(261 days)
The gammaCore Sapphire Non-invasive Vagus Nerve Stimulator is intended to provide non-invasive vagus nerve stimulation (nVNS) on the side of the neck. The gammaCore Sapphire device is indicated for:
- · Adjunctive use for the preventive treatment of cluster headache in adult patients.
- · The acute treatment of pain associated with episodic cluster headache in adult patients.
- · The acute treatment of pain associated with migraine headache in adult patients.
- · The preventive treatment of migraine headache in adult patients.
The gammaCore Sapphire (gammaCore) is a multiuse, handheld, rechargeable, portable device consisting of a rechargeable battery and signal-generating and amplifying electronics, with a slide control switch for user/operator control of the signal amplitude (relative range, 0-40 continuous).
The gammaCore Sapphire:
- . Includes a charging station incorporated into the "clam shell" storage case connected to a power adapter for charging of the device as necessary by the end user.
- Provides visible (light and display) and audible (beep) feedback regarding device and . stimulation status.
- Allows for multiple stimulations or doses; each stimulation or dose lasts 120 seconds, . after which the device automatically turns off unless turned off earlier by the user/operator.
- Delivers up to a fixed number of doses within a 24-hour period; once the maximum daily . number of doses has been reached, the device will not deliver any more doses until the following 24-hour period.
- Indicates on the display the number of remaining doses available in a 24-hour period. .
This document is a 510(k) summary for the gammaCore Sapphire device and doesn't explicitly state "acceptance criteria" in a quantitative table. Instead, it demonstrates substantial equivalence to a predicate device by proving that the expansion of indications for use (specifically, for the preventive treatment of migraine headache) does not raise new questions of safety or effectiveness.
The "acceptance criteria" effectively are that the new indication performs similarly in terms of safety and efficacy to the existing cleared indications and comparable to the predicate device, as demonstrated through clinical studies.
Here's a breakdown of the requested information based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
As noted, the document doesn't provide a formal table of "acceptance criteria" and "reported device performance" in the typical sense of a pass/fail metric for a specific algorithm. Instead, it presents clinical study results to demonstrate that the expanded indication for the gammaCore Sapphire device (preventive treatment of migraine headache) is safe and effective and does not raise new questions of safety or effectiveness compared to the predicate device and previously cleared indications.
The key "performance" is the therapeutic gain and P-value for migraine reduction and the safety profile demonstrated in the clinical trials.
| Acceptance Criteria (Inferred from Substantial Equivalence Claim) | Reported Device Performance (from Clinical Studies) |
|---|---|
| No new safety concerns for nVNS use in the intended population. | In both PREMIUM and EVENT studies, nVNS was safe and well tolerated. No serious device-related adverse events (AEs) were reported. Most adverse events were mild or moderate and transient. |
| Non-inferior efficacy for preventive migraine treatment. | PREMIUM Study (mITT population for preventive migraine): - Therapeutic gain in mean reduction of migraine days per month: -0.74 (P=0.043) - Therapeutic gain in mean reduction of headache days per month: -0.86 (P=0.045) - Therapeutic gain in mean reduction of acute medication days per month: -0.80 (P=0.039) EVENT Study (for preventive chronic migraine): - Change in mean headache days from baseline to month 2: -1.4 (nVNS) vs. -0.2 (sham) - Continued nVNS use led to significant reductions over time. |
| Technological characteristics, intended use, and other device specifications remain unchanged or demonstrably equivalent. | "There have been no changes in the technological characteristics or intended use of the gammaCore Sapphire in the proposed Indications for Use statement." (Detailed comparison in Table 5 showing no changes in patient-contacting materials, electrical classification, waveform/frequency, maximum output, load impedance, power supply, service life, controls, output regulation, display, charger, RFID, Bluetooth, and alarm signals). |
2. Sample Size Used for the Test Set and Data Provenance
The "test set" in this context refers to the clinical study populations used to support the expanded indication.
-
PREMIUM Study:
- ITT (Intent-to-treat) Population: 332 subjects (ITT nVNS: 165, ITT Sham: 167)
- mITT (modified Intent-to-treat) Population: 278 subjects (mITT nVNS: 138, mITT Sham: 140)
- Data Provenance: Prospective, randomized, double-blind, sham-controlled, multicenter study conducted at 22 European sites from June 1, 2015, to November 21, 2017.
