K Number
K191181

Validate with FDA (Live)

Date Cleared
2019-10-11

(162 days)

Product Code
Regulation Number
890.5850
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Neux NXPRO Neuromuscular Electrical Stimulation Device is intended for the stimulation of healthy muscles in order to improve or facilitate muscle performance.

The Neux NXPRO is not intended to be used in conjunction with therapy or treatment of medical diseases or medical conditions of any kind. The NeuX NXPRO is not designed for use on injured or ailing muscles and its use on such muscles is contraindicated.

The NeuX NXPRO's electrical impulses allow triggering action potential on motor nerves (excitations). These excitations of motor neurons are transmitted to the motor endplate where they generate mechanical muscle fiber responses that correspond to muscle work. Depending on the parameters of the electrical impulses (pulse frequency, duration of contraction, duration of rest, & total session duration), different types of muscle work can be imposed on the stimulated muscles.

The various types of muscle work that the NeuX NXPRPO can impose on the stimulated muscles are able to improve or facilitate muscle performance. The NeuX NXPRO is considered a technique of muscle training.

Device Description

The NeuX NXPRO Neuromuscular Electrical Stimulation Device generates electrical impulses the result in a triggering action potential on motor neurons of motor nerves (excitations). These excitations of motor neurons are transmitted to the muscle fibers through the system electrodes where they generate mechanical muscle fiber responses that correspond to muscle work. Depending on the parameters of the electrical impulses: (a) pulse frequency; (b) duration of muscle contraction; (c) duration of rest; and (d) total session duration; different types of muscle work can be imposed on the stimulated muscles. The various types of muscle work that the NXPRO can impose on the stimulated muscles are able to improve or facilitate muscle performance.

Note: the application of the NXPRO is considered a technique of muscle training.

AI/ML Overview

The provided text is a 510(k) summary for the NeuX NXPRO Neuromuscular Electrical Stimulation Device, seeking substantial equivalence to a predicate device (ARP Manufacturing – ARP Sport Device). This type of submission focuses on demonstrating that the new device is as safe and effective as a legally marketed predicate device, rather than proving efficacy through clinical outcome studies with predefined acceptance criteria for AI or diagnostic performance metrics.

Therefore, the document does not contain the information requested regarding acceptance criteria related to AI/diagnostic performance, study design (sample size for test set, data provenance, number of experts, adjudication methods, MRMC studies, standalone performance), or ground truth establishment for AI model training.

The document primarily discusses:

  • Device Description and Intended Use: The NXPRO stimulates healthy muscles for performance improvement and is not intended for treatment of medical conditions or use on injured muscles.
  • Technological Characteristics: Details on waveform, output specifications, environmental conditions, and compliance with electrical and mechanical safety standards (IEC 60601 series).
  • Comparison to Predicate Device: A detailed table comparing the NXPRO with the predicate ARP Sport device, highlighting similarities in product code, regulation, device class, intended use, and general electrical properties. Many specific output characteristics of the predicate device are listed as "Not Publically Available."
  • Non-Clinical Performance Testing: States that bench testing demonstrated the NXPRO provided the "same performance characteristics as the predicate device" and successfully passed all tests required by IEC 60601-1, 3rd Edition.
  • Non-clinical Safety Tests: Lists compliance with various IEC 60601 standards for electrical and mechanical safety.
  • Conclusion: The device is substantially equivalent to the predicate because its intended use is the same, technological differences do not raise new safety or effectiveness questions, and performance testing (bench testing) demonstrates it is as safe and effective.

