K Number
K190665

Validate with FDA (Live)

Date Cleared
2019-12-23

(283 days)

Product Code
Regulation Number
880.6850
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Neodent Instrument Kit Cases are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. Neodent Instrument Kit Cases are intended to allow sterilization of the enclosed medical devices. Neodent Instrument Kit Cases require the use of FDA cleared wrap to maintain the sterility of the enclosed devices.

The kits are to be enclosed in a sterilization wrap that is FDA-cleared for the indicated cycles, and moist heat (steam) sterilized using one of the following cycles:

Fractionated vacuum (pre-vacuum) - Exposure at 132 °C for 4 minutes, 30-minute dry time (for GM Zygomatic Surgical Kit Case) or 20-minute dry time (for GM Helix LG Compact Surgical Kit Case)

Gravity displacement - Exposure at 132 °C for 15 minutes, 20-minute dry time (only for GM Helix LG Compact Surgical Kit Case)

Neodent Instrument Kit Cases are intended for sterilization of non-porous loads.

The combined weight of the GM Zygomatic Surgical Kit Case and the instruments is 738.1 g. The weight of the empty kit case is 515 g.

The combined weight of the GM Helix LG Compact Kit Case and the instruments is 297.3 g. The weight of the empty kit case is 236 g.

Neodent Instrument Kit Cases should not to be stacked during sterilization.

Device Description

The subject device kit cases are reusable rigid containers, comprising a case bottom (or base), a removable inner tray, and tray lid (lid). The subject device kits are to be used to organize and protect instruments and accessories that are to be sterilized by the healthcare provider. The subject devices include two (2) such kit cases. The lids are manufactured from injection molded polyphenylsulfone resin, the tray and base are manufactured from injection molded polysulfone resin. The designs include grommets and inserts of various geometries manufactured from molded silicone that retain the instruments within the tray and base. The subject device kit cases are provided nonsterile to the end-user.

AI/ML Overview

The medical device in question is the Neodent Instrument Kit Cases, intended for use in sterilizing other medical devices by healthcare providers using moist heat (steam). The provided documentation details the acceptance criteria and the studies conducted to demonstrate its performance.

Here's a breakdown of the requested information:

1. A table of acceptance criteria and the reported device performance

Acceptance CriteriaReported Device Performance
Manual Cleaning Validation
Hemoglobin Test: - < 0.50 µg/mL - < 55 µg/article - < 0.0078 µg/cm²Passed
Micro BCA Protein Test: - < 1.1 µg/mL - < 220 µg/article - < 0.018 µg/cm²Passed
Sterilization Validation (Sterilant penetration and drying time)
Sterility Assurance Level (SAL) of 10⁻⁶Passed
Life Cycle (Simulated Usage) Testing (100 cycles)
Visual and functional inspection following 100 cyclesPassed
Biocompatibility Testing (Cytotoxicity)
< 30% inhibition of proliferation of protein contentPassed

2. Sample size used for the test set and the data provenance

The document does not specify the exact sample sizes used for each test. For instance, it doesn't state how many kit cases were subjected to manual cleaning validation or how many sterility tests were performed.

Regarding data provenance:

  • The tests were nonclinical performance data.
  • The document does not explicitly state the country of origin of the data, but the sponsor is "JJGC Indústria e Comércio de Materiais Dentários SA" located in Curitiba, Parana, Brazil. Therefore, the data likely originates from Brazil or a laboratory contracted by the sponsor.
  • The tests as described (validation studies) are inherently retrospective in nature, as they involve testing the manufactured device against pre-defined criteria.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This information is not provided in the document. The studies are nonclinical performance tests and do not involve human expert interpretation of device performance in a medical context where ground truth needs to be established by experts (e.g., radiologists interpreting images). The "ground truth" for these tests is based on objective laboratory measurements and industry standards.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This is not applicable/not provided. Adjudication methods like 2+1 or 3+1 are typically used in clinical studies involving multiple human readers to resolve discrepancies in their interpretations. The tests described are laboratory-based performance validations against defined physical and biological criteria, not human-in-the-loop assessments.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No, an MRMC comparative effectiveness study was not done. This type of study is relevant for AI-powered diagnostic or assistive devices, which is not what the Neodent Instrument Kit Cases are. This device is a sterilization accessory.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

No, a standalone algorithm-only performance study was not done. This device is a physical instrument kit case, not a software algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The "ground truth" for the nonclinical performance data is based on:

  • Objective laboratory measurements and chemical/biological assays for manual cleaning validation (hemoglobin and protein levels).
  • Sterility Assurance Level (SAL) of 10⁻⁶ which is a a quantitative measure of the probability of a single viable microorganism occurring on an item after sterilization. This is established through microbiological testing.
  • Visual and functional inspection for life cycle testing, against manufacturing specifications.
  • Measurement of cell proliferation inhibition percentage for cytotoxicity testing, an objective biological response measurement.

