Intended Use

MONTAGE-QS Settable, Resorbable Hemostatic Bone Putty is indicated for the control of bleeding from cut or damaged bone by acting as a mechanical barrier or tamponade.

Device Description

MONTAGE-QS Settable. Resorbable Hemostatic Bone Putty is a sterile. biocompatible, resorbable material of putty-like consistency for use in the control of bleeding from bone surfaces. The single use MONTAGE-QS device contains two separate components of putty-like consistency comprised of granular calcium phosphate, calcium stearate, vitamin E acetate, a triglyceride, polyalcohols and a mixture of a lactide-diester and polyester-based polymers. When mixed together, the components of the MONTAGE-QS device form a resorbable putty-like material that can be applied directly to bleeding bone. The resulting material is primarily comprised of calcium phosphate. MONTAGE-QS must be mixed immediately prior to use.

When applied to surgically cut or traumatically broken bone. MONTAGE-QS Settable, Resorbable Hemostatic Bone Putty achieves local control of bleeding by acting as a mechanical barrier (tamponade).

AI/ML Overview

This document (K191140) is a 510(k) Premarket Notification for a medical device called "MONTAGE-QS Settable, Resorbable Hemostatic Bone Putty." It focuses on demonstrating substantial equivalence to a previously cleared predicate device, rather than proving that the device meets specific performance acceptance criteria via a clinical study with a detailed test set, ground truth, and expert evaluation as would be the case for a diagnostic AI/ML device.

Therefore, the information requested for acceptance criteria and a study proving the device meets those criteria (especially points 1 through 9 pertaining to diagnostic performance, expert review, MRMC studies, etc.) cannot be fully extracted or accurately described from the provided text.

Here's a breakdown of what can be extracted and why other points cannot be addressed:

1. A table of acceptance criteria and the reported device performance:

This document does not present acceptance criteria for diagnostic performance outcomes (e.g., sensitivity, specificity, accuracy) or statistical thresholds for improved human reader performance with AI. Instead, it focuses on performance testing to demonstrate substantial equivalence to a predicate device in terms of physical properties, biocompatibility, and functionality.

Criteria Type (as inferred)Reported Device Performance (as inferred)
Bench TestingHandling properties, performance over temperature range, dissolution properties evaluated (relative stiffness, spreadability, stickiness, temperature sensitivity, electrocautery compatibility, dissolution, swelling).
Biocompatibility Testing (ISO 10993)Evaluated in accordance with GLP requirements for: cytotoxicity, irritation, sensitization, acute systemic toxicity, genotoxicity, implantation, systemic toxicity, hemolysis, endotoxicity, pyrogenicity.
Animal Testing (In Vivo Functionality)Demonstrated intraoperative in vivo hemostasis, resistance to irrigation, local tissue response, and characterized resorption time.
Technological Characteristics"Nearly identical" to predicate: two putty materials mixed at time of use, calcium salt + other resorbable materials, functions by mechanical tamponade, primarily calcium phosphate (~70% by weight), hardens in situ, resorbable, sterile via gamma irradiation.
Safety (Composition)Constituents shown to be "as safe as the predicate device for the intended use."

2. Sample sizes used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

  • Sample Sizes: Not explicitly stated for specific tests. The document refers to "bench studies," "biocompatibility studies," and "animal studies" without specifying the number of samples or animals used in each.
  • Data Provenance: Not specified. These are engineering and preclinical tests, not clinical data from patients.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

This information is not applicable. The "ground truth" for this device's evaluation is based on objective physical/chemical measurements (bench testing), biological reactions (biocompatibility), and observed in-vivo effects in animal models, not on human expert radiologist or clinician interpretations.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

Not applicable, as this testing does not involve human interpretation or adjudication of diagnostic findings.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

Not applicable. This device is a bone hemostatic putty, not an AI/ML diagnostic tool meant to assist human readers.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

Not applicable. This is not an AI/ML algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

The "ground truth" for this device's performance is established through:

  • Standardized physical and chemical property measurements (e.g., stiffness, dissolution rates).
  • Validated biocompatibility assays (ISO 10993 series).
  • Direct observation of hemostatic efficacy and resorption in animal models.

8. The sample size for the training set:

Not applicable. This device does not involve machine learning or a "training set."

9. How the ground truth for the training set was established:

Not applicable. This device does not involve machine learning or a "training set."

Summary of Device and Evidence Presented:

The MONTAGE-QS Settable, Resorbable Hemostatic Bone Putty is a sterile, biocompatible, resorbable material intended to control bleeding from cut or damaged bone by acting as a mechanical barrier. The 510(k) submission successfully demonstrated substantial equivalence to a predicate device (MONTAGE Settable, Resorbable Hemostatic Bone Putty K152005) through:

  • Bench Testing: Verifying handling properties (stiffness, spreadability, stickiness), temperature sensitivity, electrocautery compatibility, dissolution, and swelling.
  • Biocompatibility Testing: According to ISO 10993 recommendations (cytotoxicity, irritation, sensitization, acute systemic toxicity, genotoxicity, implantation, systemic toxicity, hemolysis, endotoxicity, pyrogenicity).
  • Animal Testing: Demonstrating intraoperative in vivo hemostasis, resistance to irrigation, local tissue response, and characterizing resorption time.
  • Technological Characteristics Comparison: Showing the device is "nearly identical" in configuration, formulation (with one exception for faster setting time), materials (primarily calcium phosphate), and principle of action (mechanical tamponade) to the predicate.