No information is provided on:

  1. A table of acceptance criteria and reported device performance related to diagnostic accuracy or AI metrics. The "acceptance criteria" here are implicitly related to meeting safety standards and demonstrating comparable performance to the predicate via electrical output measurements, not diagnostic outcomes.
  2. Sample size used for a "test set" in the context of an AI study. The testing mentioned is bench testing for electrical characteristics.
  3. Data provenance or retrospective/prospective nature of a clinical study.
  4. Number or qualifications of experts used for ground truth (as there's no diagnostic AI component).
  5. Adjudication method.
  6. MRMC comparative effectiveness study or human reader improvement.
  7. Standalone (algorithm only) performance.
  8. Type of ground truth (beyond meeting electrical safety standards).
  9. Training set sample size or ground truth establishment for a training set (as no AI model training is described).

This document describes a regulatory filing for a physical medical device, not a software as a medical device (SaMD) or an AI-powered diagnostic tool, which would typically involve the requested performance study details.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a sans-serif font.

October 11, 2019

NeuX Technologies, Inc. % Christopher Devine, PhD President Devine Guidance International, Inc. 4730 South Fort Apache Road, Suite 300 Las Vegas, Nevada 89147

Re: K191181

Trade/Device Name: NXPRO Neuromuscular Electrical Stimulation Device Regulation Number: 21 CFR 890.5850 Regulation Name: Powered Muscle Stimulator Regulatory Class: Class II Product Code: NGX Dated: July 12, 2019 Received: July 17, 2019

Dear Dr. Christopher Devine:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For Vivek Pinto, Ph.D. Director (Acting) DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K191181

Device Name

NXPRO Neuromuscular Electrical Stimulation Device

Indications for Use (Describe)

The Neux NXPRO Neuromuscular Electrical Stimulation Device is intended for the stimulation of healthy muscles in order to improve or facilitate muscle performance.

The Neux NXPRO is not intended to be used in conjunction with therapy or treatment of medical diseases or medical conditions of any kind. The NeuX NXPRO is not designed for use on injured or ailing muscles and its use on such muscles is contraindicated.

The NeuX NXPRO's electrical impulses allow triggering action potential on motor nerves (excitations). These excitations of motor neurons are transmitted to the motor endplate where they generate mechanical muscle fiber responses that correspond to muscle work. Depending on the parameters of the electrical impulses (pulse frequency, duration of contraction, duration of rest, & total session duration), different types of muscle work can be imposed on the stimulated muscles.

The various types of muscle work that the NeuX NXPRPO can impose on the stimulated muscles are able to improve or facilitate muscle performance. The NeuX NXPRO is considered a technique of muscle training.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Section Five (5) - 510(k) Summary

510(k) SUMMARY

[As Required by 21 CFR 807.92(c)]

Submitter's Name & Address:NeuX Technologies, Inc.3802 Spectrum Blvd.Suite 112FTampa, FL 33612
Contact Person:Christopher J. Devine, Ph.D.Telephone (702) 939-5507Mobile (702) 917-0585Chris.devine@devineguidanceinternational.com
Date Summary Prepared:11 October 2019
Device Name:Trade/Proprietary Name - NXPRO NeuromuscularElectrical Stimulation Device
Common/Usual Name – Powered Muscle Stimulator
Classification Name - Powered Muscle Stimulator(890.5850), Product Code NGX
Predicate Device – ARP Manufacturing – ARP SportDevice (K093999)
Reference Device - N/A

{4}------------------------------------------------

Device Description

The NeuX NXPRO Neuromuscular Electrical Stimulation Device generates electrical impulses the result in a triggering action potential on motor neurons of motor nerves (excitations). These excitations of motor neurons are transmitted to the muscle fibers through the system electrodes where they generate mechanical muscle fiber responses that correspond to muscle work. Depending on the parameters of the electrical impulses: (a) pulse frequency; (b) duration of muscle contraction; (c) duration of rest; and (d) total session duration; different types of muscle work can be imposed on the stimulated muscles. The various types of muscle work that the NXPRO can impose on the stimulated muscles are able to improve or facilitate muscle performance.

Note: the application of the NXPRO is considered a technique of muscle training.

Intended Use

The NeuX NXPRO Neuromuscular Electrical Stimulation Device is intended for the stimulation of healthy muscles in order to improve or facilitate muscle performance.

