K Number
K190932

Validate with FDA (Live)

Date Cleared
2019-09-13

(156 days)

Product Code
Regulation Number
878.4300
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Tag of the RFLS is intended for percutaneous placement in the breast to mark (>30 days) a lesion intended for surgical removal. Using image guidance (such as ultrasound or radiography) or aided by non-imaging guidance (RFLS), the RFID Tag is located and surgically removed with the target tissue. The RFLS is intended only for the non-imaging detection and localization of the Tag that has been implanted in a lesion intended for surgical removal.

Device Description

The Tag, Tag Applicator, Tag Applicator S, LOCalizer Reader, and LOCalizer Surgical Probe are components of the Health Beacons RFID Localization System (RFLS). The proposed device is a marker-with-detector localization device that employs miniature RFID tags as markers and a handheld reader that can measure distance to the Tag. The Tag, when used in conjunction with the Reader and Surgical Probe, can be used as a guide for the surgeon during the excision of tissue. The RFLS is a prescription device meant only for use by trained professionals, specifically breast surgeons and diagnostic radiologists.

AI/ML Overview

This document describes the 510(k) summary for the Health Beacons, Inc. RFID Localization System (RFLS). However, based on the provided text, there is no detailed information about acceptance criteria and a specific study proving the device meets those criteria, especially in the context of AI assistance or human reader performance improvement.

The document is a 510(k) summary for a medical device (RFID Localization System) intended to mark and locate lesions in the breast for surgical removal. While it mentions performance testing was conducted to support substantial equivalence, it does not provide the specific acceptance criteria or the results from those tests in a format that would allow filling out the requested table or answering many of the follow-up questions.

The device itself is an RFID localization system, which uses miniature RFID tags as markers and a handheld reader to measure distance to the tag. This is a physical localization device, not an AI/software-based diagnostic tool that would typically involve acceptance criteria related to sensitivity, specificity, or human reader improvement with AI assistance.

Therefore, many of the requested details, particularly those pertaining to AI/ML algorithms (e.g., acceptance criteria for diagnostic accuracy, standalone algorithm performance, MRMC studies, training/test set details, ground truth for AI models), are not applicable or not present in this specific 510(k) summary.

Here's what can be extracted and what cannot, based on the provided text:

1. A table of acceptance criteria and the reported device performance:

  • Not explicitly provided in the document. The document states that "All verification and validation activities identified as necessary were performed... and results demonstrate that predetermined acceptance criteria were met." However, the specific criteria and the numerical performance results are not tabulated or detailed.
  • The "performance testing" subsections (VIII) list categories of tests but not acceptance criteria or outcomes. These tests include:
    • Magnetic field emission testing per IEC 60601-1-2:2014
    • Delivery testing
    • Deployment Force testing
    • Needle Penetration Force testing
    • Usability testing

2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective):

  • Not provided. The document states "performance testing was provided," but offers no details on sample size, study design (retrospective/prospective), or data provenance.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Not applicable/Not provided. This is relevant for diagnostic AI/ML devices where human expert consensus often establishes ground truth. This device is a physical localization system, not a diagnostic imaging AI.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

  • Not applicable/Not provided. Same reasoning as point 3.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • No. This device is a physical localization system; it's not an AI assisting human readers with diagnostic tasks.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • Not applicable/Not provided. The device is a "marker-with-detector localization device" that involves a Tag and a Reader/Probe used by a surgeon. There isn't an "algorithm only" component in the sense of a standalone diagnostic AI.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

  • Not explicitly stated for the listed performance tests. For a physical device, ground truth for performance tests would likely involve physical measurements (e.g., actual vs. measured distance, force applied, magnetic field readings).

8. The sample size for the training set:

  • Not applicable/Not provided. This device is not an AI/ML algorithm that requires a training set.

9. How the ground truth for the training set was established:

  • Not applicable/Not provided. Same reasoning as point 8.

