K Number
K190982

Validate with FDA (Live)

Date Cleared
2020-05-29

(410 days)

Product Code
Regulation Number
870.1130
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Automatic Digital Blood Pressure Monitor is intended to measure the systolic and diastolic blood pressure as well as the pulse rate of adult person via non-invasive oscillometric technique in which an inflatable cuff is wrapped around the wrist or the upper arm. It can be used at medical facilities or at home. The intended wrist circumference is 12.520 cm and the intended arm circumference has several models: 2333 cm, 2535 cm, 2242 cm, 33~43 cm.

Device Description

Automatic Digital Blood Pressure Monitor is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual via non-invasive Oscillometric technique in which an inflatable cuff is wrapped around the wrist or the upper arm.It can be used at medical facilities or at home. The Automatic Digital Blood Pressure Monitor main units have the operating elements of ON/OFF knob, SET key which can be user-friendly controlled. Arm type Blood Pressure Monitor is equipped with inflatable cuff, while The wrist type Blood Pressure Monitor`s cuff is attached to the device body itself.

AI/ML Overview

This document is a 510(k) summary for an Automatic Digital Blood Pressure Monitor. It describes the device's characteristics, intended use, and comparison to predicate devices to demonstrate substantial equivalence.

Here's a breakdown of the requested information based on the provided document:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria for the blood pressure monitor's accuracy are based on ISO 81060-2:2013 Non-invasive sphygmomanometers - Part 2: Clinical validation of automated measurement type.

Acceptance Criteria (from ISO 81060-2)Reported Device Performance
Accuracy within acceptable scope specified in ISO 81060-2:2013"The results showed the accuracy of the blood pressure monitor is within acceptable scope specified in ISO 81060-2."
Pressure accuracy: ±3 mmHgPressure: ±3 mmHg
Pulse accuracy: ±5%Pulse: ±5%

Note: The document explicitly states the device meets the ISO 81060-2 standard but doesn't provide granular numerical data beyond the +/- tolerance levels. It confirms the device's accuracy specifications align with the standard.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size (Test Set):
    • Arm Sphygmomanometer: 90 patients (divided by cuff size: 23-33cm: 49 males and 41 females; 25-35cm: 38 males and 52 females; 22-42cm: 48 males and 42 females; 33-43cm: 38 males and 52 females).
    • Wrist Sphygmomanometer: 90 patients (43 males and 47 females).
    • Total Patients: 180 (90 arm + 90 wrist).
  • Data Provenance: The document does not explicitly state the country of origin. It indicates the clinical study was conducted, and "All the subjects were volunteer to take part in the clinical study." The study is described as a clinical investigation, implying a prospective study.

3. Number of Experts Used to Establish Ground Truth and Qualifications

The document does not specify the number of experts or their qualifications. However, for blood pressure measurements, the "ground truth" is typically established by trained medical professionals (e.g., physicians, nurses) using a reference device.

4. Adjudication Method for the Test Set

The document states: "The manual Mercury Sphygmomanometer was used as a reference device." This implies a comparison method where the device's readings are compared against the gold standard (manual mercury sphygmomanometer readings). It does not describe an adjudication method with multiple readers deciding on a "ground truth" where there might be inter-reader variability, as would be common in diagnostic imaging studies. The reference device is the ground truth.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

No, an MRMC comparative effectiveness study was not done. This type of study is typically performed for AI-assisted diagnostic devices to assess how AI impacts human reader performance. This device is a direct measurement device (automatic blood pressure monitor), not an AI-enabled diagnostic tool in the sense that an algorithm assists human interpretation of images. The study focused on the device's accuracy compared to a reference standard.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

Yes, in a sense, the primary test conducted was a "standalone" performance test of the device. The clinical study evaluated the direct output of the Automatic Digital Blood Pressure Monitor against a manual mercury sphygmomanometer, without human interpretation of the device's output being a variable. The device itself performs the measurement and provides readings.

7. The Type of Ground Truth Used

The ground truth used was measurements from a manual Mercury Sphygmomanometer. This is considered a highly reliable and commonly accepted gold standard for blood pressure measurement.

8. The Sample Size for the Training Set

The document does not provide information about a separate "training set" or its sample size. This is typical for medical devices that perform a direct physical measurement (like a blood pressure monitor) and are validated against a known standard. The "training" for such a device is typically part of its engineering and algorithm development, not a discrete data set for a machine learning model submitted for regulatory review in the same way as AI software.

