K Number
K191230

Validate with FDA (Live)

Date Cleared
2019-11-08

(184 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Halyard Sterling Nitrile Powder-Free Exam Gloves are disposable devices intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. These gloves were tested for use with the following chemotherapy drugs:
Arsenic Trioxide (1 mg/ml), No breakthrough up to 240 minutes Doxorubicin HCL (2 mg/ml), No breakthrough up to 240 minutes Paclitaxel (6 mg/ml), No breakthrough up to 240 minutes Azacitidine (Vidaza) (25 mg/ml), No breakthrough up to 240 minutes Epirubicin (Ellence) (2 mg/ml), No breakthrough up to 240 minutes Paraplatin (10 mg/ml), No breakthrough up to 240 minutes Bendamustine (5 mg/ml), No breakthrough up to 240 minutes Eribulin Mesylate (0.5 mg/ml), No breakthrough up to 240 minutes Pemetrexed (25 mg/ml), No breakthrough up to 240 minutes Bortezomib (Velcade) (1 mg/ml), No breakthrough up to 240 minutes Etoposide (20 mg/ml), No breakthrough up to 240 minutes Pertuzumab (30 mg/ml), No breakthrough up to 240 minutes Bleomycin sulfate (15 mg/ml), No breakthrough up to 240 minutes Fludarabine (25 mg/ml), No breakthrough up to 240 minutes Raltitrexed (0.5 mg/ml), No breakthrough up to 240 minutes Busulfan (6 mg/ml),No breakthrough up to 240 minutes Fluorouracil (50 mg/ml), No breakthrough up to 240 minutes, Retrovir (10 mg/ml) No breakthrough up to 240 minutes Carboplatin (10 mg/ml), No breakthrough up to 240 minutes Fulvestrant (50 mg/ml), No breakthrough up to 240 minutes Rituximab (10 mg/ml), No breakthrough up to 240 minutes Carfilzomib (2 mg/ml). No breakthrough up to 240 minutes Gemcitabine (38 mg/ml), No breakthrough up to 240 minutes Temsirolimus (25 mg/ml), No breakthrough up to 240 minutes Cetuximab (Erbitux) (2 mg/ml), No breakthrough up to 240 minutes Idarubicin (1 mg/ml), No breakthrough up to 240 minutes Trastuzumab (21 mg/ml), No breakthrough up to 240 minutes Cisplatin (1 mg/ml), No breakthrough up to 240 minutes Ifosfamide (50 mg/ml), No breakthrough up to 240 minutes Topotecan HCL (1 mg/ml), No breakthrough up to 240 minutes Cyclophosphamide (20 mg/ml), No breakthrough up to 240 minutes Irinotecan (20 mg/ml), No breakthrough up to 240 minutes Triclosan (2 mg/ml), No breakthrough up to 240 minutes Cytarabine HCL (100 mg/ml), No breakthrough up to 240 minutes Mechlorethamine HCL (1 mg/ml), No breakthrough up to 240 minutes Trisonex (1 mg/ml), No breakthrough up to 240 minutes Cytovene (10 mg/ml), No breakthrough up to 240 minutes Melphalan (5 mg/ml), No breakthrough up to 240 minutes Vincrinstine Sulfate (1 mg/ml), No breakthrough up to 240 minutes
Dacarbazine (10 mg/ml), No breakthrough up to 240 minutes Methotrexate (25 mg/ml), No breakthrough up to 240 minutes, Vinblastine (1 mg/ml), No breakthrough up to 240 minutes Daunorubicin HCL (5 mg/ml). No breakthrough up to 240 minutes Mitomycin-C (0.5 mg/ml), No breakthrough up to 240 minutes Vinorelbine (10 mg/ml), No breakthrough up to 240 minutes Decitabine (5 mg/ml), No breakthrough up to 240 minutes Mitoxantrone (2 mg/ml), No breakthrough up to 240 minutes Zoledronic Acid (0.8 mg/ml).No breakthrough up to 240 minutes Docetaxel (10 mg/ml), No breakthrough up to 240 minutes Oxaliplatin (2 mg/ml), No breakthrough up to 240 minutes
ThioTEPA (10 mg/ml), breakthrough detected at 37.1 minutes Carmustine (3.3 mg/ml), breakthrough detected at 22.9 minutes

The Halyard Sterling SG* Nitrile Sensi-Guard Powder-Free Exam Gloves are disposable devices intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. These gloves were tested for use with the following chemotherapy drugs:
Arsenic Trioxide (1 mg/ml), No breakthrough up to 240 minutes Doxorubicin HCL (2 mg/ml), No breakthrough up to 240 minutes Paclitaxel (6 mg/ml), No breakthrough up to 240 minutes Azacitidine (Vidaza) (25 mg/ml), No breakthrough up to 240 minutes Epirubicin (Ellence) (2 mg/ml), No breakthrough up to 240 minutes Paraplatin (10 mg/ml), No breakthrough up to 240 minutes Bendamustine (5 mg/ml), No breakthrough up to 240 minutes Eribulin Mesylate (0.5 mg/ml), No breakthrough up to 240 minutes Pemetrexed (25 mg/ml), No breakthrough up to 240 minutes Bortezomib (Velcade) (1 mg/ml), No breakthrough up to 240 minutes Etoposide (20 mg/ml), No breakthrough up to 240 minutes Pertuzumab (30 mg/ml), No breakthrough up to 240 minutes Bleomycin sulfate (15 mg/ml), No breakthrough up to 240 minutes Fludarabine (25 mg/ml), No breakthrough up to 240 minutes Raltitrexed (0.5 mg/ml), No breakthrough up to 240 minutes Busulfan (6 mg/ml),No breakthrough up to 240 minutes Fluorouracil (50 mg/ml), No breakthrough up to 240 minutes Retrovir (10 mg/ml) No breakthrough up to 240 minutes Carboplatin (10 mg/ml), No breakthrough up to 240 minutes Fulvestrant (50 mg/ml), No breakthrough up to 240 minutes Rituximab (10 mg/ml), No breakthrough up to 240 minutes Carfilzomib (2 mg/ml). No breakthrough up to 240 minutes Gemcitabine (38 mg/ml), No breakthrough up to 240 minutes Temsirolimus (25 mg/ml), No breakthrough up to 240 minutes Cetuximab (Erbitux) (2 mg/ml), No breakthrough up to 240 minutes Idarubicin (1 mg/ml), No breakthrough up to 240 minutes Trastuzumab (21 mg/ml), No breakthrough up to 240 minutes Cisplatin (1 mg/ml), No breakthrough up to 240 minutes Ifosfamide (50 mg/ml), No breakthrough up to 240 minutes Topotecan HCL (1 mg/ml), No breakthrough up to 240 minutes Cyclophosphamide (20 mg/ml), No breakthrough up to 240 minutes Irinotecan (20 mg/ml), No breakthrough up to 240 minutes Triclosan (2 mg/ml), No breakthrough up to 240 minutes Cytarabine HCL (100 mg/ml), No breakthrough up to 240 minutes Mechlorethamine HCL (1 mg/ml), No breakthrough up to 240 minutes Trisonex (1 mg/ml), No breakthrough up to 240 minutes Cytovene (10 mg/ml), No breakthrough up to 240 minutes Melphalan (5 mg/ml), No breakthrough up to 240 minutes Vincrinstine Sulfate (1 mg/ml), No breakthrough up to 240 minutes
Dacarbazine (10 mg/ml), No breakthrough up to 240 minutes Methotrexate (25 mg/ml), No breakthrough up to 240 minutes, Vinblastine (1 mg/ml), No breakthrough up to 240 minutes Daunorubicin HCL (5 mg/ml), No breakthrough up to 240 minutes Mitomycin-C (0.5 mg/ml), No breakthrough up to 240 minutes Vinorelbine (10 mg/ml), No breakthrough up to 240 minutes Decitabine (5 mg/ml), No breakthrough up to 240 minutes Mitoxantrone (2 mg/ml), No breakthrough up to 240 minutes Zoledronic Acid (0.8 mg/ml).No breakthrough up to 240 minutes Docetaxel (10 mg/ml), No breakthrough up to 240 minutes Oxaliplatin (2 mg/ml), No breakthrough up to 240 minutes
ThioTEPA (10 mg/ml), breakthrough detected at 23.9 minutes Carmustine (3.3 mg/ml), breakthrough detected at 14.8 minutes

