K Number
K192032

Validate with FDA (Live)

Date Cleared
2020-03-03

(217 days)

Product Code
Regulation Number
888.3030
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Fixone Biocomposite Small Anchor are intended to be used fixation to bone in the foot, ankle, knee, hand, wrist, elbow, shoulder, and hip.

Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clabicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction

Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Metatarsal Ligament Repair, Hallux Valgus reconstruction, digital tendon transfers. Mid-foot reconstruction

Knee: Medial Collateral Ligament Repair, Lateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis

Hand Wrist: Scapholunate Ligament Reconstruction, Carpal Ligament Reconstruction of collateral ligaments, Repair of Flexor and Extensor Tendons at the PP, DIP, and MCP joints for all digits, digital tendon transfers Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction

Hip: Capsular repair, Acetabular Labral repair

Device Description

The Fixone Biocomposite Small Anchor is intended for reattaching soft tissue to bone with sutures. It consists of Driver, Anchor, Non-absorbable suture and needle. A needle was attached to the end of the suture so that it could penetrate the damaged soft tissue to enable it to be used for OPEN surgery without arthroscopy. The anchor is manufactured from biodegradable materials (PLGA copolymer and beta-TCP). A non absorbable suture manufactured from co-braided UHMWPE and PET fibers is inserted into the anchor is implanted using a provided driver.

AI/ML Overview

The provided text is a 510(k) summary for the Fixone Biocomposite Small Anchor, a medical device intended for soft tissue to bone fixation. It details the device's characteristics, intended use, and performance data used to demonstrate substantial equivalence to a legally marketed predicate device.

Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided document:

Acceptance Criteria and Reported Device Performance

The acceptance criteria for the Fixone Biocomposite Small Anchor are primarily based on demonstrating substantial equivalence to a predicate device (Arthrex SutureTak Suture Anchor) through a series of bench tests and biocompatibility evaluations. The "Requirements" column in the table implicitly defines the acceptance criteria, and the "Results" column indicates whether these criteria were met.

  1. Table of Acceptance Criteria and Reported Device Performance:
Test Item / CategoryAcceptance Criteria (Defined by Standard/Comparison)Reported Performance / Result
Material Performance
External surface MeasurementNot explicitly stated (likely visual/dimensional inspection)Pass
Insertion torqueASTM F2502 and USP <28> (implies meeting specified torque ranges)Pass
Fixation strengthNot explicitly stated (implied to be equivalent to predicate or safe for intended use)Pass
Tensile strengthNot explicitly stated (implied to be equivalent to predicate or safe for intended use)Pass
Extractable colorNot explicitly stated (likely visual inspection for absence of unapproved color)Pass
Extraction testNot explicitly stated (general requirement for extraction testing)Pass
pHThe difference should be 1.5 and less.Pass
Potassium permanganate reducing substancesThe difference of the consumption of potassium permanganate should be 2.0 mL and less.Pass
Residue after evaporationRecord the weight of the residue should be 1.0mg and less.Pass
Heavy metalsAny brown color produced within 10 minutes in the tube containing the extract of the prepared sample does not exceed that in the tube containing the standard lead solutionPass
UV spectrum (250nm~350nm)Maximum absorbance between 250 to 350 nm should be 0.1 and less.Pass
PropertyWhen observing it with the naked eye, test solution should be clear and have no foreign particles.Pass
Performance Testing of Fixone® Biocomposite Anchor (Pull-out test)ASTM F1839-08 (Standard for Rigid Polyurethane Foam, implies testing in a relevant bone model with expected retention)Pass
Biocompatibility (Anchor)
CytotoxicityISO 10993-5(2009) Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicityPass
Acute systemic toxicity testISO 10993-11(2009) Biological evaluation of medical devices - Part 11: Tests for systemic toxicityPass
Pyrogen TestISO 10993-11 Test for systemic toxicity, pyrogen testPass
Intracutaneous (intradermal) reactivity testISO 10993-10(2013) Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitizationPass
Maximization test for delayed hypersensitivityISO 10993-10(2013) Test for irritation and skin sensitization, Maximization test for delayed hypersensitivityPass
Bacterial reverse mutation testISO 10993-3, Genotoxicity test OECD 471, Bacterial reverse mutation testPass
Mammalian erythrocyte micronucleus testISO 10993-3, Genotoxicity test OECD 471, Bacterial reverse mutation testPass
Implantation testISO 10993-6, Tests for local effects after implantation, Annex D test methods for implantation in bonePass
Bioabsorbable screws testASTM F2502 Standard specification and test methods for bioabsorbable plates and screws for internal fixation implantsPass
Subchronic toxicity testISO 10993-11 Biological Evaluation of Medical Devices Part 11- Test for systemic toxicityPass
Biocompatibility (Suture)
CytotoxicityISO 10993-5, Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicityPass
Systemic toxicity testISO 10993-11, Biological evaluation of medical devices - Part 11: Tests for systemic toxicityPass
Pyrogen TestISO 10993-11 Test for systemic toxicity, pyrogen testPass
Intracutaneous reactivity testISO 10993-10, Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitizationPass
Maximization sensitizationISO 10993-10, Test for irritation and skin sensitization, Maximization test for delayed hypersensitivityPass
Genotoxicity testISO 10993-3, Genotoxicity test OECD 471, Bacterial reverse mutation testPass
Implantation testISO 10993-6, Tests for local effects after implantation, Annex D test methods for implantation in bonePass
Hemolysis testISO 10993-4, Biological evaluation of medical devices - Part 4: Selection of tests for interactions with bloodPass

