K Number
K191976

Validate with FDA (Live)

Manufacturer
Date Cleared
2019-11-13

(112 days)

Product Code
Regulation Number
880.5570
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Piper GO-IO® Intraosseous Infusion System provides intraosseous access in the proximal tibia, distal tibia and humeral head (proximal humerus) of adult and pediatric patients, and the distal femur in pediatric patients when intravenous access is difficult or impossible to obtain in emergent, urgent, or medically necessary cases for up to 24 hours.

Device Description

The Piper GO-IO® Intraosseous Infusion System provides clinicians and emergency personnel with access to the intraosseous space for resuscitation and lifesaving fluid delivery for up to 24 hours. The Piper GO-IO® Intraosseous Infusion System consists of the following:

  • a single use hypodermic needle (with needle safety cap), ●
  • a powered or manual driver to assist with needle insertion, ●
  • an extension set, and;
  • an adhesive-backed securement dressing.

For insertions using the powered driver, the hypodermic needle hub that mates with a stylet connected to a drive adapter hub. The drive adapter hub includes a magnetic insert that attaches to the powered driver prior to needle insertion. The Piper GO-IO® Powered Driver is a hand-held, batterypowered device with a rechargeable lithium battery used to assist in the insertion of the subject device needle through the bone cortex. The assembly of the hypodermic needle and stylet with connected drive adapter hub is referred to as the needle set.

For insertions using the manual driver, the needle and the needle hub mate with a stylet in the same way as the needle set that is used with the powered driver, except the stylet is integrated into the handle of the manual driver instead of a drive adaptor hub (i.e. the manual driver needle assembly does not include a drive adapter hub).

The stylet was designed to include a passive safety feature to protect the placer from sharps injury. After the needle is inserted, the stylet is separated from the needle and needle hub. Upon separation of the stylet from the needle hub, the passive safety feature is released onto the stylet tip and can be safely discarded into a sharps container. Following needle insertion, the securement dressing can be applied to secure the needle hub to the skin. An extension set is available for access to the needle hub to support fluid exchange.

The subject device Piper GO-IO® Intraosseous Infusion System will be offered in needle set (for use with the powered driver) and manual driver needle kit configurations. Each kit configuration will include a securement dressing and an extension set.

AI/ML Overview

1. Table of Acceptance Criteria and Reported Device Performance

The provided document details various performance tests conducted on the Piper GO-IO® Intraosseous Infusion System. For most entries, the "acceptance criteria" are implied by the standard followed (e.g., ISO 9626:2016 for Needle Outer Diameter) and the "reported device performance" is stated as "met all predetermined acceptance criteria" within the "Summary of Substantial Equivalence" section. Specific quantitative acceptance criteria or detailed performance results are not provided in this document for individual tests.

Therefore, a table cannot be constructed with explicit quantitative acceptance criteria and reported device performance for each test. However, a summary table indicating the tests performed and the general outcome can be presented:

