(114 days)
Intended Use
The MicroLink™ All-Suture Button Fixation System, when used for fixation of bone-to-bone or soft-tissue-to-bone, is intended as a fixation post, distribution bridge, or for distributing suture tension over areas of ligament or tendon repair.
Indications for Use
The CONMED MicroLink All-Suture Button Fixation System is indicated for fixation of bone-to-bone as an adjunct in the following orthopedic surgical procedures:
Procedure: Carpometacarpal Suspension
Anatomic Location: Hand
The MicroLinkTM All-Suture Button Fixation System (Radiopaque) is an all-suture suspension device with one strand of #2 (5 metric) Hi-Fi® suture, a radiopaque flat braid Hi-Fi® Suture Button and a radiopaque flat braid Hi-Fi® suture Backstop threaded on a Loader. A suture passing drill is provided with a nitinol loop for passing. A trapezium pin is provided to facilitate manipulation of the trapezium during trapeziectomy. All four catalog numbers are provided sterile, intended for single-use.
The provided document is a 510(k) summary for a medical device called the "MicroLink™ All-Suture Button Fixation System (Radiopaque)". This document is a regulatory submission to the FDA for medical device clearance, indicating that the device is a physical product, not a software algorithm. Therefore, the questions related to AI/software aspects such as acceptance criteria for AI performance, sample sizes for test/training sets, expert ground truth, adjudication methods, MRMC studies, or standalone algorithm performance are not applicable.
The document discusses performance testing for the physical device to demonstrate its equivalence to a predicate device.
Here's the relevant information that can be extracted from the provided text regarding device performance:
1. A table of acceptance criteria and the reported device performance
The document lists types of performance testing completed, but does not provide specific acceptance criteria or quantitative performance results in a table format. It generally states that the device "performs as intended and is substantially equivalent to the predicate device" and "met the endotoxin limits."
| Performance Test | Reported Device Performance |
|---|---|
| Bacterial Endotoxin Testing | Met the endotoxin limits. |
| Sterilization | Completed (implies met standards). |
| Reliability | Completed (implies met standards). |
| Pyrogen | Completed (implies met standards). |
| Ultimate Fixation Strength | Completed (implies met standards and comparable to predicate). |
| Cyclic | Completed (implies met standards and comparable to predicate). |
| Biocompatibility | Completed (implies met standards). |
| Shelf-life | Completed (implies met standards). |
| User Validation | Completed (implies met standards). |
| Packaging | Completed (implies met standards). |
| Transportation | Completed (implies met standards). |
| Side-by-Side Testing | Completed (implies comparable to predicate). |
2. Sample sizes used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
This information is not provided in the document. The document describes a physical medical device, and the testing mentioned (e.g., sterilization, biocompatibility, fixation strength) would typically involve laboratory or bench testing on device samples, not patient data sets.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This information is not applicable and not provided. The device is a physical implant, not an diagnostic imaging or AI tool requiring expert interpretation for ground truth.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This information is not applicable and not provided. Adjudication methods are relevant for subjective assessments, typically in clinical trials or AI performance evaluations, not for the physical performance testing described.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This information is not applicable and not provided. This device is not an AI-assisted diagnostic tool.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
This information is not applicable and not provided. This device is a physical surgical implant, not a software algorithm.
7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)
This information is not applicable and not provided in the context of ground truth for an AI system. For the physical device, performance criteria are likely based on engineering standards, biomechanical principles, and regulatory requirements for medical devices of its type.
8. The sample size for the training set
This information is not applicable and not provided. This device is a physical surgical implant, not a machine learning model.
9. How the ground truth for the training set was established
This information is not applicable and not provided. This device is a physical surgical implant, not a machine learning model.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the acronym "FDA" followed by the words "U.S. Food & Drug Administration".
