K Number
K191310

Validate with FDA (Live)

Date Cleared
2019-06-10

(26 days)

Product Code
Regulation Number
892.1650
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Intended for use by a qualified/trained doctor on both adult and pediatric subjects for obtaining fluoroscopic radiographic images of the skull, spinal column, chest, abdomen, extremities, and other body parts.

Device Description

The Insight Essentials™ DRF Digital Imaging System is designed to replace the video camera and user interface on the following FDA cleared R&F fluoroscopic imaging systems employing an image intensifier: GE Advantx, GE Legacy, GE P500 and Shimadzu YSF300. The x-ray generator, beam-limiting device and patient positioner. necessary for a full fluoroscopy system are not part of the subject device.

The Insight Essentials™ DRF Digital Imaging System allows the operator to view and enhance high-definition fluoroscopy images. High resolution digital spot images may also be acquired at single or rapid acquisition rates, up to 30 fps. Images may be viewed and enhanced enabling the operator to bring out diagnostic details difficult or impossible to see using conventional imaging techniques. Images can be stored locally for short term storage. The Insight Essentials™ DRF Digital Imaging System enables the operator to produce hardcopy images with a laser printer or send images over a network for longer-term storage. The major system components include: a fluoroscopic video camera, monitors, and an image processor PC.

AI/ML Overview

This document describes a 510(k) submission for the Insight Essentials™ DRF Digital Imaging System, comparing it to a predicate device, the FluoroPRO RF Digital Imaging System (K070390). The submission focuses on demonstrating substantial equivalence rather than proving clinical performance against specific acceptance criteria for an AI algorithm.

Therefore, the requested information regarding acceptance criteria, study details, expert involvement, and ground truth for an AI device is not directly applicable to this submission, as it is a device modification submission for digital imaging system components, not an AI/ML medical device.

However, I can extract information regarding non-clinical tests performed to demonstrate substantial equivalence, which serves as the "study" for compliance in this context.

Here's a breakdown based on the provided document:

1. Table of Acceptance Criteria and Reported Device Performance:

Since this is a substantial equivalence submission for hardware/software components, the "acceptance criteria" are implied by the comparison to the predicate device. The goal is to show the new device is "as safe and effective as" the predicate. Performance is assessed through non-clinical bench testing.

CharacteristicPredicate Device Performance (FluoroPro RF Digital Imaging System - K070390)Subject Device Performance (Insight Essentials™ DRF Digital Imaging System - K191310)Acceptance Criterion (Implied)Outcome
Intended UseAs described in documentSAMEIdentical to PredicateMet
Power Source120 VAC 50/60 HZ, 2.5 ampsSAMEIdentical to PredicateMet
Image AcquisitionUp to 15 FPS (spot), up to 30 fps (fluoro)SAMEIdentical to PredicateMet
File CompatibilityDICOMSAMEIdentical to PredicateMet
Digital Video CameraThales TH 8740 - 020 CCDBasler acA1920-50gm CMOSSubstantially EquivalentMet
Digital Resolution1000 x 1000 12-bit1k x 1k 12 bitSubstantially EquivalentMet
Electrical SafetyIEC 60601-1, IEC 60601-1-2, IEC 60601-1-3SAMEConformance to StandardsMet
Image Output Characteristics (Bench Test)(Implied standard based on predicate)Substantially equivalent to predicateSubstantially EquivalentMet
Doses Used (Bench Test)(Implied standard based on predicate)Analyzed and compared to predicateSubstantially EquivalentMet

The Study:

The study conducted to demonstrate the device meets these (implied) criteria is a non-clinical bench testing study.

2. Sample Size Used for the Test Set and the Data Provenance:

  • Sample Size: Not explicitly stated as a number of distinct "cases" or "images" in the way an AI study would define a test set. The document mentions "phantom images" were acquired. The quantity of phantom images is not specified.
  • Data Provenance: The images were acquired from a "GE Legacy R&F system" with the subject device installed and compared to images from a "GE Advantx 1 R&F system" with the predicate device installed. This indicates prospective acquisition for the purpose of the comparison. The country of origin is not specified but implicitly within the context of the submitter's (Imaging Engineering, LLC in Florida, USA) operations.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts:

  • Not applicable in the context of this submission. The "ground truth" for this engineering bench test is objective measurement and comparison of technical parameters and image quality, not expert clinical interpretation.

4. Adjudication Method for the Test Set:

  • Not applicable. This was a technical comparison, not a clinical interpretation study requiring adjudication. The analysis was based on direct comparison of acquired images and doses.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • No. This is not an AI-assisted device, nor is it a clinical efficacy study involving human readers.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

  • Not applicable. This device is a digital imaging system component, not an autonomous algorithm.

