AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Xpert® CT/NG test, performed on the GeneXpert® Instrument Systems, is a qualitative in vitro real-time PCR test for the automated detection and differentiation of genomic DNA from Chlamvdia trachomatis (CT) and/or Neisseria gonorrhoeae (NG) to aid in the diagnosis of chlamydial and gonorrheal disease in the urogenital tract and extragenital sites (pharynx and rectum). The assay may be used to test the following specimens from asymptomatic and symptomatic individuals: female and male urine, patient-collected vaginal swabs (collected in a clinical setting), clinician-collected endocervical swabs, and female and male pharyngeal and rectal swabs.

Device Description

The Xpert CT/NG test is an automated in vitro diagnostic test for qualitative detection and differentiation of DNA from Chlamydia trachomatis (CT) and/or Neisseria gonorrhoeae (NG). The test is performed on the Cepheid GeneXpert Instrument Systems. The Xpert CT/NG test on the GeneXpert Instrument System automates and integrates sample purification, nucleic acid amplification and detection of the target sequences in simple or complex samples using real-time PCR. The system consists of an instrument, personal computer, and preloaded software for running the tests and viewing the results. The system requires the use of single-use disposable cartridges that hold the PCR reagents and host the PCR process. Because the cartridges are self-contained, crosscontamination between samples is minimized.

The Xpert CT/NG test includes reagents for the detection and differentiation of CT and NG. A Sample Processing Control (SPC), a Sample Adequacy Control (SAC), and a Probe Check Control (PCC) are also included. The SPC is present to control for adequate processing of the target bacteria and to monitor the presence of inhibitors in the PCR reaction. The SAC reagents detect the presence of a single copy human gene and monitor whether the specimen contains human cells. The PCC verifies reagent rehydration, PCR tube filling in the cartridge, probe integrity, and dye stability.

The GeneXpert Instrument Systems, comprised of the GeneXpert Dx Systems, the GeneXpert Infinity-48 System. GeneXpert Infinity-48s. and the GeneXpert Infinity-80 System, have 1 to 80 randomly accessible modules, depending upon the instrument, that are each capable of performing separate sample preparation and real-time PCR tests. Each module contains a syringe drive for dispensing fluids (i.e., the syringe drive activates the plunger that works in concert with the rotary valve in the cartridge to move fluids between chambers), an ultrasonic horn for lysing cells or spores, and a proprietary I-CORE® thermocycler for performing real-time PCR and detection.

The ancillary specimen collection kits for use with the Xpert CT/NG test are the Xpert Vaginal/Endocervical Specimen Collection Kit, Xpert Swab Specimen Collection kit and the Xpert Urine Specimen Collection kit.

AI/ML Overview

This looks like a 510(k) summary for the Cepheid Xpert CT/NG test. The document provides information on analytical and clinical performance studies.

Here's a breakdown of the requested information based on the provided text:

1. Table of acceptance criteria and the reported device performance:

The document doesn't explicitly state "acceptance criteria" with numerical targets for clinical performance (sensitivity, specificity) in the same way it defines LoD for analytical sensitivity. However, based on the presentation and context, the reported sensitivities and specificities from the clinical performance study appear to be the performance demonstrated to justify substantial equivalence.

Since the submission is for a 510(k), the implied "acceptance criteria" for clinical performance is that the device's performance is substantially equivalent to a predicate device. The document states "the Xpert CT/NG test performance is equivalent to the predicate" in the conclusions.

Here's a table summarizing the reported clinical performance:

Specimen TypeTarget OrganismPerformance Metric% (95% CI)
Pharyngeal SwabCTSensitivity95.9% (86.3-98.9)
Pharyngeal SwabCTSpecificity99.7% (99.4-99.8)
Rectal SwabCTSensitivity86.0% (80.9-89.9)
Rectal SwabCTSpecificity99.4% (98.9-99.6)
Pharyngeal SwabNGSensitivity94.7% (90.7-97.0)
Pharyngeal SwabNGSpecificity98.8% (98.3-99.2)
Rectal SwabNGSensitivity91.2% (86.6-94.4)
Rectal SwabNGSpecificity99.6% (99.3-99.8)

Analytical Sensitivity (LoD) - (Example of explicit acceptance criteria and performance)

  • Acceptance Criteria for LoD: The lowest concentration at which 95% of at least 20 replicates are positive.
  • Reported Device Performance (LoD - Pharyngeal Swab Matrix):
    • CT ATCC vr885 serovar D: 161 EB/mL
    • CT ATCC vr879 serovar H: 225 EB/mL
    • NG ATCC 19424: 7.1 CFU/mL
    • NG ATCC 49226: 6.4 CFU/mL
  • Reported Device Performance (LoD - Rectal Swab Matrix):
    • CT ATCC vr885 serovar D: 88 EB/mL
    • CT ATCC vr879 serovar H: 161 EB/mL
    • NG ATCC 19424: 4.9 CFU/mL
    • NG ATCC 49226: 5.3 CFU/mL

2. Sample size used for the test set and the data provenance:

  • Sample Size for Test Set:
    • Pharyngeal Swabs: 2577 specimens (eligible for inclusion in data analyses)
    • Rectal Swabs: 2538 specimens (eligible for inclusion in data analyses)
  • Data Provenance:
    • Country of Origin: United States ("multi-site prospective investigational study at 9 US institutions").
    • Retrospective or Prospective: Prospective ("multi-site prospective investigational study").