Performance Test CategoryStandard FollowedReported Device Performance
Needle Set Kit and Manual Driver Kit TestsVarious ISO standards and Internal ProtocolsMet all predetermined acceptance criteria
Needle Outer Diameter (OD)ISO 9626: 2016Met all predetermined acceptance criteria
Effective Needle LengthISO 7864: 2016Met all predetermined acceptance criteria
Needle LubricityISO 7864: 2016Met all predetermined acceptance criteria
Needle CleanlinessISO 9626: 2016Met all predetermined acceptance criteria
Needle to Hub Assembly TensileInternal Protocol/StandardMet all predetermined acceptance criteria
Stylet to Drive Adapter Hub TensileInternal Protocol/StandardMet all predetermined acceptance criteria
Needle and Stylet Disassembly ForceISO 23908: 2011Met all predetermined acceptance criteria
Safety ActivationFDA Guidance for Sharps Injury Prevention Features & ISO 23908: 2011Met all predetermined acceptance criteria
Stylet Safety Override (force to failure)ISO 23908: 2011Met all predetermined acceptance criteria
Securement Dressing - Pinch ForceInternal Protocol/StandardMet all predetermined acceptance criteria
Securement Dressing – Peel StrengthInternal Protocol/StandardMet all predetermined acceptance criteria
Securement Dressing – Liner RemovalInternal Protocol/StandardMet all predetermined acceptance criteria
Manual Driver Hub to Stylet TensileInternal Protocol/StandardMet all predetermined acceptance criteria
Cannula Needle Resistance to CorrosionISO 9626: 2016Met all predetermined acceptance criteria
Cannula Needle Surface FinishISO 7864: 2016Met all predetermined acceptance criteria
Needle Surface Finish and Visual AppearanceISO 9626: 2016Met all predetermined acceptance criteria
Needle Hub LuerISO 594-1: 1986 and ISO 594-2: 1998Met all predetermined acceptance criteria
Needle Hub CleanlinessISO 7864: 2016Met all predetermined acceptance criteria
Needle PointISO 7864: 2016Met all predetermined acceptance criteria
Needle Resistance to BreakageISO 9626: 2016Met all predetermined acceptance criteria
Needle StiffnessISO 9626: 2016Met all predetermined acceptance criteria
Gravity Flow RateInternal Protocol/StandardMet all predetermined acceptance criteria
Liquid Leak Needle HubInternal Protocol/StandardMet all predetermined acceptance criteria
Limits for Acidity or Alkalinity (Needle)ISO 7864: 2016Met all predetermined acceptance criteria
Limits for Extractable Metals (Needle)ISO 7864: 2016Met all predetermined acceptance criteria
Depth MarkingsInternal Protocol/StandardMet all predetermined acceptance criteria
Chemical ResistanceInternal Protocol/StandardMet all predetermined acceptance criteria
Needle DurabilityInternal Protocol/StandardMet all predetermined acceptance criteria
Manual DrillingInternal Protocol/StandardMet all predetermined acceptance criteria
Packaging Integrity and Seal StrengthISO 11607-1:2006, ASTM F88/F88M: 2015, ASTM F1886/F1886M: 2016, ASTM F1929: 2015Met all predetermined acceptance criteria
Sharps Injury Prevention Feature (in Simulated Clinical Use)FDA Guidance for Sharps Injury Prevention Features & ISO 23908: 2011Met all predetermined acceptance criteria
Powered Drill Performance TestsVarious IEC/ANSI/Internal StandardsMet all predetermined acceptance criteria
Needle Set CouplingInternal Protocol/StandardMet all predetermined acceptance criteria
Battery Capacity/IndicatorInternal Protocol/StandardMet all predetermined acceptance criteria
Battery Usable LifeInternal Protocol/StandardMet all predetermined acceptance criteria
Battery Indicator / State of ChargeInternal Protocol/StandardMet all predetermined acceptance criteria
Drill High Temperature Shut DownInternal Protocol/StandardMet all predetermined acceptance criteria
Use LifeInternal Protocol/StandardMet all predetermined acceptance criteria
Duty CycleInternal Protocol/StandardMet all predetermined acceptance criteria
Motor Stall/Stuck ShutoffInternal Protocol/StandardMet all predetermined acceptance criteria
UsabilityIEC 60601-1-6: 2013Met all predetermined acceptance criteria
Electrical Safety and Electromagnetic CompatibilityANSI AAMI ES60601-1:2005/(R)2012, A1:2012, C1:2009/(R)2012, A2:2010/(R)2012; IEC 60601-1-2: 2014; IEC 60601-1-12: 2014; IEC 62133: 2012Met all predetermined acceptance criteria
Firmware Verification and ValidationGuidance for the Content of Premarket Submissions for Software Contained in Medical DevicesMet all predetermined acceptance criteria
Sterilization, Packaging, and Shelf-Life TestsVarious ISO standards and ASTMMet all predetermined acceptance criteria
Sterilization Validation/AdoptionISO 11135:2014Met all predetermined acceptance criteria
Packaging/Shelf-Life ValidationsISO 11607-1 AMD 1: 2014; ASTM F88/F88M: 2015; ASTM F1886/F1886M: 2016; ASTM F1929: 2015Met all predetermined acceptance criteria
ResidualsISO 10993-7: 2008Met all predetermined acceptance criteria
Bacterial EndotoxinUSP <85>USP <161>Met all predetermined acceptance criteria
Biocompatibility TestsISO 10993 seriesMet all predetermined acceptance criteria
CytotoxicityISO 10993-05: 2009Met all predetermined acceptance criteria
SensitizationISO 10993-10: 2010Met all predetermined acceptance criteria
Irritation/Intracutaneous ReactivityISO 10993-10: 2010Met all predetermined acceptance criteria
Acute Systemic ToxicityISO 10993-11: 2006Met all predetermined acceptance criteria
Material Mediated PyrogenicityISO 10993-11: 2006Met all predetermined acceptance criteria
HemocompatibilityISO 10993-4: 2017Met all predetermined acceptance criteria