-
EVENT Study:
- ITT (Intent-to-treat) Population: 59 subjects (ITT nVNS: 30, ITT Sham: 29)
- PP (Per-protocol) Population: 49 subjects (PP nVNS: 26, PP Sham: 23)
- Data Provenance: Prospective, randomized, sham-controlled, multicenter pilot study for the prevention of chronic migraine, conducted at 6 sites in the United States from October 2012 to April 2014.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
Clinical trials for medical devices like this typically rely on patient-reported outcomes validated by medical professionals and objective measures (like headache day diaries). The "ground truth" here is less about image interpretation by experts and more about the clinical definition of migraine, headache days, and adverse events, as assessed by the investigators and patients themselves.
The document does not specify a separate panel of experts establishing ground truth in the way it might for an AI-powered diagnostic device. Instead, the "ground truth" for efficacy (e.g., reduction in migraine days) is derived from the comprehensive data collected during the trials, including patient diaries and clinical assessments by the investigators at the study sites. The inclusion criteria for patients required a diagnosis of migraine with or without aura in accordance with international classification criteria (ICHD-3 Beta for PREMIUM, ICHD-2 for EVENT), implying that the diagnosing clinicians (experts) established the baseline condition.
4. Adjudication Method for the Test Set
This type of information (e.g., 2+1, 3+1 adjudication) is typically relevant for studies where subjective assessments (like imaging reads) require a consensus for ground truth. For a clinical trial involving patient symptoms and objective measurements, this type of adjudication method is generally not applicable. The studies were randomized, sham-controlled, and double-blind, which are methods designed to reduce bias in treatment efficacy evaluation.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance
No, an MRMC comparative effectiveness study was not done. This device is a therapeutic stimulation device, not an AI-powered diagnostic tool requiring human reader assistance.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
This question is not applicable. The gammaCore Sapphire is a non-invasive vagus nerve stimulator, a physical device used by patients under a prescription, not an algorithm. The "performance" is the physiological effect of the stimulation.
7. The Type of Ground Truth Used
The "ground truth" for the efficacy endpoints was based on:
- Patient-reported outcomes: Primarily the number of migraine days per month, headache days per month, and acute medication days per month, typically recorded in diaries.
- Clinical assessment: Diagnosis of migraine based on established International Classification of Headache Disorders (ICHD) criteria by trained clinicians at the study sites.
- Safety monitoring: Assessment of adverse events by investigators.
8. The Sample Size for the Training Set
There isn't a "training set" in the context of an algorithm or AI. The device itself is a physical medical device. The "training" that occurs is the patient's and clinician's understanding of how to use the device, not an algorithmic learning process. The clinical studies described (PREMIUM and EVENT) serve as the validation datasets for safety and efficacy for the expanded indication.
9. How the Ground Truth for the Training Set Was Established
Not applicable, as there is no algorithmic "training set" for this device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 26, 2020
Electrocore, Inc. Mike Romaniw VP, Quality Assurance & Regulatory Affairs 150 Allen Road, Suite 201 Basking Ridge, New Jersey 07920
Re: K191830
Trade/Device Name: gammaCore Sapphire Regulation Number: 21 CFR 882.5892 Regulation Name: External vagal nerve stimulator for headache Regulatory Class: Class II Product Code: PKR. OAK Dated: February 25, 2020 Received: February 27, 2020
Dear Mike Romaniw:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Vivek Pinto, PhD Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K191830
Device Name gammaCore Sapphire
Indications for Use (Describe)
The gammaCore Sapphire Non-invasive Vagus Nerve Stimulator is intended to provide non-invasive vagus nerve stimulation (nVNS) on the side of the neck. The gammaCore Sapphire device is indicated for:
- · Adjunctive use for the preventive treatment of cluster headache in adult patients.
- · The acute treatment of pain associated with episodic cluster headache in adult patients.
- · The acute treatment of pain associated with migraine headache in adult patients.
