K Number
K191801

Validate with FDA (Live)

Device Name
PROBEAT-CR
Manufacturer
Date Cleared
2019-09-13

(72 days)

Product Code
Regulation Number
892.5050
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The PROBEAT-CR is a medical device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation.

Device Description

The PROBEAT-CR is a proton beam irradiation system, which provides a therapeutic proton beam for clinical treatment. It is designed to deliver a proton beam with the prescribed dose, dose distribution and directed to the prescribed patient treatment site. The PROBEAT-CR is a modification to the cleared PROBEAT-V system, for installation at a different clinical site.

The PROBEAT-CR has two main subsystems: (1) equipment necessary to generate the proton beam and direct it to the beam delivery system for patient treatment, and (2) a beam delivery system whose primary responsibility is to ensure that the desired prescription parameters are properly delivered. The PROBEAT-CR comprises the following components and subsystems:

  • Beam production system
    • Accelerator system (LINAC, Synchrotron).
    • Beam transport system (Low/High Energy Beam Transport systems).
  • Beam delivery system in 4 separate treatment rooms. Each of 3 rooms will have a rotating gantry and 1 room will have a fixed beam.
    • Gantry Room
      • Scanning Nozzle
      • Rotating Gantry
      • Patient Positioning System
      • Orthogonal X-ray system
      • Cone Beam CT
    • Fixed Beam Room
      • Patient Positioning System .
      • Orthogonal X-ray system .
  • Treatment Control and Safety System
AI/ML Overview

The provided text is a 510(k) summary for the PROBEAT-CR Proton Beam Therapy System. It describes the device, its intended use, and a comparison to predicate devices, but it does not contain information regarding software performance, acceptance criteria for an AI/algorithm, or a study that specifically proves the device meets such criteria.

The document primarily focuses on demonstrating the substantial equivalence of the PROBEAT-CR to its predicate devices (PROBEAT-V systems K151132 and K152592) in terms of hardware components, technological characteristics, and overall function as a proton beam therapy system.

The "Performance Data" section discusses:

  • Mechanical performance of the rotating gantry and patient couch.
  • Beam performance testing (dose shape and dose).
  • Safety interface testing (beam stop control, dose monitor, area safety, mechanical interlocks).
  • Electrical safety and electromagnetic compatibility (IEC 60601-1 and IEC 60601-1-2).

It states: "In all instances, the PROBEAT-CR functioned as intended and met its specifications. Testing demonstrated substantial equivalence to the predicates." However, it does not specify what those "specifications" or "acceptance criteria" were, nor does it detail the study design, sample sizes, or ground truth for any algorithmic performance evaluation.

Therefore, I cannot provide the requested information about acceptance criteria and a study proving an AI/algorithm meets those criteria based solely on the provided text. The document describes a physical medical device (proton beam therapy system), not an AI or algorithm with associated performance metrics.

{0}------------------------------------------------

September 13, 2019

Image /page/0/Picture/1 description: The image shows the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

Hitachi Ltd. % Mr. Jonathan Kahan Partner Hogan Lovells US LLP 555 Thirteenth Street, NW WASHINGTON DC 20004

Re: K191801

Trade/Device Name: PROBEAT-CR Proton Beam Therapy System Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: Class II Product Code: LHN Dated: July 3, 2019 Received: July 3, 2019

Dear Mr. Kahan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K191801

Device Name PROBEAT-CR Proton Beam Therapy System

Indications for Use (Describe)

The PROBEAT-CR is a medical device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation.

Type of Use (Select one or both, as applicable)| | Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) SUMMARY

PROBEAT-CR Proton Beam Therapy Device

Submitter Name, Address, Telephone Number, Contact Person and Date Prepared

Hitachi Ltd., Healthcare Hitachi Works 3-1-1 Saiwai-cho, Hitachi-shi lbaraki-ken, 317-8511, Japan Telephone: +81 (294) 55-5900 Facsimile: +81 (294) 55-9946

Contact Person: Tomoyuki Seino

Date Prepared: July 3, 2019

Name of Device:

PROBEAT-CR Proton Beam Therapy System

Common or Usual Name:

Proton Beam Therapy Device

Classification Name:

System, Radiation Therapy, Charged-Particle, Medical

Regulatory Class:

Class II

Product Code:

LHN

Predicate Devices:

Hitachi Ltd. PROBEAT-V (K151132) Hitachi Ltd. PROBEAT-V (K152592)

Device Description

The PROBEAT-CR is a proton beam irradiation system, which provides a therapeutic proton beam for clinical treatment. It is designed to deliver a proton beam with the prescribed dose, dose distribution and directed to the prescribed patient treatment site. The PROBEAT-CR is a modification to the cleared PROBEAT-V system, for installation at a different clinical site.

