K Number
K190939

Validate with FDA (Live)

Date Cleared
2019-12-19

(253 days)

Product Code
Regulation Number
870.1025
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ZOLL Arrhythmia Management System (AMS) is intended to continuously record, store, and transmit ECG, Heart Rate, Respiration Rate, Activity and Posture. The data provided will aid medical professionals as they diagnose and identify various clinical conditions, events, and/or trends.

AMS is intended for use in clinic and home settings. AMS is indicated for patients who are 21 years of age or older who require monitoring for the detection of non-lethal cardiac arrhythmias, such as, but not limited to, atrial flutter, ventricular ectopy, and bradyarrhythmias.

Device Description

The ZOLL Arrhythmia Management System (AMS) noninvasively monitors patients' clinical parameters (ECG, Heart Rate, Respiration Rate, Activity, and Posture). These raw data are transmitted wirelessly to a remote Server for processing into the clinical parameters. The ZOLL Arrhythmia Management System is for prescription use only. It is intended for use in outpatient clinic and home settings.

The ZOLL Arrhythmia Management System consists of the following components:

  • Sensor - a patient worn device for signal acquisition.
  • Patch – a single use, disposable adhesive piece adhered to the patient's body and allow for Sensor attachment.
  • Charger – the Charger recharges the Sensor and the Gateway. The Sensor typically requires recharging after 5 days. The Gateway typically requires recharging every day.
  • Gateway – An off-the-shelf item, the Gateway is essentially a cellphone that relays data and passes commands between the Sensor and the Server.
  • Server – Server refers to the hardware and the processing software, and resides in a cyber-secure location. The software analyzes the raw data received from the Sensor and processes the data into clinical values for eventual presentation to the physicians via an independent monitoring center (IDTF) that is regulated under 42 CFR 410.33.

Raw data from the Sensor to the Gateway is transmitted via Bluetooth; the Gateway then transmits this data to the Server via TCP/IP over WiFi or cellular network for data processing and analysis.

The ZOLL Arrhythmia Management System Patch and Sensor will be placed on the upper left chest ("front" location).

AI/ML Overview

The acceptance criteria and study proving the device meets them are described below.

1. Table of Acceptance Criteria and Reported Device Performance:

Acceptance Criteria (Performance Requirements)Reported Device Performance (Algorithm Accuracy)
Arrhythmia detection algorithm performance requirements as per ANSI/AAMI EC57:2012Met the algorithm performance requirements.
Ability to capture ECGs (P waves, QRS complexes, PR duration, QRS duration, RR interval) compared to a standard lead-II deviceDemonstrated equivalence in capturing the specified ECG waveform aspects.
Median amplitude of AMS QRS complex compared to standard lead-II QRS complexMedian amplitude of the AMS QRS complex was greater than or equal to the median amplitude of the standard lead-II of a 12-lead ECG comparator device.
Capability to monitor ECG, Heart Rate, Respiration Rate, Posture, and ActivityDemonstrated ability to collect ECG data, measure Heart Rate and Respiration Rate, and classify Posture and Activity.
Software validationSuccessfully validated the device software (API and Analysis Software).

2. Sample Size Used for the Test Set and Data Provenance:

  • Arrhythmia Detection Algorithm Test Set: Specific sample size is not explicitly stated, but the testing was done to ANSI/AAMI EC57:2012. Data provenance is not specified in terms of country of origin or retrospective/prospective nature for this particular test.
  • Equivalency of ECGs Study: 43 volunteers (22 females, 21 males) participated. This was a prospective study. The country of origin is not explicitly stated.
  • Vital Signs Validation Study (Leveraged from Predicate Device): 15 healthy volunteers (9 females, 6 males) participated. This was a prospective study. The country of origin is not explicitly stated.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications:

The document does not explicitly state the number of experts or their qualifications for establishing ground truth for the arrhythmia detection algorithm performance tests or the vital signs validation study.
For the "Equivalency of ECGs" study, the comparison was against a "standard lead-II device," implying established ground truth from a conventional medical device, but not necessarily human expert adjudication for each waveform.

