K Number
K191064

Validate with FDA (Live)

Device Name
Argon Handset
Date Cleared
2019-06-12

(51 days)

Product Code
Regulation Number
878.4400
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Argon Handset is an electrosurgical accessory intended to be used in open or laparoscopic and thoracoscopic surgical procedures (general, neurosurgical, gynecologic) where monopolar standard or argon-enhanced electrosurgery is desired. The electrosurgical energy is controlled by the electrosurgical generator during cutting and coagulation and the argon gas flow is controlled by the argon gas delivery system, via the Argon Handset, during argon-shrouded cutting and argon-enhanced coagulation. When the Argon Handset is activated in the argon-enhanced mode, argon gas plasma is created between the electrode and target tissue.

Device Description

The Argon Handset is intended to be used in electrosurgical procedures where monopolar standard or argon enhanced electrosurgery is desired. The pre-installed blade electrode is used in open surgery, while the longer argon electrodes can replace the blade electrode to be used in minimally invasive procedure. The Argon Handset has to be in conjunction with an electrosurgical generator, argon gas delivery system and patient return electrode. During the operation, the argon handset performs like a standard electrosurgical pencil, while it receives both the argon gas and high frequency current, and delivers the electrosurgical energy onto a target tissue for either a standard or argon-enhanced procedure. The Argon Handset is sold in sterile packaging and is a single use device. These devices can be used in hospitals and are used by trained professionals only. The Argon Handset is designed for fitting into the Covidien Argon Gas Delivery Unit II or Force™ GSU Argon Gas Delivery Systems, and Covidien Force EZ™ or Force FX™ series electrosurgical generators. The blade electrode of the Argon Handset is replaceable by the New Deantronics Argon Electrode Series or Covidien E35XX series Argon Electrodes depending on the surgical procedures required.

AI/ML Overview

This document is a 510(k) Premarket Notification for the Argon Handset device, a type of electrosurgical accessory. It seeks to demonstrate substantial equivalence to a predicate device, the Valleylab Force Argon II Argon Enhanced Electrosurgical System (K964636).

The document does not describe a study that proves the device meets acceptance criteria related to AI/ML algorithm performance, nor does it pertain to a diagnostic or assistive AI medical device that would involve human readers, ground truth establishment by experts, or MRMC studies.

Instead, this 510(k) submission focuses on the safety and performance of a physical medical device (an electrosurgical accessory) through bench testing, electrical safety, EMC testing, shelf-life testing, biocompatibility testing, and package testing.

Therefore, most of the requested information (sample size for test/training sets, data provenance, number/qualification of experts, adjudication methods, MRMC studies, standalone algorithm performance, ground truth types for AI) is not applicable to this type of device and submission.

However, I can extract the acceptance criteria and the methods used to prove the device meets these criteria for this physical medical device.


Acceptance Criteria and Study for the Argon Handset (Electrosurgical Accessory)

The document primarily demonstrates substantial equivalence through technical and performance characteristics compared to a predicate device, supported by extensive non-clinical testing rather than clinical or AI performance studies.

1. Table of Acceptance Criteria and Reported Device Performance

The submission states that "All the test results demonstrate the performance of Argon Handset meets the requirements of its pre-defined acceptance criteria and intended uses." Specific quantitative acceptance criteria are not explicitly tabulated in this document; rather, it details the types of tests conducted and concludes that the device passed them, implying the predefined criteria were met. The performance is reported as meeting "requirements" or "passing" standards.

Acceptance Criteria CategoryReported Device Performance (Summary)
Bench TestingAll test results met the requirements of pre-defined acceptance criteria and intended uses. Device performs as safely and effectively as the predicate. Specific tests included: - Electrode slide - Activation force - Activation overtime - Continuity - Gas filter test - Impact and back pressures - HF breakdown with argon gas - Thermal Effects on Tissue (within a system)
Electrical SafetyThe Argon Handset passed all electrical safety tests. Tested in accordance with: - IEC 60601-1:2005+AM1:2012 (General requirements for basic safety and essential performance) - IEC 60601-2-2:2017 (Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories).
Electromagnetic Compatibility (EMC)The Argon Handset passed all EMC tests. Tested in accordance with: - IEC 60601-1-2:2014 (Electromagnetic disturbances - Requirements and tests).
Shelf-LifeEstablished that the device and packaging remain functional and maintain sterility for 2 years through accelerated aging studies. Tested in accordance with FDA guidance and internal requirements.
BiocompatibilityDemonstrates that the materials in the device will not cause a biocompatibility reaction when used as intended. Tested in accordance with: - ISO 10993-1:2009/AC:2010 (Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process) - FDA guidance documents.
Package IntegrityThe sterile package integrity remains uncompromised and the sterile barrier is not adversely affected after testing. Verified and validated in accordance with associated ASTM package standards.

