K Number
K191075

Validate with FDA (Live)

Device Name
Healy
Date Cleared
2019-11-22

(213 days)

Product Code
Regulation Number
882.5890
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The device is designed to be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, arms and legs due to strain from exercise or normal household work activities and for the symptomatic relief and management of chronic, intractable pain and relief of pain associated with arthritis.

Device Description

The Healy device is a single channel wearable electrotherapy device that is designed to alleviate temporary and chronic muscle and joint pain on multiple body locations. It delivers microcurrent therapy through the simple and convenient use of the dedicated iOS or Android App, or on product controls. Reusable, selfadhesive and contouring electrodes allow for discreet and convenient placement on multiple pain locations on the body. The system contains one unit, which is rechargeable and is attached to the electrode via cables. The electrode can then be applied to intact skin at the desired location for therapy and pain relief. Control of the Healy system is completed through the available App or on product controls.

AI/ML Overview

The provided text is a 510(k) summary for the Healy device, which is a Transcutaneous Electrical Nerve Stimulator (TENS) for pain relief. The document primarily focuses on demonstrating substantial equivalence to a predicate device (Omron Avail, Model PM601, K172079) rather than providing detailed acceptance criteria and a study proving the device meets those criteria.

However, based on the information provided, we can infer some details related to acceptance criteria and the underlying studies.

1. A table of acceptance criteria and the reported device performance:

The document doesn't present a specific table of acceptance criteria with corresponding performance metrics like a typical clinical trial report. Instead, it relies on demonstrating equivalence to a predicate device by comparing various technological characteristics and performance parameters. The "acceptance criteria" can be interpreted as demonstrating that the device's characteristics are either "No Difference", "Similar", or if different, that the differences "do not raise new questions of safety or effectiveness" when compared to the predicate device.

Here's an aggregated summary table based on the "Substantial Equivalence Comparison Table" and "Output Comparison" in the provided text. The "Acceptance Criteria" column reflects the implicit criteria used for substantial equivalence, i.e., that the device characteristics should be comparable to the predicate. The "Reported Device Performance" is the value for the Healy device. The "Outcome/Compliance" indicates whether the Healy device was found to be acceptable in comparison to the predicate.

CharacteristicAcceptance Criteria (based on predicate)Reported Device Performance (Healy)Outcome/Compliance
General Characteristics
Regulation21 CFR§882.589021 CFR§882.5890No Difference
UsageOver the CounterOver the CounterNo Difference
Intended UseRelief of pain associated with sore/aching muscles (back, arms, legs, shoulders, feet) from strain; symptomatic relief/management of chronic, intractable pain; relief of pain associated with arthritis.Same as predicateNo Difference
Environment of UseClinics, hospital and home environments.Clinics, hospital and home environments.No Difference
Patient PopulationAll adults 18 years and older.All adults 18 years and older.No Difference
ContraindicationsCardiac pacemaker, implanted defibrillator, other implanted metallic/electronic device.Same as predicateNo Difference
SterilityExternal contacting device, nonsterileExternal contacting device, nonsterileNo Difference
Power SourceRechargeable Lithium Ion BatteryRechargeable Lithium Ion BatteryNo Difference
Method of Line Current IsolationN/A (internal power source)N/A (internal power source)No Difference
Automatic No Load contact TripYesYesNo Difference
Automatic Shut OffYesYesNo Difference
Compliance With Voluntary StandardsIEC 60601-1, IEC 60601-1-2, IEC 60601-2-10, IEC 60601-1-11IEC 60601-1, IEC 60601-1-2, IEC 60601-2-10, IEC 60601-1-11No Difference
Performance Characteristics (Microcurrent mode)
WaveformBiphasicBiphasicNo Difference
ShapeRectangularRectangularNo Difference
Max Output Voltage (V) @ 500 Ohms0.025 (predicate)0.030Similar (slightly higher, but deemed acceptable for equivalence)
Max Output Voltage (V) @ 2k Ohms0.1000.100No Difference
Max Output Voltage (V) @ 10k Ohms0.5000.500No Difference
Maximum Output Current (mA) @ 500 Ohm0.0500.050No Difference
Maximum Output Current (mA) @ 2kOhm0.0500.050No Difference
Maximum Output Current (mA) @ 10kOhm0.0500.050No Difference
Pulse Duration, Sec2.502.50No Difference
Frequency, Hz0.200.20No Difference
Maximum Phase Charge (uC) @ 500 Ohms125.0125.0No Difference
Maximum Average Current, mA0.0250.025No Difference
Maximum Current Density, mA/cm² @ 500 Ohms0.0008 (predicate)0.0010Similar (slightly higher, within IEC limit of 2mA/cm², deemed acceptable for equivalence)
Maximum Average Power [mW/cm²] @ 500Ohm0.00001398 (predicate)0.00002500Similar (slightly higher, deemed acceptable for equivalence)
Regulated Current or Voltage?CurrentCurrent (and Voltage)Similar (Healy provides both current and voltage, enhancing safety, deemed acceptable)
Patient Leakage Current -normal Condition, uA<100Similar (Healy somewhat lower, deemed acceptable for equivalence)
Patient Leakage Current - Fault Condition, uA<500Similar (Healy somewhat lower, deemed acceptable for equivalence)
Average DC current through electrodes when device is on but no pulses are being applied (uA)0.0<1No Difference
Number of Output Modes1 Microcurrent mode (for comparison)1 Microcurrent modeNo Difference (Healy only compares to the predicate's microcurrent mode)
Automatic Overload TripNo (predicate)YesSimilar (Healy adding an additional safety feature, deemed acceptable)

