K Number
K191493

Validate with FDA (Live)

Date Cleared
2019-10-16

(133 days)

Product Code
Regulation Number
892.2050
Age Range
0 - 21
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The VMS+ 3.0 is an adjunct to existing ultrasound imaging systems and is intended to record, analyze, store and retrieve digital ultrasound images for computerized 3-dimensional image processing.

The VMS+ 3.0 is indicated for use where Left Ventricle (LV), Right Ventricle (RV), Left Atrium (LA), and Right Atrium (RA) volumes and ejection fractions are warranted or desired

Device Description

The VMS+ was cleared under 510(k) (K173810) for use in evaluation where Right Ventricle (RV). Left Ventricle (LV). Right Atrium (RA), and Left Atrium (RA) volumes and eiection fractions are warranted or desired. The modified VMS+ (VMS+ 3.0) has the same operating principle and employs the same fundamental scientific technology to that of the cleared device.

AI/ML Overview

The Ventripoint Medical System Plus (VMS+) 3.0 is a modified version of a previously cleared device (VMS+). The FDA letter and 510(k) summary do not detail a study involving AI assistance or a multi-reader multi-case (MRMC) comparative effectiveness study, nor do they specify acceptance criteria related to a general performance benchmark table or expert-based ground truth establishment as one might find for an AI/ML device.

Instead, the documentation focuses on demonstrating substantial equivalence to the predicate device (GE EchoPAC) and the reference device (VMS+) by showing that the modifications do not introduce new questions of safety or effectiveness and that the device performs as intended and as well as the predicate device(s).

Here's an breakdown of the information that is available and a note on what is not provided in the given text:

Acceptance Criteria & Device Performance:

The document broadly states that "Predefined acceptance criteria were applied during testing and were met" for specific types of nonclinical performance bench studies. However, the specific quantitative acceptance criteria for performance metrics (e.g., accuracy, precision) of volume measurements are not explicitly provided in a table within this document. It states that the device "delivers volume measurements that are equivalent in accuracy when compared with volumes obtained using the legally marketed VMS+."

Acceptance Criteria (Generic Statement)Reported Device Performance (General Statement)
Predefined acceptance criteria for nonclinical performance bench testing were applied and met.VMS+ 3.0 delivers volume measurements that are equivalent in accuracy when compared with volumes obtained using the legally marketed VMS+. Test results demonstrate the device is as safe, as effective, performs as intended and as well as the predicate device (VMS+).
Software verification and validation test reports were successful according to acceptance criteria.The verification and validation of existing and new features demonstrate that VMS+ 3.0 performs as intended, specifications conform to user needs and intended uses, and that requirements implemented can be consistently fulfilled.
Compliance with ISO 10993-1 for biocompatibility.Patient contacting components comply with ISO 10993-1.
Compliance with IEC 60601-1 for electrical safety and essential performance.Complies with IEC 60601-1.
Compliance with IEC 60601-1-2 for electromagnetic compatibility (EMC).Complies with IEC 60601-1-2.

Study Details (Based on the provided text):

  1. Sample Size used for the test set and data provenance:

    • The document mentions "nonclinical performance bench study" and "software verification and validation testing" but does not specify the sample size (e.g., number of cases or patients) used for these tests.
    • The document does not specify the country of origin of the data or whether the data was retrospective or prospective. Given it's a bench study and software V&V, it likely refers to engineered test data or data from phantoms/previous device performance.
  2. Number of experts used to establish the ground truth for the test set and qualifications of those experts:

    • The document does not specify a number of experts, their qualifications, or their role in establishing ground truth for the test set as one might expect for a clinical performance study. The ground truth appears to be based on comparison to a previously cleared device's performance benchmarks.
  3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • The document does not mention any adjudication method for a test set based on expert review.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, an MRMC comparative effectiveness study was not explicitly mentioned or detailed. The device appears to be an image analysis system, but the submission focuses on its equivalence to a previous version and predicate, not on human-AI interaction or improvement. The document explicitly states "No clinical tests were conducted to support substantial equivalence for the subject device."
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • The "nonclinical performance bench study" and "software verification and validation" would represent standalone performance assessments of the algorithm and system. The text states these tests demonstrate the device "performs as intended" and "delivers volume measurements that are equivalent in accuracy when compared with volumes obtained using the legally marketed VMS+." This implies an algorithm-only evaluation against established benchmarks from the reference device.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • The "ground truth" for the nonclinical performance bench study was the "volumes obtained using the legally marketed VMS+" (the reference device). This implies a comparison to the established performance of a prior cleared device, rather than an independent expert consensus, pathology, or outcomes data.
  7. The sample size for the training set:

    • The document describes the VMS+ 3.0 as an updated version of a previous device utilizing a "Knowledge-Based Reconstruction (KBR) algorithm." It does not specify a separate "training set" in the context of machine learning model development. For "knowledge-based" systems, the "training" often refers to the creation and refinement of the underlying rules and models based on anatomical principles and potentially a dataset of examples. The document does not provide a sample size for a training set.
  8. How the ground truth for the training set was established:

    • As this is described as a "Knowledge-Based Reconstruction (KBR) algorithm" and an update to an existing system, the concept of "ground truth for a training set" as typically applied to large-scale deep learning models is not explicitly detailed. The "ground truth" for developing such a knowledge-based system would involve meticulously defined anatomical landmarks and their relationships, likely established through anatomical studies or prior medical imaging analysis principles. The document does not provide details on how the original knowledge base was implicitly "trained" or how its "ground truth" was established.

{0}------------------------------------------------

October 16, 2019

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Ventripoint Diagnostics, Ltd. % Desmond Hirson President 2 Sheppard Avenue East, Suite 605 Toronto, Ontario M2N 5Y7 CANADA

Re: K191493

Trade/Device Name: Ventripoint Medical System Plus (VMS+) 3.0 Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: Class II Product Code: LLZ, IYN, IYO, ITX Dated: September 13, 2019 Received: September 16, 2019

Dear Desmond Hirson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For

Thalia T Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K191493

Device Name

Ventripoint Medical System Plus (VMS+) 3.0

Indications for Use (Describe)

The VMS+ 3.0 is an adjunct to existing ultrasound imaging systems and is intended to record, analyze, store and retrieve digital ultrasound images for computerized 3-dimensional image processing.

The VMS+ 3.0 is indicated for use where Left Ventricle (LV), Right Ventricle (RV), Left Atrium (LA), and Right Atrium (RA) volumes and ejection fractions are warranted or desired

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Ventripoint Diagnostics Ltd. The logo features a geometric design on the left, resembling a network of interconnected points and lines with blue, green, and orange dots. To the right of the geometric design is the company name, "Ventripoint," in a bold, blue font, with "Diagnostics Ltd." in a smaller font size underneath.

VMS+ Ventripoint Medical System 510(k) Submission

K191493

Section 3: 510(k) Summary

Date:September 13, 2019
Submitter:Ventripoint Diagnostics Ltd.2 Sheppard Avenue East, Suite 605Toronto, Ontario M2N 5Y7
Contact Person:Desmond HirsonPresidentdhirson@ventripoint.comTel: 416-848-4156
Device NameVentripoint Medical System Plus (VMS+) 3.0
Common Name:Diagnostic Ultrasound Image Analysis System
Classification:Regulatory Class: IIReview Category: Tier IIClassification Panel: Radiology
Classification Names:Picture archiving and communications system 892.2050 90 LLZUltrasonic pulsed doppler imaging system 892.1550 90 IYNUltrasonic pulsed echo imaging system 892.1560 90 IYODiagnostic ultrasonic transducer 892.1570 90 ITX

A. LEGALLY MARKETED PREDICATE DEVICE

Device NameGE EchoPAC
510(k) NumberK150085
Common Name:Picture Archiving and Communications System
Classification:Regulatory Class: II
Review Category: Tier II
Classification Panel: Radiology
Ultrasonic Pulsed Doppler Imaging System 892.2050 90 LL

