K Number
K191698

Validate with FDA (Live)

Date Cleared
2020-07-16

(387 days)

Product Code
Regulation Number
888.3650
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

When used in the Humelock II Cemented Shoulder System: The Humelock II Cemented Shoulder System is indicated for use in total and hemishoulder replacement to treat:

  1. Proximal humeral fractures
  2. A severely painful and/or disabled joint resulting from osteoarthritis, traumatic arthritis or rheumatoid arthritis;
  3. Other difficult clinical problems where shoulder arthrodesis or resection arthroplasty are not acceptable (e.g. revision of a previously implanted primary component, a humeral plate or a humeral nail).

The humeral stem and glenoid components of the Humelock II Cemented Shoulder System are intended for cemented use only.

TiN coated Heads are not intended for hemi-shoulder arthroplasty

When used with the Humeris™ Shoulder System: In an anatomic shoulder configuration, the Humeris Shoulder System is indicated for use in total and hemi-shoulder replacement to treat:

  • A severely painful and/or disabled joint resulting from osteoarthritis or rheumatoid arthritis;
  • Other difficult clinical problems where shoulder arthrodesis or resection arthroplasty are not acceptable (e.g. revision of a previously implanted primary component, a humeral plate or a humeral nail).

TiN coated Heads are not intended for hemi-shoulder arthroplasty.

In a reverse shoulder configuration, the Humeris Shoulder is indicated for primary or revision total shoulder arthroplasty for the relief of pain and to improve function in patients with a massive and non-repairable rotator cuff tear.

The patient's joint must be anatomically and structurally suited to receive the selected implants and a functional deltoid muscle is necessary to use the device.

The humeral stem of the Humeris Cementless Shoulder is intended for cementless use only. The humeral stem of the Humeris Cemented Shoulder is intended for cemented use only. The glenoid components of the Humeris Shoulder System are intended for cemented use only. The glenoid baseplate component is intended for cementless use with the addition of screws for fixation.

Device Description

The TiN Coated Humeral Head is a new component for Humelock II total shoulder replacement systems. The TiN Coated Humeral Head has a titanium nitride (TiN) coating, which is applied to the predicate humeral heads made of cobalt chromium molybdenum (CoCr). Compatible components for use with the TiN coated humeral heads to complete the total shoulder replacement construct are the same as those previously cleared compatible components for use with the CoCr humeral heads in the respective predicate devices, including the primary predicate (K123814) and K163669 for the Humeris Shoulder System.

AI/ML Overview

This document is a 510(k) Premarket Notification from the FDA regarding a TiN Coated Humeral Head. It does not describe a study that proves a device meets acceptance criteria for an AI/ML medical device. Instead, it details the substantial equivalence determination for a physical orthopedic implant.

Therefore, I cannot extract the requested information (acceptance criteria table, sample size, expert ground truth, MRMC study, standalone performance, training set details) because this document does not pertain to an AI/ML medical device and does not contain information about a study proving an algorithm's performance.

The document states:

  • "Clinical testing was not necessary to determine substantial equivalence of the TiN Coated Humeral Head to the predicate devices." (Page 7)
  • The determination is based on the device being "identical to the primary predicate with the only modification of added TiN coating" and a review of "risk analysis, design controls, biocompatibility, wear properties of the TiN coating, and verification and validation activities to demonstrate that the TiN coating does not increase risk and does not raise new questions of safety and effectiveness." (Page 7)

{0}------------------------------------------------

July 16, 2020

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

FX Shoulder USA, Inc. Kathy Trier VP Regulatory, Quality, Clinical, Compliance 13465 Midway Road, Suite 101 DALLAS, TX 75244

Re: K191698

Trade/Device Name: TiN Coated Humeral Head Regulation Number: 21 CFR 888.3650 Regulation Name: Shoulder joint metal/polymer non-constrained cemented prosthesis Regulatory Class: Class II Product Code: KWT, PHX Dated: July 7, 2020 Received: July 8, 2020

Dear Kathy Trier:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see

{1}------------------------------------------------

https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Michael Owens Acting Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number -K191698

Device Name TiN Coated Humeral Head

Indications for Use (Describe)

When used in the Humelock II Cemented Shoulder System:

The Humelock II Cemented Shoulder System is indicated for use in total and hemi-shoulder replacement to treat:

  1. Proximal humeral fractures
    1. A severely painful and/or disabled joint resulting from osteoarthritis or rheumatoid arthritis;
    1. Other difficult clinical problems where shoulder arthrodesis or resection arthroplasty are not acceptable (e.g.

revision of a previously implanted primary component, a humeral plate or a humeral nail).

The humeral stem and glenoid components of the Humelock II Cemented Shoulder System are intended for cemented use only.

TiN coated Heads are not intended for hemi- shoulder arthroplasty.

When used in the Humeris Shoulder System:

In an anatomic shoulder configuration, the Humeris Shoulder System is indicated for use in total and hemi-shoulder replacement to treat:

  • A severely painful and/or disabled joint resulting from osteoarthritis or rheumatoid arthritis;

  • Other difficult clinical problems where shoulder arthrodesis or resection arthroplasty are not acceptable (e.g. revision of a previously implanted primary component, a humeral plate or a humeral nail).

TiN coated Heads are not intended for hemi- shoulder arthroplasty.

In a reverse shoulder configuration, the Humeris Shoulder is indicated for primary or revision total shoulder arthroplasty for the relief of pain and to improve function in patients with a massive and non-repairable rotator cuff tear.

The patient's joint must be anatomically suited to receive the selected implants and a functional deltoid muscle is necessary to use the device.