7. The Type of Ground Truth Used:

  • The ground truth in this context is objective technical performance measurements (e.g., image output characteristics, resolution, doses, adherence to safety standards) derived from the phantom images acquired during bench testing, compared against the known specifications and performance of the predicate device.

8. The Sample Size for the Training Set:

  • Not applicable. This is a hardware/software upgrade to an existing imaging system, not a machine learning model that requires a training set.

9. How the Ground Truth for the Training Set was Established:

  • Not applicable, as there is no training set for an AI/ML model.

{0}------------------------------------------------

June 10, 2019

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Imaging Engineering, LLC % Mr. George Jachode President 1318 NW 7th Avenue CAPE CORAL FL 33993

Re: K191310

Trade/Device Name: Insight Essentials™ DRF Digital Imaging System Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: Class II Product Code: JAA, LLZ Dated: May 10, 2019 Received: May 15, 2019

Dear Mr. Jachode:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see

{1}------------------------------------------------

https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K191310

Device Name

Insight Essentials™ DRF Digital Imaging System

Indications for Use (Describe)

Intended for use by a qualified/trained doctor on both adult and pediatric subjects for obtaining fluoroscopic radiographic images of the skull, spinal column, chest, abdomen, extremities, and other body parts.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Imaging Engineering, LLC. The logo consists of two overlapping squares, one blue and one orange, each made up of smaller squares. The text "Imaging Engineering, LLC" is written in a simple, sans-serif font below the squares.

1. Traditional 510(k) SUMMARY

K191310

This summary of 510(k) is being submitted in accordance with requirements of SMDA 1990 and 21 CFR Part 807.92.

Date 510K summary prepared: May 10, 2019

Submitter's Name:Imaging Engineering, LLC
Submitter's Address:1318 NW 7th Ave, Cape Coral, Florida 33993
Submitter's Telephone:(239) 223-1371

Contact person: George Jachode/President

Official Correspondent: George Jachode (jachode@aol.com)
Address:1318 NW 7th Ave, Cape Coral, Florida 33993
Telephone:(239) 223-1371

Name of the device, including the trade or proprietary name if applicable, the common or usual name and the classification name, if known:

510K Number:K191310Trade/proprietary name: Insight Essentials™ DRF Digital Imaging System
Model Number:ES-1000-00
Regulation Name:lmage-Intensified Fluoroscopic X-ray System
Regulation Number:21 CFR 892. 1650
Regulatory Class:Class II
Product Code:JAA and LLZ

Predicate Device

Trade Name:FluoroPRO RF Digital Imaging System
510(k) Clearance #:K070390
Clearance date:02/02/2007
Classification Name:Image Intensified Fluoroscopic X-ray System
Classification Panel:Radiology
CFR Section:21CFR 892.1650 (Produce Code; JAA and LLZ)
Device Class:Class II

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Imaging Engineering, LLC. The logo consists of a square made up of smaller squares in shades of blue and orange. The text "Imaging Engineering, LLC" is written in a simple, sans-serif font below the square graphic.

2. Device Description

The Insight Essentials™ DRF Digital Imaging System is designed to replace the video camera and user interface on the following FDA cleared R&F fluoroscopic imaging systems employing an image intensifier: GE Advantx, GE Legacy, GE P500 and Shimadzu YSF300. The x-ray generator, beam-limiting device and patient positioner. necessary for a full fluoroscopy system are not part of the subject device.

The Insight Essentials™ DRF Digital Imaging System allows the operator to view and enhance high-definition fluoroscopy images. High resolution digital spot images may also be acquired at single or rapid acquisition rates, up to 30 fps. Images may be viewed and enhanced enabling the operator to bring out diagnostic details difficult or impossible to see using conventional imaging techniques. Images can be stored locally for short term storage. The Insight Essentials™ DRF Digital Imaging System enables the operator to produce hardcopy images with a laser printer or send images over a network for longer-term storage. The major system components include: a fluoroscopic video camera, monitors, and an image processor PC.

3. Indications for Use

Intended for use by a qualified/trained doctor or technician on both adult and pediatric subjects for obtaining fluoroscopic radiographic images of the skull, spinal column, chest, abdomen, extremities, and other body parts.

