K Number
K192023

Validate with FDA (Live)

Date Cleared
2019-12-23

(147 days)

Product Code
Regulation Number
870.1130
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Truly Automatic Arm Blood Pressure Monitor DB series, Models DB66-1, DB68 are a series devices intended to measure the systolic and diastolic blood pressure and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm.

The devices' features include irregular pulse rhythm detection during measurement, and will display a alert signal with the reading when irregular heartbeat is detected.

The devices' feature include Bluetooth function to transmit data to an external Bluetooth device with wireless communication.

Device Description

Automatic Arm Blood Pressure Monitor DB series, Models DB66-1, DB68 are designed to measure the systolic and diastolic blood pressure and pulse rate of an individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. Our method to define systolic and diastolic pressure is similar to the auscultatory method but uses an electronic pressure sensor rather than a stethoscope and mercury manometer. The sensor converts tiny alterations in cuff pressure to electrical signals, by analyzing those signals to define the systolic and diastolic blood pressure and calculating pulse rate, which is a well-known technique in the market called the "oscillometric method".

The main components of the Truly Automatic Arm Blood Pressure Monitor DB series are the main unit and cuff unit. ABS is used to outer housing of the main unit. The preformed cuff unit, which is applicable to arm circumference approximately between 220 and 340 mm, includes the inflatable bladder and nylon shell. All models of the arm blood pressure monitor use a single size of cuff. The device consists of the microprocessor, the pressure sensor, the operation keys, the pump, the electromagnetic deflation control valve and the LCD.

The devices also compares the longest and the shortest time intervals of detected pulse waves to mean time interval and displays a warning signal with the reading to indicate the detection of irregular pulse rhythm when the difference of the time intervals is over 25%.

The devices embed a Bluetooth 4.0 Wireless network connections module that allows it to connect to nearby receiving end. Once measurement is over, the LCD of device displays results, and the device will start to send out data such as systolic, diastolic, pulse, date, time with Wireless method and protocol.

AI/ML Overview

Here's an analysis of the provided text to fulfill your request, focusing on the acceptance criteria and the study that proves the device meets them.

The document describes the submission of a "Truly Automatic Arm Blood Pressure Monitor" (Models DB66-1, DB68) for 510(k) clearance, asserting its substantial equivalence to a predicate device (DB62, K161846).

Acceptance Criteria and Reported Device Performance

The core acceptance criteria for a non-invasive blood pressure monitor are its accuracy for both pressure and pulse readings. The document refers to compliance with ISO 81060-2:2013, which sets the standards for clinical validation of automated sphygmomanometers.

Table of Acceptance Criteria and Reported Device Performance:

ParameterAcceptance Criteria (from ISO 81060-2:2013, implied)Reported Device Performance (from "Table-2: The comparison table in Design and Materials")
Accuracy PressureMean difference $\le \pm 5$ mmHg and standard deviation $\le 8$ mmHg (for both systolic and diastolic blood pressure)$\pm$ 3mmHg (This is directly stated as "Accuracy Pressure" for the device, which is well within the ISO standard's requirements for mean difference, and implies the device meets the deviation as well, as confirmed by the clinical study summary.)
Accuracy Pulse$\pm$ 5%$\pm$ 5%
Measurement Pressure Range20 ~ 280 mmHg20 ~ 280 mmHg
Measurement Pulse Range40 ~ 195 beats/min40 ~ 195 beats/min
Measurement Resolution1 mmHg1 mmHg

Study Proving Device Meets Acceptance Criteria:

The document states:

"Testing Performance testing has been carried out to demonstrate that this device meets the performance specifications for its intend use. The following tests were performed on the device. Compliant to the standard of ISO 81060-2: Second Edition 2013-05-01 Non-invasive sphygmomanometers- Part 2: Clinical validation of automated measurement type. The results of this clinical investigation show that the required limits for mean difference and standard deviation are fulfilled by the subject device."

This directly indicates that a clinical validation study, adhering to the ISO 81060-2 standard, was performed, and the device successfully met its requirements.


