K Number
K191217

Validate with FDA (Live)

Device Name
JAZZ PF
Manufacturer
Date Cleared
2019-08-02

(88 days)

Product Code
Regulation Number
888.3010
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The JAZZ systems (including the JAZZ Claw Connector) are temporary implants to be used in orthopedic surgery. The JAZZ systems are intended to provide temporary stabilization as a bone anchor during the development of solid bony fusion and aid in the repair of bone fractures.

The indications for use include the following applications:

  1. Spinal trauma surgery, used in sublaminar or facet wiring techniques;
  2. Spinal reconstructive surgery, incorporated into constructs for the purpose of correction of spinal deformities such as adolescent idiopathic scoliosis, adult scoliosis, kyphosis and spondylolisthesis;
  3. Spinal degenerative surgery, as an adjunct to spinal fusions.

The JAZZ Systems may also be used in conjunction with other medical implants made of titanium alloy or Cobalt-chromium-molybdenum alloy whenever "wiring" may help secure the attachment of other implants.

Device Description

JAZZ PF enables the passing of a JAZZ braid around or through the upper spinous process. The main difference with the JAZZ regular connector is that in the subject device the braid is locked horizontally, that is, along the spine axis, instead of vertically. On each side of the instrumented levels, a JAZZ PF connector is implanted on the union rod below the upper pedicle screw. A JAZZ braid is passed and locked through a 1st connector, above or through the spinous process and back through the second connector on the other side. The slack that can exist in the band is removed using the JAZZ PF Band Reducer (Enrouleur Instrument) without applying any tension in the band and the 2nd connector is locked. By doing so, flexion is controlled which helps in the primary stabilization of the levels where fusion has to be achieved.

AI/ML Overview

The provided text describes a medical device, JAZZ PF, and its substantial equivalence determination by the FDA. However, it does not include detailed information regarding "acceptance criteria" for a study in the context of device performance, nor does it present a study that rigorously proves the device meets specific acceptance criteria in a quantitative sense as might be expected for an AI/CADe device.

Instead, the document focuses on regulatory approval based on substantial equivalence to predicate devices. This means the device is considered safe and effective because its technological characteristics and intended use are similar to those of devices already legally marketed. The performance data mentioned is for mechanical durability testing, not for diagnostic or analytical performance against ground truth.

Therefore, most of the requested information cannot be extracted directly from this document.

Here's what can be provided based on the text:

1. A table of acceptance criteria and the reported device performance

Based on the provided text, the primary "acceptance criterion" for this type of device (spinal implant) in a 510(k) submission is typically demonstrating mechanical durability and safety comparable to predicate devices. The study mentioned is a mechanical test.

Acceptance Criteria (Inferred from 510(k) context for spinal implants)Reported Device Performance (from "Performance Data" section)
Withstand dynamic compression loading for a specified number of cycles.JAZZ PF constructs passed 5 million cycles at a load of 600N.
Demonstrate sufficient rationale for sterilization, cleaning, and biocompatibility.Detailed rationales were provided to demonstrate why the device did not constitute a new worst-case in terms of sterilization, cleaning and biocompatibility.
Substantial equivalence to predicate device.Concluded as substantially equivalent based on similar intended uses, indications, technological characteristics, and principles of operation, and performance data.

2. Sample sized used for the test set and the data provenance

  • Sample Size for Test Set: Not explicitly stated as "test set" in the context of an analytical or diagnostic study. For the dynamic compression testing, the number of devices or constructs tested is not specified, only that "JAZZ PF constructs passed."
  • Data Provenance: Not applicable in the context of clinical data for AI/CADe. The performance data is from mechanical testing of the device itself.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable. Ground truth in the context of a diagnostic or analytical AI study (e.g., expert consensus on medical images) is not relevant for this device's mechanical performance testing.

4. Adjudication method for the test set

Not applicable. There's no adjudication process mentioned for a mechanical test.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No. This is a spinal implant, not an AI/CADe device that assists human readers.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

No. This is a spinal implant, not an algorithm.

7. The type of ground truth used

Not applicable in the context of an AI/CADe device. The "ground truth" for the mechanical test is adherence to engineering specifications for load and cycles.

8. The sample size for the training set

Not applicable. This is not an AI/ML device that requires a training set.

9. How the ground truth for the training set was established

Not applicable.


Summary of Device and Study from Text:

The JAZZ PF is a temporary spinal implant for stabilization during fusion and fracture repair. Its approval (K191217) was based on demonstrating substantial equivalence to existing predicate devices (NuVasive VersaTie System - K173117 and JAZZ System - K182771). The key performance data provided in the submission was a dynamic compression test where JAZZ PF constructs "passed 5 million cycles at a load of 600N," indicating its mechanical durability. The submission also addressed sterilization, cleaning, and biocompatibility by providing rationales that the device did not constitute a new worst-case compared to predicates. The document does not describe a study involving human readers, AI algorithms, or clinical performance against a diagnostic ground truth.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

August 2, 2019

Implanet, S.A. % Janice M. Hogan Partner Hogan Lovells US LLP 1845 Market Street, Suite 230 Philadelphia, Pennsylvania 19103

Re: K191217

Trade/Device Name: JAZZ PF Regulation Number: 21 CFR 888.3010 Regulation Name: Bone fixation cerclage Regulatory Class: Class II Product Code: OWI Dated: May 6, 2019 Received: May 6, 2019

Dear Ms. Hogan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Ronald P. Jean, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 6/30/2020 See PRA Statement on last page

510(k) Number (if known)

K191217

Device Name

JAZZ PF

Indications for Use (Describe)

The JAZZ systems (including the JAZZ Claw Connector) are temporary implants to be used in orthopedic surgery. The JAZZ systems are intended to provide temporary stabilization as a bone anchor during the development of solid bony fusion and aid in the repair of bone fractures.

