K Number
K190397

Validate with FDA (Live)

Date Cleared
2019-11-15

(269 days)

Product Code
Regulation Number
862.3590
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Immunalysis Carisoprodol Metabolite / Meprobamate Urine HEIA is a homogeneous enzyme immunoassay for the qualitative analysis of carisoprodol metabolite, Meprobamate, at a cutoff of 280 ng/mL in human urine. The assay is intended for use in laboratories with automated clinical chemistry analyzers. This in vitro diagnostic device is for prescription use only.

The Immunalysis Carisoprodol Metabolite / Meprobamate Urine HEIA provides only a preliminary analytical test result. A more specific alternate chemical must be used in order to obtain a confirmed analytical result. Gas Chromatography/ Mass Spectrometry (GC-MS) or Liquid Chromatography / Tandem Mass Spectrometry (LC-MS/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used.

Device Description

The Immunalysis Carisoprodol Metabolite / Meprobamate Urine HEIA is a sensitive in vitro diagnostic test intended for use in laboratories for the qualitative analysis of Meprobamate at a cutoff of 280 ng/mL in human urine with automated clinical chemistry analyzers.

Carisoprodol (N-isopropylmeprobamate, Soma, ingredient of Soma Compound, Somadril®) is a carbamate derivative first synthesized in 1959 and used clinically as a muscle relaxant and sedative. Carisoprodol is known to be metabolized to meprobamate and hydroxyl-meprobamate.

The assay is based on the competition of carisoprodol labeled enzyme glucose-6-phosphate dehydrogenase (G6PDH) and the free Meprobamate in the urine sample for the fixed amount of sheep anti-carisoprodol antibody binding sites. In the absence of the free Meprobamate in the sample, the antibody binds the drug enzyme conjugate and enzyme activity is inhibited. This creates a dose response relationship between drug concentration in the urine and enzyme activity. The enzyme G6PDH activity is determined at 340 nm spectrophotometrically by the conversion of nicotinamide adenine dinucleotide (NAD) to NADH.

AI/ML Overview

The provided document describes the Immunalysis Carisoprodol Metabolite / Meprobamate Urine HEIA, an in vitro diagnostic device for the qualitative analysis of carisoprodol metabolite, Meprobamate, in human urine. The acceptance criteria and the study proving the device meets these criteria are detailed in the "Performance Characteristics" section (G).

Here's a breakdown of the requested information:

1. A table of acceptance criteria and the reported device performance

Based on the studies described, the acceptance criteria can be inferred from the reported performance. The document doesn't explicitly state "acceptance criteria" values in a separate table, but rather presents the study results which are implicitly considered acceptable for substantial equivalence.

