K Number
K190525

Validate with FDA (Live)

Manufacturer
Date Cleared
2019-05-03

(60 days)

Product Code
Regulation Number
874.4420
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The RELIEVA ULTIRRA® Sinus Balloon Catheter is an instrument intended to dilate sinus ostia and spaces within the paranasal sinus cavities for diagnostic procedures. It is also intended to irrigate from within a target sinus for therapeutic procedures and to facilitate diagnostic procedures.

For children aged 17 and under, the balloon catheter system is intended to dilate sinus ostia and spaces associated with the maxillary sinus for diagnostic and therapeutic procedures. It is also intended to irrigate from within a target sinus for therapeutic procedures and to facilitate diagnostic procedures.

The RELIEVA ULTIRRA® Sinus Balloon Catheter may be utilized in conjunction with TruDi™ NAV Wire and TruDi™ Navigation System to provide access to nasal and sinus spaces, and to confirm placement in the accessed anatomy. It is NOT intended to irrigate from within a target sinus for therapeutic procedures nor to facilitate diagnostic procedures with TruDi™ NAV Wire.

Device Description

The RELIEVA ULTIRRA® Sinus Balloon Catheter is a flexible catheter that is intended to dilate sinus ostia. The shaft allows for inflation of the sinus balloon and permits the passage of an electromagnetic navigable sinus guidewire or sinus illumination system to facilitate access to the target sinus ostia. A hypotube is incorporated on the proximal end to provide rigidity during insertion through a sinus guide catheter.

AI/ML Overview

Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text.

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriterionReported Device Performance
Simulated Use TestingMet all performance acceptance criteria
Dimensional SpecificationsMet all performance acceptance criteria
Balloon Burst PressureMet all performance acceptance criteria
Joint Separation ForceMet all performance acceptance criteria
Deflation TimeMet all performance acceptance criteria
Balloon Cycle FatigueMet all performance acceptance criteria
Shelf LifeEstablished per ASTM F1980-07, ASTM F88/F88M-09, ISTA 2A, and ASTM F2096-11 requirements.
BiocompatibilityBiocompatible per ISO 10993-1.
Sterilization Assurance Level (SAL)10^-6, validated per AAMI/ANSI/ISO 11135:2014 (overkill/half-cycle approach in a fixed chamber).
Ethylene Oxide ResidualsMet ISO 10993-7:2008 requirements.
Packaging Shelf LifeEstablished per ASTM F1980-07.

2. Sample Size for Test Set and Data Provenance

The document states, "Clinical data was not necessary for the RELIEVA ULTIRRA® Sinus Balloon Catheter." This indicates that no clinical test set was used, and therefore no sample size or data provenance for a clinical test set is applicable. The evaluation was based on non-clinical performance testing.

3. Number of Experts and Qualifications for Ground Truth

Since no clinical data or human-in-the-loop performance was performed for this 510(k) submission, the concept of "experts used to establish ground truth" is not applicable in the context of evaluating diagnostic accuracy or clinical decision-making. The ground truth for the non-clinical performance tests would be the established engineering and materials standards and specifications. Engineers and technicians with expertise in medical device testing would have established and verified these.

4. Adjudication Method for Test Set

Not applicable, as no clinical test set requiring adjudication was used.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No MRMC comparative effectiveness study was mentioned or performed. The submission explicitly states "Clinical data was not necessary."

6. Standalone (Algorithm Only) Performance Study

Not applicable. This device is a manual surgical instrument (catheter), not an algorithm or AI system. The performance studies conducted were non-clinical, evaluating the physical and mechanical properties of the device.

7. Type of Ground Truth Used

The ground truth for the non-clinical performance data was based on established engineering standards, material specifications, and regulatory requirements (e.g., ISO, ASTM standards) for medical device performance (e.g., burst pressure, fatigue, biocompatibility, sterilization).

8. Sample Size for Training Set

Not applicable. As a physical medical device, there is no "training set" in the context of machine learning or AI. The device's design and manufacturing processes are developed and refined through engineering principles and iterative testing, not through data-driven training sets as understood in AI systems.

