K Number
K191397

Validate with FDA (Live)

Manufacturer
Date Cleared
2019-10-02

(131 days)

Product Code
Regulation Number
886.5925
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The OxySoft (olifilcon C) Spherical Silicone Hydrogel soft contact lenses are daily wear single use soft contact lens indicated for the correction of refractive ametropia (myopia and hyperopia) in phakic persons with nondiseased eyes who exhibit refractive astigmatism of 1.00D or less where the astigmatism does not interfere with visual acuity.

The OxySoft (olifileon C) Toric Silicone Hydrogel Soft Contact Lenses are indicated as daily wear for the correction of ametropia (myopia and hyperopia) with astigmatism in aphakic persons with non-diseased eyes and whose powers are from -20.00 to +20.00 diopters and astigmatic corrections are from -0.75 to -2.25 diopters.

The OxySoft (olifilcon C) Multifocal Silicone Hydrogel Soft Contact lenses are indicated as daily wear for the correction of ametropia (myopia and hyperopia) and emmetropia with presbyopia in aphakic persons with nondiseased eyes and whose powers are from -20.00 to +20.00 diopters with add powers from +0.75 to +2.75 diopters. The lenses may be worn by persons who exhibit astigmatism of 1.00 diopters or less where the astigmatism does not interfere with visual acuity.

Eye care practitioners may prescribe the lens for daily wear disposable use (single use). The lenses are to be discarded upon removal. Therefore, no cleaning or disinfecting is required.

Device Description

The OxySoft (olifilcon C) Silicone Hydrogel Soft Contact Lens is made of silicone hydrogel material with UV blocker available as spherical lens, toric lens and multifocal lens. The composition of the lens is 46% olifilcon C and 54 % water. A light blue color tinted with "reactive Blue19" is for handling visibility purpose. A benzotriazole UV absorbing monomer is used to block UV radiation. The transmittance characteristics are 1.9% (< 5%) in the UVB range of 280-315nm and 14.7% (< 50%) in the UVA range of 316-380nm. It is supplied in a sterile state packaged in a buffered saline solution.

AI/ML Overview

The provided document is a 510(k) premarket notification for a soft contact lens called OxySoft (olifilcon C). It does not describe an AI medical device. Therefore, a response matching the format of an AI device's acceptance criteria and study cannot be generated. However, I can summarize the information provided regarding the clinical study conducted for the contact lens.

Here's a summary of the clinical study for the OxySoft (olifilcon C) silicone hydrogel soft contact lens:

Clinical Study for OxySoft (olifilcon C) Silicone Hydrogel Soft Contact Lens

1. Acceptance Criteria and Reported Device Performance (as inferred from the summary):

The document does not explicitly state quantitative acceptance criteria for device performance in the same way an AI device would have specific metrics (e.g., sensitivity, specificity). Instead, the study aimed to demonstrate safety and effectiveness by comparison to a predicate device, focusing on:

MetricDesired Outcome (Inferred Acceptance)Reported Performance
SafetyNo significant adverse events or findings requiring treatment.- No adverse event (AE) occurred in the study.- No symptom or syndrome requiring treatment for all subjects during the study period.- No significant slit lamp findings (i.e., Grade 3 or Grade 4) were found.
Effectiveness (Visual Acuity)Visual acuity comparable to the predicate device and correctable to 0.1 LogMAR or better.- The mean visual acuity (VA) for both lenses (OxySoft and predicate Si-Hy) was similar at each visit.- Visual acuity could be corrected to 0.1 LogMAR or better for all subjects.- None of the subjects had VA decreased more than 2 lines.
Wear TimeSafe for daily wear up to 9 hours a day for 30 days.Found safe for daily disposable single use, worn up to 9 hours a day during 30 days follow-up.

2. Sample Size and Data Provenance:

  • Test Set Sample Size: 36 subjects were enrolled.
    • Trial group (OxySoft): 24 subjects
    • Control group (Si-Hy predicate): 12 subjects
  • Data Provenance: The document states, "All of them are Asia people." This suggests the data was collected in an Asian population, likely Taiwan, given the manufacturer's location. The study was prospective.

