K Number
K191415

Validate with FDA (Live)

Date Cleared
2019-08-30

(94 days)

Product Code
Regulation Number
876.1500
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Pathfinder™ Endoscope Overtube is intended to be used with an endoscope to facilitate intubation, change of endoscopes, and treatment in the gastrointestinal (GI) tract in adult patients (22 years of age and older).

Device Description

The Pathfinder™ Endoscope Overtube (Pathfinder™) device consists of a flexible overtube that may be connected to vacuum for rigidization via an attached stopcock and is used with an endoscope for procedures in the gastrointestinal tract. The stopcock is connected to the vacuum line which is connected to free space within the device and is completely contained, forming the vacuumable volume. The stopcock has two positions: the first position connects the vacuumable volume within the device to atmosphere (vent) to stay in the flexible condition, and the second position connects the device to a source of vacuum to the rigid condition. When transitioned to the rigid condition, the device maintains its shape at the time of rigidization, allowing the endoscope to advance or withdraw relative to the overtube with minimal disturbance to surrounding anatomy. When transitioned to the flexible condition, the device is able to move relative to the patient anatomy and endoscope for navigation through the GI tract. The device is provided sterile (EO). After use, the device is discarded and disposed of in accordance with local regulations.

AI/ML Overview

This document is a 510(k) premarket notification for a medical device called the Pathfinder™ Endoscope Overtube. It does not describe a study involving an AI-based medical device, nor does it include information on acceptance criteria for such a device.

The document focuses on demonstrating substantial equivalence to a predicate device (KMS Medical EndoGuide) primarily through:

  • Comparison of technological characteristics: This involves comparing the design, materials, intended use, and other features of the new device to the predicate.
  • Non-clinical performance data: This typically includes biocompatibility testing (cytotoxicity, sensitization, irritation, systemic toxicity, pyrogenicity) and mechanical testing (simulated use, lubricity, insufflation, insertion/removal, steering, navigation, rigidization/de-rigidization, endoscope compatibility).

Therefore, I cannot provide the information requested in your prompt regarding acceptance criteria for an AI-based device and the study proving it meets them. The document does not contain any of the following:

  • Table of acceptance criteria and reported device performance for an AI device.
  • Sample size used for an AI test set or data provenance.
  • Number of experts or their qualifications for establishing ground truth for an AI test set.
  • Adjudication method for an AI test set.
  • MRMC comparative effectiveness study or human reader improvement with AI assistance.
  • Standalone algorithm performance.
  • Type of ground truth used for an AI algorithm.
  • Sample size for an AI training set.
  • How ground truth for an AI training set was established.

The provided text is solely for a traditional medical device (an endoscope overtube) and its regulatory clearance process, not for an AI/ML-based diagnostic or therapeutic device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

August 30, 2019

Neptune Medical, Inc. % Ian Broome, M.S. Consultant AlvaMed. Inc. 935 Great Plain Avenue, #166 Needham, MA 02492

Re: K191415

Trade/Device Name: Pathfinder™ Endoscope Overtube Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: Class II Product Code: FED Dated: August 15, 2019 Received: August 16, 2019

Dear Ian Broome:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

for

Martha Betz, Ph.D. Acting Assistant Division Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K191415

Device Name Pathfinder™ Endoscope Overtube

Indications for Use (Describe)

The Pathfinder™ Endoscope Overtube is intended to be used with an endoscope to facilitate intubation, change of endoscopes, and treatment in the gastrointestinal (GI) tract in adult patients (22 years of age and older).

Type of Use (Select one or both, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

5.0 TRADITIONAL 510(K) SUMMARY FOR PATHFINDER™ ENDOSCOPE OVERTUBE DEVICE

I. SUBMITTER

Neptune Medical, Inc. 1828 El Camino Real, Suite 508 Burlingame, CA 94010

Phone: (617) 517-4932 Fax: (617) 249-0955

Contact Person: lan Broome, AlvaMed, Inc. Date Prepared: August 28, 2019

II. DEVICE

Name of Device: Pathfinder™ Endoscope Overtube (Model GI 085140) Common or Usual Name: Endoscopic Access Overtube Classification Name: Endoscope and accessories (21 CFR 876.1500) Requlatory Class: II Product Code: FED

III. PREDICATE DEVICE

KMS Medical EndoGuide (K063654)

IV. DEVICE DESCRIPTION

The Pathfinder™ Endoscope Overtube (Pathfinder™) device consists of a flexible overtube that may be connected to vacuum for rigidization via an attached stopcock and is used with an endoscope for procedures in the gastrointestinal tract. The stopcock is connected to the vacuum line which is connected to free space within the device and is completely contained, forming the vacuumable volume. The stopcock has two positions: the first position connects the vacuumable volume within the device to atmosphere (vent) to stay in the flexible condition, and the second position connects the device to a source of vacuum to the rigid condition. When transitioned to the rigid condition, the device maintains its shape at the time of rigidization, allowing the endoscope to advance or withdraw relative to the overtube with minimal disturbance to surrounding anatomy. When transitioned to the flexible condition, the device is able to move relative to the patient anatomy and endoscope for navigation through the GI tract. The device is provided sterile (EO). After use, the device is discarded and disposed of in accordance with local regulations.