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document does not describe any clinical study with a "test set" in the context of typical AI/software performance validation involving human data. The submission is a 510(k) for a TENS device, and the "performance data" section states, "All necessary non-clinical testing was conducted on the Healy device to confirm that the device performs as intended." This primarily refers to bench testing, electrical safety, EMC, and software validation.

Therefore, there is:

  • No reported sample size for a test set.
  • No clinical data provenance (country of origin, retrospective/prospective).

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable. As no clinical "test set" based on human data or expert review was used for performance evaluation, no experts were needed to establish ground truth. The evaluation focused on engineering and software validation.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. No expert-adjudicated test set was used.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is not an AI/imaging device. The Healy device is a transcutaneous electrical nerve stimulator, and the submission does not mention any AI component or human reader assistance.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This question is typically relevant for AI/software devices. The "standalone" performance here relates to the device's physical and electrical functioning. The document states that "non-clinical, bench testing included performance verification to confirm acceptable performance of device features and functions" and "It was functionally tested and found to be in compliance with the specification." This implies standalone testing of the device's electrical and software functions.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

For the non-clinical testing, the "ground truth" was established by engineering specifications, international standards (IEC, EN), and regulatory guidance. For example:

  • Electrical safety tests against IEC 60601 series standards.
  • EMC tests against IEC 60601-1-2 and EN 301 489-1 & 17.
  • Software validation against EN 62304, EN ISO 13485, EN ISO 14971, EN 62366 and FDA's Guidance for the Content of Premarket Submissions for Software contained in Medical Devices.
  • Performance parameters (voltage, current, frequency, etc.) were compared against the predicate device's specifications and relevant safety limits (e.g., current density limit of 2mA/cm2 from IEC60601-2-10).

8. The sample size for the training set

Not applicable. The Healy device is a hardware medical device (TENS) with associated software for control. The provided documentation does not describe an AI/ML component that uses a "training set" in the typical sense for learning or pattern recognition.