9. How Ground Truth for Training Set Was Established

Not applicable, as there is no training set in this context. The "ground truth" for the device's development and manufacturing would be its intended function, design specifications, and relevant industry standards.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

May 3, 2019

Acclarent, Inc. Leena Sorathia Sr. Regulatory Affairs Specialist 33 Technology Drive Irvine, California 92618

Re: K190525

Trade/Device Name: RELIEVA ULTIRRA Sinus Balloon Catheter Regulation Number: 21 CFR 874.4420 Regulation Name: Ear, nose, and throat manual surgical instrument Regulatory Class: Class I Product Code: LRC Dated: March 1, 2019 Received: March 4, 2019

Dear Leena Sorathia:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Michael Ryan Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

510(k) Number (if known)

K190525

Device Name RELIEVA ULTIRRA® Sinus Balloon Catheter

Indications for Use (Describe)

The RELIEVA ULTIRRA® Sinus Balloon Catheter is an instrument intended to dilate sinus ostia and spaces within the paranasal sinus cavities for diagnostic procedures. It is also intended to irrigate from within a target sinus for therapeutic procedures and to facilitate diagnostic procedures.

For children aged 17 and under, the balloon catheter system is intended to dilate sinus ostia and spaces associated with the maxillary sinus for diagnostic and therapeutic procedures. It is also intended to irrigate from within a target sinus for therapeutic procedures and to facilitate diagnostic procedures.

The RELIEVA ULTIRRA® Sinus Balloon Catheter may be utilized in conjunction with TruDi™ NAV Wire and TruDi™ Navigation System to provide access to nasal and sinus spaces, and to confirm placement in the accessed anatomy. It is NOT intended to irrigate from within a target sinus for therapeutic procedures nor to facilitate diagnostic procedures with TruDi™ NAV Wire.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "Acclarent" in a stylized font. The first part of the word, "Acclar" is in black, while the "ent" is in purple. There is a purple swoosh above the "ent" part of the word.

510(K) SUMMARY

[807.92(a)(1)] Submitter Information

Sponsor/Submitter:Acclarent, Inc.33 Technology DriveIrvine, CA 92618
Contact Person:Leena SorathiaSr. Regulatory Affairs SpecialistEmail: lsorathi@its.jnj.comTel: 949-923-4118

Date Summary Prepared: May 3, 2019

[807.92(a)(2)] Name of Device

Device TradeName:RELIEVA ULTIRRA® Sinus Balloon Catheter
ClassificationName:Ear, Nose, and Throat Manual Surgical Instrument
Common Name:Sinus Balloon Catheter
DeviceClassification:Class I
RegulationNumber:21 CFR 874.4420
Review Panel:Ear, Nose, and Throat
Product Code:LRC
[807.92(a)(3)] Legally Marketed Devices
Predicate Device:RELIEVA Solo Elite™ Sinus Balloon Catheter (K111254)
Reference Device:RELIEVA UltirraNav® Sinus Balloon Catheter, (K161698)
[807.92(a)(4)] Device Description
Device Description:The RELIEVA ULTIRRA® Sinus Balloon Catheter is a flexible catheter that isintended to dilate sinus ostia. The shaft allows for inflation of the sinus balloon andpermits the passage of an electromagnetic navigable sinus guidewire or sinusillumination system to facilitate access to the target sinus ostia. A hypotube is

incorporated on the proximal end to provide rigidity during insertion through a sinus guide catheter.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Acclarent. The word "Acclarent" is written in a sans-serif font, with the first part of the word in black and the "rent" portion in purple. Above the purple part of the word is a curved line, also in purple, that arches over the text.

