K Number
K191854

Validate with FDA (Live)

Manufacturer
Date Cleared
2019-10-08

(90 days)

Product Code
Regulation Number
882.4560
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Premia Spine XL Instruments are intended to be used during the preparation and placement of Premia Spine screws during surgery to assist the surgeon in precisely locating anatomical structure in either open or minimally invasive procedures.

Premia Spine XL Instruments are specifically designed for use with the Medtronic StealthStation System, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as vertebra, can be identified relative to a CT or MR based model, fluoroscopy images, or digitized landmarks for the anatomy.

The XL Instruments are nonsterile instruments that can be used manually. These instruments are intended to be used with the Medtronic's StealthStation Navigation System.

Device Description

The XL Instruments is comprised of seven instruments:

  • Cannulated Screwdriver;
  • Cannulated Pedicle AWL;
  • Cannulated Pedicle Probe;
  • Cannulated TAP XL 4.5, 5.5 and 6.5 mm;
  • Straight Pedicle Probe.

The XL Instruments are reusable stereotaxic surgical instruments, designed to be used with the Medtronic StealthStation navigation system.

They are reusable (autoclavable) instruments, intended to be used in a sterile environment, and initially supplied non-sterile to the user and requires the user to process (i.e., clean and sterilize) the device for initial use, as well as after each use.

AI/ML Overview

The Premia Spine XL Instruments are a set of reusable stereotaxic surgical instruments intended for use with the Medtronic StealthStation navigation system. The 510(k) submission (K191854) claims substantial equivalence to the Medtronic Navigated Reusable Instruments for Use with the StealthStation® and IPC® Powerease™ Systems (Predicate K153442).

Here's an analysis of the acceptance criteria and study information provided:

1. Table of Acceptance Criteria and Reported Device Performance:

Acceptance CriteriaReported Device Performance
Registering and calibration of the instruments on the StealthStationAll tests met predefined acceptance criteria.
Failure mode testingAll tests met predefined acceptance criteria.
Interfacing of each instrument with the StealthStation, per ASTM F-2554 (Standard Specification for Image-Guided Surgical Tool Accuracy)All tests met predefined acceptance criteria.
Equivalence of the pre-submitted geometric properties of the predicate and subject devicesThe document states: - "The system features sufficiently equivalent geometries to the predicate such that the performance of the subject and predicate devices are equivalent." - "Critical dimensions: Identical for Tap 4.5 mm, Tap 5.5 mm, Tap 6.5 mm, Pedicle awl, Pedicle probe; Equivalent for screwdriver." - "The differences of geometry between the subject and the predicate devices are supported by the comparison testing and do not raise different questions of safety and effectiveness." All tests met predefined acceptance criteria.

2. Sample Size for the Test Set and Data Provenance:

The document describes bench testing, which typically does not involve human subjects or patient data. Therefore, there is no "test set" in the traditional sense of patient data. The tests were performed on the instruments themselves.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

Not applicable, as this was bench testing of surgical instruments and their compatibility with a navigation system. Ground truth would be established by engineering specifications and objective measurements based on the test protocols.

4. Adjudication Method for the Test Set:

Not applicable, as this was bench testing. Test results would be compared against predefined engineering specifications and criteria.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If So, what was the effect size of how much human readers improve with AI vs without AI assistance:

Not applicable. This device is a surgical instrument set cleared under a 510(k) for substantial equivalence to a predicate device. It is not an AI/ML-driven diagnostic or interpretative tool that would typically undergo an MRMC study.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

Not applicable. This is a physical surgical instrument and not an algorithm. The performance evaluation focuses on the instrument's mechanical properties, its ability to be registered and calibrated by the navigation system, and its interface compatibility.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

For bench testing of surgical instruments, the "ground truth" is defined by engineering specifications, design tolerances, and established performance standards (e.g., those outlined in ASTM F-2554). The instruments' ability to accurately interact with the Medtronic StealthStation system and meet dimensional requirements constitutes the ground truth.

8. The Sample Size for the Training Set:

Not applicable. This is a physical medical device (surgical instruments), not an AI/ML model that requires a training set.

9. How the Ground Truth for the Training Set was Established:

Not applicable, for the same reason as point 8.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

October 8, 2019

Premia Spine, Ltd. % Janice Hogan Partner Hogan Lovells US LLP 1735 Market Street Suite 2300 Philadelphia, Pennsylvania 19103

Re: K191854

Trade/Device Name: Premia Spine XL Instruments Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OLO Dated: July 10, 2019 Received: July 10, 2019

Dear Janice Hogan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Shumaya Ali, MPH Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

510(k) Number (if known)

K191854

Device Name

Premia Spine XL Instruments

Indications for Use (Describe)

Premia Spine XL Instruments are intended to be used during the preparation and placement of Premia Spine screws during surgery to assist the surgeon in precisely locating anatomical structure in either open or minimally invasive procedures.

Premia Spine XL Instruments are specifically designed for use with the Medtronic StealthStation System, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as vertebra, can be identified relative to a CT or MR based model, fluoroscopy images, or diqitized landmarks for the anatomy.

