K Number
K190529

Validate with FDA (Live)

Device Name
SOZO
Manufacturer
Date Cleared
2019-11-25

(266 days)

Product Code
Regulation Number
870.2770
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The SOZO system may be used as an adjunct to existing methods by aiding clinicians who are using Subjective Global Assessment (SGA) tools to assess patients at risk of protein-calorie malnutrition (PCM).

The SOZO system may be further used to estimate the following body composition parameters in humans to track clinically relevant body composition parameters over time:

  • Fat mass
  • Fat-free mass
  • Total body water
  • Intracellular fluid
  • Extracellular fluid
  • Skeletal muscle mass

The following outputs are also presented:

  • Body Mass Index (BMI)
  • Basal metabolic rate (BMR; based on Mifflin - St. Jeor's algorithm) displayed in calories per day
  • Protein and mineral (also known as 'dry lean mass') represents the content of a body that is not fat or fluid; calculated by subtracting total body water weight from fat-free mass weight.

The SOZO device measures current (I), voltage (V) and phase angle (Phi), and from these values calculates resistance (R), reactance (Xc), and impedance (Z), which are used to estimate the above body composition parameters. The device/software will also display the Cole plot, subject height, weight, age and sex.

Device Description

The SOZO system consists of a connected hand and footplate with built-in stainless steel electrodes, paired with an Android tablet over Bluetooth connection. An app ("SOZOapp"), supplied with the tablet, controls the functionality of the hardware and supplies the bioimpedance measurement data to a database ("SOZOhub") contained within the hospital/facility network.

Measurements require the patient to make contact with bare hands and feet on stainless steel electrodes. The measurement takes about 30 seconds, during which the SOZO® system applies small levels of electrical current (200µA RMS) to the body across 256 frequencies spaced logarithmically from 3kHz to 1000kHz and measures the resulting voltage levels. Established algorithms are used to analyze data and calculate various body composition parameters and present the outputs for the clinician to review.

AI/ML Overview

Here's an analysis of the acceptance criteria and study information provided in the document:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategoryReported Device Performance
Electrical Safety/EMCMeets requirements set forth in IEC 60601 (subparts -1, -1-2, and -1-6). CB certificate granted.
Software V&VMeets acceptance criteria and performs as intended, tested in accordance with ISO 62304.
BiocompatibilityPassed testing according to ISO 10993 for a low-risk, limited-contact device; no failures reported.
Functional PerformanceSystem expected to remain functional throughout its intended life; tested for design reliability by repeatedly placing weights on components.
Comparative Performance vs. Predicate Device (SFB7)Very strong correlation (r > 0.99) with outputs from the SFB7 device when tested using a fixture with multiple fixed impedance loads representing different "humans."

2. Sample Sizes Used for the Test Set and Data Provenance

The document does not explicitly state a specific "test set" sample size for a clinical study to directly prove the new indications for use. Instead, it states: "Clinical testing: clinical studies have identified the applicability of ImpediMed's BIS technology to act as an aid in body composition." This general statement implies that the underlying Bioimpedance Spectroscopy (BIS) technology has been validated through clinical studies, but it doesn't provide details about studies specific to the SOZO system's new indications or a dedicated test set with sample size.

For the comparative performance with the predicate device, a "test fixture" was used with "multiple fixed impedance loads," which isn't a human sample size but rather an engineering test.

Data Provenance: Not explicitly stated for specific clinical studies, only generally refers to "clinical studies."

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

This information is not provided in the document. The document refers to "clinical studies" for the applicability of BIS technology but does not detail how ground truth was established for these studies, nor the number or qualifications of experts involved.

4. Adjudication Method for the Test Set

This information is not provided in the document.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of Improvement with AI vs. Without AI Assistance

No MRMC study is mentioned in the document. The SOZO system is a device for measuring body composition parameters, not an AI-assisted diagnostic tool for human readers in the traditional sense of an MRMC study. Its use is described as "an adjunct to existing methods by aiding clinicians who are using Subjective Global Assessment (SGA) tools," but no study comparing human performance with and without SOZO assistance is described.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

The device itself is a standalone measurement system that takes raw electrical measurements and processes them using "established algorithms" to calculate body composition parameters. The "comparative performance vs. predicate device" test (r > 0.99 correlation) can be considered a standalone performance assessment against a validated reference.