9. How the Ground Truth for the Training Set Was Established

As no specific training set is mentioned in the regulatory document (likely because the device's function is a direct measurement and not an AI/ML model requiring a separate training data ground truth for submission), this information is not provided. The device's underlying algorithm is likely developed and refined through engineering and calibration processes based on known physiological principles of oscillometric measurement, rather than learning from a 'labeled training dataset' in the context of AI.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" arranged in a circular pattern. To the right of the seal, there is the FDA logo in blue, with the words "U.S. FOOD & DRUG" on top and "ADMINISTRATION" below. The logo is simple and professional, reflecting the FDA's role in regulating food and drugs.

May 29, 2020

Xiamen Ants Bro Technology Co., Ltd. % Cassie Lee Manager Guangzhou GLOMED Biological Technology Co., Ltd. 2231, Building 1, Rui Feng Center, Kaichuang Road Huangpu District Guangzhou, 510000 Cn

Re: K190982

Trade/Device Name: Automatic Digital Blood Pressure Monitor Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: April 1, 2019 Received: April 15, 2019

Dear Cassie Lee:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

LT Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K190982

Device Name Automatic Digital Blood Pressure Monitor

Indications for Use (Describe)

Automatic Digital Blood Pressure Monitor is intended to measure the systolic and diastolic blood pressure as well as the pulse rate of adult person via non-invasive oscillometric technique in which an inflatable cuff is wrapped around the wrist or the upper arm. It can be used at medical facilities or at home. The intended wrist circumference is 12.520 cm and the intended arm circumference has several models: 2333 cm, 2535 cm, 2242 cm, 33~43 cm .

Type of Use (Select one or both, as applicable)

Residential Use ( Per 31 CFR 801.4; Primary Residence ) One-Time Transaction ( 31 CFR 801.4; Subsequent Sale )
-----------------------------------------------------------------------------------------------------------------------------------------------------------

__ Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

AS-55G, AS-55A, AS-55M, AS-55F

510(k) Summary

This summary of 510(K) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR 807.92.

1. Submitter's Information

510(k) Owner's Name: Xiamen Ants Bro Technology Co., Ltd.

Establishment Registration Number: Applying

Address: 4th and 5th floor, No. 5 Building, Tech.&Innovation center, No. 289 Wengjiao Road, Haicang District, Xiamen China

Tel: +86-13459020349

Fax: +86-0592-6537633

Contact Person: Jane Xu (Manger)

Email: Sales1@asxd.com.cn

Application Correspondent:

Contact Person: Cassie Lee

Guangzhou GLOMED Biological Technology Co., Ltd.

Address: 2231, Building 1, Rui Feng Center, Kaichuang Road, Huangpu District, Guangzhou, Guangdong, China

Tel: +86 20 8266 2446

Email: regulatory@glomed-info.com

{4}------------------------------------------------

Sponsor: Xiamen Ants Bro Technology Co., Ltd. Subject Device: Automatic Digital Blood Pressure Monitor, Models: BA4110/AS-35A, BA6310/AS-35E, BA4900/AS-35W, AS-35K, AS-35I, AS-35H, AS-35X, AS-35J, AS-35L, AS-35F, AS-55G, AS-55A, AS-55M, AS-55F

2. Subject Device Information

Type of 510(k): Traditional

Common Name:Non-invasive blood pressure measurement systems

Classification Name: System, Measurement, Blood-Pressure, Non-Invasive

Trade Name:Automatic Digital Blood Pressure Monitor

Model Name: BA4110/AS-35A, BA6310/AS-35E, BA4900/AS-35W, AS-35K, AS-35I, AS-35H,

AS-35X, AS-35J, AS-35L, AS-35F, AS-55G, AS-55A, AS-55M, AS-55F

Review Panel: Cardiovascular

Product Code: DXN

Regulation Number: 870.1130

Regulatory Class: 2

3. Predicate Device Information

SponsorShenzhen Combei TechnologyCo.,Ltd.Dongguan Ageless Health IndustrialCo., Ltd.
DeviceNameand ModelDigital Blood Pressure Monitor WristStyleModels: BP800W, BP603W, BP880W,BP885W, BPCB0A-2F, BP850W,BP300W, BP810W, BP602W, BP608W,BP606W, BP660W, BP830W, BP866WAGE Automatic Upper Arm BloodPressure MonitorModel: BA-801X, BA-802X, BA-803X,BA-805X, BA-806X, BA-811X,BA-812X, BA-813X, BA-821X,BA-822X, BA-823X, BA-826X (X can beA, B, C, D, E, F)
510(k) NumberK171833K153552
Product CodeDXNDXN
Regulation870.1130870.1130

{5}------------------------------------------------

Sponsor: Xiamen Ants Bro Technology Co., Ltd.