Detailed Information about the Study:

Given the provided text, some of the requested information is not explicitly detailed. However, based on common practice for ISO 81060-2 clinical validation studies for blood pressure monitors, we can infer some aspects and state what is present in the document.

1. A table of acceptance criteria and the reported device performance:

  • See table above. The document explicitly states the device's accuracy for pressure ($\pm$ 3mmHg) and pulse ($\pm$ 5%), and confirms "the required limits for mean difference and standard deviation are fulfilled by the subject device" according to ISO 81060-2.

2. Sample size used for the test set and the data provenance:

  • Sample Size: The document does not specify the exact sample size. However, ISO 81060-2 typically requires a minimum of 85 subjects for clinical validation. It can be assumed that the study met this minimum requirement to claim compliance.
  • Data Provenance (Country of origin, retrospective/prospective): The document does not explicitly state the country of origin or if the study was retrospective or prospective, but clinical validation studies for regulatory submission are almost invariably prospective in nature.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • The document does not specify the number or qualifications of experts. For oscillometric blood pressure devices validated against ISO 81060-2, ground truth (reference blood pressure) is established by two trained observers using a mercury or auscultatory sphygmomanometer following a specific protocol. These observers are trained to accurately read the reference device.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

  • The document does not explicitly state the adjudication method. For ISO 81060-2, the ground truth is typically established by two independent observers taking simultaneous readings. If their readings differ by more than a specified amount, a third measurement (or a different method of reconciliation) might be employed, though two simultaneous, blinded observers are standard for the reference method.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • No, an MRMC comparative effectiveness study was not done. This type of study is relevant for AI-assisted diagnostic tools (e.g., radiology AI), not for a standalone non-invasive blood pressure monitor. The device is measuring a physiological parameter directly, not interpreting complex medical images for human assistance.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • Yes, in essence, this was a standalone performance test. The clinical validation per ISO 81060-2 assesses the accuracy of the automated device (algorithm + hardware) against a reference standard (manual auscultatory method). While a human applies the cuff and initiates the measurement, the device's measurement process itself is automated ("algorithm only" in the context of its function).

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

  • The ground truth was established by simultaneous auscultatory measurements performed by trained observers using a reference sphygmomanometer (typically mercury or validated aneroid) on the contralateral arm or simultaneously on the same arm if protocol allows, per ISO 81060-2. This is the "expert consensus" of blood pressure measurement via a gold standard clinical method.

8. The sample size for the training set:

  • The document does not provide information on a training set sample size. For medical devices primarily validated through performance testing against an ISO standard (like blood pressure monitors), the term "training set" in the context of machine learning (AI development) isn't directly applicable in the same way. The device's underlying "oscillometric method" is a long-established algorithm, not typically a deep learning model requiring a large training dataset in the same vein as image classification AI. The development of the algorithm might involve internal data, but it's not described here.

9. How the ground truth for the training set was established:

  • As noted above, the concept of a "training set" with ground truth in the AI sense isn't directly applicable or described for this traditional medical device. The "ground truth" for the validation (test set) study was established through the gold-standard auscultatory method by trained observers. If any internal algorithm refinement involved data, it would likely follow similar principles of comparison to expert-measured reference values.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 23, 2019

Truly Instrument Limited % Max Wong Official Correspondent Truly (U.S.A) Inc 2620 Concord Avenue, Suite 106 Alhambra, California 91803

Re: K192023

Trade/Device Name: Arm Blood Pressure Monitor Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: November 29, 2019 Received: November 29, 2019

Dear Max Wong:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

LT Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

510(k) Number (if known)

K192023

Device Name

Arm Blood Pressure Monitor

Indications for Use (Describe)

Truly Automatic Arm Blood Pressure Monitor DB series, Models DB66-1, DB68 are a series devices intended to measure the systolic and diastolic blood pressure and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm.

The devices' features include irregular pulse rhythm detection during measurement, and will display a alert signal with the reading when irregular heartbeat is detected.

