K Number
K191911

Validate with FDA (Live)

Manufacturer
Date Cleared
2019-11-27

(133 days)

Product Code
Regulation Number
872.5470
Panel
Dental
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

3Shape Splint Design for splints and mouthguards is intended for use as a medical front-end device providing tools for systematic inspection, detailed analysis, treatment simulation and virtual appliance design based of 3D models of the patient's dentition before the start of an orthodontic treatment.

The use of Splint Design software requires the user to have the necessary training and domain knowledge in the practice of orthodontics, as well as to have received a dedicated training in the use of the software.

Device Description

The Splint design software for splints and mouthquards/nightquards is a software system used for the virtual planning of orthodontic treatments by simulating jaw movements, and design of orthodontic appliances based on 3D scanned orthodontic models.

The realization of the above-mentioned orthodontic appliances is done using splint workflow which supports

  • Direct production using 3D printer or milling machines with FDA cleared or . listed materials included in the splint workflow.
    The actual production of appliances is done by the user and it is under their control and responsibility.

3Shape Splint Design has no patient contact being a software only device.

AI/ML Overview

I am sorry, but the provided text does not contain the detailed information needed to answer your request about acceptance criteria and the study proving the device meets them. The document is a 510(k) summary for a medical device (3Shape Splint Design) stating that it is substantially equivalent to a predicate device.

Specifically, the text does not include:

  1. A table of acceptance criteria and reported device performance.
  2. Details on sample size, data provenance, or the nature of any test set used.
  3. Information about experts used to establish ground truth or their qualifications.
  4. Adjudication methods.
  5. A multi-reader multi-case (MRMC) comparative effectiveness study, nor any effect size for human readers with and without AI assistance.
  6. Information about a standalone (algorithm only) performance study.
  7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.).
  8. The sample size for the training set.
  9. How the ground truth for the training set was established.

The document only states that "Software, hardware, and integration verification and validation testing was performed" and "All test results have been reviewed and approved, showing the Ortho System™ to be substantially equivalent to the primary predicate device." It also explicitly states that "Clinical testing is not a requirement and has not been performed."

Therefore, I cannot provide the requested information based on the given input.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 27, 2019

3Shape A/S Shreyosi Chakraborty Regulatory Affairs Specialist Holmens Kanal 7 DK-1060 Copenhagen K

Re: K191911

Trade/Device Name: 3Shape Splint Design Regulation Number: 21 CFR 872.5470 Regulation Name: Orthodontic plastic bracket Regulatory Class: Class II Product Code: PNN Dated: September 10, 2019 Received: September 13, 2019

Dear Shreyosi Chakraborty:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Srinivas Nandkumar, Ph.D. Acting Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K191911

Device Name 3Shape Splint Design

Indications for Use (Describe)

3Shape Splint Design for splints and mouthguards is intended for use as a medical front-end device providing tools for systematic inspection, detailed analysis, treatment simulation and virtual appliance design based of 3D models of the patient's dentition before the start of an orthodontic treatment.

The use of Splint Design software requires the user to have the necessary training and domain knowledge in the practice of orthodontics, as well as to have received a dedicated training in the use of the software.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(K) SUMMARY K191911

Submitter Information

Company Name:3Shape A/S
Company Address:Holmens Kanal 7DK-1060 Copenhagen K
Company Phone:+45 7027 2620
Company Fax:+45 7027 2621
Contact Person:Shreyosi ChakrabortyRegulatory Affairs Specialist
Date Summary Prepared:September 3, 2019

Device Identification

510(k) number:K191911
Trade/proprietary Name:3Shape Splint Design
Regulation Number:872.5470
Classification:Class 2
Product Code:PNN (Orthodontic Software)

Primary Predicate Device

The Splint design software for splints and mouthguards (K191911), based on the information and supporting documentation provided, and has the same intended use, scientific concept and technical characteristics as the primary predicate device 3Shape Ortho System™ - Splint (K161884) also manufactured by 3Shape A/S.

Both software devices are used by Dental Professionals in orthodontic treatment planning (before, during, after treatment) covering inspection, 2D measurement and analysis of orthodontic appliances, treatment simulation with virtual articulation, as well as virtual appliance design, handling and export, and they are both providing CAM output for 3D printers and milling machines.

Specifically, the functionality included in the new, the proposed device to realize splints and mouthguards using of digital design manufactured by approved materials are identical for the two devices. Therefore, the Splint Design software (K191911) and the predicate (K161884) are found to be similar in their intended use, supported anatomic areas and the available relevant features and functionalities.

Indications for Use

3Shape Splint Design for splints and mouthquards/nightquards is intended for use as a medical front-end device providing tools for systematic inspection, detailed

{4}------------------------------------------------

analysis, treatment simulation and virtual appliance design based on 3D models of the patient's dentition before the start of an orthodontic treatment.

The use of Splint Design software requires the user to have the necessary training and domain knowledge in the practice of orthodontics, as well as to have received a dedicated training in the use of the software.