Performance CharacteristicAcceptance Criteria (Inferred from Study Design & Results)Reported Device Performance
Precision/Cutoff CharacterizationAll samples at -100%, -75%, -50%, +25%, +50%, +75%, +100% of cutoff should yield 100% agreement (Negative or Positive). At cutoff, approx. 50% Negative/50% Positive.Lot #1:-100% to -25%: 80/80 NegativeCutoff (280 ng/mL): 40 Negative/40 Positive+25% to +100%: 80/80 PositiveLot #2:-100% to -25%: 80/80 NegativeCutoff (280 ng/mL): 38 Negative/42 Positive+25% to +100%: 80/80 PositiveLot #3:-100% to -25%: 80/80 NegativeCutoff (280 ng/mL): 39 Negative/41 Positive+25% to +100%: 80/80 Positive
Specificity and Cross-ReactivityCompounds not structurally similar to meprobamate should show <0.1% cross-reactivity and result in NEG. Meprobamate and Carisoprodol (metabolized to meprobamate) should show appropriate POS results/cross-reactivity.Meprobamate: 100% Cross-Reactivity (POS)Carisoprodol: 280% Cross-Reactivity (POS)Felbamate: 0.2% Cross-Reactivity (POS)Most other tested compounds: <0.1% Cross-Reactivity (NEG) or N/D.
Interference (Structurally Unrelated)Tested compounds should not interfere with assay performance at ±25% of cutoff.All 80+ listed compounds (e.g., Acetaminophen, Alprazolam, Caffeine, Ibuprofen, Marijuana metabolites, etc.) did not interfere when spiked at high concentrations (e.g., 100,000 ng/mL, 500,000 ng/mL).
Interference (Endogenous & Preservatives)Tested compounds and preservatives (except as noted) should not interfere.All listed compounds (e.g., Acetone, Ascorbic Acid, Creatinine, Ethanol, Glucose, Hemoglobin, Urea) did not interfere.Boric Acid: Did not interfere at 1% w/v when Meprobamate at 140 ng/mL (-50% Cutoff) or 420 ng/mL (+50% Cutoff). (Note: Initial interference at ±25% of cutoff led to testing at ±50%.)
Interference (pH)No positive or negative interference at pH 3.0, 7.0, and 11.0.No positive or negative interference observed at urine pH values 3.0, 7.0, and 11.0.
Interference (Specific Gravity)No positive or negative interference at specific gravity 1.000, 1.015, and 1.030.No positive or negative interference observed at urine specific gravity values 1.000, 1.015, and 1.030.
Calibration DurationMaintain performance for the recommended duration.The test results met acceptance criteria for up to 14 days with a two-point calibration. Recommended frequency of calibration is 14 days.
Specimen Stability (Urine)Urine samples containing meprobamate should be stable for specified durations at various temperatures.Urine samples containing carisoprodol and/or meprobamate are stable for up to 7 days stored at ambient temperature (up to 30°C) and up to 6 months stored at 2°C - 8°C.
Method ComparisonHigh agreement rate with the confirmatory method (LC-MS/MS).Total Samples: 167LC-MS/MS Negative (< 280 ng/mL): 60 samplesLC-MS/MS Positive (>= 280 ng/mL): 107 samplesAgreement Immunalysis HEIA Positive vs. LC-MS/MS Positive: 100% (107/107)Agreement Immunalysis HEIA Negative vs. LC-MS/MS Negative: 92% (55/60)False Positives: 5 samples (all contained carisoprodol and were considered clinically acceptable/not a concern).

2. Sample size used for the test set and the data provenance

  • Precision/Cutoff Characterization: For each of the three product lots, 80 determinations were made at each concentration level across 10 days (2 runs/day, 4 replicates). Total for this study: 3 lots x 9 concentrations x 80 determinations/concentration = 2160 determinations (or 3 lots x 80 samples at each of 9 levels = 2160 total samples, if each replicate is considered a sample). The data provenance is not specified, but it involved "drug free urine" spiked with meprobamate. This suggests laboratory-prepared samples.
  • Specificity and Cross-Reactivity: The sample size for each compound tested is not explicitly stated, but compounds were "spiked into drug free urine."
  • Interference (Structurally Unrelated / Endogenous & Preservatives / pH / Specific Gravity): The sample size for each compound/condition is not explicitly stated, but compounds were "spiked into drug-free negative urine" containing meprobamate at specific concentrations.
  • Urine Elimination Study: 10 subjects were used. The samples were "self-reported single dose users" and collected their own urine. Data provenance is not explicitly stated in terms of country, but implies prospective collection for the study.
  • Method Comparison: 167 de-identified, unaltered leftover clinical urine samples. Data provenance is "obtained from clinical testing laboratories," indicating retrospective use of existing clinical samples. Country of origin is not specified.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

For the quantitative results in the precision and method comparison studies, the ground truth was established by Liquid Chromatography-Tandem Mass Spectrometry (LC-MS/MS), a highly accurate analytical chemistry method. This is a laboratory instrumental method, not dependent on human experts for qualitative interpretation in the way imaging studies might be. While trained laboratory personnel operate and validate LC-MS/MS, the "ground truth" itself is based on the chemical analysis, not expert consensus reading of a visual output.

For the qualitative interpretation of the Immunalysis assay results (Positive/Negative call at cutoff), this is determined automatically by the instrument based on the measured signal relative to the cutoff value, not by human experts adjudicating results.

