K Number
K191612

Validate with FDA (Live)

Date Cleared
2020-03-20

(277 days)

Product Code
Regulation Number
878.4400
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

These AEM Instruments incorporate the use of AEM technology and are intended for use in delivering monopolar electrosurgical energy during laparoscopic procedures only. AEM Instruments are intended for use with the AEM Monitoring System and electrosurgical generators having compatibility with the AEM Monitor. Scissors Inserts are intended for use on soft tissue only.

Device Description

Encision AEM® Monopolar Laparoscopic Instruments and Accessories incorporate the use of AEM technology and are intended for use in delivering monopolar electrosurgical energy during laparoscopic procedures only. Encision AEM® Monopolar Laparoscopic Instruments and Accessories are intended for use with the AEM Monitoring System and electrosurgical generators having compatibility with the AEM Monitor. These Encision AEM® Monopolar Laparoscopic Instruments and Accessories include reusable handles and reusable & disposable insert electrodes plus disposable sheath. Encision AEM® Monopolar Laparoscopic Instruments and Accessories, in conjunction with an AEM Monitor properly connected to the electrosurgical generator (ESU), continuously monitor and dynamically manage "stray energy" (insulation failure and capacitive coupling), which are likely out of the surgeon's field of view.

AI/ML Overview

This document does not contain explicit acceptance criteria and corresponding performance data tables for a diagnostic device, nor does it describe a study that directly proves a device meets such criteria in terms of diagnostic effectiveness (e.g., sensitivity, specificity, accuracy).

Instead, this document is an FDA 510(k) premarket notification for Encision AEM® Monopolar Laparoscopic Instruments and Accessories. The purpose of a 510(k) submission is to demonstrate that a new device is "substantially equivalent" to a legally marketed predicate device, meaning it is as safe and effective.

Therefore, the information provided focuses on demonstrating substantial equivalence rather than fulfilling specific performance acceptance criteria for diagnostic output.

However, I can extract information related to the device's characteristics and the studies performed to support its
FDA clearance, framed within the context of substantial equivalence.

Here's an analysis based on the provided document:


1. Table of Acceptance Criteria and Reported Device Performance

As this is a 510(k) for electrosurgical instruments, the "acceptance criteria" are primarily related to safety, functionality, and equivalence to predicate devices, rather than diagnostic performance metrics like sensitivity or specificity.

Acceptance Criteria Category (Implied by Substantial Equivalence Demonstration)Reported Device Performance / Evaluation
Intended UseIdentical to predicate devices: "delivering monopolar electrosurgical energy during laparoscopic procedures only."
Operating PrincipleIdentical to predicate devices.
MaterialsSubstantially equivalent. Material changes (e.g., lubricant) are equivalent to predicate devices and do not add new biocompatibility concerns. Tested for biocompatibility per ANSI/AAMI/ISO 10993-1:2009.
UseNo changes to use. Prescription use only, identical to predicate devices.
TechnologyOnly minor changes to technology. Design changes have not affected the fundamental technology of the AEM Instruments.
SterilizationNo changes to sterilization methods. Disposable components E-beam sterilized (10⁻⁶ SAL) per ANSI/AAMI/ISO 11137-1:2006. Reusable components steam sterilized (10⁻⁶ SAL) per ANSI/AAMI/ISO 17665-1:2006.
Storage & TransportationNo change.
Performance (Functionality & Safety)Demonstrated through bench testing to perform as intended and as well as predicate devices under same/similar conditions, including typical procedure settings. The device's requirements for electrode function within the AEM Monitoring System have not changed.
Electromagnetic Compatibility (EMC)AEM® Monitoring System (with these electrodes as an accessory) tested per ANSI/AAMI/IEC 60601-1-2:2014.
Electrical SafetyAEM® Monitoring System (with these electrodes as an accessory) tested per ANSI/AAMI/IEC 60601-2-2:2017.
LabelingAssociated labeling in accordance with ANSI/AAMI/ISO 15223-1:2016.
Risk ManagementActivities conducted per ANSI/AAMI/ISO 14971:2007.
Shelf LifeDisposable Scissors and Sheaths increased from 1 year to 5 years (in sterile packaging), supported by testing for sterility and packaging integrity.

2. Sample Size Used for the Test Set and Data Provenance

The document states, "All testing required per design control procedures was conducted using bench testing." It also mentions "No animal or clinical testing is required."