The XL Instruments are nonsterile instruments that can be used manually. These instruments are intended to be used with the Medtronic's StealthStation Navigation System.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K191854 Premia Spine Ltd.'s XL Instruments

1. Submitter's Identification

Premia Spine Ltd. 7 Giborey Israel Street Ramat Poleg, 42504 Israel Phone: +972-72-228-1200 Contact Person: Ron Sacher

Date Prepared: October 7, 2019

2. Name of Device

Name of Device: XL Instruments Common or Usual Name: XL navigation instruments Classification: 21 CFR 882.4560 Classification Name: Stereotaxic instrument Regulatory Class: II Product Code: OLO Product Code Name: Orthopedic Stereotaxic Instrument

3. Predicate and Reference Device Information

Predicate: K153442, Medtronic Navigated Reusable Instruments for Use with the StealthStation® and IPC® Powerease™ Systems

4. Device Description

The XL Instruments is comprised of seven instruments:

  • · Cannulated Screwdriver;
  • Cannulated Pedicle AWL;
  • · Cannulated Pedicle Probe;
  • · Cannulated TAP XL 4.5, 5.5 and 6.5 mm;
  • · Straight Pedicle Probe.

The XL Instruments are reusable stereotaxic surgical instruments, designed to be used with the Medtronic StealthStation navigation system.

They are reusable (autoclavable) instruments, intended to be used in a sterile environment, and initially supplied non-sterile to the user and requires the user to process (i.e., clean and sterilize) the device for initial use, as well as after each use.

5. Indications for Use

Premia Spine XL Instruments are intended to be used during the preparation and placement of Premia Spine screws during surgery to assist the surgeon in precisely locating anatomical structure in either open or minimally invasive procedures.

{4}------------------------------------------------

Premia Spine XL Instruments are specifically designed for use with the Medtronic StealthStation System, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as vertebra, can be identified relative to a CT or MR based model, fluoroscopy images, or digitized landmarks for the anatomy.

The XL Instruments are nonsterile instruments that can be used manually. These instruments are intended to be used with the Medtronic's StealthStation Navigation System.

6. Summary of Technological Characteristics and Comparison

The clinical set-up and mode of operation of the subject device are similar to the proposed predicate (K153442).

Both the predicate and the subject devices are composed of several instruments designed to be registered and then identified by the Medtronic StealthStation to allow the surgeon using them to precisely locate anatomical structures during stereotaxic surgical procedures.

The principles of operation of the XL instruments are identical to those of the predicate.

The technical specifications of the subject and predicate devices are similar. The system features sufficiently equivalent geometries to the predicate such that the performance of the subject and predicate devices are equivalent.

ParameterXL InstrumentsMedtronic Instruments
Indication forUsePremia Spine XL Instruments areintended to be used during thepreparation and placement of PremiaSpine screws during surgery to assistthe surgeon in precisely locatinganatomical structure in either open orminimally invasive procedures.Medtronic Navigated ReusableInstruments are intended to be usedduring the preparation and placement ofMedtronic screws during spinal surgeryto assist the surgeon in preciselylocating anatomical structures in eitheropen, or minimally invasive,procedures.
Premia Spine XL Instruments arespecifically designed for use with theMedtronic StealthStation System, whichis indicated for any medical condition inwhich the use of stereotactic surgerymay be appropriate, and wherereference to a rigid anatomicalstructure, such as vertebra, can beidentified relative to a CT or MR basedmodel, fluoroscopy images, or digitizedlandmarks for the anatomy.Medtronic Navigated ReusableInstruments are specifically designedfor use with the StealthStation® System,which is indicated for any medicalcondition in which the use ofstereotactic surgery may beappropriate, and where reference to arigid anatomical structure, such as askull, a long bone, or vertebra can beidentified relative to a CT or MR-basedmodel, fluoroscopy images, or digitizedlandmarks of the anatomy.
The XL Instruments are nonsterileinstruments that can be used manually.These instruments are intended to beused with the Medtronic'sStealthStation Navigation System.

{5}------------------------------------------------

ParameterXL InstrumentsMedtronic Instruments
ClassIIII
Regulation21 CFR §882.456021 CFR §882.4560
CodeOLOOLO
510k numberSubjectK153442
EnvironmentSurgery operating roomSurgery operating room
PrescriptionRx PrescriptionRx Prescription
Critical dimensionsIdentical for Tap 4.5 mm, Tap 5.5 mm, Tap 6.5 mm, Pedicle awl, Pedicle probe;Equivalent for screwdriver
Instrument RegistrationIdentical
Instrument CalibrationIdentical
Instrument InterfaceIdentical
Tracking methodIdentical
MaterialStainless steelStainless steel
SterilizationTo be steam sterilized before useTo be steam sterilized before use

Table VII-1: Comparison Table

7. Summary of Performance Data

Bench testing was performed per the following voluntary performance standards or FDA guidance:

  • Registering and calibration of the instruments on the StealthStation testing; ●
  • Failure mode testing; ●
  • . Interfacing of each instrument with the StealthStation testing, per ASTM F-2554;
  • Equivalence of the pre-submitted geometric properties of the predicate and subject devices.

All tests met the predefined acceptance criteria.

8. Conclusions

The XL Instruments set has the same intended use and the same basic technological characteristics of the predicate device, Medtronic Navigated Reusable Instruments for Use with the StealthStation® and IPC® Powerease™ Systems. The differences of geometry between the subject and the predicate devices are supported by the comparison testing and do not raise different questions of safety and effectiveness. Performance testing has demonstrated that the XL Instruments set is as safe and effective as the predicate. Therefore, the XL Instruments set is substantially equivalent to the predicate device.

§ 882.4560 Stereotaxic instrument.

(a)
Identification. A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.(b)
Classification. Class II (performance standards).