K Number
K191789

Validate with FDA (Live)

Date Cleared
2019-07-25

(22 days)

Product Code
Regulation Number
876.5010
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Stonetome Stone Removal Device is indicated for use in diagnostic or therapeutic endoscopic retrograde cholangiopancreatography (ERCP), transendoscopic sphincterotomy of the Papilla of Vater and/or the Sphincter of Oddi, to remove stones from the biliary system, and to facilitate injection of contrast medium while occluding the duct with the balloon.

Device Description

The Stonetome Stone Removal Device is a 200 cm tapered 7F (2.3 mm) to 5.5F (1.8 mm) triple lumen catheter that incorporates a sphincterotome with a latex retrieval balloon. The Stonetome Stone Removal Device is available in several configurations: (1) with the latex balloon mounted distal or proximal to the cutting wire, (2) with a 20- or 30-mm wire, and (3) with a 5 mm or 20 mm distal tip.

AI/ML Overview

This FDA 510(k) clearance summary for the Stonetome Stone Removal Device (K191789) does not describe a study involving an AI/Machine Learning device. Instead, it focuses on the substantial equivalence of a modified medical device (a catheter for stone removal) to a predicate device.

Therefore, many of the requested criteria related to AI/ML model evaluation (such as sample size for test/training sets, data provenance, expert ground truth, MRMC studies, standalone performance, etc.) are not applicable to the information provided in this document.

The document discusses performance data in the context of bench testing for a physical medical device, not a software algorithm.

Here's an attempt to answer the questions based only on the provided document, highlighting what is present and what is absent for this type of medical device clearance:


Acceptance Criteria and Study for the Stonetome Stone Removal Device (K191789)

This document pertains to the clearance of a physical medical device, not an AI/Machine Learning algorithm. Therefore, the "acceptance criteria" and "study" described here relate to the device's physical and functional performance, as well as its safety and biocompatibility, as compared to a predicate device.

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategorySpecific Test/EvaluationReported Device Performance (Met/Not Met)
Performance Testing (Bench)Balloon Inflation (OD)Successfully completed (Met)
Balloon Inflation (Leakage)Successfully completed (Met)
Simulated Stone RetrievalSuccessfully completed (Met)
Balloon DistensibilitySuccessfully completed (Met)
BiocompatibilityCytotoxicityAll acceptance criteria were met (Met)
IrritationAll acceptance criteria were met (Met)
SensitizationAll acceptance criteria were met (Met)
Acute Systemic ToxicityAll acceptance criteria were met (Met)
Substantial EquivalenceDesign VerificationMet all acceptance criteria

Note: The document states "Non-clinical testing was successfully performed" and "Performance testing (bench) was successfully completed to establish substantial equivalence" for the listed tests. For Biocompatibility, it explicitly states "All acceptance criteria were met."

2. Sample Size Used for the Test Set and Data Provenance:
This information is not provided in the document. For bench testing of a physical device, "sample size" would refer to the number of device units tested. The document does not specify this, nor does it specify the provenance of data as it's not a clinical study on human subjects or a dataset for AI.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:
Not applicable. This is not an AI/ML device requiring expert-established ground truth for image interpretation or diagnosis. The "ground truth" for this physical device's performance is its ability to meet engineering specifications and safety standards during bench testing.

4. Adjudication Method for the Test Set:
Not applicable. This is not an AI/ML device requiring human adjudication of results. Testing involved objective engineering measurements and biocompatibility evaluations.

5. Was a Multi-Reader Multi-Case (MRMC) comparative effectiveness study done? If so, what was the effect size of how much human readers improve with AI vs without AI assistance?
No, an MRMC study was not done. This device is a physical catheter, not an AI-assisted diagnostic or therapeutic tool for which human reader performance with and without AI assistance would be relevant.

6. Was a standalone (i.e., algorithm only without human-in-the-loop performance) study done?
Not applicable. This is a physical device, not a standalone algorithm. The performance described is of the device itself through bench testing.

7. The Type of Ground Truth Used:
The 'ground truth' for this device's acceptance is based on engineering specifications and pre-defined performance standards met during non-clinical (bench) testing, and compliance with biocompatibility standards (ISO 10993-1). It is not based on expert consensus, pathology, or outcomes data in the typical sense for AI/ML or clinical trials.

8. The Sample Size for the Training Set:
Not applicable. This is not an AI/ML device that requires a training set.

9. How the Ground Truth for the Training Set Was Established:
Not applicable. There is no training set for this type of device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

July 25, 2019

Boston Scientific Corporation Laura Kuroski Senior Regulatory Affairs Specialist 100 Boston Scientific Way Marlborough, Massachusetts 01752

Re: K191789

Trade/Device Name: Stonetome Stone Removal Device Regulation Number: 21 CFR 876.5010 Regulation Name: Biliary Catheter and Accessories Regulatory Class: Class II Product Code: LQR Dated: July 2, 2019 Received: July 3, 2019

Dear Laura Kuroski:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. However, you are responsible to determine that the medical devices you use as components in the kit have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were legally on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit/tray. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Daniel Walter Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICESForm Approved: OMB No. 0910-0120
Food and Drug AdministrationExpiration Date: 06/30/2020
Indications for Use
510(k) Number (if known)
Device Name
Stonetome Stone Removal Device
Indications for Use (Describe)
The Stonetome Stone Removal Device is indicated for use in diagnostic or therapeutic endoscopic retrograde cholangiopancreatography (ERCP), transendoscopic sphincterotomy of the Papilla of Vater and/or the Sphincter of Oddi, to remove stones from the biliary system, and to facilitate injection of contrast medium while occluding the duct with the balloon.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services

Food and Drug Administration

Office of Chief Information Officer

Paperwork Reduction Act (PRA) Staff

PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."