Therefore, for this type of in vitro diagnostic device, the concept of "experts establishing ground truth for a test set" with qualifications like "radiologist with 10 years of experience" is not directly applicable.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. This is an analytical immunoassay where the result (positive/negative) is determined by the instrument against an established cutoff, or by a confirmatory lab method (LC-MS/MS). There is no human interpretative step requiring adjudication.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is an IVD immunoassay, not an AI-assisted diagnostic imaging device that involves human reader interpretation.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Yes, the device's performance characteristics (precision, specificity, interference, calibration, stability) and the method comparison were evaluated for the "Immunalysis Carisoprodol Metabolite / Meprobamate Urine HEIA" as a standalone analytical test. The device itself is an automated immunoassay intended for use with automated clinical chemistry analyzers. As an IVD, its function is essentially "algorithm only" in the sense that it relies on chemical reactions and optical readings, with the interpretation (positive/negative) being determined by the instrument based on programmed parameters (e.g., cutoff).

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The ground truth for the quantitative concentration of meprobamate/carisoprodol in urine samples was established using Liquid Chromatography-Tandem Mass Spectrometry (LC-MS/MS), which is considered the gold standard confirmatory method for drug testing. This falls under the category of a highly accurate, analytical laboratory test.

8. The sample size for the training set

The document does not describe a separate "training set" for the device development. This is typical for an immunoassay where the assay's chemical formulation, antibody properties, and cutoff are developed based on laboratory analytical studies and calibration, rather than on a large dataset of patient samples in the way a machine learning algorithm is trained. The studies described are performance validation studies.

9. How the ground truth for the training set was established

As there's no explicitly described "training set" in the context of an algorithm or machine learning, the concept of establishing ground truth for such a set is not applicable here. The ground truth for the analytical and validation studies was established through precisely prepared spiked samples (for precision, specificity, interference) and confirmatory LC-MS/MS analysis for clinical samples (method comparison, urine elimination study).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters.

November 15, 2019

Immunalysis Corporation Yang Yang Regulatory Affairs Specialist II 829 Towne Center Drive Pomona, CA 91767

Re: K190397

Trade/Device Name: Immunalysis Carisoprodol Metabolite / Meprobamate Urine HEIA Regulation Number: 21 CFR 862.3590 Regulation Name: Meprobamate Test System Regulatory Class: Class II Product Code: QBK Dated: October 3, 2019 Received: October 4, 2019

Dear Yang Yang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Marianela Perez-Torres, Ph.D. Acting Deputy Director Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K190397

Device Name

Immunalysis Carisoprodol Metabolite / Meprobamate Urine HEIA

Indications for Use (Describe)

The Immunalysis Carisoprodol Metabolite / Meprobamate Urine HEIA is a homogeneous enzyme immunoassay for the qualitative analysis of carisoprodol metabolite, Meprobamate, at a cutoff of 280 ng/mL in human urine. The assay is intended for use in laboratories with automated clinical chemistry analyzers. This in vitro diagnostic device is for prescription use only.

The Immunalysis Carisoprodol Metabolite / Meprobamate Urine HEIA provides only a preliminary analytical test result. A more specific alternate chemical must be used in order to obtain a confirmed analytical result. Gas Chromatography/ Mass Spectrometry (GC-MS) or Liquid Chromatography / Tandem Mass Spectrometry (LC-MS/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used.

Type of Use (Select one or both , as applicable)
----------------------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "IMMUNALYSIS" in white letters on a red background. The word is written in all capital letters and is centered in the image. The red background is a solid color and there are no other objects in the image.

510(k) SUMMARY: K190397

A. GENERAL INFORMATION

Applicant Name:Immunalysis Corporation829 Towne Center DrivePomona, CA 91767Establishment # 2020952
Company Contact:Yang YangRegulatory Affairs Specialist IIPhone: (909) 451-6665Email: yyang@immunalysis.com
Date Prepared:November 13, 2019

B. DEVICE IDENTIFICATION

Trade or Proprietary Names:Immunalysis Carisoprodol Metabolite / Meprobamate UrineHEIA
Common Name:Carisoprodol Metabolite / Meprobamate Urine EnzymeImmunoassay

C. REGULATORY INFORMATION

Device Classification Name:Meprobamate Test System
Product Codes:QBK
Regulatory Class:II
Classification Regulation:21 CFR 862.3590, Meprobamate Test System
Panel:Toxicology (91)
Predicate Device:Lin-Zhi Carisoprodol Metabolite (Meprobamate) EnzymeImmunoassay, Meprobamate Drugs of Abuse Calibrators andControls [DEN170010]

D. DEVICE DESCRIPTION

The Immunalysis Carisoprodol Metabolite / Meprobamate Urine HEIA is a sensitive in vitro diagnostic test intended for use in laboratories for the qualitative analysis of Meprobamate at a cutoff of 280 ng/mL in human urine with automated clinical chemistry analyzers.