Device Description

Halyard Sterling SG* Nitrile Sensi-Guard Powder-Free Exam Gloves and Halyard STERLING* Nitrile Powder-Free Exam Gloves tested for Use with Chemotherapy Drugs are disposable, grey-colored, chlorinated, nitrile, powder-free, textured fingertip, ambidextrous, non-sterile patient examination gloves that are packed in a cardboard dispenser box.

AI/ML Overview

The provided document is a 510(k) summary for Halyard Sterling* Nitrile Powder-Free Exam Gloves and Halyard Sterling SG* Nitrile Sensi-Guard Powder-Free Exam Gloves. It details the safety and performance characteristics of these medical gloves, particularly their resistance to permeation by chemotherapy drugs.

Here’s a breakdown of the acceptance criteria and study information:

1. Table of Acceptance Criteria and Reported Device Performance

The device (Halyard Sterling SG* Nitrile Sensi-Guard Powder-Free Exam Gloves and Halyard Sterling* Nitrile Powder-Free Exam Gloves) performance was evaluated against several standards for medical gloves and their resistance to chemotherapy drugs.

StandardAcceptance CriteriaReported Device Performance (Halyard Sterling SG* Nitrile Sensi-Guard Powder-Free Exam Gloves)Reported Device Performance (Halyard Sterling* Nitrile Powder-Free Exam Gloves)
ASTM D6978-05: Assessment of Medical Gloves to Permeation by Chemotherapy DrugsFor most drugs, "No breakthrough up to 240 minutes." For specific drugs, breakthrough times are expected.- No breakthrough up to 240 minutes for: Arsenic Trioxide, Doxorubicin HCL, Paclitaxel, Azacitidine, Epirubicin, Paraplatin, Bendamustine, Eribulin Mesylate, Pemetrexed, Bortezomib, Etoposide, Pertuzumab, Bleomycin sulfate, Fludarabine, Raltitrexed, Busulfan, Fluorouracil, Retrovir, Carboplatin, Fulvestrant, Rituximab, Carfilzomib, Gemcitabine, Temsirolimus, Cetuximab, Idarubicin, Trastuzumab, Cisplatin, Ifosfamide, Topotecan HCL, Cyclophosphamide, Irinotecan, Triclosan, Cytarabine HCL, Mechlorethamine HCL, Trisonex, Cytovene, Melphalan, Vincrinstine Sulfate, Dacarbazine, Methotrexate, Vinblastine, Daunorubicin HCL, Mitomycin-C, Vinorelbine, Decitabine, Mitoxantrone, Zoledronic Acid, Docetaxel, Oxaliplatin. - Breakthrough detected at 23.9 minutes for ThioTEPA (10 mg/ml). - Breakthrough detected at 14.8 minutes for Carmustine (3.3 mg/ml).- No breakthrough up to 240 minutes for: Arsenic Trioxide, Doxorubicin HCL, Paclitaxel, Azacitidine, Epirubicin, Paraplatin, Bendamustine, Eribulin Mesylate, Pemetrexed, Bortezomib, Etoposide, Pertuzumab, Bleomycin sulfate, Fludarabine, Raltitrexed, Busulfan, Fluorouracil, Retrovir, Carboplatin, Fulvestrant, Rituximab, Carfilzomib, Gemcitabine, Temsirolimus, Cetuximab, Idarubicin, Trastuzumab, Cisplatin, Ifosfamide, Topotecan HCL, Cyclophosphamide, Irinotecan, Triclosan, Cytarabine HCL, Mechlorethamine HCL, Trisonex, Cytovene, Melphalan, Vincrinstine Sulfate, Dacarbazine, Methotrexate, Vinblastine, Daunorubicin HCL, Mitomycin-C, Vinorelbine, Decitabine, Mitoxantrone, Zoledronic Acid, Docetaxel, Oxaliplatin. - Breakthrough detected at 37.1 minutes for ThioTEPA (10 mg/ml). - Breakthrough detected at 22.9 minutes for Carmustine (3.3 mg/ml).
ASTM D5151-06: Detection of Holes in Medical GlovesMust meet an AQL of 2.5% for leakage.Pass: Meets the 2.5% AQL requirement for leakage.Pass: Meets the 2.5% AQL requirement for leakage.
ASTM D6124-06: Residual Powder on Medical GlovesLess than 2 mg maximum powder per glove.Pass: Average of 0.03 mg/glove, within the powder-free limit.Pass: Average of 0.03 mg/glove, within the powder-free limit.
ASTM D6319-10: Specification for Nitrile Examination Gloves for Medical ApplicationsWithin the limits of the standard for tensile strength and elongation before and after aging.Pass: Average Tensile Strength (before aging) = 35.2 MPa; Average Tensile Strength (after aging) = 44.1 MPa; Average Elongation (before aging) = 606%; Average Elongation (after aging) = 553%.Pass: Average Tensile Strength (before aging) = 37.2 MPa; Average Tensile Strength (after aging) = 41.6 MPa; Average Elongation (before aging) = 606%; Average Elongation (after aging) = 567%.
ISO 10993: Systemic Toxicity Study in MiceNo animals treated with test extracted exhibited greater reaction than control animals.Pass: No mortality or evidence of systemic toxicity from the extracts.Pass: No mortality or evidence of systemic toxicity from the extracts.
ISO 10993: Dermal IrritationPrimary Irritation Index ≤ Grade 1.Pass: Considered non-irritating.Pass: Considered non-irritating.
ISO 10993: Dermal SensitizationStimulation Index ≤ 3.Pass: Showed no evidence of causing delayed dermal contact sensitization.Pass: Showed no evidence of causing delayed dermal contact sensitization.