[807.92(a)(5)] Intended Use

Indications for Use:The RELIEVA ULTIRRA® Sinus Balloon Catheter is an instrument intended todilate sinus ostia and spaces within the paranasal sinus cavities for diagnostic andtherapeutic procedures. It is also intended to irrigate from within a target sinus fortherapeutic procedures and to facilitate diagnostic procedures.
For children aged 17 and under, the balloon catheter system is intended to dilate sinusostia and spaces associated with the maxillary sinus for diagnostic and therapeuticprocedures. It is also intended to irrigate from within a target sinus for therapeuticprocedures and to facilitate diagnostic procedures.
The RELIEVA ULTIRRA® Sinus Balloon Catheter may be utilized in conjunctionwith TruDi™ NAV Wire and TruDi™ Navigation System to provide access to nasaland sinus spaces, and to confirm placement in the accessed anatomy. It is NOTintended to irrigate from within a target sinus for therapeutic procedures nor tofacilitate diagnostic procedures with TruDi™ NAV Wire.
Difference inIndications fromPredicate DeviceThe indications for use statement of the subject device, RELIEVA ULTIRRA® SinusBalloon Catheter, is similar to predicate device, RELIEVA Solo Elite™ Sinus BalloonCatheter (K111254). The subject and predicate devices are both intended to access anddilate the sinus ostia and paranasal spaces using a balloon.
The proposed revised indications for use for the subject device are largely unchangedfrom the predicate device with the exception of the additional statement that thedevice may be used in conjunction with TruDi™ NAV Wire and TruDi™ NavigationSystem. In addition, the irrigation function of the subject device is not intended to beused in conjunction with the TruDi™ NAV Wire.
The changes in indications for use of the subject device do not raise any significantissues of safety and effectivenessof the device when used as labeled, as demonstrated bperformance testing.
The difference in indications for use between the subject device and thepredicate/reference devices is supported is presented in Table 1 of this summary.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the word "Acclarent" in a stylized font. The first part of the word, "Acclar" is in black, while the "ent" is in purple. There is a purple curved line above the "ent" portion of the word.

[807.92(a)(6)] Technical Characteristics

TechnologicalCharacteristics:The design and materials of the subject device, RELIEVA ULTIRRA® SinusBalloon Catheter, are identical to the predicate device RELIEVA Solo Elite™ SinusBalloon Catheter (K111254). The added capability of the subject device to be usedwith the TruDi™ NAV Wire and TruDi™ Navigation System did not require anydesign changes from the predicate device.
Like the predicate device, the subject device has the same ballon dilation pressure,material composition, and size specifications. However, the subject device does notallow for the irrigation feature.
The subject device is substantially equivalent in technological characteristics, asthere are no significant differences in design, fundamental scientific technology, orother features of the device as compared to the predicate device.
See Table 1 for a comparison of the technological characteristics between thesubject device and the predicate/reference devices.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for "Acclarent." The word "Acclarent" is written in a sans-serif font, with the first part of the word in black and the "ent" portion in purple. There is a purple curved line above the "ent" portion of the word, and a trademark symbol to the right of the "t".