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

The document does not specify the number of experts or their qualifications for establishing the ground truth. It states that the ground truth was established using an "anatomic site infected status (ASIS) algorithm based on combined results from two NAAT tests, with a tiebreaker NAAT test if applicable." This suggests a reliance on laboratory test results rather than human expert interpretation of raw data.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

The adjudication method used for establishing the ASIS (ground truth) was a form of 2+1 rule:

  • The anatomic site was considered infected if both primary reference NAAT test results were positive.
  • The anatomic site was considered not infected if both primary reference NAAT test results were negative.
  • If there was discordance between the two primary reference tests, an additional (tiebreaker) NAAT was performed. In this case, agreement of 2 out of 3 (two primary reference tests + tiebreaker) determined the ASIS result.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This study is an analytical and clinical performance evaluation of an in vitro diagnostic (IVD) nucleic acid amplification test (NAAT), not an AI-assisted diagnostic imaging or interpretation device that would involve human readers.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

Yes, a standalone performance study was done. The study directly compares the results of the Xpert CT/NG test (algorithm only) to the established ASIS ground truth for each specimen. There is no human-in-the-loop component mentioned for the interpretation of the Xpert CT/NG results themselves.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

The ground truth used was Anatomic Site Infected Status (ASIS), which was determined by a composite reference method involving multiple nucleic acid amplification tests (NAATs). This is a type of laboratory-based diagnostic ground truth, not pathology, expert consensus on images, or long-term outcomes data.

8. The sample size for the training set:

The document does not provide information on the sample size for a training set. This is typical for traditional in vitro diagnostic (IVD) submissions like this one, as the assay development and validation generally don't utilize "training sets" in the same way machine learning algorithms do. The studies described are for validation of the final device.

9. How the ground truth for the training set was established:

Not applicable, as a distinct training set and its ground truth establishment are not discussed in this document, which focuses on the validation of the Xpert CT/NG assay.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the symbol of the Department of Health & Human Services on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out next to it.

May 23, 2019

Cepheid Sudhakar Marla Senior Director, Regulatory Affairs 904 Caribbean Drive Sunnyvale, California 94089

Re: K190441

Trade/Device Name: Xpert CT/NG, GeneXpert Dx System, GeneXpert Infinity-48s and GeneXpert Infinity-80 Systems, GeneXpert Infinity-48 System, Xpert Vaginal/Endocervical Specimen Collection, Xpert Urine Specimen Collection Kit, Xpert Swab Specimen Collection Kit Regulation Number: 21 CFR 866.3393 Regulation Name: Device to detect nucleic acids from non-viral microorganism(s) causing sexually transmitted infections and associated resistance marker(s) Regulatory Class: Class II Product Code: QEP, MKZ, LSL, OOI, QBD Dated: February 22, 2019 Received: February 25, 2019

Dear Sudhakar Marla:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for

Uwe Scherf, Ph.D. Director Division of Microbiology Devices OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

510(k) Summary

As required by 21 CFR Section 807.92(c).

Submitted by:Cepheid904 Caribbean DriveSunnyvale, CA 90489Phone number: (408) 548-8946Fax number: (408) 541-4192
Contact:Sudhakar Marla, Ph.D.
Date Submitted:May 20, 2019
Proprietary/Trade name:Xpert® CT/NG
Common name:Xpert CT/NG
Type of Test:Automated, multiplex real-time polymerase chain reaction(PCR) assay intended for the in vitro qualitative detection anddifferentiation of DNA from Chlamydia trachomatis (CT)and/or Neisseria gonorrhoeae (NG).
Regulation NumberClassification Name:21 CFR 866.3393 Device to detect nucleic acids from non-viralmicroorganism(s) causing sexually transmitted infections andassociated resistance marker(s)
Primary Product code:QEP; Class II
Secondary Product code:LSL, MKZ, OOI, QBD
Classification AdvisoryPanelMicrobiology (83)
Prescription UseYes
Predicate Device -Assay:Aptima Mycoplasma genitalium Assay[De Novo #DEN180047]
Predicate Devices -Ancillary SpecimenCollection Kits:Xpert Vaginal/Endocervical Specimen Collection Kit[510(k) #K173840]BD MAX UVE Specimen Collection Kit[510(k) #K043144]

{3}------------------------------------------------

Device Description:

The Xpert CT/NG test is an automated in vitro diagnostic test for qualitative detection and differentiation of DNA from Chlamydia trachomatis (CT) and/or Neisseria gonorrhoeae (NG). The test is performed on the Cepheid GeneXpert Instrument Systems. The Xpert CT/NG test on the GeneXpert Instrument System automates and integrates sample purification, nucleic acid amplification and detection of the target sequences in simple or complex samples using real-time PCR. The system consists of an instrument, personal computer, and preloaded software for running the tests and viewing the results. The system requires the use of single-use disposable cartridges that hold the PCR reagents and host the PCR process. Because the cartridges are self-contained, crosscontamination between samples is minimized.

The Xpert CT/NG test includes reagents for the detection and differentiation of CT and NG. A Sample Processing Control (SPC), a Sample Adequacy Control (SAC), and a Probe Check Control (PCC) are also included. The SPC is present to control for adequate processing of the target bacteria and to monitor the presence of inhibitors in the PCR reaction. The SAC reagents detect the presence of a single copy human gene and monitor whether the specimen contains human cells. The PCC verifies reagent rehydration, PCR tube filling in the cartridge, probe integrity, and dye stability.

The GeneXpert Instrument Systems, comprised of the GeneXpert Dx Systems, the GeneXpert Infinity-48 System. GeneXpert Infinity-48s. and the GeneXpert Infinity-80 System, have 1 to 80 randomly accessible modules, depending upon the instrument, that are each capable of performing separate sample preparation and real-time PCR tests. Each module contains a syringe drive for dispensing fluids (i.e., the syringe drive activates the plunger that works in concert with the rotary valve in the cartridge to move fluids between chambers), an ultrasonic horn for lysing cells or spores, and a proprietary I-CORE® thermocycler for performing real-time PCR and detection.

The ancillary specimen collection kits for use with the Xpert CT/NG test are the Xpert Vaginal/Endocervical Specimen Collection Kit, Xpert Swab Specimen Collection kit and the Xpert Urine Specimen Collection kit.

