K Number
K190788

Validate with FDA (Live)

Date Cleared
2019-04-18

(22 days)

Product Code
Regulation Number
876.1500
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Standard Clamp, Disposable is indicated for use in laparoscopic procedures to grasp, clamp, and manipulate soft tissues. The Clamp is indicated for use in bariatric procedures such as sleeve gastrectomy as a guide, to clamp and manipulate flat tissue and organs, such as the stomach.

Device Description

The Standard Clamp, Disposable is a disposable non-energized, hand held surgical instrument designed for use in laparoscopic procedures to clamp long planes of soft tissue such as the stomach. The Standard Clamp, Disposable is provided sterile (gamma). There are no accessories provided with the device. The instrument is comprised of three main sections: handle, shaft, and end effector. The handle is a pistol grip with a trigger and release button which are manually activated in order to open and close the end effector. The shaft is sized to fit a standard 12 mm trocar. The clamp is comprised of an upper and lower jaw that close to grasp and hold tissue. The environment of use is in a hospital during laparoscopic surgical procedures. The current Standard Clamp, Disposable has a jaw length of 25 cm. This Special 510(k) is being filed to add a new shorter jaw length (22 cm) to accommodate patients with smaller abdomens.

The Standard Clamp, Disposable is intended to be used by a surgeon on a single patient during a single laparoscopic surgery under normal operating conditions. The Standard Clamp, Disposable is intended to be inserted into the peritoneal cavity through a 12mm trocar and used in up to six cycles of opening and closure to grasp, clamp and manipulate long, 1.2 to 4.5mm thick, planes of soft tissues such as the stomach. The Standard Clamp, Disposable is not intended for direct contact with the cardiovascular system, lymphatic system or cerebrospinal fluid. The Standard Clamp, Disposable can be used while a surgeon uses devices such as a stapler adjacent to tissue that is clamped with the Standard Clamp, Disposable, e.g., to help guide endoscopic staplers during resection of tissue such as in sleeve gastrectomy.

AI/ML Overview

The provided text is a 510(k) summary for the Standard Clamp, Disposable surgical instrument. The document discusses the addition of a new, shorter jaw length (22 cm) to an existing device (Standard Clamp, Disposable 25 cm), and it asserts substantial equivalence to the predicate device.

However, the document does not contain the detailed type of study information requested in your prompt. Specifically, it does not include:

  • A table of acceptance criteria and reported device performance (beyond a general statement that it "meets the same specifications").
  • Sample sizes for test sets or data provenance.
  • Number and qualifications of experts or adjudication methods.
  • Multi-reader multi-case (MRMC) comparative effectiveness study results.
  • Standalone (algorithm-only) performance data.
  • Specific ground truth types for testing or training sets.
  • Sample size for a training set.
  • How ground truth for the training set was established.

Instead, the document states:

"Standard Clamp, Disposable performance bench testing was provided in K170379. Testing for the Standard Clamp, Disposable 22 cm is provided in Section 018 - Performance Testing - Bench of this submission to the same specifications as the Standard Clamp, Disposable 25 cm previously provided in K 170379. The data demonstrates that the Standard Clamp, Disposable 22 cm meets the same specifications for clamping and slip force as the Standard Clamp, Disposable 25 cm."

This indicates that bench testing was performed to compare the new device to the predicate device based on previously established specifications for "clamping and slip force." The specifics of these specifications, the acceptance criteria, and the detailed results are not present in this 510(k) summary but are referenced as being in another section (Section 018) and a previous submission (K170379).

Therefore, I cannot populate the table or answer the specific questions about clinical study details, expert involvement, or AI performance, as this information is not provided in the given text.

The document primarily focuses on demonstrating substantial equivalence based on:

  • Identical Indications for Use.
  • Similar Technological Characteristics (with the only difference being the jaw length).
  • Bench Testing to ensure the new jaw length performs to the "same specifications for clamping and slip force" as the predicate.
  • Validation of a packaging change.

This type of 510(k) submission, especially a "Special 510(k)" as mentioned for the jaw length change, often relies heavily on engineering bench tests rather than extensive clinical studies or expert-driven evaluations, particularly when the change is considered minor and does not significantly alter the device's fundamental technology or safety profile. The absence of details about clinical studies, expert reviews, or AI integration further supports the conclusion that this is not the type of device or submission that would typically include such data.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for Standard Bariatrics. The logo consists of a stylized red letter "S" enclosed in a rounded square on the left, followed by the words "STANDARD BARIATRICS" in black, sans-serif font on the right. The trademark symbol is located to the right of the word "BARIATRICS".

510(K) Summary

I. SUBMITTER

Standard Bariatrics 4362 Glendale Milford Road Cincinnati, OH 45242 USA

Phone: 513-304-7971 Email: alison(@standardbariatrics.com

Contact Person: Alison Sathe Date Prepared: March 15, 2019

II. DEVICE

Name of Device: Standard Clamp, Disposable Common or Usual Name: Laparoscopic Surgical Instrument Classification Name: Endoscope and Accessories (21 CFR 876.1500) Regulatory Class: II Product Code: GCJ

III. PREDICATE DEVICE

Standard Bariatrics Standard Clamp, Disposable K181608

IV. DEVICE DESCRIPTION

The Standard Clamp, Disposable is a disposable non-energized, hand held surgical instrument designed for use in laparoscopic procedures to clamp long planes of soft tissue such as the stomach. The Standard Clamp, Disposable is provided sterile (gamma). There are no accessories provided with the device. The instrument is comprised of three main sections: handle, shaft, and end effector. The handle is a pistol grip with a trigger and release button which are manually activated in order to open and close the end effector. The shaft is sized to fit a standard 12 mm trocar. The clamp is comp is comprised of an upper and lower jaw that close to grasp and hold tissue. The environment of use is in a hospital during laparoscopic surgical procedures. The current Standard Clamp, Disposable has a jaw length of 25 cm. This Special 510(k) is being filed to add a new shorter jaw length (22 cm) to accommodate patients with smaller abdomens.

