(100 days)
The Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Model BP3GY1-2N is a device intended to measure the systolic and diastolic blood pressure, pulse rate of an adult individual with arm circumference sizes ranging from 22 -42 cm by using a non-invasive oscillometric technique in which an inflatable cuff is being wrapped around the upper arm.
The device is also validated for all adult diabetic users.
The device detects the appearance of irregular heartbeat during measurement, and gives a warning signal with the reading once the irregular heartbeat is detected.
The device can be used in connection with a smart phone running the APP. The memory data can be transferred to the smart phone via Bluetooth.
The blood pressure monitor is a fully automatic digital blood presure measuring device for use by adults (also applicable for diabetic patients) on the upper arm at home or in your doctor's/nurse's office.
Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Model BP3GY1-2N is designed to measure systolic and diastolic blood pressure, pulse rate of an individual with arm circumference sizes ranging from 22 -42 cm by using a noninvasive technique in which one inflatable cuff is wrapped around the single upper arm. Our method to define systolic and diastolic pressure is similar to the auscultatory method but using a semiconductor sensor rather than a stethoscope and mercury manometer. The sensor converts tiny alterations in cuff pressure to electrical signals, by analyzing those signals to define the systolic and diastolic blood pressure and calculating pulse rate, which is a well - known technique in the market called the "oscillometric method".
The device detects the appearance of irregular heartbeat during measurement, and the symbol " * " is displayed after the measurement. In addition, the device can be used in connection with smart mobile devices running the APP and via Bluetooth.
The device is also validated for all adult diabetic users.
The blood pressure monitor is a fully automatic digital blood pressure measuring device for use by adults(also applicable to the diabetic patients) on the upper arm at home or in your doctor's/nurse's office.
Here's an analysis of the acceptance criteria and study information for the Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Model BP3GY1-2N, based on the provided text:
Important Note: The provided document is a 510(k) summary for a new device (BP3GY1-2N) claiming substantial equivalence to a predicate device (BP3GX1-5X). Therefore, the "study that proves the device meets the acceptance criteria" refers to the predicate device's clinical validation study, which the subject device is leveraging. The new device itself did not undergo a separate clinical validation study for blood pressure accuracy.
1. Table of Acceptance Criteria and Reported Device Performance
The core acceptance criterion for blood pressure monitors is typically defined by standards like ANSI/AAMI/IEC 81060-2. While the exact numerical criteria are not explicitly stated in this document (they would be detailed in the standard itself), it is generally understood to involve accuracy metrics for systolic and diastolic blood pressure measurements compared to a reference method.
The document indicates that the device's performance is demonstrated by its substantial equivalence to a clinically validated predicate device.
| Acceptance Criteria (from ANSI/AAMI/IEC 81060-2) | Reported Device Performance (via predicate) |
|---|---|
| Blood Pressure Accuracy: | Blood Pressure Accuracy: |
| Not explicitly detailed in the provided text, but generally involves: - Mean difference (device vs. reference) and standard deviation for systolic and diastolic BP - Within-subject differences (e.g., % of readings within 5 mmHg) | The subject device (BP3GY1-2N) is technically identical to the primary predicate device (BP3GX1-5X, K183469) concerning detection algorithm and fundamental scientific technology. The predicate device was clinically validated according to ANSI/AAMI/IEC 81060-2 and proved accurate for blood pressure detection. Also validated for adult diabetic users. |
| Irregular Heartbeat Detection: | Detects and signals irregular heartbeat during measurement. |
| Functional / Safety Standards Compliance: | Complies with numerous IEC, ISO, and AAMI standards for electrical safety, EMC, risk management, biocompatibility, and home healthcare environments (listed in section 7). |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: Not explicitly stated in this document for the predicate device's clinical study. The document refers to the predicate's 510(k) (K183469) where this information would be found. Standards like ANSI/AAMI/IEC 81060-2 require specific sample sizes (e.g., at least 85 subjects).
- Data Provenance: Not specified in the provided text. The location of the original clinical validation study for the predicate device is not mentioned.
- Retrospective or Prospective: Not specified in the provided text for the predicate device's clinical study.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications
- Number of Experts: Not explicitly stated. For clinical validation of blood pressure devices based on standards like ANSI/AAMI/IEC 81060-2, multiple qualified observers (e.g., physicians or specially trained technicians) are typically used for the auscultatory reference measurements.