2. Sample Size for Test Set and Data Provenance

The document does not provide specific sample sizes for the test sets (e.g., number of needles, drivers, or tests performed on each). The provenance of the data is not specified; it refers to "performance tests completed on the subject device system" and implies the data was generated internally as part of the regulatory submission. There is no information regarding country of origin or whether the studies were retrospective or prospective beyond the stated date of preparation of the summary (November 8, 2019).

3. Number of Experts and Qualifications for Ground Truth

This information is not applicable. The device is a medical instrument (intraosseous infusion system), and the performance testing described is engineering and biological in nature, evaluating physical and material properties, and adherence to established standards. It does not involve diagnostic interpretation or human reader performance that would require experts to establish ground truth in the context of image analysis or similar AI/algorithm-driven tasks.

4. Adjudication Method for Test Set

This information is not applicable, as the performance tests are objective engineering and biological assessments against established standards, not subjective interpretations requiring adjudication.

5. Multi Reader Multi Case (MRMC) Comparative Effectiveness Study

No MRMC comparative effectiveness study was done. The document describes premarket notification for a medical device, which typically involves demonstrating substantial equivalence to a predicate device through engineering, biocompatibility, and performance testing, rather than comparative effectiveness studies with human readers and AI.

6. Standalone (Algorithm Only) Performance Study

No standalone (algorithm only) performance study was done. The device is a physical medical instrument, not an AI or software algorithm.

7. Type of Ground Truth Used

The "ground truth" for the performance tests outlined in the document is established by adherence to recognized international and national standards (e.g., ISO, ASTM, FDA guidance, USP) and internal protocols. These standards define the methods, parameters, and acceptable ranges for various tests related to material properties, mechanical performance, electrical safety, biocompatibility, and sterilization.

8. Sample Size for Training Set

This information is not applicable. The device is a hardware medical device, not an AI or machine learning model that requires a training set.

9. How Ground Truth for Training Set Was Established

This information is not applicable, as there is no training set for a hardware medical device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

November 13, 2019

Piper Access, LLC % Mark Job Official Correspondent Regulatory Technology Services, LLC 1000 Westgate Drive, Suite 510k Saint Paul, Minnesota 55114

Re: K191976

Trade/Device Name: Piper GO-IO Intraosseous Infusion System Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic single lumen needle Regulatory Class: Class II Product Code: FMI Dated: September 5, 2019 Received: September 17, 2019

Dear Mark Job:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Alan Stevens Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K191976

Device Name Piper GO-IO® Intraosseous Infusion System

Indications for Use (Describe)

The Piper GO-IO® Intraosseous Infusion System provides intraosseous access in the proximal tibia, distal tibia and humeral head (proximal humerus) of adult and pediatric patients, and the distal femur in pediatric patients when intravenous access is difficult or impossible to obtain in emergent, urgent, or medically necessary cases for up to 24 hours.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K191976

510(k) Summary 21 CFR 807.92(a)

Applicant

Name:Piper Access, LLC
Address:3981 South 700 East Suite #15Salt Lake City, UT 84107
Ph:801-210-2886
Manufacturer Contact:Jay Muse, President and CEO
Email:Jay.muse@piperaccess.com
Application Correspondent:Jacob Lee
Email:Jacob.Lee@bd.com
Date prepared:November 08, 2019

Subject Device

Trade Name:Piper GO-IO® Intraosseous Infusion System
Common Name:Intraosseous Infusion System
Classification Name:Hypodermic single lumen needle
Class:II
Regulation:21 CFR 880.5570
Product Code:FMI
Panel:General Hospital

Predicate Device

510(k):K141117 (Clearance Date July 8, 2014)
Trade Name:EZ-IO® Intraosseous Infusion System
Manufacturer:Arrow/Teleflex Medical
Common Name:Intraosseous Infusion System
Classification Name:Hypodermic single lumen needle
Class:II
Regulation:21CFR 880.5570
Product Code:FMI
Panel:General Hospital

Reason for Submission

This is a new device.