The devices' feature include Bluetooth function to transmit data to an external Bluetooth device with wireless communication

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (7/17)

Page 1 of 1

PSC Publishing Services (301) 443-6740

5-1

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the text 'K192023' in a simple, sans-serif font. The text is horizontally oriented and appears to be a code or identifier. The characters are uniformly sized and spaced, making the text easily readable.

Image /page/3/Picture/1 description: The image shows the word "TRULY" in red, bold letters. To the right of the word is a circled R, indicating that the word is a registered trademark. A black line is located underneath the word.

510(k) Summary

V1.0

Date of Summary Preparation: May 21.2019

1. Submitter's Identifications

Submitter's Name:Truly Instrument Limited
Address:Site 2, Truly Industrial Area, Shanwei City,Guangdong Province, China
Contact Person:Manager Su-HaiSen
Telephone:86-0660-3363561
Fax:86-0660-3380377

2. Name of the Device

Device Classification Name:System, Measurement, Blood-Pressure, Non-invasive
Trade Name:Arm Blood Pressure Monitor
Models:DB66-1, DB68
Classification Panel:cardio-vascular
Common/Usual Name:Automatic Arm Blood Pressure Monitor
Product Code:DXN
Device Classification:Class II
Contraindications :N/A

3. The Predicate Devices

TRULY Instrument Limited.

Truly Automatic Arm Bluetooth Blood Pressure Monitor DB Series, Model DB62 510 K Number : K161846

4. Device Description

Automatic Arm Blood Pressure Monitor DB series, Models DB66-1, DB68 are designed to measure the systolic and diastolic blood pressure and pulse rate of an individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. Our method to define systolic and diastolic pressure is similar to the auscultatory method but uses an electronic pressure sensor rather than a stethoscope and mercury manometer. The sensor converts tiny alterations in cuff pressure to electrical signals, by analyzing those signals to define the systolic and diastolic blood pressure and calculating pulse rate, which is a well-known technique in the market called the "oscillometric method".

The main components of the Truly Automatic Arm Blood Pressure Monitor DB series are the main unit and cuff unit. ABS is used to outer housing of the main unit. The preformed cuff unit, which is applicable to arm circumference approximately between 220 and 340 mm, includes the inflatable bladder and nylon shell. All models of the arm blood pressure monitor use a single size of cuff. The device consists of the microprocessor, the pressure sensor, the operation keys, the pump, the electromagnetic deflation control valve and the LCD.

The devices also compares the longest and the shortest time intervals of detected pulse waves to mean time interval and displays a warning signal with the reading to indicate the detection of

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image contains the word "TRULY" in large, bold, red letters. To the right of the word, there is a circled "R" symbol, indicating that the word is a registered trademark. The word is slightly slanted to the right, giving it a dynamic appearance. The background is plain white.

irregular pulse rhythm when the difference of the time intervals is over 25%.

The devices embed a Bluetooth 4.0 Wireless network connections module that allows it to connect to nearby receiving end. Once measurement is over, the LCD of device displays results, and the device will start to send out data such as systolic, diastolic, pulse, date, time with Wireless method and protocol.

5. Intended use of device

Automatic Arm Blood Pressure Monitor DB series, Models DB66-1, DB68 are a series devices intended to measure the systolic and diastolic blood pressure and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm.

The devices' features include irregular pulse rhythm detection during measurement, and will display a alert signal with the reading when irregular heartbeat is detected.

The devices' feature include Bluetooth function to transmit data to an external Bluetooth device with wireless communication .

6. Technological Characteristics of our new DEVICE COMPARED TO THE

PREDICATE DEVICE:

The technological characteristics of Automatic Arm Blood Pressure Monitor, models DB66-1, DB68 are substantially equivalent to Truly Automatic Arm Blood Pressure Monitor DB Series, Model DB62.( K161846). There is the same Owner, TRULY instrument Limited. Which FDA owner number is 9054344. In addition, there are the same design specifications, the same form and intended to be used in the same manner that means the new devices are same as the predicate devices.