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state the sample sizes (e.g., number of gloves tested for each drug or physical property) for the test sets. It describes the tests conducted according to the "Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs" (ASTM D6978-05), and other ASTM and ISO standards, which inherently include methodologies for sample selection.

The data provenance is not specified in terms of country of origin, nor whether it was retrospective or prospective. It refers to the testing conducted on the subject device, implying this testing occurred directly for this submission.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This type of information (number and qualifications of experts) is not applicable to the studies described. The "ground truth" here is established by the standardized test methods themselves, which are objective measurements of physical and chemical properties of the gloves (e.g., breakthrough time for chemotherapy drugs, presence of holes, powder content, tensile strength, biological reaction in animal models). These are not studies that require expert interpretation of images or clinical outcomes.

4. Adjudication Method for the Test Set

Not applicable. The tests are based on objective, standardized measurements, not human interpretation requiring adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

No, an MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic devices where human readers interpret medical images or data. The device in question is a medical exam glove, and its effectiveness is determined by its physical and chemical properties under specified test conditions, not by human reader performance.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

Not applicable. This is not an AI/algorithm-based device.

7. The Type of Ground Truth Used

The "ground truth" for the performance evaluation of these gloves is defined by:

  • Standardized Test Methods: Adherence to established industry standards like ASTM D6978-05 for chemotherapy drug permeation, ASTM D5151-06 for hole detection, ASTM D6124-06 for residual powder, ASTM D6319-10 for physical properties, and ISO 10993 for biological evaluation.
  • Objective Measurements: Direct measurements of physical properties (e.g., breakthrough time, amount of powder, tensile strength, elongation) and biological responses (e.g., systemic toxicity, irritation, sensitization).

8. The Sample Size for the Training Set

Not applicable. There is no "training set" in the context of evaluating medical gloves through physical and chemical performance testing. This is not an AI or machine learning model.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no training set for this type of device evaluation.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 8, 2019

O & M Halyard, Inc. Steve Dowdley Associate Director of Regulatory Affairs 9120 Lockwood Blvd Mechanicsville, Virginia 23116

Re: K191230

Trade/Device Name: Halyard Sterling* Nitrile Powder-Free Exam Gloves, Halyard Sterling SG* Nitrile Sensi-Guard Powder-Free Exam Gloves Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZC Dated: July 19, 2019 Received: August 13, 2019

Dear Steve Dowdley:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For Elizabeth Claverie, M.S. Assistant Director for THT4B2 Acting Assistant Director for THT4B1 DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K191230

Device Name

Halyard Sterling Nitrile Powder-Free Exam Gloves

Indications for Use (Describe)

The Halyard Sterling Nitrile Powder-Free Exam Gloves are disposable devices intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. These gloves were tested for use with the following chemotherapy drugs:

Arsenic Trioxide (1 mg/ml), No breakthrough up to 240 minutes Doxorubicin HCL (2 mg/ml), No breakthrough up to 240 minutes Paclitaxel (6 mg/ml), No breakthrough up to 240 minutes Azacitidine (Vidaza) (25 mg/ml), No breakthrough up to 240 minutes Epirubicin (Ellence) (2 mg/ml), No breakthrough up to 240 minutes Paraplatin (10 mg/ml), No breakthrough up to 240 minutes Bendamustine (5 mg/ml), No breakthrough up to 240 minutes Eribulin Mesylate (0.5 mg/ml), No breakthrough up to 240 minutes Pemetrexed (25 mg/ml), No breakthrough up to 240 minutes Bortezomib (Velcade) (1 mg/ml), No breakthrough up to 240 minutes Etoposide (20 mg/ml), No breakthrough up to 240 minutes Pertuzumab (30 mg/ml), No breakthrough up to 240 minutes Bleomycin sulfate (15 mg/ml), No breakthrough up to 240 minutes Fludarabine (25 mg/ml), No breakthrough up to 240 minutes Raltitrexed (0.5 mg/ml), No breakthrough up to 240 minutes Busulfan (6 mg/ml),No breakthrough up to 240 minutes Fluorouracil (50 mg/ml), No breakthrough up to 240 minutes, Retrovir (10 mg/ml) No breakthrough up to 240 minutes Carboplatin (10 mg/ml), No breakthrough up to 240 minutes Fulvestrant (50 mg/ml), No breakthrough up to 240 minutes Rituximab (10 mg/ml), No breakthrough up to 240 minutes Carfilzomib (2 mg/ml). No breakthrough up to 240 minutes Gemcitabine (38 mg/ml), No breakthrough up to 240 minutes Temsirolimus (25 mg/ml), No breakthrough up to 240 minutes Cetuximab (Erbitux) (2 mg/ml), No breakthrough up to 240 minutes Idarubicin (1 mg/ml), No breakthrough up to 240 minutes Trastuzumab (21 mg/ml), No breakthrough up to 240 minutes Cisplatin (1 mg/ml), No breakthrough up to 240 minutes Ifosfamide (50 mg/ml), No breakthrough up to 240 minutes Topotecan HCL (1 mg/ml), No breakthrough up to 240 minutes Cyclophosphamide (20 mg/ml), No breakthrough up to 240 minutes Irinotecan (20 mg/ml), No breakthrough up to 240 minutes Triclosan (2 mg/ml), No breakthrough up to 240 minutes Cytarabine HCL (100 mg/ml), No breakthrough up to 240 minutes Mechlorethamine HCL (1 mg/ml), No breakthrough up to 240 minutes Trisonex (1 mg/ml), No breakthrough up to 240 minutes Cytovene (10 mg/ml), No breakthrough up to 240 minutes Melphalan (5 mg/ml), No breakthrough up to 240 minutes Vincrinstine Sulfate (1 mg/ml), No breakthrough up to 240 minutes