4. Adjudication Method for the Test Set:

The document does not describe any specific adjudication methods (e.g., 2+1, 3+1) for establishing ground truth in any of the studies mentioned. The comparisons appear to be against established standards or conventional medical devices.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:

No, an MRMC comparative effectiveness study was not reported. The studies described focus on the standalone performance of the device or its equivalence to established monitoring methods, not on how human readers improve with AI assistance.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done:

Yes, a standalone (algorithm only) performance evaluation was done for the "ZOLL Arrhythmia Management System detection algorithm" against ANSI/AAMI EC57:2012. Additionally, the equivalency and vital signs validation studies assessed the device's ability to capture and measure physiological parameters independently.

7. The Type of Ground Truth Used:

  • Arrhythmia Detection Algorithm: The ground truth was established by adherence to the performance requirements outlined in ANSI/AAMI EC57:2012. This is a recognized industry standard for arrhythmia detectors.
  • Equivalency of ECGs: The ground truth was established by comparison to a standard lead-II ECG system.
  • Vital Signs Validation Study: The ground truth for ECG, Heart Rate, Respiration Rate, Posture, and Activity was likely derived from accepted measurement methods for these parameters, though not explicitly detailed. The study validated the device's capability to monitor these parameters, implying a comparison to known or referent values.

8. The Sample Size for the Training Set:

The sample size for the training set for the arrhythmia detection algorithm is not provided in the document.

9. How the Ground Truth for the Training Set Was Established:

The document does not describe how the ground truth for the training set was established for the arrhythmia detection algorithm.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

December 19, 2019

ZOLL Manufacturing Corporation % Michael Nilo President and Principal Consultant Nilo Medical Consulting Group 5737 Kentucky Avenue #3 Pittsburgh, Pennsylvania 15232

Re: K190939

Trade/Device Name: ZOLL Arrhythmia Management System Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector and Alarm (Including ST-Segment Measurement and Alarm) Regulatory Class: Class II Product Code: MHX. DSI Dated: November 22, 2019 Received: November 25, 2019

Dear Michael Nilo:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Jessica Paulsen Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K190939

Device Name ZOLL Arrhythmia Management System

Indications for Use (Describe)

The ZOLL Arrhythmia Management System (AMS) is intended to continuously record, store, and transmit ECG, Heart Rate, Respiration Rate, Activity and Posture. The data provided will aid medical professionals as they diagnose and identify various clinical conditions, events, and/or trends.

AMS is intended for use in clinic and home settings. AMS is indicated for patients who are 21 years of age or older who require monitoring for the detection of non-lethal cardiac arrhythmias, such as, but not limited to, atrial flutter, ventricular ectopy, and bradyarrhythmias.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image features the word "ZOLL" in a bold, sans-serif typeface. The letters are a vibrant blue color, which contrasts against the white background. To the bottom right of the word, there is a small registered trademark symbol.

510(k) Summary

510(k) Owner:ZOLL Manufacturing Corporation121 Gamma DrivePittsburgh, PA 51238U. S. A.
Contact:Zachary NelsonSr. Regulatory Affairs Engineer - PremarketPhone: 412-968-3333 x14814Fax: 412-567-9574Email: znelson@zoll.com
Date Summary Prepared:December 17, 2019
Trade Name:ZOLL Arrhythmia Management System
Common Name:Management and Monitoring System
Device ClassificationName:Monitor, Physiological, Patient(with arrhythmia detection or alarms)Arrhythmia Detector and Alarm
Product Code:MHX, DSI
Classification Regulation:870.1025
Device Classification:II
Classification Panel:Circulatory System Devices Panel (74)
Predicate Device:K172510 - μCor Heart Failure and Arrhythmia Management System,ZOLL Manufacturing Corporation
Reference Device:K133701 - NUVANT MCT System, Corventis, Inc.
Comparator Devices:K102507 - Aera CT, TZ Medical, Inc (used for validation)K972795 - Cardio-Card Management System II, NasiffAssociates, Inc. (used for validation)

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the word "ZOLL" in large, bold, blue letters. The letters are sans-serif and appear to be a custom font. A registered trademark symbol is located to the right of the second "L" in the word.