In summary of the provided text:

This FDA 510(k) summary describes a physical medical device (RFID Localization System), not an AI/ML diagnostic software. The document asserts that performance testing was conducted and met predetermined acceptance criteria, which supported the substantial equivalence determination to a predicate device. However, it does not provide the specific numerical acceptance criteria or the detailed results of those performance tests. Information regarding sample sizes, data provenance, expert involvement for ground truth, or MRMC studies (which are highly relevant for AI/ML diagnostic devices) is absent as these concepts are not directly applicable to the type of device described.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

September 13, 2019

Health Beacons, Inc. % Ms. Carol Vierling Sr. Principal Advisor R&Q Solutions 2790 Mosside Blvd, Suite 800 Monroeville, Pennsylvania 15146

Re: K190932

Trade/Device Name: RFID Localization System Regulation Number: 21 CFR 878.4300 Regulation Name: Implantable Clip Regulatory Class: Class II Product Code: NEU Dated: May 3, 2019 Received: August 13, 2019

Dear Carol Vierling:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For Cindy Chowdhury, Ph.D., M.B.A. Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K190932

Device Name RFID Localization System (RFLS)

Indications for Use (Describe)

The Tag of the RFLS is intended for percutaneous placement in the breast to mark (>30 days) a lesion intended for surgical removal. Using image guidance (such as ultrasound or radiography) or aided by non-imaging guidance (RFLS), the RFID Tag is located and surgically removed with the target tissue. The RFLS is intended only for the non-imaging detection and localization of the Tag that has been implanted in a lesion intended for surgical removal.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) SUMMARY

This 510(k) Summary is provided per the requirements of section 21 CFR 807.92 on April 8, 2019.

I. Submitter

Submitter's Name:Health Beacons, Inc.
Contact Person:Ms. Nancy ConfreyChief Executive Officer
Address:34 Walden St., #753Concord, MA 01742
Telephone:(978) 287-4635
Fax:(978) 246-6019
Email:nconfrey@healthbeacons.com

II. Application Correspondent

Contact's Name:Regulatory and Quality Solutions, LLC
Contact Person:Ms. Carol Vierling, RAC (Consultant)Senior Principal Advisor
Address:2790 Mosside Blvd #800Monroeville, PA 15146
Telephone:877-652-0830
Email:cvierling@rqteam.com

III. Device

Trade Name:RFID Localization System (RFLS)
Common Name:Marker, Radiographic, Implantable
Classification Name:Implantable clip.
Product Classification:Class II, §878.4300, Product Code NEU

Health Beacons, Inc.

{4}------------------------------------------------

IV. Predicate Device

  • . Health Beacons RFID Localization System - 5cm Tag Applicator, 7cm Tag Applicator, 10 Cm Tag Applicator, 5 Cm Tag Applicator (10 Pack), 7cm Tag Applicator (10 Pack), 10cm Tag Applicator (10 Pack), LOCalizer Surgical Probe, LOCalizer Surgical Probe (5 Pack)
    • K181692 (Health Beacons Inc.), FDA cleared on September 25th, 2018 O

V. Device Description

The Tag, Tag Applicator, Tag Applicator S, LOCalizer Reader, and LOCalizer Surgical Probe are components of the Health Beacons RFID Localization System (RFLS). The proposed device is a marker-with-detector localization device that employs miniature RFID tags as markers and a handheld reader that can measure distance to the Tag. The Tag, when used in conjunction with the Reader and Surgical Probe, can be used as a guide for the surgeon during the excision of tissue. The RFLS is a prescription device meant only for use by trained professionals, specifically breast surgeons and diagnostic radiologists.

VI. Indications for Use

The Tag of the RFLS is intended for percutaneous placement in the breast to mark (>30 days) a lesion intended for surgical removal. Using image guidance (such as ultrasound or radiography) or aided by non-imaging guidance (RFLS), the RFID Tag is located and surgically removed with the target tissue. The RFLS is intended only for the non-imaging detection and localization of the Tag that has been implanted in a lesion intended for surgical removal.

There have been no changes to the indication for use.