9. How the ground truth for the training set was established

Not applicable, as no training set for an AI/ML component is mentioned or relevant to this submission.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 22, 2019

TimeWaver Production GmbH % Douglas Herrington Principal Consultant Herrington Consulting LLC 2885 Sanford Ave, SW #43083 Grandville, Michigan 49418

Re: K191075

Trade/Device Name: Healy Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous electrical nerve stimulator for pain relief Regulatory Class: Class II Product Code: NUH, NYN Dated: October 30, 2019 Received: November 12, 2019

Dear Douglas Herrington:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Pamela Scott Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K191075

Device Name Healy

Indications for Use (Describe)

The device is designed to be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, arms and legs due to strain from exercise or normal household work activities and for the symptomatic relief and management of chronic, intractable pain and relief of pain associated with arthritis.

Type of Use (Select one or both, as applicable)| | Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary: K191075

General Information

Date: May 26, 2019

Applicant & Manufacturing Address:

TimeWaver Production GmbH Schloss Kränzlin, Darritzer Straße 6 Kränzlin, Germany 16818 Phone: +49-3391-40022-11 FAX: +49-3391-40022-99 Registration Number: 3009411292

Contact Name

Douglas Herrington Principal Consultant Herrington Consulting LLC 2885 Sanford Ave SW#43083 Grandville, MI 49418-1342 Telephone: 1.248.369.5564 Fax: 1.877.881.4412 E-mail: dgls herrington@yahoo.com

Regulatory Information

Device Name: Healy 510(K) Number: K191075 Classification: Class II Panel: Neurological Therapeutic Devices Product Code: NUH, NYN Regulation Number: 21 CFR 882.5890, Transcutaneous electrical nerve stimulator for pain relief

Indication for Use Statement

The device is designed to be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, arms and legs due to strain from exercise or normal household work activities. And for the symptomatic relief and management of chronic, intractable pain and relief of pain associated with arthritis.

Environments of Use: Clinics, hospital and home environments.

Patient Population: All adults 18 years and older.

{4}------------------------------------------------

510(k) Summary: K191075

Substantial Equivalence Summary

Table 1: Substantial
Equivalence Comparison TableMicrocurrentMicrocurrentComparison
510(k) NumberK191075K172079
Regulation21 CFR§882.589021 CFR§882.5890No Difference
Product CodeNUH, NYNNUHSimilar, Healy adds product codeNYN, however Omron's IFU statesarthritis pain treatment but the 510kapproval is for only NUH.
Device NameHealyAvail, Model PM601
ManufacturerTimeWaver Production GmbHOmron Healthcare Inc
UsageOver the CounterOver the CounterNo Difference
Intended UseThe Healy is intended for the reliefof pain associated with sore oraching muscles of the lower back,arms, legs, shoulders or feet due tostrain from exercise or normalhousehold work activities.For the symptomatic relief andmanagement of chronic, intractablepain and relief of pain associatedwith arthritis.The Avail is intended for the reliefof pain associated with sore oraching muscles of the lower back,arms, legs, shoulders or feet due tostrain from exercise or normalhousehold work activities.When used for the symptomaticrelief and management of chronic,intractable pain and relief of painassociated with arthritis, use theTap, Shoulder, Arm or Leg modeof stimulation.No Difference
Environment of UseEnvironments of Use: Clinics,hospital and home environments.Environments of Use: Clinics,hospital and home environments.No Difference
Patient PopulationPatient Population: All adults 18years and older.Patient Population: All adults 18years and older.No Difference
Contraindications/Warnings/PrecautionsContraindications: Do not use thisdevice if you have a cardiacpacemaker, implanted defibrillator,or other implanted metallic orContraindications: Do not use thisdevice if you have a cardiacpacemaker, implanted defibrillator,or other implanted metallic orNo Difference

