K Number
K191742

Validate with FDA (Live)

Device Name
ARIES MRSA Assay
Date Cleared
2019-09-25

(86 days)

Product Code
Regulation Number
866.1640
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ARIES® MRSA Assay is an integrated, real-time, polymerase chain reaction (PCR) based qualitative in vitro diagnostic test for the direct detection of methicillinresistant Staphylococcus aureus (MRSA) DNA from nasal swabs in patients at risk for nasal colonization.

The ARIES® MRSA Assay is intended to aid in the prevention and control of MRSA infections in healthcare settings.

The assay is not intended to guide, diagnose, or monitor treatment for MRSA infections. It is not intended to provide results of susceptibility to oxacillin/methicillin. A negative result does not preclude MRSA nasal colonization. Concomitant cultures are necessary to recover organisms for epidemiological typing or for further susceptibility testing.

The ARIES® MRSA Assay is indicated for use with ARIES® Systems.

Device Description

The ARIES® MRSA Assay is a polymerase chain reaction (PCR)-based qualitative in vitro diagnostic test system which consists of the ARIES® System or the ARIES® M1 System with their included ARIES® Software, a sample processing tube, an assay-specific cassette, and an assay-specific protocol file. The ARIES® MRSA Assay cassette is a disposable, single-use cassette containing nucleic acid purification reagents, internal sample process control (SPC), and an assay-specific master mix capable of performing the designated assay on one sample. The ARIES® MRSA Assay cassette directly detects methicillin-resistant Staphylococcus aureus (MRSA) from nasal swabs in patients at risk of nasal colonization. Specifically, the ARIES® MRSA Assay cassette detects the methicillin resistance genes (mecA and mecC), Staphylococcus aureus orfX gene, the SCCmec junction region, and a DNA Sample Processing Control.

Nasal swab specimens are collected using the Liquid Amies Elution Swab (ESwab™) Collection and Transport System, or equivalent. A portion of the sample is transferred to the provided 2 mL ARIES MRSA Sample Processing Tube and vortexed. The processed sample is then transferred to the ARIES® MRSA Assay cassette.

The specimen is lysed and nucleic acid is extracted using an ARIES® instrument. An extractable sample processing control (SPC) target present in the ARIES® MRSA Assay cassette is processed with the specimen. The SPC controls for recovery of extracted nucleic acid, for inhibitory substances and for PCR reagent and instrument integrity. The Ct value of the SPC is designed to verify nucleic acid extraction, to identify PCR inhibition, if any, and verify proper function of the extraction system and real-time instrument.

The extracted nucleic acid and SPC are transferred via magnetic beads through the cassette to the ARIES® MRSA Assay lyophilized PCR reagents in the PCR tube that contain primer pairs and probes specific to mecA/mecC, orfX, SCCmec and the SPC sequence. Each probe is labeled with a distinct fluorophore and detected in a distinct channel of an ARIES® System. PCR amplification is performed and assay fluorescence is monitored. Hybridization of a fluorescently labeled probe to the amplified target results in the release of quenching and generation of fluorescence signal that is indicative of PCR generated amplicon. Following amplification, the reaction is heated to separate the fluorescentlabeled probe from the amplified target, a process that results in a decrease in the fluorescence signal. The reaction fluorescence is measured during this process and the temperature at which the change in fluorescence is the maximum is the Tm of the amplicon. The instrument fluorescence output is analyzed and test results are determined using the ARIES® System software and the ARIES® MRSA Assay protocol and run files. ARIES® MRSA Assay results may be reported from the ARIES® Software or from the optional SYNCT® Software.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the ARIES® MRSA Assay, extracted from the provided FDA 510(k) summary:

Acceptance Criteria and Reported Device Performance

ParameterAcceptance Criteria (Implied)Reported Device Performance
Analytical Studies
Within-Lab Precision100% expected MRSA results for positive and negative samples100% expected MRSA results for all samples (Positive and Negative)
Site-to-Site Precision100% expected MRSA results for positive and negative samples100% expected MRSA results for all samples (Positive and Negative)
Inclusivity100% MRSA positivity for representative MRSA strains (at 3X LoD)100% MRSA positivity (3/3 replicates) for all 55 tested MRSA strains
Challenge StudyExpected MRSA positive for MRSA strains; 0% MRSA positive for non-MRSA strainsMRSA positive for all MRSA strains (some retested at 5X LoD); 0% MRSA positive for BORSA, MRSE, and empty cassette SA variants
Cross-Reactivity100% expected MRSA positive for MRSA strains; 100% expected MRSA negative for negative samples100% expected MRSA positive for MRSA strains; 100% expected MRSA negative for negative samples (99 organisms tested)
Interfering SubstancesNo interference observed at tested concentrationsNo interference detected for any of the 24 tested substances
Swab Equivalency100% expected MRSA positivity for MRSA strains; 0% positivity for negative samples100% expected MRSA positivity for MRSA strains; 0% positivity for negative samples (two swab types)
Clinical Performance
Clinical SensitivityNot explicitly stated, but high percentage expected for clinical utility93.3% (95% CI: 87% - 97%) compared to direct and enriched bacterial culture
Clinical SpecificityNot explicitly stated, but high percentage expected for clinical utility93.5% (95% CI: 92% - 95%) compared to direct and enriched bacterial culture
Positive Percent AgreementNot explicitly stated, but high percentage expected for clinical utility93.5% (95% CI: 87% - 97%) compared to direct bacterial culture
Negative Percent AgreementNot explicitly stated, but high percentage expected for clinical utility92.9% (95% CI: 92% - 94%) compared to direct bacterial culture

Study Information

The provided document describes studies primarily for analytical and clinical performance.

  1. Sample sizes used for the test set and the data provenance:

    • Analytical Studies (Test Set):

      • Precision/Reproducibility: A blinded panel of 5 samples (2 MRSA strains at 1x LoD and 5x LoD, and 1 negative sample). Each tested in triplicate by 2 operators over 5 days (within-lab) or at 3 sites by 2 operators over 5 days (site-to-site). Total replicates: 30 per condition for within-lab, 90 per condition for site-to-site.
      • Limit of Detection (LoD): 20 replicates for each MRSA strain at suspected LoD concentration.
      • Inclusivity: 55 MRSA strains, each tested in triplicate. Total: 165 replicates.
      • Challenge Study: 16 MRSA strains with high MICs (some retested), 17 MRSA strains with low MICs (one retested), 4 BORSA strains, 16 empty cassette SA variants, 1 MRSE strain. Each tested in triplicate.
      • Cross-Reactivity: 99 organisms, each tested in triplicate with MRSA positive and negative samples.
      • Interfering Substances: 24 substances, each tested in triplicate with MRSA positive and negative samples.
      • Nasal Swab Equivalency: 1 MRSA strain at 3 concentrations, 1 negative sample. Each tested with two swab types, likely in replicates (e.g., 6/6 for MRSA positivity suggests 6 replicates for each condition).
      • Data Provenance: Not explicitly stated for analytical studies, but often performed in-house by the manufacturer. The strains used are from established collections (BEI, ATCC, ARLG, CDC AR Bank, Lyon University), suggesting controlled laboratory conditions.
    • Clinical Performance Study (Test Set):