Carisoprodol (N-isopropylmeprobamate, Soma, ingredient of Soma Compound, Somadril®) is a carbamate derivative first synthesized in 1959 and used clinically as a muscle relaxant and sedative. Carisoprodol is known to be metabolized to meprobamate and hydroxyl-meprobamate.

The assay is based on the competition of carisoprodol labeled enzyme glucose-6-phosphate

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the word "IMMUNALYSIS" in white letters on a red background. The word is written in all capital letters and is centered in the image. The background is a solid red color. The image is simple and straightforward, with the focus on the word itself.

dehydrogenase (G6PDH) and the free Meprobamate in the urine sample for the fixed amount of sheep anti-carisoprodol antibody binding sites. In the absence of the free Meprobamate in the sample, the antibody binds the drug enzyme conjugate and enzyme activity is inhibited. This creates a dose response relationship between drug concentration in the urine and enzyme activity. The enzyme G6PDH activity is determined at 340 nm spectrophotometrically by the conversion of nicotinamide adenine dinucleotide (NAD) to NADH.

E. INTENDED USE

The Immunalysis Carisoprodol Metabolite / Meprobamate Urine HEIA is a homogenous enzyme immunoassay for the qualitative analysis of carisoprodol metabolite, Meprobamate, at a cutoff of 280 ng/mL in human urine. The assay is intended for use in laboratories with automated clinical chemistry analyzers. This in vitro diagnostic device is for prescription use only.

The Immunalysis Carisoprodol Metabolite / Meprobamate Urine HEIA provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. Gas Chromatography/ Mass Spectrometry (GC-MS) or Liquid Chromatography-Tandem Mass Spectrometry (LC-MS/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used.

AttributePredicate DeviceLin-Zhi Carisoprodol Metabolite(Meprobamate) EnzymeImmunoassay [DEN170010]Candidate DeviceImmunalysis CarisoprodolMetabolite / Meprobamate UrineHEIA
Similarities
Intended UseFor the qualitative and semi-quantitative determination of thepresence of carisoprodol metabolite(meprobamate) in human urine withautomated clinical chemistryanalyzers.For the qualitative analysis ofcarisoprodol metabolite,meprobamate, in human urine withautomated clinical chemistryanalyzers.
Test PrincipleHomogeneous EnzymeImmunoassaySame
User EnvironmentFor use in laboratoriesSame
Sample MatrixHuman UrineSame
Mass SpectrometryConfirmationRequired for preliminary positiveanalytical resultsSame
Reagent Storage2-8°C until expiration dateSame
Assay Materials2 analytical reagents:Antibody/Substrate Reagent andEnzyme Labeled ConjugateSame
Differences
Assay Cutoff Level100 ng/mL280 ng/mL

F. COMPARISON WITH PREDICATE

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the word "IMMUNALYSIS" in white letters on a red background. The word is written in all capital letters and is centered in the image. The background is a solid red color. The image is simple and easy to read.

AttributePredicate DeviceLin-Zhi Carisoprodol Metabolite(Meprobamate) EnzymeImmunoassay [DEN170010]Candidate DeviceImmunalysis CarisoprodolMetabolite / Meprobamate UrineHEIA
AntibodyMouse monoclonal anti-meprobamatePolyclonal sheep antibodies tocarisoprodol
CalibratorMeprobamateCarisoprodol

G. PERFORMANCE CHARACTERISTICS

The following laboratory performance studies were performed to determine substantial equivalence of the Immunalysis Carisoprodol Metabolite / Meprobamate Urine HEIA to the predicate device. Assay performance was established using the Olympus AU400e analyzer.