Device Intended Use:

The Xpert® CT/NG test, performed on the GeneXpert® Instrument Systems, is a qualitative in vitro real-time PCR test for the automated detection and differentiation of genomic DNA from Chlamvdia trachomatis (CT) and/or Neisseria gonorrhoeae (NG) to aid in the diagnosis of chlamydial and gonorrheal disease in the urogenital tract and extragenital sites (pharynx and rectum). The assay may be used to test the following specimens from asymptomatic and symptomatic individuals: female and male urine,

{4}------------------------------------------------

patient-collected vaginal swabs (collected in a clinical setting), clinician-collected endocervical swabs, and female and male pharyngeal and rectal swabs.

Ancillary Collection Kits:

Xpert Vaginal/Endocervical Specimen Collection Kit

The Cepheid Xpert® Swab Specimen Collection Kit is designed to collect, preserve, and transport Chlamydia trachomatis, Neisseria gonorrhoeae, and Trichomonas vaginalis DNA in endocervical swab specimens (collected by a clinician) and patient-collected vaginal swab specimens (collected in a clinical setting) from symptomatic and asymptomatic women prior to analysis with the Xpert CT/NG Assay and the Xpert TV Assay.

Xpert Urine Specimen Collection Kit

The Cepheid Xpert® Urine Specimen Collection Kit is designed to preserve and transport Chlamydia trachomatis, Neisseria gonorrhoeae, and Trichomonas vaginalis DNA in first-catch female and male urine specimens from symptomatic and asymptomatic individuals prior to analysis with the Xpert CT/NG Assay and the Xpert TV Assay.

Xpert Swab Specimen Collection Kit

The Xpert® Swab Specimen Collection Kit is designed to collect, preserve, and transport endocervical, vaginal, pharyngeal, and rectal swab specimens. This transport system is for use for testing with Xpert tests.

Substantial Equivalence:

The Xpert CT/NG is substantially equivalent to the predicate device, Aptima Mycoplasma genitalium Assay [De Novo #DEN180047: January 23, 2019].

Table 8-1 shows the similarities and differences between Xpert CT/NG and the predicate device.

{5}------------------------------------------------

Similarities
DevicePredicate Device
ItemCepheid Xpert CT/NGAptima Mycoplasmagenitalium Assay
Nucleic AcidExtractionYesYes
Assay ResultsSameQualitative
Indication for UseSameAsymptomatic andsymptomatic patients

Table 8-1: Comparison of Similarities and Differences of Xpert CT/NG with the Predicate Device

Differences
ItemDeviceCepheid Xpert CT/NGPredicate DeviceAptima Mycoplasmagenitalium Assay
Technology/ DetectionMultiplex real-time polymerasechain reaction(PCR)Transcription-mediated nucleicacid amplification andhybridization
Specimen TypeFemale and male urine,endocervical swab, patient-collected vaginal swab(collected in a clinical setting),and female and malepharyngeal and rectal swabsClinician-collected and self-collected vaginal swabs (in aclinical setting), clinician-collected endocervical swabs,female and male urine,clinician-collected maleurethral swabs, and self-collected penile meatal swabs(in a clinical setting)
Intended UseThe Xpert® CT/NG test,performed on the GeneXpert®Instrument Systems, is aqualitative in vitro real-timePCR test for the automateddetection and differentiationof genomic DNA fromChlamydia trachomatis (CT)and/or Neisseria gonorrhoeaeThe Aptima Mycoplasmagenitalium assay is an invitro nucleic acidamplification test (NAAT)for the qualitative detectionof ribosomal RNA (rRNA)from Mycoplasmagenitalium on the fullyautomated Panther system. It
Differences
ItemDeviceCepheid Xpert CT/NGPredicate DeviceAptima Mycoplasmagenitalium Assay
(NG) to aid in the diagnosis ofchlamydial and gonorrhealdisease in the urogenital tractand extragenital sites(pharynx and rectum). Theassay may be used to test thefollowing specimens fromasymptomatic andsymptomatic individuals:female and male urine,patient- collected vaginalswabs (collected in a clinicalsetting), clinician-collectedendocervical swabs, andfemale and male pharyngealand rectal swabs.is intended for use as an aidin the diagnosis of M.genitalium urogenitalinfections in male andfemale patients suspected ofM. genitalium infection.The assay may be used totest the following specimens:clinician-collected and self-collected vaginal swabs (in aclinical setting), clinician-collected endocervicalswabs, female and maleurine, clinician-collectedmale urethral swabs, andself-collected penile meatalswabs (in a clinical setting).For females, a vaginal swabis the preferred specimentype due to higher clinicalsensitivity for detecting M.genitalium than otherspecimen types; however,female urine or clinician-collected endocervical swabsmay be used as alternativespecimens when vaginalswab specimens are notavailable. If female urine orclinician-collectedendocervical swabspecimens test negative,testing with a vaginal swabmay be indicated, if M.genitalium infection issuspected.
Differences
ItemDeviceCepheid Xpert CT/NGPredicate DeviceAptima Mycoplasmagenitalium Assay
Assay TargetsDNA from Chlamydiatrachomatis (CT) and/orNeisseria gonorrhoeae(NG)Ribosomal RNA fromMycoplasma genitalium
Instrument SystemCepheid GeneXpert InstrumentSystemPanther System
Assay ControlsInternal sample processingcontrol (SPC), sample adequacycontrol (SAC), and probe checkcontrol (PCC).External controls available.Internal Control (IC). Externalcalibrators available.

{6}------------------------------------------------

{7}------------------------------------------------

The Xpert Swab Specimen Collection Kit is substantially equivalent to the following predicate devices:

  • . Cepheid Xpert Vaginal/Endocervical Specimen Collection Kit [510(k) #K173840]
  • Becton Dickinson BD MAX UVE Specimen Collection Kit [510(k) . #K043144]

Similarities and differences between the Cepheid Xpert Swab Specimen Collection Kit and the predicate collection devices are shown in Table 8-2.