B. LEGALLY MARKETED REFERENCE DEVICE

Device NameVentripoint Medical System Plus (VMS+)
510(k) NumberK173810
Common Name:Diagnostic Ultrasound Image Analysis System
Classification:Regulatory Class: II
Review Category: Tier II
Classification Panel: Radiology
Ultrasonic Pulsed Doppler Imaging System 892.1550 90-IYM

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Ventripoint Diagnostics Ltd. The logo features a geometric design on the left, resembling a network of interconnected points and lines with different colored dots. To the right of the geometric design, the company name "Ventripoint" is displayed in a blue, sans-serif font, with "Diagnostics Ltd." written in a smaller font size below.

C. DEVICE DESCRIPTION

The VMS+ was cleared under 510(k) (K173810) for use in evaluation where Right Ventricle (RV). Left Ventricle (LV). Right Atrium (RA), and Left Atrium (LA) volumes and eiection fractions are warranted or desired. The modified VMS+ (VMS+ 3.0) has the same operating principle and employs the same fundamental scientific technology to that of the cleared device.

D. INTENDED USE/INDICATIONS FOR USE

The VMS+ 3.0 system is an adjunct to existing ultrasound imaging systems and is intended to record, analyze, store and retrieve digital ultrasound images for computerized 3dimensional image processing.

The VMS+ 3.0 system is indicated for use where Left Ventricle (LV), Right Ventricle (RV), Left Atrium (LA), and Right Atrium (RA) volumes and ejection fractions are warranted or desired.

E. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE REFERENCE DEVICE

The VMS+ 3.0 is substantially equivalent to the primary predicate device, GE EchoPAC (K150085) and the reference device, Ventripoint Medical System Plus (VMS+) (K173810) with regard to both the intended use and technological characteristics.

As with the primary predicate device, VMS+ 3.0 provides annotation, analysis, measurement, report generation, communication, storage and retrieval of ultrasound images that are acquired via any ultrasound imaging system, for cardiology ultrasound applications.

Using the Knowledge-Based Reconstruction (KBR) algorithm and capturing 2D ultrasound images at specific angles and locations to recreate the shape of a heart is the technological principle for both the subject and predicate devices. It is based on the use of the constructed heart information to calculate volumes of any one of the four chambers, either at end-diastolic and/or end-systolic.

The VMS+ 3.0 has the same intended use and provides the same technological characteristics in terms of operational environment, software, operating system, operating principle, and performance when compared to the reference device.

At a high level, the subject and reference devices are based on the following same technological elements:

  • Knowledge-based reconstruction algorithm. Requires placement of anatomic landmarks on the 2D images. Landmarks are then used to build an accurate 3D constructed shape mesh (defined by vertices, edges, and faces).
  • 3D visualization and calcs: Generate a 3D surface model of the 4 chambers of the heart and calculate accompanying volume measurement.
  • Executing on a computer, processes data acquired by a sensor system in conjunction

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Ventripoint Diagnostics Ltd. The logo features a geometric design on the left, resembling a network of interconnected points and lines with different colored dots. To the right of the geometric design, the company name "Ventripoint" is written in a blue, sans-serif font, with "Diagnostics Ltd." in a smaller font size below it.

with the output from the ultrasound equipment, to enable tracking anatomic landmark points from the 2D images and conversion of those points into 3D data points.

  • Windows OS 7 and 10 based analysis system ●
  • . 3D tracking/positional system is magnetic-based; free-hand scanning
  • Console (system) and Workstation (standalone) software
  • Patient contacting components for image acquisition ●
  • Measurement and analysis technique: Record the ultrasound image sequences from ● the secondary video output of the 2D ultrasound machine and tracks the ultrasound transducer's 3D spatial coordinates during the imaging session by utilizing a 3D tracking system connected to the ultrasound transducer.
  • 3D Echo and MRI functionality
  • Original VMS+ and DICOM image data formats
  • . Software functionality: The software components are responsible for providing the user with controls for managing the hardware operation, capturing ultrasound images, marking up images with key anatomical features, invoking the reconstruction algorithm, displaying the resulting construction and generating the corresponding report.

