K Number
K191571

Validate with FDA (Live)

Date Cleared
2020-02-06

(237 days)

Product Code
Regulation Number
878.4630
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The UV Radiation Treatment System models YK-6000A-T, YK-6000B, andYK-6000B-T, are intended for use under the direction of a physician, for the treatment of vitilizo, psoriasis, and eczema. The devices are intended for treatment of these conditions on Fitzpatrick Skin Types (I-VI).

Device Description

The proposed device, UV Radiation Treatment System is a portable medical device that consists of tubes, irradiator, control circuit. It is a therapeutic product under the direction of a physician for individuals who require ultraviolet radiation for diagnosed skin disorders. It is for the partial treatment. It has four models: YK-6000A,YK-6000A-T,YK-6000B,YK-6000B-T,in which the model YK-6000A and YK-6000A-T emit UVA light and the model YK-6000B and YK-6000B-T emit UVB light.

This product has the features are as follows:

A. Use Philips special UV lamp as light source, with high stability and long service life:

B. Small bulk, lightweight, operation simply, patients can take with themselves;

C. There is glisten board inner radiation unit, which can raise radiant efficiency.

The UV Radiation Treatment System only can be used in Hospital, Clinic, Medical Center, Private Medical Practice, or Other Professional Medical Environments under direction of physician.

AI/ML Overview

The provided document is a 510(k) summary for a UV Radiation Treatment System and specifies that no clinical study was performed to prove the device meets acceptance criteria. The submission relies solely on non-clinical (lab bench) testing and comparison to a predicate device. Therefore, many of the requested sections (2-5, 7-9) cannot be answered with information from this document because there was no study involving patient data, ground truth, or expert review.

Here's a breakdown of the available information:

1. A table of acceptance criteria and the reported device performance

The document does not explicitly state "acceptance criteria" in the traditional sense of performance metrics against a defined standard (e.g., sensitivity, specificity for an AI device). Instead, it presents a comparison of the proposed device's characteristics against a predicate device, focusing on substantial equivalence. The "performance" is demonstrated through compliance with various safety and electrical standards via non-clinical bench testing.

Characteristic / Acceptance Criterion (Implicitly "Substantially Equivalent")Proposed Device PerformancePredicate Device (K181805) Performance
Product Code (FTC)FTCFTC
Regulation No. (21 CFR 878.4630)21 CFR 878.463021 CFR 878.4630
Class (II)IIII
Intended UseTreatment of vitiligo, psoriasis, and eczema under direction of a physician, for Fitzpatrick Skin Types (I-VI).Treatment of psoriasis, vitiligo, and atopic dermatitis (eczema) under direction of a physician, for all skin types (I-VI). Can be used in hospitals, clinics, and households.
Prescriptive (Yes)YesYes
Mode of operationPortableHandheld
Shell material (ABS)ABSABS
Treatment AreaPartial treatment excluding eyesPartial treatment excluding eyes
Treatment Time5~20 min0~30min
UV spectral outputUVA or UVBUVA or UVB
Lamp configuration1 (9W UV tube)1 or 2 (9W UV tube)
Power SourceAC outletAC outlet or DC jack
Wavelength range (nm)UVA: 320400; UVB: 300320UVA: 350400; UVB: 310315
IPX – Rating / water resistanceIPX0IP22
Application EnvironmentHospital, Clinic, Medical Center, Private Medical Practice, or Other Professional Medical Environments under direction of physicianHospital, Clinic, Medical Center, Private Medical Practice, or Other Professional Medical Environments, Household use under direction of physician.
Electrical Safety/PerformanceComply with IEC60601-1 and IEC 60601-2-57Comply with IEC60601-1 and IEC 60601-2-57
Home UseNot appliedComply with IEC60601-1-11
SterileN/AN/A
Single UseNoNo
EMCComply with IEC60601-1-2Comply with IEC60601-1-2
BiocompatibilityCytotoxicity (ISO 10993-5:2009), Sensitization (ISO 10993-10:2010), Irritation (ISO 10993-10:2010), Pyrogen (ISO 10993-11:2017), Acute Systemic Toxicity (ISO 10993-11:2017)Under the conditions of the study, Comply with the requirements
Label and LabelingConforms to FDA Regulatory RequirementsConforms to FDA Regulatory Requirements

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • No test set was used. As explicitly stated in Section 8.0, "No clinical study is included in this submission." The evaluation for substantial equivalence was based on non-clinical bench testing and comparison to a predicate device.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

  • Not applicable. No clinical study was performed, and therefore no ground truth needed to be established by experts for a test set.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Not applicable. No clinical study was performed.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not applicable. This device is a UV radiation treatment system, not an AI diagnostic or assistance tool. No MRMC study was performed.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Not applicable. This device is a physical UV treatment system, not an algorithm.

