbial Aerosol Challenge: Absence of microbial growth after exposure.
  • Sterility Maintenance: Absence of microbial growth in stored test articles.
  • Material Compatibility: Absence of degradation or impact on functionality.
  • Cytotoxicity: Response to article not greater than 2 (mild reactivity) via ISO Elution Method.
  • Cleaning Validation: Absence of visible soil and protein/hemoglobin levels below predetermined amounts.

8. The sample size for the training set

This information is not applicable/provided. A "training set" refers to data used to train machine learning models. This device is a physical product, not an AI algorithm, so there is no training set in this context.

9. How the ground truth for the training set was established

This information is not applicable/provided for the same reason as point 8.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services - USA. To the right of this symbol, there is a blue square with the letters "FDA" in white. Next to the blue square, the words "U.S. FOOD & DRUG ADMINISTRATION" are written in blue.

March 27, 2020

Aesculap, Inc. Sierra Mertz Regulatory Affairs Specialist 3773 Corporate Parkway Center Valley, Pennsylvania 18034

Re: K192056

Trade/Device Name: Aesculap JJ Series Container System Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: Class II Product Code: KCT Dated: February 25, 2020 Received: February 27, 2020

Dear Sierra Mertz:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For CAPT Elizabeth Claverie, M.S. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K192056

Device Name

Aesculap JJ Series Container System

Indications for Use (Describe)

The Aesculap JJ Series Container System is a reusable rigid sterilization container system intended to be used to enclose another medical device that is to be sterlized by a healthcare provider. It is intended to allow sterilization of the enclosed device and also maintain sterility of the enclosed device during transport and until used. This container system is compatible for use in steam sterilization according to the configurations listed in the tables below.

The JJ Series Container System container styles: a solid bottom container and enhanced drying system (EDS) container.

The Aesculap JJ Series Container System includes accessories such as baskets, trays, silicone mats, instrument organizational accessories, holders, indicator cards, tamper evident locks, lid covers, faceplate holders, and tags.

The attached tables identifies the load configurations.

Table 1. JJ Series Container System Lumen Configurations

SterilizationCycleSizeLumens
PreVac Steamwith dry time(270°F4 min Exposure)Full Size4 stainless steel lumens with ≥ 1mm ID x≤500mm L
Three-Quarter4 stainless steel lumens with ≥ 1mm ID x≤400mm L
Half Size4 stainless steel lumens with ≥ 1mm ID x≤200mm L
Prevac Steamwithout dry time(270°F3 min Exposure)Non-PorousFull SizeNon lumened instruments
Three-QuarterSizeNon lumened instruments
Half SizeNon lumened instruments
Prevac Steamwithout dry time(270°F4 min Exposure)PorousFull Size4 stainless steel lumens with ≥ 1mm ID x≤500mm L
Three-QuarterSize4 stainless steel lumens with ≥ 1mm ID x≤400mm L
Half Size4 stainless steel lumens with ≥ 1mm ID x≤125mm L

Validation testing was performed using the JJ612 single use paper filter.

{3}------------------------------------------------

Table 2. JJ Series Container System Configurations, Without Silicone Mat and Lid cover*

ContainerTypeSterilizationMethodContainerBottom Part #ContainerLid Part #Maximum Load Weight (lbs)*
Solid BottomContainerPrevac Steam270°F4 min exposure15 minute*dry timeJJ140JJ41025
JJ13024
JJ12023
JJ11022
JJ340JJ43023
JJ33022
JJ32020
JJ31018
JJ240JJ42018
JJ23018
JJ22015
JJ21013
EDS ContainerPrevac Steam270°F4 min exposure8 minute*dry timeJJ141JJ41025
JJ13124
JJ12123
JJ11122
JJ341JJ43023
JJ33122
JJ32120
JJ31118
JJ241JJ42018
JJ23118
JJ22115
JJ21113
Solid BottomContainerPrevac SteamWithout dry time270°F3 min Exposure(Non-porous)ANDPrevac SteamWithout dry time270°F4 min Exposure(Porous)JJ140JJ41025
JJ130
JJ120
JJ110
JJ340JJ43025
JJ330
JJ320
JJ310
JJ240JJ42025
JJ230JJ220JJ210
EDS ContainerPrevac SteamWithout dry time270°F3 min Exposure(Non-porous)ANDPrevac SteamWithout dry time270°F4 min Exposure(Porous)JJ141JJ41025
JJ131
JJ121
JJ111
JJ341JJ43025
JJ331
JJ321
JJ311
JJ241JJ42025
JJ231JJ221
ContainerTypeSterilizationMethodContainerBottom Part #ContainerLid Part #Maximum LoadWeight (lbs)*
Solid BottomContainerPrevac Steam270°F4 min exposure30 minute*dry timeJJ140JJ41025
JJ13024
JJ12023
JJ11022
JJ340JJ43023
JJ33022
JJ32020
JJ31018
JJ240JJ42018
JJ23018
JJ22015
JJ21013
EDS ContainerPrevac Steam270°F4 min exposure30 minute*dry timeJJ141JJ41025
JJ13124
JJ12123
JJ11122
JJ341JJ43023
JJ33122
JJ32120
JJ31118
JJ241JJ42018
JJ23118
JJ22115
JJ21113
Solid BottomContainerPrevac SteamWithout dry time270°F3 min Exposure(Non-porous)ANDJJ140JJ41025
JJ130
JJ120
JJ110
JJ340JJ43025
JJ330
Prevac SteamWithout dry time270°F4 min Exposure(Porous)JJ320
JJ310
JJ240JJ42025
JJ230
JJ220
JJ210
EDS ContainerPrevac SteamWithout dry time270°F3 min Exposure(Non-porous)ANDJJ141JJ41025
JJ131
JJ121
JJ111
JJ341JJ43025
JJ331
Prevac SteamWithout dry time270°F4 min Exposure(Porous)JJ321
JJ311
JJ241JJ42025
JJ231
JJ221
JJ211

