K Number
K190744

Validate with FDA (Live)

Date Cleared
2019-08-29

(160 days)

Product Code
Regulation Number
884.1720
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

This instrument has been designed to be used with a video system center, documentation equipment, monitor, hand instruments, electrosurgical unit, and other ancillary equipment for endoscopic surgery within the thoracic and abdominal cavities including the female reproductive organs.

Device Description

The ENDOEYE HD II - High Definition Digital Video Laparoscope is a rigid video telescope used for endoscopic diagnosis, treatment, video observation and surgery within the thoracic and abdominal cavities including female reproductive organs. For laparoscopic applications, the video telescope is inserted via a trocar into the patient. The ENDOEYE HD II is used with a video system center, light source and monitor to achieve its intended function. In addition, the ENDOEYE HD II can be inserted in compatible instrument trays for reprocessing. The ENDOEYE HD II can provide an image with either white light or narrow band imaging and WA50040A and WA50042A include a heater function at the distal tip to reduce fogging of the lens.

AI/ML Overview

This document describes the premarket notification (510(k)) for the Olympus ENDOEYE HD II, a rigid video telescope used for endoscopic diagnosis, treatment, video observation, and surgery. This is primarily a regulatory submission for a medical device and not a study proving an AI/ML device meets acceptance criteria.

Therefore, many of the requested details, such as acceptance criteria for AI/ML performance metrics, sample sizes for AI/ML test/training sets, expert adjudication methods for AI/ML ground truth, and MRMC studies, are not applicable to this document. This document focuses on demonstrating substantial equivalence to a predicate device through engineering performance data rather than sophisticated clinical or AI/ML performance studies.

However, based on the provided text, I can infer and state the following regarding "acceptance criteria" and "study" in the context of this device:

1. A table of acceptance criteria and the reported device performance:

The document doesn't provide a table of specific numerical acceptance criteria for each performance test, nor does it present detailed quantitative performance results in a table. Instead, it states that "All samples tested met their predefined acceptance criteria." The acceptance criteria would be linked to the standards mentioned for each test.

Acceptance Criteria Category (Inferred from tests)Reported Device Performance (Summary Statement)
Electrical Safety (IEC 60601-1 Ed 3.1, IEC 60601-2-18 Ed 3.0)All samples tested met their predefined acceptance criteria.
Electromagnetic Compatibility (EMC) (IEC 60601-1-2 Ed 4.0)All samples tested met their predefined acceptance criteria.
Thermal Safety (IEC 60601-1 Ed 3.1, IEC 60601-2-18 Ed 3.0)All samples tested met their predefined acceptance criteria.
Mechanical Performance (ISO 8600-3 Ed 1, ISO 8600-4 Ed 2, ISO 8600-5 Ed 1)All samples tested met their predefined acceptance criteria.
Transport and Shipping (ASTM D4169-16)All samples tested met their predefined acceptance criteria.
Expected Service Life (IEC 60601-1 Ed 3.1, IEC 60601-2-18 Ed 3.0)All samples tested met their predefined acceptance criteria.
Illumination System (IEC 60601-1 Ed 3.1, IEC 60601-2-18 Ed 3.0)All samples tested met their predefined acceptance criteria.
Software Verification (IEC 62304 Ed 1.1)All samples tested met their predefined acceptance criteria.
Usability Validation for Instructions for Use (FDA Guidance)All samples tested met their predefined acceptance criteria.
Risk Analysis (ISO 14971:2007)Risk analysis carried out in accordance with established internal acceptance criteria.

2. Sample size used for the test set and the data provenance:

  • Sample size: The document does not specify the exact sample sizes (number of units) used for each individual test. It generally refers to "All samples tested."
  • Data provenance: Not explicitly stated regarding country of origin or specific patient data. The tests are lab-based design verification and validation tests rather than clinical data tests. It's a regulatory submission from a German manufacturer (Olympus Winter & Ibe GmbH) to the US FDA. The tests are "Performance Data" and are likely performed in a controlled, prospective manner as part of the design and manufacturing process.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Not applicable. This device is a hardware endoscopic camera. The "ground truth" for its performance is established by meeting engineering specifications and recognized international standards (e.g., IEC, ISO, ASTM) through objective physical and software testing, not by expert human interpretation of medical images or conditions.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

  • Not applicable. As the "ground truth" is established by adherence to engineering standards and objective measurements, there is no need for expert adjudication of medical findings.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • No. This is not an AI-enabled device. Therefore, no MRMC study looking at human reader improvement with or without AI assistance was conducted or is relevant for this regulatory submission.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • Not applicable. This device is a camera system, not an algorithm that performs a standalone diagnostic or analytical function.