Study Details:

The provided document describes a bench testing and biocompatibility testing study conducted to demonstrate the substantial equivalence of the Fixone Biocomposite Small Anchor to its predicate device. This is a common approach for Class II medical devices seeking 510(k) clearance, where clinical trials are often not required if substantial equivalence can be demonstrated through non-clinical data.

  1. Sample Size used for the test set and the data provenance:

    • Sample Size: The document does not explicitly state the specific number of units/samples used for each bench test (e.g., number of anchors for pull-out tests, number of extracts for chemical analysis). It simply states that "Bench test were performed." and "The tests demonstrated that the device performs in a substantially equivalent manner to the predicate device." For biocompatibility, it states "Test result" for each test, implying a sufficient number of test articles were used as per the specified ISO/ASTM standards.
    • Data Provenance: The tests were performed by AJU Pharm Co.,Ltd. in Korea, which is the manufacturer. The data is retrospective, as it's presented after the tests were conducted.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not Applicable. This type of study (bench testing and biocompatibility) does not involve human experts establishing "ground truth" for a test set in the way an AI diagnostic study would where radiologists label images. The "ground truth" here is established by adherence to recognized international standards (ISO, ASTM, USP, OECD) and their inherent testing methodologies, which are designed to be objective and reproducible.
  3. Adjudication method for the test set:

    • Not Applicable. Adjudication methods (like 2+1 or 3+1 by experts) are pertinent to studies involving human interpretation or clinical outcomes, typically in AI/diagnostic device evaluations. For bench testing and biocompatibility, the results are typically quantitative or qualitative "pass/fail" outcomes based on pre-defined standard criteria.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, an MRMC comparative effectiveness study was not done. This type of study is relevant for AI-powered diagnostic devices where human readers (e.g., radiologists) use the device to assist in interpretation. The Fixone Biocomposite Small Anchor is a physical implant, not a diagnostic tool.
  5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Not Applicable. As a physical medical device (biocomposite anchor), the concept of "standalone algorithm only performance" does not apply. The device's performance is measured directly through physical and chemical properties and biological interactions, not through an algorithm.
  6. The type of ground truth used:

    • The "ground truth" for this device is based on established industry standards and regulatory requirements for medical device safety and performance. This includes:
      • Biocompatibility Standards: ISO 10993 series for biological evaluation of medical devices (e.g., cytotoxicity, systemic toxicity, irritation, sensitization, genotoxicity, implantation).
      • Material and Mechanical Testing Standards: ASTM (e.g., F2502 for bioabsorbable devices, F1839 for foam testing) and USP (pharmacopeial standards).
      • Comparison to Predicate Device: The performance and safety profile of the predicate device (Arthrex SutureTak Suture Anchor) acts as a de facto "ground truth" for establishing substantial equivalence in terms of intended use, principle of operation, and general safety/performance characteristics.
  7. The sample size for the training set:

    • Not Applicable. There is no "training set" in the context of this device's evaluation. This is not an AI/machine learning device that requires training data. The device's design and manufacturing processes are established based on engineering principles and regulatory guidance, not learned from data.
  8. How the ground truth for the training set was established:

    • Not Applicable. Since there is no training set, there is no ground truth established for one.