      • Total specimens collected: 2254 nasal swab specimens.
      • Specimens meeting eligibility criteria and included in analysis: 1762 unique specimens.
      • Data Provenance: Prospective collection from August 2018 to February 2019 at four (4) geographically distinct clinical sites within the United States.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Analytical Studies: Ground truth was established by known organism concentrations and characteristics (e.g., specific MRSA strains, non-MRSA organisms, substances). No human experts are typically involved in establishing ground truth for these types of analytical tests.
    • Clinical Study: The "reference method" for ground truth was direct and enriched bacterial culture. This process is laboratory-based and while it involves trained microbiologists, it does not typically involve "experts" in the sense of clinical reviewers or adjudicators for each case. The results of the culture are the ground truth.
  3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Analytical Studies: No adjudication method described as ground truth is based on known laboratory preparations and characteristics.
    • Clinical Study: No explicit adjudication method for the reference method (bacterial culture) is described. The culture results are presented as the definitive ground truth.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No MRMC comparative effectiveness study was mentioned. The ARIES® MRSA Assay is a diagnostic molecular assay that directly detects target DNA; it is not an AI-based imaging or interpretive device that assists human readers.
  5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    • Yes, the ARIES® MRSA Assay is a standalone diagnostic device. The performance data presented (sensitivity, specificity, agreement) directly reflects the algorithm's performance (the assay and its associated software) compared to the reference method (bacterial culture). There is no "human-in-the-loop" component for interpretation of the assay results, as the software determines whether the result is Positive, Negative, or Invalid based on predefined internal cut-offs.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • Analytical Studies: Ground truth was based on known bacterial strains, concentrations, and characteristics (e.g., presence or absence of specific genes like mecA/mecC, orfX, SCCmec; known categories of non-MRSA organisms or interfering substances).
    • Clinical Study: Ground truth was established by direct and enriched bacterial culture for MRSA.
  7. The sample size for the training set:

    • The document implies that an "Initial Assay Protocol File parameters were set during internal optimization and benchmarking studies" and "The final Assay Protocol File parameters were then established during internal verification studies using data from optimization, benchmarking and verification." However, specific sample sizes for these internal training or development phases are not explicitly provided in the 510(k) summary. The listed analytical studies are effectively "test sets" for various performance metrics, not necessarily for training.
  8. How the ground truth for the training set was established:

    • As detailed above, specific training sets and their ground truth establishment methods are not explicitly described in this summary. For molecular diagnostic assays like this, the "training" typically involves empirical testing with known positive and negative controls, spiked samples, and characterized clinical samples during assay development to optimize PCR parameters, probe design, and define interpretation algorithms (Ct cut-offs, Tm windows). The "ground truth" during this development phase would be based on well-characterized strains and samples, similar to the analytical studies described.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

September 25, 2019

Luminex Corporation Kate Linak Manager, Regulatory Affairs 12212 Technology Blvd Austin, Texas 78727

Re: K191742

Trade/Device Name: ARIES MRSA Assay Regulation Number: 21 CFR 866.1640 Regulation Name: Antimicrobial Susceptibility Test Powder Regulatory Class: Class II Product Code: NQX Dated: June 28, 2019 Received: July 1, 2019

Dear Kate Linak:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Steven Gitterman, M.D., Ph.D. Deputy Director Division of Microbiology Devices OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

510(k) Summary

This Summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.

Date of Preparation: 09/13/2019

A. 510(k) Number:

K191742

B. Purpose for Submission:

Traditional 510(k), New Device

C. Measurand:

Target DNA sequences for

  • (1) mecA and mecC genes for methicillin resistance
  • (2) orfX complex gene of Staphylococcus aureus
  • (3) SCCmec junction region

D. Type of Test:

Qualitative Real Time Polymerase Chain Reaction (PCR)

E. Applicant:

Kate Linak Luminex Corporation 12212 Technology Blvd Austin, TX 78727

Tel: 512-381-4397

F. Proprietary and Established Names:

ARIES® MRSA Assay

G. Regulatory Information:

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "Luminex" in a bold, sans-serif font. There is a gray dot above the "i" in "Luminex". The word is slightly slanted to the right. There is a small circle with an "R" inside of it to the right of the "x".

ProductCodeClassificationRegulation SectionPanel
NQXII21 CFR 866.1640—Antimicrobial susceptibility testpowderMicrobiology

H. Intended Use:

1. Intended use(s):

The ARIES® MRSA Assay is an integrated, real-time, polymerase chain reaction (PCR) based qualitative in vitro diagnostic test for the direct detection of methicillinresistant Staphylococcus aureus (MRSA) DNA from nasal swabs in patients at risk for nasal colonization.

The ARIES® MRSA Assay is intended to aid in the prevention and control of MRSA infections in healthcare settings.

The assay is not intended to guide, diagnose, or monitor treatment for MRSA infections. It is not intended to provide results of susceptibility to oxacillin/methicillin. A negative result does not preclude MRSA nasal colonization. Concomitant cultures are necessary to recover organisms for epidemiological typing or for further susceptibility testing.

The ARIES® MRSA Assay is indicated for use with ARIES® Systems.

    1. Indication(s) for use:
      Same as intended use.
    1. Special conditions for use statement(s):
      For prescription use only.
    1. Special instrument requirements:
      For use with ARIES® Systems.

Device Description: l.

The ARIES® MRSA Assay is a polymerase chain reaction (PCR)-based qualitative in vitro diagnostic test system which consists of the ARIES® System or the ARIES® M1 System with their included ARIES® Software, a sample processing tube, an assay-specific cassette, and an assay-specific protocol file. The ARIES® MRSA Assay cassette is a disposable, single-use cassette containing nucleic acid purification reagents, internal sample process control

{4}------------------------------------------------

(SPC), and an assay-specific master mix capable of performing the designated assay on one sample. The ARIES® MRSA Assay cassette directly detects methicillin-resistant Staphylococcus aureus (MRSA) from nasal swabs in patients at risk of nasal colonization. Specifically, the ARIES® MRSA Assay cassette detects the methicillin resistance genes (mecA and mecC), Staphylococcus aureus orfX gene, the SCCmec junction region, and a DNA Sample Processing Control.

Nasal swab specimens are collected using the Liquid Amies Elution Swab (ESwab™) Collection and Transport System, or equivalent. A portion of the sample is transferred to the provided 2 mL ARIES MRSA Sample Processing Tube and vortexed. The processed sample is then transferred to the ARIES® MRSA Assay cassette.

The specimen is lysed and nucleic acid is extracted using an ARIES® instrument. An extractable sample processing control (SPC) target present in the ARIES® MRSA Assay cassette is processed with the specimen. The SPC controls for recovery of extracted nucleic acid, for inhibitory substances and for PCR reagent and instrument integrity. The Ct value of the SPC is designed to verify nucleic acid extraction, to identify PCR inhibition, if any, and verify proper function of the extraction system and real-time instrument.