These are all objective, measurable criteria derived from recognized standards rather than subjective expert consensus, pathology reports, or patient outcomes data.

8. The sample size for the training set

This information is not applicable/not provided. The "Neodent Instrument Kit Cases" is a physical medical device, not an AI, machine learning, or software device that requires a training set.

9. How the ground truth for the training set was established

This information is not applicable/not provided as the device is not an AI/ML system requiring a training set.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of the seal is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font size below it. The logo is simple and professional, and it is easily recognizable.

December 23, 2019

JJGC Industria e Comercio de Materiais Dentarios S.A. % Jennifer Jackson Director of Regulatory Affairs Straumann USA, LLC 60 Minuteman Road Andover, Massachusetts 01810

Re: K190665

Trade/Device Name: Neodent Instrument Kit Cases Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization wrap Regulatory Class: Class II Product Code: KCT Dated: October 11, 2019 Received: October 15, 2019

Dear Jennifer Jackson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Sreekanth Gutala, Ph.D. Acting Assistant Director for the Sterility Devices Team DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K190665

Device Name Neodent Instrument Kit Cases

Indications for Use (Describe)

Indications for Use for GM Zygomatic Surgical Kit Case:

Neodent Instrument Kit Cases are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. Neodent Instrument Kit Cases are intended to allow sterilization of the enclosed medical devices. Neodent Instrument Kit Cases require the use of FDA cleared wrap to maintain the sterility of the enclosed devices.

The kits are to be enclosed in a sterilization wrap that is FDA-cleared for the indicated cycles, and moist heat (steam) sterilized using one of the following cycles:

Fractionated vacuum (pre-vacuum) - Exposure at 132 °C for 4 minutes, 20 minute dry time

Neodent Instrument Kit Cases are intended for sterilization of non-porous loads.

The GM Zygomatic Surgical Kit Case maximum load weight is 220 grams.

Neodent Instrument Kit Cases should not to be stacked during sterilization.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Indications for Use

510(k) Number (if known)

K190665 Device Name

Neodent Instrument Kit Cases

Indications for Use (Describe)

Indications for Use for GM Helix LG Compact Surgical Kit Case:

Neodent Instrument Kit Cases are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. Neodent Instrument Kit Cases are intended to allow sterilization of the enclosed medical devices. Neodent Instrument Kit Cases require the use of FDA cleared wrap to maintain the sterility of the enclosed devices.

The kits are to be enclosed in a sterilization wrap that is FDA-cleared for the indicated cycles and moist heat (steam) sterilized using one of the following cycles:

Fractionated vacuum (pre-vacuum) - Exposure at 132 °C for 4 minutes, 20-minute dry time

Gravity displacement - Exposure at 132 °C for 15 minutes, 20-minute dry time

Neodent Instrument Kit Cases are intended for sterilization of non-porous loads.

The combined weight of the GM Helix LG Compact Kit Case and the instruments is 297.3 g. The weight of the empty kit case is 236 g.

Neodent Instrument Kit Cases should not to be stacked during sterilization.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary

K190665

ADMINISTRATIVE INFORMATION

SponsorJJGC Indústria e Comércio de Materiais Dentários SAAv. Juscelino Kubitschek de Olivera, 3291Curitiba, Parana, Brazil 81270-200Registration No.: 3008261720Owner/Operator No.: 10031702
Contact PersonJennifer M. Jackson, MSDirector of Regulatory AffairsStraumann USAjennifer.jackson@straumann.comTelephone (978) 747-2509
Date Prepared17/Dec/2019
Preparer / Alternate ContactLuiza Vaccari ToppelRegulatory Affairs Coordinatorluiza.toppel@neodent.com

DEVICE NAME AND CLASSIFICATION

Trade/ Proprietary NameNeodent Instrument Kits
Common NameInstrument Sterilization Trays
Classification NameSterilization Wrap Containers, Trays, Cassettes & Other Accessories
Classification Regulations21 CFR 880.6850, Class II
Product CodeKCT
Classification PanelGeneral Hospital
Reviewing BranchInfection Control Devices Branch