{4}------------------------------------------------

Subject Device Description

The Piper GO-IO® Intraosseous Infusion System provides clinicians and emergency personnel with access to the intraosseous space for resuscitation and lifesaving fluid delivery for up to 24 hours. The Piper GO-IO® Intraosseous Infusion System consists of the following:

  • a single use hypodermic needle (with needle safety cap), ●
  • a powered or manual driver to assist with needle insertion, ●
  • an extension set, and;
  • an adhesive-backed securement dressing.

For insertions using the powered driver, the hypodermic needle hub that mates with a stylet connected to a drive adapter hub. The drive adapter hub includes a magnetic insert that attaches to the powered driver prior to needle insertion. The Piper GO-IO® Powered Driver is a hand-held, batterypowered device with a rechargeable lithium battery used to assist in the insertion of the subject device needle through the bone cortex. The assembly of the hypodermic needle and stylet with connected drive adapter hub is referred to as the needle set.

For insertions using the manual driver, the needle and the needle hub mate with a stylet in the same way as the needle set that is used with the powered driver, except the stylet is integrated into the handle of the manual driver instead of a drive adaptor hub (i.e. the manual driver needle assembly does not include a drive adapter hub).

The stylet was designed to include a passive safety feature to protect the placer from sharps injury. After the needle is inserted, the stylet is separated from the needle and needle hub. Upon separation of the stylet from the needle hub, the passive safety feature is released onto the stylet tip and can be safely discarded into a sharps container. Following needle insertion, the securement dressing can be applied to secure the needle hub to the skin. An extension set is available for access to the needle hub to support fluid exchange.

The subject device Piper GO-IO® Intraosseous Infusion System will be offered in needle set (for use with the powered driver) and manual driver needle kit configurations. Each kit configuration will include a securement dressing and an extension set.

Intended Use

The Piper GO-IO® is intended to provide clinicians and emergency personnel with access to the intraosseous space.

Indications for Use

The Piper GO-IO® Intraosseous Infusion System provides intraosseous access in the proximal tibia, distal tibia and humeral head (proximal humerus) of adult and pediatric patients, and the distal femur in pediatric patients when intravenous access is difficult or impossible to obtain in emergent, urgent, or medically necessary cases for up to 24 hours.

Technological Characteristics

The technological characteristics of the subject Piper GO-IO® Intraosseous Infusion System are substantially equivalent with respect to the basic design and function as compared to the predicate EZ-IO® Intraosseous Infusion System. The differences between the subject and predicate devices are not critical to the intended use of the device and do not raise different questions of safety and effectiveness.

{5}------------------------------------------------

The following table provides a comparison of the technological characteristics between the subject and predicate device in sufficient detail to provide an understanding of the basis for a determination of substantial equivalence.