  • Test Set Sample Size: Not explicitly stated as a number of "cases" or "patients" since this is bench testing of physical devices. The "sample" would be the number of devices or components tested. This detail is not provided.
  • Data Provenance: Bench testing data, originating from the manufacturer's internal studies. It is retrospective in the sense that it's evaluating the newly designed devices against established performance of predicate devices and recognized standards. There's no mention of country of origin for patients, as no patient data was used.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

  • This question is not applicable (N/A) for this submission. The "ground truth" for electrosurgical instrument performance is established by recognized consensus standards, safety regulations, and functional specifications, rather than expert interpretation of medical images or clinical outcomes. The device's performance is measured objectively against these established engineering and safety benchmarks. There are no "experts" establishing a "ground truth" in the clinical diagnostic sense.

4. Adjudication Method for the Test Set

  • This question is not applicable (N/A). Adjudication methods (like 2+1, 3+1) are used in clinical studies, particularly for diagnostic imaging or outcome evaluation where there might be disagreement among human readers. Since this submission relies on bench testing and engineering standards, there is no human adjudication process involved for the "test set" performance.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, If So, What was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

  • This question is not applicable (N/A). This device is an electrosurgical instrument, not an AI-powered diagnostic tool. Therefore, no MRMC study or evaluation of AI assistance for human readers was performed or would be relevant.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

  • This question is not applicable (N/A). This device is a medical instrument used directly in surgery, not a standalone algorithm. Its function is to deliver electrosurgical energy safely, monitored by an AEM Monitoring System. There is no "algorithm-only" performance to evaluate in this context.

7. The Type of Ground Truth Used

The "ground truth" for the device's safety and effectiveness is largely defined by:

  • Engineering Specifications: Internal design requirements for the device's mechanical, electrical, and material properties.
  • Recognized Consensus Standards: Compliance with FDA-recognized standards for sterilization (e.g., ANSI/AAMI/ISO 11137-1, 17665-1), biocompatibility (ANSI/AAMI/ISO 10993-1), electrical safety (ANSI/AAMI/IEC 60601-2-2), electromagnetic compatibility (ANSI/AAMI/IEC 60601-1-2), packaging (ANSI/AAMI/ISO 11607-1), and reprocessing instructions (ANSI/AAMI/ISO 17664).
  • Safety and Performance of Established Predicate Devices: The primary ground truth for a 510(k) is demonstrating that the new device performs "as well as" or is "as safe and effective as" the predicate device(s).

8. The Sample Size for the Training Set

  • This question is not applicable (N/A). This is not an AI/machine learning device that requires a "training set."

9. How the Ground Truth for the Training Set was Established

  • This question is not applicable (N/A). As there is no training set for an AI algorithm, there is no ground truth established for a training set.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left side of the image is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

March 20, 2020

Encision Incorporated % Charles Hart Principle Consultant HART Consulting LLC 615 Reid Place Castle Rock. Colorado 80108

Re: K191612

Trade/Device Name: Encision AEM Monopolar Laparoscopic Instruments and Accessories Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting & Coagulation Devices and Accessories Regulatory Class: Class II Product Code: GEI Dated: February 18, 2020 Received: February 21, 2020

Dear Charles Hart:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K191612

Device Name

Encision AEM® (Active Electrode Monitoring) Monopolar Laparoscopic Instruments and Accessories

Indications for Use (Describe)

These AEM Instruments incorporate the use of AEM technology and are intended for use in delivering monopolar electrosurgical energy during laparoscopic procedures only. AEM Instruments are intended for use with the AEM Monitoring System and electrosurgical generators having compatibility with the AEM Monitor. Scissors Inserts are intended for use on soft tissue only.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the text "K191612" in a simple, sans-serif font. The text is horizontally oriented and appears to be a code or identifier. The characters are uniformly sized and spaced, making the text easily readable.

Image /page/3/Picture/1 description: The image shows the logo for Encision. The logo features the word "ENCISION" in green, with a stylized orange circle to the right of the word. Below the word "ENCISION" is the phrase "DELIVERING ENERGY SAFELY" in a smaller font size.

K191612 510(k) Summary

Device: Encision AEM® (Active Electrode Monitoring) Monopolar Laparoscopic Instruments and Accessories

Owner: Encision Inc.