2. Sample size used for the test set and the data provenance:

  • Test Set Sample Size: Not explicitly stated as a number of devices. Testing typically involves a representative sample of devices, but the exact quantity is not detailed in this summary.
  • Data Provenance: The testing was conducted by New Deantronics Taiwan Ltd. ("New Deantronics") and its "extended team," implying internal testing conducted by the manufacturer. The document doesn't specify data origin by country other than the manufacturer's location (Taiwan). It's not relevant whether it's retrospective or prospective, as this applies to clinical data, not bench/lab testing.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • N/A. This device is a physical electrosurgical accessory. The "ground truth" (i.e., whether the device performs as intended and is safe) is established through engineering and laboratory testing against defined standards and specifications, not through expert human interpretation of data like in an AI medical device.

4. Adjudication method for the test set:

  • N/A. Adjudication methods like 2+1 or 3+1 are used for human review of medical images or data in studies involving expert consensus on ground truth. This is not applicable to an electrosurgical accessory undergoing physical and electrical bench testing.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • No. An MRMC study is not relevant to this device. This is a physical electrosurgical tool, not an AI or imaging-based diagnostic/assistive device that would involve human readers.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • No. This is not an algorithm. Therefore, "standalone performance" in the context of an algorithm is not applicable.

7. The type of ground truth used:

  • For this device, the "ground truth" is defined by international consensus standards (e.g., IEC, ISO, ASTM) and the predetermined design specifications and acceptance criteria of the manufacturer, all validated through bench, electrical, and materials testing. It is not expert consensus, pathology, or outcomes data in the sense used for diagnostic/AI devices.

8. The sample size for the training set:

  • N/A. There is no "training set" for this type of device, as it does not involve machine learning or AI.

9. How the ground truth for the training set was established:

  • N/A. As there is no training set, this question is not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

June 12, 2019

New Deantronics Taiwan Ltd % Craig Coombs President Coombs Medical Device Consulting, Inc 1193 Sherman St Alameda, California 94501

Re: K191064

Trade/Device Name: Argon Handset Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: April 19, 2019 Received: April 22, 2019

Dear Craig Coombs:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K191064

Device Name Argon Handset

Indications for Use (Describe)

The Argon Handset is an electrosurgical accessory intended to be used in open or laparoscopic and thoracoscopic surgical procedures (general, neurosurgical, gynecologic) where monopolar standard or argon-enhanced electrosurgery is desired. The electrosurgical energy is controlled by the electrosurgical generator during and coagulation and the argon gas flow is controlled by the argon gas delivery system, via the Argon Handset, during argon-shrouded cutting and argonenhanced coagulation. When the Argon Handset is activated in the argon-enhanced mode, argon gas plasma is created between the electrode and target tissue.

Type of Use (Select one or both, as applicable)
-------------------------------------------------
☒ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)
-----------------------------------------------------------------------------------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for "DEANTRONICS". The logo consists of the word "NEW" in blue, followed by a blue circle with a red dot in the center. To the right of the circle is the word "DEANTRONICS" in blue. The logo is simple and professional, and it is likely used to represent a company or organization.