7. The Type of Ground Truth Used

  • For Body Composition Parameters: The document states that "established algorithms are used to analyze data and calculate various body composition parameters." For the "comparative performance vs. predicate device," the predicate device (SFB7) is essentially the reference standard (ground truth surrogate). However, for the initial validation of the BIS technology to derive these parameters, it is generally understood that such systems are validated against gold standard body composition assessment methods (e.g., DEXA, isotope dilution, direct chemical analysis), though this specific detail is not provided for the SOZO system in this document. The document refers to "clinical studies have identified the applicability of ImpediMed's BIS technology to act as an aid in body composition," implying prior validation work.
  • For Protein-Calorie Malnutrition (PCM) Assessment: The device is used as an adjunct to Subjective Global Assessment (SGA) tools, implying that SGA is the primary assessment method clinicians use.

8. The Sample Size for the Training Set

No training set sample size is explicitly mentioned. The device uses "established algorithms," suggesting that the algorithms were developed and potentially trained on datasets prior to this submission. This document focuses on the performance and equivalence of the SOZO device itself, not the detailed development of its underlying algorithms.

9. How the Ground Truth for the Training Set Was Established

This information is not provided in the document. As with the training set size, the document does not delve into the historical development and validation of the "established algorithms" that process the bioimpedance data.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA acronym with the full name of the agency on the right. The symbol on the left is a stylized representation of a human figure, while the FDA acronym is in blue, followed by the full name of the agency, "U.S. Food & Drug Administration," in a smaller font size.

November 25, 2019

ImpediMed Limited % Reuben Lawson Senior Directory, Regulatory Affairs and Clinical ImpediMed, Inc. 5900 Pasteur Court, Unit 125 Carlsbad, CA 92008

Re: K190529

Trade/Device Name: SOZO® Regulation Number: 21 CFR 870.2770 Regulation Name: Impedance plethysmograph Regulatory Class: II Product Code: OJB, DSB, MNW Dated: October 24, 2019 Received: October 24, 2019

Dear Reuben Lawson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of

U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov

{1}------------------------------------------------

Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Carolyn Y. Neuland, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K190529

Device Name SOZO®

Indications for Use (Describe)

The SOZO system may be used as an adjunct to existing methods by aiding clinicians who are using Subjective Global Assessment (SGA) tools to assess patients at risk of protein-calorie malnutrition (PCM).

The SOZO system may be further used to estimate the following body composition parameters in humans to track clinically relevant body composition parameters over time:

  • Fat mass
  • · Fat-free mass
  • Total body water
  • Intracellular fluid
  • · Extracellular fluid
  • · Skeletal muscle mass

The following outputs are also presented:

  • · Body Mass Index (BMI)
  • · Basal metabolic rate (BMR; based on Mifflin St. Jeor's algorithm) displayed in calories per day

· Protein and mineral (also known as 'dry lean mas') represents the content of a body that is not fat or fluid; calculated by subtracting total body water weight from fat-free mass weight.

The SOZO device measures current (I), voltage (V) and from these values calculates resistance (R), reactance (Xc), and impedance (Z), which are used to estimate the above body composition parameters. The device/ software will also display the Cole plot, subject height, weight, age and sex.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) SUMMARY lmpediMed's SOZO® system

Submitter:

ImpediMed LimitedUnit 150 Parker CourtPinkenba Qld 4008Australia
Phone:760 585 2104
Facsimile:760 804 9245
Contact Person:Reuben Lawson
Date Prepared:November 25, 2019
Name of Device:SOZO®
Common or Usual Name:Body Fluid Analyzer
Regulation Number:21 CFR§870.2770
Regulation Name:Impedance Plethysmograph
Regulatory Class:II
Product Codes:QJB, DSB, MNW
Predicate Device:ImpediMed Limited's SOZO (K172507); ImpediMed Limited's IMPSFB7 (K052319)
Reference Devices:ImpediMed Limited's SOZO (K172122; K180126)

Device Description

The SOZO system consists of a connected hand and footplate with built-in stainless steel electrodes, paired with an Android tablet over Bluetooth connection. An app ("SOZOapp"), supplied with the tablet, controls the functionality of the hardware and supplies the bioimpedance measurement data to a database ("SOZOhub") contained within the hospital/facility network.

{4}------------------------------------------------

Measurements require the patient to make contact with bare hands and feet on stainless steel electrodes. The measurement takes about 30 seconds, during which the SOZO® system applies small levels of electrical current (200µA RMS) to the body across 256 frequencies spaced logarithmically from 3kHz to 1000kHz and measures the resulting voltage levels. Established algorithms are used to analyze data and calculate various body composition parameters and present the outputs for the clinician to review.