7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

  • Not applicable. No clinical study involving patient data and ground truth was conducted. The "ground truth" for the non-clinical tests would be the established performance specifications and regulatory standards (e.g., measured electrical safety, biocompatibility results).

8. The sample size for the training set

  • Not applicable. As a physical medical device, there is no "training set" in the context of an AI/ML algorithm.

9. How the ground truth for the training set was established

  • Not applicable. There is no training set for this device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Xuzhou Yongkang Electronic Science Technology Co., Ltd % Boyle Wang Official Correspondent Shanghai Truthful Information Technology Co., Ltd. RM. 608, No. 738, Shangcheng Rd., Pudong Shanghai, 200120 Cn

Re: K191571

Trade/Device Name: UV Radiation Treatment System Regulation Number: 21 CFR 878.4630 Regulation Name: Ultraviolet Lamp For Dermatologic Disorders Regulatory Class: Class II Product Code: FTC Dated: June 14, 2019 Received: June 14, 2019

Dear Boyle Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jessica Mavadia-Shukla Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K191571

Device Name UV Radiation Treatment System

Indications for Use (Describe)

The UV Radiation Treatment System models YK-6000A-T, YK-6000B, andYK-6000B-T, are intended for use under the direction of a physician, for the treatment of vitilizo, psoriasis, and eczema. The devices are intended for treatment of these conditions on Fitzpatrick Skin Types (I-VI).

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Tab #06 510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR 807.92.

1.0 Submitter's information

Name: Xuzhou Yongkang Electronic Science Technology Co., Ltd Address: 4F Building C8, 40 Jingshan Road, Economic and Technological Development Zone, Xuzhou, Jiangsu,221000,China Tel: 86-516-87892766-601 Fax: 86-516-87892766-606 Contact: Kai Li Date of Preparation: Feb.5,2020

Designated Submission Correspondent

Mr. Boyle Wang Shanghai Truthful Information Technology Co., Ltd. Address: Room 608, No. 738 Shangcheng Rd., Pudong Shanghai, 200120 China Tel: +86-21-50313932 Email: Info@truthful.com.cn

2.0 Device information

Trade name: UV Radiation Treatment System Common name: Ultraviolet Lamp for Dermatologic Disorders Classification name: Light, Ultraviolet, Dermatological Model(s): YK-6000A,YK-6000A-T,YK-6000B,YK-6000B-T

3.0 Classification

Production code: FTC Regulation number: 21 CFR 878.4630 Classification: Class II Review Panel: General& Plastic Surgery

4.0 Predicate device information

Manufacturer: Xuzhou Kernel Medical Equipment Co., LTD. UV Phototherapy Device: 510(k) number: K181805

{4}------------------------------------------------

5.0 Indication for Use Statement

The UV Radiation Treatment System models YK-6000A, YK-6000A-T, YK-6000B, andYK-6000B-T, are intended for use under the direction of a physician, for the treatment of vitiligo, psoriasis, and eczema. The devices are intended for treatment of these conditions on Fitzpatrick Skin Types (I-VI).

6.0 Device Description

The proposed device, UV Radiation Treatment System is a portable medical device that consists of tubes, irradiator, control circuit. It is a therapeutic product under the direction of a physician for individuals who require ultraviolet radiation for diagnosed skin disorders. It is for the partial treatment. It has four models: YK-6000A,YK-6000A-T,YK-6000B,YK-6000B-T,in which the model YK-6000A and YK-6000A-T emit UVA light and the model YK-6000B and YK-6000B-T emit UVB light.

This product has the features are as follows:

A. Use Philips special UV lamp as light source, with high stability and long service life:

B. Small bulk, lightweight, operation simply, patients can take with themselves;

C. There is glisten board inner radiation unit, which can raise radiant efficiency.

The UV Radiation Treatment System only can be used in Hospital, Clinic, Medical Center, Private Medical Practice, or Other Professional Medical Environments under direction of physician.

7.0 Non-Clinical Test Conclusion

Non clinical tests were conducted by lab bench testing to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:

IEC 60601-1:2012, Medical electrical equipment-Part 1: General requirements for basic safety, and essential performance.