{3}------------------------------------------------

Dacarbazine (10 mg/ml), No breakthrough up to 240 minutes Methotrexate (25 mg/ml), No breakthrough up to 240 minutes, Vinblastine (1 mg/ml), No breakthrough up to 240 minutes Daunorubicin HCL (5 mg/ml). No breakthrough up to 240 minutes Mitomycin-C (0.5 mg/ml), No breakthrough up to 240 minutes Vinorelbine (10 mg/ml), No breakthrough up to 240 minutes Decitabine (5 mg/ml), No breakthrough up to 240 minutes Mitoxantrone (2 mg/ml), No breakthrough up to 240 minutes Zoledronic Acid (0.8 mg/ml).No breakthrough up to 240 minutes Docetaxel (10 mg/ml), No breakthrough up to 240 minutes Oxaliplatin (2 mg/ml), No breakthrough up to 240 minutes

ThioTEPA (10 mg/ml), breakthrough detected at 37.1 minutes Carmustine (3.3 mg/ml), breakthrough detected at 22.9 minutes

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Indications for Use

510(k) Number (if known) K191230

Device Name

Halyard Sterling SG * Nitrile Sensi-Guard Powder-Free Exam Gloves

Indications for Use (Describe)

The Halyard Sterling SG* Nitrile Sensi-Guard Powder-Free Exam Gloves are disposable devices intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. These gloves were tested for use with the following chemotherapy drugs:

Arsenic Trioxide (1 mg/ml), No breakthrough up to 240 minutes Doxorubicin HCL (2 mg/ml), No breakthrough up to 240 minutes Paclitaxel (6 mg/ml), No breakthrough up to 240 minutes Azacitidine (Vidaza) (25 mg/ml), No breakthrough up to 240 minutes Epirubicin (Ellence) (2 mg/ml), No breakthrough up to 240 minutes Paraplatin (10 mg/ml), No breakthrough up to 240 minutes Bendamustine (5 mg/ml), No breakthrough up to 240 minutes Eribulin Mesylate (0.5 mg/ml), No breakthrough up to 240 minutes Pemetrexed (25 mg/ml), No breakthrough up to 240 minutes Bortezomib (Velcade) (1 mg/ml), No breakthrough up to 240 minutes Etoposide (20 mg/ml), No breakthrough up to 240 minutes Pertuzumab (30 mg/ml), No breakthrough up to 240 minutes Bleomycin sulfate (15 mg/ml), No breakthrough up to 240 minutes Fludarabine (25 mg/ml), No breakthrough up to 240 minutes Raltitrexed (0.5 mg/ml), No breakthrough up to 240 minutes Busulfan (6 mg/ml),No breakthrough up to 240 minutes Fluorouracil (50 mg/ml), No breakthrough up to 240 minutes Retrovir (10 mg/ml) No breakthrough up to 240 minutes Carboplatin (10 mg/ml), No breakthrough up to 240 minutes Fulvestrant (50 mg/ml), No breakthrough up to 240 minutes Rituximab (10 mg/ml), No breakthrough up to 240 minutes Carfilzomib (2 mg/ml). No breakthrough up to 240 minutes Gemcitabine (38 mg/ml), No breakthrough up to 240 minutes Temsirolimus (25 mg/ml), No breakthrough up to 240 minutes Cetuximab (Erbitux) (2 mg/ml), No breakthrough up to 240 minutes Idarubicin (1 mg/ml), No breakthrough up to 240 minutes Trastuzumab (21 mg/ml), No breakthrough up to 240 minutes Cisplatin (1 mg/ml), No breakthrough up to 240 minutes Ifosfamide (50 mg/ml), No breakthrough up to 240 minutes Topotecan HCL (1 mg/ml), No breakthrough up to 240 minutes Cyclophosphamide (20 mg/ml), No breakthrough up to 240 minutes Irinotecan (20 mg/ml), No breakthrough up to 240 minutes Triclosan (2 mg/ml), No breakthrough up to 240 minutes Cytarabine HCL (100 mg/ml), No breakthrough up to 240 minutes Mechlorethamine HCL (1 mg/ml), No breakthrough up to 240 minutes Trisonex (1 mg/ml), No breakthrough up to 240 minutes Cytovene (10 mg/ml), No breakthrough up to 240 minutes Melphalan (5 mg/ml), No breakthrough up to 240 minutes Vincrinstine Sulfate (1 mg/ml), No breakthrough up to 240 minutes

{5}------------------------------------------------

Dacarbazine (10 mg/ml), No breakthrough up to 240 minutes Methotrexate (25 mg/ml), No breakthrough up to 240 minutes, Vinblastine (1 mg/ml), No breakthrough up to 240 minutes Daunorubicin HCL (5 mg/ml), No breakthrough up to 240 minutes Mitomycin-C (0.5 mg/ml), No breakthrough up to 240 minutes Vinorelbine (10 mg/ml), No breakthrough up to 240 minutes Decitabine (5 mg/ml), No breakthrough up to 240 minutes Mitoxantrone (2 mg/ml), No breakthrough up to 240 minutes Zoledronic Acid (0.8 mg/ml).No breakthrough up to 240 minutes Docetaxel (10 mg/ml), No breakthrough up to 240 minutes Oxaliplatin (2 mg/ml), No breakthrough up to 240 minutes

ThioTEPA (10 mg/ml), breakthrough detected at 23.9 minutes Carmustine (3.3 mg/ml), breakthrough detected at 14.8 minutes

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Halyard, which is part of the Owens & Minor Family. The logo consists of a circular emblem with a stylized white and green shape inside, followed by the word "HALYARD" in bold, dark blue letters. Below the company name, the text "Part of the Owens & Minor Family" is written in a smaller, lighter font.