Product FeaturesModified DeviceHealth Beacons, Inc. RFIDLocalization SystemPredicate DeviceHealth Beacons RFID LocalizationSystem (K181692)
Device DescriptionThe proposed RFID Localization Systemis a marker-with-detector localizationdevice that employs miniature RFID tagsas markers and a hand-held reader thatcan measure distance to the tag. TheRFLS is comprised of a Tag, TagApplicator, Tag Applicator S, LOCalizerReader (Reader), and LOCalizerSurgical Probe (Surgical Probe). TheTag, when used in conjunction with theReader and Surgical Probe, can be usedas a guide for the surgeon to refer to inthe excision of tissue. The RFID tagemits no radiation, the Reader providesan accurate distance measurement to thetag, and the RFID Tag can be placed at aconvenient time before surgery withoutmigration prior to removal.Same except that there is only one TagApplicator
Product FeaturesModified DeviceHealth Beacons, Inc. RFIDLocalization SystemPredicate DeviceHealth Beacons RFID LocalizationSystem (K181692)
Intended UseThe Tag of the RFLS is intended forpercutaneous placement in the breast tomark a lesion intended for surgicalremoval.-same-
Indications for UseThe Tag of the RFLS is intended forpercutaneous placement in the breast tomark (>30 days) a lesion intended forsurgical removal. Using image guidance(such as ultrasound or radiography) oraided by non-imaging guidance (RFLS),the RFID Tag is located and surgicallyremoved with the target tissue. TheRFLS is intended only for the non-imaging detection and localization of theTag that has been implanted in a lesionintended for surgical removal.-same-
ClassificationClass II-same-
Product CodeNEU-same-
Regulation Number§878.4300-same-
Regulation NameMarker, Radiographic, Implantable-same-
Anatomical LocationsBreast Tissue-same-
Breast TissuePenetration /DeliveryDeviceNeedle Implanter-same-
VisibilityX-ray, Ultrasound and MRI (artifactonly)-same-
Primary DeviceComponentsRFID Tag pre-loaded in Tag Applicator,Reader and Surgical Probe-same-
Principle of OperationRadiofrequency wave technology todetect tissue marker-same-
IndicatorsVisual and Audible-same-
Implant DeliveryDevice Size/WorkingLength• 12 Gauge/5cm• 12 Gauge/7cm• 12 Gauge/10cm-same-
Implant DeliveryDevice NeedleMarking1 cm marker increments-same-
Tag Applicator TipGeometry22° bevel28° bevel
Tag ImplantDimensionsApproximately 0.08" (2.29mm) diameterx 0.43" (10.8mm) longApproximately 0.09" (2.26 mm)diameter x 0.37" (9.52 mm) long
Product FeaturesModified DeviceHealth Beacons, Inc. RFIDLocalization SystemPredicate DeviceHealth Beacons RFID LocalizationSystem (K181692)
Patient ContactingMaterials• Tag: Soda lime Bioglass, Kimble R6,Biobond™, (a.k.a. Pro-fax PF-531)Polypropylene Homopolymer• Tag Applicator: 304 Stainless Steelwith silicone coating• Tag Applicator S: 304 Stainless Steelwith silicone coating• Surgical Probe: AcrylonitrileButadiene Styrene (ABS), SabicCycolac HMG94-8H7D195, 304Stainless Steel, Thermoplasticpolyurethane (TPU) IROGRANR A75E 5040, white, Silicone Elastomer• Reader: Polycarbonate, LTL ColorCompounds Colorfast PC200,Polyester, Flexcon Compucal Excel10442 Label Stock• Tag: -same-• Tag Applicator: 304 Stainless Steel• Surgical Probe: -same-• Reader: -same-
Sterilization Method(sterile, single-usecomponents)• Tag Applicator and RFID Tag:Ethylene Oxide• Surgical Probe: Gamma-same-
Tissue MarkerLocatorReusable, non-sterile Console-same-

VII. Comparison of Technological Characteristics with the Predicate Device

{5}------------------------------------------------

{6}------------------------------------------------

VIII. Design Control Activities

A risk analysis was performed to identify new risks based on the device modification. Risks are mitigated as far as possible. All verification and validation activities identified as necessary were performed by designated individuals and results demonstrate that predetermined acceptance criteria were met. The manufacturing facility fulfills design procedure requirements.

The following performance testing was provided to support the substantial equivalence determination.

  • Magnetic field emission testing per IEC 60601-1-2:2014
  • . Delivery testing
  • Deployment Force testing .
  • Needle Penetration Force testing .
  • . Usability testing

IX. Conclusion

The proposed RFID Localization System (RFLS) has the same indications for use, principles of operation and fundamental scientific technology as the predicate device, Health Beacons RFID Localization System. The proposed changes do not raise any new questions regarding safety and effectiveness of the RFLS. Therefore, the information provided in this submission supports the RFLS being as safe and effective as the predicate device for its intended use and demonstrates that the device is substantially equivalent to its predicate.

Health Beacons, Inc.

§ 878.4300 Implantable clip.

(a)
Identification. An implantable clip is a clip-like device intended to connect internal tissues to aid healing. It is not absorbable.(b)
Classification. Class II.