The following technological differences exist between the subject and reference devices:

  • Knowledge-based reconstruction located locally on the system whereas for the predicate . device, the knowledge-based reconstruction is located on a Microsoft Windows Server and accessible to the data acquisition tool over the internet
  • Use of touchscreen and mouse as input devices
  • . Use of touchscreen for image capture and general user interface
  • . Medical grade computer mounted to a medical grade variable height roll stand
  • . User manually selects the End Diastolic and End Systolic frames whereas the predicate device automatically determined the first End Diastolic frame.
  • . The magnetic transmitter which acts as the reference point from which position and orientation has been moved closer to the patient.
  • .
  • Scan plane adjustment for minor adjustment of the plane position obtained form the position sensors.
  • User can both send to PACS and retrieve from PACS .

F. PERFORMANCE DATA

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Ventripoint Diagnostics Ltd. The logo features a geometric design on the left, resembling a network of interconnected points and lines with different colored dots. To the right of the design is the company name, "Ventripoint," in a blue, sans-serif font, with "Diagnostics Ltd." in a smaller font size below it.

The following performance data were provided in support of the substantial equivalence determination:

Biocompatibility Testing

Biocompatibility testing was conducted on patient contacting components of the VMS+ 3.0 device, including evaluation tests for cytotoxicity, sensitization of intracutaneous reactivity. The components comply with the ISO 10993-1 standard.

Electrical safety and electromagnetic compatibility (EMC)

Electrical safety and EMC testing were conducted on the VMS+ 3.0 device. The system complies with the IEC 60601-1 standard for safety and essential performance and IEC 60601-1-2 standard for electromaqnetic compatibility.

Software Verification and Validation

Software verification and validation testing were conducted, and documentation was provides as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "moderate" level of concern, since failures or latent design flaws are unlikely to result in minor injury to the patient or operator through incorrect or delayed information or through the action of a care provider.

The verification and validation of the existing and new features of the VMS+ 3.0 demonstrate that the VMS+ 3.0 performs as intended, specifications conform to user needs and intended uses, and that the requirements implemented can be consistently fulfilled.

All test reports were successful according to the acceptance criteria. The verification and validation were performed with software versions and hardware units that are considered equivalent to the final version of the product, as warranted by 21 CFR 820.30(g) and with the user interface as planned for the release.

Nonclinical Performance Bench Study

Performance bench testing of the modified device was completed to verify that the modified device was substantially equivalent in performance specifications to the previously cleared reference device. Predefined acceptance criteria were applied during testing and were met. The test results generated demonstrate that the device is as safe, as effective, performs as intended and as well as the predicate device (i.e. delivers volume measurements that are equivalent in accuracy when compared with volumes obtained using the legally marketed VMS+.

Animal Study

No animal study was required to support substantial equivalence for the subject device.

Clinical Studies

No clinical tests were conducted to support substantial equivalence for the subject device.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for Ventripoint Diagnostics Ltd. The logo features a geometric design on the left, resembling a network of interconnected points and lines with different colored dots. To the right of the geometric design is the company name, "Ventripoint," in a blue, sans-serif font, with "Diagnostics Ltd." in a smaller font size underneath.

G. Conclusions

Device Similarities

Intended use and other key features are consistent with traditional clinical practice and FDA guidance. The VMS+ 3.0 product conforms to applicable medical device safety standards and compliance is verified through independent evaluation. The design and development process of the manufacturer conforms to 21 CFR 820 Quality System Regulation and ISO 13485:2016 quality system standards.

Safety and Effectiveness

VMS+ 3.0 is a non-invasive technology. The modifications to the cleared VMS+ software and hardware do not introduce new questions concerning safety or effectiveness. and is therefore substantially equivalent to the predicate. Ventripoint Diagnostics Ltd. believes that VMS+ 3.0 is as safe and effective as the reference device and performs as well as the reference device.

N/A