510(k) Summary

A. Device Information:

CategoryComments
Sponsor:New Deantronics Taiwan Ltd.12F., No.51, Sec. 4, Chong Yang Rd.,Tu Cheng District,New Taipei City 23675, Taiwan R.O.C.Tel: (886) 2-2268-1726Fax: (886) 2-2268-3800
Correspondent ContactInformation:Craig CoombsCoombs Medical Device Consulting1193 Sherman StreetAlameda, CA 94501Tel: 510-337-0140
Device Common Name:Electrosurgical accessory
Device Classification Number:21 CFR 878.4400
Device Classification &Product Code:Class 2,GEI
Device Proprietary Name:Argon Handset

Predicate Device Information:

Predicate Device:Valleylab Force Argon II Argon EnhancedElectrosurgical System
Predicate Device Manufacturer:Valleylab, Inc.
Predicate Device Common Name:Electrosurgical accessory
Predicate Device Premarket Notification #K964636
Predicate Device Classification:21 CFR 878.4400Electrosurgical, Cutting & CoagulationDevice and Accessories
Predicate Device Class & Product Code:Class 2, GEI

B. Date Summary Prepared

11 June 2019

C. Description of Device

The Argon Handset is intended to be used in electrosurgical procedures where monopolar standard or argon enhanced electrosurgery is desired. The pre-installed blade electrode is used in open surgery, while the longer argon electrodes can replace the blade electrode to be used in minimally invasive procedure. The Argon Handset has to be in conjunction with an electrosurgical generator, argon gas delivery system and patient return electrode. During the operation, the argon handset performs like a standard electrosurgical pencil, while it receives

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for New Deantronics. The word "NEW" is in blue, and next to it is a blue circle with a red dot in the center. The word "DEANTRONICS" is in blue and is to the right of the circle.

both the argon gas and high frequency current, and delivers the electrosurgical energy onto a target tissue for either a standard or argon-enhanced procedure.

The Argon Handset is sold in sterile packaging and is a single use device. These devices can be used in hospitals and are used by trained professionals only.

The Argon Handset is designed for fitting into the Covidien Argon Gas Delivery Unit II or Force™ GSU Argon Gas Delivery Systems, and Covidien Force EZ™ or Force FX™ series electrosurgical generators.

The blade electrode of the Argon Handset is replaceable by the New Deantronics Argon Electrode Series or Covidien E35XX series Argon Electrodes depending on the surgical procedures required.

D. Indications for Use

The Argon Handset is an electrosurgical accessory intended to be used in open or laparoscopic and thoracoscopic surgical procedures (general, neurosurgical, gynecologic) where monopolar standard or argon-enhanced electrosurgery is desired. The electrosurgical energy is controlled by the electrosurgical generator during cutting and coagulation and the argon gas flow is controlled by the argon gas delivery system, via the Argon Handset, during argon-shrouded cutting and argon-enhanced coagulation. When the Argon Handset is activated in the argonenhanced mode, argon gas plasma is created between the electrode and target tissue.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for New Deantronics. The word "NEW" is in blue, and next to it is a blue circle with a red dot in the center. The word "DEANTRONICS" is in blue and is to the right of the circle.

E. Comparison to Predicate Device

As described below, the application Argon Electrodes is substantially equivalent in intended use, technology, design and physician use to the predicate Force Argon II Argon Enhanced Electrosurgical System (K964636).