- · The preventive treatment of migraine headache in adult patients.
| Type of Use (Select one or both, as applicable) |
|---|
| ------------------------------------------------- |
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(K) SUMMARY 5
The following information is provided as required by 21 CFR §807.87 for the electroCore gammaCore Sapphire 510(k) premarket notification. In response to the Safe Medical Devices Act of 1990, the following is a summary of the information upon which the substantial equivalence determination is based.
| Applicant: | electroCore, Inc. |
|---|---|
| 150 Allen Road, Suite 201 | |
| Basking Ridge, New Jersey 07920 | |
| Ph: 973-290-0097 | |
| Fax: 973-290-9171 |
Establishment Registration Number: 3009060963
| Contact: | Mike Romaniw |
|---|---|
| VP, Quality Assurance & Regulatory Affair | |
| Office: 973-355-6702 | |
| Fax: 973-290-9171 | |
| Mike.Romaniw@electroCore.com | |
| Alternate Contact: | Eric J. Liebler |
| SVP, Neurology | |
| Mobile: 908-938-9780 | |
| Eric.Liebler@electroCore.com | |
| Alternate Contact: | Deborah Lavoie Grayeski |
| Sr. Project Manager | |
| M Squared Associates, Inc. | |
| Office: 347-954-0418 | |
| Fax: 703-562-9797 | |
| dgrayeski@MSquaredAssociates.com |
{4}------------------------------------------------
| Alternate Contact: | Janice M. Hogan, Esq. |
|---|---|
| Partner | |
| Hogan Lovells US LLP | |
| Direct: 267-675-4611 | |
| Janice.Hogan@HoganLovells.com | |
| Date of Submission: | March 24, 2020 |
| Proprietary Name: | gammaCore Sapphire |
| Common Name: | External vagal nerve stimulator for headache |
| Classification Status: | Class II |
| Product Codes: | PKR, QAK |
| Predicate Device: | gammaCore Sapphire, K182369 |
Indications for Use: The gammaCore Sapphire Non-invasive Vagus Nerve Stimulator is intended to provide non-invasive vagus nerve stimulation (nVNS) on the side of the neck. The gammaCore Sapphire device is indicated for:
- Adjunctive use for the preventive treatment of cluster headache in adult patients. .
- The acute treatment of pain associated with episodic cluster headache in adult patients. .
- . The acute treatment of pain associated with migraine headache in adult patients.
- The preventive treatment of migraine headache in adult patients.
Device Description: The gammaCore Sapphire (gammaCore) is a multiuse, handheld, rechargeable, portable device consisting of a rechargeable battery and signal-generating and amplifying electronics, with a slide control switch for user/operator control of the signal amplitude (relative range, 0-40 continuous).
The gammaCore Sapphire:
- . Includes a charging station incorporated into the "clam shell" storage case connected to a power adapter for charging of the device as necessary by the end user.
- Provides visible (light and display) and audible (beep) feedback regarding device and . stimulation status.
{5}------------------------------------------------
-
Allows for multiple stimulations or doses; each stimulation or dose lasts 120 seconds, . after which the device automatically turns off unless turned off earlier by the user/operator.
Note: One dose is defined as one stimulation cycle lasting 120 seconds (2 minutes). -
Delivers up to a fixed number of doses within a 24-hour period; once the maximum daily . number of doses has been reached, the device will not deliver any more doses until the following 24-hour period.
-
Indicates on the display the number of remaining doses available in a 24-hour period. .
The device will be provided to the patient/user with an initial 10-, 31-, or 93-day radio-frequency identification (RFID) card on the basis of the healthcare provider's prescription. Additional (refill/reload) cards will be provided in response to a user/patient request based on a prescription from his or her healthcare provider. The refill/reload RFID cards will be programmed using the gammaCore Dispensing and Ordering Terminal (gammaCore DOT) by electroCore or its authorized agent. This is a specialized application for dispensing the device therapy.
When a 10-. 31-. or 93-day refill/reload card is requested by a patient/user (in accordance with a prescription from a healthcare provider) for a unique device serial number, an RFID card is encoded with the appropriate dosage according to the prescription. The encoded RFID card is matched to a specific gammaCore device serial number residing in a database maintained by electroCore. The gammaCore DOT application, running on a Sony Xperia® tablet, uses a proprietary encoding algorithm to encrypt the therapy days and doses per day on the refill/reload RFID card using near field communication (NFC) protocols.
The encoding algorithm is based on a seed-value pair of numbers specific to a device ID (unique device serial number) that is registered in the gammaCore DOT database. The gammaCore DOT application ensures that only legitimate seed values allow refilling/reloading of the device through validation of the prescription and seed values in the gammaCore DOT database using the unique device and patient IDs.