{8}------------------------------------------------

ItemDevicePredicate Devices
Cepheid Xpert SwabSpecimen Collection KitCepheid XpertVaginal/EndocervicalSpecimen Collection KitBD MAX UVE SpecimenCollection Kit
510(k)NumberK190441K173840K151589
Intended UseThe Xpert® Swab SpecimenCollection Kit is designed tocollect, preserve, andtransport endocervical,vaginal, pharyngeal, andrectal swab specimens. Thistransport system is for usefor testing with Xpert tests.The Cepheid® Xpert®Vaginal/EndocervicalSpecimen Collection Kit isdesigned to collect, preserve,and transport Chlamydiatrachomatis, Neisseriagonorrhoeae, andTrichomonas vaginalis DNAin endocervical swabspecimens (collected by aclinician) and patient-collected vaginal swabspecimens (collected in aclinical setting) fromsymptomatic andasymptomatic women prior toanalysis with the XpertCT/NG Assay and the XpertTV Assay.The BD MAX UVESpecimen Collection Kit isintended to be used in clinicalsettings according to theinstructions provided forcollection and transport ofvaginal and endocervicalswab specimens, in additionto the preservation of urinefrom males and females.This transport system is foruse for testing with the BDMAX products.
DescriptionContains an individuallypackaged sterile largecleaning swab (forendocervical samples) and apackage containing anindividually packaged sterilecollection swab (for vaginaland endocervical sampling)and an Xpert Swab TransportReagent tube. The collectionswab is placed into theTransport Reagent Tube afterswab sampling to stabilizethe nucleic acid until samplepreparation.Contains an individuallypackaged sterile largecleaning swab (forendocervical samples) and apackage containing anindividually packaged sterilecollection swab (for vaginaland endocervical sampling)and an Xpert Swab TransportReagent tube. The collectionswab is placed into theTransport Reagent Tube afterswab sampling to stabilize thenucleic acid until samplepreparation.Contains a BD MAXSpecimen Collection Swab, agraduated transfer pipette andaBD MAX UVE SampleBuffer Tube. The BD MAXSpecimen Collection Swab isa sterile polyester swab.The swab is placed into theBD MAX UVE SampleBuffer Tube. For urinespecimens, a sample of urineis transferred directly into theBD MAX UVE SampleBuffer Tube using thegraduated transfer pipette.

Table 8-2: Comparison of Similarities and Differences of the Xpert Swab Specimen Collection Kit with the Predicate Device Collection Kits

{9}------------------------------------------------

Non-Clinical Studies:

Analytical Sensitivity

Pooled Pharyngeal Swab Matrix

Purified and titered elementary bodies from two CT serovars. ATCC vr885 serovar D and ATCC vr879 serovar H. were each tested in a sample matrix of negative pooled clinical pharyngeal swab matrix. Replicates of 20 were evaluated at five concentrations for CT serovar D and for CT serovar H and LoDs were estimated by probit analysis. The claimed LoDs were confirmed by analyzing at least 20 replicate samples with elementary bodies diluted to the estimated LoD concentrations. For this study, the claimed LoD is defined as the lowest concentration at which 95% of at least 20 replicates are positive.

The claimed LoD for purified CT serovar D elementary bodies (EB) in pharvngeal swab matrix is 161 EB/mL (Table 8-3). The claimed LoD for purified CT serovar H elementary bodies in pharyngeal swab matrix is 225 EB/mL (Table 8-3).

Table 8-3: LoD of Two CT Serovars in Pooled Pharyngeal Swab Matrix

OrganismLoD
CT ATCC vr885 serovar D (EB/mL)161
CT ATCC vr879 serovar H (EB/mL)225

Two NG strains (ATCC 19424 and ATCC 49226) were tested. Replicates of 20 were evaluated at five concentrations. The LoD was estimated by probit analysis.

The LoD for NG, estimated by probit analysis, is 6.4 – 7.1 CFU/mL in a pooled pharyngeal swab matrix (Table 8-4).

Table 8-4: LoD of Two NG Strains in Pooled Pharyngeal Swab Matrix

OrganismLoD
NG ATCC 19424 (CFU/mL)7.1
NG ATCC 49226 (CFU/mL)6.4

{10}------------------------------------------------

Pooled Rectal Swab Matrix

Purified and titered elementary bodies from two CT serovars, ATCC vr885 serovar D and ATCC vr879 serovar H, were each tested in a sample matrix of negative pooled clinical rectal swab matrix. Replicates of 20 were evaluated at five concentrations for CT serovar D and for CT serovar H and LoDs were estimated by probit analysis. The claimed LoDs were confirmed by analyzing at least 20 replicate samples with elementary bodies diluted to the estimated LoD concentrations. For this study, the claimed LoD is defined as the lowest concentration at which 95% of at least 20 replicates are positive.

The claimed LoD for purified CT serovar D elementary bodies (EB) in rectal swab matrix is 88 EB/mL (Table 8-5). The claimed LoD for purified CT serovar H elementary bodies in rectal swab matrix is 161 EB/mL (Table 8-5).

Table 8-5: LoD of Two CT Serovars in Pooled Rectal Swab Matrix

OrganismLoD
CT ATCC vr885 serovar D (EB/mL)88
CT ATCC vr879 serovar H (EB/mL)161

Two NG strains (ATCC 19424 and ATCC 49226) were tested. Replicates of 20 were evaluated at five concentrations. The LoD was estimated by probit analysis.

The LoD for NG, estimated by probit analysis, is 4.9 - 5.3 CFU/mL in a pooled rectal swab matrix (Table 8-6).