ParameterPredicate Devices DB62DB66-1DB68Explanation of thedifferences
Intended useMeasuring systolic and diastolic bloodpressure and pulse rate of adult individualMeasuring systolic and diastolic bloodpressure and pulse rate of adult individualMeasuring systolic and diastolic bloodpressure and pulse rate of adult individualIdentical
The trade nameAutomatic Arm Bluetooth Blood PressureMonitorArm Blood Pressure MonitorIdentical
510(K) NumberK161846
ApplicantTruly Instrument Limited.Truly Instrument Limited.Identical
Regulatory ClassClass IIClass IIIdentical
PanelCardiovascularCardiovascularIdentical
Product CodeDXNDXNIdentical
Indications foruseTruly Automatic Arm Bluetooth BloodPressure Monitor DB62 is intended tomeasure the systolic and diastolic bloodTruly Automatic Arm Bluetooth BloodPressure Monitor DB62 is intended tomeasure the systolic and diastolic bloodIdentical

Table-1: The comparison table

Page 2 of 6

{5}------------------------------------------------

TRULY®r
---------------
ParameterPredicate Devices DB62DB66-1DB68Explanation of the differences
pressure and pulse rate of an adultindividual by using a non-invasivetechnique in which an inflatable cuff iswrapped around the upper arm.The devices' features include irregularpulse rhythm detection duringmeasurement, and will display a alertsignal with the reading when irregularheartbeat is detected.The devices' feature include Bluetoothfunction to transmit data to an externalBluetooth device with wirelesscommunication .. Over-The-Counter Usepressure and pulse rate of an adultindividual by using a non-invasivetechnique in which an inflatable cuff iswrapped around the upper arm.The devices' features include irregularpulse rhythm detection duringmeasurement, and will display a alertsignal with the reading when irregularheartbeat is detected.The devices' feature include Bluetoothfunction to transmit data to an externalBluetooth device with wirelesscommunication .. Over-The-Counter Use
TargetPopulationAdultAdultIdentical
Anatomical sitesUpper ArmUpper ArmIdentical
Where used(hospital, home,ambulance. etc)HomeHomeIdentical
Energy used and/ or delivered4x 1.5V AA BatteryPolymer batteryDC3.7V,500mAh4x 1.5V AABatteryEquivalentThe change in thespecification isdocumented andtested,does notaffect the intendeduse or thefundamentalscientifictechnology.
Human factorsBlood pressureBlood pressureIdentical
performanceMeasuring systolic and diastolic bloodpressure and pulse rate of adult individual,Including irregular pulse rhythm detectionMeasuring systolic and diastolic bloodpressure and pulse rate of adultindividual, Including irregular pulserhythm detectionIdentical
biocompatibilityCuffAccording to ISO-10993CuffAccording to ISO-10993Identical
Compatibility withthe environmentand other devicesOperation Environment:10°C~40°C,15%90%RHStorage Environment:-20°C60°C,10%95%RHAtmospheric:70KPa106KPaOperation Environment:10°C~40°C,15%90%RHStorage Environment:-20°C60°C, 10%95%RHAtmospheric 70KPa106KPaIdentical

{6}------------------------------------------------

ParameterPredicate Devices DB62DB66-1DB68Explanation of thedifferences
Electrical safetyAccording to IEC60601-1-2According to IEC60601-1According to IEC60601-1-2According to IEC60601-1Identical
SterilizationN/AN/AIdentical
Chemical safetyN/AN/AIdentical
sterilityN/AN/AIdentical
Thermal safetyN/AN/AIdentical
designRefer to Table-2
materialsRefer to Table-2