- Qualifications of Experts: Not explicitly stated in this document. However, standard clinical validation protocols require observers to be trained and certified in auscultatory blood pressure measurement.
4. Adjudication Method for the Test Set
- Adjudication Method: Not explicitly stated. In auscultatory reference measurements, it is common to have multiple observers taking measurements simultaneously, and their readings are often averaged or assessed for agreement. The specific method would be detailed in the predicate device's clinical study report.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- Was an MRMC study done? No, an MRMC comparative effectiveness study for human readers with and without AI assistance was not conducted. This device is a standalone blood pressure monitor, not an AI-assisted diagnostic imaging device that would typically involve human readers interpreting cases.
- Effect size of human readers improved with AI vs. without AI assistance: Not applicable.
6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study
- Was a standalone study done? Yes. The clinical validation of the predicate device (BP3GX1-5X) against the ANSI/AAMI/IEC 81060-2 standard represents a standalone performance study. This standard assesses the accuracy of the automated device (algorithm only) compared to a reference standard (auscultation), without human interpretation of the device's output influencing the measurement itself. The subject device BP3GY1-2N is leveraging this standalone clinical data from the predicate.
7. Type of Ground Truth Used
- Type of Ground Truth: The ground truth for blood pressure accuracy in studies following ANSI/AAMI/IEC 81060-2 is established using auscultatory measurements performed by trained human observers using a stethoscope and manometer. This is considered the clinical reference standard for non-invasive blood pressure.
8. Sample Size for the Training Set
- Sample Size for Training Set: This information is not provided in the document. For a medical device like this, the "training set" typically refers to the data used to develop and refine the oscillometric algorithm. This data is usually proprietary to the manufacturer and not typically disclosed in 510(k) summaries unless specifically required for algorithm-based devices where the training data directly impacts performance claims.
9. How the Ground Truth for the Training Set Was Established
- How Ground Truth for Training Set Was Established: This information is not provided in the document. Similar to the sample size, the process for establishing ground truth for the internal development and training of the oscillometric algorithm is generally proprietary. It would likely involve similar auscultatory reference measurements to those used in the final clinical validation, but potentially in a more controlled or varied dataset for algorithm optimization.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 10, 2019
Microlife Intellectual Property GmbH % Susan Goldstein-Falk Official Correspondent for Microlife Intellectual Property. GmbH Mdi Consultants Inc. 55 Northern Blvd Great Neck. New York 11021
Re: K190818
Trade/Device Name: Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Model BP3GY1-2N Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: June 10, 2019 Received: June 11, 2019
Dear Susan Goldstein-Falk:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K K190818
Device Name
Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Model BP3GY 1-2N
Indications for Use (Describe)
The Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Model BP3GY1-2N is a device intended to measure the systolic and diastolic blood pressure, pulse rate of an adult individual with arm circumference sizes ranging from 22 -42 cm by using a non-invasive oscillometric technique in which an inflatable cuff is being wrapped around the upper arm.
The device is also validated for all adult diabetic users.
The device detects the appearance of irregular heartbeat during measurement, and gives a warning signal with the reading once the irregular heartbeat is detected.
The device can be used in connection with a smart phone running the APP. The memory data can be transferred to the smart phone via Bluetooth.
The blood pressure monitor is a fully automatic digital blood presure measuring device for use by adults (also applicable for diabetic patients) on the upper arm at home or in your doctor's/nurse's office.
| Type of Use (Select one or both, as applicable) | |
|---|---|
| Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. | |
| DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW. |
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
VOL 4, 001
510(k) SUMMARY
The assigned 510(k) number is:
1. Submitter's Identification:
Microlife Intellectual Property GmbH, Switzerland Espenstrasse 139 9443 Widnau / Switzerland
Date Summary Prepared: June 7, 2019
Contact: Mr. Gerhard Frick Vice President of Technical and Service Microlife Intellectual Property GmbH, Switzerland Tel: +41 79 216 0070 E-Mail: gerhard.frick@microlife.ch
2. Name of the Device:
Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Model BP3GY1-2N
Regulation Number: 21 CFR Part 870.1130 Regulation Name: Non-Invasive Blood Pressure Measurement System Regulatory Class: II Product Code: DXN
3. Information for the 510(k) Cleared Device (Predicate Device):
Primary Predicate:
- a. Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Model BP3GX1-5X (BP A3 PC), K183469, Microlife Intellectual Property GmbH.