OrganismLoD
NG ATCC 19424 (CFU/mL)4.9
NG ATCC 49226 (CFU/mL)5.3

Table 8-6: LoD of Two NG Strains in Pooled Rectal Swab Matrix

Analytical Reactivity (Inclusivity)

Fourteen CT serovars and twenty NG strains were tested in this study. Testing was performed using CT and NG cultures that were diluted in pooled clinical pharyngeal and pooled clinical rectal swab matrices at levels near the analytical LoD. Three replicates were tested for each strain. Results are shown in Table 8-8 for CT serovars and NG strains, respectively. All 14 CT serovars and all 20 NG strains were correctly reported using the Xpert CT/NG test.

{11}------------------------------------------------

C. trachomatisConcentrationTested in PharyngealSwab MatrixConcentrationTested in RectalSwab MatrixAssay Result
serovarCTNG
A1800 EB/mL1800 EB/mLPOSNEG
B9 EB/mL8.1 EB/mLPOSNEG
Ba0.9 EB/mL0.81 EB/mLPOSNEG
C900 EB/mL322 EB/mLPOSNEG
E450 EB/mL322 EB/mLPOSNEG
E/SW20.9 IFU/mLª0.81 IFU/mLªPOSNEG
F450 EB/mL322 EB/mLPOSNEG
G900 EB/mL644 EB/mLPOSNEG
I0.18 EB/mL0.16 EB/mLPOSNEG
J900 EB/mL644 EB/mLPOSNEG
K900 EB/mL644 EB/mLPOSNEG
LGV I450 EB/mL322 EB/mLPOSNEG
LGV II450 EB/mL322 EB/mLPOSNEG
LGV III450 EB/mL644 EB/mLPOSNEG

Table 8-7: Analytical Reactivity Results of the Xpert CT/NG Assay with CT Serovars in Pooled Pharyngeal and Rectal Swab Matrices

a. IFU/mL = Infectious units per mL.

Table 8-8: Analytical Reactivity Results of the Xpert CT/NG Assav with
NG Strains in Pharyngeal and Rectal Swab Matrices
N. gonorrhoeaeStrainConcentrationTested inPharyngeal SwabMatrix (CFU/mL)ConcentrationTested in RectalSwab Matrix(CFU/mL)Assay Result
979314.210.6CTNG
983014.210.6NEGPOS
1999914.210.6NEGPOS
2762914.210.6NEGPOS
2763014.210.6NEGPOS
2763114.210.6NEGPOS
3114814.210.6NEGPOS
3139714.210.6NEGPOS
3139914.210.6NEGPOS

{12}------------------------------------------------

N. gonorrhoeaeStrainConcentrationTested inPharyngeal SwabMatrix (CFU/mL)ConcentrationTested in RectalSwab Matrix(CFU/mL)Assay Result
CTNG
3140014.210.6NEGPOS
117014.242.4NEGPOS
639514.210.6NEGPOS
1328114.210.6NEGPOS
3444714.210.6NEGPOS
3754114.210.6NEGPOS
1022614.210.6NEGPOS
1022714.210.6NEGPOS
1093214.210.6NEGPOS
1147214.210.6NEGPOS
5034814.210.6NEGPOS

Analytical Specificity (Cross-reactivity and Competitive Interference)

Forty-one microorganisms potentially present in pharyngeal flora (Table 8-9) and fortythree microorganisms potentially present in rectal flora (see Table 8-10) were tested using Xpert CT/NG. The microorganisms were tested in the presence (competitive interference) and absence (cross-reactivity) of 2X LoD CT (Serovar D) and NG (ATCC 49226) organisms and were diluted into pooled clinical negative pharyngeal swab matrix or pooled clinical negative rectal swab matrix for testing. Bacterial strains were tested in triplicate at a concentration of at least 10° CFU/mL except for Treponema denticola, which was tested at 1.92 x 106 genome equivalents/mL. Parasites were tested at 1 x 100 cells/mL except for Entamoeba histolytica, which was tested at 1 x 105 CFU/mL and viruses were tested at 1 x 105 TCID50/ml or 1 x 105 IFU/mL. Positive and negative controls were included in the study. All CT and NG positive samples remained positive and all CT and NG negative samples remained negative, indicating that there was no interference or cross-reactivity with the results of the Xpert CT/NG test for these microorganisms.

{13}------------------------------------------------

Pooled Pharyngeal Swab Matrix
Actinobacillus actinomycetemcomitansKlebsiella pneumoniaeStaphylococcus aureus
AdenovirusLactobacillusacidophilusStaphylococcusepidermidis
Arcanobacterium haemolyticumLactobacillus lactisStreptococcus anginosus
Bordetella pertussisMoraxella catarrhalisStreptococcusdysgalactiae
Campylobacter rectusMycoplasma pneumoniaeStreptococcus mitis
Candida albicansNeisseria flavescensStreptococcus mutans
CoronavirusPeptostreptococcusmicrosStreptococcuspneumoniae
Corynebacterium diphtheriaePorphyromonasgingivalisStreptococcus pyogenes
Fusobacterium necrophorumPrevotella biviaStreptococcus salivarius
Haemophilus influenzaePrevotella oralisaStreptococcus sanguinis
Herpes virusPseudomonasaeruginosaTannerella forsythiab
Human influenza virus ARespiratory syncytialvirusTreponema denticolac
Human influenza virus BRhinovirusVeillonella parvula
Human metapneumovirusSaccharomycescerevisiae

Table 8-9: Potential Cross-reacting/Competitive Interfering Microorganisms in Pooled Pharyngeal Swab Matrix

a. Bacteroides oralis is Prevotella oralis.

b. Bacteriodes forsythus is Tannerella forsythia.

c. Genomic DNA tested.