510(k) Summary

Date Summarywas PreparedNov 4, 2019
510(k) SubmitterOwens & Minor Halyard, Inc9120 Lockwood BlvdMechanicsville, VA 23116
Primary Contact forthis 510(k) SubmissionSteven Dowdley, RACTel: 678-451-8062Email: steven.dowdley@hyh.com
Device Trade NameHalyard Sterling SG* Nitrile Sensi-Guard Powder-Free Exam GlovesandHalyard STERLING* Nitrile Powder-Free Exam Gloves
Device Common NameMedical Exam Gloves
Device Product Codeand Classification NameLZCClass I, 21 CFR §880.6250Patient Examination Glove
Predicate DeviceK180646Halyard Lavender Nitrile Powder-Free Exam Gloves
Subject Device DescriptionHalyard Sterling SG* Nitrile Sensi-Guard Powder-FreeExam Gloves and Halyard STERLING* Nitrile Powder-FreeExam Gloves tested for Use with Chemotherapy Drugs aredisposable, grey-colored, chlorinated, nitrile, powder-free,textured fingertip, ambidextrous, non-sterile patientexamination gloves that are packed in a cardboarddispenser box.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the Halyard logo. The logo consists of a blue circle with a white and green shape inside, followed by the word "HALYARD" in blue, block letters. Below the word "HALYARD" is the phrase "Part of the Owens & Minor Family" in a smaller font.

Indications for Use forHalyard Sterling SG* NitrileSensi-Guard Powder-FreeExam GlovesThe Halyard Sterling SG* Nitrile Sensi-Guard Powder-Free Exam Gloves aredisposable devices intended for medical purposes that is worn on theexaminer's hand to prevent contamination between patient and examiner.These gloves were tested for use with the following chemotherapy drugs:
Arsenic Trioxide (1 mg/ml), No breakthrough up to 240 minutes
Doxorubicin HCL (2 mg/ml), No breakthrough up to 240 minutes
Paclitaxel (6 mg/ml), No breakthrough up to 240 minutes
Azacitidine (Vidaza) (25 mg/ml), No breakthrough up to 240 minutes
Epirubicin (Ellence) (2 mg/ml), No breakthrough up to 240 minutes
Paraplatin (10 mg/ml), No breakthrough up to 240 minutes
Bendamustine (5 mg/ml), No breakthrough up to 240 minutes
Eribulin Mesylate (0.5 mg/ml), No breakthrough up to 240 minutes
Pemetrexed (25 mg/ml), No breakthrough up to 240 minutes
Bortezomib (Velcade) (1 mg/ml), No breakthrough up to 240 minutes
Etoposide (20 mg/ml), No breakthrough up to 240 minutes
Pertuzumab (30 mg/ml), No breakthrough up to 240 minutes
Bleomycin sulfate (15 mg/ml), No breakthrough up to 240 minutes
Fludarabine (25 mg/ml), No breakthrough up to 240 minutes
Raltitrexed (0.5 mg/ml), No breakthrough up to 240 minutes
Busulfan (6 mg/ml), No breakthrough up to 240 minutes
Fluorouracil (50 mg/ml), No breakthrough up to 240 minutes, Nobreakthrough up to 240 minutes
Retrovir (10 mg/ml) No breakthrough up to 240 minutes
Carboplatin (10 mg/ml), No breakthrough up to 240 minutes
Fulvestrant (50 mg/ml), No breakthrough up to 240 minutes
Rituximab (10 mg/ml), No breakthrough up to 240 minutes
Carfilzomib (2 mg/ml), No breakthrough up to 240 minutes
Gemcitabine (38 mg/ml), No breakthrough up to 240 minutes
Temsirolimus (25 mg/ml), No breakthrough up to 240 minutes
Cetuximab (Erbitux) (2 mg/ml), No breakthrough up to 240 minutes
Idarubicin (1 mg/ml), No breakthrough up to 240 minutes
Trastuzumab (21 mg/ml), No breakthrough up to 240 minutes
Cisplatin (1 mg/ml), No breakthrough up to 240 minutes
Ifosfamide (50 mg/ml), No breakthrough up to 240 minutes
Topotecan HCL (1 mg/ml), No breakthrough up to 240 minutes
Cyclophosphamide (20 mg/ml), No breakthrough up to 240 minutes
Irinotecan (20 mg/ml), No breakthrough up to 240 minutes
Triclosan (2 mg/ml), No breakthrough up to 240 minutes
Cytarabine HCL (100 mg/ml), No breakthrough up to 240 minutes
Mechlorethamine HCL (1 mg/ml), No breakthrough up to 240 minutes
Trisonex (1 mg/ml), No breakthrough up to 240 minutes
Cytovene (10 mg/ml), No breakthrough up to 240 minutes
Melphalan (5 mg/ml), No breakthrough up to 240 minutes
Vincrinstine Sulfate (1 mg/ml), No breakthrough up to 240 minutes
Dacarbazine (10 mg/ml) No breakthrough up to 240 minutes
Methotrexate (25 mg/ml), No breakthrough up to 240 minutes,Vinblastine (1 mg/ml), No breakthrough up to 240 minutesDaunorubicin HCL (5 mg/ml), No breakthrough up to 240 minutesMitomycin-C (0.5 mg/ml), No breakthrough up to 240 minutesVinorelbine (10 mg/ml), No breakthrough up to 240 minutesDecitabine (5 mg/ml), No breakthrough up to 240 minutesMitoxantrone (2 mg/ml), No breakthrough up to 240 minutesZoledronic Acid (0.8 mg/ml),No breakthrough up to 240 minutesDocetaxel (10 mg/ml), No breakthrough up to 240 minutesOxaliplatin (2 mg/ml), No breakthrough up to 240 minutesThioTEPA (10 mg/ml), breakthrough detected at 23.9 minutesCarmustine (3.3 mg/ml), breakthrough detected at 14.8 minutes
Indications for Use forHalyard Sterling NitrilePowder-Free Exam GlovesThe Halyard STERLING* Nitrile Powder-Free Exam Gloves Tested for Use withChemotherapy Drugs are disposable devices intended for medical purposesthat is worn on the examiner's hand to prevent contamination betweenpatient and examiner. These gloves were tested for use with the followingchemotherapy drugs:
Arsenic Trioxide (1 mg/ml), No breakthrough up to 240 minutesDoxorubicin HCL (2 mg/ml), No breakthrough up to 240 minutesPaclitaxel (6 mg/ml), No breakthrough up to 240 minutesAzacitidine (Vidaza) (25 mg/ml), No breakthrough up to 240 minutesEpirubicin (Ellence) (2 mg/ml), No breakthrough up to 240 minutesParaplatin (10 mg/ml), No breakthrough up to 240 minutesBendamustine (5 mg/ml), No breakthrough up to 240 minutesEribulin Mesylate (0.5 mg/ml), No breakthrough up to 240 minutesPemetrexed (25 mg/ml), No breakthrough up to 240 minutesBortezomib (Velcade) (1 mg/ml), No breakthrough up to 240 minutesEtoposide (20 mg/ml), No breakthrough up to 240 minutesPertuzumab (30 mg/ml), No breakthrough up to 240 minutesBleomycin sulfate (15 mg/ml), No breakthrough up to 240 minutesFludarabine (25 mg/ml), No breakthrough up to 240 minutesRaltitrexed (0.5 mg/ml), No breakthrough up to 240 minutesBusulfan (6 mg/ml),No breakthrough up to 240 minutes
Cisplatin (1 mg/ml), No breakthrough up to 240 minutesIfosfamide (50 mg/ml), No breakthrough up to 240 minutesTopotecan HCL (1 mg/ml), No breakthrough up to 240 minutesCyclophosphamide (20 mg/ml), No breakthrough up to 240 minutesIrinotecan (20 mg/ml), No breakthrough up to 240 minutesTriclosan (2 mg/ml), No breakthrough up to 240 minutesCytarabine HCL (100 mg/ml), No breakthrough up to 240 minutesMechlorethamine HCL (1 mg/ml), No breakthrough up to 240 minutesTrisonex (1 mg/ml), No breakthrough up to 240 minutesCytovene (10 mg/ml), No breakthrough up to 240 minutesMelphalan (5 mg/ml), No breakthrough up to 240 minutesVincrinstine Sulfate (1 mg/ml), No breakthrough up to 240 minutes
Methotrexate (25 mg/ml), No breakthrough up to 240 minutes
Vinblastine (1 mg/ml), No breakthrough up to 240 minutes
Daunorubicin HCL (5 mg/ml), No breakthrough up to 240 minutes
Mitomycin-C (0.5 mg/ml), No breakthrough up to 240 minutes
Vinorelbine (10 mg/ml), No breakthrough up to 240 minutes
Decitabine (5 mg/ml), No breakthrough up to 240 minutes
Mitoxantrone (2 mg/ml), No breakthrough up to 240 minutes
Zoledronic Acid (0.8 mg/ml),No breakthrough up to 240 minutes
Docetaxel (10 mg/ml), No breakthrough up to 240 minutes
Oxaliplatin (2 mg/ml), No breakthrough up to 240 minutes
ThioTEPA (10 mg/ml), breakthrough detected at 37.1 minutes
Carmustine (3.3 mg/ml), breakthrough detected at 22.9 minutes
Summary of comparison oftechnological characteristicsBoth the subject devices and the predicate devices are sterile,powder-free nitrile exam gloves tested for resistance to permeationby chemotherapy drugs.

