K Number
K191131

Validate with FDA (Live)

Date Cleared
2022-08-03

(1192 days)

Product Code
Regulation Number
880.6850
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Solidence Multipockets is a sterilization pouch intended to provide dentists with an effective method to enclose stainless steel dental burs of up to 0.5 grams intended for sterilization in steam autoclaves. The validated gravity steam sterilization cycle parameters are 10 minutes at 275°F (135°C), 30 minutes drying time, with a maximum weight of 18 grams, with no more than one bur in each pocket. The sterilization pockets maintain the sterility of the enclosed devices for one year post sterilization.

Device Description

The Solidence Multipockets is a single use sterilization pouch that features 36 pockets, designed to be used by dentists to enclose small medical devices intended for sterilization in steam autoclaves. The design of the Solidence Multipockets allows the dentists to sterilize 36 dental burs individually, using one sterilization pouch. Each pocket can then be opened independently, without compromising the sterility of the items contained in the other pockets of the sterilization pouch.

AI/ML Overview

The provided text is a 510(k) Summary for a medical device called "Solidence Multipockets," a sterilization pouch. This document details non-clinical testing performed to demonstrate the device's safety and effectiveness compared to a predicate device.

Here's the breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The "SUMMARY OF NON-CLINICAL TESTING" section directly provides this information.

TestPurposeAcceptance CriteriaResult
Sterilization efficacy half cycle sterilizationDemonstrate adequate sterilization performanceNo growthPass
Full cycle sterilizationDemonstrate adequate sterilant penetrationNo growthPass
Drying timeDemonstrate adequate drying performanceThe mass of the device after sterilization shall not be increased more than 3%.Pass
ASTM 10993-5 (Cytotoxicity)Demonstrate pouch contact during sterilization will not produce a cytotoxic effect in sterilized devicesUnder the conditions of the testing, non-cytotoxicPass
ASTM F88 Seal StrengthDemonstrate adequate sealing performanceSeal strength greater than 0.57 lbf/inPass
ASTM F1929 Dye PenetrationDemonstrate adequate resistance to seal leakageNo channels detectedPass

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state the specific sample sizes used for each non-clinical test (e.g., how many pouches were tested for seal strength). It only states that the tests were performed and passed.

Data Provenance:

  • Country of Origin of the Data: Not explicitly stated, though the manufacturer is Solidence Surgical Corporation in San Fernando, California, USA, and the representative/consultant is Acknowledge Regulatory Strategies, LLC. The testing itself could have been conducted in a variety of locations.
  • Retrospective or Prospective: All non-clinical testing described (sterilization efficacy, drying time, seal strength, dye penetration, biocompatibility, accelerated aging) is inherently prospective testing, as it involves newly manufactured devices undergoing specific controlled procedures to assess their performance.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

The information provided describes non-clinical, laboratory-based testing, not studies involving human experts establishing ground truth for medical image interpretation or similar AI applications. Therefore, this question is not applicable to this document. The "ground truth" for these tests comes from the physical and chemical properties of the materials and the validated performance metrics for sterilization and packaging.

4. Adjudication Method for the Test Set

Since this is non-clinical performance testing for a physical device (sterilization pouch), rather than an AI or diagnostic imaging study requiring expert interpretation and potential disagreement, there is no adjudication method described or necessary in the context of this document. The results are based on objective measurements and established standards (e.g., "no growth," "mass not increased more than 3%," "no channels detected").

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, a MRMC comparative effectiveness study was not done. The document explicitly states: "No clinical testing was used in support of this submission." MRMC studies are typically performed for diagnostic devices (e.g., AI for medical imaging) to assess reader performance with and without AI assistance. This device is a sterilization pouch.

6. If a Standalone (i.e., Algorithm Only Without Human-in-the-Loop Performance) Was Done

This question is not applicable as the device is a physical sterilization pouch, not an algorithm or AI system.

7. The Type of Ground Truth Used

For this physical device, the "ground truth" for the non-clinical testing is based on:

  • Objective Laboratory Measurements: e.g.,
    • Absence/presence of microbial growth (for sterilization efficacy)
    • Change in mass (for drying time)
    • Force measurements (for seal strength)
    • Visual inspection for dye penetration (for seal leakage)
    • Cell viability assays (for cytotoxicity/biocompatibility)
  • Established Industry Standards and Protocols: The tests reference ASTM and ISO standards (e.g., ASTM F88, ASTM F1929, ISO 10993-1, ISO 11138-1), indicating that the methodologies and criteria are widely accepted and validated within the medical device industry for sterilization and packaging.