The FDA clearance (K191140) implies that based on these non-clinical tests, the device is considered as safe and effective as the legally marketed predicate device for its stated Indications for Use.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG" on top and "ADMINISTRATION" below.

July 18. 2019

Orthocon, Inc. Howard Schrayer Official Correspondent 1 Bridge Street, Suite 121 Irvington, New York 10533

Re: K191140

Trade/Device Name: MONTAGE-QS Settable, Resorbable Hemostatic Bone, HBP5, Montage Rapid Set (RS) and Montage Fast Set (FS) Regulatory Class: Unclassified Product Code: MTJ Dated: April 29, 2019 Received: April 30, 2019

Dear Howard Schrayer:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Nina Mezu-Nwaba, PharmD., MPH. MSC. CAPT USPHS Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K191140

Device Name

MONTAGE-QS Settable, Resorbable Hemostatic Bone Putty

Indications for Use (Describe)

MONTAGE-QS Settable, Resorbable Hemostatic Bone Putty is indicated for the control of bleeding from cut or damaged bone by acting as a mechanical barrier or tamponade.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

Contact:Howard SchrayerOrthocon, Inc.1 Bridge Street, Suite 121Irvington, NY 10533Telephone: 914-357-2600Fax: 914-231-7884hs.ss@verizon.net
Date Prepared:June 25, 2019
Device Trade Name:MONTAGE-QS Settable, Resorbable Hemostatic BonePutty
Manufacturer:Orthocon, Inc.1 Bridge Street, Suite 121Irvington, NY 10533
Common Name:Bone Wax
Classification:Unclassified
Product Code:MTJ
Predicate DevicesPrimary Predicate:Orthocon, Inc. MONTAGE Settable, Resorbable HemostaticBone Putty510(k) K152005
Reference Device:Orthocon, Inc. HBP4 Hardening, Resorbable Hemostatic BonePutty510(k) K141502

{4}------------------------------------------------

Indications for Use:

MONTAGE-QS Settable, Resorbable Hemostatic Bone Putty is indicated for the control of bleeding from cut or damaged bone by acting as a mechanical barrier or tamponade.

Device Description:

MONTAGE-QS Settable. Resorbable Hemostatic Bone Putty is a sterile. biocompatible, resorbable material of putty-like consistency for use in the control of bleeding from bone surfaces. The single use MONTAGE-QS device contains two separate components of putty-like consistency comprised of granular calcium phosphate, calcium stearate, vitamin E acetate, a triglyceride, polyalcohols and a mixture of a lactide-diester and polyester-based polymers. When mixed together, the components of the MONTAGE-QS device form a resorbable putty-like material that can be applied directly to bleeding bone. The resulting material is primarily comprised of calcium phosphate. MONTAGE-QS must be mixed immediately prior to use.

When applied to surgically cut or traumatically broken bone. MONTAGE-QS Settable, Resorbable Hemostatic Bone Putty achieves local control of bleeding by acting as a mechanical barrier (tamponade).

Substantial Equivalence and Predicate Devices:

The device was shown to be substantially equivalent to the previously cleared bone hemostasis predicate device, Montage Settable, Resorbable Hemostatic Bone Putty (K152005).

Performance Testing:

Bench testing, biocompatibility and animal functionality testing performed on MONTAGE-QS Settable. Resorbable Hemostatic Bone Putty demonstrate that the device is substantially equivalent to the predicate device in intended use, technological characteristics, and performance. This testing included the following:

Bench Testing was conducted to verify the device's handling properties, to characterize the device's performance over a range of temperatures and to evaluate the device's dissolution properties. The following bench studies were completed: relative stiffness, spreadability, stickiness, temperature sensitivity, electrocautery compatibility, dissolution and swelling.

{5}------------------------------------------------

Biocompatibility Testing was conducted to evaluate the device's biocompatibility in accordance with the recommendations of ISO 10993. The following biocompatibility studies were conducted on the final, finished, gamma-irradiated sterile device in accordance with the GLP requirements: cvtotoxicity, irritation, sensitization, acute systemic toxicity, genotoxicity, implantation, systemic toxicity, hemolysis, endotoxicity and pyrogenicity.

Animal Testing included animal studies to demonstrate intraoperative in vivo hemostasis, resistance to irrigation, local tissue response and to characterize resorption time.

Technological Characteristics of MONTAGE-QS and the predicate Orthocon, Inc. MONTAGE Settable, Resorbable Hemostatic Bone Putty are nearly identical. MONTAGE-QS Settable, Resorbable Hemostatic Bone Putty and the referenced predicate use similar configurations and formulations (i.e., two putty materials that are mixed together at the time of use, including a calcium salt and other resorbable materials for ease of application). Both function by the same principle (i.e., mechanical tamponade). MONTAGE-QS Settable, Resorbable Hemostatic Bone Putty and the predicate are formulated from the same materials with a single exception (that accelerates setting time). Both are primarily comprised (~70% by weight) of calcium phosphate: and rely on a reaction between two packaged components to produce a putty that hardens in situ. These devices are resorbable devices provided sterile by gamma irradiation.

The constituents of the MONTAGE-QS Settable. Resorbable Hemostatic Bone Putty device and the predicate device formulations have been shown to be as safe as the predicate device for the intended use. The devices are provided in packages designed for individual patient use.

Conclusion

MONTAGE-QS is substantially equivalent to the previously cleared bone wax predicate device [Montage Settable, Resorbable Hemostatic Bone Putty (K152005)] with respect to intended use, general technological characteristics and performance.