7. The type of ground truth used:

  • The ground truth is based on engineering specifications, recognized international standards (e.g., IEC, ISO, ASTM), and internal company acceptance criteria for hardware and software performance. This includes electrical safety, EMC, thermal safety, mechanical performance, transport resistance, service life, illumination, and software functionality.

8. The sample size for the training set:

  • Not applicable. This is not an AI/ML device that requires a "training set."

9. How the ground truth for the training set was established:

  • Not applicable. See point 8.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a stacked format.

August 29, 2019

Olympus Winter & Ibe GmbH Sheri Musgnung Manager, Regulatory Affairs Olympus Corporation of the Americas 3500 Corporate Parkway Center Valley, PA 18034-061

Re: K190744

Trade/Device Name: ENDOEYE HD II (WA50040A, WA50042A, WA50050A, WA50052A) Regulation Number: 21 CFR 884.1720 Regulation Name: Gynecologic laparoscope and accessories Regulatory Class: II Product Code: HET, GCJ, NWB Dated: July 29, 2019 Received: July 30, 2019

Dear Sheri Musgnung:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Sharon Andrews Assistant Division Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K190744

Device Name ENDOEYE HD II (WA50040A, WA50042A, WA50050A, WA50052A)

Indications for Use (Describe)

This instrument has been designed to be used with a video system center, documentation equipment, monitor, hand instruments, electrosurgical unit, and other ancillary equipment for endoscopic surgery within the thoracic and abdominal cavities including the female reproductive organs.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary of Safety and Effectiveness

August 26, 2019

General Information 1

Manufacturer:Olympus Winter & Ibe GmbH
Kuehnstr. 61
22045 Hamburg
Germany
Establishment Registration Number: 9610773
Official Correspondent:Sheri L Musgnung
Manager, Regulatory Affairs
Olympus Corporation of the Americas
3500 Corporate Parkway
Center Valley PA 18034-061
Phone: 484 896-3147
Email: sheri.musgnung@olympus.com
Establishment Registration Number: 2429304

Device Identification 2

Proprietary /Trade Name:ENDOEYE HD II (WA50040A, WA50042A,WA50050A, WA50052A)
Classification name:Laparoscope, Gynecologic and accessories
Regulations Number:21 CFR 884.1720
Regulatory class:II
Product code:HETGCJ / NWB
Review Panel:Obstetrics and Gynecology

Predicate Device 3

Olympus K11788, ENDOEYE HD II

The predicate has been subject to a design related recall.

{4}------------------------------------------------

4 Product Description

The ENDOEYE HD II - High Definition Digital Video Laparoscope is a rigid video telescope used for endoscopic diagnosis, treatment, video observation and surgery within the thoracic and abdominal cavities including female reproductive organs. For laparoscopic applications, the video telescope is inserted via a trocar into the patient.

The ENDOEYE HD II is used with a video system center, light source and monitor to achieve its intended function.

In addition, the ENDOEYE HD II can be inserted in compatible instrument trays for reprocessing.

The ENDOEYE HD II can provide an image with either white light or narrow band imaging and

WA50040A and WA50042A include a heater function at the distal tip to reduce fogging of the lens.

5 Indications for Use

This instrument has been designed to be used with a video system center, light source, documentation equipment, monitor, hand instruments, electrosurgical unit, and other ancillary equipment for endoscopy and endoscopic surgery within the thoracic and abdominal cavities including the female reproductive organs.

The ENDOEYE HD II has the same intended use as the predicate device.

Technological Characteristics 6

The ENDOEYE HD II has the same technological characteristics as the predicate device.