FeatureApplication Device:Argon HandsetPredicate Device:Force Argon II ArgonEnhanced ElectrosurgicalSystem (K964636)Pertinence ofFeature toConsideration ofSubstantialEquivalence.
Indications for UseThe Argon Handset is anelectrosurgical accessoryintended to be used in open orlaparoscopic and thoracoscopicsurgical procedures (general,neurosurgical, gynecologic)where monopolar standard orargon enhanced electrosurgeryis desired. The electrosurgicalenergy is controlled by theelectrosurgical generator duringcutting and coagulation and theargon gas flow is controlled bythe argon gas delivery system,via the Argon Handset, duringargon-shrouded cutting andargon-enhanced coagulation.When the Argon Handset isactivated in the argon-enhancedmode, argon gas plasma iscreated between the electrodeand target tissue.The Force Argon IIEnhanced ElectrosurgicalSystem is intended for usein both open, laparoscopicand thoracoscopic surgicalprocedures (general,neurosurgical, gynecologic)where monopolarelectrosurgery (cutting andcoagulation) is normallyused. The Force Argon IIArgon EnhancedElectrosurgical Systemprovides a controlled flowof argon to electrosurgicalhandset during cutting andcoagulation. When thehandset is activated in thegas enhanced mode, anargon gas plasma is createdbetween the electrode andthe tissue.The Indications forUse of the applicationand predicate devicesare the same, bothdevices are forstandard cutting orcoagulation; argonenhancedelectrosurgery is alsoavailable when thehandset is activatedin the gas enhancedmode.The Argon shroudedmode was included inthe devicedescription of thepredicate. It allowsthe use of argon gasto blow smoke out ofthe surgical fieldduring RF cutting.
Product CodeGEIGEIIdentical
Technology
Mechanism of StandardElectrosurgeryThe Argon Handset is designedto connect with electrosurgicalgenerator and argon gas deliverysystem. When standardelectrosurgery is activated, theelectrosurgical energy will bedelivery from electrosurgicalgenerator to the handsetelectrode, to cut or coagulate insurgical proceduresSame for K964636Identical
FeatureApplication Device:Argon HandsetPredicate Device:Force Argon II ArgonEnhanced ElectrosurgicalSystem (K964636)Pertinence ofFeature toConsideration ofSubstantialEquivalence.
Mechanism of Argon-Enhanced or ShroudedOperationThe Argon Handset is designedto connect with electrosurgicalgenerator and argon gas deliverysystem. The argon gas tubeserves a gas path for argon gasto flow through to the surgicalfield.At the distal end of electrode,the argon gas will turn intoplasma state and then emit tothe target site when the ArgonGas Delivery System is in the"enhanced" mode or the gas willblow out smoke formed duringRF cutting when in "shrouded"mode.Same for K964636Identical
Energy UsedRadiofrequency ElectricalCurrentRadiofrequency ElectricalCurrentIdentical
Operation PrincipleMonopolar ElectrosurgeryMonopolar ElectrosurgeryIdentical
Operation GasArgon GasArgon GasIdentical
Equipment MatedElectrosurgical Generator:Covidien Force EZTM or ForceFXTM seriesArgon Gas Delivery System:Covidien Argon Gas Delivery UnitII or ForceTM GSU Argon GasDelivery SystemSame for K964636Identical
Design - Handset Mechanism
Electrode ConfigurationBladeBladeIdentical
Electrode Length~17.5mm~17.5mmIdentical
ElectrodeExtension/RetractionSlide GuideSlide GuideIdentical
ElectrodeInterchangeableYesYesIdentical
Control TypeHand ControlHand ControlIdentical
Activation MechanismRocker SwitchRocker SwitchIdentical
Gas On/Off ControlSlide SwitchSlide SwitchIdentical
Cable Set Length10 ft10 ftIdentical
FeatureApplication Device:Argon HandsetPredicate Device:Force Argon II ArgonEnhanced ElectrosurgicalSystem (K964636)Pertinence ofFeature toConsideration ofSubstantialEquivalence.
Connection InterfaceESU: banana leadArgon Gas Delivery: gas filterSignal Control: 3-pin connectorSame for K964636Identical
Argon Gas Flow Rateup to 12LPMup to 12LPMIdentical
Handset Insulationelectrosurgical energy line:double layer of polymerhandset body: polymer shellelectrosurgical energy line:double layer of polymerhandset body: polymer shellIdentical
Electrical SafetyWithstand 5.6kVp for HighFrequency breakdownWithstand 5.6kVp AC at MainFrequency (60Hz) breakdown(unknown, but electricalsafety complies withconsensus standard)Identical
Handset ElectrodeStainless Steel SUS 304Stainless Steel SUS 302/SUS304Functionally Identical
Handset HandleNozzle: Aluminum CeramicNozzle: Aluminum CeramicElectrode Extension Guide:HDPEIdentical
Electrode Extension Guide: HDPEActivation Switch: PCGas Switch: PCHandset Body: HIPSActivation Switch: PCGas Switch: PCHandset Body: HIPS
Cable SetOuter Jacket: PVCArgon Gas Tube: PVCElectrosurgical Energy Line:PE+EVA, CopperSignal Control Line: PVC, CopperOuter Jacket: PVCArgon Gas Tube: PVCElectrosurgical Energy Line:PE+EVA, CopperSignal Control Line: PVC,CopperIdentical
ConnectorsESU: CopperArgon Gas Delivery: AcrylicSignal Control: PVC mold,CopperESU: CopperArgon Gas Delivery: AcrylicSignal Control: PVC mold,CopperIdentical
Single Use or ReusableSingle useSingle useIdentical
SterilizationGammaGammaIdentical
Performance/ SafetyTesting in accordancewith:IEC 60601-1IEC 60601-1-2IEC 60601-2-2IEC 60601-1IEC 60601-1-2IEC 60601-2-2Nearly Identical,Application device inconformance withlatest version of thelisted Standards