There are no associated device accessories.

V. INDICATIONS FOR USE

The Pathfinder™ Endoscope Overtube is intended to be used with an endoscope to facilitate intubation, change of endoscopes, and treatment in the gastrointestinal (GI) tract in adult patients (22 years of age and older).

{4}------------------------------------------------

The Indications for Use statement for the Pathfinder™ Endoscope Overtube device above is similar to the predicate device; any differences do not alter the intended therapeutic use of the device nor do they affect the safety and effectiveness of the device relative to the predicate. Both the subject and predicate devices have the same intended use for endoscopic treatment in the GI tract.

The table below provides an overview of the intended patient population, environment of use, and contraindications.

Intended Population, Environment of Use, and Contraindications

Intended Population:Adults (≥ 22 years old) requiring gastrointestinal endoscopictreatment
Intended Environment ofUse:Gastrointestinal tract
Contraindications:Those who have had extensive abdominal surgeries may be poorcandidates because of adhesions or altered anatomy.Use contraindicated in patients with esophageal bleeding, lesion(s),and/or laceration; esophageal strictures and/or varices; laryngealperforation; trauma to teeth, gums, and/or pharynx; aspirationpneumonia; or any other condition that may preclude endoscopy.

VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

Subject Device:Pathfinder™ Endoscope OvertubePredicate Device:KMS EndoGuide
ManufacturerNeptune Medical, Inc.KMS Medical, LLC
510(k) Number(to be determined)K063654
Product CodeFEDFDF, subsequently FED
RegulationNumber21 CFR 876.150021 CFR 876.1500
RegulationDescriptionEndoscope and accessories.Endoscope and accessories.
CommonNameEndoscopic Access Overtube,Gastroenterology-urologyColonoscope and Accessories,Flexible/Rigid
Intended UseThe Pathfinder™ Endoscope Overtubeis intended to be used with anendoscope to facilitate intubation,change of endoscopes, and treatmentin the gastrointestinal (GI) tract in adultpatients (22 years of age and older).The EndoGuide is intended to be usedwith an endoscope to facilitateintubation, change of endoscopes andremoval of multiple polyps and/orforeign bodies.
SterilityEthylene Oxide (EO) SterilizationUnknown, believed to be EO
Single-UseYesYes

{5}------------------------------------------------

Subject Device:Pathfinder™ Endoscope OvertubePredicate Device:KMS EndoGuide
DesignCharacteristicsVacuum-assisted rigidizing overtubefor endoscopic procedures in the GlTractInternal wire-reinforced member●Vacuum-assisted rigidizing overtube●for endoscopic procedures in the GITract· Internal wire-reinforced member

Both devices are intended as assistive aids to the medical practitioner for endoscopic treatment in the GI tract.

The subject and predicate devices are based on the following same technological elements:

  • . an internal wire-reinforced member to the device;
  • switchable vacuum-based rigidization/de-rigidization; and ●
  • dimensions and design characteristics intended for use with pediatric endoscopes. ●

VII. PERFORMANCE DATA

The following performance data are provided in support of the substantial equivalence determination.

Biocompatibility Testing

The biocompatibility evaluation for the Pathfinder™ was conducted in accordance with the guidance document "Use of International Standard ISO 10993-1, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing Within a Risk Management Process," June 16, 2016, and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA and other applicable standards. The battery of testing included the following tests:

  • . Cytotoxicity
  • Sensitization
  • Irritation
  • Systemic Toxicity
  • . Pyrogenicity

The Pathfinder™ is considered to contact breached or compromised surfaces for a duration of less than 24 hours.

Mechanical Testinq

  • . Simulated use testing
  • Lubricity
  • Insufflation ●
  • Insertion/Removal ●
  • Steering ●
  • Navigation ●
  • Rigidization/De-Rigidization ●
  • Endoscope Compatibility

{6}------------------------------------------------

VIII. CONCLUSIONS

The non-clinical data support the safety of the device and the hardware verification and validation demonstrate that the Pathfinder™ shall perform as intended in the specified use conditions. The data demonstrate that the Pathfinder™ performs comparably to the predicate device currently marketed for the same intended use.

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.