{14}------------------------------------------------

Acinetobacter baumanniiFusobacteriumnecrophorumProvidencia stuartii
Anaerococcus tetradiusFusobacterium nucleatumPseudomonas aeruginosa
Anaerococcus hydrogenalisGiardia lambliaSalmonella enterica sb enterica svminnesota
Bacteroides fragilisHelicobacter pyloriSalmonella enterica sb enterica svtyphimurium
Bifidobacterium adolescentKlebsiella oxytocaShigella flexneri
Campylobacter jejuniLactobacillus acidophilusShigella sonnei
Candida albicansLactobacillus delbreueckiiStaphylococcus aureus
Citrobacter freundiiListeria monocytogenesStaphylococcus epidermidis
Clostridium difficileMorganella morganiiStreptococcus agalactiae
Entamoeba histolyticaNorovirusStreptococcus dysgalactiae
Enterobacter cloacaePeptostreptococcusanaerobiusVibrio cholerae
Enterococcus faecalisPlesiomonas shigelloidesVibrio parahaemolyticus
Enterococcus faeciumPrevotella biviaYersinia enterocolitica
EnterovirusPrevotella oralis
Escherichia coliProteus mirabilis

Table 8-10: Potential Cross-reacting/Competitive Interfering Microorganisms in Pooled Rectal Swab Matrix

Interfering Substances

Potentially interfering exogenous substances were diluted in pooled clinical pharyngeal swab and pooled clinical rectal swab matrices containing two different mixtures of CT and NG cells. The first mixture contained 3x LoD CT serovar D and NG strain ATCC 49226. The second mixture contained 3x LoD CT serovar H and NG strain ATCC 19424.

There was no assay interference in the presence of the substances at the concentrations tested for pharyngeal swab matrix (Table 8-11) and rectal swab matrix (Table 8-12).

{15}------------------------------------------------

Potentially Interfering Substances to be EvaluatedConcentrationTested
Mucin (pig gastric mucin)25 mg/mL
Whole human blood5% v/v
Mouthwash (Cool Mint Listerine, antiseptic)5% v/v
Cough Medicine Guaifenesin (Guaiacol glyceryl)5 mg/mL
Cough Medicine Dextromethorphan HBr100 µg/mL
Antibiotic (Penicillin G)1.2 mg/mL
Antibiotic (Erythromycin)15 µg/mL
Sugar-containing cold and flu remedies (Acetaminophen)5% v/v
Chloraseptic5% v/v
Salt-modifying remedy (sodium chloride)50% v/v
Foods/drinks that increase salivary viscosity (milk)5% v/v
pH Modifying Remedy (orange juice)5% v/v
Cold Sore medication Abreva5% v/v
Potentially Interfering Substances to be EvaluatedConcentrationTested
Barium sulfate0.25% w/v
Ciprofloxacin0.25% w/v
Condom1 condom (#)
Cortizone0.25% w/v
ExLax0.25% w/v
Fecal fat (Stearic acid/Palmitic acid/Cholesterol)0.25% w/v
Imodium0.25% w/v
K-Y Jelly0.25% w/v
Milk of Magnesia0.25% w/v
Mineral Oil0.25% w/v
Neosporin(Polymixin B/ Neomycin/Bacitracin)0.25% w/v
Nystatin0.25% w/v
Pepcid0.25% w/v
Pepto-Bismol0.25% w/v
Preparation H0.25% w/v
Prilosec0.25% w/v
Saline0.25% w/v
Tagamet0.25% w/v
Vagisil0.25% w/v

Table 8-11: Potentially Interfering Substances Tested in Pooled Pharyngeal Swab Matrix

{16}------------------------------------------------

Table 8-12: Potentially Interfering Substances Tested in Pooled Rectal Swab Matrix

Carry-Over Contamination Study

Please refer to the previously FDA-cleared 510(k) #K121710 for additional information.

{17}------------------------------------------------

Linearity

Please refer to the previously FDA-cleared 510(k) #K121710 for additional information.

Clinical Performance Characteristics:

Reproducibility and Precision

Please refer to the previously FDA-cleared 510(k) #K121710 for additional information.

Clinical Performance Study

Performance characteristics of Xpert CT/NG were determined in a multi-site prospective investigational study at 9 US institutions by comparing Xpert CT/NG to the anatomic site infected status (ASIS) algorithm based on combined results from two NAAT tests, with a tiebreaker NAAT test if applicable.

The anatomic site was considered to be infected if both of the reference test results were positive. The anatomic site is considered to be not infected when both reference test results were negative. If there was discordance between the reference tests, an additional NAAT was tested as a tiebreaker. In this case, agreement of 2/3 of the reference NAATs determined the ASIS result. If two tests were equivocal or one equivocal and one not run, the third test result stood as the ASIS if positive or negative. If two tests were not run, the ASIS was considered invalid and excluded from the analysis. The tiebreaker test was run by the lab if any NAAT was not concordant with the others and interpreted only in the case of discordant results between the two planned reference tests for each assay. As the tiebreaker test was not a combination test, the tiebreaker was only run for the organism with disagreement (e.g., if NG disagrees and CT agrees, the tiebreaker was only run for NG).

Study participants included consenting adults seeking sexually transmitted disease (STD) testing at the participating clinics, which included clinics focused on sexually transmitted diseases, women's health, student health and family planning , as well as clinics specializing in lesbian, gav, bisexual, and transgender (LGBT) health. Potential subjects were identified, assessed for eligibility and approached for informed consent. Both symptomatic and asymptomatic individuals were included in the study population. The study specimens consisted of prospectively collected rectal and pharyngeal swabs. Performance of Xpert CT/NG was calculated relative to the ASIS for each of the two sample types.

A total of 2767 study participants were enrolled in the study, of which 2577 pharvngeal swab and 2538 rectal swab specimens were eligible for inclusion in the data analyses. One hundred and ninety (190) pharyngeal specimens were excluded from the data analyses due to the following reasons: 167 for temperature excursions during shipment, 4 participants withdrew consent, 2 speciment errors, 2 post-swab collection

{18}------------------------------------------------

errors, 1 specimen not collected, 1 participant receiving antibiotics, and 13 specimens with Xpert results not available or non-determinate. Two hundred and twenty-nine (229) rectal specimens were excluded from the data analyses due to the following reasons: 167 for temperature excursions during shipment, 6 participants withdrew consent, 5 specimens shipment errors. 2 post-swab collection errors, 1 specimen not collected, 1 participant receiving antibiotics, and 46 specimens with Xpert results not available or non-determinate.