8. The Sample Size for the Training Set

This question is not applicable as the device is a physical medical device, not an AI model that requires a training set.

9. How the Ground Truth for the Training Set Was Established

This question is not applicable as the device is a physical medical device, not an AI model that requires a training set and corresponding ground truth establishment.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

August 3, 2022

Solidence Surgical Corporation Morris Azad President 2251 San Diego Avenue, Suite B-257 San Fernando, California 91340

Re: K191131

Trade/Device Name: Solidence Multipockets Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: Class II Product Code: FRG Dated: July 3, 2022 Received: July 7, 2022

Dear Morris Azad:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Clarence W. Murray, III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K191131

Device Name Solidence Multipockets

Indications for Use (Describe)

The Solidence Multipockets is a sterilization pouch intended to provide dentists with an effective method to enclose stainless steel dental burs of up to 0.5 grams intended for sterilization in steam autoclaves. The validated gravity steam sterilization cycle parameters are 10 minutes at 275°F (135°C), 30 minutes drying time, with a maximum weight of 18 grams, with no more than one bur in each pocket. The sterilization pockets maintain the sterility of the enclosed devices for one year post sterilization.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary K191131

DATE PREPARED Aug 1, 2022

MANUFACTURER AND 510(k) OWNER

Solidence Surgical Corporation 1104 San Fernando Road, San Fernando, CA 91340, USA Telephone: Official Contact: Morris Azad, D.D.S., President

REPRESENTATIVE/CONSULTANT

Allison C. Komiyama, Ph.D., R.A.C.
Lucie Dalet, Ph.D.
Acknowledge Regulatory Strategies, LLC
Telephone:+1 (619) 458-9547
Email:akomiyama@acknowledge-rs.com
ldalet@acknowledge-rs.com
Website:acknowledge-rs.com

DEVICE INFORMATION

Proprietary Name/Trade Name:Solidence Multipockets
Common Name:Wrap, Sterilization
Regulation Number:21 CFR 880.6850
Class:II
Product Code:FRG

PREDICATE DEVICE IDENTIFICATION

Proprietary Name/Trade Name:BH Sterilization Pouch
510(k) Number:172280
Product Code:FRG
Manufactured by:BH Medical Products Co., Ltd.No.90 Shenjiang Villagers' group, Zhangjiacunwei, Xilin StreetZhonglou District, Changzhou City, Jiangsu Province,People's Republic of China

DEVICE DESCRIPTION

The Solidence Multipockets is a single use sterilization pouch that features 36 pockets, designed to be used by dentists to enclose small medical devices intended for sterilization in steam autoclaves. The design of the Solidence Multipockets allows the dentists to sterilize 36 dental burs individually, using one sterilization pouch. Each pocket can then be opened independently, without compromising the sterility of the items contained in the other pockets of the sterilization pouch.

{4}------------------------------------------------

INDICATIONS FOR USE

The Solidence Multipockets is a sterilization pouch intended to provide dentists with an effective method to enclose stainless steel dental burs of up to 0.5 grams intended for sterilization in steam autoclaves. The validated gravity steam sterilization cycle parameters are 10 minutes at 275°F (135°C), 30 minutes drying time, with a maximum weight of 18 grams, with no more than one bur in each pocket. The sterilization pockets maintain the sterility of the enclosed devices for one year post sterilization.