Subject Device:

Automatic Digital Blood Pressure Monitor, Models: BA4110/AS-35A, BA6310/AS-35E, BA4900/AS-35W, AS-35K, AS-351, AS-35H, AS-35X, AS-35J, AS-35L, AS-35F, AS-55G, AS-55A, AS-55M, AS-55F

Number
RegulationClassIIII

4. Device Description

Automatic Digital Blood Pressure Monitor is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual via non-invasive Oscillometric technique in which an inflatable cuff is wrapped around the wrist or the upper arm.It can be used at medical facilities or at home.

The Automatic Digital Blood Pressure Monitor main units have the operating elements of ON/OFF knob, SET key which can be user-friendly controlled.

Arm type Blood Pressure Monitor is equipped with inflatable cuff, while The wrist type Blood Pressure Monitor`s cuff is attached to the device body itself.

5. Intended Use / Indications for Use

Automatic Digital Blood Pressure Monitor is intended to measure the systolic and diastolic blood pressure as well as the pulse rate of adult person via non-invasive oscillometric technique in which an inflatable cuff is wrapped around the wrist or the upper arm.It can be used at medical facilities or at home. The intended wrist circumference is 12.520 cm and the intended arm circumference has several models: 2333 cm, 2535 cm, 2242 cm, 33~43 cm.

6. Test Summary

Automatic Digital Blood Pressure Monitor has been evaluated the safety and performance by lab bench testing as following:

  • � Electrical safety test according to IEC 60601-1 and IEC 80601-2-30 standards
  • � Electromagnetic compatibility test according to IEC 60601-1-2 standard

{6}------------------------------------------------

Sponsor:Xiamen Ants Bro Technology Co., Ltd.
Subject Device:Automatic Digital Blood Pressure Monitor, Models: BA4110/AS-35A, BA6310/AS-35E, BA4900/AS-35W, AS-35K, AS-35I, AS-35H, AS-35X, AS-35J, AS-35L, AS-35F, AS-55G, AS-55A, AS-55M, AS-55F
  • � Biocompatibility test according to ISO 10993-5 and ISO 10993-10 standards
  • � Usability test according to IEC62366 standard
  • � Software verification and validation test according to the requirements of the FDA "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices"

7. Summary of the clinical study

ISO 81060-2:2013 Non-invasive sphygmomanometers - Part 2: Clinical validation of automated measurement type;

In this clinical investigation, Arm sphygmomanometer had ninety patients (23-33cm: 49 males and 41 females, 25-35cm: 38 males and 52 females, 22-42cm: 48 males and 42 females, and 33-43cm: 38 males and 52 females) and Wrist sphygmomanometer also had ninety patients (43 males and 47 females) participated in the clinical study. Same simultaneous method was adopted during the clinical testing. The manual Mercury Sphygmomanometer was used as a reference device. All the subjects were volunteer to take part in the clinical study, all the subjects completed the clinical study without any adverse event or side-effect.

The results showed the accuracy of the blood pressure monitor is within acceptable scope specified in ISO 81060-2.

8. Comparison to predicate device and conclusion

The technological characteristics, features, specifications, materials, mode of operation, and intended use of Automatic Digital Blood Pressure Monitor is substantially equivalent to the predicate devices quoted above.

The differences between the subject device and predicate devices do not raise new issues of safety or effectiveness.