Among the study participants included in the data analyses for pharyngeal swab performance 20.8% were female at birth and 79.2% were male at birth. The average age was 33.8 years (range = 18 to 76 years).

Among the study participants included in the data analyses for rectal swab performance 20.9% were female at birth and 79.1% were male at birth. The average age was 33.7 years (range = 18 to 76 years).

Of the 2572 study participants eligible for inclusion in the pharyngeal and rectal swab analyses for CT detection, 0.9% (22/2572) were positive for CT by pharyngeal swab and rectal swab by ASIS. Of the 2573 study participants eligible for inclusion in the pharyngeal and rectal swab analyses for NG detection. 3.7% (95/2573) were positive for NG by pharyngeal swab and rectal swab.

Among the 5163 tests performed, 198 (3.8%) had to be retested due to ERROR, INVALID or NO RESULT outcomes. Of those. 151 specimens vielded valid results upon repeat assay (2 specimens were not retested). The overall valid reporting rate of the assay was 99.1% (5116/5163).

Chlamydia trachomatis Performance Results - Pharyngeal and Rectal Swabs Results from Xpert CT/NG were compared to the ASIS algorithm for determination of sensitivity and specificity. Results for CT by symptomatic status are shown in Table 8-13.

{19}------------------------------------------------

SpecimenStatusnTPFPTNFNPrev%Sensitivity %(95 CI)Specificity %(95 CI)
PSSym3069029702.9100.0%(70.1-100.0)100.0%(98.7%-100.0)
Asym2269388222121.895.0%(83.5-98.6)99.6%(99.3-99.8)
All2575478251821.995.9%(86.3-98.9)99.7%(99.4-99.8)
RSSym188221160514.481.5%(63.3-91.8)99.4%(96.6-99.9)
Asym2347175142131278.686.6%(81.3-90.7)99.4%(98.9-99.6)
All2535197152291329.086.0%(80.9-89.9)99.4%(98.9-99.6)

Table 8-13: Xpert CT/NG vs. ASIS for CT Detection by Symptomatic Status -Pharyngeal Swabs and Rectal Swabs

TP=true positive, FP=false positive, TN=true negative, PN=false negative, PS=pharyngeal swab

Neisseria gonorrhoeae Performance Results – Pharyngeal and Rectal Swabs

Results from Xpert CT/NG were compared to the ASIS algorithm for determination of sensitivity and specificity. Results for NG by symptomatic status are shown in Table 8-14.

Table 8-14: Xpert CT/NG vs. ASIS for NG Detection by Symptomatic Status -Pharyngeal Swabs and Rectal Swabs

SpecimenStatusnTPFPTNFNPrev %Sensitivity %(95 CI)Specificity %(95 CI)
PSSym306393261313.792.9%(81.0-97.5)98.9%(96.7-99.6)
Asym226915626207987.295.1%(90.7-97.5)98.8%(98.2-99.2)
All2575195292340118.094.7%(90.7-97.0)98.8%(98.3-99.2)
RSSym188380149120.797.4%(86.8-99.6)100.0%(97.5-100.0)
Asym234814992173177.189.8%(84.2-93.5)99.6%(99.2-99.8)
All253618792322188.191.2%(86.6-94.4)99.6%(99.3-99.8)

TP=true positive, FP=false positive, TN=true negative, PN=false negative, PS=pharyngeal swab, RS=rectal swab

{20}------------------------------------------------

Table 8-15 and Table 8-16 show the number of results designated as infected or not infected with CT based on the ASIS algorithm for pharyngeal and rectal specimens, respectively.

ASISaNAAT1NAAT2TiebreakerNAATXpertTotal
NI--NAb-2504
NINRc---6
NI--NA+4
NI-+--5
NI+---2
NI-+-+1
NI+--+1
NIEQd---1
INDe+-EQ+1
INDNR-++1
Total Not Infected2526
I++NA+40
I-+++5
I+-++2
I++NA-1
I-++-1
Total Infected49

Table 8-15: Anatomic Site Infected Status – Pharyngeal CT

ASIS = Anatomic Site Infected Status: NI=Not Infected; IND=indeterminate, considered not a. infected.

b. NA=Not applicable; both reference NAAT tests agreed.

c. NR = Not run

EQ=Equivocal d.

IND=Indeterminate. Considered if Xpert Negative and not infected if Xpert positive to evaluate e. under worst-case scenario.

{21}------------------------------------------------

ASISaNAAT1NAAT2TiebreakerNAATXpertTotal
NI--NAb-2221
NINRc---47
NI--NA+12
NI+---11
NI-+--10
NI-+-+2
NI-EQd--2
INDe+EQ-+1
Total Not Infected2306
I++NA+172
I-+++14
I-++-11
I++NA-9
I+-++6
I+-+-5
I+EQ++3
I-EQ+-2
INR+++2
I+EQ+-1
I+EQNR-1
INRE+-1
IND-NR+-1
IND+-NR-1
Total Infected229

Table 8-16: Anatomic Site Infected Status – Rectal CT

ASIS = Anatomic Site Infected Status: NI=Not Infected; IND=indeterminate, considered not a. infected.

b. NA=Not applicable; both reference NAAT tests agreed.

C. NR = Not run

d. EQ=Equivocal

IND=Indeterminate. Considered if Xpert negative and not infected if Xpert positive to evaluate under e. worst-case scenario.

{22}------------------------------------------------

Table 8-17 and Table 8-18 show the number of results designated as infected or not infected with NG based on the ASIS algorithm for pharyngeal and rectal specimens, respectively.