{6}------------------------------------------------
The encoded refill/reload RFID card is then provided to the user/patient who requested the refill/reload of the device, along with 1 to 6 additional tubes of conductive gel (the number of conductive gel tubes provided is based on the 10-, 31-, or 93-day refill/reload being provided). On receipt of the RFID card, the user/patient refills/reloads his or her gammaCore device by placing the RFID card across the face of the device (with the device turned on). The device will display "rd" and the "refill" icon as the device reads the RFID card. The device will signal (beeping twice) when it has been loaded with the programmed doses. The device will now be ready for use as treatment. The RFID card can be used for only one refill/reload; upon completion of the device refill/reload, the card can be thrown away.
In addition, a Bluetooth® feature will be enabled to facilitate diagnostics of any devices returned by patients/users to the manufacturer, to allow determination of the number of days the device was used and/or the number of doses, as well as any days/doses remaining on the device. The Bluetooth feature will not be accessible to the patient/user; it is accessible only to the device manufacturer.
Summary of Technological Characteristics:
There are no changes to the technological characteristics of the gammaCore Sapphire for this expanded indication.
Summary of Non-clinical Testing:
There are no changes to the technological characteristics of the gammaCore Sapphire for this expanded indication; no additional non-clinical or performance testing is required.
Summary of Clinical Data:
Clinical data demonstrating the safety and effectiveness of the gammaCore Sapphire for the prophylactic/preventive treatment of migraine headache were collected from the PREMIUM study and the EVENT study. The PREMIUM study was a prospective, randomized, doubleblind, sham-controlled, multicenter study in patients with episodic migraine conducted at 22 European sites from June 1, 2015, to November 21, 2017. It consisted of a 4-week run-in period of no study treatment, which was followed by a 12-week double-blind phase of randomly
{7}------------------------------------------------
assigned preventive treatment with nVNS or sham and a 24-week open-label phase during which all participants received nVNS therapy. Post hoc analysis of the modified intent-to-treat (mITT) population, which included subjects with ≥67% adherence per month, showed significant differences between groups in favor of nVNS for the study's primary endpoint, mean reduction in the number of migraine days per month (therapeutic gain, 0.74; P=0.043), as well as for headache days per month (therapeutic gain, 0.86; P=0.045) and acute medication days per month (therapeutic gain, 0.80; P=0.039). Response to treatment with nVNS during the double-blind period was maintained with further nVNS therapy during the open-label period. The sham device used in the PREMIUM study was recently proven to have activated the vagus nerve. This likely mitigated the therapeutic gain measured in the PREMIUM study. Table 3 summarizes the key characteristics of the PREMIUM study.
| Test device | gammaCore R-30 (10015-00302) | |
|---|---|---|
| Study design | Randomized, sham-controlled, double-blind trial | |
| n | ITT332 | mITT278 |
| Sites | 22 | |
| Trial duration (persubject) | 10 months (1-month screening, 3-month randomization phase, and 6-month open-labelphase) | |
| Treatment paradigm | Each treatment consisted of 2 consecutive 2-minute stimulations administered on thesame side of the neck 3 times per day: (1) upon waking and (2, 3) 6 to 8 hours after thefirst and second daily treatments | |
| Primary efficacyendpoint | Mean reduction in the number of migraine days from the 4-week run-in period(baseline) to the last 4 weeks of the 12-week double-blind period | |
| Secondary efficacyendpoints | Reductions in number of headache daysSubjects with ≥50% response rate | |
| Primary safetyendpoint | Assessment of onset, type, severity, and frequency of adverse events (anticipated andunanticipated), including determination of device relatedness | |