{4}------------------------------------------------

Table 3. JJ Series Container System Configurations, with Silicone Mat and Lid Cover*

*Dry times and load weights are based on container with silicone mats and/or lid cover.

{5}------------------------------------------------

Table 4: Sterilization Cycle Compatible Accessories - Prevac Steam
--------------------------------------------------------------------------------
AccessoriesPrevac Steamwith dry timePrevac SteamWithout dry time
Baskets, trays, indicator cards, tamper evidentlocks, faceplate holders, and tags.YesYes
Instrument Organization System (Silicone andStainless Steel racks, brackets, holders, andclamps)YesYes
Silicone matsYes¹Yes
Lid CoversYes¹Yes

1 Using silicone mats and/or lid covers may require a longer drying time.

When used with PreVac Steam with dry time (270°F, 4 minute exposure), the maintenance of sterility of the Aesculap JJ Series Container System is 365 days.

Type of Use (Select one or both, as applicable)Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)Prescription Use (Part 21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration

Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff

PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (7/17)Page 1 of 1PSC Publishing Services (301) 443-
6740EF

{6}------------------------------------------------

ട. 510(k) SUMMARY (as required by 21 CFR 807.92)

Aesculap JJ Series Container System for Steam

March 27, 2020

COMPANY:Aesculap, Inc.3773 Corporate ParkwayCenter Valley, PA 18034Establishment Registration Number: 2916714
CONTACT:Sierra M. Mertz610-984-9076 (phone)Sierra.mertz@aesculapimplants.com610-791-6882 (fax)
TRADE NAME:Aesculap® JJ Series Container System
COMMON NAME:Sterilization Container Wrap
CLASSIFICATION NAME:Wrap, Sterilization
REGULATION NUMBER:880.6850
PRODUCT CODE:KCT
DEVICE CLASS:Class II per 21 CFR 880.6850

DEVICE DESCRIPTION

The Aesculap JJ Series Container System is a reusable rigid sterilization container system intended to be used to enclose another medical device that is to be sterilized by a healthcare provider. It is intended to allow sterilization of the enclosed device and also maintain sterility of the enclosed device until used. This container system is compatible for use in Prevac steam with and without dry time.

The JJ Series Container System consists of two different container styles: a solid bottom container and enhanced drying system (EDS) container.

The containers are made from anodized aluminum and utilize a single-use filter. The JJ Series Container System includes accessories such as baskets, trays, silicone mats, instrument organizational accessories, holders, indicator cards, tamper evident locks, lid covers, faceplate holders, and tags.

INDICATIONS FOR USE

The Aesculap JJ Series Container System is a reusable rigid sterilization container system intended to be used to enclose another medical device that is to be sterilized by a healthcare

{7}------------------------------------------------

provider. It is intended to allow sterilization of the enclosed device and also maintain sterility of the enclosed device during transport and until used. This container system is compatible for use in steam sterilization according to the configurations listed in the tables below.

The JJ Series Container System consists of two different container styles: a solid bottom container and enhanced drying system (EDS) container.

The Aesculap JJ Series Container System includes accessories such as baskets, trays, silicone mats, instrument organizational accessories, holders, indicator cards, tamper evident locks, lid covers, faceplate holders, and tags.

The attached tables identifies the load configurations.