ASISaNAAT1NAAT2TiebreakerNAATXpertTotal
NI--NAb-2317
NI--NA+19
NI-+--14
NI-+-+4
NI+---4
NI+--+4
NINRc---5
NI-EQd-+1
INDe-+EQ+1
Total Not Infected2369
I++NA+175
I++NA-4
I-+++16
I-++-5
I+-++2
INR+++2
IND+EQ--1
IND-EQ+-1
Total Infected206

Table 8-17: Anatomic Site Infected Status – Pharyngeal NG

-ASIS = Anatomic Site Infected Status: NI=Not Infected; IND=indeterminate, considered not a. infected.

b. NA=Not applicable; both reference NAAT tests agreed.

NR = Not run C.

EQ=Equivocal d.

IND=Indeterminate. Considered if Xpert negative and not infected if Xpert positive to evaluate under e. worst-case scenario.

{23}------------------------------------------------

ASISaNAAT1NAAT2TiebreakerNAATXpertTotal
NI--NAb-2261
NINRc---49
NI--NA+6
NI+---5
NI-+--4
NI+--+2
NI-EQd--2
NI-NR--1
INDe+EQ-+1
Total Not Infected2331
I++NA+172
I-+++13
I++NA-8
I-++-8
I+-++1
I+EQ++1
INR++-1
IND-EQ+-1
Total Infected205

Table 8-18: Anatomic Site Infected Status - Rectal NG

ASIS = Anatomic Site Infected Status: NI=Not Infected; IND=indeterminate, considered not a. infected.

b. NA=Not applicable; both reference NAAT tests agreed.

с. NR = Not run

EQ=Equivocal d.

e. IND=Indeterminate. Considered if Xpert negative and not infected if Xpert positive to evaluate under worst-case scenario.

Conclusions

The results of the nonclinical analytical and clinical performance studies summarized above demonstrate that the Xpert CT/NG test performance is equivalent to the predicate.

§ 866.3393 Device to detect nucleic acids from non-viral microorganism(s) causing sexually transmitted infections and associated resistance marker(s).

(a)
Identification. A device to detect nucleic acids from non-viral microorganism(s) causing sexually transmitted infections and associated resistance marker(s) is an in vitro diagnostic device intended for the detection and identification of nucleic acids from non-viral microorganism(s) and their associated resistance markers in clinical specimens collected from patients suspected of sexually transmitted infections. The device is intended to aid in the diagnosis of non-viral sexually transmitted infections in conjunction with other clinical and laboratory data. These devices do not provide confirmation of antibiotic susceptibility since mechanisms of resistance may exist that are not detected by the device.(b)
Classification. Class II (special controls). The special controls for this device are:(1) The intended use for the labeling required under § 809.10 of this chapter must include a detailed description of targets the device detects, the results provided to the user, the clinical indications appropriate for test use, and the specific population(s) for which the device is intended.
(2) Any sample collection device used must be FDA-cleared, -approved, or -classified as 510(k) exempt (standalone or as part of a test system) for the collection of specimen types claimed by this device; alternatively, the sample collection device must be cleared in a premarket submission as a part of this device.
(3) The labeling required under § 809.10(b) of this chapter must include:
(i) A detailed device description, including reagents, instruments, ancillary materials, all control elements, and a detailed explanation of the methodology, including all pre-analytical methods for processing of specimens;
(ii) Detailed discussion of the performance characteristics of the device for all claimed specimen types based on analytical studies, including Limit of Detection, inclusivity, cross-reactivity, interfering substances, competitive inhibition, carryover/cross contamination, specimen stability, within lab precision, and reproducibility, as appropriate;
(iii) Detailed descriptions of the test procedure, the interpretation of test results for clinical specimens, and acceptance criteria for any quality control testing;
(iv) Limiting statements indicating that:
(A) A negative test result does not preclude the possibility of infection;
(B) The test results should be interpreted in conjunction with other clinical and laboratory data available to the clinician;
(C) Reliable results are dependent on adequate specimen collection, transport, storage, and processing. Failure to observe proper procedures in any one of these steps can lead to incorrect results; and
(D) If appropriate (
e.g., recommended by the Centers for Disease Control and Prevention, by current well-accepted clinical guidelines, or by published peer reviewed research), that the clinical performance is inferior in a specific clinical subpopulation or for a specific claimed specimen type; and(v) If the device is intended to detect antimicrobial resistance markers, limiting statements, as appropriate, indicating that:
(A) Negative results for claimed resistance markers do not indicate susceptibility of detected microorganisms, as resistance markers not measured by the assay or other potential mechanisms of antibiotic resistance may be present;
(B) Detection of resistance markers cannot be definitively linked to specific microorganisms and the source of a detected resistance marker may be an organism not detected by the assay, including colonizing flora;
(C) Detection of antibiotic resistance markers may not correlate with phenotypic gene expression; and
(D) Therapeutic failure or success cannot be determined based on the assay results, since nucleic acid may persist following appropriate antimicrobial therapy.
(4) Design verification and validation must include:
(i) Detailed device description documentation, including methodology from obtaining sample to result, design of primer/probe sequences, rationale for target sequence selection, and computational path from collected raw data to reported result (
e.g., how collected raw signals are converted into a reported result).(ii) Detailed documentation of analytical studies, including, Limit of Detection, inclusivity, cross-reactivity, microbial interference, interfering substances, competitive inhibition, carryover/cross contamination, specimen stability, within lab precision, and reproducibility, as appropriate.
(iii) Detailed documentation and performance results from a clinical study that includes prospective (sequential) samples for each claimed specimen type and, when determined to be appropriate by FDA, additional characterized clinical samples. The study must be performed on a study population consistent with the intended use population and compare the device performance to results obtained from FDA accepted comparator methods. Documentation from the clinical studies must include the clinical study protocol (including a predefined statistical analysis plan) study report, testing results, and results of all statistical analyses.
(iv) A detailed description of the impact of any software, including software applications and hardware-based devices that incorporate software, on the device's functions.