K Number
K191179

Validate with FDA (Live)

Device Name
Myo Plus
Date Cleared
2019-09-04

(125 days)

Product Code
Regulation Number
882.1320
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Myo Plus is to be used exclusively for exoprosthetic fittings of the upper limbs.

Device Description

MYO PLUS measures the electromyographic control signals of the user and assigns them to the movements of the prosthesis, such as the hand or rotation unit. The product is the central control unit of a myoelectric prosthesis. By calibrating via the Myo Plus App, the control unit learns to assign muscle signals from the user to the various types of movement. This calibration can be performed by the users themselves, and repeated at regular intervals if necessary.

Description of the components:
Myo Plus TR
The central control unit is installed in the forearm of the prosthesis and is supplied with energy by the battery system of the prosthesis. Up to 8 remote electrodes can be connected. The electronics handles the signal processing of the electrode signals and controls the compatible prosthesis components.

Remote Electrode
The remote electrodes take the muscle signal from the skin surface and transmit it to MyoPlus TR. The electrode contacts are located on a flexible cable and are individually screwed into the socket by the prosthesist.

Myo Plus App
The Myo Plus App is the central interface for the user and/or professionals to make settings on the control system or monitor operating states. The data transfer to mobile setting systems on Android or mobile iOS devices shall be carried out by a wireless connection according to the state of the art.

Mvo Cuff
The Myo Cuff is used to evaluate the muscle signals of a potential user of a Myo Plus controller. The Myo Cuff forms a flexible wristband which is temporarily applied to the user's forearm. Thus, the signal patterns of the user can be evaluated with the help of the app without the preparation of a complex test socket. If a user has problems separating the patterns at the beginning, he can also use the Myo Cuff for practicing with an professional.

Myo Plus trial kit
The Myo Plus trial kit provides the components of Myo Plus as a loaner kit for the prosthesist.

AI/ML Overview

The provided text describes the Myo Plus device and its substantial equivalence to a predicate device, but it does not contain specific details about acceptance criteria or a study that proves the device meets those criteria in the context of a comparative effectiveness study involving AI assistance for human readers or standalone algorithm performance.

The document primarily focuses on demonstrating substantial equivalence based on:

  • Identical Indications for Use
  • Identical Principles of Operation
  • Similar Technological Characteristics
  • Compliance with recognized safety and performance standards (electrical safety, EMC, biocompatibility, battery safety)
  • Results from general functional and performance tests.

There is no mention of an AI component, human readers, effect sizes, multi-reader multi-case studies, or the types of ground truth typically established for AI-driven diagnostic or assistive devices. The "Pattern recognition" mentioned refers to the device's ability to interpret muscle signals for prosthetic control, not image analysis or AI interpretation of medical data in a diagnostic context.

Therefore, I cannot extract the requested information from the provided text for the following reasons:

  1. Acceptance criteria and reported device performance related to AI/human reader studies: Not present. The acceptance criteria mentioned are related to functional performance (e.g., "Functional testing: Passed," "Pattern recognition: Passed") and compliance with safety standards.
  2. Sample size for test set and data provenance: Not mentioned.
  3. Number of experts and their qualifications for ground truth: Not mentioned.
  4. Adjudication method for test set: Not mentioned.
  5. MRMC comparative effectiveness study and effect size: No such study is described.
  6. Standalone (algorithm only) performance: No such performance evaluation is described. Myo Plus is a physical device (control unit for prosthetic limbs) that uses signal processing, not a standalone AI algorithm for medical image analysis.
  7. Type of ground truth used (expert consensus, pathology, outcomes data, etc.): Ground truth for the device's function would be the user's intended muscle movements and the device's accurate interpretation and execution of those movements, but this is not defined in terms of a clinical study or labeled dataset.
  8. Training set sample size and how ground truth was established for training: Not mentioned. The device "learns to assign muscle signals from the user to the various types of movement" through a calibration process via the Myo Plus App, which can be done by users themselves or with a professional. This is a user-specific calibration, not a large-scale AI model training process using a distinct training dataset.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the FDA acronym in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.