1. Precision/Cutoff Characterization - Meprobamate

Precision/Cutoff Characterization study for meprobamate was performed for ten days using three product lots with two runs per day in replicates of four on drug free urine (N=80) spiked with meprobamate to concentrations of ±25%, ±50%, ±75%, and ±100% of the cutoff. The spiked concentrations were confirmed by Liquid Chromatography-Tandem Mass Spectrometry (LC-MS/MS). The study verified that the cutoff serves as a boundary between a negative and positive interpretation of a qualitative result. Precision test results in qualitative mode are presented in Table 9 - 11.

Concentration (ng/mL)% of cutoff# of determinationsResult
0-100%8080 Negative
70-75%8080 Negative
140-50%8080 Negative
210-25%8080 Negative
280Cutoff8040 Neg/40 Pos
350+25%8080 Positive
420+50%8080 Positive
490+75%8080 Positive
560+100%8080 Positive

Table 1. Meprobamate Precision - Lot#1

Table 2. Meprobamate Precision - Lot#2
----------------------------------------
Concentration (ng/mL)% of cutoff# of determinationsResult
0-100%8080 Negative
70-75%8080 Negative
140-50%8080 Negative
210-25%8080 Negative
280Cutoff8038 Neg/42 Pos
350+25%8080 Positive
420+50%8080 Positive

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image contains the word "IMMUNALYS" in white letters on a red background. The word is written in all capital letters and is centered in the image. The background is a solid red color, and the letters are a bright white color. The image is simple and easy to read.

Concentration (ng/mL)% of cutoff# of determinationsResult
490+75%8080 Positive
560+100%8080 Positive

Table 3. Meprobamate Precision - Lot#3

Concentration (ng/mL)% of cutoff# of determinationsResult
0-100%8080 Negative
70-75%8080 Negative
140-50%8080 Negative
210-25%8080 Negative
280Cutoff8039 Neg/41 Pos
350+25%8080 Positive
420+50%8080 Positive
490+75%8080 Positive
560+100%8080 Positive

2. Specificity and Cross-Reactivity

Structurally and functionally similar compounds to meprobamate were spiked into drug free urine at levels that will yield a result that is equivalent to the cutoff. The study verified assay performance relative to the ability of the device to exclusively determine certain drugs in the qualitative mode. Cross-reactivity test results in qualitative mode are presented in Table 4.

Table 4. Cross-Reactivity - Qualitative
---------------------------------------------
CompoundCompoundConc.(ng/mL)ResultCross-Reactivity (%)
Meprobamate280POS100%
Carisoprodol100POS280%
Buprenorphine100,000NEG<0.1%
Codeine100,000NEG<0.1%
Darunavir200,000NEGN/D
Dihydrocodeine100,000NEG<0.1%
Efavirenz200,000NEGN/D
Felbamate120,000POS0.2
Hydrocodone100,000NEG<0.1%
Hydromorphone100,000NEG<0.1%
Meperidine100,000NEG<0.1%
Methocarbamol200,000NEGN/D
Mitomycin C200,000NEGN/D
Morphine100,000NEG<0.1%
Morphine-3-glucuronide100,000NEG<0.1%
Morphine-6-glucuronide100,000NEG<0.1%

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the word "IMMUNALYSIS" in white letters on a red background. The word is centered in the image and is in all capital letters. The red background is in the shape of a rectangle with an arrow on each end.

CompoundCompoundConc.(ng/mL)ResultCross-Reactivity (%)
Naloxone100,000NEG<0.1%
Naltrexone100,000NEG<0.1%
Neostigmine200,000NEGN/D
Norbuprenorphine100,000NEG<0.1%
Norcodeine100,000NEG<0.1%
Normorphine100,000NEG<0.1%
Oxycodone100,000NEG<0.1%
Oxymorphone100,000NEG<0.1%
Propoxyphene100,000NEG<0.1%
Retigabine200,000NEGN/D
Ritonavir200,000NEGN/D
Rivastigmine200,000NEGN/D
Tramadol100,000NEG<0.1%
Trazadone100,000NEG<0.1%
Venlafaxine100,000NEG<0.1%
Zafirlukast200,000NEGN/D

3. Interference – Meprobamate – Structurally Unrelated Compounds

Structurally unrelated compounds were evaluated in qualitative mode by spiking the potential interfering compound into drug-free negative urine containing a meprobamate concentration equivalent to ±25% of the assay cutoff. All potential interferents analyzed verified that assay performance is unaffected by externally ingested compounds. The levels of structurally unrelated compounds that did not interfere in the assay are presented in Table 12.