The extracted nucleic acid and SPC are transferred via magnetic beads through the cassette to the ARIES® MRSA Assay lyophilized PCR reagents in the PCR tube that contain primer pairs and probes specific to mecA/mecC, orfX, SCCmec and the SPC sequence. Each probe is labeled with a distinct fluorophore and detected in a distinct channel of an ARIES® System. PCR amplification is performed and assay fluorescence is monitored. Hybridization of a fluorescently labeled probe to the amplified target results in the release of quenching and generation of fluorescence signal that is indicative of PCR generated amplicon. Following amplification, the reaction is heated to separate the fluorescentlabeled probe from the amplified target, a process that results in a decrease in the fluorescence signal. The reaction fluorescence is measured during this process and the temperature at which the change in fluorescence is the maximum is the Tm of the amplicon. The instrument fluorescence output is analyzed and test results are determined using the ARIES® System software and the ARIES® MRSA Assay protocol and run files. ARIES® MRSA Assay results may be reported from the ARIES® Software or from the optional SYNCT® Software.

Substantial Equivalence Information: J.

    1. Predicate device name(s):
      BD MAX™ MRSA XT

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the word "Luminex" in a bold, sans-serif font. The word is black, and there is a small, gray circle above the "i" in "Luminex." There is a small registration mark to the right of the "x" in "Luminex."

2. Predicate 510(k) number(s):

K133605

3. Comparison with predicate:

The following tables compare Luminex's ARIES® MRSA Assay to GeneOhm Sciences Canada, Inc. (BD Diagnostics) BD MAX™ MRSA XT assay (K133605).

ItemNew DevicePredicate: K133605
Intended UseThe ARIES® MRSA Assay is anintegrated, real-time,polymerase chain reaction (PCR)based qualitative in vitrodiagnostic test for the directdetection of methicillin-resistant Staphylococcus aureus(MRSA) DNA from nasal swabsin patients at risk for nasalcolonization.The ARIES® MRSA Assay isintended to aid in theprevention and control of MRSAinfections in healthcare settings.The assay is not intended toguide, diagnose, or monitortreatment for MRSA infections.It is not intended to provideresults of susceptibility tooxacillin/methicillin. A negativeresult does not preclude MRSAnasal colonization. Concomitantcultures are necessary torecover organisms forepidemiological typing or forfurther susceptibility testing.The ARIES® MRSA Assay isindicated for use with ARIES®Systems.The BD MAX™ MRSA XT assayperformed on the BD MAX™ Systemis an automated qualitative in vitrodiagnostic test for the directdetection of methicillin-resistantStaphylococcus aureus (MRSA) DNAfrom nasal swabs in patients at riskfor nasal colonization. The testutilizes real-time polymerase chainreaction (PCR) for the amplificationof MRSA DNA and fluorogenic target-specific hybridization probes for thedetection of the amplified DNA. TheBD MAX™ MRSA XT assay is intendedto aid in the prevention and controlof MRSA infections in healthcaresettings. It is not intended todiagnose MRSA infections nor guideor monitor treatment for MRSAinfections. A negative result does notpreclude nasal colonization.Concomitant cultures are necessaryto recover organisms forepidemiological typing or for furthersusceptibility testing.
Assay Targets(1) orfX , S. aureus complexgene(2) SCCmec junction region(1) SCCmec/orfX junction area ofmethicillin-resistantStaphylococcus aureus (i.e.,MREJ for SCCmec Right)
(3) mecA and mecC genes, formethicillin resistanceExtremity Junction). The BDMAX™ MRSA XT assay isdesigned to detect MREJ types i,ii, iii, iv, v, vi, vii, ix, xiii, xiv, andxxi; and(2) mecA and mecC genes formethicillin resistance.
Sample TypesNasal swabsNasal swabs
Assay TypeReal-time PCRAmplification: PCR
Assay ResultsQualitativeQualitative
Assay ControlsSample Processing Control (SPC)Specimen Processing Control (SPC)

Table 1: Similarities between New Device and Predicate

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the word "Luminex" in a bold, sans-serif font. There is a small, gray circle above the "i" in "Luminex". The word is in black and the background is white. The registered trademark symbol is to the right of the "x".

Table 2: Differences between New Device and Predicate

ItemNew DevicePredicate: K133605
Extraction MethodAutomated by the ARIES® SystemsAutomated by the BD MAX™ System
Assay InstrumentARIES® SystemsBD MAX™ System
DetectionFluorescent reporter probes for each target, and melting curve analysis.Fluorogenic target-specific hybridization.

K. Test Principle:

The ARIES®MRSA Assay PCR amplification and detection reagents contain primers and probes specific to the methicillin resistance genes (mecA and mecC), a Staphylococcus aureus complex gene (orfX), the SCCmec region, and the Sample Processing Control (SPC) sequence. Each of the probes are labeled with a distinct fluorophore and detected in distinct channels of the ARIES® System. The probes contain a fluorophore on the 5' end of the oligonucleotide sequence and a quencher at the 3′ end of the oligonucleotide sequence such that in random coil state when the probe is not hybridized to the target sequence, the fluorescent signal is quenched. PCR amplification is performed and assay fluorescence is monitored. Hybridization of a fluorescently labeled probe to the amplified target results in the release of quenching and generation of fluorescence signal that is indicative of PCR generated amplicon.

L. Performance Characteristics:

    1. Analytical performance:
    • Precision/Reproducibility: a.

Precision

{7}------------------------------------------------

Within-laboratory precision/repeatability of the ARIES® MRSA Assay was conducted by testing a blinded panel of samples by two operators from a single site using a single lot of reagents and assessing the variabilities of the results. The panel consisted of five samples: two methicillin-resistant Staphylococcus aureus (MRSA) strains each prepared at 1x and 5x Limit of Detection (LoD) in simulated nasal matrix in Modified Liquid Amies (SNM+LA) and a negative sample (SNM+LA only). Each sample was tested in triplicate by each of the two operators each day for five non-consecutive days. The results showed that each operator generated 100% expected MRSA results, inclusive of MRSA positive and MRSA negative calls demonstrating that the ARIES® MRSA Assay is reproducible between independent operators. The results of the precision/repeatability study are presented in Table 3.

Strain/LevelPositive/Tested (%)
MRSA mecA+Moderate Positive5X LoD130/30(100)
MRSA mecA+Low Positive1X LoD30/30(100)
MRSA mecC+Moderate Positive5X LoD30/302(100)
MRSA mecC+Low Positive1X LoD30/30(100)
Negative0/30(0.0)

Table 3: Precision Study

LoD; Limit of detection for MRSA Strains BEI NR-46232 and ATCC BAA-2313 5X LoD MRSA mecA+ strain BEI NR-46232: 9.75E+04 CFU/mL; 1X LoD MRSA mecA+ BEI NR-46232: 1.95E+04 CFU/mL; 5X LoD MRSA strain mecC+ ATCC BAA-2313: 3.88E+05 CFU/mL; 1X LoD MRSA mecC+ ATCC BAA-2313: 7.75E+04 CFU/mL

2 A single sample was reported as Invalid on initial testing; reported as Positive upon repeat testing.

Reproducibility

Site-to-site reproducibility of the ARIES® MRSA Assay was conducted by testing a blinded panel of samples at 3 independent sites. The panel consisted of 5 samples: 2 methicillin-resistant Staphylococcus aureus (MRSA) strains each prepared at 1x and 5x Limit of Detection (LoD) in simulated nasal matrix in

{8}------------------------------------------------

modified Liguid Amies (SNM+LA) and a negative sample (SNM+LA only). Each sample was tested in triplicate by 2 operators for 5 non-consecutive days using a single lot of assay reagents at each site. The ARIES® MRSA Assay generated 100% expected MRSA positive results for MRSA positive samples and 100% expected MRSA negative results (0% MRSA positivity) for MRSA negative samples demonstrating reproducibility of the assay. The results of the reproducibility study are presented in Table 4.