September 27, 2019

Otto Bock Healthcare Products GmbH % Rene Urtz Regulatory Affairs Manager Otto Bock Healthcare Products GmbH Brehmstrabe 16 1110 Wien Vienna, AT

Re: K191179

Trade/Device Name: Myo Plus Regulation Number: 21 CFR 882.1320 Regulation Name: Cutaneous electrode Regulatory Class: Class II Product Code: GXY, IQZ Dated: August 6, 2019 Received: August 9, 2019

Dear Rene Urtz:

This letter corrects our substantially equivalent letter of September 4, 2019.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of

{1}------------------------------------------------

Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Heather L. Dean -S

For: Vivek Pinto, Ph.D. Acting Division Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K191179

Device Name Myo Plus

Indications for Use (Describe)

Myo Plus is to be used exclusively for exoprosthetic fittings of the upper limbs.

Type of Use (Select one or both, as applicable)

Construction (Permit, SEP, NOI, Other, Please Specify) Post-Construction (Permit, SEP, NOI, Other, Please Specify)
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image: ottobock logoPREMARKETNOTIFICATIONMYO PLUSProject Report No.PD-PP00120325A-200Attachment XPage1 of 4
---------------------------------------------------------------------------------------------------------------------------------------

510(K) SUMMARY

1 Submitter Information

510(k) owner:Otto Bock Healthcare Products GmbH
Establishment Registration:3005190268
Address:Brehmstraße 161110 ViennaAustria
Contact Person:Rene UrtzRegulatory Affairs Managerrene.urtz@ottobock.com
Telephone:+43 1 523 37 86 707
Fax Number:+43 1 523 22 64
Date Prepared:September 03, 2019

2 Device Information

Trade Name:Myo Plus
Common or Usual Name:Powered, External Upper Limb Prosthetic System
Classification Name:Cutaneous Electrode (21 CFR 882.1320)
Classification Product Code:GXY (Electrode, Cutaneous)
Subsequent Product Code:IQZ (Hand, External Limb Component, Powered)

3 Predicate Devices

510(k) Number:K182112
Device Name:Sense System with IBT Electrodes

4 Indications for Use

Myo Plus is to be used exclusively for exoprosthetic fittings of the upper limbs.

{4}------------------------------------------------

ottobock.PREMARKETNOTIFICATIONMYO PLUSProject Report No.PD-PP00120325A-200Attachment XPage2 of 4
----------------------------------------------------------------------------------------------------------------------------

5 Device Description

MYO PLUS measures the electromyographic control signals of the user and assigns them to the movements of the prosthesis, such as the hand or rotation unit. The product is the central control unit of a myoelectric prosthesis. By calibrating via the Myo Plus App, the control unit learns to assign muscle signals from the user to the various types of movement. This calibration can be performed by the users themselves, and repeated at regular intervals if necessary.

Description of the components:

Myo Plus TR

The central control unit is installed in the forearm of the prosthesis and is supplied with energy by the battery system of the prosthesis. Up to 8 remote electrodes can be connected. The electronics handles the signal processing of the electrode signals and controls the compatible prosthesis components.

Remote Electrode

The remote electrodes take the muscle signal from the skin surface and transmit it to MyoPlus TR. The electrode contacts are located on a flexible cable and are individually screwed into the socket by the prosthesist.

Myo Plus App

The Myo Plus App is the central interface for the user and/or professionals to make settings on the control system or monitor operating states. The data transfer to mobile setting systems on Android or mobile iOS devices shall be carried out by a wireless connection according to the state of the art.

Mvo Cuff

The Myo Cuff is used to evaluate the muscle signals of a potential user of a Myo Plus controller. The Myo Cuff forms a flexible wristband which is temporarily applied to the user's forearm. Thus, the signal patterns of the user can be evaluated with the help of the app without the preparation of a complex test socket. If a user has problems separating the patterns at the beginning, he can also use the Myo Cuff for practicing with an professional.

Myo Plus trial kit

The Myo Plus trial kit provides the components of Myo Plus as a loaner kit for the prosthesist.

{5}------------------------------------------------

PREMARKETProject Report No.Page
ottobock.NOTIFICATIONMYO PLUSPD-PP00120325A-200Attachment X3 of 4

6 Substantial Equivalence

The following table compares Myo Plus to the predicate device with respect to indications for use, principle of operation, technological characteristics and safety and performance testing.