The NeuX NXPRO is not intended to be used in conjunction with therapy or treatment of medical diseases or medical conditions of any kind. The NeuX NXPRO is not designed for use on injured or ailing muscles and its use on such muscles is contraindicated.

The NeuX NXPRO's electrical impulses allow triggering action potential on motor neurons of motor nerves (excitations). These excitations of motor neurons are transmitted to the muscle fibers via the motor endplate where they generate mechanical muscle fiber responses that correspond to muscle work. Depending on the parameters of the electrical impulses (pulse frequency, duration of contraction, duration of rest, & total session duration), different types of muscle work can be imposed on the stimulated muscles.

{5}------------------------------------------------

The various types of muscle work that the NeuX NXPRPO can impose on the stimulated muscles are able to improve or facilitate muscle performance. The NeuX NXPRO is considered a technique of muscle training.

Technological Characteristics/Principles of Operation

Technological Characteristics

Storage and Transportation Conditions

The NXPRO may be used either indoors or outdoors. However, if used outdoors, the NXPRO should not be exposed to rain, snow, condensing fog or condensing humidity, and should not be subjected to long-term exposure to direct sunlight.

Storage Environmental Conditions:

    • 10°C to +60°C
  • +5°C to +35°C at a relative humidity up to 90%, non-condensing ●
  • +35°C to 70°C at a water vapor pressure up to 50 hPa .

Operating Environmental Conditions:

  • · +5°C to +40°C
  • Relative humidity range of 15% to 90%. non-condensing, but not requiring a . water vapor partial pressure greater than 50 hPa
  • Atmospheric pressure range of 700 hPa to 1060 hPa

Disposal:

  • . Electronics must be disposed of in compliance with relevant national regulatory requirements.
    Note: the NeuX NXPRO device and components used in the manufacture of the NXPRO are RoHS compliant.

Standards:

{6}------------------------------------------------

  • . The NXPRO complies with current medical standards.
  • The NXPRO also complies with the IEC 60601-1 standard on general safety . requirements for electro-medical devices, IEC 60601-1-11 general requirements for basic safety and essential performance, ty requirements for electro-medical devices, the IEC 60601-1-2 standard on electromagnetic compatibility, and the IEC 60601-2-10 standard on particular safety requirements for nerve and muscle stimulators.

Output Waveform:

  • A continuous, medium amplitude, sensory waveform, periodically broken by high . amplitude, rectangular stimulation pulses. Transformer isolation ensures that no direct current (DC) is present in the outputs.
  • Channels: 2 channels are electrically isolated from each other and floating. .

Output Specifications:

  • Main Pulse duration 400us .
  • . Main Pulse Shape: Rising and falling Exponential (RC) with a time constant of 50 microseconds
  • Main Pulse Frequency 50 to 250 pulses per second (manually adjustable) ●
  • Maximum Main Pulse Intensity: 75V peak-to-peak with 500-Ohm load
  • Waveform Ramp Up Time: 2.0 seconds ●
  • Waveform Ramp Down Time: 1.0 seconds
  • Output Power - 0.707 watts RMS Max with 500-Ohm load
  • Polarity Waveform polarity can be inverted .
  • Sensory Waveform Shape: Exponential (RC) ●
  • Sensory Waveform Frequency 10kHz (fixed) ●
  • Maximum Sensory Waveform Intensity: 14.0V peak-to-peak with 500-Ohm load .

Unit Characteristics (without the stand):

  • Body: Plastic .
  • Weight: 1.59kg / 3.5lb .

510(k) Summary (NeuX – NXPRO Device)

{7}------------------------------------------------

  • . Length: 240mm / 9.45in
  • . Width: 160mm / 6.30in
  • Height: 40mm / 1.57in ●

Power Supply

  • +12Vdc, 3A, AC to DC power supply. P/N: PEAMD36-12-B2 .