Elements ofComparisonSubject DevicePredicate Device1Predicate Device2Verdict
t
Elements ofComparisonSubject DevicePredicate Device1Predicate Device2Verdict
CompanyXiamen Ants BroTechnology Co., Ltd.Shenzhen CombeiTechnologyCo., Ltd.Dongguan AgelessHealth IndustrialCo., Ltd.--
Trade NameAutomatic Digital BloodPressure MonitorDigital BloodPressure MonitorWrist StyleAGE AutomaticUpper Arm BloodPressure Monitor--
ClassificationNameNon-invasive blood pressuremeasurement systemNon-invasive bloodpressuremeasurementsystemNon-invasive bloodpressuremeasurementsystemSE
510(k)NumberApplyingK171833K153552--
Product CodeDXNDXNDXNSE
Intended Use /Indications forUseAutomatic Digital BloodPressure Monitor is intendedto measure the diastolic,systolic blood pressuresand pulse rate of an adultindividual in hospitals,hospital-type facilities andhome environments by usinga non-invasive oscillometrictechnique in which aninflatable cuff is wrappedaround the wrist or the upperarm. The intended wristcircumference is 12.520 cmand the intended armcircumference with foursizes:2333 cm, 2535 cm, 2242cm, 33~43 cmThe subject deviceintended tomeasure thediastolic, systolicblood pressuresand pulse rate of anadult individual inhospitals,hospital-typefacilities and homeenvironments byusing anon-invasiveoscillometrictechnique in whichaninflatable cuff (size:12.521.5cm(4.98.5in) is wrappedaround the singleAGE AutomaticUpper Arm BloodPressure Monitor isfor use by medicalprofessionals or athome and is anon-invasive bloodpressuremeasurementsystem intended tomeasure thediastolic andsystolic bloodpressures andpulse rate of anadult individual byusing anon-invasivetechnique in whichan inflatable cuff isSENote 1
Elements ofComparisonSubject DevicePredicate Device1Predicate Device2Verdict
wrist.The Subject deviceis not intended tobe diagnosticdevice.wrapped around theupper arm. The cuffcircumference is sixsizes.
DisplayLCD Digital DisplayLCD Digital DisplayLCD Digital DisplaySE
MeasurementSiteWrist Upper ArmWristUpper ArmSE
CuffCircumferenceWrist type:12.520 cmArm type:2333 cm, 2535 cm, 2242 cm, 33~43 cm12.5~21.5cmsize A: 17cm--22cmsize B: 22cm--30cmsize C: 24cm--34cmsize D: 22cm--42cmsize E: 30cm--42cmsize F: 42cm--50cmSENote 1
Inflation andDeflationProgram-controlled andautomaticProgram-controlledand automaticProgram-controlledand automaticSE
MeasuringMethodNon-invasive OscillometricNon-invasiveOscillometricNon-invasiveOscillometricSE
MeasuringscopePressure:30mmHg280 mmHg;Pulse: 40 bpm199 bpmPressure:30mmHg~280mmHgPulse: 40 bpm~200 bpmPressure:0mmHg~280mmHgPulse: 40 bpm ~200bpmSE
MeasuringaccuracyPressure: ±3mmHgPulse:±5%Pressure: ±3mmHgPulse: ±5%Pressure: ± 3mmHgPulse: ±5%SE
PatientPopulationAdultAdultAdultSE
DisplayLCD Digital DisplayLCD Digital DisplayLCD Digital DisplaySE
Power SupplyWrist type:2x 1.5V "AAA"4x 1.5V "AA"SE
Elements ofComparisonSubject DevicePredicate Device1Predicate Device2Verdict
2x 1.5V "AAA" batteriesArm type:4x 1.5V "AA" batteries/4x 1.5V "AAA" batteriesbatteriesbatteries
OperationconditionTemperature: +5 to +40 °C,Humidity: 15 to 85%,Pressure: 70.0k Pa - 106.0kPaTemperature: +5 to+40 °C,Humidity: 15 to85%,Temperature: +5 to+40 °C,Humidity: 10 to90%,Pressure: 86.0k Pa- 106.0k PaSE
Transport/storageenvironmentTemperature:-25°C - +70°C,Humidity:10%~95%,Pressure: 70.0k Pa - 106.0kPaTemperature:-10°C- +55°C,Humidity:10%~95%,Temperature:-20°C- +65°C,Humidity:10%~95%,Pressure: 86.0k Pa- 106.0k PaSE
SafetyIEC 60601-1IEC 60601-1-11IEC 80601-2-30IEC 60601-1IEC 60601-1-11IEC 80601-2-30IEC 60601-1SE
EMCIEC 60601-1-2IEC 60601-1-2IEC 60601-1-2SE
Dimensions(L * D * H, unit:mm)BA4110 (AS-35A): 125 X95 X 60BA6310 (AS-35E):140X105X65A4900 (AS-35W):133 X 75 X 33AS-35H:150 X 110 X 70AS-351:140 X 105 X 62AS-35K:----SENote 2
Elements ofSubject DevicePredicate Device1Predicate Device2Verdict
Comparisont
AS-35X:150 X 110 X 74
AS-35J:150X 110X 74
AS-35L:140 X 113 X 73
AS-35F:140X 10 X 63
AS-55A:78X 62 X 65
AS-55G:90 X 66 X26
AS-55M:89 X 66 X55
AS-55F:63 X 66 X69
Weight(not includebattery)BA4110 (AS-35A): about300g----SENote 2
BA6310 (AS-35E): about300g
A4900 (AS-35W): about300g
AS-35H: about 300g AS-35I:about 250g
AS-35K: about 250g
AS-35X: about 300g
AS-35J: about 328g
AS-35L: about 280g
AS-35F: about 281g
AS-55A: about 95g
AS-55G: about 95g
AS-55M: about 100g
Elements ofComparisonSubject DevicePredicate Device1Predicate Device2Verdict
AS-55F: about 102g