SterilizationCycleSizeLumens
PreVac Steamwith dry time(270°F4 min Exposure)Full Size4 stainless steel lumens with > 1mm ID x ≤500mm L
Three-QuarterSize4 stainless steel lumens with > 1mm ID x ≤400mm L
Half Size4 stainless steel lumens with > 1mm ID x ≤200mm L
Prevac Steamwithout dry time(270°F3 min Exposure)Non-PorousFull SizeNon lumened instruments
Three-QuarterSizeNon lumened instruments
Half SizeNon lumened instruments
Prevac Steamwithout dry time(270°F4 min Exposure)PorousFull Size4 stainless steel lumens with > 1mm ID x <500mm L
Three-QuarterSize4 stainless steel lumens with > 1mm ID x ≤400mm L
Half Size4 stainless steel lumens with > 1mm ID x <125mm L

Validation testing was performed using the JJ612 single use paper filter.

{8}------------------------------------------------

ContainerTypeSterilizationMethodContainerBottom Part #ContainerLid Part #Maximum LoadWeight (lbs)
Solid BottomContainerPrevac Steam270°F4 min exposure15 minute*dry timeJJ140JJ41025
JJ13024
JJ12023
JJ11022
JJ340JJ43023
JJ33022
JJ32020
JJ31018
JJ240JJ42018
JJ23018
JJ220JJ2101513
EDS ContainerPrevac Steam270°F4 min exposure8 minute*dry timeJJ141JJ131JJ121JJ111JJ41025242322
JJ341JJ43023
JJ331JJ321JJ311222018
JJ241JJ231JJ4201818
JJ221JJ2111513
Solid BottomContainerPrevac SteamWithout dry time270°F3 min Exposure(Non-porous)JJ140JJ130JJ120JJ110JJ41025
ANDJJ340JJ330JJ43025
Prevac SteamWithout dry timeJJ320JJ310
270°F4 min Exposure(Porous)JJ240JJ230JJ220JJ210JJ42025
EDS ContainerPrevac SteamWithout dry time270°F3 min Exposure(Non-porous)JJ141JJ131JJ121JJ111JJ43025
ANDJJ341JJ331JJ31025
Prevac SteamWithout dry timeJJ321
270°FJJ241JJ42025
4 min Exposure(Porous)JJ231
JJ221
JJ211

Table 2. JJ Series Container System Configurations, Without Silicone Mat and Lid cover*

{9}------------------------------------------------

Table 3. JJ Series Container System Configurations, with Silicone Mat and Lid Cover*
------------------------------------------------------------------------------------------
ContainerTypeSterilizationMethodContainerBottom Part #ContainerLid Part #Maximum LoadWeight (lbs)*
Solid BottomContainerPrevac Steam270°F4 min exposure30 minute*dry timeJJ140JJ41025
JJ13024
JJ12023
JJ11022
JJ340JJ43023
JJ33022
JJ32020
JJ31018
JJ240JJ42018
JJ23018
JJ22015
JJ21013
EDS ContainerPrevac Steam270°F4 min exposure30 minute*dry timeJJ141JJ41025
JJ13124
JJ12123
JJ11122
JJ341JJ43023
JJ33122
JJ32120
JJ31118
JJ241JJ42018
JJ23118
JJ22115
JJ21113
Solid BottomContainerPrevac SteamWithout dry time270°F3 min Exposure(Non-porous)JJ140JJ41025
JJ130
JJ120
JJ110
ANDPrevac SteamWithout dry time270°F4 min Exposure(Porous)JJ340JJ43025
JJ330
JJ320
JJ310
JJ240JJ42025
270°F4 min Exposure(Porous)JJ230
JJ220
JJ210

{10}------------------------------------------------

EDS ContainerPrevac SteamWithout dry time270°F3 min Exposure(Non-porous)ANDPrevac SteamWithout dry time270°F4 min Exposure(Porous)
JJ141JJ131JJ121JJ111JJ41025
JJ341JJ331JJ321JJ311JJ43025
JJ241JJ231JJ221JJ211JJ42025

*Dry times and load weights are based on container with silicone mats and/or lid cover.

Table 4: Sterilization Cycle Compatible Accessories - Prevac Steam
--------------------------------------------------------------------
AccessoriesPrevac Steam withdry timePrevac SteamWithout dry time
Baskets, trays, indicator cards, tamper evidentlocks, faceplate holders, and tags.YesYes
Instrument Organization System (Silicone andStainless Steel racks, brackets, holders, andclamps)YesYes
Silicone matsYes1Yes
Lid CoversYes1Yes

1 Using silicone mats and/or lid covers may require a longer drying time.

When used with PreVac Steam with dry time (270°F, 4 minute exposure), the maintenance of sterility of the Aesculap JJ Series Container System is 365 days.

{11}------------------------------------------------

TECHNOLIGICALLCHARACTERISTICS (compared to predicate device)

The Aesculap JJ Series Container System has the same intended use as the Aesculap predicate device. The subject device is offered in similar sizes and is comprised of similar materials of construction as the predicate device.