{8}------------------------------------------------

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for Halyard. The logo consists of a circular emblem with a stylized image of a house in white and green, followed by the word "HALYARD" in bold, dark blue letters. Below the name, it says "Part of the Owens & Minor Family" in a smaller font.

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the Halyard logo. The logo consists of a blue circle with a white and green shape inside, followed by the word "HALYARD" in blue. Below the word "HALYARD" is the phrase "Part of the Owens & Minor Family" in a smaller font.

Technological Characteristics Table
Subject DeviceHalyard SterlingSG* Nitrile Sensi-Guard Powder-Free Exam GlovesSubject DeviceHalyard SterlingNitrile Powder-Free Exam GlovesPredicateK180646Comparison
FDA Product CodeLZCLZCLZCSame
FDA ClassificationClass IClass IClass ISame
Regulation880.6250880.6250880.6250Same
Common NameMedical ExamMedical Exam GloveMedical ExamSame
Device Trade NameHalyard SterlingSG* Nitrile Sensi-Guard Powder-Free Exam GlovesHalyard SterlingNitrile Powder-FreeExam GlovesHalyard LavenderNitrile Powder-Free Exam GlovesSimilar
CharacteristicsGrey- colored, 9.5inch, nitrile, 0.08mm thick at palm,powder-free,textured fingertips,ambidextrous, non-sterile patientexamination gloveGrey- colored, 9.5inch, nitrile, 0.08mm thick at palm,powder-free,textured fingertips,ambidextrous, non-sterile patientexamination gloveLavender-colored,9.5 inch, nitrile,0.08 mm thick atpalm, powder-free,textured fingertips,ambidextrous, non-sterile patientexamination gloveSimilar
Sizes of glovesXS, S, M, L, XLXS, S, M, L, XLXS, S, M, L, XLSame

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image shows the Halyard logo. The logo consists of a blue circle with a white and green shape inside, followed by the word "HALYARD" in blue. Below the word "HALYARD" is the text "Part of the Owens & Minor Family".

TextureTextured fingertipsTextured finger tipsTextured finger tipsSame
SterilityNon-sterileNon-sterileNon-sterileSame
Summary ofcomparison oftechnologicalcharacteristicsThere are no different technological characteristics of the subject devices,(Halyard of technological Sterling SG* Nitrile Sensi-Guard Powder-Free ExamGloves and Halyard STERLING* Nitrile Powder-Free Exam Gloves) compared tothe predicate device. They are powder-free nonsterile nitrile exam glovestested for resistance to permeation by chemotherapy drugs. The subjectdevices were tested for use with 52 drugs and the predicate device was testedfor use with 12 drugs. The color and thickness profile are different betweenthe subject devices and the predicate device.
Performance Data for Chemotherapy Drugs
StandardSubject Device Halyard Sterling SG* Nitrile Sensi-Guard Powder-Free Exam GlovesSubject Device Halyard Sterling Nitrile Powder-Free Exam GlovesPredicate K180646Comparison
ASTM D6978-05Standard Practice forAssessment of MedicalGloves to Permeationby Chemotherapy DrugsArsenic Trioxide (1 mg/ml), No breakthrough up to 240 minutesArsenic Trioxide (1 mg/ml), No breakthrough up to 240 minutesNo signs of breakthrough after 4 hours for 10 drugs.Similar
Doxorubicin HCL (2 mg/ml), No breakthrough up to 240 minutesDoxorubicin HCL (2 mg/ml), No breakthrough up to 240 minutesCarmustine showed no signs of breakthrough until 0.3 minutes
Paclitaxel (6 mg/ml), No breakthrough up to 240 minutesPaclitaxel (6 mg/ml), No breakthrough up to 240 minutesThiotepa showed no signs of breakthrough until 30.9 minutes.
Azacitidine (Vidaza) (25 mg/ml), No breakthrough up to 240 minutesAzacitidine (Vidaza) (25 mg/ml), No breakthrough up to 240 minutesCyclophosphamide (20.0 mg/ml) No breakthrough up to 240 minutes
Epirubicin (Ellence) (2 mg/ml), No breakthrough up to 240 minutesEpirubicin (Ellence) (2 mg/ml), No breakthrough up to 240 minutesDoxorubicin HCI (2.0 mg/ml) No breakthrough up to 240 minutes

{12}------------------------------------------------

Image /page/12/Picture/0 description: The image contains the Halyard logo. The logo consists of a blue circle with a white and green shape inside, followed by the word "HALYARD" in large, bold, blue letters. Below the word "HALYARD" is the text "Part of the Owens & Minor Family" in a smaller font size.