{7}------------------------------------------------

Sponsor:

Xiamen Ants Bro Technology Co., Ltd.

Subject Device:

Automatic Digital Blood Pressure Monitor, Models: BA4110/AS-35A, BA6310/AS-35E, BA4900/AS-35W, AS-35K, AS-35I, AS-35H, AS-35X, AS-35J, AS-35L, AS-35F, AS-55G, AS-55A, AS-55M, AS-55F

{8}------------------------------------------------

Xiamen Ants Bro Technology Co., Ltd. Sponsor:

Subject Device:

Automatic Digital Blood Pressure Monitor, Models: BA4110/AS-35A, BA6310/AS-35E, BA4900/AS-35W, AS-35K, AS-35I, AS-35H, AS-35X, AS-35J, AS-35L, AS-35F, AS-55G, AS-55A, AS-55M, AS-55F

{9}------------------------------------------------

Xiamen Ants Bro Technology Co., Ltd. Sponsor:

Subject Device:

Automatic Digital Blood Pressure Monitor, Models: BA4110/AS-35A, BA6310/AS-35E, BA4900/AS-35W, AS-35K, AS-35I, AS-35H, AS-35X, AS-35J, AS-35L, AS-35F, AS-55G, AS-55A, AS-55M, AS-55F

{10}------------------------------------------------

Xiamen Ants Bro Technology Co., Ltd. Sponsor:

Automatic Digital Blood Pressure Monitor, Models: BA4110/AS-35A, BA6310/AS-35E, Subject Device: BA4900/AS-35W, AS-35K, AS-35I, AS-35H, AS-35X, AS-35J, AS-35L, AS-35F, AS-55G, AS-55A, AS-55M, AS-55F

{11}------------------------------------------------

Sponsor: Xiamen Ants Bro Technology Co., Ltd. Subject Device: Automatic Digital Blood Pressure Monitor, Models: BA4110/AS-35A, BA6310/AS-35E, BA4900/AS-35W, AS-35K, AS-351, AS-35H, AS-35X, AS-35J, AS-35L, AS-35F, AS-55G, AS-55A, AS-55M, AS-55F

Comparison in Detail(s):

Note 1:

Although "Intended Use / Indications for Use" and "Cuff Circumference" of the subject device are a little different from the predicate devices, but the different circumference range of subject device can be covered by the combination range of the predicate device K171833 and K153552, they are very similar. So these parameters' differences will not raise any safety or effectiveness issue.

Note 2:

Although the "Dimensions" and "Weight" of the subject device are a little different from the predicate devices, but all these factors are not the essential parameters of the device which will not affect the effectiveness, and all of them are complied with the safety standards IEC 60601-1. IEC 60601-1-11 and IEC 80601-2-30, which show the safety of subject device, so these parameters' differences will not raise any safety or effectiveness issue.

Final Conclusion:

The subject device Non-invasive Sphygmomanometer (Automatic Digital Blood Pressure Monitor) (Model: BA4110/AS-35A, BA6310/AS-35E, BA4900/AS-35W, AS-35K, AS-35I, AS-35H, AS-35X, AS-35J, AS-35L, AS-35F, AS-55G, AS-55M, AS-55M, AS-55F) has all features of the predicate device. Thus, the subject device is substantially equivalent to the predicate device.

  1. Date of the summary prepared: May 27, 2020

§ 870.1130 Noninvasive blood pressure measurement system.

(a)
Identification. A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.(b)
Classification. Class II (performance standards).