Tabular Comparison of the Argon Handset to Predicate Device

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for New Deantronics. The word "NEW" is in blue, and to the right of it is a blue circle with a red dot in the center. To the right of the circle is the word "DEANTRONICS" in blue.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for New Deantronics. The word "NEW" is in blue, and there is a blue circle with a red dot in the middle. The word "DEANTRONICS" is in blue and is to the right of the circle.

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for New Deantronics. The logo features the word "NEW" in a smaller font size, followed by a stylized letter "D" with a red dot in the center. The word "DEANTRONICS" is written in a larger, bold font to the right of the "D" symbol. The logo is primarily in shades of blue.

F. Summary of Supporting Data.

New Deantronics has conducted extensive testing to ensure that the subject device met design specifications, functions as intended and conforms to the internationally recognized standards.

Bench Testing

All the test results demonstrate the performance of Argon Handset meets the requirements of its pre-defined acceptance criteria and intended uses. The results of the non-clinical testing demonstrate that the Argon Handset is as safe and effective as the predicate device and reference device.

Performance bench testing was conducted in accordance with FDA's guidance Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery issued August 15, 2016. Besides the tests derived from FDA guidance, some internal requirements were conducted to verify the design specification (e.g. electrode slide, activation force, activation overtime, and continuity); and some testing was performed to confirm the integrity and safety of argon gas delivery as well (e.g. gas filter test, impact and back pressures, and HF breakdown with argon gas). Thermal Effects on Tissue testing were conducted to demonstrate adequate performance within a system.

Electrical Safety and Electromagnetic Compatibility Testing

Electrical safety testing was conducted in accordance with: IEC 60601-1:2005+AM1:2012 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance

IEC 60601-2-2:2017 Medical electrical equipment - Part 2-2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories.

Electromagnetic Compatibility Testing (EMC) was conducted in accordance with IEC 60601-1-2:2014 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests.

The Argon Handset passed all electrical safety and EMC testing.

Shelf-Life Testing

Shelf-life testing was conducted in accordance with FDA's guidance document Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery issued August 15, 2016 and internal requirements. The Argon Handset was subjected to accelerated aging. The aging studies established that the device and packaging remain functional and maintain sterility for 2 years.

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for "NEW DEANTRONICS". The word "NEW" is in blue, and to the right of it is a blue circle with a red dot in the center. The word "DEANTRONICS" is in blue and is to the right of the circle.

Biocompatibility Testing

Biocompatibility testing was conducted in accordance with ISO 10993-1:2009/AC:2010, Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process, and FDA's guidance documents, Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" issued June 16, 2016 and Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery issued August 15, 2016. This testing demonstrates that the materials in the device will not cause a biocompatibility reaction when used as intended.

Package Testing

The argon handset package has been verified and validated in accordance to the associated ASTM package standards. The test result demonstrates the sterile package of argon handset is adequate. After testing, the package integrity remains uncompromised and the sterile barrier is not adversely affected.

Animal Studies

No animal studies were performed as appropriate verification of the device was achieved from the results of the bench performance testing, biocompatibility evaluation, and electrical/safety testing.

Clinical Studies

No clinical studies were performed as appropriate verification of the device was achieved from the results of the bench performance testing, biocompatibility evaluation, and electrical/safety testing.

Conclusion

New Deantronics concludes that the validations demonstrate that the application Argon Handset is in conformance with the latest bench testing and biocompatibility standards and is substantially equivalent to the predicate Valleylab Force Argon II Argon Enhanced Electrosurgical System (K964636).

§ 878.4400 Electrosurgical cutting and coagulation device and accessories.

(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.