| Key inclusion criteria | 1. Between the ages of 18 and 75 years2. Previously diagnosed as suffering from migraine with or without aura inaccordance with the International Classification of Headache Disorders-3 Betacriteria3. Experienced between 5 and 12 headache days per month (over the last 4 months)with at least 2 of the migraines lasting more than 4 hours4. No use of any migraine prevention treatments and/or medications (includingBotox® [onabotulinumtoxinA] injections)5. Agreed to refrain from changing the type or dosage of any prophylacticmedications for indications other than chronic migraine that in the opinion of theclinician might interfere with the study objectives (eg, antidepressants,anticonvulsants, beta-adrenergic blockers) |
Table 3. Key Characteristics of the PREMIUM Study
{8}------------------------------------------------
| Key exclusioncriteria | 1. A history of aneurysm, intracranial hemorrhage, brain tumors, or significant headtrauma | ||||||
|---|---|---|---|---|---|---|---|
| 2. Known or suspected severe atherosclerotic cardiovascular disease, severe carotidartery disease (eg, bruits or history of transient ischemic attack or cerebrovascularaccident), congestive heart failure, known severe coronary artery disease, or recentmyocardial infarction | |||||||
| 3. Uncontrolled high blood pressure | |||||||
| 4. Current implantation of an electrical and/or neurostimulator device, including butnot limited to a cardiac pacemaker or defibrillator, vagal neurostimulator, deepbrain stimulator, spinal stimulator, bone growth stimulator, or cochlear implant | |||||||
| 5. Current implantation of metal cervical spine hardware or a metallic implant nearthe gammaCore stimulation site | |||||||
| Subject baselinedemographics | Device | Population | Sex, No. (%) | Age, y,mean(SD) | Headachedays, mean(SD) | Migrainedays,mean (SD) | |
| nVNS | ITT | 142 (86.1) | 23(13.9) | 43.5(11.1) | 8.9 (2.6) | 7.9 (2.2) | |
| mITT | 119 (86) | 19 (14) | 43.9(11) | 8.9 (2.6) | 8.1 (2.1) | ||
| Sham | ITT | 138 (82.6) | 29(17.4) | 41.4(12.3) | 9.1 (2.6) | 8.1 (2.0) | |
| mITT | 115 (82) | 25 (18) | 41.8(12) | 9.0 (2.6) | 8.1 (2.1) | ||
| Primary safetyendpoint results | Preventive nVNS therapy was safe and well tolerated | ||||||
| Across all study periods, the most common adverse device effects were rash, pain,erythema, discomfort at the application site, and dizziness | |||||||
| No serious adverse device effects were reported during the study | |||||||
| Primary efficacyendpoint results | Mean changefrom baseline inno. of migrainedays per month | Therapeuticgain | P value | ||||
| Population | n(nVNS/Sham) | nVNS | Sham | ||||
| ITT | -2.3 | -1.8 | -0.47 | 0.15 | |||
| mITT | -2.3 | -1.5 | -0.74 | 0.04 | |||
| Control formedication use | The use of migraine prevention medication was not allowed | ||||||
| Medications were allowed for the acute relief of migraine symptoms |
Diener HC, Goadsby PJ, Ashina M, et al. Non-invasive vagus nerve stimulation (nVNS) for the preventive treatment of episodic migraine: the multicentre, double-blind, randomised, sham-controlled PREMIUM trial. Cephalalgia. 2019;39(12):1475-1487.
Abbreviations: ITT, intent-to-treat; mITT, modified intent-to-treat; nVNS, non-invasive vagus nerve stimulator; SD, standard deviation.
The EVENT study, a prospective, randomized, sham-controlled, multicenter pilot study for the prevention of chronic migraine, was conducted at 6 sites in the United States from October 2012 to April 2014. It consisted of a 4-week run-in phase of no study treatment, an 8-week doubleblind phase of randomly assigned preventive treatment with nVNS or sham, and a 6-month openlabel phase of nVNS treatment for all participants. The primary efficacy measure, mean change from baseline to month 2 in the number of headache days per month, was – 1.4 in the nVNS group and -0.2 in the sham group. Continued nVNS use led to significant reductions from
{9}------------------------------------------------
baseline in the number of headache days per month during the open-label period. Among subjects who completed the study, the proportion who had a ≥50% reduction from baseline in the number of headache days per month increased with greater duration of treatment. Such findings suggest that the benefits of nVNS therapy may accrue over time. Table 4 summarizes the key characteristics of the EVENT study.