4. Summary of Design Control Risk Management

The Insight Essentials™ DRF Digital Imaging System has been developed to provide medical professionals optimized workflow when imaging patients while meeting critical functional requirements and international safety standards. The risks and the hazardous impact of the device design were analyzed with FMEA method. The specific risk control and protective measures to mitigate the risks from the device design and production phase were reviewed and implemented in the new product design phase. The overall assessment concluded that all risks and hazardous conditions identified arising from the design and production were successfully mitigated and accepted.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Imaging Engineering, LLC. The logo consists of two overlapping squares, one orange and one blue. The squares are made up of smaller squares, giving them a pixelated appearance. The text "Imaging Engineering, LLC" is located below the logo.

5. Substantial Equivalence

The Insight Essentials™ DRF Digital Imaging System conforms to the FDA recognized standards as like the predicate device. Based on the recognized standard conformity evidences related to the electro-, mechanical, software-, clinical-and risk management, it's the sponsor's opinion that the subject device is a safe and effective device.

CharacteristicFluoroPro RF DigitalImaging System(K070390)Insight EssentialsDRF DigitalImaging System(K191310)
Intended Use:Intended for use by aqualified/trained doctor ortechnician on both adultand pediatric subjects forobtaining fluoroscopicradiographic images of theskull, spinal column, chest,abdomen, extremities, andother body parts.SAME
Power source120 VAC 50/60 HZ2.5 amps120 VAC 50/60 HZ2.5 amps
Image acquisitionUp to 15 FPS (spot), up to30 fps (fluoro)SAME
File compatibilityDICOMSAME
Digital Video CameraThales TH 8740 - 020CCDBasler acA1920-50gmCMOS
Digital Resolution1000 x 1000 12-bit1k x 1k 12 bit
Electrical safetyIEC 60601-1:IEC 60601-1-2IEC 60601-1-3SAME

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows a logo and the text "Imaging Engineering, LLC". The logo consists of two overlapping squares, one in shades of blue and the other in shades of orange. The blue square is positioned slightly below and to the left of the orange square, creating a layered effect.

6. Difference Discussion

There are four primary differences between the two systems are due to the amount of time that has passed (and the technological advancements made) since the predicate device was first cleared in 2007.

  • a. The CCD camera in the predicate device has been replaced with a CMOS camera with substantially equivalent features.
  • b. The personal computer (pc) used in the predicate device uses Windows XP operating system. The subject device has a pc that uses Windows 10 Professional operating system.
  • c. The Digital Interface Board (DIB) was updated to incorporate a modern microcontroller.
  • d. The predicate device did not have a Dose Area Product (DAP) meter to record and display patient exposure dose. The subject device includes a DAP meter.

7. Summary of the technological characteristics of the device compared to the predicate device

Both the subject device and the predicate device have the identical software program that provides a Graphical User Interface (GUI), image processing and patient workflow functions.

The CCD camera in the predicate device is replaced with a CMOS camera. The image output characteristics of both cameras are substantially equivalent. The CCD camera technology used in the predicate device is at the end of its life.

The DIB has been updated with a modern microcontroller as the predicate device microcontroller is no longer in production.

The pc used in the predicate device has a Windows XP operating system that is obsolete and no longer in production. The pc of the subject device has a Windows 10 Professional operating system that includes modern cybersecurity protection and is compliant with recently established guidelines.

The subject device includes a DAP meter to record and provide DAP and Air Kerma measurements that are displayed and stored in the DICOM header of patient studies. The fluoro exam time and timer are maintained by the host R&F system.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for Imaging Engineering, LLC. The logo consists of two overlapping squares made up of smaller squares. The top right square is orange, and the bottom left square is blue. The text "Imaging Engineering, LLC" is located below the logo.

8. Description of non-clinical tests.

Bench testing was performed on the subject device installed on a GE Legacy R&F system to assess substantial equivalence of the device. Phantom images were acquired with the subject device and compared to the predicate device, installed on a GE Advantx 1 R&F system. These images and the doses used to acquire them were analyzed and compared. In conclusion, the tests obtained demonstrated substantial equivalence to the predicate device.

9. Description of clinical tests.

No clinical data is necessary to evaluate safety or effectiveness for purposes of determining substantial equivalence of the proposed modification. Bench testing was performed to assess the device safety and effectiveness.

10. Conclusion as to Substantial Equivalence.

The Insight Essentials™ DRF Digital Imaging System, the subject device is substantially equivalent to the predicate (K070390). The intended use, the design principle, and the applicable standards for the subject device are identical to those of the predicate device. The performance test and non-clinical consideration result demonstrate that these differences do not raise any new questions of safety and effectiveness. Therefore, it is the sponsor's opinion that the subject device appears to be as safe and effective as the predicate device.

§ 892.1650 Image-intensified fluoroscopic x-ray system.

(a)
Identification. An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.