Paraplatin (10 mg/ml), No breakthrough up to 240 minutesParaplatin (10 mg/ml), No breakthrough up to 240 minutesEtoposide (20.0 mg/ml) No breakthrough up to 240 minutes
Bendamustine (5 mg/ml), No breakthrough up to 240 minutesEribulin Mesylate (0.5 mg/ml), No breakthrough up to 240 minutesBendamustine (5 mg/ml), No breakthrough up to 240 minutesEribulin Mesylate (0.5 mg/ml), No breakthrough up to 240 minutes5-Fluorouracil (50.0 mg/ml) No breakthrough up to 240 minutesPaclitaxel (Taxol) (6.0 mg/ml) No breakthrough up to 240 minutes
Pemetrexed (25 mg/ml), No breakthrough up to 240 minutesPemetrexed (25 mg/ml), No breakthrough up to 240 minutesCisplatin (1.0 mg/ml) No breakthrough up to 240 minutes
Bortezomib (Velcade) (1 mg/ml), No breakthrough up to 240 minutesBortezomib (Velcade) (1 mg/ml), No breakthrough up to 240 minutesDacarbazine (10.0 mg/ml) No breakthrough up to 240 minutes
Etoposide (20 mg/ml), No breakthrough up to 240 minutesEtoposide (20 mg/ml), No breakthrough up to 240 minutesIfosfamide (50.0 mg/ml) No breakthrough up to 240 minutes
Pertuzumab (30 mg/ml), No breakthrough up to 240 minutesPertuzumab (30 mg/ml), No breakthrough up to 240 minutesMitoxantrone (2.0 mg/ml) No breakthrough up to 240 minutes
Bleomycin sulfate (15 mg/ml), No breakthrough up to 240 minutesBleomycin sulfate (15 mg/ml), No breakthrough up to 240 minutesVincristine sulfate (1.0 mg/ml) No breakthrough up to 240 minutes
Fludarabine (25 mg/ml), No breakthrough up to 240 minutesFludarabine (25 mg/ml), NoCarmustine (3.3 mg/ml) No breakthrough up to 0.3 minutes
Raltitrexed (0.5 mg/ml), No breakthrough up to 240 minutesBusulfan (6 mg/ml),No breakthrough up to 240 minutesFluorouracil (50 mg/ml), No breakthrough up to 240 minutesRetrovir (10 mg/ml) No breakthrough up to 240 minutesCarboplatin (10 mg/ml), No breakthrough up to 240 minutesFulvestrant (50 mg/ml), No breakthrough up to 240 minutesRituximab (10 mg/ml), No breakthrough up to 240 minutesCarfilzomib (2 mg/ml), No breakthrough up to 240 minutesGemcitabine (38 mg/ml), No breakthrough up to 240 minutesbreakthrough up to 240 minutesRaltitrexed (0.5 mg/ml), No breakthrough up to 240 minutesBusulfan (6 mg/ml),No breakthrough up to 240 minutesFluorouracil (50 mg/ml), No breakthrough up to 240 minutes,Retrovir (10 mg/ml) No breakthrough up to 240 minutesCarboplatin (10 mg/ml), No breakthrough up to 240 minutesFulvestrant (50 mg/ml), No breakthrough up to 240 minutesRituximab (10 mg/ml), No breakthrough up to 240 minutesCarfilzomib (2 mg/ml), No breakthrough up to 240 minutesGemcitabine (38 mg/ml) NoThioTEPA (10.0 mg.ml) No breakthrough up to 30.9 minutes
Temsirolimus (25 mg/ml), No breakthrough up to 240 minutesbreakthrough up to 240 minutesTemsirolimus (25 mg/ml), No breakthrough up to 240 minutes
Cetuximab (Erbitux) (2 mg/ml), No breakthrough up to 240 minutesCetuximab (Erbitux) (2 mg/ml), No breakthrough up to 240 minutes
Idarubicin (1 mg/ml), No breakthrough up to 240 minutesIdarubicin (1 mg/ml), No breakthrough up to 240 minutes
Trastuzumab (21 mg/ml), No breakthrough up to 240 minutesTrastuzumab (21 mg/ml), No breakthrough up to 240 minutes
Cisplatin (1 mg/ml), No breakthrough up to 240 minutesCisplatin (1 mg/ml), No breakthrough up to 240 minutes
Ifosfamide (50 mg/ml), No breakthrough up to 240 minutesIfosfamide (50 mg/ml), No breakthrough up to 240 minutes
Topotecan HCL (1 mg/ml), No breakthrough up to 240 minutesTopotecan HCL (1 mg/ml), No breakthrough up to 240 minutes
Cyclophosphamide (20 mg/ml), No breakthrough up to 240 minutesCyclophosphamide (20 mg/ml), No breakthrough up to 240 minutes
Irinotecan (20 mg/ml), No
breakthrough upto 240 minutesIrinotecan (20 mg/ml), Nobreakthrough upto 240 minutes
Triclosan (2 mg/ml), Nobreakthrough upto 240 minutesTriclosan (2 mg/ml), Nobreakthrough upto 240 minutes
Cytarabine HCL(100 mg/ml), Nobreakthrough upto 240 minutesCytarabine HCL(100 mg/ml), Nobreakthrough upto 240 minutes
MechlorethamineHCL (1 mg/ml), Nobreakthrough upto 240 minutesMechlorethamineHCL (1 mg/ml), Nobreakthrough upto 240 minutes
Trisonex (1 mg/ml), Nobreakthrough upto 240 minutesTrisonex (1 mg/ml), Nobreakthrough upto 240 minutes
Cytovene (10 mg/ml), Nobreakthrough upto 240 minutesCytovene (10 mg/ml), Nobreakthrough upto 240 minutes
Melphalan (5 mg/ml), Nobreakthrough upto 240 minutesMelphalan (5 mg/ml), Nobreakthrough upto 240 minutes
VincrinstineSulfate (1 mg/ml),No breakthroughup to 240 minutesVincrinstineSulfate (1 mg/ml),No breakthroughup to 240 minutes
Dacarbazine (10mg/ml), Nobreakthrough upto 240 minutesDacarbazine (10mg/ml), Nobreakthrough upto 240 minutes
Methotrexate (25mg/ml) No
breakthrough upto 240 minutes,Methotrexate (25 mg/ml), Nobreakthrough upto 240 minutes,
Vinblastine (1mg/ml), Nobreakthrough upto 240 minutesVinblastine (1mg/ml), Nobreakthrough upto 240 minutes
Daunorubicin HCL(5 mg/ml), Nobreakthrough upto 240 minutesDaunorubicin HCL(5 mg/ml), Nobreakthrough upto 240 minutes
Mitomycin-C (0.5mg/ml), Nobreakthrough upto 240 minutesMitomycin-C (0.5mg/ml), Nobreakthrough upto 240 minutes
Vinorelbine (10mg/ml), Nobreakthrough upto 240 minutesVinorelbine (10mg/ml), Nobreakthrough upto 240 minutes
Decitabine (5mg/ml), Nobreakthrough upto 240 minutesDecitabine (5mg/ml), Nobreakthrough upto 240 minutes
Mitoxantrone (2mg/ml), Nobreakthrough upto 240 minutesMitoxantrone (2mg/ml), Nobreakthrough upto 240 minutes
Zoledronic Acid(0.8 mg/ml),Nobreakthrough upto 240 minutesZoledronic Acid(0.8 mg/ml),Nobreakthrough upto 240 minutes
Docetaxel (10mg/ml), Nobreakthrough upto 240 minutesDocetaxel (10mg/ml), Nobreakthrough upto 240 minutes
Oxaliplatin (2mg/ml) No
breakthrough upto 240 minutesThioTEPA (10 mg/ml),breakthroughdetected at 23.9minutesCarmustine (3.3mg/ml),breakthroughdetected at 14.8minutesOxaliplatin (2mg/ml), Nobreakthrough upto 240 minutesThioTEPA (10mg/ml),breakthroughdetected at 37.1minutesCarmustine (3.3mg/ml),breakthroughdetected at 22.9minutes