Positive/Number (%)
Strain/LevelSite 1Site 2Site 3Overall
MRSA mecA+Moderate Positive5X LoD¹30/30(100)30/30(100)30/30(100)90/90(100)
MRSA mecA+Low Positive1X LoD30/30(100)30/30(100)30/30(100)90/90(100)
MRSA mecC+Moderate Positive5X LoD30/30(100)30/30(100)30/302(100)90/90(100)
MRSA mecC+Low Positive1X LoD30/302(100)30/30(100)30/30(100)90/90(100)
Negative0/30(0.0)0/303(0.0)0/30(0.0)0/90(0.0)

Table 4: Site-to-Site Reproducibility

LoD; Limit of detection for MRSA Strains BEI NR-46232 and ATCC BAA-2313 5X LoD MRSA mecA+ strain BEI NR-46232: 9.75E+04 CFU/mL; 1X LoD MRSA mecA+ BEI NR-46232: 1.95E+04 CFU/mL; 5X LoD MRSA strain mecC+ ATCC BAA-2313: 3.88E+05 CFU/mL; 1X LoD MRSA mecC+ ATCC BAA-2313: 7.75E+04 CFU/mL

2 A single sample was reported as Invalid on initial testing; reported as Positive upon repeat testing.

က A single sample was reported as Invalid on initial testing; reported as Negative upon repeat testing.

  • b. Linearity/assay reportable range:
    Not applicable. The ARIES® MRSA Assay is a qualitative assay.

  • Traceability, Stability, Expected values (controls, calibrators, or methods): C.

Controls:

Process Control

Each ARIES® MRSA Assay cassette contains a Sample Processing Control (SPC), which is processed with the sample and analyzed during the amplification reaction. The SPC verifies nucleic acid extraction, and proper reagent, cassette, ARIES® System, and assay protocol performance. The SPC has a known melting

{9}------------------------------------------------

temperature (Tm) range and Ct range. Each time an assay is run, the system measures the temperature and fluorescence intensity of the SPC control to ensure the thermal and optical subsystems have remained in calibration.

External Controls

External Positive and Negative Controls were tested on a daily basis during the prospective Clinical Study using a total of 5 ARIES systems and 3 ARIES MRSA Assay cassette lots. The Positive External Control comprised a standardized suspension of a strain of MRSA at 1.95E+05 CFU/mL (10X LoD). The Negative External Control comprised a standardized suspension of a strain of S. epidermidis at 4.23E+04 CFU/mL (10X LoD). On initial testing, 182/183 (99.5%) Positive and 178/183 (97.3%) Negative External Controls produced the expected results. Upon repeat testing, all controls produced the expected results.

External controls should be tested according to guidelines or requirements of local, provincial and/or federal regulations or accreditation organizations. A reference methicillin-resistant Staphylococcus aureus strain or well characterized methicillin-resistant Staphylococcus aureus clinical isolates may be used as positive controls. The ARIES® MRSA Assay Kit does not include external positive or negative controls.

Stability:

Specimen Stability

Specimen stability was evaluated using contrived specimens, stored at 2°C and 30°C and tested with the ARIES® MRSA Assay. This was assessed by testing six replicates of each of two methicillin-resistant Staphylococcus aureus (MRSA) strains prepared at 3x and 10x Limit of Detection (LoD) in Natural Nasal Matrix (NNM), as well as a negative sample (NNM only). Following testing on the day of sample preparation, specimens stored at 2±2°C were tested at 5 time points over 10 days and specimens stored at 30±1℃ were tested at 7 time points over 10 days. The results demonstrated that MRSA specimens stored at both temperatures (2°C and 30°C) generated 100% expected MRSA positive results across all time points tested. An overall 0.53% (4/757) cassette invalid rate was observed for the testing of MRSA strains and NNM samples in this study. MRSA specimens are stable for up to 10 days when stored at 2-30°C.

Shelf-Life Stability

A real time stability study was performed to evaluate the stability of the ARIES® MRSA Assay Cassette in order to establish a shelf life. This was assessed by testing six replicates of positive target and six replicates of negative target on

{10}------------------------------------------------

Luminex

three different lots of ARIES® MRSA Assay Cassettes stored at two different temperatures (4°C and 25°C/room temperature) at ten different time points extending out to 19 months. ARIES® MRSA Assay Cassettes generated the expected MRSA positive calls for all positive replicates and MRSA negative calls for all negative replicates stored at both 4°C and 25°C at all time points across all cassette lots.

Cassette open box stability evaluated the performance of the ARIES® MRSA Assay Cassettes after removal from the cassette pouch and exposed to ambient temperatures, humidity and light for 10 hours. The results showed that all MRSA positive and MRSA negative samples through the 10 hour study duration generated 100% expected results at all 5 time points. Three lots of cassettes were used to assess open box stability. Over the course of 10 hours, all three lots of cassettes produced expected results, showing that ARIES® MRSA Assay Cassettes are stable in ambient temperatures for up to 10 hours after they have been removed from the storage pouch.

  • d. Detection Limit:
    A Limit of Detection (LoD) study was performed to evaluate the analytical sensitivity of the ARIES MRSA Assay using one strain of MRSA mecA+ (NR-46232) and one strain of MRSA mecC+ (BAA-2312). The preliminary LoD for each strain was determined by performing a six point, five-fold dilution series in natural nasal matrix (NNM). The preliminary LoD was identified as the lowest concentration that had 100% positivity (out of at least 5 replicates). Concentrations in colony forming units per milliliter (CFU/mL) and CFU per cassette were determined for dilutions of each strain by plating the dilutions used in the preliminary study and colony counting.

The LoD concentrations determined in the preliminary study were confirmed with the same strains diluted in NNM to concentrations at, below, and above the LoD and tested with 20 replicates. The confirmed LoD concentrations, identified as the lowest concentration detected in ≥ 95% of replicates, for each strain are presented in Table 5.

SpeciesStrainMatrix TypeLoD(CFU/mL)LoD(CFU/cassette)ARIES MRSAPositivity95% C.I.
MRSAmecA+NR-46232NNM1.95E+043.89E+03100% (20/20)83.9% - 100%
SNM+LA1.95E+043.89E+0395.0% (19/20)76.4% - 99.1%
MRSAmecC+BAA-2312NNM7.75E+041.55E+04100% (20/20)83.9% - 100%

Table 5: LoD Results for MRSA strains

{11}------------------------------------------------

Image /page/11/Picture/1 description: The image shows the word "Luminex" in a bold, sans-serif font. There is a small, filled circle above the "i" in "Luminex". A small registration mark is present to the right of the "x".