{4}------------------------------------------------

Intended Use / Indications for Use

The PROBEAT-CR is a medical device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation.

Summary of Technological Characteristics

The PROBEAT-CR has two main subsystems: (1) equipment necessary to generate the proton beam and direct it to the beam delivery system for patient treatment, and (2) a beam delivery system whose primary responsibility is to ensure that the desired prescription parameters are properly delivered. The PROBEAT-CR comprises the following components and subsystems:

  • . Beam production system
    • o Accelerator system (LINAC, Synchrotron).
    • Beam transport system (Low/High Energy Beam Transport systems). O
  • Beam delivery system in 4 separate treatment rooms. Each of 3 rooms will have a rotating gantry and 1 room will have a fixed beam.
    • o Gantry Room
      • Scanning Nozzle ■
      • 트 Rotating Gantry
      • 트 Patient Positioning System
      • 트 Orthogonal X-ray system
      • Cone Beam CT
    • o Fixed Beam Room
      • Patient Positioning System .
      • Orthogonal X-ray system .
  • . Treatment Control and Safety System

The PROBEAT-CR is a modification to the cleared PROBEAT-V systems for use at a difference clinical facility. The system comprises the same components and subsystems as outlined above, with minor differences in design as summarized in the following table.

Comparison ItemHitachi PROBEAT-CRHitachi PROBEAT-VHitachi PROBEAT-V
510(k) Number- Pending -K151132K152592
Intended Use/Indications for UseThe PROBEAT-CR is amedical devicedesigned to produceand deliver a protonbeam for the treatmentof patients with localizedtumors and otherconditions susceptibleto treatment byradiation.The PROBEAT-V is amedical devicedesigned to produceand deliver a protonbeam for the treatmentof patients with localizedtumors and otherconditions susceptibleto treatment byradiation.The PROBEAT-V is amedical devicedesigned to produceand deliver a protonbeam for the treatmentof patients with localizedtumors and otherconditions susceptibleto treatment byradiation.
AcceleratorSynchrotronSynchrotronSynchrotron

{5}------------------------------------------------

Comparison ItemHitachi PROBEAT-CRHitachi PROBEAT-VHitachi PROBEAT-V
ParticleProtonsProtonsProtons
Variable energy70-230 MeV70-230 MeV70-230 MeV
NozzlesDiscrete Spot ScanningDiscrete Spot ScanningDiscrete Spot Scanning
Support for patient positioningCone Beam CT Conventional x-ray system A computer assisted patient position systemConventional x-ray system A computer assisted patient position systemCone Beam CT Conventional x-ray system A computer assisted patient position system
Treatment rooms3 rotating gantry rooms and 1 fixed beam room (Maximum 5 total)4 rotating gantry rooms and 1 fixed beam room (Maximum 5 total)2 rotating gantry rooms and 1 fixed beam room (Maximum 5 total)
Gantry rotating angle360 degrees (-180 - 180 degrees)190 degrees (-5 - 185 degrees)190 degrees (-5 - 185 degrees)

Performance Data

The company performed testing, as follows:

  • The mechanical performance of the rotating gantry and patient couch o
  • Beam performance testing to evaluate beam dose shape and beam dose ●
  • Safety interface testing to evaluate beam stop control, dose monitor, area safety, and 0 mechanical interlocks.

Further, electrical safety and electromagnetic compatibility testing was also performed in accordance with IEC 60601-1 and IEC 60601-1-2.

In all instances, the PROBEAT-CR functioned as intended and met its specifications. Testing demonstrated substantial equivalence to the predicates.

Conclusions

The PROBEAT-CR has the same intended use and indications for use, as well as substantially similar principles of operation and technological characteristics, as compared to Hitachi's cleared PROBEAT-V (K151132, K152592). Each of the systems comprises the same components of the beam production and beam delivery subsystems. The minor differences between the PROBEAT-CR that is the subject of this submission and the cleared PROBEAT-V do not raise different questions of safety or effectiveness. Performance data demonstrate that the PROBEAT-CR is as safe and effective as the cleared PROBEAT-V systems Thus, the PROBEAT-CR is substantially equivalent to its predicates.

§ 892.5050 Medical charged-particle radiation therapy system.

(a)
Identification. A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.(b)
Classification. Class II. When intended for use as a quality control system, the film dosimetry system (film scanning system) included as an accessory to the device described in paragraph (a) of this section, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.