Reference Predicate:
b. Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Model BP3MW1-4B, K153077, Microlife Intellectual Property GmbH.
4. Device Description:
Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Model BP3GY1-2N is designed to measure systolic and diastolic blood pressure, pulse rate of an individual with arm circumference sizes ranging from 22 -42 cm by using a noninvasive technique in which one inflatable cuff is wrapped around the sinqle
{4}------------------------------------------------
upper arm. Our method to define systolic and diastolic pressure is similar to the auscultatory method but using a semiconductor sensor rather than a stethoscope and mercury manometer. The sensor converts tiny alterations in cuff pressure to electrical signals, by analyzing those signals to define the systolic and diastolic blood pressure and calculating pulse rate, which is a well - known technique in the market called the "oscillometric method".
The device detects the appearance of irregular heartbeat during measurement,
and the symbol " *** *is displayed after the measurement. In addition, the device can be used in connection with smart mobile devices running the APP and via Bluetooth.
The device is also validated for all adult diabetic users.
The blood pressure monitor is a fully automatic digital blood pressure measuring device for use by adults(also applicable to the diabetic patients) on the upper arm at home or in your doctor's/nurse's office.
5. Indications for Use:
The Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Model BP3GY1-2N is a device intended to measure the systolic and diastolic blood pressure, pulse rate of an adult individual with arm circumference sizes ranging from 22 -42 cm by using a non-invasive oscillometric technique in which an inflatable cuff is being wrapped around the single upper arm.
The device is also validated for all adult diabetic users.
The device detects the appearance of irregular heartbeat during measurement, and gives a warning signal with the reading once the irregular heartbeat is detected.
The device can be used in connection with a smart phone running the APP. The memory data can be transferred to the smart phone via Bluetooth.
The blood pressure monitor is a fully automatic digital blood pressure measuring device for use by adults (also applicable for diabetic patients) on the upper arm at home or in your doctor's/nurse's office.
6. Comparison to the 510(k) Cleared Device (Predicate Device):
Subject (Modified) Device Compared to Primary Predicate BP3GX1-5X (BP A3 PC) (K183469):
The subject BP3GY1-2N uses the same oscillometric method as the predicate BP3GX1-5X (BP A3 PC) with the same algorithm to determine the systolic and diastolic blood pressure, pulse rate. Upper arm cuff is inflated automatically by pump, the deflation rate is controlled by factory set exhaust valve and the deflation pressures are transferred via tubing to a sensor in these two units.
{5}------------------------------------------------
The subject BP3GY1-2N and the predicate original BP3GX1-5X (BP A3 PC) both have MAM function, IHD function traffic light function and they are both validated for all adult diabetic users. They differ by the PC-link function. Bluetooth function and the way to operate the device. The PC-link function is removed from the subject device. The Bluetooth function is added to the subject device, but it's only a way to transfer the data. The way to operate the predicate device BP3GX1-5X (BP A3 PC) is to press the button, and the subject device requires a light touch of the finger on the screen. However, those differences do not affect the accuracy and normal use of this device because they use the same fundamental scientific technology. The declaration of identity and features comparison table (VOL 10, 001, VOL 10, 002) demonstrate that the subject device BP3GY1-2N can leverage the clinical test report of BP3GX1-5X (BP A3 PC) which was proved in K183469. Therefore, repeated clinical testing in accordance with the standard ANSI/AAMI/IEC 81060-2 for BP3GY1-2N is not warranted.
Based upon the aforementioned information, the two devices are substantially equivalent.
Subject (Modified) Device Compared to Reference Predicate BP3MW1-4B (K153077):
The modified device model BP3GY1-2N uses the same oscillometric method as the predicate device BP3MW1-4B with the same algorithm to determine the systolic and diastolic blood pressure, pulse rate. Upper arm cuff is inflated automatically by pump, the deflation rate is controlled by factory set exhaust valve and the deflation pressures are transferred via tubing to a sensor in these two units.
They both have MAM function, IHD function, traffic light function and Bluetooth function. They have the same fundamental scientific technology.