{13}------------------------------------------------

Image /page/13/Picture/0 description: The image shows the Halyard logo. The logo consists of a blue circle with a white and green design inside, followed by the word "HALYARD" in large, bold, dark blue letters. Below the word "HALYARD" is the text "Part of the Owens & Minor Family" in a smaller font size.

{14}------------------------------------------------

{15}------------------------------------------------

{16}------------------------------------------------

Image /page/16/Picture/0 description: The image shows the Halyard logo. The logo consists of a blue circle with a white and green shape inside, followed by the word "HALYARD" in blue. Below the word "HALYARD" is the text "Part of the Owens & Minor Family" in a smaller font.

{17}------------------------------------------------

Image /page/17/Picture/0 description: The image contains the Halyard logo. The logo features a blue circle with a white and green shape inside, followed by the word "HALYARD" in blue, block letters. Below the word "HALYARD" is the text "Part of the Owens & Minor Family" in a smaller, lighter font.

{18}------------------------------------------------

Image /page/18/Picture/0 description: The image shows the Halyard logo. The logo consists of a blue circle with a white and green shape inside, followed by the word "HALYARD" in blue. Below the word "HALYARD" is the text "Part of the Owens & Minor Family" in a smaller font size.

Summary of Non-Clinical Data
StandardSubject DeviceHalyard Sterling SG*Nitrile Sensi-GuardPowder-Free ExamGlovesSubject DeviceHalyard SterlingNitrile Powder-FreeExam GlovesAcceptanceCriteriaResults ofSubjectDevices
ASTM D5151-06Standard TestMethod forDetection ofHoles in MedicalGlovesTesting of the subjectdevice shows it meetsthe 2.5% AQLrequirement in thestandards for leakage.The device meets theacceptance criteria ofthe standard.Testing of the subjectdevice shows itmeets the 2.5% AQLrequirement in thestandards forleakage. The devicemeets theacceptance criteriaof the standard.Must meet an AQLof 2.5% forleakage.Pass
ASTM D6124-06Standard TestMethod forResidual Powderon MedicalGlovesResidual powder onthe subject device isan average of 0.03mg/glove within thepowder-free limit of <2 mg maximumpowder per glove andmeets the acceptancecriteria for powder-free.Residual powder onthe subject device isan average of 0.03mg/glove within thepowder-free limit of< 2 mg maximumpowder per gloveand meets theacceptance criteriafor powder- free.Less than 2 mgmaximum powderper glove.Pass
ASTM D6319-10StandardSpecification forNitrileExaminationGloves forMedicalApplicationsAverage TensileStrength (beforeaging) = 35.2 MPaAverage TensileStrength (after aging)= 44.1 MPaAverage Elongation(before aging) = 606%Average Elongation(after aging) = 553%Average TensileStrength (beforeaging) = 37.2 MPaAverage TensileStrength (after aging)= 41.6 MPaAverage Elongation(before aging) =606%Average Elongation(after aging) = 567%Within the limits ofthe standard andthe physicalproperties of thesubject device metthe requirementsfor tensile strengthand elongationbefore and afteraging.Pass
ISO 10993 Biological evaluation of medical devices
SystemicToxicity Study inMice -Saline andSesame OilExtractsNo mortality ofevidence of systemictoxicity from theextracts - PASSNo mortality ofevidence of systemictoxicity from theextracts - PASSNo animals treatedwith test extractedexhibited greaterreaction thancontrol animals.Pass
Dermal IrritationBased on the criteriaand conditions of thestudy, the test articlewas considered non-irritating - PASSBased on the criteriaand conditions of thestudy, the test articlewas considered non-irritating - PASSPrimary IrritationIndex $\leq$ Grade 1Pass
DermalSensitizationBased on the criteriaand conditions of thestudy, the test articleshowed no evidenceof causing delayeddermal contactsensitization - PASSBased on the criteriaand conditions of thestudy, the test articleshowed no evidenceof causing delayeddermal contactsensitization - PASSStimulation Index$\leq$ 3Pass
Conclusion:The conclusions drawn from the nonclinical tests demonstrate that the subjectdevices (Halyard Sterling SG* Nitrile Sensi-Guard Powder-FreeExam Gloves and Halyard STERLING* Nitrile Powder-FreeExam Gloves are as safe, as effective, and performs as well as or better thanthe legally marketed device cleared under K180646.

{19}------------------------------------------------

Image /page/19/Picture/0 description: The image shows the Halyard logo. The logo consists of a blue circle with a white and green shape inside, followed by the word "HALYARD" in blue. Below the word "HALYARD" is the text "Part of the Owens & Minor Family".

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.