Inclusivity (Analytical Reactivity) e.

Inclusivity (Analytical Reactivity) of ARIES® MRSA Assay was evaluated by testing fifty-five (55) methicillin-resistant Staphylococcus aureus (MRSA) strains including the two strains tested in the Limit of Detection (LoD) study. These strains are representative SCCmec, Pulse-Field Gel Electrophoresis (PFGE), and MREJ variants from diverse geographic locations. Each strain was diluted in simulated nasal matrix in Modified Liquid Amies (SNM+LA) to a final concentration of 3X LoD and tested in triplicate. The results demonstrated that all 55 strains generated 100% (3/3 replicates) MRSA positivity as expected. The results of the analytical reactivity (inclusivity) study are presented in Table 6.

StrainDescriptionSourceCatalog #Strain/LocationPFGE TypeSCCmecTypeMRSAPositivity (%)
MRSA mecA+BEINR-46232(LoD strain)NRS703/MinnesotaUSA300IV100% (3/3)
ATCCBAA-38N/AN/AI100% (3/3)
ATCCBAA-1686N/AN/AII100% (3/3)
ATCCBAA-1687¹N/AN/AII100% (3/3)
ATCCBAA-1692N/AUSA100II100% (3/3)
ATCCBAA-1681N/AUSA100II100% (3/3)
ATCCBAA-1682N/AUSA100II100% (3/3)
ATCCBAA-1699N/AUSA100II100% (3/3)
BEINR-46250NRS721/OregonUSA100II100% (3/3)
ATCCBAA-1761N/AUSA100II100% (3/3)
ATCCBAA-1750N/AUSA200II100% (3/3)
BEINR-46251NRS722/OregonUSA200II100% (3/3)
ATCCBAA-39N/AN/AIII100% (3/3)
ATCCBAA-40N/AN/AIIIa100% (3/3)
ATCCBAA-1717N/AUSA300IVa100% (3/3)
ATCCBAA-1762N/AUSA300IVb100% (3/3)
ATCCBAA-1680N/AUSA300IVa100% (3/3)
BEINR-46070NRS384/MississippiUSA300IV100% (3/3)
ATCCBAA-1683N/AUSA400IVa100% (3/3)
ATCCBAA-1707N/AUSA400IV100% (3/3)
ATCCBAA-1752N/AUSA400IV100% (3/3)
ATCCBAA-1757N/AUSA400IV100% (3/3)
ATCCBAA-1696N/AUSA400IVa100% (3/3)
StrainDescriptionSourceCatalog #Strain/LocationPFGE TypeSCCmecTypeMRSAPositivity (%)
BEINR-46207NRS678/ConnecticutUSA500IV100% (3/3)
ATCCBAA-1689N/AUSA500IV100% (3/3)
BEINR-46220NRS691/GeorgiaUSA500IV100% (3/3)
ATCCBAA-1688N/AN/AV100% (3/3)
ATCCBAA-42N/AN/AVI100% (3/3)
BEINR-46177NRS648/CaliforniaUSA600II100% (3/3)
ATCCBAA-1751N/AUSA600II100% (3/3)
BEINR-46218NRS689/GeorgiaUSA700IV100% (3/3)
ATCCBAA-1766N/AUSA700V100% (3/3)
BEINR-46221NRS692/GeorgiaUSA800IV100% (3/3)
ATCCBAA-1771N/AUSA800IV100% (3/3)
BEINR-46197NRS668/ColoradoUSA800N/A100% (3/3)
ATCCBAA-1747N/AUSA1000IV100% (3/3)
ATCCBAA-1769N/AUSA1000IV100% (3/3)
ATCCBAA-1764N/AUSA1100IV100% (3/3)
ATCCBAA-1767N/AUSA1100IV100% (3/3)
ATCC33592N/AN/AIII100% (3/3)
ATCC33593N/AN/AIII100% (3/3)
ATCC43300N/AN/AII100% (3/3)
ATCC700698N/AN/AII100% (3/3)
ATCC700699N/AN/AII100% (3/3)
ATCC700787N/AN/AII100% (3/3)
ATCC700788N/AN/AII100% (3/3)
ATCC700789N/AN/AII100% (3/3)
ATCCBAA-41N/AN/AII100% (3/3)
ATCCBAA-43N/AN/AIIIa100% (3/3)
ATCCBAA-44N/AN/AI100% (3/3)
ATCCBAA-1720N/AN/AII100% (3/3)
ATCCBAA-2094N/AN/AV100% (3/3)
ATCCBAA-2096N/AN/AIV100% (3/3)
MRSA mecC+ATCCBAA-2313(LoD strain)N/AN/AXI100% (3/3)
ATCCBAA-2312N/AN/AXI100% (3/3)

Table 6: Inclusivity (Analytical Reactivity) Results

{12}------------------------------------------------

Image /page/12/Picture/0 description: The image shows the title of a document. The title is "ARIES® MRSA Assay 510(k) Summary". The text is black and the background is white.

1 One replicate was reported as Invalid on initial testing; reported as Positive upon repeat testing.

{13}------------------------------------------------

Image /page/13/Picture/0 description: The image shows the word "Luminex" in a bold, sans-serif font. The word is written in black, and there is a small, dark gray circle above the "i" in "Luminex". There is a small trademark symbol to the right of the "x".

Challenge Study f.

An additional analytical study was carried out to evaluate the analytical performance of the ARIES® MRSA Assay using a panel of challenge strains. The challenge panel included 16 methicillin-resistant Staphylococcus aureus (MRSA) strains with high minimum inhibitory concentration (MIC) values of ≥ 16 µg/mL oxacillin and 17 MRSA strains with low MIC values of ≤ 8 µg/mL oxacillin, four borderline oxacillin-resistant Staphylococcus aureus (BORSA) strains, 16 empty cassette variants of Staphylococcus aureus strains, and one methicillin-resistant Staphylococcus epidermidis (MRSE) strain. The MRSA strains were tested at 3X LoD and all other strains were tested at a concentration of 1E+06 CFU/mL. Each strain was tested in triplicate. Fourteen of the sixteen (14/16) MRSA with high MICs and 16/17 of the MRSA with low MICs generated the expected MRSA positive results at 3X LoD (100% MRSA Positive). Two strains of MRSA with high MIC values and one strain of MRSA with low MIC value that did not generate the expected MRSA positive results when tested at 3X LoD were re-tested at 5X LoD and generated the expected MRSA positive results (100% MRSA Positive). The four BORSA strains, the MRSE strain, and all empty cassette variants of Staphylococcus aureus generated the expected 0% MRSA positive results (100% MRSA Negative). Results of the challenge study are presented in Table 7.

