K Number
K190536

Validate with FDA (Live)

Device Name
Mediracer NCS
Manufacturer
Date Cleared
2019-09-25

(205 days)

Product Code
Regulation Number
882.1550
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Mediracer NCS is intended to measure sensory and motor neripheral nerves. The measured data can be utilized in evaluating patients suspected of having focal neuropathies. The measured data must be used in the context of other patient information and must be reviewed and interpreted by a physician.

Device Description

The Mediracer NCS is a hand-held NCS Device to stimulate and record the nerve responses from patient. The NCS Device is connected to the disposable surface electrodes with cables provided with the NCS Device. The NCS Device communicates via Bluetooth with a computer which is using The Mediracer Analysis Center (MAC) software. The computer uses the MAC software to store and handle the patient measurement data. With the NCS Device comes a charger and a docking station for charging the NCS Device. Also, solution includes a test module for checking the functionality of the whole test system.

AI/ML Overview

The Mediracer NCS is a nerve conduction velocity measurement device. The provided text outlines the performance data for the device, including bench testing and clinical studies, to support its substantial equivalence determination.

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are not explicitly stated as quantitative thresholds in the provided text. Instead, the document focuses on demonstrating substantial equivalence to a predicate device and high agreement with traditional nerve conduction studies. The "performance data" section primarily describes the results of studies rather than pre-defined acceptance criteria with specific numerical targets typical for a 510(k) submission.

However, based on the summary provided, we can infer the reported device performance:

Performance MetricReported Device Performance (Mediracer NCS)
Bench Testing - AccuracyResults "show significantly similar nerve response action potential values" compared to a conventional EMG device, and "matching inside the given deviation."
Clinical Study - Positive AgreementAround 90% (with traditional NCS devices for detecting median nerve lesion in CTS)
Clinical Study - Negative Agreement99-100% (with traditional NCS devices for detecting median nerve lesion in CTS)
Clinical Study - Method ConcordanceHigh concordance between used latency difference parameters between traditional and new device study.
Clinical Study - SafetyNon-invasive, safe, no adverse effects or complications reported.
BiocompatibilityHydrogel passed all ISO 10993 testing (Cytotoxicity, Irritation, Sensitization).
Electrical Safety & EMCComplies with IEC 60601-1, IEC 60601-2-40, and EN 60601-1-2.
Software Verification & ValidationSoftware considered "Moderate" level of concern; V&V conducted as per FDA guidance.

2. Sample Size Used for the Test Set and Data Provenance

The provided text describes two clinical studies used for performance evaluation, which can be considered the test sets.

  • Finnish Multicenter Study:

    • Sample Size: 194 patients and 95 healthy control volunteers.
    • Data Provenance: Finland (multicenter study).
    • Retrospective or Prospective: The description strongly suggests a prospective study ("recruited patients," "went through a clinical examination, Mediracer NCS measurement and the traditional NCS").
  • United Kingdom Study:

    • Sample Size: 63 patients (selected from 104 referrals).
    • Data Provenance: United Kingdom (Leicester Carpal Tunnel Service).
    • Retrospective or Prospective: Description indicates patients "were recruited from the normal referral stream," "fulfilled the clinical CTS diagnosis after symptom inquiry and clinical assessment," and "went through clinical assessment, Mediracer NCS measurement and the traditional NCS," suggesting a prospective study.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

The text states that in both clinical studies, participants underwent "traditional nerve conduction studies (NCS)" and "clinical assessment." The "traditional NCS" likely serves as the ground truth. It also mentions that "The data measured is reviewed and interpreted by a specially trained physician or specialist in neurophysiology."

  • Number of Experts: Not explicitly stated how many distinct experts were involved in interpreting the traditional NCS for ground truth establishment within the context of the studies. It refers to "a specially trained physician or specialist in neurophysiology" in general terms for interpreting Mediracer NCS data, but does not specify for the ground truth of the studies.
  • Qualifications of Experts: Assumed to be "specially trained physician or specialist in neurophysiology," as they are the ones who perform and interpret traditional NCS. Specific experience levels (e.g., "10 years of experience") are not provided.

4. Adjudication Method for the Test Set

The adjudication method for determining the true disease status based on the "traditional NCS" is not explicitly described. It only states that the traditional NCS was performed. It doesn't mention if multiple readers reviewed the traditional NCS, or if there was a consensus process (e.g., 2+1, 3+1).