CONMED Corporation Diana Nader-Martone Senior Regulatory Affairs Specialist 11311 Concept Boulevard Largo, Florida 33773
Re: K190582
Trade/Device Name: MicroLink™ All-Suture Button Fixation System (Radiopaque) Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: HTN Dated: May 28, 2019 Received: May 29, 2019
Dear Ms. Nader-Martone:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance)and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Laurence D. Coyne, Ph.D. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known)
Device Name
MicroLink™ All-Suture Button Fixation System (Radiopaque)
Indications for Use (Describe)
Intended Use
The MicroLink™ All-Suture Button Fixation System, when used for fixation of bone-to-bone or soft-tissue-to-bone, is intended as a fixation post, distribution bridge, or for distributing suture tension over areas of ligament or tendon repair.
Indications for Use
The CONMED MicroLink All-Suture Button Fixation System is indicated for fixation of bone-to-bone as an adjunct in the following orthopedic surgical procedures:
Procedure Carpometacarpal Suspension Anatomic Location Hand
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Conmed Corporation. The logo features a blue square with a white curved shape inside on the left. To the right of the square is the word "CONMED" in black, with a blue line underneath. Below the line is the word "CORPORATION" in smaller, black letters.
510(k) SUMMARY
In accordance with the requirements of the Safe Medical Device Act of 1990 and 21 CFR 807.92, CONMED Corporation is hereby submitting the 510(k) Summary of Safety and Effectiveness for 510(k) Number K190582.
I. SUBMITTER
CONMED Corporation 11311 Concept Blvd Largo, Florida 33773
Phone: 727-399-5425 Fax: 727-399-5264
Contact Persons: Diana L. Nader-Martone or Kathy Reddig Date Prepared: April 25, 2019
II. DEVICE NAME
| Device Name: | MicroLink™ All-Suture Button Fixation System(Radiopaque) |
|---|---|
| Common Name: | Nonabsorbable Suture Fixation System |
| Classification Name: | Single/multiple component metallic bone fixation appliancesand accessories |
| Regulatory Class: | Class II, per 21 CFR Part 888. 3030 |
| Product Codes: | HTN |
III. PREDICATE/ LEGALLY MARKET DEVICE
| Device Name: | CMC Mini TightRope |
|---|---|
| Company Name: | Arthrex, Inc. |
| 510(k) #: | K140328 |
IV. DEVICE DESCRIPTION
The MicroLinkTM All-Suture Button Fixation System (Radiopaque) is an all-suture suspension device with one strand of #2 (5 metric) Hi-Fi® suture, a radiopaque flat braid Hi-Fi® Suture Button and a radiopaque flat braid Hi-Fi® suture Backstop threaded on a Loader. A suture passing drill is provided with a nitinol loop for passing. A trapezium pin is provided to facilitate manipulation of the trapezium during trapeziectomy. All four catalog numbers are provided sterile, intended for single-use.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Conmed Corporation. The logo features a blue abstract shape on the left, followed by the word "CONMED" in black, bold letters. Below "CONMED" is the word "CORPORATION" in smaller, blue letters.
INTENDED USE/ INDICATIONS FOR USE V.
The MicroLinkTM All-Suture Button Fixation System, when used for fixation of bone-tobone or soft-tissue-to-bone, is intended as a fixation post, distribution bridge, or for distributing suture tension over areas of ligament or tendon repair.
The CONMED MicroLink All-Suture Button Fixation System is indicated for fixation of bone-to-bone as an adjunct in the following orthopedic surgical procedures:
| Procedure | Anatomic Location |
|---|---|
| Carpometacarpal Suspension | Hand |
COMPARISON OF THE TECHNOLOGICAL CHARACTERISTICS WITH THE VII. PREDICATE DEVICE
The following table represents a summary of the technological characteristics between the proposed and the predicate device.