{0}------------------------------------------------

March 3, 2020

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The FDA text logo is in blue and includes the letters "FDA" followed by "U.S. FOOD & DRUG ADMINISTRATION" in a stacked format.

AJU Pharm Co., Ltd. % Peter Chung President Plus Global 300 Atwood St. Pittsburgh, Pennsylvania 15213

Re: K192032

Trade/Device Name: Fixone Biocomposite Small Anchor Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: MAI, MBI Dated: January 28, 2020 Received: February 3, 2020

Dear Peter Chung:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Laura C. Rose, Ph.D. Acting Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K192032

Device Name

Fixone Biocomposite Small Anchor

Image /page/2/Picture/6 description: The image shows the words "Indications for Use (Describe)". The words are in black font and are centered on the page. There is a line above the words.

The Fixone Biocomposite Small Anchor are intended to be used fixation to bone in the foot, ankle, knee, hand, wrist, elbow, shoulder, and hip.

Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clabicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction

Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Metatarsal Ligament Repair, Hallux Valgus reconstruction, digital tendon transfers. Mid-foot reconstruction

Knee: Medial Collateral Ligament Repair, Lateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis

Hand Wrist: Scapholunate Ligament Reconstruction, Carpal Ligament Reconstruction of collateral ligaments, Repair of Flexor and Extensor Tendons at the PP, DIP, and MCP joints for all digits, digital tendon transfers Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction

Hip: Capsular repair, Acetabular Labral repair

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary [K192032]

1. Applicant

    1. Company : AJU Pharm Co.,Ltd.
    1. Address : A-207, 697, Pangyo-ro, Seongnam-si, Gyeonggi-do, Korea
    1. Tel : 82-31-765-4420
    1. Fax : 82-31-602-7818
  • ട) Prepared date : July 01, 2019
  • (() Contact person : Peter Chung, 412-512-8802
    1. Contact person address : 300, Atwood Street, Pittsburgh, PA, 15213, USA
    1. Submission date : Feb. 25, 2020
    1. Prior related submissions : K171299

2. Device Information

    1. Trade name : Fixone Biocomposite Small Anchor
    1. Common name : Biodegradable Orthopedic Bone Screw
    1. Regulation name : Single/multiple component metallic bone fixation appliances and accessories
    1. Product code : MAI, MBI
    1. Regulation number : 888.3080
  • Class of device : Class II ର)
    1. Panel : Orthopedic

3. The legally marketed device to which we are claiming equivalence

K140855, Arthrex SutureTak Suture Anchor

4. Device description

The Fixone Biocomposite Small Anchor is intended for reattaching soft tissue to bone with sutures. It consists of Driver, Anchor, Non-absorbable suture and needle. A needle was attached to the end of the suture so that it could penetrate the damaged soft tissue to enable it to be used for OPEN surgery without arthroscopy. The anchor is manufactured from biodegradable materials (PLGA copolymer and β-ΤCP). A non absorbable suture manufactured from co-braided UHMWPE and PET fibers is inserted into the anchor is implanted using a provided driver.

This device is to be used with instrument that manufactured by Aju Pharm Co.,Ltd listed in the table below. The instruments are provided non-sterile(user must sterilization before use).

Product nameDrill bit and Drill guide with handle / Fixone.I.S-D/B300 and Fixone.I.S-D/G300
Intended useThe instruments are used to make the hole in the bone for anchor placement.
ManufacturerAJU Pharm Co.,Ltd. / Korea
CharacteristicThis device is provided as Non-sterile and Reusable device.This device is user-sterilized.
Cleaning andSterilization processBefore use, user must proceed cleaning and sterile process.Sterilization process : Steam sterilizationTemperature 132°C (270°F) / Operating time 4min / Drying 20 min

5. Intended Use :

The Fixone Biocomposite Small Anchor are intended to be used for suture (soft tissue) fixation to bone in the foot, ankle, knee, hand, wrist, elbow, shoulder, and hip.

Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clabicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction

Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Metatarsal Ligament Repair, Hallux Valgus reconstruction, digital tendon transfers, Mid-foot reconstruction

Knee: Medial Collateral Ligament Repair, Lateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis

{4}------------------------------------------------

Hand/Wrist: Scapholunate Ligament Reconstruction, Carpal Ligament Reconstruction, Repair/Reconstruction of collateral ligaments, Repair of Flexor and Extensor Tendons at the PIP, DIP, and MCP joints for all digits, digital tendon transfers

Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction Hip: Capsular repair, Acetabular Labral repair

6. Performance data:

  • Bench test were performed. Bench testing included biocompatibility, mechanical testing 1) including EO residues. The tests demonstrated that the device performs in a substantially equivalent manner to the predicate device. The following bench testing is performed to demonstrate the functionality is substantially equivalent.
Test itemRequirementsResults
External surfaceMeasurement
Insertion torqueASTM F2502 and USP<28>
Fixation strengthPass
Tensile strength
Extractable color
Extraction test
pHThe difference should be 1.5 and less.
Potassium permanganatereducing substancesThe difference of the consumption of potassium permanganate shouldbe 2.0 mL and less.
Residue after evaporationRecord the weight of the residue should be 1.0mg and less.
Heavy metalsAny brown color produced within 10 minutes in the tube containing theextract of the prepared sample does not exceed that in the tubecontaining the standard lead solutionPass
UV spectrum(250nm~350nm)Maximum absorbance between 250 to 350 nm should be 0.1 and less.
PropertyWhen observing it with the naked eye, test solution should be clear andhave no forign particles.
Performance Testing of Fixone® Biocomposite Anchor (Pull-out test)Pull-out test by immersion time of saline solution to evaluate two bioabsorbable suture anchors.[Absorption time (soaking time) : 4 weeks, 6 weeks, 12 weeks and 26 weeksASTM F1839-08, Standard specification for RGID Polyurethane Foam for Use as a Standard Material forTesting Orthopaedic Devices and Instruments
Pass

2) Biocompatibility

Anchor

#Test itemTest method / Test criteriaTest result
1CytotoxicityISO 10993-5(2009) Biological evaluation of medical devices - Part5: Tests for in vitro cytotoxicityPass
2Acute systemic toxicitytestISO 10993-11(2009) Biological evaluation of medical devices -Part 11: Tests for systemic toxicityPass
3Pyrogen TestISO 10993-11 Test for systemic toxicity, pyrogen testPass
4Intracutaneous(intradermal) reactivity testISO 10993-10(2013) Biological evaluation of medical devices -Part 10: Tests for irritation and skin sensitizationPass
5Maximization test fordelayed hypersensitivityISO 10993-10(2013) Test for irritation and skin sensitization,Maximization test for delayed hypersensitivityPass
6Bacterial revers mutationtestISO 10993-3, Genotoxicity test OECE 471, Bacterial reversemutation testPass
7Mammalian erythrocytemicronucleus testISO 10993-3, Genotoxicity test OECE 471, Bacterial reversemutation testPass
8Implantation testISO 10993-6, Tests for local effects after implantation, Annex Dtest methods for implantation in bonePass
9Bioabsorbable screws testASTM F2502 Standard specification and test methods forbioabsorbable plates and screws for internal fixation implantsPass
10Subchronic toxicity testISO 10993-11 Biological Evaluation of Medical Devices Part 11-Pass

{5}------------------------------------------------

Test for systemic toxicity
#Test itemTest method / Test criteriaTest result
Suture
1CytotoxicityISO 10993-5, Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicityPass
2Systemic toxicity testISO 10993-11, Biological evaluation of medical devices - Part 11: Tests for systemic toxicityPass
3Pyrogen TestISO 10993-11 Test for systemic toxicity, pyrogen testPass
4Intracutaneous reactivity testISO 10993-10, Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitizationPass
5Maximization sensitizationISO 10993-10, Test for irritation and skin sensitization, Maximization test for delayed hypersensitivityPass
6Genotoxicity testISO 10993-3, Genotoxicity test OECE 471, Bacterial reverse mutation testPass
8Implantation testISO 10993-6, Tests for local effects after implantation, Annex D test methods for implantation in bonePass
9Hemolysis testISO 10993-4, Biological evaluation of medical devices - Part 4: Selection of tests for interactions with bloodPass

The performance tests demonstrated that Fixone Biocomposite Small Anchor performs in a substantially equivalent manner to the predicate device.