TargetDescriptionSourceStrain IDTestConcentrationARIES® MRSAPositivity
ARLGARLG-16433X LoD100% (3/3)
ARLGARLG-16443X LoD100% (3/3)
ARLGARLG-16453X LoD100% (3/3)
ARLGARLG-16463X LoD100% (3/3)
ARLGARLG-16623X LoD100% (3/3)
ARLGARLG-166913X LoD66.7% (2/3)
ARLGARLG-16043X LoD100% (3/3)
MRSA with HighOxacillin MICARLGARLG-161313X LoD66.7% (2/3)
CDC AR BankAR-02153X LoD100% (3/3)
CDC AR BankAR-02183X LoD100% (3/3)
CDC AR BankAR-02193X LoD100% (3/3)
CDC AR BankAR-02203X LoD100% (3/3)
CDC AR BankAR-02233X LoD100% (3/3)
CDC AR BankAR-02243X LoD100% (3/3)
CDC AR BankAR-02273X LoD100% (3/3)
CDC AR BankAR-02283X LoD100% (3/3)
MRSA withLow OxacillinMICARLGARLG-164213X LoD66.7% (2/3)
Lyon University201302373X LoD100% (3/3)
Lyon University201305243X LoD100% (3/3)
TargetDescriptionSourceStrain IDTestConcentrationARIES® MRSAPositivity
CDC AR BankAR-02163X LoD100% (3/3)
CDC AR BankAR-02173X LoD100% (3/3)
CDC AR BankAR-02213X LoD100% (3/3)
CDC AR BankAR-02253X LoD100% (3/3)
CDC AR BankAR-02263X LoD100% (3/3)
ATCCBAA-16883X LoD100% (3/3)
CDC AR BankAR-4723X LoD100% (3/3)
CDC AR BankAR-4733X LoD100% (3/3)
CDC AR BankAR-4743X LoD100% (3/3)
CDC AR BankAR-4753X LoD100% (3/3)
CDC AR BankAR-4763X LoD100% (3/3)
CDC AR BankAR-4773X LoD100% (3/3)
CDC AR BankAR-4783X LoD100% (3/3)
CDC AR BankAR-4793X LoD100% (3/3)
CDC AR BankAR-04891E+06 CFU/mL0% (0/3)
BORSACDC AR BankAR-04901E+06 CFU/mL0% (0/3)
CDC AR BankAR-04911E+06 CFU/mL0% (0/3)
CDC AR BankAR-04921E+06 CFU/mL0% (0/3)
Lyon University201012701E+06 CFU/mL0% (0/3)
Lyon University201128961E+06 CFU/mL0% (0/3)
Lyon University201129111E+06 CFU/mL0% (0/3)
Lyon University201205561E+06 CFU/mL0% (0/3)
Lyon University201208441E+06 CFU/mL0% (0/3)
Lyon University2012087121E+06 CFU/mL0% (0/3)
Lyon University201209841E+06 CFU/mL0% (0/3)
Empty CassetteVariant of SALyon University201214691E+06 CFU/mL0% (0/3)
Lyon University2012154421E+06 CFU/mL0% (0/3)
Lyon University201216351E+06 CFU/mL0% (0/3)
Lyon University201218911E+06 CFU/mL0% (0/3)
Lyon University201307691E+06 CFU/mL0% (0/3)
Lyon University201311901E+06 CFU/mL0% (0/3)
Lyon University201312731E+06 CFU/mL0% (0/3)
Lyon University201317271E+06 CFU/mL0% (0/3)
Lyon University201408521E+06 CFU/mL0% (0/3)
MRSEATCC516251E+06 CFU/mL0% (0/3)

Table 7: Challenge Study Results

{14}------------------------------------------------

1 Two strains of MRSA with high MIC values (ARLG-1669, ARLG-1613) and one strain of MRSA with low MIC value (ARLG-1642) that did not generate the expected MRSA positive results (100% MRSA positive) when tested at 3X LoD were re-tested at 5X LoD and generated the expected MRSA positive results (100% MRSA positive; 3/3).

² One replicate was reported as Invalid on initial testing; reported as Negative upon repeat testing.

{15}------------------------------------------------

Image /page/15/Picture/0 description: The image shows the word "Luminex" in a bold, sans-serif font. There is a gray dot above the "i" in "Luminex". The word is in black and the background is white.

g. Analytical specificity

Cross-Reactivity and Microbial Interference:

Analytical specificity of the ARIES® MRSA Assay in the presence of potentially cross-reacting or interfering organisms was evaluated. Ninety-nine (99) organisms commonly present in nasal specimen collection sites were tested at concentrations of 1E+06 CFU/mL for non-viral organisms, 1E+05 TCID50/mL for viruses and 5 µg/mL for human genomic DNA. The study was performed using one strain of mecA+ methicillin-resistant Staphylococcus aureus (MRSA) (NR-46232) and one strain of mecC+ MRSA (BAA-2313) at 3x Limit of Detection (LoD), as well as a negative sample consisting of simulated nasal matrix in Modified Liquid Amies (SNM+LA). Target samples at an intermediate concentration prepared in SNM+LA were spiked with potentially cross-reactive organisms prepared in SNM+LA before being tested on the ARIES® MRSA Assay in triplicates. The test results demonstrated that all replicates of both MRSA strains in absence and presence of the potentially cross-reactive organisms generated 100% expected MRSA positive results; and all replicates of SNM+LA tested in absence and presence of the cross-reacting organisms generated 100% expected MRSA negative results (0% MRSA positivity). A summary of all samples tested are presented in Table 8.

Non-Staphylococcal organisms
Acinetobacter baumannii3Listeria monocytogenes
Acinetobacter haemolyticusLegionella pneumophila
Bacillus cereusMoraxella catarrhalis
Bordetella pertussisMicrococcus luteus
Candida albicansMycoplasma pneumoniae
Citrobacter freundiiMycobacterium tuberculosis avirulent
Candida glabrataNeisseria meningitidis
Citrobacter koseriListeria monocytogenes
Chlamydia pneumoniaePasteurella aerogenes
Corynebacterium bovisPseudomonas aeruginosa
Corynebacterium flavescensPseudomonas fluorescens1
Corynebacterium genitaliumProteus mirabilis
Cryptococcus neoformansProvidencia stuartii
Enterobacter aerogenesProteus vulgaris
Enterobacter cloacae1Salmonella enterica subsp. Enterica
Enterococcus faecalisSerratia marcescens
Enterococcus faeciumStreptococcus agalactiae
Escherichia coli (0157:H7)Streptococcus anginosus
Enterococcus flavescensStreptococcus mitis
Enterococcus gallinarumStreptococcus mutans
Enterococcus hiraeStreptococcus pneumoniae

Table 8: Cross-reactivity/Microbial Interference Study

{16}------------------------------------------------

Image /page/16/Picture/0 description: The image shows the word "Luminex" in a bold, sans-serif font. There is a gray dot above the "i" in "Luminex". A small circle with an "R" inside is located to the right of the "x" in "Luminex", indicating that the word is a registered trademark.