| Test device | gammaCore-150 (10007-00305) | |||
|---|---|---|---|---|
| Study design | Randomized, sham-controlled, double-blind trial | |||
| n | ITT | PP | ||
| 59 | 49 | |||
| Sites | 6 | |||
| Trial duration (per subject) | 9 months (1-month screening, 2-month randomization phase, and 6-month open-label phase) | |||
| Treatment paradigm | Each treatment consisted of two 2-minute self-administered stimulations delivered 5 to 10 minutes apart to the right side of the neck at 3 prespecified times every day: (1) within 1 hour of awakening; (2) 6 to 8 hours after the first treatment; and (3) 6 to 8 hours after the second treatment | |||
| Primary efficacy endpoint | Change in the mean number of headache days at the end of month 2 of the randomized phase compared with baseline | |||
| Secondary efficacy endpoint | Severity of each headache day | |||
| Primary safety endpoint | Assessment of the onset, type, severity, and frequency of adverse events (anticipated and unanticipated), including the determination of device relatedness | |||
| Key inclusion criteria | 1. | Between the ages of 18 and 65 years | ||
| 2. | Previously diagnosed as suffering from migraine with or without aura in accordance with the International Classification of Headache Disorders-2 criteria | |||
| 3. | Experienced at least 15 headache days per month (over the last 3 months) | |||
| 4. | No use of any migraine prevention treatments and/or medications (including Botox® [onabotulinumtoxinA] injections) | |||
| 5. | Agreement to refrain from changing the type or dosage of any prophylactic medications for indications other than chronic migraine that in the opinion of the clinician might interfere with the study (eg, antidepressants, anticonvulsants, beta-adrenergic blockers) | |||
| Key exclusion criteria | 1. | A history of aneurysm, intracranial hemorrhage, brain tumors, or significant head trauma | ||
| 2. | Known or suspected severe atherosclerotic cardiovascular disease, severe carotid artery disease (eg, bruits or history of transient ischemic attack or cerebrovascular accident), congestive heart failure, known severe coronary artery disease, or recent myocardial infarction | |||
| 3. | Uncontrolled high blood pressure | |||
| 4. | Current implantation of an electrical and/or neurostimulator device, including but not limited to cardiac pacemaker or defibrillator, vagal neurostimulator, deep brain stimulator, spinal stimulator, bone growth stimulator, or cochlear implant | |||
| 5. | Current implantation of metal cervical spine hardware or a metallic implant near the gammaCore stimulation site |
| Table 4. Kev Characteristics of the EVENT Study | ||
|---|---|---|
| े |
{10}------------------------------------------------
| Subject baselinedemographics | Device | Population | Sex, No. (%) | Age, y,mean(SD) | Headache days, mean(SD) | ||
|---|---|---|---|---|---|---|---|
| ITT | Female | Male | 40.5(14.2) | 20.8 (5.0) | |||
| nVNS | PP | 22 (85) | 4 (15) | 42.0(14.4) | 22.7 (6.0) | ||
| ITT | 27 (93) | 2 (7) | 38.8(11.1) | 22.3 (4.9) | |||
| Sham | PP | 22 (96) | 1 (4) | 39.3(11.1) | 25.2 (5.0) | ||
| Primary safetyendpoint results | nVNS had a benign tolerability profile that was generally similar to that of the sham treatmentMost adverse events were mild or moderate and transientThe most commonly reported adverse events were upper respiratory tract infections and gastrointestinal symptoms | ||||||
| Primary efficacyendpoint results | Population | n(nVNS/Sham) | Mean change frombaseline in no. ofmigraine days permonthnVNS | Sham | Therapeuticgain | P value | |
| ITT | 30/29 | -1.4 | -0.2 | -1.2 | 0.56 | ||
| PP | 26/23 | -2.0 | -0.1 | -1.9 | 0.44 | ||
| Control formedication use | The use of migraine prevention medication was not allowedMedications were allowed for the acute relief of migraine symptoms |
Silberstein SD, Calhoun AH, Lipton RB, et al. Chronic migraine headache prevention with noninvasive vagus nerve stimulation: the EVENT study. Neurology. 2016;87(5):529-538.
Abbreviations: ITT, intent-to-treat; nVNS, non-invasive vagus nerve stimulator; PP, per-protocol; SD, standard deviation.
Consistent with results from previous trials. nVNS was shown to be safe and well tolerated in both the PREMIUM and EVENT studies. No serious device-related adverse events (AEs) were reported, and no new safety concerns for nVNS use in the intended population were apparent.