Table 5. Non-Interfering Structurally Unrelated Compounds to Meprobamate

CompoundConc. Tested (ng/mL)
4-Bromo-2,5,Dimethoxyphenethylamine100,000
Acetaminophen500,000
Acetylsalicylic Acid500,000
6-Acetylcodeine100,000
Alphenal100,000
6-Acetylmorphine100,000
Alprazolam100,000
7-Aminoclonazepam50,000
7-Aminoflunitrazepam100,000
7-Aminonitrazepam100,000
Amitriptyline100,000
Amobarbital100,000
S-(+) Amphetamine100,000
Aprobarbital100,000
CompoundConc. Tested (ng/mL)
Barbital100,000
Benzoylecgonine100,000
Benzylpiperazine100,000
Bromazepam100,000
Bupropion100,000
Butabarbital100,000
Butalbital100,000
Caffeine500,000
Cannabidiol100,000
Cannabinol100,000
Carbamazepine100,000
Chlordiazepoxide100,000
Chlorpromazine100,000
cis-Tramadol100,000
Clobazam100,000
Clomipramine100,000
Clonazepam100,000
Clozapine100,000
Cocaine100,000
Cotinine100,000
Cyclobenzaprine100,000
Cyclopentobarbital100,000
Demoxepam100,000
Desakylflurazepam100,000
Desipramine100,000
Dextromethorphan100,000
Diazepam100,000
Digoxin100,000
Diphenhydramine500,000
Dehydronorketamine50,000
Delta-9-THC100,000
Doxepin100,000
Doxylamine100,000
Ecgonine100,000
Ecgonine methyl ester100,000
EDDP100,000
EMDP100,000
1R,2S(-)-Ephedrine100,000
1S,2R(+)-Ephedrine100,000
Ethyl glucuronide100,000
CompoundConc. Tested (ng/mL)
Ethylmorphine100,000
Fenfluramine100,000
Fentanyl100,000
Flunitrazepam100,000
Fluoxetine100,000
Flurazepam100,000
Haloperidol100,000
Heroin100,000
Hexobarbital100,000
11-hydroxy-delta-9-THC100,000
Ibuprofen500,000
Imipramine100,000
Ketamine100,000
Labetalol100,000
Lamotrigine100,000
Levorphanol tartrate100,000
Lidocaine100,000
Lorazepam100,000
Lorazepam Glucuronide50,000
Lormetazepam100,000
LSD100,000
Maprotiline100,000
MDA100,000
MDEA100,000
MDMA100,000
S(+)-Methamphetamine100,000
Methadone500,000
Methaqualone100,000
Methoxetamine100,000
Methylone100,000
Methylphenidate100,000
Midazolam100,000
N-desmethyl tapentadol100,000
N-desmethyl venlafaxine100,000
Nalorphine100,000
Naproxen100,000
Nitrazepam100,000
11-nor-9 carboxy THC100,000
Nordiazepam100,000
Norketamine100,000
CompoundConc. Tested (ng/mL)
Norpropoxyphene100,000
Norpseudoephedrine100,000
Nortriptyline100,000
O-desmethyl tramadol100,000
O-desmethyl venlafaxine100,000
Olanzapine100,000
Oxazepam100,000
PCP100,000
Pentobarbital100,000
Pentazocine100,000
Phenazepam100,000
Phenobarbital100,000
Phentermine100,000
Phenylephedrine100,000
Phenytoin100,000
Phenylpropanolamine100,000
PMA100,000
Prazepam100,000
Propranolol100,000
Protriptyline100,000
R,R(-)-Pseudoephedrine100,000
S,S(+)-Pseudoephedrine100,000
Ritalinic Acid100,000
Salicylic Acid100,000
Secobarbital100,000
Sertraline100,000
Sufentanil Citrate50,000
Talbutal100,000
Tapentadol100,000
Temazepam100,000
Theophylline100,000
Thiopental100,000
Thioridazine100,000
Triazolam100,000
Trifluoromethylphenyl-piperazine100,000
Trimipramine100,000
Verapamil100,000
Zolpidem Tartrate100,000

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the word "IMMUNALYSIS" in white letters on a red background. The word is centered and in all capital letters. The background is a red arrow shape.

