K Number
K191494

Validate with FDA (Live)

Date Cleared
2019-08-12

(68 days)

Product Code
Regulation Number
888.3070
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Facilis™ Spinal System is intended for posterior, non-cervical fixation to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, and/or lordosis); spinal tumor; pseudoarthrosis; and failed previous fusion. The Facilis™ Spinal System is to be used with autograft and/or allograft.

Device Description

The Facilis™ Spinal System is a system that is indicated for multiple types of spinal fusion procedures (please see Section IV). All components are made from titanium alloy Titanium-6 Aluminum-4 Vanadium (Ti-6AL-4V), a biocompatible material which complies with ASTM F136. The components, which are included as part of the system, include screws, rods, hooks and accessory connection components. The Facilis™ Spinal Systems are supplied "Non-Steriled" and must be sterilized before use.

AI/ML Overview

This document describes the Facilis™ Spinal System, a medical device, and its acceptance criteria as proven by a study.

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance
Mechanical TestingAll testing results passed acceptance criteria in accordance with ASTM F1717 and ASTM F1798.
Substantial Equivalence to Predicate DeviceThe device demonstrates substantial equivalence to the predicate device (Facilis™ Spinal System K161231) based on shared operating principles, fundamental scientific technology, materials, shelf life, packaging, and manufacturing process, with additional component types.

2. Sample Size Used for the Test Set and Data Provenance

The document describes mechanical testing performed for the device. The "test set" in this context refers to the physical samples of the Facilis™ Spinal System components used for these mechanical durability tests.

  • Sample Size: Not explicitly stated as a number of devices or components. It refers to "the subject Facilis Spinal System with its components and the predicate devices."
  • Data Provenance: The tests were conducted according to established ASTM standards (F1717 and F1798), implying laboratory-based testing rather than clinical data from human patients. The origin of the specific tested devices is not specified beyond being the subject device and predicate devices. This would be considered prospective data generation for the purpose of demonstrating mechanical performance.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of those Experts

This information is not applicable to a mechanical performance study. "Ground truth" established by experts typically refers to clinical diagnoses or interpretations. For mechanical testing, the "ground truth" is defined by the physical properties and performance limits set forth in the ASTM standards.

4. Adjudication Method for the Test Set

This information is not applicable to a mechanical performance study. Adjudication methods like 2+1 or 3+1 are used for clinical studies involving expert consensus on diagnoses or interpretations. For mechanical testing, the results are objectively measured against predefined criteria in the ASTM standards.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

No, an MRMC comparative effectiveness study was not done. This type of study is relevant for evaluating the impact of an AI system on human reader performance, which is not the subject of this medical device submission.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

No, a standalone performance study was not done. This submission is for a physical medical device (spinal system), not an AI algorithm.

7. The Type of Ground Truth Used

The "ground truth" for the mechanical testing was the acceptance criteria established by recognized industry standards (ASTM F1717 and ASTM F1798). These standards define the mechanical properties and performance requirements for spinal implant systems. The devices were tested to ensure they met these predefined physical and mechanical thresholds.

8. The Sample Size for the Training Set

This information is not applicable. This is a submission for a physical medical device. The concept of a "training set" is relevant for AI/machine learning models, not for traditional medical devices undergoing mechanical performance testing.

9. How the Ground Truth for the Training Set Was Established

This information is not applicable for the reasons stated above.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is a symbol representing the Department of Health & Human Services-USA, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The FDA logo is a symbol of the agency's mission to protect and promote public health through the regulation and supervision of food safety, tobacco, pharmaceuticals, medical devices, and other related products.

August 12, 2019

Baui Biotech Co., Ltd. Mr. Herman Jhan Regulatory Affairs Supervisor 6F., No. 8, Sec. 1, Zhongxing Rd., Wugu Dist. 24872 New Taipei City Taiwan (R.O.C.)

Re: K191494

Trade/Device Name: Facilis™ Spinal System Regulation Number: 21 CFR 888.3070 Regulation Name: Thoracolumbosacral pedicle screw system Regulatory Class: Class II Product Code: NKB, KWP Dated: July 30, 2019 Received: July 31, 2019

Dear Mr. Jhan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Ronald P. Jean, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K191494

Device Name Facilis™ Spinal System

Indications for Use (Describe)

The Facilis™ Spinal System is intended for posterior, non-cervical fixation to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, and/or lordosis); spinal tumor; pseudoarthrosis; and failed previous fusion. The Facilis™ Spinal System is to be used with autograft and/or allograft.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) SUMMARY

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 C.F.R. §807.92.

Preparation Date:May 28, 2019
Applicant/Sponsor:BAUI BIOTECH CO., LTD.6F., No.8, Sec.1, Zhongxing Rd., Wugu Dist., 24872New Taipei City, Taiwan(R.O.C.)
Contact Person:Herman JhanPhone: +886-2-8976-9538 #154Fax: +886-2-8976-9608Email: ra.bu@baui.com.tw
Proprietary Name:Facilis™ Spinal System
Common Name:Pedicle Screw SystemNon-pedicle spinal fixation system
Classification Name:Thoracolumbosacral pedicle screw system (21 CFR888.3070)
Classification Identification: Class IIProduct codes:NKB, KWP
Predicate Device:Facilis™ Spinal System (K161231)

Device Description:

The Facilis™ Spinal System is a system that is indicated for multiple types of spinal fusion procedures (please see Section IV). All components are made from titanium alloy Titanium-6 Aluminum-4 Vanadium (Ti-6AL-4V), a biocompatible

{4}------------------------------------------------

material which complies with ASTM F136. The components, which are included as part of the system, include screws, rods, hooks and accessory connection components. The Facilis™ Spinal Systems are supplied "Non-Steriled" and must be sterilized before use.

The safety and effectiveness of this device has not been established when used in conjunction with bone cement or for use in patients with poor bone quality (e.g., osteoporosis, osteopenia). This device is intended only to be used with saline or radiopaque dye.

Indications for Use:

The FacilisTM Spinal System is intended for posterior, non-cervical fixation to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); spinal tumor; pseudoarthrosis; and failed previous fusion. The Facilis™ Spinal System is to be used with autograft and/or allograft.

Substantial Equivalence:

  • Predicate device name: Facilis™ Spinal System A.
  • Predicate K number: K161231 B.
  • Comparison with predicate: C.

Added two kinds screw to Facilis™ Spinal System : cannulated poly axial screw, pedicle long screw, and one transverse link and inside hex locking screw. FacilisTM Spinal System added new components, has the following similarities to the predicate device:

  • Same operating principle
  • Same fundamental scientific technology
  • Incorporate the same materials
  • Same shelf life
  • Packaged using the same materials
  • Manufactured by the same process

{5}------------------------------------------------

The modifications encompass:

  • Addition of screws type
  • Addition of transverse link type
  • Addition of inside hex locking screw ■

Performance Characteristics:

The mechanical testing performed for the subject Facilis Spinal System with its components and the predicate devices is in accordance with ASTM F1717 and ASTM F1798. Specifically, compression bending, fatigue bending, and static torsion of the subject and the predicate systems were tested. All testing results passed acceptance criteria and are equivalent to predicate devices.

    1. Conclusion:
      Based on the information provided in this submission, the Facilis™ Spinal System is substantially equivalent to the predicate Facilis™ Spinal System.

§ 888.3070 Thoracolumbosacral pedicle screw system.

(a)
Identification. (1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.
(b)
Classification. (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
“Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.