IEC 60601-2-57: 2011 Medical electrical equipment Part2-57: Particular requirements for the basic safety and essential performance of non-laser light source equipment intended for therapeutic, diagnostic, monitoring and cosmetic/ aesthetic use.

IEC 60601-1-2:2014, Medical electrical equipment -Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests.

IEC 62471:2006, Photobiological safety of lamps and lamp systems

ISO 10993-5:2009 Biological evaluation of medical devices- Part 5: Test for in

{5}------------------------------------------------

vitro cytotoxicity

ISO 10993-10:2010 Biological evaluation of medical devices- Part 10: Test for irritation and delayed-type hypersensitivity

ISO 10993-11:2017 Biological evaluation of medical devices- Part 11: Test for systemic toxicity

8.0 Clinical Test Conclusion

No clinical study is included in this submission.

9.0 Substantial Equivalence Comparison

ItemProposed devicePredicated deviceK181805Remark
Product CodeFTCFTCSE
Regulation No.21 CFR 878.463021 CFR 878.4630SE
ClassIIIISE
Intended UseThe UV RadiationTreatment Systemmodels YK-6000A,YK-6000A-T, YK-6000B,andYK-6000B-T, areintended for use underthe direction of aphysician, for thetreatment of vitiligo,psoriasis, and eczema.The devices are intendedfor treatment of theseconditions on FitzpatrickSkin Types (I-VI).The UV Phototherapyis intended for use, byor under the directionof a physician for thetreatment of psoriasis,vitiligo, and atopicdermatitis (eczema). Itis intended for use onall skin types (I-VI). Itcan be used inhospitals, clinics andhouseholds.SE
PrescriptiveYesYesSE
Mode of operationPortableHandheldAnalysis
Shell materialABSABSSE
Treatment AreaPartial treatmentexcluding eyesPartial treatmentexcluding eyesSE
Treatment Time5~20 min0~30minAnalysis
UV spectral outputUVA or UVBUVA or UVBSE
Lamp configuration19W UV tube1 or 29W UV tubeAnalysis
Power SourceAC outletAC outlet or DC jackAnalysis
Wavelength range(nm)UVA:320400UVB: 300320UVA: 350400UVB: 310315Analysis

Table1- Technological Characteristic Comparison

{6}------------------------------------------------

IPX – Rating / waterresistanceIPX0IP22Analysis
ApplicationEnvironmentHospital, Clinic, MedicalCenter, Private MedicalPractice, or OtherProfessional MedicalEnvironments underdirection of physicianHospital, Clinic,Medical Center,Private MedicalPractice, or OtherProfessional MedicalEnvironments,Household use underdirection of physician.Analysis
ElectricalSafety/PerformanceComply with IEC60601-1and IEC 60601-2-57Comply withIEC60601-1 and IEC60601-2-57SE
Home UseNot appliedComply with IEC60601-1-11Analysis
SterileN/AN/ASE
Single UseNoNoSE
EMCComply with IEC60601-1-2Comply with IEC60601-1-2SE
BiocompatibilityCytotoxicity (ISO10993-5:2009)Under the conditions ofthe study, Comply withthe requirementsSE
Sensitization (ISO10993-10:2010)
Irritation (ISO10993-10:2010)
Pyrogen(ISO10993-11:2017)
Acute SystemicToxicity(ISO10993-11:2017)
Label and LabelingConforms to FDARegulatoryRequirementsConforms to FDARegulatoryRequirementsSE

Analysis:

The predicate device has the models which can support household use. The proposed device is not recommended household use, it just can be used in the professional medical environments under the guidance of doctors. In addition, the proposed device is substantially equivalent to the predicate device. The differences between both devices are insignificant in terms of safety and effectiveness. Based on the nonclinical and clinical tests performed, the subject device is as safe, as effective, and performs as well as the legally marketed predicate device.

{7}------------------------------------------------

10.0 Conclusion

The conclusions drawn from the comparison and analysis above demonstrate that the proposed device is as safe, as effective, and performs as well as the legally marketed predicated device and raises no new questions of safety or effectiveness. The differences between both devices are insignificant in terms of safety and effectiveness.

§ 878.4630 Ultraviolet lamp for dermatologic disorders.

(a)
Identification. An ultraviolet lamp for dermatologic disorders is a device (including a fixture) intended to provide ultraviolet radiation of the body to photoactivate a drug in the treatment of a dermatologic disorder if the labeling of the drug intended for use with the device bears adequate directions for the device's use with that drug.(b)
Classification. Class II.