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and the effect size of how much human readers improve with AI vs without AI assistance

  • MRMC Comparative Effectiveness Study: No, an MRMC comparative effectiveness study involving human readers with vs. without AI assistance was not done. The study compares the Mediracer NCS device (which is also used by humans) against traditional NCS devices, not the effect of AI on human reader performance. The Mediracer NCS itself is a measurement device, and while it has accompanying software (MAC), the study is evaluating the device's diagnostic capability, not an AI's interpretative assistance to a clinician.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) was done

  • Standalone Performance: No, a standalone (algorithm only) performance study was not done in the context of an AI algorithm. The Mediracer NCS is a device that requires a human operator ("used by physician or by technician on the order of physician") to take measurements, and a physician to review and interpret the data. The "Mediracer Analysis Center (MAC) software" is for storing and handling data, not an AI for autonomous interpretation or diagnosis.

7. The Type of Ground Truth Used

The ground truth used in the clinical studies was based on traditional nerve conduction studies (NCS) and clinical assessment for diagnosing Carpal Tunnel Syndrome (CTS). This would fall under expert consensus/clinical diagnosis derived from established diagnostic procedures.

8. The Sample Size for the Training Set

The document does not mention a training set for the Mediracer NCS device. This is likely because the Mediracer NCS is presented as a measurement device, not an AI/machine learning algorithm that requires a training phase. Its equivalence is established through direct comparison to existing measurement techniques and devices.

9. How the Ground Truth for the Training Set Was Established

Since no training set is mentioned for an AI/machine learning model, the method for establishing its ground truth is not applicable here.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.

Mediracer Ov Lassi Laitinen Product Specialist R&D Rantakatu 4 Oulu, 90100 FI

Re: K190536

Trade/Device Name: Mediracer NCS Regulation Number: 21 CFR 882.1550 Regulation Name: Nerve Conduction Velocity Measurement Device Regulatory Class: Class II Product Code: JXE Dated: August 26, 2019 Received: August 30, 2019

Dear Lassi Laitinen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for

Jay Gupta Assistant Director DHT5A: Division of Neurosurgical. Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K190536

Device Name Mediracer NCS

Indications for Use (Describe)

The Mediracer NCS is intended to measure sensory and motor neripheral nerves. The measured data can be utilized in evaluating patients suspected of having focal neuropathies. The measured data must be used in the context of other patient information and must be reviewed and interpreted by a physician.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image is a logo for a company called "mediracer". The logo consists of a pink abstract shape on the left, followed by the word "mediracer" in black and gray letters. Below the word "mediracer" is the phrase "POINT OF CARE DIAGNOSTIC TESTING" in smaller, gray letters. The logo is simple and modern, and the colors are eye-catching.

510(K) SUMMARY

Submission type:Traditional 510(k)
Basis for submission:New device introduced
Submitter information
Sponsor:Mediracer Oy (Ltd.)
Address:Rantakatu 4
90100 Oulu
Finland
Contact person:Georg Meissner, CEO
Phone:+358 40 517 0182
Submission date:19/09/2019

Device information

Trade name:Mediracer NCS
Common name:Nerve Conduction Testing System
Classification name:Nerve Conduction Study Measuring
Device Class:Device Class II (21 CFR 882.1550)
Product Code:JXE
Predicate device:
NeuroMetrix ADVANCEK070109

1 Device Description

The Mediracer NCS is intended to measure sensory and motor nerve conduction from peripheral nerves, when there is reason to suspect nerve damage. The measured data can be utilized in evaluating patients suspected of having focal neuropathies. Current application allows examination of Carpal Tunnel Syndrome related median nerve entrapment and ulnar nerve entrapment at the elbow. Test results can be used to determine if the findings are abnormal and in CTS, grade the severity of the nerve entrapment.

The Mediracer NCS is a hand-held NCS Device to stimulate and record the nerve responses from patient. The NCS Device is connected to the disposable surface electrodes with cables provided with the NCS Device. The NCS Device communicates via Bluetooth with a computer which is using The Mediracer Analysis Center (MAC) software. The computer

{4}------------------------------------------------

uses the MAC software to store and handle the patient measurement data. With the NCS Device comes a charger and a docking station for charging the NCS Device. Also, solution includes a test module for checking the functionality of the whole test system.

Mediracer NCS is used by physician or by technician on the order of physician for measuring sensory or motor nerve conductions from distal nerves. The measured data is transferred during the measurement to a computer via Bluetooth connection. The data measured is reviewed and interpreted by a specially trained physician or specialist in neurophysiology.