Subject and Predicate Device Comparison Table
AttributePiper GO-IO® IntraosseousInfusion System(Subject Device)EZ-IO® IntraosseousInfusion System(Predicate)
510(k)Subject of this PremarketNotificationK141117
Product CodeSame as predicateFMI
Intended UseSame as predicateIntended to provide clinicians andemergency personnel with access tothe intraosseous space.
Indications for UseSame as predicate with theexception of the subject devicebrand nameThe EZ-IO® Intraosseous InfusionSystem provides intraosseous accessin the proximal tibia, distal tibia andhumeral head (proximal humerus) ofadult and pediatric patients, and thedistal femur in pediatric patientswhen intravenous access is difficultor impossible to obtain in emergent,urgent, or medically necessary casesfor up to 24 hours.
Target PatientPopulationSame as predicateAdults and Pediatrics
AnatomicalInsertion SiteSame as predicateAdults: Proximal tibia, distal tibia,proximal humerusPediatrics: Proximal tibia, distaltibia, proximal humerus, distalfemur
Primary IO SystemComponentsSame as predicate with theexception of the subject stylet,which includes a passive safetysharps prevention feature• Hypodermic Needle w/Stylet· Needle Safety Cap· Securement Dressing· Extension Set• Powered Driver· Manual Driver
Needle: Dwell TimeSame as predicate24 hours or less
Needle: UseSame as predicateSingle Use
Needle Lengths15mm (3-39kg)25mm (>3kg)45mm (>40kg)15 mm (3-39kg)25 mm (3kg and over)45 mm (40kg and over)
Subject and Predicate Device Comparison Table
AttributePiper GO-IO® IntraosseousInfusion System(Subject Device)EZ-IO® IntraosseousInfusion System(Predicate)
Needle: OuterDiameterSame as predicate15 gauge
Needle: MaterialsSame as predicate304 Stainless Steel
Needle: Tip DesignTouhy/Huber Style Needle TipFaceted Tip
Needle: DepthMarkersSame as predicateDepth markers every 1 cm
Needle: HubMaterialSame as predicateMedical grade polycarbonate
Needle: HubConnectionSame as predicateStandard Luer Lock
Stylet: MaterialsSame as predicateStainless Steel
Stylet: SharpsInjury PreventionFeatureIncludes a stylet tip safety featureDoes not include a sharpsprevention feature on the stylet
Drive Adapter Hub:MaterialsSame as predicatePolycarbonate and stainless steel
Inclusion of aNeedle ProtectiveCoverYes, includes a cover made ofpolypropyleneYes, includes a cover made ofunknown material
Needle SetSterilization Method& SALSame as predicateEO, 10-6
Manual DriverAttachmentManual driver handle withintegrated stylet mates with internallumen of needle and needle hubattaches to manual driverManual handle attaches toneedle set (i.e. assembly of needleand stylet) with magnet
Manual DriverComponentMaterialsHandle: ABSStylet: Stainless Steel (304)Handle: PolycarbonateStylet: The manual handle does notinclude an integrated stylet
Manual DriverSterilization MethodSame as predicateEO
Manual Driver SALSame as predicate10-6
Powered DriverFeaturesSame as predicateCordless, Battery-powered
Powered Driver UseSame as predicateReusable
Powered DriverMaterialsCopolyesterPolycarbonate
Powered DriverEnergy SourceRechargeable Lithium BatteriesLithium Batteries
Subject and Predicate Device Comparison Table
AttributePiper GO-IO® IntraosseousInfusion System(Subject Device)EZ-IO® IntraosseousInfusion System(Predicate)
Powered DriverBattery LightIndicator4 battery light indicators torepresent battery charge levelOne battery light indicator
Powered DriverNeedle AttachmentSame as predicateMagnetic
Powered DriverCleaning MethodSame as predicateHigh level disinfectant
Powered Driver:Type of protectionagainst electricshockInternally Powered and Class IIInternally Powered
Powered Driver:Degree of protectionagainst electricshockSame as predicateType BF applied part
Powered Driver:Degree of protectionagainst ingress ofwaterIP33, Spraying water and solidforeign body protection (objects>2.5mm diameter)IPX0 Ordinary Protection
Means to InsertNeedleSame as predicateManual or Powered Driver
General Method ofInsertionSame as predicatePush needle through soft tissue untilit contacts bone. Confirm depthmarkings. Insert needle set throughbone until change in pressure is feltor to desired depth. Remove stylet.Connect IV extension set.
Degree of safetyor application inthe presence of aflammableanestheticmixture:Not intended for use in an oxygenrich environmentEquipment not suitable for use inthe presence of a flammableanesthetic mixture with air or withoxygen or nitrous oxide

{6}------------------------------------------------

{7}------------------------------------------------

{8}------------------------------------------------

Subject and Predicate Device Comparison Table
AttributePiper GO-IO® IntraosseousInfusion System(Subject Device)EZ-IO® IntraosseousInfusion System(Predicate)
Expected ServiceLife (PoweredDriver)Expected service life isapproximately 400 insertions.Service life expectancy isdependent on actual usage (bonedensity and average insertion time),storage, and frequency of testing.The predicate device is warranted toa service life of 500 insertions.Service Life expectancy isdependent on actual usage (bonedensity and average insertiontime), storage, and frequency oftesting.
ElectromagneticImmunity andEmissions (IEC60601-1-2)Same as predicateComplies with standard
Maximum RPMSame as predicate1587 or less

{9}------------------------------------------------

Performance Testing

The following tables identify the performance tests completed on the subject device system, including the standard followed for each test.