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the word "IMMUNALYSIS" in white letters on a red background. The word is centered in the image and is written in all capital letters. The red background is in the shape of a rectangle with an arrow on the left side.

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the word "IMMUNALYSIS" in white letters on a red background. The word is centered in the image and is written in all capital letters. The background is a solid red color, and the letters are a bright white color. The image is simple and easy to read.

{11}------------------------------------------------

4. Interference - Endogenous Compounds and Urine Preservatives

Endogenous compounds and urine preservatives were evaluated in qualitative mode by spiking the potential interferent into drug free urine containing meprobamate concentration equivalent to ±25% of the assay cutoff. Due to the interference of boric acid observed at ±25% of the cutoff, potential interference was also evaluated at ±50% of the cutoff. Other than boric acid, assay performance is unaffected by all the other internally existing physiological conditions or urine preservatives tested. Compounds tested that did not interfere in the assay are presented in Table 13. Boric acid interference test results in qualitative mode are presented in Table 14.

CompoundConc. Tested (ng/mL)
Acetone1.0 g/dL
Ascorbic Acid1.5 g/dL
Bilirubin0.002 g/dL
Creatinine0.5 g/dL
Ethanol1.0 g/dL
Galactose0.01 g/dL
y-Globulin0.5 g/dL
Glucose2.0 g/dL
Hemoglobulin0.300 g/dL
Human Serum Albumin0.5 g/dL
Oxalic Acid0.1 g/dL
Riboflavin0.0075 g/dL
Sodium Azide1% w/v
Sodium Chloride6.0 g/dL
Sodium Fluoride1% w/v
Urea6.0 g/dL

Table 6. Non-Interfering Endogenous Compounds and Urine Preservatives

Table 7. Boric Acid Interference

CompoundConcentrationTestedQualitative Result
-50% Cutoff(140 ng/mL)+50% Cutoff(420 ng/mL)
Boric Acid1% w/vNegativeNegative

5. Interference - pH

To evaluate potential interference from the effect of urine pH, device performance in the qualitative mode was tested at pH values 3.0, 7.0 and 11.0. All test samples were prepared in drug-free urine containing meprobamate at the concentration equivalent to ±25% of the carisoprodol assay cutoff. No positive or negative interference was observed at urine pH values 3.0, 7.0 and 11.0 for each test mode.

6. Interference - Specific Gravity

To evaluate potential interference from the specific gravity of urine, device performance in the

{12}------------------------------------------------

Image /page/12/Picture/0 description: The image shows the word "IMMUNALYSIS" in white letters on a red background. The word is written in all capital letters and is centered in the image. The background is a solid red color. The image is simple and straightforward, with the focus being on the word itself.

qualitative mode was tested at physiologically relevant urine specific gravity values 1.000, 1.015 and 1.030. All test samples were prepared in drug free urine containing meprobamate at the concentration equivalent to ±25% of the carisoprodol assay cutoff. No positive or negative interference was observed at urine specific gravity values 1.000, 1.015 and 1.030 for each test mode.

7. Calibration Duration

Drug free negative urine spiked with carisoprodol at ±25% of the cutoff were tested in qualitative mode at time points up to 14 days. At the initial time point, a two-point calibration curve was established. This calibration was used through the duration of this study. The test results met acceptance criteria at each time point. The recommended frequency of calibration is 14 days.

8. Specimen Stability (Urine)

De-identified, unaltered clinical urine samples containing carisoprodol and/or meprobamate obtained from clinical testing were tested by LC-MS/MS at each time point at 22°C - 30°C and at 2°C - 8°C. Test results indicated that urine samples containing carisoprodol and/or meprobamate are stable for up to 7 days stored at ambient temperature up to 30°C and up to 6 months stored at 2°C - 8°C.