3. Number of Experts and Qualifications for Ground Truth:

  • This information is not provided in the document. For contact lens studies, "ground truth" usually refers to clinical measurements (visual acuity, slit lamp evaluations by optometrists/ophthalmologists) rather than expert consensus on images. The clinical evaluations were likely performed by the study investigators or their staff.

4. Adjudication Method for the Test Set:

  • This information is not provided in the document. Clinical trials typically have protocols for handling discrepancies in measurements or adverse events, but a specific "adjudication method" in the context of expert review for ground truth (as relevant to AI) is not applicable here.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

  • No, an MRMC study was not done. This type of study model is generally used for diagnostic devices that involve human interpretation of images, often with AI assistance. This document describes a clinical study for a physical medical device (contact lens).

6. Standalone (Algorithm Only) Performance:

  • No, this is not applicable. The device is a physical contact lens, not an algorithm.

7. Type of Ground Truth Used:

  • The "ground truth" for this study was established through a combination of clinical measurements and observations by healthcare professionals. This included:
    • Visual acuities
    • Slit lamp findings (e.g., for corneal health, lens fit)
    • Subjective symptoms, problems, and complaints (patient-reported outcomes)
    • Adverse reactions

8. Sample Size for the Training Set:

  • This is not applicable. The study describes a clinical trial for a physical medical device, not an AI model that requires a "training set."

9. How Ground Truth for the Training Set Was Established:

  • This is not applicable as there is no AI training set.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

October 2, 2019

Visco Vision Inc. % Ms. Jennifer Ting Manager Jens Medical Consulting Ltd. 3F, No. 364, Jixian Rd., Luzhou Dist. 247 New Taipei City, Taiwan R.O.C.

Re: K191397

Trade/Device Name: OxySoft (olifilcon C) silicone hydrogel soft contact lens Regulation Number: 21 CFR 886.5925 Regulation Name: Soft (Hydrophilic) Contact Lens Regulatory Class: Class II Product Code: LPL, MVN Dated: August 26, 2019 Received: August 27, 2019

Dear Ms. Ting:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801): medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-devicereporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

J. Angelo Green, Ph.D. Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K191397

Device Name

OxySoft (olifilcon C) Silicone Hydrogel Soft Contact Lens

Indications for Use (Describe)

The OxySoft (olifilcon C) Spherical Silicone Hydrogel soft contact lenses are daily wear single use soft contact lens indicated for the correction of refractive ametropia (myopia and hyperopia) in phakic persons with nondiseased eyes who exhibit refractive astigmatism of 1.00D or less where the astigmatism does not interfere with visual acuity.

The OxySoft (olifileon C) Toric Silicone Hydrogel Soft Contact Lenses are indicated as daily wear for the correction of ametropia (myopia and hyperopia) with astigmatism in aphakic persons with non-diseased eyes and whose powers are from -20.00 to +20.00 diopters and astigmatic corrections are from -0.75 to -2.25 diopters.

The OxySoft (olifilcon C) Multifocal Silicone Hydrogel Soft Contact lenses are indicated as daily wear for the correction of ametropia (myopia and hyperopia) and emmetropia with presbyopia in aphakic persons with nondiseased eyes and whose powers are from -20.00 to +20.00 diopters with add powers from +0.75 to +2.75 diopters. The lenses may be worn by persons who exhibit astigmatism of 1.00 diopters or less where the astigmatism does not interfere with visual acuity.

Eye care practitioners may prescribe the lens for daily wear disposable use (single use). The lenses are to be discarded upon removal. Therefore, no cleaning or disinfecting is required.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)
----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDEDThis section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

Preparation Date: July 16, 2019

5.1 Establishment Information:

NameVisco Vision Inc.
AddressNo. 1, Xingye St., Guishan Dist., Taoyuan City 33341 Taiwan
ContactTed TT Huang
Phone No.886-3-3596868
Fax No.886-3-3490202

5.2 Contact Person:

NameEvan Huang
Phone No.886-3-3596868
Fax No.886-3-3490202
e-mail:evan.huang@viscovision.com.tw