Haemophilus influenzaeShigella sonnei
Klebsiella oxytocaStreptococcus pyogenes
Klebsiella pneumoniae (ESBL-producing)Streptococcus salivarius
Klebsiella pneumoniae (KPC-producing)Streptococcus sanguinis
Lactobacillus crispatusStreptococcus suis
Yersinia enterocolitica
Viruses
Adenovirus Type 40Measles virus
Adenovirus Type 7Mumps virus
Coronavirus 229EParainfluenza 1
Coronavirus OC43Parainfluenza 2
CytomegalovirusParainfluenza 3
Epstein Barr VirusRhinovirus type 1A3
Influenza ARSV A
Influenza BRSV B
Human metapneumovirus
Coagulase Negative Staphylococci (CoNS)
Staphylococcus arlettaeStaphylococcus equorum
Staphylococcus captisStaphylococcus felis
Staphylococcus carnosusStaphylococcus gallinarum2
Staphylococcus chromogenesStaphylococcus haemolyticus Z067
Staphylococcus epidermidis (255-01B)Staphylococcus kloosii
Staphylococcus epidermidis (RP12 CIP 106510)Staphylococcus lentus
Staphylococcus epidermidis (MRSE;RP62A)Staphylococcus pulvereri
Staphylococcus epidermidis (CCF 15990)Staphylococcus simulans Z032
Staphylococcus epidermidis (CCF 16471)Staphylococcus warneri Z1133
Coagulase Positive Staphylococci
Staphylococcus pseudointermediusStaphylococcus delphini
Staphylococcus scheleiferi subsp. CoagulansStaphylococcus intermedius
Staphylococcus scheleiferi subsp. SchleiferiStaphylococcus lutrae
Methicillin-susceptible Staphylococcus aureus (MSSA)
Staphylococcus aureus BAA-1749Staphylococcus aureus 29213
Staphylococcus aureus BAA-1765Staphylococcus aureus 0801675
Staphylococcus aureus BAA-1718
Other

| Human genomic DNA

1 One replicate of SNM+LA negative matrix was reported as Invalid on initial testing; reported as Negative upon repeat testing.

² One replicate containing mecA+ MRSA was reported as Invalid on initial testing; reported as Positive upon repeat testing.

3 One replicate containing mecC+ MRSA strain was reported as Invalid on initial testing; reported as Positive upon repeat testing.

Potentially Interfering Substances:

{17}------------------------------------------------

Luminex

Twenty-four (24) potentially interfering non-microbial substances that are commonly present at the site of specimen collection were tested with the ARIES® MRSA Assay. The study was performed using one strain of mecA+ methicillin-resistant Staphylococcus aureus (MRSA) (NR-46232) and one strain of mecC+ MRSA (BAA-2313) at 3x LoD prepared in simulated nasal matrix in Modified Liquid Amies (SNM+LA), as well as SNM+LA alone as a negative control. Samples at intermediate concentrations prepared in SNM+LA were spiked with the non-microbial substances prepared in the appropriate diluent to reach the final concentrations and tested in triplicate. All replicates of both MRSA strains tested in the absence and presence of the potentially interfering substances generated 100% MRSA positivity as expected; and all replicates of SNM+LA alone generated 0% MRSA positivity (100% negativity) as expected. No interference was detected for any of the substances at the concentration tested. Potentially interfering substances that were tested are presented in Table 9.

SubstanceTest Concentration
Whole Blood5% (v/v)
Mucin5 mg/mL
Phenylephrine0.03 µg/mL
Drixoral (Oxymetazoline)10% (v/v)
Benzalkonium chloride0.12%
Propylene glycol20% (v/v)
Sorbitol16.45%
Benzyl alcohol10.5% (v/v)
Hypromellose0.10%
Phosphoric acid1.282 mg/mL
Beclomethasone8.4 µg/mL
Dexamethasone12 µg/mL
Flunisolide5 µg/mL
Triamcinolone22 µg/mL
Budesonide6.30E-03 µg/mL
Mometasone4.50E-04 µg/mL
Flonase (Fluticasone)1.26E-03 µg/mL
ZICAM2 (Galphimia glauca, Histaminum hydrochloricum)10% (v/v)
Benzocaine75 µg/mL
Menthol0.5 mg/mL
Zanamivir31 mg/mL
Mupirocin50 µg/mL

Table 9: Potentially Interfering Substances

{18}------------------------------------------------

Image /page/18/Picture/0 description: The image shows the word "Luminex" in a bold, sans-serif font. There is a small, dark gray circle above the "i" in "Luminex". The word is slightly slanted to the right. The registered trademark symbol is located to the right of the "x".

Tobramycin33 µg/mL
FluMist® (Live intranasal influenza virus vaccine)10% (v/v)
  • 1 One replicate of SNM+LA negative matrix was reported as Invalid on initial testing; reported as Negative upon repeat testing.
  • 2 One replicate containing mecA+ MRSA strain was reported as Invalid on initial testing; reported as Positive upon repeat testing.
  • 3 One replicate containing mecC+ MRSA strain was reported as Invalid on initial testing; reported as Positive upon repeat testing.

Competitive Interference:

Competitive interference was tested with methicillin-resistant Staphylococcus aureus (MRSA) at 1x limit of detection (LoD) and the co-infecting agent, methicillin-susceptible Staphylococcus aureus (MSSA) or methicillin-resistant coagulase-negative staphylococci (MRCoNS), at increasing concentrations. Each combination was tested in triplicate. The results showed that the ARIES® MRSA Assay detected MRSA at 1x LoD in the presence of increasing concentrations of MSSA or MRCoNS. No competitive interference in the ARIES® MRSA Assay was observed for co-infections of MRSA with MSSA or MRCoNS.

Carry-Over/Cross-Contamination:

Carry-over and cross contamination for the ARIES® MRSA Assay was evaluated by testing sixty (60) high positive (1.0E+07 CFU/mL) MRSA samples prepared in simulated nasal matrix in modified Liquid Amies (SNM+LA) and sixty (60) MRSA negative samples consisting of SNM+LA alone. The high positive samples were run adjacent to negative samples in an alternating pattern across twenty (20) consecutive module runs using two ARIES® instruments. Overall percent agreement with expected results was 100% for both high positive and negative samples. One replicate was reported Invalid on initial testing; reported as Positive upon repeat testing. No carry-over or cross contamination was observed in the study.

h. Assay cut-off

For the ARIES® MRSA Assay, each target (mecA/mecC, orfX, and SCCmec) has a Ct cut-off, Tm window, and Tm Peak Threshold. In addition, the internal sample processing control (SPC) also has a corresponding Ct cut-off, Tm window, and Tm Peak Threshold. Collectively, the cut-off values compose the assay protocol file parameters, which are used to determine the assay result for the detection of the target as Positive, Negative, or Invalid. These values are hard-coded into the ARIES® MRSA Assay Protocol File and are not modifiable. The Assay Protocol File parameters were determined, and their performance in the ARIES® MRSA Assay were evaluated according to the following general procedure:

{19}------------------------------------------------

  • Initial Assay Protocol File parameters were set during internal optimization . and benchmarking studies.
  • . The final Assay Protocol File parameters were then established during internal verification studies using data from optimization, benchmarking and verification.
  • . The selected Assay Protocol File parameter values were utilized in the determination of assay performance in the multi-site clinical trial conducted for the ARIES MRSA Assay.

The specific assay parameters for the ARIES® MRSA Assay are considered confidential and proprietary.

    1. Comparison Studies
    • Method comparison with predicate device: a.

Not applicable.