9. Urine Elimination Study

The ten subjects were self-reported single dose users of either 250 mg or 350 mg of carisoprodol and have taken the single dose of carisoprodol within 48 hours of enrollment and did not take carisoprodol for 96 hours after enrollment. The subjects were instructed to collect his/her first urine at the time of enrollment and every 12-24 hours from time of enrollment. The collected samples were screened with Immunalysis Carisoprodol Metabolite / Meprobamate Urine HEIA in qualitative mode in singlicate. Carisoprodol and Meprobamate mass spectrometry analysis was performed for all samples. The earliest time post ingestion of drug that the carisoprodol or meprobamate in the urine specimen decreased below cutoff level is 45.5 hours.

10. Method Comparison

A total of one hundred and sixty seven (167) de-identified, unaltered leftover clinical urine samples obtained from clinical testing laboratories were analyzed by the Immunalysis Carisoprodol Metabolite / Meprobamate Urine HEIA and by LC-MS/MS. The cutoff of 280 ng/mL of meprobamate was used to distinguish positive results from negative results for LC-MS/MS. Fifty-five (55) specimens showed negative results by both methods, and one hundred and seven (107) specimens showed positive results by both methods. Results are shown in Table 8. The five false positive samples (indicated by an asterisk) each contained carisoprodol and are not a cause for clinical concern (see Table 9). The instruments used were an Olympus AU 400e and an Agilent 6430 Liquid Chromatography-Tandem Mass Spectrometry.

{13}------------------------------------------------

Image /page/13/Picture/0 description: The image shows the word "IMMUNALYSIS" in white letters on a red background. The word is in all capital letters and is bolded. The background is a solid red color. The image is simple and straightforward, with the focus on the word "IMMUNALYSIS".

Table 8. Method Comparison

LC-MS/MS Total Meprobamate Concentration
ImmunalysisMeprobamateUrine HEIA Result< 140 ng/mL(less than-50% cutoff)140-279 ng/mL(between -50%cutoff andcutoff)280-420 ng/mL(between cutoffand +50%cutoff)> 420 ng/mL(greater than+50% cutoff)Agreement (%)
Positive323104100% (107/107)
Negative5140092% (55/60)

Table 9 Discrepant Samples

Qualitative ResultMeprobamate(ng/mL)Carisoprodol(ng/mL)
POS0147
POS0148
POS0750
POS20135,156
POS230120,593

H. CONCLUSION

The information provided in this pre-market notification demonstrates that the Immunalysis Carisoprodol Metabolite / Meprobamate Urine HEIA is substantially equivalent to the legally marketed predicate device for its intended use.

§ 862.3590 Meprobamate test system.

(a)
Identification. A meprobamate test system is a device intended to measure meprobamate in human specimens. Measurements obtained by this device are used to detect the presence of meprobamate to diagnose the use or overdose of meprobamate or structurally-related drug compounds (e.g., prodrugs).(b)
Classification. Class II (special controls). The special controls for this device are:(1) Design verification and validation must include:
(i) Robust data demonstrating the accuracy of the device when used in the intended specimen matrix. The accuracy data must include a comparison between the meprobamate test system results and meprobamate results that are measured on an FDA-accepted measurement method that is specific and accurate (e.g., gas or liquid chromatography combined with tandem mass spectrometry).
(ii) Robust analytical data demonstrating the performance characteristics of the device, including, but not limited to, specificity, cross-reactivity to relevant endogenous and exogenous substances, and the reproducibility of analyte detection around the cutoff(s).
(2) The intended use of the device must not include an indication for use in monitoring therapeutic drug concentrations or informing dosing adjustment decisions.
(3) Your 21 CFR 809.10 labeling must include the following:
(i) If indicated for use as a screening test to identify preliminary results for further confirmation, the intended use must state “This assay provides only a preliminary analytical result. A more specific alternative chemical confirmatory method (e.g., gas or liquid chromatography and mass spectrometry) must be used to obtain a confirmed analytical result. Clinical consideration and professional judgment must be exercised with any drug of abuse test, particularly when the preliminary test result is positive.”
(ii) A limiting statement that reads as follows: “This test should not be used to monitor therapeutic drug concentrations or to inform dosing adjustment decisions.”