2 Intended Use

The Mediracer NCS is intended to measure sensory and motor nerve conduction from peripheral nerves. The measured data can be utilized in evaluating patients suspected of having focal neuropathies. The measured data must be used in the context of other patient information and must be reviewed and interpreted by a physician.

3 Comparison of technological characteristics with the predicate device

Mediracer NCS which is the subject of this Premarket Notification is substantially equivalent in sensory nerve conduction studies to the predicate device. NeuroMetrix ADVANCE. All the other functions of the predicate device are ignored in the comparison. Neuromuscular signal measurement is the technological principle for both the subject and predicate devices. It is based on stimulating the nerve with electric stimulus on skin above the nerve and the activated nerve response is measured on skin from the same nerve at some distance of the stimulus.

Difference between the devices are mainly in the design and the technological specifications due to the design. None of the different features of Mediracer NCS compared to predicate device have altered the basic performance, operation or intended use. At high level, the subject and predicate devices are based on the following same technological elements:

  • stimulating and recording electrodes
  • stimulation and recording cables
  • hand-held, point-of-care, device that performs the measurement
  • . Bluetooth for data transfer

The following technological differences exist between the subject and predicate devices:

  • Electrodes differ in design and placement. ●
  • The cables differ in design, the predicate uses the Proximal adapter in which the stimulation cable and recording cable are connected whereas the subject device has two separate cables which are connected straight to the device.
  • . Hand-held device differs from the display of data, the predicate uses touch screen for display of data whereas the subject device uses small LCD screen.
  • . The subject device must be used in conjunction with the Mediracer Analysis Center (MAC) whereas predicate device can be used without any external software.

{5}------------------------------------------------

Table 1. Device Comparison

Advance NCSMediracer NCS
Intended useDevice for measuringneuromuscular signals thatare useful as an aid indiagnosing and evaluatingpatients suspected ofhaving focal or systemicneuropathies.Device for stimulate andmeasure neuromuscularsignals that are useful indiagnosing and evaluatingsystemic and entrapmentneuropathies.
Power supplyHigh Capacity Lithium IonBatteryBattery: Nickel MetalHybrid NiHM battery 6V,730 mAh
Common Mode RejectionRatio90 dB min95...100 dB at 1 kHz
Noise<2µV RMS1.0 µV RMS (from 2 Hz to10 kHz)
Sampling Frequency20 kHz14.6 kHz
A/D Resolution24 bits10 bits
Stimulus TypeMonophasic or biphasicsquare waveDual phase, rectangularwave
Stimulator Max Voltage420 V163 V
Stimulation Frequency2 Hz2 Hz
Recording Channels11
ConnectivityBluetoothBluetooth
LatenciesMotor and sensorylatenciesMotor and sensorylatencies
Dimensions18.034 x 8.128 cm18.99 x 6.971 cm
Nerves TestedMedian and ulnar nervesMedian and ulnar nerves
Test Signal ControlDuring TestAutomatic or manualadjustment depending onthe used electrodesManual adjustment
Assessment time10 minutes15 minutes (both hands)
Access to test dataadjustmentCursor placing adjustmentCursor placing adjustment
ElectrodesNerve specific wrist anddigit electrodes or standardelectrodes (also elbow andfoot electrodes available)Stimulation Ring Electrodeand Recording Electrode
Patient Data Handlingand StorageOn the Device. All patientand test data included. Canbe uploaded to a DataServer.On the software includedon the system. All patientand test data included.
Operator SkillRequirementsTrained techniciansupervised by a physicianor a physicianTrained techniciansupervised by a physicianor a physician
Interpretation of the TestResultsPhysician or specialistlocallyPhysician or specialistlocally or remotely
TemperatureTemperature integrated sensorTemperature measurementwith a separate device
RegulationsCE -Marked; 60601-1complianceCE -Marked; 60601-1compliance

{6}------------------------------------------------

4 Performance data

The following performance data were provided in support of the substantial equivalence determination.

Biocompatibility testing

The Mediracer NCS component relevant for biocompatibility evaluation is the Electrode conductive hydrogel, which is in direct contact of patient skin while conducting the NCS test. Conductive hydrogel was assessed according to International Organization of Standardization 10993-1: Biological Evaluation of Medical Devices Part 1: Evaluation and Testing.

Performance Testing includes the biocompatibility test, bench test and clinical tests. Detailed descriptions of the biocompatibility test, bench test and clinical tests are included in the submission. The testing included the following tests:

  • Cytotoxicity ●
  • Irritation
  • Sensitization

The hydrogel passed all above ISO 10993 testing.