AttributePrimary Predicate Device:Reference Device:Subject Device:
RELIEVA Solo Elite™ Sinus GuideRELIEVA ULTIRRA® Nav SinusRELIEVA ULTIRRA® Sinus Balloon
Catheter (K111254)Balloon Catheter (K161698)Catheter (K190525)
ManufacturerAcclarentAcclarentAcclarent
Common NameSinus Balloon Dilation SystemSinus Balloon Dilation SystemSinus Balloon Dilation System
ClassIII
Product CodeClassificationSectionLRC21 CFR 874.4420LRC21 CFR 874.4420LRC21 CFR 874.4420
Indications for UseThe RELIEVA Ultirra® Sinus BalloonCatheter is an instrument intended to dilatesinus ostia and spaces within the paranasalsinus cavities for diagnostic andtherapeutic procedures. It is also intendedto irrigate from within a target sinus fortherapeutic procedures and to facilitatediagnostic procedures.For children aged 17 and under, theballoon catheter system is intended todilate sinus ostia and spaces associatedwith the maxillary sinus for diagnostic andtherapeutic procedures. It is also intendedto irrigate from within a target sinus fortherapeutic procedures and to facilitatediagnostic procedures.The RELIEVA ULTIRRA® Nav SinusBalloon Catheter is an instrument intendedto dilate the sinus ostia and spaces withinthe paranasal sinus cavities for diagnosticand therapeutic procedures.For children aged 17 and under, the ballooncatheter system is intended to dilate sinusostia and spaces associated with themaxillary sinus for diagnostic andtherapeutic procedures.The RELIEVA ULTIRRA® Sinus BalloonCatheter may be utilized in conjunctionwith the Acclarent NavWire™ SinusNavigation Guidewire and ACCLARENT®ENT Navigation System to provide access tonasal and sinus spaces, and to confirmplacement in the accessed anatomy.The RELIEVA ULTIRRA® Sinus BalloonCatheter is an instrument intended to dilatesinus ostia and spaces within the paranasalsinus cavities for diagnostic and therapeuticprocedures. It is also intended to irrigate fromwithin a target sinus for therapeutic proceduresand to facilitate diagnostic procedures.For children aged 17 and under, the ballooncatheter system is intended to dilate sinus ostiaand spaces associated with the maxillary sinusfor diagnostic and therapeutic procedures. It isalso intended to irrigate from within a targetsinus for therapeutic procedures and tofacilitate diagnostic procedures.The RELIEVA ULTIRRA® Sinus BalloonCatheter may be utilized in conjunction withTruDi™ NAV Wire and TruDi™ NavigationSystem to provide access to nasal and sinusspaces, and to confirm placement in theaccessed anatomy. It is NOT intended toirrigate from within a target sinus fortherapeutic procedures nor to facilitatediagnostic procedures with TruDi™ NAV Wire.
Indicated forChildrenYes, for children 17 years and under inmaxillary sinus onlyYes, for children 17 years and under inmaxillary sinus onlyYes, for children 17 years and under inmaxillary sinus only
AttributePrimary Predicate Device:RELIEVA Solo Elite™ Sinus GuideReference Device:RELIEVA ULTIRRA® Nav SinusSubject Device:RELIEVA ULTIRRA® Sinus Balloon
Catheter (K111254)Balloon Catheter (K161698)Catheter (K190525)
Single Patient UseYesYesYes
Direct PatientContactYesYesYes
Labeled as Non-Pyrogenic?NoNoNo
TechnologicalCharacteristicsAllows for dilation of sinus ostia with theadded capability to irrigate.Allows for dilation of sinus ostia.Allows for dilation of sinus ostia with theadded capability to irrigate when used with theRELIEVA LUMA SENTRY®. It is NOTintended to irrigate from within a target sinusfor therapeutic procedures nor to facilitatediagnostic procedures with TruDi™ NAV Wire.
Constructed ofMaterialsCommonly Usedin PatientContactingMedical DevicesYesYesYes
Balloon MaterialNylonNylonNylon
Balloon Diameterand Length5x16mm, 6x16mm, 7x16mm and 7x24mm5x16mm5x16mm, 6x16mm, 7x16mm and 7x24mm
MaximumInflation Pressure12 ATM12 ATM12 ATM
Flexible ShaftYesYesYes
Deflation Time< 5 seconds< 3.6 seconds< 5 seconds
Catheter Length250 ± 2 mm250 ± 2mm250 ± 2mm
Shaft DesignDual LumenDual LumenDual Lumen
Used withGuidewireYesYesYes
GuidewireCompatibility0.035"0.035"0.035"
AttributePrimary Predicate Device:RELIEVA Solo Elite™ Sinus GuideCatheter (K111254)Reference Device:RELIEVA ULTIRRA® Nav SinusBalloon Catheter (K161698)Subject Device:RELIEVA ULTIRRA® Sinus BalloonCatheter (K190525)
BalloonRadiopaqueMarker BandsYesYesYes
Balloon SlideMechanismYesYesYes
IrrigationCapabilityYesNoYes, if used with RELIEVA LUMA SENTRY®No, if used TruDi™ NAV Wire
IncorporatedSuctionNoNoNo
Sinuses in whichdevice is intendedfor use: [Sphenoid,Maxillary, Frontal]Sphenoid, Maxillary, FrontalSphenoid, Maxillary, FrontalSphenoid, Maxillary, Frontal
EtO SterilizedYesYesYes
PackagingHDPE backer card in Tyvek/Nylon pouchHDPE backer card in Tyvek/Nylon pouchHDPE backer card in Tyvek/Nylon pouch
Accessory DevicesPacked withDeviceStyletStyletStylet
Principles ofOperationManually operated device. Balloon inflatedwith sterile saline or water to mechanicallydilate sinus ostia.Manually operated device. Balloon inflatedwith sterile saline or water to mechanicallydilate sinus ostia.Manually operated device. Balloon inflatedwith sterile saline or water to mechanicallydilate sinus ostia.
Labeled forcompatibility withImage NavigationSystemsNoYesYes

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for Acclarent. The word "Acclarent" is written in black font. To the right of the word is a purple design that looks like a swoosh.