The Standard Clamp, Disposable is intended to be used by a surgeon on a single patient during a single laparoscopic surgery under normal operating conditions. The Standard Clamp, Disposable is intended to be inserted into the peritoneal cavity through a 12mm trocar and used in up to six cycles of opening and closure to grasp, clamp and manipulate long, 1.2 to 4.5mm thick, planes of soft tissues such as the stomach. The Standard Clamp, Disposable is not intended for direct contact with the cardiovascular system, lymphatic system or cerebrospinal fluid. The Standard Clamp, Disposable can be used while a surgeon uses devices such as a stapler adjacent to tissue that is clamped with the Standard Clamp, Disposable, e.g., to help guide endoscopic staplers during resection of tissue such as in sleeve gastrectomy.

V. INDICATIONS FOR USE

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image is a logo for Standard Bariatrics. The logo features a red square with a stylized white "S" inside of it. To the right of the square is the text "STANDARD BARIATRICS" in black, bold letters. The letters are all capitalized, and there is a trademark symbol after the word "BARIATRICS".

The Standard Clamp, Disposable is indicated for use in laparoscopic procedures to grasp, clamp, and manipulate soft tissues. The Clamp is indicated for use in bariatric procedures such as sleeve gastrectomy as a guide, to clamp and manipulate flat tissue and organs, such as the stomach.

COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE VI. DEVICE

The Standard Clamp, Disposable has the same technological characteristics as the predicate device with the exception of jaw length.

Table 1: Technological Characteristics
ProductPredicate Standard Clamp, Disposable(product code SCD25)(K181608)Standard Clamp, Disposable(product code SCD22)(Subject)
Indications for UseThe Standard Clamp, Disposable isindicated for use in laparoscopicprocedures to grasp, clamp, andmanipulate soft tissues. The Clamp isindicated for use in bariatric proceduressuch as sleeve gastrectomy as a guide, toclamp and manipulate flat tissue andorgans, such as the stomach.Same
Disposable/ReusableDisposableSame
MaterialsEnd Effector: aluminumShaft: stainless steelHandle: polycarbonateSame
Overall Length26.3 inches (66.8 cm)25.1 inches (63.8 cm)
End Effector Length10.6 inches (27 cm)9.4 inches (24 cm)
End Effector Clamping Length9.8 inches (25 cm)8.9 inches (22 cm)
End Effector Width0.47 inches (12 mm)Same
Shaft Width0.39 inches (10 mm)Same
Weight0.5 lbs.0.5 lbs.
ArticulationNoneSame
End effector designAluminum jaws with non-piercingridges to grasp tissueSame
End effector closureDistal to proximal closureSame
HandlePistol gripSame

{2}------------------------------------------------

Sterilization MethodGamma SterilizationSame
Anatomical Site UsedVarious soft tissues accessible duringlaparoscopic proceduresSame
Types of proceduresLaparoscopic proceduresSame
Energy Delivered?NoSame
Compatible with 12 mm trocar?YesSame
BiocompatibilityBiocompatible for blood/bone/tissuecontact for limited durationSame
PackagingTyvek pouch in a cardboard boxTyvek covered tray in cardboard box

VII. PERFORMANCE DATA

Standard Clamp, Disposable performance bench testing was provided in K170379. Testing for the Standard Clamp, Disposable 22 cm is provided in Section 018 - Performance Testing - Bench of this submission to the same specifications as the Standard Clamp, Disposable 25 cm previously provided in K 170379. The data demonstrates that the Standard Clamp, Disposable 22 cm meets the same specifications for clamping and slip force as the Standard Clamp, Disposable 25 cm. The packaging change has been validated and data is provided in Section 014 - Sterilization and Shelf Life.

VIII. CONCLUSIONS

The Standard Clamp, Disposable has the same intended use and principles of operation as its predicate device. The only difference in technological characteristics is the jaw being offered in a shorter length in order to accommodate varying anatomy. The Standard Clamp, Disposable 22 cm meets the same specifications as the previously cleared Standard Clamp, Disposable 25 cm. There are no new issues of safety or effectiveness. Standard Clamp, Disposable is substantially equivalent to the predicate device.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym and name on the right. The Department of Health & Human Services logo is a stylized depiction of a human figure, while the FDA acronym and name are written in blue, with the acronym in a square and the name written out as "U.S. Food & Drug Administration".

Standard Bariatrics Alison Sathe, Regulatory 4362 Glendale Milford Rd. Cincinnati. Ohio 45242

April 18, 2019

Re: K190788

Trade/Device Name: Disposable Standard Clamp Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: Class II Product Code: GCJ Dated: March 21, 2019 Received: March 27, 2019

Dear Alison Sathe:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see

{4}------------------------------------------------

https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Please wait...

If this message is not eventually replaced by the proper contents of the document, your PDF viewer may not be able to display this type of document.

You can upgrade to the latest version of Adobe Reader for Windows®, Mac, or Linux® by visiting http://www.adobe.com/go/reader_download.

For more assistance with Adobe Reader visit http://www.adobe.com/go/acrreader.

Windows is either a registered trademark of Microsoft Corporation in the United States and/or other countries. Mac is a trademark of Apple Inc., registered in the United States and other countries. Linux is the registered trademark of Linus Torvalds in the U.S. and other countries.

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.