Based upon the aforementioned information, the two devices are substantially equivalent.
7. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence are as follows:
Testing information demonstrating safety and effectiveness of the Microlife Upper Arm Automatic Digital Blood Pressure Monitor. ModelBP3GY1-2N in the intended environment of use is supported by testing that was conducted in accordance with the FDA November 1993 Draft "Reviewer Guidance for Premarket Notification Submissions", DCRND, which outlines Electrical, Mechanical and Environmental Performance requirements.
The following testing was conducted to prove safety and effectiveness as well as substantial equivalence to the predicate devices:
The following National and International Standards were utilized for testing the
{6}------------------------------------------------
subject device:
-
- IEC 60601-1 Medical electrical equipment Part 1: General requirements for basic safety and essential performance AAMI / ANSI ES60601-1:2005/(R) 2012 And A1:2012, C1:2009/(R) 2012 And A2:2010/(R) 2012
-
- IEC 60601-1-2:2014 Medical electrical equipment Part 1-2: General requirements for safety and essential performance - Collateral standard: Electromagnetic Disturbances - Requirements And Tests.
-
- ISO 14971: 2007 Medical devices Application of risk management o medical devices.
-
- AAMI/ANSI/ISO 10993-1:2009/(R)2013, Biological evaluation of medical devices - Part 1: Evaluation And Testing Within A Risk Management Process.
-
- AAMI/ANSI/ISO 10993-5:2009/(R)2014, Biological evaluation of medical devices - Part 5: Tests for In Vitro Cytotoxicity.
-
- AAMI / ANSI / ISO 10993-10:2010/(R)2014,, Biological evaluation of medical devices - Part 10: Tests for Irritation and Skin Sensitization
-
- AAMI/ANSI/IEC 80601-2-30 Medical electrical equipment Part 2-30: Particular requirements for the basic safety and essential performance of automated noninvasive sphygmomanometers, 2013
-
- IEC 60601-1-11:2015 Medical Electrical Equipment Part 1-11: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Requirements For Medical Electrical Equipment And Medical Electrical Systems Used In The Home Healthcare Environment
-
- AAMI/ANSI/ISO 81060-2 Non-Invasive Sphygmomanometers Part 2: Clinical Validation of Automated Measurement Type. 2013
None of the testing demonstrated any design characteristics that violated the requirements of the Reviewer Guidance or resulted in any safety hazards. It was our conclusion that Microlife Upper Arm Automatic Digital Blood Pressure Monitor, ModelBP3GY1-2N tested met all relevant requirements of the aforementioned tests.
8. Discussion of Clinical Tests Performed:
From a clinical validation standpoint, the subject device is identical to the 510(k) cleared primary predicate device, BP3GX1-5X (BP A3 PC), K183469, in detection.
Regarding clinical validation concerning the compliance of ANSI/AAMI/IEC 81060-2, the subject blood pressure monitor Model BP3GY1-2N is, from a technical point of view, identical to the predicate blood pressure Monitor Model BP3GX1-5X (BP A3 PC).
The only differences between the two models are in (please see VOL 10, 002 for more details):
{7}------------------------------------------------
- Model name and shape ●
- Touch screen technology ●
- Bluetooth function ●
- No PC-link function .
The subject device BP3GY1-2N has added the function of transferring the memory data to the smart phone via Bluetooth. However, this function is only a way to transfer the data and will not affect the clinical accuracy. The other differences listed above do not affect the accuracy and normal use of this device because they use the same fundamental scientific technology. That said, BP3GY1-2N can leverage the clinical data of the primary predicate device BP3GX1-5X (BP A3 PC) which was proven in K183469 according to the standard ANSI/AAMI/IEC 81060-2 for the accuracy of blood pressure detection and validated in diabetics function. Therefore, repeated clinical testing in accordance with the standard ANSI/AAMI/IEC 81060-2 for BP3GY1-2N is not warranted.
9. Software information:
Software validation was conducted in accordance with a moderate level of concern designation in accordance with the FDA November 2005 document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices".
10. Conclusions:
Conclusions drawn from the non-clinical and clinical tests demonstrate that the subject device is as safe, effective, and performs as well as the predicate devices.
§ 870.1130 Noninvasive blood pressure measurement system.
(a)
Identification. A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.(b)
Classification. Class II (performance standards).