The humeral stem of the Humeris Cementless Shoulder is intended for cementless use only. The humeral stem of the Humeris Cemented Shoulder is intended for cemented use only. The Humeris Shoulder System are intended for cemented use only. The glenoid baseplate component is intended for cementless use with the addition of screws for fixation.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary
Applicant/Sponsor:FX Shoulder USA, Inc.13465 Midway Road, Suite 101Dallas, Texas 75244Establishment Registration No: 3014128390
Manufacturer:FX Solutions1663 Rue de MajornasViriat, France 01440Establishment Registration No: 3009532798
Contact Person:Kathy Trier, Ph.D.VP Regulatory, Clinical, Quality, Compliancektrier@fxshoulder.com800.280.0775
Date:July 15, 2020
Proprietary Name:TiN Coated Humeral Heads
Common Name:Anatomic Shoulder Prosthesis
Product Code(s):KWT, PHX
Classification Name:21 CFR 888.3650: shoulder joint metal/polymernon-constrained cemented prosthesis - Class II
Substantially EquivalentDevices:Primary Predicate:Humelock II Cemented Shoulder System (K123814)Predicate:Humeris™ Shoulder System (K163669)Reference Device:LINK® Endo-Model® Knee System with PorEx®;LINK® Sled Knee System with PorEx® (K152431)

Device Description

The TiN Coated Humeral Head is a new component for Humelock II total shoulder replacement systems. The TiN Coated Humeral Head has a titanium nitride (TiN) coating, which is applied to the predicate humeral heads made of cobalt chromium molybdenum (CoCr). Compatible components for use with the TiN coated humeral heads

{5}------------------------------------------------

to complete the total shoulder replacement construct are the same as those previously cleared compatible components for use with the CoCr humeral heads in the respective predicate devices, including the primary predicate (K123814) and K163669 for the Humeris Shoulder System.

Intended Use / Indications

When used in the Humelock II Cemented Shoulder System: The Humelock II Cemented Shoulder System is indicated for use in total and hemishoulder replacement to treat:

    1. Proximal humeral fractures
    1. A severely painful and/or disabled joint resulting from osteoarthritis, traumatic arthritis or rheumatoid arthritis;
    1. Other difficult clinical problems where shoulder arthrodesis or resection arthroplasty are not acceptable (e.g. revision of a previously implanted primary component, a humeral plate or a humeral nail).

The humeral stem and glenoid components of the Humelock II Cemented Shoulder System are intended for cemented use only.

TiN coated Heads are not intended for hemi-shoulder arthroplasty

When used with the Humeris™ Shoulder System: In an anatomic shoulder configuration, the Humeris Shoulder System is indicated for use in total and hemi-shoulder replacement to treat:

  • A severely painful and/or disabled joint resulting from osteoarthritis or rheumatoid arthritis;

  • Other difficult clinical problems where shoulder arthrodesis or resection arthroplasty are not acceptable (e.g. revision of a previously implanted primary component, a humeral plate or a humeral nail).

TiN coated Heads are not intended for hemi-shoulder arthroplasty.

In a reverse shoulder configuration, the Humeris Shoulder is indicated for primary or revision total shoulder arthroplasty for the relief of pain and to improve function in patients with a massive and non-repairable rotator cuff tear.

The patient's joint must be anatomically and structurally suited to receive the selected implants and a functional deltoid muscle is necessary to use the device.

The humeral stem of the Humeris Cementless Shoulder is intended for cementless use only. The humeral stem of the Humeris Cemented Shoulder is intended for cemented use only. The glenoid components of the Humeris Shoulder System are intended for cemented use only. The glenoid baseplate component is intended for cementless use with the addition of screws for fixation.

{6}------------------------------------------------

Summary of Technologies / Substantial Equivalence

The TiN Coated Humeral Head is substantially equivalent to the primary predicate in that it is identical to the primary predicate on indications, design, dimensions, packaging, single use, sterilization, shelf life, pyrogen testing, biocompatibility, compatible components, instrumentation and surgical technique. It is identical to the primary predicate material CoCr. The change, subject of this 510(k), is to add the titanium nitride (TiN) coating to the surface of the humeral head, which is substantially equivalent to the technology of coating on the reference device based upon the characterization and testing in accordance with the same scientific methods and standards. Differences between the subject device and the predicate devices do not raise new questions of safety and effectiveness.

Non-Clinical Testing

Biocompatibility testing has been completed in accordance with FDA Guidance titled. "Use of International Standard ISO 10993-1. "Biological evaluation of medical devices -Part 1: Evaluation and testing within risk management process" (September 2016). Nonclinical biocompatibility testing is the same as the reference device and includes Acute System Toxicity Study; Cytotoxicity; GC/MS Fingerprint Study; Irritation Study; Sensitization Study. Characterization of the TiN coating demonstrates substantial equivalence of the subject device to the coating of the reference device - Bone Implantation Study, 28 Day Muscle Implantation Study; 90 Day Muscle Implantation Study are referenced for the subject device. Wear testing under worst case loading and worst case environment with analysis of surface roughness and both UHMWPE and metal particle analysis have been completed.

Clinical Testing

Clinical testing was not necessary to determine substantial equivalence of the TiN Coated Humeral Head to the predicate devices.

Conclusion

The subject device is identical to the primary predicate with the only modification of added TiN coating. The evidence reviewed to demonstrate substantial equivalence includes risk analysis, design controls, biocompatibility, wear properties of the TiN coating, and verification and validation activities to demonstrate that the TiN coating does not increase risk and does not raise new questions of safety and effectiveness of the TiN Coated Humeral Head.