Electromagnetic Compatibility (EMC)

The NXPRO has been tested for EMC Class B emissions per IEC 60601-1-2:2014. The NXPRO is suitable for use in any establishment, including a private dwelling and a place connected directly to the low voltage mains, which powers residential buildings. Please observe the following precautions when operating the NXPRO.

  • Operation in close proximity to shortwave or microwave therapy may produce instability . in the output of the device.
  • . Simultaneous connection to high-frequency surgical equipment may result in burns at the site of stimulator electrodes and possible damage to the device.
  • . To ensure proper use and to mitigate the possibility of interference, avoid placing in close proximity to other electromagnetic devices.
Load ResistanceOutput VoltageWatt RMS toLoad
500 ohm18.8 Vrms0.707W
2 Kohm33.7 Vrms0.568W
10 Kohm45.5 Vrms0.207W

Maximum Output Voltage & Power vs. Load Resistance

Principles of Operation

The principle of electrostimulation is to stimulate nerve fibers by means of electrical impulses transmitted by electrodes. Neuromuscular Electrical Stimulation (NMES) is applied to the muscles and specifically targets the motor nerve with electrical pulses that can activate the junction between the nerves and muscles being treated and produce a contraction of the muscle fibers. A stimulation-induced contraction can give an effective workout for the muscle. The electrical pulses generated by the NXPRO stimulator are high quality pulses - offering safety, 510(k) Summary (NeuX – NXPRO Device)

{8}------------------------------------------------

comfort and efficiency. The quantity and the benefits obtained depend on the stimulation parameters.

Non-Clinical Performance Testing

Performance testing consisted of bench testing that has demonstrated that the output of

the NeuX NXPRO Neuromuscular Electrical Stimulation Device provided the same

performance characteristics as the predicate device. The system successfully passed all tests

required by IEC 60601-1, 3td Edition.

Non-clinical Safety Tests

The NeuX NXPRO Neuromuscular Electrical Stimulation Device has been designed and

constructed to meet the following electrical and mechanical safety standards:

  • IEC 60601-1: Medical electrical equipment Part 1: General requirements o for basic safety and essential performance (3td edition)
  • IEC 60601-1-2: Medical electrical equipment Part 1-2: General o requirements for safety and essential performance - Collateral Standard: Electromagnetic disturbances - requirements and tests
  • IEC 60601-1-11: Medical electrical equipment Part 1-11: o General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
  • IEC 60601-2-10: Medical equipment Part 2-10: O Particular requirements for the basic safety and essential performance of nerve and muscle stimulators

There are no significant functional differences between the NXPRO Neuromuscular Electrical Stimulation Device and the ARP POV Sport Device. In accordance with FDA recommended guidance, NeuX Technologies has provided a detailed comparison of the NXPRO

{9}------------------------------------------------

Neuromuscular Electrical Stimulation Device versus the previously cleared ARP POV Sport Device in Tables 5.1 and 5.2.