Subject Device -Aesculap JJ Series Container System(K192056)Product Code: KCTPredicate Device-Aesculap SterilContainer(K112671)Product Code: KCT
Intended UseA device intended to be used to enclose anothermedical device that is to be sterilized by a health careprovider. It is intended to allow sterilization of theenclosed medical device and also to maintain sterilityof the enclosed device until used.A device intended to be used to enclose anothermedical device that is to be sterilized by a health careprovider. It is intended to allow sterilization of theenclosed medical device and also to maintain sterilityof the enclosed device until used.Same
SterilizationModalitiesPreVac Steam with dry timePreVac Steam without dry timePreVac SteamIUSSEthylene OxideSimilar
MaterialContainer: Anodized aluminumLid: Anodized aluminumGasket: SiliconeContainer: Anodized aluminumLid: Anodized aluminumGasket: SiliconeSame
Filter TypeSingle use (Paper)Single use (Paper) and reusable (PTFE)Similar
ContainerDesignSolid bottom orSolid bottom with enhanced drying systemPerforated LidSolid or perforated bottomPerforated LidSimilar
SizesFull sizeThree-Quarter SizeHalf SizeThree-Quarter SizeSimilar
AccessoriesBaskets, trays, silicone mats, instrumentorganizational accessories, holders, indicator cards,tamper evident locks, lid covers, faceplate holders,and tags.Silicone mats, instrument organization system, lidcovers, baskets, trays, and racks, indicator cards, locksSimilar
Maintenanceof SterilityPrevac Steam with dry time - 365 daysPrevac Steam & Ethylene Oxide- 360 daysIUSS - N/ASimilar

{12}------------------------------------------------

SUMMARY OF NONCLINICAL PERFORMANCE DATA

The Aesculap JJ Series Container System is a reusable rigid container system that will allow for sterilization and storage of other medical devices. This container system is compatible for use in the following sterilization modalities:

  • Prevac steam with dry time (Exposure: 270°F for 4 minutes) ●
  • Prevac steam without dry time (Exposure: 270°F for 4 minutes Porous) ●
  • Prevac steam without dry time (Exposure: 270°F for 3 minutes - Non-Porous)

Testing was conducted in accordance with applicable FDA guidance and standards. The table below outlines the testing which was performed on the subject device.

PerformanceTestingPurposeAcceptance CriteriaResults
SterilizationEfficacyTo determinesterilizationeffectiveness of testdevice after processingin a sterilization cycle.A sterility assurance level (SAL) of10-6 will be achieved post sterilization using the BI overkillmethod and half cycle validation indicated in section 5.6 ofAAMI ST-77. Biological indicators must be negative forgrowth after incubation period.Pass
Dry TimeTo determine the properdrying time required forthe JJ Series ContainerSystem.Per AAMI ST-77 section 4.4.2, (in association with EN868-8section G.3/G.4 and EN 285 section G20.3), the systemshall demonstrate an average pre and post sterilizationweight difference of less than 0.2% within five 5 minutes ofcycle completion using final validated parameters and befree of visible moisture following a cooling period.Pass
MicrobialAerosolChallengeTo analyze the packageintegrity and microbialbarrier properties of thetest device.The container load maintains sterility after exposure to adefined amount of aerosol microorganisms per ST-77sections 5.9.1.1 & 5.9.1.2. No presence of growth afterincubation period.Pass
SterilityMaintenanceTo demonstrate that aprocessed test devicecan maintain a sterilebarrier for a definedperiod of time.Sterility of container contents is maintained under conditionswhich simulate hospital sterile package handling and storageconditions and were tested per ISO 11167-1 section 6.4.Test articles stored for their post processing shelf life.Biological indicators must be negative for growth afterincubation period.Pass
MaterialCompatibilityTo assess effects of fulluse cycles on devicecomponents and theirintended functionality.No degradation or impact to functionality at the completionmultiple sterilization cycles according to section 5.0 ofTIR17.Pass
CytotoxicityTo determine thepotential of a test deviceto cause cytotoxicity.Testing completed in accordance with ISO 10993-5:2009.Using the ISO Elution Method, the response to the articlemust not be greater than 2 (mild reactivity).Pass
CleaningValidationTo verify theeffectiveness of thedevice cleaningprocedure.In accordance with Annex A and section 7.5 of TIR30, testsamples shall show no visible soil after cleaning and shallhave protein or hemoglobin levels less than thepredetermined amounts.Pass

Results of the testing demonstrate that the proposed device met the acceptance criteria for each performance test.

CONCLUSION

The conclusions drawn from the nonclinical tests demonstrate that Aesculap JJ Series Container System the device is as safe, as effective, and performs as well as or better than the legally marketed device.

§ 880.6850 Sterilization wrap.

(a)
Identification. A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.(b)
Classification. Class II (performance standards).