Predicate DeviceSubject DeviceComparison
BH Medical Products Co., Ltd.Solidence Surgical Corp.
BH Sterilization PouchSolidence Multipockets
K172280K191131
Product CodeFRGFRGSame
Indications For UseThe BH sterilization pouch is intended to provide dentists with an effective method to enclose devices intended for sterilization in steam autoclaves. The recommended pre- vacuum steam sterilization cycle parameters are 4 minutes at 132°C (270°F). The sterilization pouch maintains the enclosed devices up until 180 days post sterilization. Lastly, the pouch's external chemical ink indicators are designed to indicate to the user that the pouch has undergone a steam sterilization process. Only one instrument can be sterilized in one pouch and only instruments that are be metal, hinged, or knurled can be sterilized in the pouch.The Solidence Multipockets is a sterilization pouch intended to provide dentists with an effective method to enclose stainless steel dental burs of up to 0.5 grams intended for sterilization in steam autoclaves. The validated gravity steam sterilization cycle parameters are 10 minutes at 275°F (135°C), 30 minutes drying time, with a maximum weight of 18 grams, with no more than one bur in each pocket. The sterilization pockets maintain the sterility of the enclosed devices for one year post sterilization.Different
Material CompositionMedical Grade Paper, laminate film, polyurethane adhesive, Steam Process Indicator Print InkPolymeric laminate film (RPA-202 and RPA-204)Different
Sterilization ParametersPre-vacuum steam sterilization: 4 minutes at 132°C (270°F).Gravity steam sterilization: 10 minutes at 275°F (135°C), 30 minutes drying timeDifferent
Process IndicatorYesNoDifferent
DesignThe pouches are made from a medical grade paper and plastic film that is heat sealed on three sides. The fourth side has an adhesive strip that is used to seal the paper to the film prior to sterilization of the enclosed medical device. The medical grade paper conforms to recognized material standards and can be steam sterilized.The Solidence Multipockets is composed of two sheets of laminated film that are heat sealed together to form 36 pockets. The design of the Solidence Multipockets allows the dentist to sterilize 36 items individually, using one sterilization sheet. Each pocket can then be opened independently. withoutDifferent
The process indicator ink printed on the medical paper will exhibit a color change after the pouch is exposed to steam sterilization.compromising the sterility of the items enclosed in the other pockets of the sterilization sheet.
UsageSingle useSingle useSame
Configurations /DimensionsSingle pouch configuration:#990613: 110 mm x 300 mm#990616: 90 mm x 230 mm#990617: 135 mm x 260 mm#990618: 190 mm x 330 mmMultiple pockets on one sheet.Sheet dimensions: 265 mm x 308 mmPockets dimensions: 25 mm x 40 mmDifferent
Shelf Life2 years5 yearsDifferent
Maintenance ofSterility180 days1 yearDifferent
BiocompatibilityUnder the conditions of the testing, noncytotoxicUnder the conditions of the testing, non-cytotoxicSame

COMPARISON OF TECHNOLOGICAL CHARACTERISTICS

{5}------------------------------------------------

SUMMARY OF NON-CLINICAL TESTING

The following tests were performed to demonstrate safety based on current industry standards:

TestPurposeAcceptance CriteriaResult
Sterilization efficacyhalf cycle sterilizationDemonstrateadequate sterilizationperformanceNo growthPass
Full cycle sterilizationDemonstrateadequate sterilantpenetrationNo growthPass
Drying timeDemonstrateadequate dryingperformanceThe mass of the deviceafter sterilization shallnot be increased morethan 3%.Pass
ASTM 10993-5Demonstrate pouchcontact duringsterilization will notproduce a cytotoxiceffect in sterilizeddevicesUnder the conditions ofthe testing,non-cytotoxicPass
ASTM F88 SealStrengthDemonstrateadequate sealingperformanceSeal strength greaterthan 0.57 Ibf/inPass
ASTM F1929 DyePenetrationDemonstrateadequate resistanceto seal leakageNo channels detectedPass

Sterilant penetration per ISO 11138-1 Sterilization of health care products - Biological indicators - Part 1: General requirements

{6}------------------------------------------------

Package integrity and maintenance of package integrity per ASTM F88 Standard Test Method for Seal Strength of Flexible Barrier Materials and ASTM F1929 Standard Test Method for Detecting Seal Leaks in Porous Medical Packaging by Dye Penetration

Accelerated aging per ASTM F1980 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices

Biocompatibility per ISO 10993-1 Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process

SUMMARY OF CLINICAL TESTING

No clinical testing was used in support of this submission.

CONCLUSION

The conclusions drawn from the non-clinical testing demonstrate that the subject device, Solidence MultiPockets (K191131), is as safe, as effective, and performs as well as or better than the legally marketed predicate (K1712280).

§ 880.6850 Sterilization wrap.

(a)
Identification. A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.(b)
Classification. Class II (performance standards).