Device NameNXPRO DevicePredicateARP SportEquivalencyDiscussion
510(k) Number-K093999-
Device Name, ModelNXPROPOV Sport-
ManufacturerNeuX TechnologiesArp Manufacturing-
Product CodeNGXNGXIdentical
Regulation890.5850890.5850Identical
Device Class22Identical
Indications For UseThe NeuX NXPRONeuromuscularElectrical StimulationDevice is intended forthe stimulation ofhealthy muscles in orderto improve or facilitatemuscle performance.The NeuX NXPRO is notintended to be used inconjunction withtherapy or treatment ofmedical diseases ormedical conditions ofany kind. The NeuXNXPRO is not designedfor use on injured orailing muscles and itsuse on such muscles iscontraindicated.The NeuX NXPRO'selectrical impulses allowtriggering actionpotential on motorThe ARP POV Sport isintended to stimulatehealthy muscles inorder to improve orfacilitate muscleperformance.The ARP POV Sport isnot intended to be usedin conjunction withtherapy or treatment ofmedical diseases ormedical conditions ofany kind. The ARP POVSport is not designedfor use on injured orailing muscles and itsuse on such muscles iscontraindicated.The POV Sport'selectrical impulsesallow triggering actionpotential on motorneurons ofEquivalent
Device NameNXPRO DevicePredicateARP SportEquivalencyDiscussion
neurons of motorMotor nerves
nerves (excitations).(excitations). These
These excitations ofexcitations of motor
motor neurons areneurons are
transmitted to thetransmitted to the
muscle fibers via themuscle fibers via the
motor endplate wheremotor endplate where
they generatethey generate
mechanical muscle fiber mechanical muscle fiber
responses thatresponses that
correspond to musclecorrespond to muscle
work. Depending on the work. Depending on the
parameters of theparameters of the
electrical impulseselectrical impulses
(pulse frequency,(pulse frequency,
duration of contraction, duration of contraction,
duration of rest, & total duration of rest, total
session duration),session duration),
different types ofdifferent types of
muscle work can bemuscle work can be
imposed on theimposed on the
stimulated muscles.stimulated muscles.
The various types ofThe various types of
muscle work that themuscle work that the
NeuX NXPRPO canimpose on thePOV Sport can imposeon the stimulated
stimulated muscles aremuscles are able to
able to improve orimprove or facilitate
facilitate musclemuscle performance.
performance. The NeuX The POV Sport is
NXPRO is considered aconsidered a technique
technique of muscletraining.of muscle training.
Power Source(s)External DC PowerSupplyNot PublicallyAvailableEquivalent
- Method of Line Current IsolationPower Supply IsolationNot PublicallyAvailableIdentical
- Patient Leakage Current<100μANot PublicallyAvailableIdentical
- Normal condition<100μANot PublicallyIdentical
Device NameNXPRO DevicePredicateARP SportEquivalencyDiscussion
Available
- Single fault condition$<100μA$Not PublicallyAvailableIdentical
Number of Output Modes2Not PublicallyAvailableIdentical
Number of Output Channels2Not PublicallyAvailableIdentical
- Synchronous or Alternating?SynchronousNot PublicallyAvailableIdentical
- Method of Channel IsolationIsolation TransformerNot PublicallyAvailableIdentical
Regulated Current or RegulatedVoltage?VoltageNot PublicallyAvailableIdentical
Software/Firmware/MicroprocessorControl?YesNot PublicallyAvailableEquivalent
Automatic Overload Trip?YesNot PublicallyAvailableIdentical
Automatic No-Load Trip?NoNot PublicallyAvailableIdentical
Automatic Shut Off?YesNot PublicallyAvailableEquivalent
Patient Override Control?YesNot PublicallyAvailableEquivalent
Indicator Display:YesNot PublicallyAvailableEquivalent
- On/Off Status?YesNot PublicallyAvailableEquivalent
- Low Battery?N/A (No Batt)Not PublicallyAvailableNXPRO does nothave a battery
- Voltage/Current Level?Intensity Level(0 to 100%)Not PublicallyAvailableEquivalent
Timer Range (minutes)1 to 20 minutesNot PublicallyAvailableEquivalent
Device NameNXPRO DevicePredicateARP SportEquivalencyDiscussion
Compliance with VoluntaryStandards?IEC 60601-1 (3rd Ed)IEC 60601-1-11IEC 60601-2-10EN ISO 14971Not PublicallyAvailableEquivalentNote: IEC 60601-1--was withdrawn (IEC60601-1-11 is thereplacementstandard)
Compliance* with 21 CFR 898?(*Becomes mandatory beginningMay 9, 2000)YesNot PublicallyAvailableIdentical
Weight3.0lbNot PublicallyAvailableEquivalent
Dimensions (in.) [W x H x D]10"x7"x2"Not PublicallyAvailableEquivalent
Housing Materials and ConstructionPlastic InjectionMoldingNot PublicallyAvailableIdentical