Needle Set Kit and Manual Driver KitPerformance TestsStandard Followed
Needle Outer Diameter (OD)ISO 9626: 2016
Effective Needle LengthISO 7864: 2016
Needle LubricityISO 7864: 2016
Needle CleanlinessISO 9626: 2016
Needle to Hub Assembly TensileInternal Protocol/Standard
Stylet to Drive Adapter Hub TensileInternal Protocol/Standard
Needle and Stylet Disassembly ForceISO 23908: 2011
Safety ActivationFDA Guidance for Sharps Injury PreventionFeatures & ISO 23908: 2011
Stylet Safety Override (force to failure)ISO 23908: 2011
Securement Dressing - Pinch ForceInternal Protocol/Standard
Securement Dressing – Peel StrengthInternal Protocol/Standard
Securement Dressing – Liner RemovalInternal Protocol/Standard
Manual Driver Hub to Stylet TensileInternal Protocol/Standard
Cannula Needle Resistance to CorrosionISO 9626: 2016
Cannula Needle Surface FinishISO 7864: 2016
Needle Surface Finish and Visual AppearanceISO 9626: 2016
Needle Hub LuerISO 594-1: 1986 and ISO 594-2: 1998
Needle Hub CleanlinessISO 7864: 2016
Needle PointISO 7864: 2016
Needle Resistance to BreakageISO 9626: 2016
Needle StiffnessISO 9626: 2016
Gravity Flow RateInternal Protocol/Standard
Liquid Leak Needle HubInternal Protocol/Standard
Limits for Acidity or Alkalinity (Needle)ISO 7864: 2016
Limits for Extractable Metals (Needle)ISO 7864: 2016
Depth MarkingsInternal Protocol/Standard
Chemical ResistanceInternal Protocol/Standard
Needle DurabilityInternal Protocol/Standard
Manual DrillingInternal Protocol/Standard
Packaging Integrity and Seal StrengthISO 11607-1:2006ASTM F88/F88M: 2015ASTM F1886/F1886M: 2016ASTM F1929: 2015
Sharps Injury Prevention Feature (in SimulatedClinical Use)FDA Guidance for Sharps Injury PreventionFeatures & ISO 23908: 2011

{10}------------------------------------------------

Powered Drill Performance TestsStandards Followed
Needle Set CouplingInternal Protocol/Standard
Battery Capacity/IndicatorInternal Protocol/Standard
Battery Usable LifeInternal Protocol/Standard
Battery Indicator / State of ChargeInternal Protocol/Standard
Drill High Temperature Shut DownInternal Protocol/Standard
Use LifeInternal Protocol/Standard
Duty CycleInternal Protocol/Standard
Motor Stall/Stuck ShutoffInternal Protocol/Standard
UsabilityIEC 60601-1-6: 2013
Electrical Safety and ElectromagneticCompatibility• ANSI AAMI ES60601-1:2005/(R)2012,A1:2012, C1:2009/(R)2012,A2:2010/(R)2012• IEC 60601-1-2: 2014• IEC 60601-1-12: 2014• IEC 62133: 2012
Firmware Verification and ValidationGuidance for the Content ofPremarket Submissions for Software Containedin Medical Devices
Sterilization, Packaging, and Shelf-LifeStandard Followed
Sterilization Validation/AdoptionISO 11135:2014
Packaging/Shelf-Life ValidationsISO 11607-1 AMD 1: 2014ASTM F88/F88M: 2015ASTM F1886/F1886M: 2016ASTM F1929: 2015
ResidualsISO 10993-7: 2008
Bacterial EndotoxinUSP <85>USP <161>

A biocompatibility evaluation was conducted on the subject device per ISO 10993-1:2009, Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process. According to the evaluation, the biological tests in the table below were conducted.

Biological EndpointStandard Followed
CytotoxicityISO 10993-05: 2009
SensitizationISO 10993-10: 2010
Irritation/Intracutaneous Reactivity
Acute Systemic ToxicityISO 10993-11: 2006
Material Mediated Pyrogenicity
HemocompatibilityISO 10993-4: 2017

{11}------------------------------------------------

Summary of Substantial Equivalence

The subject device, Piper GO-IO® Intraosseous Infusion System, has the same intended use and the same fundamental scientific technology as the predicate device, EZ-IO Intraosseous Infusion System. The results of performance and biological tests conducted on the Piper GO-IO® Intraosseous Infusion System met all predetermined acceptance criteria and demonstrated that the different technological characteristics of the subject device do not raise different questions of safety and effectiveness. Based on the intended use, technological characteristics, performance and biological test results, the Piper GO-IO® Intraosseous Infusion System can be considered substantially equivalent to the cited predicate device.

§ 880.5570 Hypodermic single lumen needle.

(a)
Identification. A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.(b)
Classification. Class II (performance standards).