  • b. Nasal Swab Comparison:
    A nasal swab equivalency study was performed to evaluate the reproducibility of the ARIES® MRSA Assay with two different nasal swab types, Regular Nylon Flocked Swab (Copan Catalog Number: 480C) and Flexible Minitip Nylon Flocked Swab (Copan Catalog Number: 482C). The swabs were evaluated using one strain of Methicillin-resistant Staphylococcus aureus (MRSA) mecA+ (NR-46232) at three concentrations, as well as a negative sample (Simulated nasal matrix in Modified Liquid Amies, SNM+LA). Samples at intermediate concentrations prepared in SNM+LA were transferred to Modified Liquid Amies using each of the two nasal swab types to reach the final testing concentrations at 3x LoD, 5x LoD and 10x LoD, respectively, and then tested on the ARIES® MRSA Assay. The test results demonstrated that both swab types generated 100% expected MRSA positivity for each strain of the concentrations tested. Both swab types also generated 0% positivity (100% negativity) for negative samples. Swab equivalency study results are presented in Table 10.
Assay TargetPartNumberTest Concentration(CFU/mL)MRSA Positivity
MRSA $mecA+$NR-462323x LoD (5.85E+04)5x LoD (9.75E+04)100% (6/6)100% (6/6)
MRSA $mecA+$NR-4623210x LoD (1.95E+05)Negative(SNM+LA)100% (6/6)100% (6/6)
Negative(SNM+LA)N/AN/AN/A0% (0/6)0% (0/6)
Table 10: ARIES® MRSA Assay Nasal Swab Equivalency Results

{20}------------------------------------------------

3. Clinical Performance:

Clinical performance of the ARIES® MRSA Assay for nasal swab specimens collected from patients at risk for methicillin-resistant Staphylococcus aureus (MRSA) colonization was established through a clinical study.

Performance of the ARIES® MRSA Assay was evaluated prospectively from August 2018 to February 2019 at four (4) geographically distinct clinical sites within the United States using the ARIES® System. Specimens included in the clinical study consisted of excess leftover de-identified, nasal clinical specimens collected using the Liquid Amies Elution Swab (Eswab™) Collection and Transport system, or equivalent, from patients at risk for nasal colonization. All eligible clinical specimens were tested by both the reference method (direct and enriched bacterial culture) and ARIES® MRSA Assay and the results compared. Reference method testing was performed at a centralized testing facility while ARIES® MRSA Assay testing was performed at each clinical site on their own clinical specimens.

A total of 2254 nasal swab specimens from subjects at risk for MRSA nasal colonization were collected. Of these 2254 specimens, 472 were excluded from the study based on inclusion/exclusion criteria leaving a total of 1,782 unique specimens that met the pre-determined eligibility criteria. Of the 472 excluded specimens, 275 specimens were from a subject who had taken a systemic or topical nasal antibiotics within 48 hours to 1 week prior to specimen collection, 93 specimens were not tested by the reference method due to either a delay in shipment or because testing could not start within 48 hours of collection, 27 specimens had an Indeterminate culture result, 24 specimens had insufficient volume for testing by both or either the ARIES MRSA Assay or reference method, 20 specimens were from a subject that had a known positive lab result for MRSA in the past 90 days of specimen collection, 9 specimens were not tested according to the reference method protocol, 8 specimens had Proteus mirabilis swarming on the blood agar plate for the reference method, 6 specimens did not have passing external controls for ARIES MRSA/SA Assay testing, 4 specimens were processed using the incorrect ARIES MRSA/SA Assay reagents, 3 specimens were from subjects previously enrolled in the study, 3 specimens were tested by the culture method after 48 hours from collection.

1,782 specimens were enrolled in the study and tested for methicillin-resistant Staphylococcus aureus by both the reference method and the ARIES® MRSA Assay. There were 20 specimens (20/1,782, 1.1%) that when tested with ARIES® MRSA Assay yielded an invalid result due to run failure or instrument error. None of these specimens were re-tested due to insufficient specimen volume.

For the 1,762 eligible specimens that were included in the device performance calculations, clinical sensitivity of the ARIES® MRSA Assay was 93.3% (97/104) with a lower bound 95% confidence interval of 87%, when compared to direct and enriched

{21}------------------------------------------------

Luminex

bacterial culture. Clinical specificity of the ARIES® MRSA Assay was 93.5% (1550/1658) with a lower bound 95% confidence interval of 92%, when compared to direct and enriched bacterial culture. Of the 108 specimens that were MRSAnegative by culture but MRSA-positive by the ARIES MRSA Assay, culture showed that 63 specimens were S. aureus and 45 were negative (no growth). Positive percent agreement of the ARIES® MRSA Assay against direct bacterial culture was 93.5% (87/93) with a lower bound 95% confidence interval of 87%. Negative percent agreement of the ARIES® MRSA Assay against direct bacterial culture was 92.9% (1551/1669) with a lower bound 95% confidence interval of 92%.

Table 11: ARIES® MRSA Assay Performance Compared to Direct and Enriched
Culture
ARIES® MRSA AssayDirect and Enriched Bacterial Culture for MRSA
PositiveNegativeTOTAL
Positive97108205
Negative715501557
TOTAL10416581762
Sensitivity (95% CI)93.3% (87% - 97%)
Specificity (95% CI)93.5% (92% - 95%)
Table 12: ARIES® MRSA Assay Performance Compared to Direct Culture
ARIES® MRSA AssayDirect Bacterial Culture for MRSA
PositiveNegativeTOTAL
Positive87118205
Negative615511557
TOTAL9316691762
Positive PercentAgreement (95% CI)93.5% (87% - 97%)
Negative PercentAgreement (95% CI)92.9% (92% - 94%)

The study results demonstrate that the diagnostic accuracy of the ARIES® MRSA Assay is acceptable for the detection of methicillin-resistant Staphylococcus aureus (MRSA) DNA from nasal swabs in patients at risk for nasal colonization.

4. Expected values/Reference range:

Table 13 provides a summary of the expected values of the 1762 prospectively collected nasal swab specimens that were included in the prospective analysis.

{22}------------------------------------------------

Image /page/22/Picture/0 description: The image shows the logo for Luminex. The logo is in black and white, with the word "Luminex" in a bold, sans-serif font. There is a small, gray circle above the "i" in "Luminex". A registered trademark symbol is present to the right of the word.

Table 13: ARIES® MRSA Assay Expected Values of Clinical Study Population

Number ofSubjects% Distribution
Gender
Male12212.8% (122/954)
Female8310.3% (83/808)
Overall20511.6% (205/1762)
Age (yrs)
<2215.2% (21/405)
2 - 11218.4% (21/251)
12 - 211910.1% (19/189)
22 - 597415.4% (74/481)
≥607016.1% (70/436)
Overall20511.6% (205/1762)

M. Proposed Labeling:

The labeling provided in the submission satisfies the requirements of 21 CFR 809.10.

N. Conclusion:

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

§ 866.1640 Antimicrobial susceptibility test powder.

(a)
Identification. An antimicrobial susceptibility test powder is a device that consists of an antimicrobial drug powder packaged in vials in specified amounts and intended for use in clinical laboratories for determining in vitro susceptibility of bacterial pathogens to these therapeutic agents. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.(b)
Classification. Class II (performance standards).