Traditional 510(k) Premarket Notification

RELIEVA ULTIRRA® Sinus Balloon Catheter

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the word "Acclarent" in black font. To the right of the word is a purple logo that appears to be a stylized letter "A" or a swoosh-like design. The logo is slightly curved and has a gradient effect, with the color fading from a darker purple to a lighter shade.

Traditional 510(k) Premarket Notification

RELIEVA ULTIRRA® Sinus Balloon Catheter

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for Acclarent. The word "Acclarent" is written in a sans-serif font. The first part of the word, "Accla," is in black, while the "rent" portion is in purple. There is a curved line above the purple portion of the word.

[807.92(b) (1)] Determination of Substantial Equivalence

Non-Clinical PerformanceData:The RELIEVA ULTIRRA® Sinus Balloon Catheter met all performanceacceptance criteria including simulated use testing, dimensionalspecifications, balloon burst pressure, joint separation force, deflationtime, and balloon cycle fatigue.Shelf life was established per ASTM F1980-07 ASTM F88/F88M-09,ISTA 2A, and ASTM F2096-11 requirements.Testing also showed that the RELIEVA ULTIRRA® Sinus BalloonCatheter is biocompatible per ISO 10993-1.The sterilization process has been validated per AAMI/ANSI/ISO11135:2014 and demonstrated a sterility assurance level of 10-6. Themethod used for sterilization validation was the overkill (half-cycleapproach) in a fixed chamber. Ethylene oxide residuals were tested andmet ISO 10993-7:2008 requirements. The subject device is not testednor labeled as “non-pyrogenic”.Packaging shelf life has been established per ASTM F1980-07.Clinical data was not necessary for the RELIEVA ULTIRRA® SinusBalloon Catheter. The performance data demonstrated that the deviceperforms as intended.The RELIEVA ULTIRRA® Sinus Balloon Catheter passed all intendedcriteria in accordance with appropriate test criteria and standards.
[807.92(b) (2)] Determination of Substantial Equivalence
Clinical Performance DataClinical data was not necessary for the RELIEVA ULTIRRA® SinusBalloon Catheter. The performance data demonstrated that the deviceperforms as intended.
[807.92(b) (3)] Conclusion
Conclusion from Non-Clinical and Clinical TestsBased on the information provided in this premarket notification,Acclarent concludes that the RELIEVA ULTIRRA® SinusBalloon Catheter is as safe and effective and substantiallyequivalent to the predicate device.

§ 874.4420 Ear, nose, and throat manual surgical instrument.

(a)
Identification. An ear, nose, and throat manual surgical instrument is one of a variety of devices intended for use in surgical procedures to examine or treat the bronchus, esophagus, trachea, larynx, pharynx, nasal and paranasal sinus, or ear. This generic type of device includes the esophageal dilator; tracheal bistour (a long, narrow surgical knife); tracheal dilator; tracheal hook; laryngeal injection set; laryngeal knife; laryngeal saw; laryngeal trocar; laryngectomy tube; adenoid curette; adenotome; metal tongue depressor; mouth gag; oral screw; salpingeal curette; tonsillectome; tonsil guillotine; tonsil screw; tonsil snare; tonsil suction tube; tonsil suturing hook; antom reforator; ethmoid curette; frontal sinus-rasp; nasal curette; nasal rasp; nasal rongeur; nasal saw; nasal scissors; nasal snare; sinus irrigator; sinus trephine; ear curette; ear excavator; ear rasp; ear scissor, ear snare; ear spoon; ear suction tube; malleous ripper; mastoid gauge; microsurgical ear chisel; myringotomy tube inserter; ossici holding clamp; sacculotomy tack inserter; vein press; wire ear loop; microrule; mirror; mobilizer; ear, nose, and throat punch; ear, nose and throat knife; and ear, nose, and throat trocar.(b)
Classification Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 874.9.