Device Description

The ZOLL Arrhythmia Management System (AMS) noninvasively monitors patients' clinical parameters (ECG, Heart Rate, Respiration Rate, Activity, and Posture). These raw data are transmitted wirelessly to a remote Server for processing into the clinical parameters. The ZOLL Arrhythmia Management System is for prescription use only. It is intended for use in outpatient clinic and home settings.

The ZOLL Arrhythmia Management System consists of the following components:

  • Sensor - a patient worn device for signal acquisition.
  • . Patch – a single use, disposable adhesive piece adhered to the patient's body and allow for Sensor attachment.
  • . Charger – the Charger recharges the Sensor and the Gateway. The Sensor typically requires recharging after 5 days. The Gateway typically requires recharging every day.
  • Gateway – An off-the-shelf item, the Gateway is essentially a cellphone that relays data and passes commands between the Sensor and the Server.
  • Server – Server refers to the hardware and the processing software, and resides in a cyber-secure location. The software analyzes the raw data received from the Sensor and processes the data into clinical values for eventual presentation to the physicians via an independent monitoring center (IDTF) that is regulated under 42 CFR 410.33.

Raw data from the Sensor to the Gateway is transmitted via Bluetooth; the Gateway then transmits this data to the Server via TCP/IP over WiFi or cellular network for data processing and analysis.

The ZOLL Arrhythmia Management System Patch and Sensor will be placed on the upper left chest ("front" location).

Indications for Use Statement

The ZOLL Arrhythmia Management System (AMS) is intended to continuously record, store, and transmit ECG, Heart Rate, Respiration Rate, Activity and Posture. The data provided will aid medical professionals as they diagnose and identify various clinical conditions, events, and/or trends.

AMS is intended for use in clinic and home settings. The AMS is indicated for patients who are 21 years of age or older who require monitoring for the detection of non-lethal cardiac arrhythmias, such as, but not limited to, atrial fibrillation, atrial flutter, ventricular ectopy, and bradyarrhythmias.

Technological Characteristics and Substantial Equivalence Discussion

The AMS technological characteristics are substantially equivalent to K172510 - μCor Heart Failure and Arrhythmia Management System. The devices are non-invasive, prescription use, and indicated for home setting. Both devices have similar intended uses and are identical in form and function. The ZOLL Arrhythmia Management System represents an

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the word "ZOLL" in a bold, sans-serif font. The color of the text is a bright, medium blue. The letters are large and take up most of the image space. The background is plain white.

iterative change from the predicate ZOLL device. The new device also uses a body-worn sensor to acquire the physiological data of interest, with the Gateway to relay the physiological data, and a remote server to process the data. The proposed device prescribes placement on the upper left chest to monitor heart activity only; whereas the predicate device uses a side location to monitor heart activity and thoracic fluid index (TFI). The removal of TFI monitoring and the difference in device placement between the ZOLL Arrhythmia Management System and the uCor Heart Failure and Arrhythmia Management System does not raise new types of safety and effectiveness questions. These differences have been assessed in bench and clinical testing. Results established that the AMS performs as intended and is substantially equivalent to its predicate device. Table 1 provides a comparison chart between the subject and predicate devices.

Table 1: Comparison Chart between Arrhythmia Management System with (AMS) and µCor Heart Failure and Arrhythmia Management System (HFMAS)