PREDICATE DEVICE INFORMATION

Primary Predicate DeviceK171713 – Neodent Instrument Kits, JJGC Indústria e Comércio deMateriais Dentários AS
Reference DevicesK142529, Genesis™ Reusable Rigid Sterilization Container System,CareFusion 2200 Inc.K180915, Sonicision Reusable Sterilization Tray, Covidien

{5}------------------------------------------------

INDICATIONS FOR USE

Indications for Use for GM Zygomatic Surgical Kit Case:

Neodent Instrument Kit Cases are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. Neodent Instrument Kit Cases are intended to allow sterilization of the enclosed medical devices. Neodent Instrument Kit Cases require the use of FDA cleared wrap to maintain the sterility of the enclosed devices.

The kits are to be enclosed in a sterilization wrap that is FDA-cleared for the indicated cycles, and moist heat (steam) sterilized using one of the following cycles:

Fractionated vacuum (pre-vacuum) – Exposure at 132 °C for 4 minutes, 30-minute dry time

Neodent Instrument Kit Cases are intended for sterilization of non-porous loads.

The combined weight of the GM Zygomatic Surgical Kit Case and the instruments is 738.1 g. The weight of the empty kit case is 515 g.

Neodent Instrument Kit Cases should not to be stacked during sterilization.

Indications for Use for GM Helix LG Compact Surgical Kit Case:

Neodent Instrument Kit Cases are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. Neodent Instrument Kit Cases are intended to allow sterilization of the enclosed medical devices. Neodent Instrument Kit Cases require the use of FDA cleared wrap to maintain the sterility of the enclosed devices.

The kits are to be enclosed in a sterilization wrap that is FDA-cleared for the indicated cycles, and moist heat (steam) sterilized using one of the following cycles:

Fractionated vacuum (pre-vacuum) – Exposure at 132 °C for 4 minutes, 20-minute dry time

Gravity displacement - Exposure at 132 °C for 15 minutes, 20-minute dry time

Neodent Instrument Kit Cases are intended for sterilization of non-porous loads.

The combined weight of the GM Helix LG Compact Kit Case and the instruments is 297.3 g. The weight of the empty kit case is 236 g.

Neodent Instrument Kits should not to be stacked during sterilization.

SUBJECT DEVICE DESCRIPTION

The subject device kit cases are reusable rigid containers, comprising a case bottom (or base), a removable inner tray, and tray lid (lid). The subject device kits are to be used to organize and protect instruments and accessories that are to be sterilized by the healthcare provider. The subject devices include two (2) such kit cases. The lids are manufactured from injection molded polyphenylsulfone resin, the tray and base are manufactured from injection molded polysulfone resin. The designs include grommets and inserts of various geometries manufactured from molded silicone that retain the instruments within the tray and base. The subject device kit cases are provided nonsterile to the end-user. The dimensions for each part of the model and the overall dimensions are presented in the table below:

{6}------------------------------------------------

AssembledKit CaseDescriptionAssembled KitCase Dimension(L x W x H)Component NumberComponent Dimension(L x W x H)
110.299GM Zygomatic SurgicalKit Case264 x 163 x 58mm212.327 (Lid)264 x 163 x 49, mm
212.345 (Tray)248 x 149 x 16, mm
212.347 (Base)256 x 155 x 20.5, mm
110.300GM Helix LG CompactSurgical Kit Case195 x 90 x 64mm212.293 (Lid)195 x 90 x 36, mm
212.346 (Tray)180 x 76 x 26.5, mm
212.348 (Base)188 x 84 x 42, mm

Note: The instrument and accessory devices that are sterilized and stored within the subject Kit Cases are not the subject devices of this submission.