Device Description

The Splint design software for splints and mouthquards/nightquards is a software system used for the virtual planning of orthodontic treatments by simulating jaw movements, and design of orthodontic appliances based on 3D scanned orthodontic models.

The realization of the above-mentioned orthodontic appliances is done using splint workflow which supports

  • Direct production using 3D printer or milling machines with FDA cleared or . listed materials included in the splint workflow.
    The actual production of appliances is done by the user and it is under their control and responsibility.

3Shape Splint Design has no patient contact being a software only device.

Scientific Concept

The underlying scientific concept of the Splint Design software is to apply digital imaging tools for orthodontic treatment planning and CAD design of customized appliances.

The system supports the following types of digital data: dcm and STL

Summary of the technological characteristics

Ortho System™ is a software only device programmed in C# and Delphi and has the following PC/laptop hardware requirements equivalent to the reference devices:

ItemMinimum RequirementsSplint Design(K19XXXX)Minimum Requirements3Shape Ortho System™(K161884)
OS:Windows 7, 8 or 10 64-bitWindows 7 or 8, 64-bit
RAM:8 GB8 GB
MonitorResolution:1280x800 or similar1280x800 or similar
Video CardMemory:1 GB1 GB
AvailableHDD Space:250 GB250 GB
CPU:Intel Core i5 or equivalentIntel Core i5 or equivalent
Network:Network Internet connectionNetwork Internet connection

{5}------------------------------------------------

Mouse:With wheel buttonWith wheel button
----------------------------------------------

351

The Splint Design software has the same intended uses and technical characteristics as the Ortho System™ software (K161884) also manufactured by 3Shape A/S:

Feature ID & NameSplintDesignK191911OrthoSystem™K161884Applicable for SplintandMouthguard/Nightguard TreatmentSimulation
Supported anatomic areasMaxillaMandibleMaxillaMandibleYesYes
Intended use
1. Managing patient and case basedataNoYesNo
2. Collection of study materialYesYesYes
3. Alignment of study materialNoYesNo
4. Measuring study materialYesYesYes
5. Analyzing study materialYesYesYes
6. Treatment simulationYesYesYes
7. Virtual appliance designYesYesYes
8. Supported PC formatsWindowsWindows
Managing patient and case base data
1.1 Creating, editing, deleting andcopying patient dataNoYesNo
1.2 Creating, editing, deleting andcopying case dataNoYesNo
Collection of study material
2.1 Surface scan for intra-oralscannerYesYesYes
2.2 Surface scan from STL fileYesYesYes
2.3 CT image dataNoDICOMNo
2.4 2D overlayNo (PNG, JPG,BMP)PNG, JPG,BMPNo
Alignment of study material
3.1 Aligning surface scan and CTimageNoYesNo
3.2 Aligning cephalometric imagesNoYesNo
3.3 Alignment of 2D overlays (e.g.ideal arch)NoYesNo
3.4 Ability to check/adjust DICOMvisibilityNoYesNo
3.5 DICOM scan segmentationNoNoNo
Measuring study material
4.1 2D measurement toolboxYesYesYes
4.2 3D measurement toolboxNoYesNo
Analyzing study material
5.1 Arch shapeNoYesNo
5.2 Wire lengthNoYesNo
5.3 Tooth widthNoYesNo
5.4 BoltonNoYesNo
5.5 Space analysisNoYesNo
5.6 Overjet/overbiteNoYesNo
5.7 Occlusion mapYesYesYes
Treatment simulation
6.1 2D & 3D simulationNoYesNo
6.2 Virtual articulatorYesYesYes
Virtual appliance design
7.1 Orthodontic appliance searchNoYesNo
7.2 Orthodontic appliance virtualpreparationNoYesNo
7.3 Orthodontic appliance designYesYesYes
7.4 Orthodontic appliance exportYesYesYes

{6}------------------------------------------------

3SHAPE SPLINT DESIGN SOFTWARE - 510(K) SUBMISSION

ssha

{7}------------------------------------------------

ssha

Nonclinical Testing

Software, hardware, and integration verification and validation testing was performed in accordance with the FDA Guidance Document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" (Issued on May 11, 2005).

The validation suite includes validation of implemented mitigations related to device hazards identified in the risk management procedures.

All test results have been reviewed and approved, showing the Ortho System™ to be substantially equivalent to the primary predicate device.

Clinical Testing

Clinical testing is not a requirement and has not been performed.

Conclusion

Based on a comparison of intended use, indications, principle of operations, features and technical data, and the test results, the Splint Design software for splints and mouthguards/nightguards is found to be as safe and as effective as the primary predicate device. Therefore, intended use and performance are found to be substantially equivalent to those of the primary predicate device.

§ 872.5470 Orthodontic plastic bracket.

(a)
Identification. An orthodontic plastic bracket is a plastic device intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.(b)
Classification. Class II.