Intended Use / Indications for Use

The SOZO® Body Fluid Analyzer is intended for the following uses:

The SOZO system may be used as an adjunct to existing methods by aiding clinicians who are using Subjective Global Assessment (SGA) tools to assess patients at risk of proteincalorie malnutrition (PCM).

The SOZO system may be further used to estimate the following body composition parameters in humans to track clinically relevant body composition parameters over time:

  • Fat mass
  • . Fat-free mass
  • Total body water
  • . Intracellular fluid
  • Extracellular fluid
  • . Skeletal muscle mass

The following outputs are also presented:

  • Body Mass Index (BMI) ●
  • . Basal metabolic rate (BMR; based on Mifflin - St. Jeor's algorithm) displayed in calories per day
  • . Protein and mineral (also known as 'dry lean mass') represents the content of a body that is not fat or fluid; calculated by subtracting total body water weight from fat-free mass weight.

The SOZO device measures current (1), voltage (V) and phase angle (Phi), and from these values calculates resistance (R), reactance (Xc), and impedance (Z), which are used to estimate the above body composition parameters. The device/software will also display the Cole plot, subject height, weight, age and sex.

Summary of Technological Characteristics

Bioimpedance spectroscopy is the technological principle for both the subject and predicate devices. The subject and predicate devices are based on the following same fundamental technological elements:

{5}------------------------------------------------

  • Use of electrodes to take measurements: two 'drive' and two 'sense' channels are . used to measure each side of the body
  • 'Drive' channels deliver very low levels of current (200µA RMS) across 256 . frequencies logarithmically spaced from 3kHz to 1000kHz;
  • 'Sense' channels measure current (I), voltage (V) and phase angle (Ph), and . calculate three bioimpedance parameters: impedance (Z), resistance (R) and reactance (Xc) to estimate impedance levels at different frequencies and use established algorithms to calculate body composition parameters;
  • Data is stored in and accessed from a local database (SOZOhub) utilizing separate . software installed on a network connected PC.

Thus, SOZO is the same as its predicate SOZO device. The technological characteristics of the SOZO system and equivalence to the reference SFB7 system have previously been reviewed by FDA in K172122, K172507, and K180126.

Performance Data

The SOZO system has gone through appropriate testing per design controls to confirm functionality and performance of the new indications.

Electrical safety/EMC: testing was performed according to the requirements set forth in IEC 60601 (subparts -1. -1-2. and -1-6). It was determined that the SOZO device meets electrical safety and EMC requirements, and CB certificate was granted for the system.

Software V&V: the same 'level of concern' software documentation as the predicate device was created and testing performed in accordance with ISO 62304. The software was verified and validated to meet acceptance criteria and perform as intended.

Biocompatibility: testing was performed by an accredited third party according to the requirements set forth in ISO 10993 for a low risk, limited contact device. It was determined that the SOZO system passed biocompatibility testing with no failures reported.

Functional performance: multiple SOZO systems were tested for design reliability by repeatedly placing weights on the components that encounter the most physical stress. Testing showed that the system is expected to remain functional throughout its intended life.

Clinical testing: clinical studies have identified the applicability of ImpediMed's BIS technology to act as an aid in body composition.

Comparative performance vs. predicate device: using a test fixture to create multiple fixed impedance loads representing different 'humans', a SOZO system was compared against ImpediMed's SFB7 system to verify correlation in outputs. The SOZO system showed a very strong correlation (r > 0.99) compared to outputs from the SFB7 device.

{6}------------------------------------------------

In all instances, the SOZO system functioned as intended and all results observed were as expected.

Conclusions

The SOZO system is as safe and effective as its predicate device. The SOZO system has the same intended uses and similar indications, technological characteristics, and principles of operation as its predicate device. In addition, the minor software and labeling differences between the SOZO system and its predicate device raise no new issues of safety or effectiveness. Performance data demonstrate that the SOZO system is as safe and effective as its predicate device. Thus, the SOZO system is substantially equivalent.

§ 870.2770 Impedance plethysmograph.

(a)
Identification. An impedance plethysmograph is a device used to estimate peripheral blood flow by measuring electrical impedance changes in a region of the body such as the arms and legs.(b)
Classification. Class II (special controls). The device, when it is a body composition analyzer which is not intended to diagnose or treat any medical condition, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.