  • Contact: Greg Trudel President & CEO 6797 Winchester Circle Boulder, CO, 80301, USA gtrudel@encision.com 303-444-2600
    Date: 20 Mar 2020

Subject Device:

Trade NameSee table below
Common NameMonopolar laparoscopic accessory
Regulation NameElectrosurgical, Cutting & Coagulation,Accessories
Regulation Number21 CFR 878.4400
Product CodeGEI
Device Class2
Review PanelGeneral and Plastic Surgery
Prior SubmissionsNo prior submissions for the subject device

Primary Predicate Devices:

Trade NameSee table below
Regulation NameElectrosurgical, Cutting & Coagulation,Accessories
ManufacturerEncision Inc.
510(k) NumberK912780

Secondary Predicate Devices:

Trade NameES03XX Series
Regulation NameElectrosurgical, Cutting & Coagulation,Accessories
ManufacturerEncision Inc.
510(k) NumberK091074

{4}------------------------------------------------

ENCISIGN

Device Description:

Encision AEM® Monopolar Laparoscopic Instruments and Accessories incorporate the use of AEM technology and are intended for use in delivering monopolar electrosurgical energy during laparoscopic procedures only. Encision AEM® Monopolar Laparoscopic Instruments and Accessories are intended for use with the AEM Monitoring System and electrosurgical generators having compatibility with the AEM Monitor. These Encision AEM® Monopolar Laparoscopic Instruments and Accessories include reusable handles and reusable & disposable insert electrodes plus disposable sheath. Encision AEM® Monopolar Laparoscopic Instruments and Accessories, in conjunction with an AEM Monitor properly connected to the electrosurgical generator (ESU), continuously monitor and dynamically manage "stray energy" (insulation failure and capacitive coupling), which are likely out of the surgeon's field of view.

Model Numbers and Trade Names:

Model Number(subject device)Trade NameProduct FamilyPredicate Device(K912780)
ES0101AEM e·Edge® CurvedScissors, 1/2"InsertDisposable ScissorsES0101
ES0102AEM e·Edge® CurvedScissors, 3/4" InsertDisposable ScissorsES0102
ES0110AEM e-Edge® Hook ScissorsInsertDisposable ScissorsES0110
ES0120AEM e·Edge® Scissors, 3/4",Low Profile InsertDisposable ScissorsES0120
ES8000 SeriesAEM enTouch® HandlesAEM® Handles(Reusable)ES5700
ES8200 SeriesAEM enTouch® Handles withIndexing and LockingAEM® Handles(Reusable)ES5700
ES0004AEM® Right Angle DissectorInsertReusable InsertsES0004
ES0008AEM® Tapered Right AngleDissector InsertReusable InsertsES0008
ES0009AEM® Blunt Nose GrasperInsertReusable InsertsES0009
ES0011AEM® Short Right AngleDissector InsertReusable InsertsES0011
ES0012AEM® Bottle Nose GrasperInsertReusable InsertsES0012
ES0013AEM® Bullet Nose GrasperInsertReusable InsertsES0013
ES0014AEM® Fine Tooth FenestratedGrasper InsertReusable InsertsES0014
ES0501AEM® Curved MarylandDissector InsertReusable InsertsES0501
ES0506AEM® 90° Grasper InsertReusable InsertsES0506
ES0507AEM® Fenestrated GraspingForceps InsertReusable InsertsES0507
ES0508AEM® Round Nose GrasperInsertReusable InsertsES0508
Model Number(subject device)Trade NameProduct FamilyPredicate Device(K912780)
ES0509AEM® Pointed Nose Grasper InsertReusable InsertsES0509
ES0510AEM® Petelin Dissector InsertReusable InsertsES0510
ES0511AEM® Dolphin Nose Grasper InsertReusable InsertsES0511
ES0512AEM® Long Dolphin Nose Grasper InsertReusable InsertsES0512
ES0513AEM® Straight Dissector InsertReusable InsertsES0513
ES0514AEM® Standard Grasper InsertReusable InsertsES0514
ES0521AEM® Bowel Grasper InsertReusable InsertsES0521
ES0522AEM® Fenestrated Bowel Grasper InsertReusable InsertsES0522
ES0526AEM® Tapered Maryland Dissector InsertReusable InsertsES0526
ES0533AEM® Strong Curved Maryland, 7-8mm InsertReusable InsertsES0533
ES0535AEM® Endo Cinch Extreme Atraumatic Serrated InsertReusable InsertsES0535
ES0537AEM® Wave Grasper InsertReusable InsertsES0537
ES0538AEM® Mixter Clamp, 90° Long InsertReusable InsertsES0538
ES0541AEM® Kelly Forceps InsertReusable InsertsES0541
ES0543AEM® Maxi Grasper InsertReusable InsertsES0543
ES0547AEM® Maryland Dissector, Diamond Serrations InsertReusable InsertsES0547
ES0548AEM® Dissecting Forceps, Right Angled InsertReusable InsertsES0548
ES0549AEM® Maryland Dissector, Aggressive InsertReusable InsertsES0549
ES0552AEM® Dissecting Forceps, Right Angled, Cross Serrated InsertReusable InsertsES0552
ES0553AEM® Straight Micro-Grasper InsertReusable InsertsES0553
ES0557AEM® Micro-Fenestrated Grasper InsertReusable InsertsES0557
ES0558AEM® Tapered Micro-Fenestrated Grasper InsertReusable InsertsES0558
ES0559AEM® Right Angle Dissector, 7mm, Diagonal Serrations InsertReusable InsertsES0559
ES0565AEM® Maryland Dissector with 90° Tooth InsertReusable InsertsES0565
ES0566AEM® Beveled Maryland Dissector, Tapered InsertReusable InsertsES0566
ES0570AEM® Insert, Atraumatic Grasper (Single Action)Reusable InsertsES0570
Model Number(subject device)Trade NameProduct FamilyPredicate Device(K912780)
ES0571AEM® Insert, Traumatic GrasperReusable InsertsES0571
ES0573AEM® Insert, Traumatic Grasper (Single Action)Reusable InsertsES0573
ES0574AEM® Insert, Fenestrated Bowel GrasperReusable InsertsES0574
ES0001AEM® Curved Scissors, 1/2" InsertReusable InsertsES0001
ES0002AEM® Curved Scissors, 3/4" InsertReusable InsertsES0002
ES0010AEM® Hook Scissors InsertReusable InsertsES0010
ES0150 SeriesAEM® Disposable SheathDisposable SheathsES0150