| MicroLink™ All-Suture Button Fixation System(Radiopaque)Proposed Device | CMC Mini TightRopePredicate Device | |
|---|---|---|
| DeviceDescription | The MicroLink™ All-Suture Button FixationSystem (Radiopaque is an all-suture suspensiondevice with one strand of #2 (5 metric) Hi-Fi®suture, a radio flat braid Hi-Fi® Suture Buttonand a radiolucent flat braid Hi-Fi® sutureBackstop threaded on a Loader. A suture passingdrill is provided with a nitinol loop for passing.A trapezium pin is provided to facilitatemanipulation of the trapezium duringtrapeziectomy. All four catalog numbers areprovided sterile, intended for single-use. | |
| Intended Use | The MicroLink™ All-Suture Button FixationSystem, when used for fixation of bone-to-boneor soft-tissue-to-bone, is intended as a fixationpost, distribution bridge, or for distributing suturetension over areas of ligament or tendon repair. | |
| Indication for Use | The CONMED MicroLink All-Suture ButtonFixation System is indicated for fixation ofbone-to-bone as an adjunct in the followingorthopedic surgical procedures:Procedure Anatomic LocationCarpometacarpal Suspension Hand | |
| Contraindications | 1. Pathological conditions of bone whichwould adversely affect the device.2. Physical conditions that would eliminate, ortend to eliminate, adequate implant supportor retard healing.3. Conditions which tend to limit the patient'sability or willingness to restrict activities orfollow directions during the healing period. |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for CONMED Corporation. The logo features a stylized blue graphic on the left, resembling a curved shape. To the right of the graphic, the word "CONMED" is displayed in a bold, sans-serif font, with a horizontal line underneath. Below the line, the word "CORPORATION" is written in a smaller, sans-serif font.
| MicroLink™ All-Suture Button Fixation System(Radiopaque)Proposed Device | CMC Mini TightRopePredicate Device | |
|---|---|---|
| 4. Attachment of artificial ligaments or other implants.5. Foreign body sensitivity, known or suspected allergies to implant and/or instrument materials.6. This device is not approved for screw attachment or fixation to the posterior elements (pedicles) of the cervical, thoracic or lumbar spine.7. Patients with active sepsis or infection.8. | 6. Conditions which tend to limit the patient's ability or willingness to restrict activities or follow directions during the healing period.7. The use of this device may not be suitable for patients with insufficient or immature bone. The physician should carefully assess bone quality before performing orthopedic surgery on patients who are skeletally immature. The use of this medical device and the placement of hardware or implants must not bridge, disturb, or disrupt the growth plate.8. Do not use for surgeries other than those indicated. | |
| Components | All-Suture Button (Radiopaque) with sutureAll-Suture BackstopLoaderTrapezium PinSuture Passing Drill | TightRope Suture Construct with sutureOblong Button for Mini TightRopeLoaderTrapeziectomy Tool w/ HandleSuture Passing K-wire, shortSuture Passing K-wire, longSuture Passing Wire, 8 inches |
| TechnologicalCharacteristics | Flat-braid suture buttonFlat-braid suture backstopOne #2 sutureRadiopaque implantSuture Passing Drill to pull suture through one or more bonesInstrument to maneuver and remove bone | Stainless steel buttonStainless steel backstopOne #2 sutureRadiopaque implantSuture passing drill to pull suture through one or more bonesInstrument to maneuver and remove bone |
VIII. PERFORMANCE DATA
Testing has been completed to demonstrate that the MicroLinkTM All-Suture Button Fixation System (Radiopaque) performs as intended and is substantially equivalent to the predicate device. Bacterial endotoxin testing was conducted and met the endotoxin limits.
Completed testing includes the following:
Verification Testing Sterilization
- . Reliability
- Pyrogen
●
IX.
-
Ultimate Fixation Strength
Side-by-Side Testing -
●
-
Biocompatibility
- Shelf-life
-
. Cyclic
-
User Validation ● Packaging ●
- Transportation ●
Validation Testing
- CONCLUSION
The MicroLink™ All-Suture Button Fixation System (Radiopaque) is either substantially equivalent or identical in design, materials, intended use, principles of operation, and technical characteristics to the predicate CMC Mini TightRope. Based upon the findings of our performance testing, the differences present no new issues of safety and efficacy, and the MicroLinkTM All-Suture Button Fixation System (Radiopaque) is substantially equivalent to the CMC Mini TightRope (K140328).
N/A