5.3 Device Identification:

Proprietary NameOxySoft (olifilcon C) Silicone Hydrogel Soft ContactLens
Common NameSoft (hydrophilic) Contact Lenses
Classification NameLenses, Soft Contact, Daily Wear,(21 CFR 886.5925, Product Code LPL)Lenses, Soft Contact, Daily Wear (Disposable),(21 CFR 886.5925, Product Code MVN)
ClassificationII

5.4 Legally Marketed Equivalent Device:

Predicate Device NameSi-Hy (olifilcon B) Silicone Hydrogel Soft ContactLens
ManufacturerVisco Vision Inc.
510(k) NumberK160344
Product CodeMVN

{4}------------------------------------------------

5.5 Device Description

The OxySoft (olifilcon C) Silicone Hydrogel Soft Contact Lens is made of silicone hydrogel material with UV blocker available as spherical lens, toric lens and multifocal lens. The composition of the lens is 46% olifilcon C and 54 % water. A light blue color tinted with "reactive Blue19" is for handling visibility purpose. A benzotriazole UV absorbing monomer is used to block UV radiation. The transmittance characteristics are 1.9% (< 5%) in the UVB range of 280-315nm and 14.7% (< 50%) in the UVA range of 316-380nm. It is supplied in a sterile state packaged in a buffered saline solution.

5.6 Indication for Use:

The OxySoft (olifilcon C) Spherical Silicone Hydrogel soft contact lenses are daily wear single use soft contact lens indicated for the correction of refractive ametropia (myopia and hyperopia) in phakic or aphakic persons with nondiseased eyes who exhibit refractive astigmatism of 1.00D or less where the astigmatism does not interfere with visual acuity.

The OxySoft (olifilcon C) Toric Silicone Hydrogel Soft Contact Lenses are indicated as daily wear for the correction of ametropia (myopia and hyperopia) with astigmatism in aphakic and non-aphakic persons with non-diseased eyes and whose powers are from -20.00 to +20.00 diopters and astigmatic corrections are from -0.75 to -2.25 diopters.

The OxySoft (olifilcon C) Multifocal Silicone Hydrogel Soft Contact lenses are indicated as daily wear for the correction of ametropia (myopia and hyperopia) and emmetropia with presbyopia in aphakic and non-aphakic persons with nondiseased eyes and whose powers are from -20.00 to +20.00 diopters with add powers from +0.75 to +2.75 diopters. The lenses may be worn by persons who exhibit astigmatism of 1.00 diopters or less where the astigmatism does not interfere with visual acuity.

Eye care practitioners may prescribe the lens for daily wear disposable use (single use). The lenses are to be dscarded upon removal. Therefore, no cleaning or disinfecting is required.

{5}------------------------------------------------

5.7 Technological characteristic

OxySoft SPHERICAL Soft Contact Lenses characteristics:

  • Diameter Range:13.0 to 15.0 mm .
  • Base Curve:8.0 to 9.2 mm .
  • . Center Thickness:0.08 mm for -3.00D (varies with power)
  • Power:+20.00 to -20.00 D o

OxySoft TORIC Soft Contact Lenses characteristics:

  • . Diameter Range:13.0 to 15.0 mm
  • . Base Curve:8.0 to 9.2 mm
  • Center Thickness:0.08 mm for -3.00D (varies with power) .
  • Power:+20.00 to -20.00 D .
  • Cylinder: -0.75D ~ -2.25D .
  • . Axis: 10° to 180° (in 10° increments)

OxySoft MULTIFOCAL Soft Contact Lenses characteristics:

  • Diameter Range:13.0 to 15.0 mm
  • . Base Curve:8.0 to 9.2 mm
  • Center Thickness:0.08 mm for -3.00D (varies with power) .
  • Power:+20.00 to -20.00 D .
  • Additional Powers: +0.75D ~ +1.25D (LOW) .
    • +1.50D ~ +2.00D (MID)
    • +2.25D ~ +2.75D (HIGH)

5.8 Comparison table:

The characteristic comparison to predicate device is summarized in the following table.