AMS(Proposed Device)HFAMSK172510(Predicate Device)ComparisonComments
Intended UseAmbulatory recordingand monitoring ofphysiologicalparametersAmbulatory recording andmonitoring of physiologicalparametersSame
Indications forUse (IFU)The ZOLL ArrhythmiaManagement System(AMS) is intended tocontinuously record,store, and transmitECG, Heart Rate,Respiration Rate,Activity and Posture.The data provided willaid medicalprofessionals as theydiagnose and identifyvarious clinicalconditions, events,and/or trends.The ZOLL ArrhythmiaManagement Systemis indicated forpatients who are 21years of age or olderThe $μ$ Cor Heart Failure andArrhythmia ManagementSystem is intended toperiodically record, store,and transmit Thoracic FluidIndex. The $μ$ Cor HeartFailure and ArrhythmiaManagement System is alsointended to continuouslyrecord and store, andperiodically transmit ECG,Heart Rate, Respiration Rate,Activity and Posture. Thedata provided will aidmedical professionals asthey diagnose and identifyvarious clinical conditions,events, and/or trends.The $μ$ Cor Heart Failure andArrhythmia ManagementIndications foruse have beenupdated toremovemonitoring ofThoracic FluidIndex. Clearedreferencedevices also donot monitorThoracic FluidIndex.
who requiremonitoring for thedetection of non-lethal cardiacarrhythmias, such as,but not limited to,atrial fibrillation, atrialflutter, ventricularectopy, andbradyarrhythmias.System is intended for use inclinic and home settings andis indicated for patients whoare 21 years of age or older(i) who require monitoringfor the detection of non-lethal cardiac arrhythmias,such as, but not limited to,atrial fibrillation, atrialflutter, ventricular ectopy,and bradyarrhythmias or, (ii)require fluid management.
Intended UseEnvironmentHome, clinicHome, clinicSame.
ParametersMonitoredECGECGSame.
Heart RateHeart Rate
Respiration RateRespiration Rate
ActivityActivity
PosturePosture
n/aThoracic FluidThe proposeddevice has thecapability tomonitor ThoracicFluid; however,the API has beenupdated todisable access tothis data.
System Configuration & Components Characteristics
Components- Sensor- Patch- Charger- Gateway- Server- Sensor- Patch- Charger- Gateway- ServerSame.
ProductConfigurationThe Sensor acquiresthe physiological dataof interest. Data istransmitted to theGateway embeddedwithin the Charger,which is thenThe Sensor acquires thephysiological data ofinterest. Data is transmittedto the Gateway embeddedwithin the Charger, which isthen forwarded to theServer for derivation.Same.
forwarded to theServer for derivation.
Body WornComponentForm Factorand WearLocationThe Sensor is attachedto the (adhesive)Patch, which enablesthe patient worneffect.Image: ZOLL patchThe device is placedon the upper leftchest, just below thecollarbone.Image: Chest placementThe Sensor is attached to the(adhesive) Patch, whichenables the patient worneffect.Image: ZOLL patchThe device is placed on the"side" location: below theleft armpit with the nipplealigned anywhere betweenthe top and middle of theSensor.Image: Side placementSimilar. Thebody worncomponent isidentical. ZOLL isproposingplacement in theupper chestlocation.Corventisreference deviceis also worn inthe upper chestwith the sameconfiguration.
DisposablePatch orElectrodesWear Duration5 days for each Patch.5 days for each Patch.Same.
DataTransmissionTechnology-- Sensor toGateway:Bluetooth- Gateway toServer: WiFi orCellular-- Sensor to Gateway:Bluetooth- Gateway to Server: WiFior CellularSame.
Clinical Parameters
ClinicalParametersMonitored-- ECG (script)- Heart Rate (beatsper minute)- Respiration Rate(breaths perminute)-Thoracic Fluid Index- ECG (script)- Heart Rate (beats perminute)- Respiration Rate(breaths per minute)Similar. ThoracicFluid Index is nolonger included.
- Activity ('active'vs. 'rest')- Posture ('supine','recline', or'upright')- Activity ('active' vs.'rest')- Posture ('supine','recline', or 'upright')
ArrhythmiadetectionDetect non-lethalcardiac arrhythmias1,such as, but notlimited to, atrialfibrillation, atrialflutter, ventricularectopy andbradyarrhythmiasDetect non-lethal cardiacarrhythmias, such as, but notlimited to, atrial fibrillation,atrial flutter, ventricularectopy andbradyarrhythmias.

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the word "ZOLL" in blue font. The letters are bold and sans-serif. There is a registered trademark symbol to the right of the second "L".

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the word "ZOLL" in large, blue, sans-serif font. The letters are bold and evenly spaced. A registered trademark symbol is located to the lower right of the second "L".

{8}------------------------------------------------

Image /page/8/Picture/1 description: The image shows the word "ZOLL" in a bold, sans-serif font. The color of the text is a bright, medium blue. A small registered trademark symbol is located to the right of the second "L".

Performance Data

The ZOLL Arrhythmia Management System was evaluated through non-clinical and clinical testing, as well as leveraged applicable data from the predicate device, as summarized below.