Table 5.1 - Comparison (Basic Device Characteristics)

{10}------------------------------------------------

{11}------------------------------------------------

{12}------------------------------------------------

Table 5.2 – Comparison (Output Specifications)

Device NameNXPRO DevicePredicateARP SportEquivalencyDiscussion
Waveform (e.g., pulsed monophasic,biphasic)Asymmetric biphasicNot PublicallyAvailableIdentical
Shape (e.g., rectangular, spike,rectified sinusoidal)RectangularNot PublicallyAvailableIdentical
Maximum Output Voltage (specifyunits)18.8Vrms @ 500 ΩNot PublicallyAvailableEquivalent
(+/- 5%)33.7Vrms @ 2 kΩNot PublicallyAvailableEquivalent
45.5Vrms @ 10 kΩNot PublicallyAvailableEquivalent
Maximum Output Current (specifyunits)37.6 mArms @ 500 ΩNot PublicallyAvailableEquivalent
(+/- 5%)16.9 mArms @ 2 kΩNot PublicallyAvailableEquivalent
Device NameNXPRO DevicePredicateARP SportEquivalencyDiscussion
4.6 mArms @ 10 kΩNot PublicallyAvailableEquivalent
Pulse Width (specify units)314 microsecondsNot PublicallyAvailableEquivalent
Frequency (Hz)Main: 50-250pps.Background: 10kHzNot PublicallyAvailableEquivalent
For interferential modes only:N/ANot PublicallyAvailable-
- Beat Frequency (Hz)N/ANot PublicallyAvailable-
For multiphasic waveforms only:
- Symmetrical phases?NoNot PublicallyAvailableIdentical
- Phase Duration (include units)(state range, if applicable)(both phases, if asymmetrical)N/ANot PublicallyAvailable-
Net Charge (mC per pulse) (If zero,state method of achieving zero netcharge.)0 mC @ 500 ΩOutput is AC coupledvia isolationtransformers to resultin zero net charge.Not PublicallyAvailableEquivalent
Maximum Phase Charge, (mC)11.8 uC @ 500 ΩNot PublicallyAvailableEquivalent
Maximum Current Density,(mA/cm²)1.46 mA/cm² @ 500ΩNot PublicallyAvailableEquivalent
Maximum Power Density, (W/cm²)(using smallest electrode conductivesurface area)2.15 mW/cm² @ 500ΩNot PublicallyAvailableEquivalent
Burst Mode (i.e., pulse trains)a. Pulses per burstb. Bursts per secondc. Burst duration (seconds)d. Duty Cycle [Line (b) x Line (c)]'Intermittent' Mode:50 to 250 pulses/burst1 burst/second0 to 60 second duration0 to 100%Not PublicallyAvailableEquivalent
ON Time (seconds)0 to 60 secondsNot PublicallyAvailableEquivalent
OFF Time (seconds)0 to 60 secondsNot PublicallyAvailableEquivalent
Device NameNXPRO DevicePredicateARP SportEquivalencyDiscussion
Available
Additional Features (if applicable)N/ANot PublicallyAvailable-

{13}------------------------------------------------

{14}------------------------------------------------

Conclusion Statement

The NeuX NXPRO – Neuromuscular Electrical Stimulation Device has the same intended use as the predicate device (ARP POV Sport). Any minor technological differences to the device do not raise new questions of safety or effectiveness. Performance testing, along with verification and validation activities demonstrate that the NeuX NXPRO – Neuromuscular Electrical Stimulation Device is as safe and effective in its intended use, and performs as well as the predicate device. Therefore, the NeuX NXPRO – Neuromuscular Electrical Stimulation Device can be considered substantially equivalent to the ARP Manufacturing POV Sport Device.

§ 890.5850 Powered muscle stimulator.

(a)
Identification. A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.(b)
Classification. Class II (performance standards).