DeviceSubject devicePredicate device
Device NameEndoEYE HD IIEndoEYE HD II
Model NumberWA50040A, WA50042AWA50040A, WA50042A
WA50050A, WA50052AWA50050A, WA50052A
510(k) numberNot yet knownK111788

{5}------------------------------------------------

DeviceSubject devicePredicate device
Field of viewWA50040A, WA50042A:90° $\pm$ 9° (at 16:9 aspectratio)WA50050A, WA50052A:80.4° $\pm$ 8° (at 5:4 aspectratio)WA50040A, WA50042A:90° $\pm$ 9° (at 16:9 aspectratio)WA50050A, WA50052A:80.4° $\pm$ 8° (at 5:4 aspectratio)
Depth of field21 to 200 mm21 to 200 mm
Direction ofviewing (forward)WA50040A, WA50050A:0° $\pm$ 10°WA50042A, WA50052A:30° $\pm$ 10°WA50040A, WA50050A:0° $\pm$ 10°WA50042A, WA50052A:30° $\pm$ 10°
Type of CCD chipColor CCDColor CCD
Number of CCDchip22
Outer diameter ofdistal endØ 10 mm (WA50040A,WA50042A)Ø 5.4 mm (WA50050A,WA50052A)Ø 10 mm (WA50040A,WA50042A)Ø 5.4 mm (WA50050A,WA50052A)
Outer diameter ofinsertion tubemaximumWA50040A: 10.07 mmWA50042A: 10.2 mmWA50050A: 5.45 mmWA50052A: 5.45 mmWA50040A: 10.07 mmWA50042A: 10.2 mmWA50050A: 5.45 mmWA50052A: 5.45 mm
Working lengthWA50040A: 325 $\pm$ 5 mmWA50042A: 330 $\pm$ 1 mmWA50050A: 300 $\pm$ 0.5 mmWA50052A: 302 $\pm$ 1 mmWA50040A: 325 $\pm$ 5 mmWA50042A: 330 $\pm$ 1 mmWA50050A: 300 $\pm$ 0.5 mmWA50052A: 302 $\pm$ 1 mm
NBI observationavailableavailable
HeaterfunctionalityWA50040A,WA50042A: YesWA50050A,WA50052A: NoWA50040A,WA50042A: YesWA50050A,WA50052A: No
Autoclavabilityyesyes

The subject devices have equivalent technology, performance, dimensions and materials. The differences to the predicate device ENDOEYE HD II are:

  • . Addition of a printed circuit board (PCB) containing a supervisor circuitry and a second temperature sensor
  • Updated Instructions for Use. .

The above described differences in technological characteristics do not raise different questions of safety and effectiveness.

{6}------------------------------------------------

7 Performance Data

The following Design verification and validation testing was performed on the subject device:

  • Electrical safety as per IEC 60601-1 Edition 3.1 and IEC 60601-2-18 Edition 3.0
  • Electromagnetic compatibility (EMC) as per IEC 60601-1-2 Edition 4.0
  • Thermal Safety as per IEC 60601-1 Edition 3.1 and IEC 60601-2-18 Edition ● 3.0
  • Mechanical performance testing as per ISO 8600-3 Edition 1, ISO 8600-4 Edition 2 and ISO 8600-5 Edition 1
  • Transport and shipping testing as per ASTM D4169-16 ●
  • Tests related to the expected service life as per IEC 60601-1 Edition 3.1 and . IEC 60601-2-18 Edition 3.0
  • Testing of the illumination system as per IEC 60601-1 Edition 3.1 and IEC 60601-2-18 Edition 3.0
  • Software Verification as per IEC 62304 Edition 1.1 ●
  • . Usability Validation of the Instructions for Use as per FDA Guidance Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling

Risk analysis was carried out in accordance with established internal acceptance criteria based on ISO 14971:2007.

All samples tested met their predefined acceptance criteria.

8 Conclusion

The performance data support the safety of the device and demonstrate that the subject devices comply with the recognized standards as specified and support a substantial equivalence determination.

§ 884.1720 Gynecologic laparoscope and accessories.

(a)
Identification. A gynecologic laparoscope is a device used to permit direct viewing of the organs within the peritoneum by a telescopic system introduced through the abdominal wall. It is used to perform diagnostic and surgical procedures on the female genital organs. This generic type of device may include: Trocar and cannula, instruments used through an operating channel, scope preheater, light source and cables, and component parts.(b)
Classification. (1) Class II (performance standards).(2) Class I for gynecologic laparoscope accessories that are not part of a specialized instrument or device delivery system, do not have adapters, connector channels, or do not have portals for electrosurgical, lasers, or other power sources. Such gynecologic laparosope accessory instruments include: the lens cleaning brush, biopsy brush, clip applier (without clips), applicator, cannula (without trocar or valves), ligature carrier/needle holder, clamp/hemostat/grasper, curette, instrument guide, ligature passing and knotting instrument, suture needle (without suture), retractor, mechanical (noninflatable), snare, stylet, forceps, dissector, mechanical (noninflatable), scissors, and suction/irrigation probe. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 884.9.