Electrical safety and electromagnetic compatibility (EMC)

Electrical safety and EMC testing were conducted on the Mediracer NCS device, consisting of the NCS device (ENG-CC-02), patient stimulator (PS-01), stimulation cable (SC-01) and recording cable (CC-02). The system complies with the IEC 60601-1, IEC 60601-2-40 standards for safety and the EN 60601-1-21 standard for EMC.

Software Verification and Validation Testing

Software verification and validation testing for the Mediracer NCS and MAC, were conducted, and documentation was provided as recommended by FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software of Mediracer NCS device and the MAC software was considered "Moderate" level of concern, since there is a risk of minor adverse events associated with Mediracer NCS and the MAC software is an accessory to a Medical Device (Mediracer NCS).

4 The European version (60601-1-2:2015) of the standard and the IEC-version (60601-1-2:2014) are equivalent to each other. In the European version the differences to its IEC counterpart are the exception of references of the 61000-4-x series and the addition of an Essential Requirements annex. The technical content of the IEC version is identical to the EN version.

{7}------------------------------------------------

Performance testing

Performance testing as a bench-testing, was conducted to the subject device to show equivalence to a previously approved device and to verify Mediracer NCS device accuracy. The performed test was same for both devices and the testing was performed by a trained operator.

The testing was performed to seven people, which six was within the measurement criteria set before testing. The six results taken into account with proper placement show significantly similar nerve response action potential values. The results show that the Mediracer NCS device results compared to conventional EMG device were matching inside the given deviation. Based on this comparison test it can be said that Mediracer NCS device is equivalent to a conventional EMG device and shows accurate readings.

Clinical Studies

The clinical validation was made against traditional nerve conduction test devices in two separate studies; one in Finland and the other in the United Kingdom. All the subjects examined in these tests had AANEM 2002 recommended set of traditional nerve conduction studies (NCS) and measurement with the Mediracer NCS done.

In the Finnish multicenter study, there were 194 recruited patients and 95 healthy control volunteers tested. The participants went through a clinical examination, Mediracer NCS measurement and the traditional NCS. In the United Kingdom study, the devices were compared only in the patient population, total of 63 patients participated in the study. The participants went through clinical assessment, Mediracer NCS measurement and the traditional NCS.

In the Finnish multicenter study, the number of patients studied were 194, aged 17-85. mean 49 years. Of these patients 147 (75.77 %) were female. Exclusion criteria for the study were prior surgery for CTS (the measured hand), neurological disorder that may produce numbness or paraesthesias in hand and those with diagnosed diabetes mellitus. For the study control values of 95 healthy volunteers aged from 16-80, mean 41 years, were examined. Of these volunteers 52 (54.73 %) were female. In the United Kingdom the patients were recruited from the normal referral stream to Leicester Carpal Tunnel Service, total of 104 referrals were selected and invited to a clinical study. No other exclusion criteria. Of the invited patients 65 fulfilled the clinical CTS diagnosis after symptom inquiry and clinical assessment. On the day of the study two participants cancelled their involvement, thus the total participation number were 63 patients. Of these patients 47 (74.60 %) were female.

Safety and effectiveness

In both studies positive per cent agreement between the traditional and the Mediracer NCS devices was around 90 % and negative per cent agreement 99-100 %. The new device study missed abnormalities mainly in cases with very mild conduction slowing and due to technical (one broken cable) or biological causes (patient was unable to relax). A high concordance between the used latency difference parameters between the traditional and the new device study were obtained. Moreover, these parameters were not dependent on skin temperature as absolute conduction velocities clearly are. The method is non-invasive and is safe to preform for the patients, no adverse effects or complications were reported during these studies.

Summary

{8}------------------------------------------------

Based on the two studies performed, the Mediracer NCS measurement device seems to reliably detect the median nerve lesion in CTS. In the studies were found high degree of agreement between the systems compared, the new and traditional methods of NCS testing.

5 Conclusions

The non-clinical data support the safety of the device and the Mediracer NCS and MAC software verification and validation demonstrate that the Mediracer NCS device should perform as intended in the normal use conditions. The clinical data and bench-testing demonstrates that nerve conduction measurements obtained using the Mediracer NCS are comparable to those obtained using conventional nerve conduction measurement equipment.

§ 882.1550 Nerve conduction velocity measurement device.

(a)
Identification. A nerve conduction velocity measurement device is a device which measures nerve conduction time by applying a stimulus, usually to a patient's peripheral nerve. This device includes the stimulator and the electronic processing equipment for measuring and displaying the nerve conduction time.(b)
Classification. Class II (performance standards).