Table-2: The comparison table in Design and Materials

ParameterPredicate Devices DB62DB66-1DB68
Measurement algorithmMethodOscillometric methodNo change ,all sameNo change ,all same
Measurement site ofbodyArmNo change ,all sameNo change ,all same
Max Cuff pressure300mmHgNo change ,all sameNo change ,all same
BluetoothBG312H04J08C
IrregularheartbeatdetectionMore than $\pm$ 25% to themean interval of pulseintervals.About the more detaileddescription of the IH detectionalgorithm, please refer to"Software validation reportI-5. Algorithm description 4.Determination method ofirregular heartbeat".No change ,all sameNo change ,all same
Measurement PressureRange20 ~ 280 mmHgNo change ,all sameNo change ,all same
Measurement PulseRange40 ~ 195 beats/minNo change ,all sameNo change ,all same
Mesauring resolution1 mmHgNo change ,all sameNo change ,all same
Accuracy Pressure$\pm$ 3mmHgNo change ,all sameNo change ,all same
Accuracy Pulse$\pm$ 5%No change ,all sameNo change ,all same
Cuff DeflationAutomatic deflationNo change ,all sameNo change ,all same
Pressurization SourceAutomatic internal pumpDC6VAutomatic internal pumpDC3VAutomatic internal pumpDC6V
Power Voltage4X 1.5VPolymer batteryDC3.7V,500mAh4x 1.5V AA Battery
Operating Environment1040℃, 1590%RHNo change ,all sameNo change ,all same
Storage Environment-20°C~60°C,10%~95%RHNo change ,all sameNo change ,all same

{7}------------------------------------------------

Truly Instrument Limited
ParameterPredicate Devices DB62DB66-1DB68
Memory Size4 x 999999
Display TypeLiquid crystal displayLiquid crystal display ,Only difference size
CoverDifference

7. Summary of Clinical study

Testing Performance testing has been carried out to demonstrate that this device meets the performance specifications for its intend use. The following tests were performed on the device. Compliant to the standard of ISO 81060-2: Second Edition 2013-05-01 Non-invasive sphygmomanometers- Part 2: Clinical validation of automated measurement type. The results of this clinical investigation show that the required limits for mean difference and standard deviation are fulfilled by the subject device.

8. Non-Clinical Tests Performed:

    1. Test Summary:
      8-1. Electric Safety , EMC and FCC test reports.
General safetyIEC 60601-1:2005+A1:2012Medical electrical equipment-Part1: General requirements for safety and essential performanceIEC60601-1-11:2015Medical electrical equipment-Part 1-11: General Requirement forbasic safety and essential performance- Collateral Standard:Requirements for medical electrical systems used in the homehealthcare environmentEN60950-1:2013Information technology equipment-Safety- Part1:Generalrequirements
EMC conformityIEC 60601-1-2:2014Medical Electrical Equipment - Part 1-2: General requirements forsafety - collateral standard: Electromagnetic compatibility -Requirements and Tests
FCC conformityFCC 47 part 15 subject B class B
ERM conformityEN301489-1:2017;EN301489-17:2017
RF conformityEN300328:2016
HealthEN62479:2010

8-2. Reliability & Clinical Test

IEC80601-2-30:20019+A1:2013Medical electrical equipment -Part 2-30: Particular requirements for the basic safety and essentialperformance of automated non-invasive sphygmomanometer
ISO 81060-2:2013Non-invasive sphygmomanometers – Part 2: Clinical validation

{8}------------------------------------------------

Image /page/8/Picture/1 description: The image contains the word "TRULY" in large, red, bold letters. To the right of the word is a small, red circle with the letter R inside, indicating a registered trademark. There is a black line underneath the word "TRULY".

of automated measurement type

8.3 Biocompatibility Test

ISO10993-1:2009Biological evaluation of medical devices-Part 1: evaluation andtesting
ISO10993-5:2009Biological evaluation of medical devices-Part 5: Test forcytotoxicity in vitro methods
ISO10993-10:2010Biological evaluation of medical devices-Part 10: Tests for irriationand sensitization

9. Conclusions

The new subject devices of Arm Blood Pressure Monitor continue to follow principles of hardware and software design of the predicate device DB62(K161846), and the feature, safety, effectiveness are also as same as DB62., Thus, the subject devices are substantially equivalent to the predicate devices.

§ 870.1130 Noninvasive blood pressure measurement system.

(a)
Identification. A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.(b)
Classification. Class II (performance standards).