Substantial Equivalence Discussion:
gammaCore Sapphire technology is identical to the device technology cleared under K182369. There have been no changes in the technological characteristics or intended use of the gammaCore Sapphire in the proposed Indications for Use statement. The instructions for the device associated with the dosing for the prevention of migraine headache differ from those for preventive or acute treatment of cluster headache and those for the acute treatment of migraine headache; however, for all indications, use of more than 24 stimulations per day has not been formally evaluated and continues to be listed as a precaution in the labeling.
{11}------------------------------------------------
Summary:
Table 5 establishes the substantial equivalence of the subject device to that of the predicate device.
{12}------------------------------------------------
| gammaCore Sapphire(Subject Device) | gammaCore Sapphire(Predicate Device) | Substantial Equivalence | |
|---|---|---|---|
| 510(k) number | TBD | K182369 | |
| Intended use | The gammaCore Sapphire is a device thatprovides non-invasive vagus nerve stimulation(nVNS) when applied to the side of the neck.This is a mild electrical stimulation of thevagus nerve, which runs through the neck andcarries information to the central nervoussystem. Each stimulation with gammaCorelasts 2 minutes. The patient controls thestimulation strength. | The gammaCore Sapphire is a device thatprovides nVNS when applied to the side ofthe neck. This is a mild electrical stimulationof the vagus nerve, which runs through theneck and carries information to the centralnervous system. Each stimulation withgammaCore lasts 2 minutes. The patientcontrols the stimulation strength. | No change in intended use |
| Indications for use | The gammaCore Sapphire is indicated for:• Adjunctive use for the preventive treatmentof cluster headache (CH) in adult patients• The acute treatment of pain associated withepisodic cluster headache (eCH) in adultpatients• The acute treatment of pain associated withmigraine headache in adult patients• The preventive treatment of migraineheadache in adult patients | The gammaCore Sapphire is indicated for:• Adjunctive use for thepreventive treatment of CH in adultpatients• The acute treatment of pain associated witheCH in adult patients• The acute treatment of pain associated withmigraine headache in adult patients | The expansion of the indicationdoes not alter the intendedtherapeutic effect or otherwisecreate a new intended use, asexplained previously. Supported byclinical data in this submission. |
| Rx vs OTC | Prescription use only | Prescription use only | No change |
| gammaCore Sapphire(Subject Device) | gammaCore Sapphire(Predicate Device) | Substantial Equivalence | |
| Treatment recommendation | Preventive treatment of migraine:120-second stimulation cycle,2 consecutive stimulations on either side of the neck as follows:• First daily treatment: within 1 hour of waking• Second daily treatment: 4-6 hours after the first daily treatment• Third daily treatment: within 1 hour before going to sleepPreventive treatment of CH:120-second stimulation cycle,3 consecutive stimulations on either side of the neck as follows:• First daily treatment: within 1 hour of waking• Second daily treatment: 7-10 hours after the first daily treatment | Preventive treatment of CH:120-second stimulation cycle,3 consecutive stimulations on either side of the neck as follows:• First daily treatment: within 1 hour of waking• Second daily treatment: 7-10 hours after the first daily treatmentAcute treatment of eCH:120-second stimulation cycle,3 consecutive stimulations up to 8 times a dayAcute treatment of migraine:120-second stimulation cycle,2 bilateral stimulations up to 3 times a day | Change in treatment protocol to reflect different forms of primary headache |
| Acute treatment of eCH:120-second stimulation cycle,3 consecutive stimulations up to 8 times a day | |||
| Acute treatment of migraine:120-second stimulation cycle,2 bilateral stimulations up to 3 times a day | |||
| Patient-contacting materials | SS, ABS-PC, SignaGel® electrode gel | SS, ABS-PC, SignaGel electrode gel | No change in materials |
| Electrical classification | UL 60601-1 Class III Type BF Applied Part | UL 60601-1 Class III Type BF Applied Part | No change in classification |
| Waveform/frequency | Sinusoidal wave, symmetrical biphasic 5000-Hz pulses at a rate of 25 Hz | Sinusoidal wave, symmetrical biphasic 5000-Hz pulses at a rate of 25 Hz | No change in waveform or frequency |
| gammaCore Sapphire(Subject Device) | gammaCore Sapphire(Predicate Device) | Substantial Equivalence | |
| Maximum output | 30 V (peak), 60 mA (peak) | 30 V (peak), 60 mA(peak) | No change in outputs |
| Load impedance | 450-550 ohms | 450-550 ohms | No change in impedance |