{5}------------------------------------------------

ENCISI&N

{6}------------------------------------------------

Indications for Use:

These AEM Instruments incorporate the use of AEM technology and are intended for use in delivering monopolar electrosurgical energy during laparoscopic procedures only.

AEM Instruments are intended for use with the AEM Monitoring System and electrosurgical generators having compatibility with the AEM Monitor.

Scissors Inserts are intended for use on soft tissue only.

Sterility/Packaging:

The Disposable Scissors and Disposable Sheaths are supplied sterile per ANSVAAMI/ISO 11137-1:2006 and ANSI/AAMI/ISO 11607-1:2006. The products are sterilized using E-beam radiation sterilization to achieve 106 SAL. Product is packaged in a Tyvek mylar pouch (10 pouches in a box) with a 5 year shelf life in the sterile packaging.

The AEM® Handles and Reusable Inserts are reusable and steam sterilized for repeated use per ANSI/AAMI/ISO 17664:2017 and ANSI/AAMI/ISO 17665-1:2006. These instruments are cleaned using manual, automatic or combined manual/automatic cleaning methods. They are sterilized through Prevac steam sterilization using FDA-cleared sterile wraps to achieve 10th SAL. These products are shipped non-sterile and require sterilization before use.

Biocompatibility:

All patient contact materials are biocompatible. Material testing demonstrates conformance with ANSI/AAMI/ISO 10993-1:2009, 4th edition. Gap analysis 05927 demonstrates that testing to a previous revision of the standard is consistent with the current recognized revision of ISO 10993-1.

Standards:

The AEM® Monopolar Laparoscopic Instruments and Accessories have been tested to show conformity to the following FDA recognized standards:

{7}------------------------------------------------

ENCISI&N®

TopicStandard UsedApplication
E-beam sterilizationANSI/AAMI/ISO 11137-1:2006Sterilization of Health Care Products – Radiation – Part 1: Requirements for Development, Validation and Routine Control of a Sterilization Process for Medical DevicesDisposable Scissors and Disposable Sheaths are sterilized to 10-6 SAL using E-beam sterilization in accordance with this standard.
Sterile packagingANSI/AAMI/ISO 11607-1:2006Packaging for Terminally Sterilized Medical Devices – Part 1: Requirements for Materials, Sterile Barrier Systems and Packaging SystemsDisposable Scissors and Disposable Sheaths are packaged with a sterile barrier system in accordance with this standard.
Reprocessing – cleaning and sterilizationANSI/AAMI/ISO 17664:2017Processing of Health Care Products – Information to be Provided by the Medical Device Manufacturer for the Processing of Medical DevicesAEM® Handles and Reusable Inserts are reprocessed for reuse in accordance with this standard
Reprocessing – steam sterilizationANSI/AAMI/ISO 17665-1:2006Sterilization of Health Care Products – Moist Heat - Part 1: Requirements for the Development, Validation and Routing Control of a Sterilization Process for Medical DevicesAEM® Handles and Reusable Inserts are reprocessed to 10-6 SAL using steam sterilization in accordance with this standard.
BiocompatibilityANSI/AAMI/ISO 10993-1:2009Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing within a Risk Management ProcessAll patient-contact materials for the AEM® Monopolar Laparoscopic Instruments and Accessories have been tested for biocompatibility in accordance with this standard.
Electromagnetic compatibilityANSI/AAMI/IEC 60601-1-2:2014Medical Electrical Equipment – Part 1-6: General Requirements for Basic Safety and Essential Performance. Collateral Standard. Electromagnetic Compatibility. Requirements and TestsThe AEM® Monitoring System in which these electrodes are an accessory has been tested for electromagnetic compatibility in accordance with this standard.
Electrical safetyANSI/AAMI/IEC 60601-2-2:2017Medical Electrical Equipment – Part 2-2: Particular Requirements for the Basic Safety and Essential Performance of High Frequency Surgical Equipment and High Frequency Surgical AccessoriesThe AEM® Monitoring System in which these electrodes are an accessory has been tested for electrical safety in accordance with this standard.
LabelingANSI/AAMI/ISO 15223-1:2016Medical Devices – Symbols to be Used with Medical Device Labels, Labelling and Information to be Supplied – Part 1: General RequirementsThe AEM® Monopolar Laparoscopic Instruments and Accessories have associated labeling in accordance with this standard.
TopicStandard UsedApplication
Risk managementANSI/AAMI/ISO 14971:2007Medical Devices – Application of RiskManagement to Medical DevicesRisk management activities for theAEM® Monopolar LaparoscopicInstruments and Accessories havebeen conducted in accordance withthis standard.

{8}------------------------------------------------

Technological Characteristics:

The technological characteristics of the subject device are identical to the predicate devices with regards to fundamental design, materials, operating principle, use, compatibility, packaging and sterilization. The accessories have both 35cm and 45cm configurations. The changes over time to the predicate devices do not introduce new concerns of safety or effectiveness. The differences between the subject and predicate device are:

  • . Use of Metrilube Instrument Lubricant for Disposable and Reusable Scissors for effective cutting performance. The lubricant is a patient-contact material and is biocompatible.
  • . Change from a removable trigger to a permanent trigger on the AEM® Handles for improved handle performance. The fundamental technology and operating principle is identical to the predicate devices.
  • . Shelf life of Disposable Scissors and Disposable Sheaths has increased from 1 year to 5 years in the sterile packaging.

Substantial Equivalence:

The subject device is substantially equivalent to the predicate device, as there is no change to the Intended Use, Operating Principle, Patient Contact Materials, Fundamental Technology, Sterilization and Performance.

TopicSubject Device (this submission) compared to Predicate Devices
(K912780)
Indications for UseNo change to Intended Use.
Operating PrincipleNo change to Operating Principle.
MaterialsSubstantially equivalent. Material changes are equivalent to predicatedevices and do not add new biocompatibility concerns.
UseNo changes to Use. The AEM Instruments are prescription use only.
TechnologyOnly minor changes to Technology. The design changes made to theproducts have not affected the fundamental technology of the AEMInstruments.
SterilizationNo changes to Sterilization. Disposable Scissors and Disposable Sheaths areE-beam sterilized and supplied sterile. AEM Handles and Reusable Insertsare reusable and supplied non-sterile.
Storage &TransportationNo change to Storage and Transportation.
PerformanceNo change to Performance.

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the word "ENCISION" in a bold, sans-serif font. The letters are all capitalized and evenly spaced. To the right of the letters is a stylized graphic that resembles a crescent shape with a curved line running through it.

Bench Testing:

All testing required per design control procedures was conducted using bench testing. Performance analysis has shown the Encision AEM® Monopolar Laparoscopic reusable handpieces and Encision AEM® Monopolar Laparoscopic reusable & disposable insert electrodes plus disposable sheath to perform as intended and as well as the predicate devices under the same and/or similar conditions. These devices have been tested with settings set equal to those seen in typical procedures where these devices would be indicated for use. The requirements for electrodes function in the AEM Monitoring System have not changed from the original 510(k) submissions. The tests conducted show that the changes since the original 510(k) submissions demonstrate conformity to the original internal and regulatory requirements. No differences have been observed between the performance of the predicate devices and the device subject of this 510(k) Notification submission. No animal or clinical testing is required.

Conclusion:

The AEM® Monopolar Laparoscopic Instruments and Accessories are substantially equivalent to the predicate devices. The design changes do not introduce new safety or effectiveness concerns and do not alter the fundamental technology or operating principle of the devices. The Intended Use is identical. Bench testing data has confirmed substantial equivalence to the predicate devices.

§ 878.4400 Electrosurgical cutting and coagulation device and accessories.

(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.