{5}------------------------------------------------

510(k) Summary: TimeWaver Production GmbH

electronic device. Such use could cause electric shock, burns, electrical interference, or death. Patient-contacting Pads are for single patient useelectronic device. Such use could cause electric shock, burns, electrical interference, or death. Patient-contacting Pads are for single patient useNo Difference
Single Use
SterilityExternal contacting device, nonsterileExternal contacting device, nonsterileNo Difference
Over-the-Counter: (OTC)YesYesNo Difference
Power Source: VoltageRechargeable Lithium Ion BatteryRechargeable Lithium Ion BatteryNo Difference
Method of Line Current IsolationN/A (internal power source)N/A (internal power source)No Difference
Patient Leakage Current
-normal Condition, uA0<10Similar with Healy having somewhat lower leakage current
- Fault Condition, uA0<50Similar with Healy having somewhat lower leakage current
Average DC Current Through electrodes when device is on but no pulses are being applied (uA)<10.0No Difference
Number of Output Modes1 Microcurrent mode1 Microcurrent mode9 TENS modesNo Difference as Healy only compares to the Microcurrent mode.
Number of Output Channels11No Difference
Synchronous or AlternatingN/AN/ANo Difference
Method of Channel IsolationnonenoneNo Difference
Regulated Current or Regulated Voltageboth (Software control)currentSimilar with Healy providing both current and voltage regulation.
Software/Firmware/Microprocessor Control?MicroprocessorMicroprocessorNo Difference
Automatic Overload TripYesnoSimilar with Healy adding an additional safety feature.
Automatic No Load contact TripYesyesNo Difference
Automatic Shut OffYesyesNo Difference
User Override Control?Yes, power On/Off button on the device and in the App softwareYes, power On/Off button on the device and in the App softwareNo Difference
Indicator Display

{6}------------------------------------------------

510(k) Summary: TimeWaver Production GmbH

On/Off StatusYes, on the App and LED indicatoron Main unitYes, on the App and LEDindicator on Main unitNo Difference
Low Battery?Yes on app and LED indicator ondeviceYES, ON APPNo Difference
Voltage/CurrentYES, ON APPYES, ON APPNo Difference
Timer Range (minutes)1min - 180min5-60 minutes and30-180 minutes for MicrocurrentSimilar with Heal having additionaltiming control of 1- 30 minutes
Compliance With Voluntarystandards?IEC 60601-1IEC 60601-1-2IEC 60601-2-10IEC 60601-1-11IEC 60601-1IEC 60601-1-2IEC 60601-2-10IEC 60601-1-11No Difference
Compliance with 21 CFR 898YesN/A no patient cableDifferent with Healy complying with21 CFR 898.
Weightdevice 34gelectrode 2gDevice: Approx. 42 gPad-L: Approx. 21 gPad-M: Approx. 17.5gCharger: Approx. 100gHealy is lighter than the SE as thepads are smaller and no charger isprovided.
Dimensions (W x H x D). mmdevice : 55 x 55 x 7pads: 32mm round x 5mm heightDevice: Approx. 60 × 72 × 15.5Charger: Approx. 158 x 90 x 20.5Pad-L: Approx. 219 × 83.5 × 9.3Pad-M: Approx. 180 × 79.5 × 9.3Healy pads are smaller than thepredicate because they do not deliverTENS voltages/currents, and nocharger is provided. See the technicaldiscussion.
Operating Conditions5°C - 40°C15-93%RH700-1060hpa10 to 40 °C30 to 80 %RH700 to 1060 hPa (noncondensing)Similar with Healy having slighthigher relative humidity
Charging Conditions5-40°C15-93% RH.5 to 35 °C30 to 80 %RHSimilar with Healy having slighthigher relative humidity
Storage Conditions5-40°C15-93% RH.0 to 40 °C30 to 80 % RHSimilar with Healy having slighthigher relative humidity and slightlylower temperature
Transporting Conditions-20 to 60 °C15-93% RH-20 to 60 °C10 to 90 % RHSimilar with Healy having slighthigher relative humidity and slightlylower temperature
Electrode StyleReusableReusableNo Difference