| Power supply | 3-V LiFePo4 battery | 3-V LiFePo4 battery | No change in power supply voltage |
| Service life | 3 years from date of manufacture | 3 years from date of manufacture | No change in service life |
| Controls | Control slideIncrease slide up/decrease slide down | Control slideIncrease slide up/decrease slide down | |
| Output regulation | Device software and control slide | Device software and control slide | |
| Device statusdisplay | LED screen | LED screen | |
| Battery charger | Qi-compatible wireless charger in clam shellstorage case | Qi-compatible wireless charger in clam shellstorage case | No change in circuitry or controls ofthe subject and predicate device. |
| RFID refill/reloadcapability | Allows refilling/reloading of the number ofdays/doses for which the device can providetreatment; allows for continued use of samedevice for extended periods of time | Allows refilling/reloading of the number ofdays/doses for which the device can providetreatment; allows for continued use of samedevice for extended periods of time | |
| Device diagnostics,Bluetooth | Provides for diagnostics by manufacturer ofreturned devices, including number of daysdevice was used, number of doses delivered,and remaining days/doses | Provides for diagnostics by manufacturer ofreturned devices, including number of daysdevice was used, number of doses delivered,and remaining days/doses | |
| Start-up | Yes | Yes | |
| Session complete | Yes | Yes | No change in available alarm |
| Errors/depletedbattery | Yes | Yes | signals |
| No doses left | Yes | Yes | |
| gammaCore Sapphire(Subject Device) | gammaCore Sapphire(Predicate Device) | Substantial Equivalence | |
| Expired/no daysleft | Yes | Yes | |
| Start-up(powered on) | Light on | Light on | |
| Unit ready(powered on) | LED doses remaining for 24-hour period | LED doses remaining for 24-hour period | |
| Dose complete | LED days, doses remaining, and lastamplitude | LED days, doses remaining, and lastamplitude | No change to display/message in thesubject and predicate device. |
| Errors/depletedbattery | E# display | E# display | |
| No dosesremaining | LED doses 00 | LED doses 00 | |
| Expired/no daysleft | LED doses/days remaining | LED doses/days remaining | |
| Low battery | LED display battery charge indicator | LED display battery charge indicator | |
| Reloading error | LED display if refill process fails | LED display if refill process fails | |
| Card error | LED display if refill card fails | LED display if refill card fails |
Table 5. Substantial Equivalence Comparison Table
{13}------------------------------------------------
electroCore, Inc. March 24, 2020
{14}------------------------------------------------
{15}------------------------------------------------
Abbreviations: ABS-PC, acrylonitrile-butade; LED, light-emitting diode; OTC, over the counter; RFID, radio-frequency identification; Rx, prescription; SS, stainless steel; TBD, to be determined.
{16}------------------------------------------------
Conclusion:
There have been no changes in the technological characteristics or intended use of the gammaCore Sapphire. The addition of the preventive treatment of migraine headache to the Indications for Use does not raise new or different questions of safety or effectiveness compared to those raised with the predicate device. Therefore, the presented information demonstrates that the subject device is substantially equivalent to the predicate device.
§ 882.5892 External vagal nerve stimulator for headache.
(a)
Identification. An external vagal nerve stimulator for headache is a prescription device used to apply an electrical current to a patient's vagus nerve through electrodes placed on the skin for the treatment of headache.(b)
Classification. Class II (special controls). The special controls for this device are:(1) The technical parameters of the device, including waveform, output modes, maximum output voltage and current (with 500, 2,000, and 10,000 ohm loads), pulse duration, frequency, net charge (µC) per pulse, maximum phase charge at 500 ohms, maximum current density (mA/cm
2 , r.m.s.), maximum average current (mA), maximum average power density (W/cm2 ), and the type of impedance monitoring system shall be fully characterized through non-clinical performance testing.(2) Software verification, validation, and hazard analysis shall be performed.
(3) Biocompatibility evaluation of the patient-contacting components of the device shall be performed.
(4) The device shall be tested for electrical, thermal, and mechanical safety, and for electromagnetic compatibility (EMC).
(5) The labeling must include:
(i) Instructions for proper use of the device, including placement of the device on the patient; and
(ii) Instructions on care and cleaning of the device.