{7}------------------------------------------------

TECHNOLOGICAL CHARACTERISTIC COMPARISON TABLE

SUBJECT DEVICEPRIMARY PREDICATE DEVICECOMPARISON
K190665Neodent Instrument Kit CasesJJGC Indústria e Comércio de MateriaisDentários S.A.K171713Neodent Instrument KitsJJGC Indústria e Comércio de MateriaisDentários S.A.
Indications forUse StatementIndications for Use for GM ZygomaticSurgical Kit Case:Neodent Instrument Kits are intended tobe used to enclose other medical devicesSimilar
Neodent Instrument Kit Cases are intendedto be used to enclose other medical devicesthat are to be sterilized by a health care
that are to be sterilized by a health careprovider. Neodent Instrument Kit Cases areprovider. Neodent Instrument Kits are
intended to allow sterilization of theintended to allow sterilization of the
enclosed medical devices. Neodentenclosed medical devices. Neodent
Instrument Kit Cases require the use of FDAInstrument Kits require the use of FDA
cleared wrap to maintain the sterility of thecleared wrap to maintain the sterility of
enclosed devices.the enclosed devices.
The kits are to be enclosed in a sterilizationwrap that is FDA-cleared for the indicatedThe kits are to be enclosed in asterilization wrap that is FDA-cleared for
cycles, and moist heat (steam) sterilizedthe indicated cycles, and moist heat(steam) sterilized using one of the
using one of the following cycles:following cycles:
Fractionated vacuum (pre-vacuum) -Exposure at 132 °C for 4 minutes, 30-minuteFractionated vacuum (pre-vacuum) -Exposure at 132 °C for 4 minutes, 20-
dry timeminute dry time
Neodent Instrument Kit Cases are intendedfor sterilization of non-porous loads.Gravity displacement - Exposure at 132 °Cfor 15 minutes, 20-minute dry timeNeodent Instrument Kits are intended for
The combined weight of the GM ZygomaticSurgical Kit Case and the instruments issterilization of non-porous loads.
$738.1 g$ . The weight of the empty kit case isThe GM/WS Surgical Kit Case maximum
$515 g$ .load weight is 125 grams.
Neodent Instrument Kit Cases areThe GM Surgical Kit Case maximum load
recommended not to be stacked duringweight is 113 grams.
sterilization.Indications for Use for GM Helix LGCompact Surgical Kit Case:Neodent Instrument Kits arerecommended not to be stacked duringsterilization.
Neodent Instrument Kit Cases are intendedto be used to enclose other medical devices
that are to be sterilized by a health care
provider. Neodent Instrument Kit Cases are
intended to allow sterilization of theenclosed medical devices. Neodent
Instrument Kit Cases require the use of FDA
cleared wrap to maintain the sterility of the
enclosed devices.
The kits are to be enclosed in a sterilizationwrap that is FDA-cleared for the indicatedcycles, and moist heat (steam) sterilized
using one of the following cycles:
Fractionated vacuum (pre-vacuum) -Exposure at 132 °C for 4 minutes, 20-minute
dry time
Gravity displacement – Exposure at 132 °Cfor 15 minutes, 20-minute dry time
Neodent Instrument Kit Cases are intendedfor sterilization of non-porous loads.
The combined weight of the GM Helix LGCompact Kit Case and the instruments is
$297.3 g$ . The weight of the empty kit case is
$236 g$ .
Neodent Instrument Kit Cases are
recommended not to be stacked duringsterilization.
SUBJECT DEVICEPRIMARY PREDICATE DEVICE
K190665Neodent Instrument Kit CasesJJGC Indústria e Comércio de MateriaisDentários S.A.K171713Neodent Instrument KitsJJGC Indústria e Comércio de MateriaisDentários S.A.COMPARISON
Intended UseNeodent Instrument Kit Cases are intendedto be used to enclose other medical devicesthat are to be sterilized by a health careprovider. Neodent Instrument Kit Cases areintended to allow sterilization of theenclosed medical devices. NeodentInstrument Kit Cases require the use of FDAcleared wrap to maintain the sterility of theenclosed devices.Neodent Instrument Kits are intended tobe used to enclose other medical devicesthat are to be sterilized by a health careprovider. Neodent Instrument Kits areintended to allow sterilization of theenclosed medical devices. NeodentInstrument Kits require the use of FDAcleared wrap to maintain the sterility of theenclosed devices.Same
DesignRigid polysulfone polymer base andremovable inner tray with apolyphenylsulfone lid.Retention grommets of medical gradesilicone.Rigid polysulfone polymer base andremovable inner tray with apolyphenylsulfone lid.Retention grommets of medical gradesilicone.Same
PerforatedYes; allows moist heat (steam) penetrationto achieve sterilizationYes; allows moist heat (steam) penetrationto achieve sterilizationSame
ReusableYesYesSame
OveralldimensionsGM Zygomatic Surgical Kit Case:264 L x 163 W x 56 H, mmGM Helix LG Compact Surgical Kit Case:195 L x 90 W x 64 H, mm264 L x 163 W x 54 H, mmSimilar
Vent to VolumeRatioGM Zygomatic Surgical Kit Case:0.0093 cm² / cm³ (0.0236 in²/in³)GM Helix LG Compact Surgical Kit Case:0.0157 cm² / cm³ (0.0400 in² / in³)0.0102 cm² / cm³ (0.0259 in² / in³)Similar
Useful LifeYes, reusable up to 100 cyclesAssembled/disassembled, cleaned, sterilizedVisual inspectionComponent dimensional fit verificationFunctional closure (lid-base latch)verificationYes, reusable up to 100 cyclesAssembled/disassembled, cleaned,sterilizedVisual inspectionComponent dimensional fit verificationFunctional closure (lid-base latch)verificationSame
BiocompatibilityThe kit cases are not intended to be incontact with the patient.The kit cases are not intended to be incontact with the patient.Same
SterilizationMethodMoist heat (steam)Moist heat (steam)Same
CyclesGravity displacement (only for GM Helix LGCompact Surgical Kit Case)Fractionated vacuum (pre-vacuum)Gravity displacementFractionated vacuum (pre-vacuum)Same
ParametersGravity (only for model number 110.300)Sterilization temperature: 132 °CSterilization time: 15 minutesDrying time: 20 minutesPre-VacuumSterilization temperature: 132 °CSterilization time: 4 minutes;Drying time: 30 minutes (model number110.299) or 20 minutes (model number110.300)GravitySterilization temperature: 132 °CSterilization time: 15 minutes;Drying time: 20 minutesPre-VacuumSterilization temperature: 132 °CSterilization time: 4 minutes;Drying time: 20 minutes.SimilarAll validated cycles providesimilar sterilizationefficacy.(minimum SAL of 10⁻⁶)