{7}------------------------------------------------

510(k) Summary: TimeWaver Production GmbH
-------------------------------------------------
Table 2: Output Comparison
Microcurrent mode
510(k) NumberK191075K172079Comparison
Product CodeNUH, NYNNUH
Device NameHealyAvail, Model PM601
ManufacturerTimeWaver Production GmbHOmron Healthcare Inc
WaveformBiphasicBiphasicNo Difference
ShapeRectangularRectangularNo Difference
Max Output Voltage (V)
500 Ohms0.0300.025Similar with Healy havingslightly higher output current.
2k Ohms0.1000.100No Difference
10k Ohms0.5000.500No Difference
Maximum Output Current (mA)
500Ohm0.0500.050No Difference
2kOhm0.0500.050No Difference
10kOhm0.0500.050No Difference
Pulse Duration, Sec2.502.50No Difference
Frequency, Hz0.200.20No Difference
Net Charge (uC per charge) @ 500ohms00No Difference
Maximum Phase Charge (uC), 500Ohms125.0125.0No Difference
Maximum Average Current, mA0.0250.025No Difference
Maximum Current Density,mA/cm²
500 Ohms0.00100.0008Similar with Healy havingslightly higher current density.
2k Ohms0.001---Unknown as the predicate doesnot give a value
10k Ohms0.001---Unknown as the predicate doesnot give a value

{8}------------------------------------------------

510(k) Summary: TimeWaver Production GmbH

Maximum Average Power[mW/cm²]
500Ohm0.000025000.00001398Similar with Healy havingslightly higher maximumaverage power.
2kOhm0.0000100----The predicate does not mentionthe value
10kOhm0.0000500----The predicate does not mentionthe value
Regulated Current or Voltage?CurrentCurrentNo Difference

{9}------------------------------------------------

510(k) Summary: K191075

Discussion of Healy and Avail similarities and differences

PREDICATE DEVICE(S) [807.92(a)(3)]

The Healy device is substantially equivalent to the primary predicate device, the Omron Avail Microcurrent mode (K172079) with regard to product labeling, intended use, anatomical sites, patient population, performance testing, technological and safety characteristics.

DEVICE DESCRIPTION [807.92(a)(4)]

The Healy device is a single channel wearable electrotherapy device that is designed to alleviate temporary and chronic muscle and joint pain on multiple body locations. It delivers microcurrent therapy through the simple and convenient use of the dedicated iOS or Android App, or on product controls. Reusable, selfadhesive and contouring electrodes allow for discreet and convenient placement on multiple pain locations on the body.

The system contains one unit, which is rechargeable and is attached to the electrode via cables. The electrode can then be applied to intact skin at the desired location for therapy and pain relief. Control of the Healy system is completed through the available App or on product controls. The Healy will be packaged with an Instruction Manual and quick start guide that provides details on setting up the device for use, installing of the App, setting and controlling therapy modes, and troubleshooting. The system includes the unit, adhesive gel electrodes, USB charging cable, electrode connection cable, and storage case.

COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICES, [807.92(a)(6)]

In regard to technological characteristics, the Healy device is similar to the predicate with only minor technological differences. Like the predicate Avail device is paired and controlled by a dedicated smartphone App and on product controls with treatment duration and intensity being controlled by the App or product controls. However, both proposed and predicate devices can be turned off by pressing the on/off button on the main unit or the App. The Healy device and the Avail both offer one microcurrent mode. The Healy and Avail both use biphasic rectangular waveform. Both products use microcurrent lowlevel electrical stimulation and share the same regulation number (21 CFR 882.5890) the same product codes of NUH and NYN. The Avail instruction booklet (pg 26 of the attached Avail IB) states. "Any of the modes can be used on body parts or pains described in this instruction manual of the device", which includes the microcurrent mode.

The microcurrent mode is considered a very low current TENS (rectangular biphasic waveform) delivered at a low frequency. A comparison of the Healy operating characteristics to the primary predicate is included in the 510k summary.

In general, the Healy output parameters fall within the range of output parameters for Avail. For example, the maximum current density (mA/cm2) range for Healy @ 500 ohms is 0.0010, while the Avail is 0.0008. This range is also well below the IEC60601-2-10:2012 (Clause 201.4.2) limit of less than 2mA/cm2. The maximum average power density (W/cm2) @500 ohms is 0.00002500 for Healy, whereas the Avail maximum average power density @500 ohms is 0.00001398.