Non-Clinical Testing

  • The ZOLL Arrhythmia Management System detection algorithm was tested to ANSI/AAMI EC57:2012. The test results met the algorithm performance requirements and can be found in the Clinical Manual.
  • . Software validation testing of the API and Analysis Software. The testing successfully validated the device software.

Non-Clinical Testing leveraged from the predicate device

The testing below was leveraged directly from the predicate device, as the ZOLL Arrhythmia Management System uses identical hardware and much of the same software as the predicate device.

  • Mechanical bench testing of the Sensor and Patch
  • Respiration Rate accuracy bench testing
  • Safety testing to: .
    • o AAMI/ANSI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012, and A2:2010/(R)2012
    • o IEC 60601-1-11 Ed. 2.0 2015-01
    • o AAMI/ANSI/IEC 60601-2-47:2012
  • Electromagnetic Compatibility testing to AAMI/ANSI/IEC 60601-1-2:2014
  • . Wireless co-existence testing

1 AMS detects arrhythmias and 2nd and 3rd degree AV nodal conduction disorders. This single lead ECG device should not be considered a substitute for a 12-lead ECG for assessing ECG intervals (ex. intraventricular conduction delays, QT intervals).

{9}------------------------------------------------

Image /page/9/Picture/1 description: The image contains the word "ZOLL" in a large, sans-serif font. The word is a bright, solid blue color. The letters are bold and evenly spaced, creating a clear and easily readable logo. The background is plain white.

  • Biocompatibility testing
  • Shelf-Life Testing
  • Shipping and Packaging Testing
  • Software Testing ●

Clinical Testing

  • Equivalency of Electrocardiograms (ECGs) Recorded by the ZOLL Arrhythmia Management System to that of a Standard Lead II ECG System This prospective, non-significant risk, premarket study demonstrated the ability of the ZOLL Arrhythmia Management System to capture ECGs as compared to a standard lead-II device. During the study 43 volunteers (22 females, 21 males) wore the AMS and a lead-II device simultaneously in sitting, supine, and standing positions. There were no adverse effects of complications during the study. The study demonstrated that the AMS was equivalent in capturing the below aspects of the ECG waveform, as compared to the standard lead II device:
    • P waves O
    • O QRS complexes
    • PR duration O
    • QRS duration O
    • O RR interval

The median amplitude of the AMS QRS complex was greater than or equal to the median amplitude of the standard lead-II of a 12-lead ECG comparator device.

Clinical Testing leveraged from the predicate device

  • . Vital Signs Validation Study of the uCor System (ViVUS)
    This prospective, non-significant risk, premarket study validated the capability of the device to monitor ECG, Heart Rate, Respiration Rate, Posture, and Activity. During the study, 15 healthy volunteers (9 females, 6 males) wore the device and were asked to perform the following activities: breathing, walking, and resting. Respiration Rate, ECG, Heart Rate, Activity, and Posture were collected during these activities. There were no adverse effects or complications during the study. The results demonstrated the ability of the device to collect ECG data, measure Heart Rate and Respiration Rate, and classify Posture and Activity.

The results from this study apply to the ZOLL Arrhythmia Management System, as the Arrhythmia Management System utilizes the same accelerometer and ECG acquisition circuitry as the predicate device.

Conclusion

The ZOLL Arrhythmia Management System is substantially equivalent to the μCor Heart Failure and Arrhythmia Management System (K172510). Both devices are both intended to

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the word "ZOLL" in large, blue, sans-serif font. The letters are bold and evenly spaced. A registered trademark symbol is located to the right of the second "L".

continuously record and store, and periodically transmit ECG, Heart Rate, Respiration Rate, Activity and Posture, as well as monitor and detect non-lethal cardiac arrhythmias. Both devices are identical in form. The ZOLL Arrhythmia Management System has removed the ability to analyze Thoracic Fluid Index in order to allow for placement of the device on the upper chest, which allows for appropriate detection of non-lethal arrhythmias.

§ 870.1025 Arrhythmia detector and alarm (including ST-segment measurement and alarm).

(a)
Identification. The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.(b)
Classification. Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.