7. Predicate device comparison table

ManufacturerAJU Pharm Co.,Ltd.Arthrex Inc.Remark
510(k) No.K140855N/A
Indication for useThe Fixone Biocomposite Small Anchor areintended to be used for suture (soft tissue)fixation to bone in the foot, ankle, knee,hand, wrist, elbow, shoulder, and hip.Shoulder: Rotator Cuff Repair, BankartRepair, SLAP Lesion Repair, Biceps Tenodesis,Acromio-Clavicular Separation Repair,Deltoid Repair, Capsular Shift orCapsulolabral ReconstructionFoot/Ankle: Lateral Stabilization, MedialStabilization, Achilles Tendon Repair,Metatarsal Ligament Repair, Hallux Valgusreconstruction, digital tendon transfers, Mid-foot reconstructionKnee: Medial Collateral Ligament Repair,Lateral Collateral Ligament Repair, PatellarTendon Repair, Posterior Oblique LigamentRepair, Iliotibial Band TenodesisHand/Wrist: Scapholunate LigamentReconstruction, Carpal LigamentReconstruction, Repair/Reconstruction ofcollateral ligaments, Repair of Flexor andExtensor Tendons at the PIP, DIP, and MCPjoints for all digits, digital tendon transfersElbow: Biceps Tendon Reattachment, Ulnaror Radial Collateral Ligament ReconstructionHip: Capsular repair, Acetabular Labral repairThe Arthrex SutureTak Suture Anchors areintended to be used for suture (soft tissue)fixation to bone in the foot, ankle, knee,hand, wrist, elbow, shoulder, and hip.Shoulder: Rotator Cuff Repair, BankartRepair, SLAP Lesion Repair, Biceps Tenodesis,Acromio-Clavicular Separation Repair,Deltoid Repair, Capsular Shift orCapsulolabral ReconstructionFoot/Ankle: Lateral Stabilization, MedialStabilization, Achilles Tendon Repair,Metatarsal Ligament Repair, Hallux Valgusreconstruction, digital tendon transfers, Mid-foot reconstructionKnee: Medial Collateral Ligament Repair,Lateral Collateral Ligament Repair, PatellarTendon Repair, Posterior Oblique LigamentRepair, Iliotibial Band TenodesisHand/Wrist: Scapholunate LigamentReconstruction, Carpal LigamentReconstruction, Repair/Reconstruction ofcollateral ligaments, Repair of Flexor andExtensor Tendons at the PIP, DIP, and MCPjoints for all digits, digital tendon transfersElbow: Biceps Tendon Reattachment, Ulnaror Radial Collateral Ligament ReconstructionHip: Capsular repair, Acetabular Labral repairSame
ClassificationnameSingle/multiple component metallic bone fixation appliances and accessoriesSingle/multiple component metallic bone fixation appliances and accessoriesSame
Trade nameFixone Biocomposite Small AnchorArthrex SutureTak Suture AnchorN/A
Model/type9 model codes including SAB-30001a3 model codes including 2.4mmx6.5mmMicro SutureTakN/A

{6}------------------------------------------------

006_510(k) Summary

ManufacturerAJU Pharm Co.,Ltd.Arthrex Inc.Remark
AppearanceImage: AJU Pharm Co., Ltd. AnchorImage: Arthrex Inc. AnchorSimilar
ProductconfigurationDriver, Anchor, Suture and NeedleDriver, Anchor, Suture and NeedleSame
MaterialPLGA + β-TCPPLDLA + β-TCPDifferent
Anchor
Outside diameter3.0mm2.4mm, 3.0mmSimilar
Length of anchor11.9mm6.5mm, 8.5mm, 14mmDifferent
Suture
AbsorbableNon-absorbableNon-absorbableSame
Suture diameterUSP size #0USP size 0 and #1Same
BiodegradableYesYesSame
Principle ofoperationManualManualSame

9. Conclusion

The device is investigated for function to compare the operation of function between Fixone Biocomposite Small Anchor and predicate devices.

Comparison results demonstrate that the specifications and performance of the device are substantially equivalent to the legally marketed predicate device.

Therefore, it is concluded that Fixone Biocomposite Small Anchor is substantially equivalent to the legally marketed predicate device.

§ 888.3030 Single/multiple component metallic bone fixation appliances and accessories.

(a)
Identification. Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.