FORM FDA 3881 (7/17) WITH THE STONETOME STONE REMOVAL DEVICE. THE CHECKBOX FOR PRESCRIPTION USE IS CHECKED. THE CHECKBOX FOR OVER-THE-COUNTER USE IS NOT CHECKED.

Special 510(k) Premarket Notification Stonetome™ Stone Removal Device
Proprietary and Confidential Information of Boston Scientific Corporation

{3}------------------------------------------------

Boston Scientific Corporation 300 Boston Scientific Way Marlborough, MA 01752-1234 (508) 683-4000

www.bostonscientific.com

1. Submitter

Boston Scientific Corporation 100 Boston Scientific Way Marlborough, MA 01752

Laura Kuroski Contact: Sr. Regulatory Affairs Specialist Telephone: (508) 683-4406 Laura.Kuroski@bsci.com E-mail:

Date Prepared: July 2, 2019

2. Proposed Device

Trade Name:Stonetome Stone Removal Device
Common Name:Biliary Stone Dislodger
Product Code:LQR
Device Class and Panel:Class II, Gastroenterology/Urology
Classification Regulation:21 CFR 876.5010

3. Predicate Device

Trade Name:Endoscopic Biliary Catheter
Manufacturer:Boston Scientific Corporation
Clearance Number:K946358
Common Name:Biliary Stone Dislodger
Product Code:LQR
Device Class and Panel:Class II, Gastroenterology/Urology
Classification Regulation:21 CFR 876.5010

4. Device Description

The Stonetome Stone Removal Device is a 200 cm tapered 7F (2.3 mm) to 5.5F (1.8 mm) triple lumen catheter that incorporates a sphincterotome with a latex retrieval balloon. The Stonetome Stone Removal Device is available in several configurations: (1) with the latex balloon mounted distal or proximal to the cutting wire, (2) with a 20- or 30-mm wire, and (3) with a 5 mm or 20 mm distal tip. See Stonetome Stone Removal Device configuration summary table below:

UPN*Mounted Balloon LocationDistal Tip LengthCut Wire Length
M00535110Above Cut Wire (Distal)20mm Long Nose20mm Cut Wire
M00535130Above Cut Wire (Distal)20mm Long Nose30mm Cut Wire
M00535150Below Cut Wire (Proximal)5mm Short Nose20mm Cut Wire
M00535170Below Cut Wire (Proximal)20mm Long Nose20mm Cut Wire
M00535190Below Cut Wire (Proximal)5mm Short Nose30mm Cut Wire
M00535210Below Cut Wire (Proximal)20mm Long Nose30mm Cut Wire

Stonetome Stone Removal Device Configurations Table

*Note: Bolded numbers within UPN are Catalog Numbers referenced within predicate 510(k), K946358

{4}------------------------------------------------

The Stonetome Stone Removal Device is capable of accepting a 0.035 in (0.89 mm) guidewire in one lumen while inflating the balloon via a second lumen. When connected to a monopolar electrosurgical generator, passed through a duodenoscope and activated, it will deliver a monopolar current to incise the Papilla of Vater and/or the Sphincter of Oddi. After sphincterotomy, the balloon may be passed beyond the stone, inflated, and withdrawn to remove the stone.

5. Indications for Use

The Stonetome Stone Removal Device is indicated for use in diagnostic or therapeutic endoscopic retrograde cholangiopancreatography (ERCP), transendoscopic sphincterotomy of the Papilla of Vater and/or the Sphincter of Oddi, to remove stones from the biliary system, and to facilitate injection of contrast medium while occluding the duct with the balloon.

6. Technological Characteristics

The modified Stonetome Stone Removal Device shares the same intended use, indications for use, and fundamental scientific technology as the predicate Endoscopic Biliary Catheter (K946358), (the current trade name had not been finalized at the time of the original submission). The modified device and the predicate device share nearly identical technological characteristics, with the modified device incorporating new retrieval balloon material.

7. Performance Data

Non-clinical testing was successfully performed on the modified Stonetome Stone Removal Device.

Performance testing (bench) was successfully completed to establish substantial equivalence between the modified Stonetome Stone Removal Device and the predicate device. This testing included the following:

  • Balloon Inflation (OD) ●
  • Balloon Inflation (Leakage)
  • Simulated Stone Retrieval ●
  • . Balloon Distensibility

Biocompatibility of the Stonetome Stone Removal Device was evaluated in accordance with a risk management process. ISO 10993-1:2009 Biological evaluation of medical devices Part 1: Evaluation and the FDA Guidance for Industry and Food and Drug Administration Staff titled Use of International Standard ISO 10993-1. "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" dated June 16, 2016. The following tests were performed: Cytotoxicity, Irritation, Sensitization, and Acute Systemic Toxicity. All acceptance criteria were met.

The results of non-clinical testing demonstrate that the modified Stonetome Stone Removal Device is considered safe and effective for its intended use.

8. Conclusion

The modified Stonetome Stone Removal Device met all acceptance criteria for design verification as specified by applicable standards, guidance, and test protocols. Boston Scientific has demonstrated that the modified Stonetome Stone Removal Device is substantially equivalent to the predicate Endoscopic Biliary Catheter (K946358).

§ 876.5010 Biliary catheter and accessories.

(a)
Identification. A biliary catheter and accessories is a tubular flexible device used for temporary or prolonged drainage of the biliary tract, for splinting of the bile duct during healing, or for preventing stricture of the bile duct. This generic type of device may include a bile collecting bag that is attached to the biliary catheter by a connector and fastened to the patient with a strap.(b)
Classification. Class II (special controls). The device, when it is a bile collecting bag or a surgical biliary catheter that does not include a balloon component, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.