Similarities and differences
ItemDevicePredicate (K160344)
Product NameOxySoft (olifilcon C) SiliconeHydrogel soft contact lensSi-Hy (olifilcon B) SiliconeHydrogel soft contact lens
ManufacturerVISCO VISION Inc.VISCO VISION Inc.
Intended UseMyopia, Hyperopia,astigmatism, PresbyopiaThe same
Lens DesignSpherical, toric, or multifocalThe same
Replacement ScheduleDaily Disposable (Single use)The same
Chemical compositionolifilcon Colifilcon B

{6}------------------------------------------------

ClassificationGroup II(Nonionic, High water)Group I (Nonionic, Low water)
Water Content54 %47 % (<50%),
Oxygen Permeability(DK, 35°C )100(Fatt method)120(Fatt method)
Base Curve Range (mm)8.0~9.28.0~9.2
Diameter (mm)13.0~15.013.0~15.0
Center ThicknessVaries with design and power(0.08 mm at -3.00D)Varies with design and power (0.08 mm at -3.00D)
Powers-20.00D to +20.00D-20.00D to +20.00D
Refractive Index1.3961.410
Light Transmittance94%94%
Blue handling tintReactive Blue19Reactive Blue19
Method of ManufactureMoldedMolded

5.9 Nonclinical Tests Performed

  • Physiochemical studies were conducted according to ISO 18369 First 5.9.1 edition 2006-08-15, Ophthalmic optics - Contact lenses (Ophthalmic). The physical, optical and chemical properties of the lens are within established specifications for the lenses.
  • 5.9.2 Toxicology studies report shows that the lenses are non-toxic and biocompatibility result is acceptable in ocular environment.

5.10 Clinical Studies

A one-month clinical study was conducted to demonstrate the safety and efffectiveness of the OxySoft (olifilcon C) silicone hydrogel soft contact lens by comparison with Si-Hy (olifilcon B) silicone hydrogel soft contact lens when worn on a daily wear basis. At least 30 evaluable subjects were participated. Parameters measured include visual acuities, slit lamp findings, symptom, problem and complaints, adverse reactions, as well as lens wearing time.

The study evaluated thirty six (36) male and female subjects who were randomized and dispensed lenses in a 2:1 ratio with 24 subjects dispensed into the Trial lenses and 12 Subjects dispensed into the Control lenses. Of the 36 subjects who were enrolled in the study, 35 subjects were female and 1 subject was male. All of them are Asia people. The subject ranged in age from 22 years to 59 years. The gender ratio is 1 male to 35 females. One subject is new contact lens wearer (2.78%) and 35 subjects (97.22%) have successful experience with daily wear soft contact lens.

{7}------------------------------------------------

In trial group, 24 subjects range in age from 22 years to 59 years with average 33.29±10.02 years. All of the subjects are female. The gender ratio is 24 to 0 (24 female/0 male). In control group, 12 subjects range in age from 25 years to 46 years with average 35.58±7.09 years. One subject is male and 11 subjects female. The gender ratio was 11 to 1 (11 female/1 male).

It was found that the OxySoft silicone hydrogel soft contact lens is safe as a daily disposable single use lens to wear up to 9 hours a day during 30 days follow up by comparing to Si-Hy (olifilcon B) silicone hydrogel soft contact lens. There is no symptom or syndrome that requires treatment for all subjects during study period.

The mean visual acuity (VA) for both lenses was similar at each visit. The visual acuity could be corrected to 0.1 LogMAR or better. None of the subjects had VA decreased more than 2 lines. No significant slit lamp findings (i.e. Grade 3 or Grade 4) were found. No adverse event (AE) occurred.

5.11 Conclusion

Comparison to the predicate device for chemical composition, physical and optical properties, it shows that "OxySoft silicone hydrogel soft contact lens" is as safe, as effective and performs as well as the predicate device.

§ 886.5925 Soft (hydrophilic) contact lens.

(a)
Identification. A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.(b)
Classification. (1) Class II if the device is intended for daily wear only.(2) Class III if the device is intended for extended wear.
(c)
Date PMA or notice of completion of a PDP is required. As of May 28, 1976, an approval under section 515 of the act is required before a device described in paragraph (b)(2) of this section may be commercially distributed. See § 886.3.