{8}------------------------------------------------

{9}------------------------------------------------

SUBJECT DEVICEPRIMARY PREDICATE DEVICECOMPARISON
K190665Neodent Instrument Kit CasesJJGC Indústria e Comércio de MateriaisDentários S.A.K171713Neodent Instrument KitsJJGC Indústria e Comércio de MateriaisDentários S.A.
Sterile BarrierSterilization wrap, FDA-cleared for indicatedmethod and cyclesSterilization pouch, FDA-cleared forindicated method and cyclesSame

The subject devices and the primary predicate device per K171713 have the same intended use and have equivalent Indications for Use Statements. The subject devices (and the primary predicate device per K171713) are reusable rigid containers used to organize and protect dental surgical instruments that are sterilized by the healthcare provider. The subject device and the primary predicate device K171713 components are perforated to allow for penetration of the moist heat (steam) sterilant and require the use of an FDA-cleared wrap or pouch to maintain sterility.

NON-CLINICAL PERFORMANCE DATA

Manual cleaning instructions were validated using microbiological, protein, and hemoglobin assays. Sterilization validation, including sterilant penetration and drying time, was performed according to ANSI/AAMI/ISO 17665-1 and 17665-2. Life cycle (simulated usage) testing was performed, which included visual inspection, component dimensional fit verification, and functional closure (lid- bottom latch) verification. Biocompatibility testing was performed using methods described in ANSI/AAMI/ISO 10993-5 (cytotoxicity). No clinical data were included in this submission.

Standard or Test MethodPurpose of the TestingAcceptance CriteriaResults
CustomManual cleaning validationHemoglobin Test:• <0.50 $µg/mL$• <55 $µg/article$• <0.0078 $µg/cm²$Micro BCA Protein Test:• <1.1 $µg/mL$• <220 $µg/article$• <0.018 $µg/cm²$Passed
ANSI/AAMI/ISO 17665-1ANSI/AAMI/ISO 17665-2Sterilization validation, including sterilant penetration and drying timeSAL of $10^{-6}$Passed
Reprocessing MedicalDevices in Health CareSettings: ValidationMethods and LabelingGuidance for Industryand Food and DrugAdministration StaffLife cycle (simulate usage) testingVisual and functional inspection following 100 cyclesPassed
ANSI/AAMI/ISO 10993-5(Cytotoxicity)Cytotoxicity testing<30% inhibition of proliferation of protein contentPassed

{10}------------------------------------------------

CONCLUSION

The conclusions drawn from the nonclinical and clinical tests that demonstrate that the device is as safe, as effective, and performs as well as or better than the legally marketed device K17173.

§ 880.6850 Sterilization wrap.

(a)
Identification. A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.(b)
Classification. Class II (performance standards).