Myo PlusPredicate deviceComparison
510k numberTo be determinedK182112
Device name, ModelMyo PlusSense System withIBT Electrodes
ManufacturerOtto Bock HealthCareProducts GmbHInfinite BiomedicalTechnologies, LLC
Classification, Indications for Use and Intended Use
Classification ProductCodeGXYGXYSame
Subsequent ProductCodeIQZIQZSame
Indications for UseMyo Plus is to be usedexclusively for exoprostheticfittings of the upper limbs.Sense System with IBTElectrodes is to be usedexclusively for externalprosthetic fittings of upperlimbs.Same
Principle of operationDetect, process, andtransmit physiologicalsignals for use with aprosthesis.Detect, process, andtransmit physiologicalsignals for use with aprosthesis.Same
Environment of UseProfessional healthcarefacility and home useProfessional healthcarefacility and home useSame
Assembling procedureComponents are assembledby a prosthetist.Components are assembledby a prosthetist.Same
Technological Characteristics - System
Signal acquisitionEMG electrodeEMG electrodeSame
Adjustment softwareYesMyo Plus AppYesUser Interface ApplicationSame
Software/Firmware/Microprocessor Control?YesYesSame
Input voltage5,5-12 VDC5 to 10 VDC (system)Similar
Output signal0 - 4,5V digital and 0-8.2Vmotor (analog)0-3.3Vdigital and 0-7.4VmotorSimilar
Terminal device (e.g.hand, wrist or elbow)included?NoNoSame
Wireless CommunicationYesBluetoothYesBluetoothSame
Component for trainingand capabilitiesassessmentYesMyo CuffNoDiffers
Power source included?Included in componentMyoCuffNoDiffers
Technological Characteristics - Processing unit
Dimensions67 x 27 x 9.2 mm59 x 27.8 x 9.8 mmSimilar
Control OptionsPattern recognition,State switchingPattern recognitionSimilar
Input ButtonNoneNoneSame

{6}------------------------------------------------

PREMARKETNOTIFICATIONMYO PLUSProject Report No.PD-PP00120325A-200Attachment XPage4 of 4
Image: ottobock logo
Technological Characteristics - Electrode
Dimensions40.8 x 13.8 x 5.3 mm28.8 x 16.8 x 6.7mmSimilar
Temperature range (use)+5°C to +40°C (41°F to104°F)-10°C to +50°C (14°F to122°F)Similar
Housing materialTECHNOMELT PA 638BLACKPolyamidPlastics( ABS/PC Blend)Polycarbonate/AcrylonitrileButadien StyrolDiffers
Contact areaTitanium (grade 2)Titanium (grade 1)Similar
Signal processingDigitalDigitalSame
Frequency bandwidth80 - 500 Hz90-500HzSimilar
AdjustmentnoneDigital gain 1-7Differs
InstallationSuspension arms/suctionsocketSuspension arms/suctionsocketSame
Safety and Performance Testing
Electrical SafetyIEC 60601-1:2005/A1:2012IEC 60601-1-11:2015IEC 60601-1:2005/A1:2012IEC 60601-1-11:2010Similar
ElectromagneticCompatibilityIEC 60601-1-2:2012IEC 60601-1-2:2007Similar
BiocompatibilityISO 10993-1:2009ISO 10993-1:2009Same
Battery SafetyIEC 62133-2:2017ST/SG/AC.10/11/Rev.5/Amend.2, Sub-section 38.3IEC 62133:2002ST/SG/AC.10/11/Rev.5/Amend.2, Sub-section 38.3Similar

The following table shows a brief summary of additional non-clinical performance testing conducted to support substantial equivalence.

TestResult
Functional testingPassed
Adaptation to the user through calibrationPassed
Commissioning by userPassed
Pattern recognitionPassed
Wireless data transmissionPassed
Product compatibilityPassed
Product lifetimePassed
Battery lifePassed

Myo Plus passed all the safety and performance testing in accordance with internal requirements and international standards, including those shown herein to demonstrate its ability to achieve its intended use and to support the substantial equivalence.

7 Conclusions

Myo Plus has the same indications for use, the same principle of operation and similar technological characteristics as the predicated device cleared under 510(k) no. K182112.

Based on the supporting data and the discussion provided herein, Otto Bock Healthcare GmbH considers its Myo Plus to be substantially equivalent to the predicate device.

§ 882.1320 Cutaneous electrode.

(a)
Identification. A cutaneous electrode is an electrode that is applied directly to a patient's skin either to record physiological signals (e.g., the electroencephalogram) or to apply electrical stimulation.(b)
Classification. Class II (performance standards).