The Healy and Avail differ in how the electrode pads are configured. The Avail has pads that do not use lead wires, where the Healy does use lead wires. Additionally, the Healy pads are 32mm round where the Avail is 180 x 80mm because Healy only delivers microcurrent stimulation whereas the Avail also delivers TENS voltages and currents. The Healy electrodes and cables comply with 21 CFR 898 so thev do not raise new questions of safety or effectiveness.

{10}------------------------------------------------

The proposed device, predicate device and reference devices are all intended for use as transcutaneous electrical nerve stimulation in adult populations for use in clinic, hospital or home settings. TimeWaver has completed comprehensive design verification testing, electrical safety and electromagnetic compatibility testing, software verification and validation to ensure that the Healy device performs as intended. The Healy also passed all testing requirements for electrical safety and EMC, and the device electrode has been previously cleared by FDA. The minor differences in labeling and technological characteristics between the proposed device and the predicate device have been evaluated and determined to not raise different questions of safety or effectiveness. As such, the proposed Healy is substantially equivalent to the predicate device. A comparison table summarizing the specifications and features of the proposed Healy device and the predicate device is included in the 510k summary.

Performance Data [807.92(b)]

All necessary non-clinical testing was conducted on the Healy device to confirm that the device performs as intended.

Nonclinical Testing Summarv:

The nonclinical, bench testing included performance verification to confirm acceptable performance of device features and functions. Other nonclinical safety testing included:

  • · Electrical safety and electromagnetic compatibility testing
  • · Software verification and validation

The collective results of the nonclinical testing demonstrate that the materials chosen, the manufacturing processes, and design of the Healy meet the established specifications.

Performance Data: Electromagnetic Compatibility and Electrical Safety

The Healy device has been thoroughly tested against applicable EN and IEC standards by a third p arty and found to be in compliance with the applicable sections. It was functionally tested and found to be in compliance with the specification.

The applicable electrical and safety standards met are:

IEC 60601-1:Medical electrical equipment - Part 1: General requirements for basic safety andessential performance
IEC 60601-1-2:Medical electrical equipment - Part 1-2: General requirements for basic safety andessential performance - Collateral Standard: Electromagnetic disturbances -Requirements and tests
IEC 60601-1-11:Medical electrical equipment - Part 1-11: General requirements for basic safety andessential performance - Collateral Standard: Requirements for medical electricalequipment and medical electrical systems used in the home healthcare environment
IEC 60601-2-10:Medical electrical equipment - Part 2-10: Particular requirements for the basic safetyand essential performance of nerve and muscle stimulators

EN 301 489-1 &17: Electromagnetic compatibility and Radio spectrum Matters (ERM); Electromagnetic Compatibility (EMC) standard for radio equipment

TCRL 2015-2: Bluetooth RF Compliance

{11}------------------------------------------------

510(k) Summary: TimeWaver Production GmbH

Software Validation

The Healy device software was validated following FDA's Guidance for the Content of Premarket Submissions for Software contained in Medical Devices.

The applicable software validation standards met are: EN 62304 Medical device software - Software life cycle processes EN ISO 13485: 2012 Medical devices-Quality management systems-Requirements for regulatory purposes EN ISO 14971: 2012 Medical devices - Application of risk management to medical devices EN 62304: 2006 Medical device software - Software life cycle processes EN 62366: 2008 Medical devices - Application of usability engineering to medical devices

Biocompatibility Data

Biocompatibility testing is not required because the only patient contact surface is through the electrode which was previously cleared by FDA.

Non-Clinical Testing - Other

In addition to electrical safety testing and software validation, human factors analysis and testing was also conducted and no user issues were identified.

Conclusion

Based upon the intended use and technical information provided in this pre-market notification, the Healy device has been shown to be safe and effective and substantially equivalent to the currently marketed predicate device.

§ 882.5890 Transcutaneous electrical nerve stimulator for pain relief.

(a)
Identification. A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.(b)
Classification. Class II (performance standards).