The indications for use include the following applications:

    1. Spinal trauma surgery, used in sublaminar or facet wiring techniques;
    1. Spinal reconstructive surgery, incorporated into constructs for the purpose of correction of spinal deformities such as adolescent idiopathic scoliosis, adult scoliosis, kyphosis and spondylolisthesis;
  • Spinal degenerative surgery, as an adjunct to spinal fusions. 3.

The JAZZ Systems may also be used in coniunction with other medical implants made of titanium allov or Cobaltchromium-molybdenum alloy whenever "wiring" may help secure the attachment of other implants.

Type of Use (Select one or both, as applicable)

区 Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) SUMMARY JAZZ PF

Submitter

Implanet, S.A. Technopole Bordeaux Montesquieu Allée Francois Magendi 33650 Martillac Phone: (+33) 557 995 555 Facsimile: (+33) 557 995 700

Contact Person: Fabienne Larquey Cadiere

Date Prepared: August 1, 2019

Name of Device: JAZZ PF

Classification Name: Bone fixation cerclage

Regulatory Class: 21 CFR 888.3010

Product Code: OWI

Primary Predicate and Reference Devices

  • . Primary Predicate: NuVasive VersaTie System (K173117)
  • Additional Predicate: ISS-JAZZ Screw System and JAZZ CAP SP (K182771)

Device Description

JAZZ PF enables the passing of a JAZZ braid around or through the upper spinous process. The main difference with the JAZZ regular connector is that in the subject device the braid is locked horizontally, that is, along the spine axis, instead of vertically. On each side of the instrumented levels, a JAZZ PF connector is implanted on the union rod below the upper pedicle screw. A JAZZ braid is passed and locked through a 1st connector, above or through the spinous process and back through the second connector on the other side. The slack that can exist in the band is removed using the JAZZ PF Band Reducer (Enrouleur Instrument) without applying any tension in the band and the 2nd connector is locked. By doing so, flexion is controlled which helps in the primary stabilization of the levels where fusion has to be achieved.

Intended Use / Indications for Use

The JAZZ systems (including the JAZZ Claw Connector) are temporary implants to be used in orthopedic surgery. The JAZZ systems are intended to provide temporary stabilization as a bone anchor during the development of solid bony fusion and aid in the repair of bone fractures.

{4}------------------------------------------------

The indications for use include the following applications:

    1. Spinal trauma surgery, used in sublaminar or facet wiring techniques;
    1. Spinal reconstructive surgery, incorporated into constructs for the purpose of correction of spinal deformities such as adolescent idiopathic scoliosis, adult scoliosis, kyphosis and spondylolisthesis;
    1. Spinal degenerative surgery, as an adjunct to spinal fusions.

The JAZZ Systems may also be used in conjunction with other medical implants made of titanium alloy or Cobalt-chromium-molybdenum alloy whenever "wiring" may help secure the attachment of other implants.

Comparison of Technological Characteristics

The JAZZ PF is an additional component of the JAZZ System that was cleared under K182771. The proposed device uses similar technological characteristics of braided bands as the primary predicate, VersaTie System, to stabilize the spine. Specifically, the subject device, similar to the NuVasive VersaTie System, is secured around posterior vertebral structures such as the lamina, facet, transverse processes, and spinous process from T1-L5. The system is comprised of braided bands and clamps designed to attach to c rods.

The following table compares the technological characteristics of the JAZZ PF with the predicate devices:

Subject DeviceJAZZ PFPrimary PredicateDeviceNuVasive VersaTieSystemAdditional PredicateDeviceJAZZ SystemComparison
510(k)SubjectK173117K182771
Braid ConnectorYesYesYesSame
BraidYesYesYesSame
MaterialsTi alloy, polyesterTi or CoCr alloy,polyesterTi alloy, polyesterSimilar biocompatiblematerials
Connection Methodto SpineBraid, rods and pediclescrewsBraid, rod and pediclescrewsBraid, rod, and pediclescrewsSame
Braid LockingMethodAs internal screw istightened, the connectorclamps the braidAs internal screw istightened, the connectorclamps the braidAs internal screw istightened, the connectorclamps the braidSame
Sizes5.5-6mm5-6mm5.5-6mmSimilar range of sizes

Performance Data

Dynamic compression testing of the subject device was performed on a lumbar construct mounted on a test block and attached to the fixed test machine jaw. The JAZZ PF constructs passed 5 million cycles at a load of 600N. Detailed rationales were provided to demonstrate why the device did not constitute a new worst-case in terms of sterilization, cleaning and biocompatibility.

{5}------------------------------------------------

Conclusions

The JAZZ PF is substantially equivalent to predicate devices. The subject devices have the same intended uses and similar indications, technological characteristics, and principles of operation as the predicate devices. In addition, the minor technological differences between the subject device and the predicate devices raise no new issues of safety or effectiveness. In particular, the JAZZ PF provides an alternate way to allow for the braid attachment to the spine, similar to how the ISS-JAZZ Screw System and the JAZZ CAP SP System (cleared under K182771) provides an additional method for stabilization. In particular, the NuVasive VersaTie System (K173117) uses similar technological characteristics of braided bands and clamps in the same manner as the subject device around the same intended anatomical location. Performance data demonstrate that the JAZZ PF is as safe and effective as the predicate devices. Thus, the JAZZ PF is substantially equivalent to the predicate devices.

N/A